Professional Documents
Culture Documents
PRACTICAL 1 & 2
Group members:
Lecture: MR LW BENEKE
DEFLECTION OF A BEAM
Object of Experiment
Apparatus
PRACTICAL 1. 1
Procedure
Let L = 600 mm and W =? N (see table) on each end. Place a stirrup on the
beam between the beam and the flat end of the dial gauge stem. Position the dial
gauge in the center of the beam. Use the position indicator to set the position of
all the stirrups. With no load on the hangers set the dial gauge bezel to zero.
Apply a load to each hanger and note the dial gauge reading. Repeat for load,
see table, remembering to ensure that the load on each hanger is identical.
W W
SUPPORT SUPPORT
3
Results
STEEL
W (N) 0 1N 2N 3.5N 5N
δ(mm) 0 0.02 0.045 0.08 0.11
BRASS
W (N) 0 1N 2N 3.5N 5N
δ(mm) 0 0.04 0.075 0.12 0.17
ALUMINUM
W (N) 0 1N 2N 3.5N 5N
δ(mm) 0 0.05 0.1 0.16 0.235
W (N) 3 Steel
Brass
2 Aluminum
0
0 0.05 0.1 0.15 0.2 0.25
d (mm)
4
The values agree with the theory behind because of the Young’s Modulus of
each material.
PRACTICAL 1.2
DEFLECTION OF A CANTILEVER
Procedure
Let L = 600 mm and W =? N (see table) on each end. Place a stirrup on the
beam between the beam and the flat end of the dial gauge stem. Position the dial
gauge in the center of the beam. Use the position indicator to set the position of
all the stirrups. With no load on the hangers set the dial gauge bezel to zero.
Apply a load to each hanger and note the dial gauge reading. Repeat for load,
see table, remembering to ensure that the load on each hanger is identical.
5
Results
STEEL
W (N) 0 1N 2N 3.5N 5N
BRASS
W (N) 0 1N 2N 3.5N 5N
WALL
L
6
δ(mm) 0 0.46 0.78 1.36 1.97
ALUMINUM
W (N) 0 1N 2N 3.5N 5N
δ(mm) 0 0.49 1.08 1.89 2.76
Chart Title
6
4
W(N)
3 STEEL
BRASS
2 ALUMINIUM
0
0 0.5 1 1.5 2 2.5 3
d(mm)
Slope of graph
PRACTICAL 1.3
Procedure
Let L = 600 mm, a = 200 mm and W =? N (see table) on each end. Place a
stirrup on the beam between the beam and the flat end of the dial gauge stem.
Position the dial gauge in the center of the beam. Use the position indicator to
set the position of all the stirrups. With no load on the hangers set the dial gauge
bezel to zero. Apply a load to each hanger and note the dial gauge reading.
Repeat for load, see table, remembering to ensure that the load on each hanger
is identical.
WALL
W W
L
Results
STEEL
W (N) 0 1N 2N 3.5N 5N
8
δ(mm) 0 0.015 0.03 0.05 0.08
BRASS
W (N) 0 1N 2N 3.5N 5N
δ(mm) 0 0.03 0.05 0.09 0.11
ALUMINUM
W (N) 0 1N 2N 3.5N 5N
δ(mm) 0 0.05 0.08 0.14 0.18
Slope of graph
= 61.54 = 50 = 30.77
The values agree with the theory behind because of the Young’s Modulus of
each material.
Discussion
Conclusion
Reference
Aims:
Apparatus:
Procedure:
Note-Note that this experiment only tests the gauges nearest to the center of the
cylinder; this is because these gauges register the larges strain (as suggested in
the theory and proven in the second experiment) and will produce more
meaningful results.
Pressure Strain
11
−2
(𝑀𝑁. 𝑚 ) (10−6)
Gauge Gauge Gauge Gauge Gauge Gauge Gauge Gauge Gauge Gauge Gauge
1 2 3 4 5 6 7 8 9 10 11
0 0 -1 0 -2 0 0 0 -1 0 1 0
1 9 -9 5 -7 3 -4 3 -4 2 0 20
2 17 -19 10 -12 7 -7 5 -4 3 -1 41
3 28 -28 17 -17 12 -10 7 -6 5 -2 63
4 37 -40 22 -22 15 -13 10 -8 8 -5 83
5 45 -51 27 -27 19 -16 11 -10 9 -6 107
6 56 -60 34 -32 24 -18 17 -12 13 -7 131
7 65 -71 40 -37 28 -23 18 -13 14 -8 154
Table 1
1. Switch on the power to the thick Cylinder and leave it for at least thirty
minutes before you do the experiment. This allows the strain gauges to
reach a stable temperature to give you accurate readings.
2. Unscrew the pump hand wheel until the pressure gauge shows zero
pressure.
3. Use the ‘Press & hold to zero’ button to zero the strain gauge display
readings. All the strain gauge readings should now read 0 µɛ (+⁄− 5 µɛ),
and the pressure meter should read 0 𝑀𝑁. 𝑚−2 (+⁄− 0.05 𝑀𝑁. 𝑚−2 ).
Note-Use the ‘Scroll Readings’ button to scroll through the strain gauge
reading from 1 to 13.
The strain gauge readings may vary slightly (+⁄− 2 µɛ) as you take the
readings. This is normal.
7. Undo (turn anticlockwise) the pump hand wheel to reduce the indicated
pressure back to 0 𝑀𝑁. 𝑚−2.
12
Results Analysis:
200
150 Gauge 1
Gauge 2
Gauge 3
100
Gauge 4
Gauge 5
Strain 50
(x10^-6) Gauge 6
Gauge 7
0
Gauge 8
0 1 2 3 4 5 6 7 8
Gauge 9
-50 Gauge 10
Gauge 11
-100
Pressure (MN.m^-2)
Conclusion
The objective of the experiment was achieved; according to the reading obtained
from the experiment all the gauges have a linear property against increasing
pressure. Observing from the graph we can say that the hoop strain has a
positive gradient and the radial strain has a negative gradient.
13
Aims:
1. To find the hoop and radial strains in the cylinder and compare them with
theoretical values.
2. To prove that the longitudinal strain in this Thick Cylinder is very small
compared with the other strain values.
Procedure:
Note – Note that this experiment only tests the gauges nearest to the center of
the cylinder; this is because these gauges register the largest strain (as
suggested in the theory and proven in the second experiment) and will produce
more meaningful results.
1. Switch on the power to the thick Cylinder and leave it for at least thirty
minutes before you do the experiment. This allows the strain gauges to
reach a stable temperature to give you accurate readings.
2. Unscrew the pump hand wheel until the pressure gauge shows zero
pressure.
3. Use the ‘Press & hold to zero’ button to zero the strain gauge display
readings. All the strain gauge readings should now read0 µɛ (+⁄− 5 µɛ),
and the pressure meter should read0 𝑀𝑁. 𝑚−2 (+⁄− 0.05 𝑀𝑁. 𝑚−2 ).
Note-Use the ‘Scroll Readings’ button to scroll through the strain gauge
reading from 1 to 13.
The strain gauge readings may vary slightly (+⁄− 2 µɛ) as you take the
readings. This is normal.
7. Undo (turn anticlockwise) the pump hand wheel to reduce the indicated
pressure back to 0 𝑀𝑁. 𝑚−2.
Theory:
15
16
17
Calculations
Sample calculations at 36 mm
𝑝 𝑅2
𝜕𝑅 = 2 (1 − 2 )
(𝐾 − 1) 𝑟
7 × 106 (0.075)2
= (1 − ) = −1.515 𝑀𝑃𝑎
((4.054)2 − 1) (0.036)2
𝑝 𝑅2
𝜕𝐻 = (1 + )
(𝐾 2 − 1) 𝑟2
7 × 106 (0.075)2
= (1 + ) = 2.42 𝑀𝑃𝑎
((4.054)2 − 1) (0.036)2
𝜕𝑅 𝑣. 𝜕𝐻
𝜀𝑅 = −
𝐸 𝐸
200 × 109
= (0.3 × 14.37 × 10−6 − (−11.205 × 10−6 )) = −3.897 𝑀𝑃𝑎
1 − 0.32
𝐸
𝜕𝑅 = (𝑣𝜀𝐻 − 𝜀𝑅 )
1 − 𝑣2
200 × 109
(0.3 × (−11.205 × 10−6 ) − 14.37 × 10−6) = 3.41 𝑀𝑃𝑎
1 − 0.32
𝑝. 𝐾 2
𝜏=
(𝐾 2 − 1)
7 × 106 × 4.0542
= = 7.454 𝑀𝑃𝑎
(4.0542 − 1)
40−14.37
Error difference= × 100 = 64.08 %
40
37 − 11.205
× 100 = 69.7 %
37
19
Gauge 1 2 3 4 5 6 7 8 9 10 11 12 13
Number
Radius at 28 36 45 56 63 18. 75
Gauge 5
(mm)
Type of Ԑ𝐻 Ԑ𝑅 Ԑ𝐻 Ԑ𝑅 Ԑ𝐻 Ԑ𝑅 Ԑ𝐻 Ԑ𝑅 Ԑ𝐻 Ԑ𝑅 Ԑ𝐻 Ԑ𝐿 Ԑ𝐻
strain
Measured 65 -71 40 -37 28 -23 18 -13 14 -8 154 0 9
Strain
(x10^-6)
Calculated 22. - 14.3 - 9.7 -6.6 6.87 -3.7 5.7 - 50 - 4.53
Strain 7 19.5 7 11.20 7 5 6 2.5 4.72 5
(x10^-6) 5 5 9 5
Error 65. 72.5 64.1 69.7 65. 70. 61.8 71. 58. 67. 67. 100 49.6
Difference 1 1 1 5 9 6 5
(%)
Cylinder 75 mm Cylinder Internal radius: 18.5 mm
outside
radius:
Cylinder 7MPa
Pressure
Graph of Hoop & Radial strain Vs diameter for measured & calculated readings
20
200
150
100
Hoop strain
Strain (x10^-6) 50 (measured)
Radial
0 strain
0 20 40 60 80 (measured)
Hoop strain
-50 (calculated)
-100
d (mm)
Graph of hoop & radial stress vs diameter for the theoretical & derived readings
21
10
4
Theoretical
2 hoop stress
Stress (x10^6)
0 Theoretical
0 20 40 60 80 radial stress
-2
Derived hoop
-4 stress
Derived radial
-6
stress
-8
d (mm)
Discussion
Conclusion
The objective of the experiment was achieved even though obtained values
slightly differ to the calculated one’s, this difference in results was because of
factors discussed above. Overall the shape of the experimental graph had a
similar shape as the calculated graph; lastly longitudinal strain came out to be the
smallest strain obtained.
Reference