Professional Documents
Culture Documents
X-Ray Imaging
X-Ray Imaging
J.Jeyasugiththan PhD
Department of Nuclear Science
University of Colombo
X-ray production
ü Radiation Dose
• Should be kept as low as possible.
1
6/18/21
Projection imaging
X-ray image (Projection radiography) produces 2-D projections of a
3-D object on film. X-ray shadow inside the body onto a photographic
film or digital detector: film
body
Dark
X-Ray
source Light
bone
X-rays are travelling through soft tissues in body with little attenuation and are
only “stopped” by high density tissues such as bone.
3
Projection imaging
2
6/18/21
Projection imaging
X-ray machine
(source) X-rays absorbed by
dense parts of
patients body
Need partial
attenuation Photographic plate or
digital detector
Film processing
White indicates
dense parts of body
X-ray Spectrums
20 to 150 kV for
diagnostic imaging
3
6/18/21
X-ray Spectrums
X-ray Spectrums
4
6/18/21
X-ray Spectrums
9
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
X-ray Spectrums
(a) generation of
low energy
electromagnetic
radiations (infra red,
visible, ultraviolet and
very soft X-rays). All
these are rapidly
converted into heat.
10
Physics for Diagnostic Radiology
5
6/18/21
Bremsstrahlung Spectrum
Greater probability of
low-energy x-ray photon
production which is
linearly decreased up to
maximum energy
No x-rays below
about 10 keV.
The average energy of the spectrum, which is typically 1/3 to 1/2 the
maximal energy .
11
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
Bremsstrahlung Spectrum
Bremsstrahlung radiation is a continues spectrum corresponding
to electrons is loosing different fraction of their initial kinetic
energy.
6
6/18/21
X-ray Spectrums
13
X-ray Spectrums
Tungsten target
14
7
6/18/21
Characteristic X-rays
15
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
Characteristic X-rays
16
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
8
6/18/21
Characteristic X-rays
17
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
X-rays –tube
The production of both Bremsstrahlung and Characteristic
Radiation requires energetic electrons hitting a target
Principle components
of an X ray tube are
an Electron Source
from a heated
tungsten filament with
a focusing cup serving
as the tube Cathode,
an Anode or Target
and a Tube Envelope
to maintain an interior
vacuum
18
Review of Radiology Physics: A Handbook for Teachers and Students/ The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
9
6/18/21
Figure 2.14 (see later) also shows, on the extreme left, an autotransformer. An autotrans-
X-rays –tube
former comprises one winding only and works on the principle of self-induction. Since the
primary and secondary circuits are in contact, it cannot transform high voltages or step up
from low to high voltages. However, it does give a variable secondary output on the low
voltage side of the transformer and hence controls kV directly.
AC
voltage Anode
supply Bridge
X-ray tube
of four
rectifiers Cathode
FIGURE 2.11
Essential features of a full wave rectified supply. (Solid) and (dotted) arrows show that irrespective of whether
A or B is at aThe alternating
positive potential
potential, the must be
current always rectified
flows through before
the X-rayittube
is applied to direction.
in the same
an X-ray set
20
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
10
increased by a factor of 100, say from 50 Hz to 5 kHz, nA may be reduced by a
similar amount. Since the efficiency has been improved the transformer is much
smaller, perhaps one-tenth the size of a three phase 12 peak generator.
4. Rapid response—The high voltage is switched on and off, and its level may be regu-
6/18/21
lated even during exposure, under feed-back control of the inverter. The rise time
of the tube voltage can be less than 200 μs.
5. Long-term stability—The tube current is more stable at the higher frequency f 2 and
is independent of the voltage.
6. Timer precision—The precision of the exposure timer can be improved.
7. Voltage range—The generator Frequency
may be usedConverter
across the full kV range from mammog-
raphy to CT.
Stage 1 Stage 2 Stage 3
Rectifer Inverting rectifier
u1 f1 u0 u2 f2
Converter
+ +
Bridge V X-ray
Capacitor
circuit volts tube
– –
Flow of electrons
+ +
Bridge
Zero volts Capacitor
circuit
– –
Flow of electrons
FIGURE 2.13 22
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
Illustration of the use of a capacitor for voltage smoothing. For explanation see text.
TABLE 2.2
Nominal Values of Voltage Ripple for Different Generators
Voltage Wave Form Ripple (%)
Half wave/full wave 100
Full wave + capacitor smoothing 20 11
Three phase – 6 pulse unsmoothed 15
Three phase – 12 pulse unsmoothed 6
High frequency generatora (typical) 1–4
High frequency generatora (attainable but expensive) 0.1
6/18/21
Exposure switch
V
Mains in AC Earthed
414 V
KV
AF
selector
Autotransformer Transformer mA control
Timer
Filament
supply volts
FIGURE 2.14
Simplified representation of the position of kV and current meters in the electrical circuit.
23
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
AF measures the filament supply current (IF) which may be adjusted to give the required
thermionic emission before exposure starts. The actual tube current flowing during expo-
sure (Ic) is measured by ammeter Ac.
The envelope also provides a vacuum seal to the metallic components that protrude 12
through it. Great care must be taken at the manufacturing stage to achieve a very high
level of vacuum before the tube is finally sealed. The electrons have a mean free path of
several metres. If residual gas molecules are bombarded by electrons, the electrons may
6/18/21
X-rays –Cathode
The numbers of
electrons are adjusted
by the filament
current and filament
temperature.
Small changes in
the filament current
can produce relatively
large changes in tube
current .
25
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
X-rays –Cathode
Above 40 kV, the filament current limits the tube current, below 40 KV,
further increases in filament current do not increase the tube
current
26
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
13
6/18/21
X-rays –Cathode
For high anode voltages all electrons boiled off the filament
are accelerated to the anode giving an anode current fairly
independent of tube voltage (saturation current)
27
Review of Radiology Physics: A Handbook for Teachers and Students/
X-rays –Cathode
Unbiased
28
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
14
6/18/21
X-rays –Cathode
Biased
Focusing cup totally insulated from
the filament wires (its voltage is
independent of the filament).
29
Review of Radiology Physics: A Handbook for Teachers and Students/ The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
X-rays –Cathode
Grid-biased
A greater negative applied voltage (about -
4,000 V) to the focusing cup actually stops
the flow of electrons, providing a means to
rapidly switch the x-ray beam on and off
30
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
15
6/18/21
X-rays –Anode
Choice of Material
For common radiographic applications a high Bremsstrahlung
yield is mandatory requiring materials with high atomic
numbers (Z)
31
Review of Radiology Physics: A Handbook for Teachers and Students/
X-rays –Anode
Choice of Material
For Mammography other anode materials such as
molybdenum (Z=42) and rhodium (Z=45) are frequently used
32
Review of Radiology Physics: A Handbook for Teachers and Students/
16
6/18/21
X-rays –Anode
Anode Configuration: Stationary
33
Review of Radiology Physics: A Handbook for Teachers and Students/ The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
X-rays –Anode
34
Review of Radiology Physics: A Handbook for Teachers and Students/ The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
17
6/18/21
X-rays –Anode
35
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
X-rays –Anode
36
Review of Radiology Physics: A Handbook for Teachers and Students/ The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
18
6/18/21
X-rays –Anode
Focal spot
For measurement purposes the Focal Spot Size is defined along the
central beam projection
37
Review of Radiology Physics: A Handbook for Teachers and Students/ The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
X-rays –Anode
Focal spot
38
Review of Radiology Physics: A Handbook for Teachers and Students/ The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
19
6/18/21
X-rays –Anode
Large Focal spot
39
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
X-rays –Anode
Anode angle
40
Review of Radiology Physics: A Handbook for Teachers and Students/ The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
20
6/18/21
X-rays –Anode
Anode angle
41
The reduction in
the x-ray beam
intensity toward the
anode side of the x-
ray field is called
heel effect.
It caused by the
greater attenuation
of x-rays in the
anode side of the
field by the anode
itself .
42
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
21
6/18/21
Since the x-ray beam intensity is greater on the cathode side of the
field, the orientation of the x-ray tube cathode over thicker parts of
the patient can result in a better balance of x-ray photons transmitted
through the patient and onto the image receptor.
43
Review of Radiology Physics: A Handbook for Teachers and Students
X-rays
X-ray Tube Insert
It contains the cathode, anode, rotor assembly, and support
structures sealed in a glass or metal enclosure under a high
vacuum.
44
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed./ https://www.worthpoint.com
22
6/18/21
The x-ray tube housing supports, insulates, and protects the x-ray tube
insert from the environment. Between the x-ray tube insert and housing
is oil that provides heat conduction and electrical insulation.
45
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed./ https://www.worthpoint.com
46
The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed./ https://www.worthpoint.com
23
6/18/21
47
Review of Radiology Physics: A Handbook for Teachers and Students/ The essential physics of medical imaging / Jerrold T. Bushberg. — 3rd ed.
X-rays –Collimators
24
6/18/21
Filtering
Inherent filtering
25
6/18/21
Inherent filtering
§ Anode
§ Tube envelope exit port (glass or metal)
§ Insulating oil
§ Window of the tube housing
Inherent filtering
26
6/18/21
Added filtering
Added filtering
This disks remove the x-rays that had enough energy to get
through the inherent filtration but are still not energetic to
contribute to image formation.
27
6/18/21
Added filtering
Beam filtering
Lower energy photons contribute to radiation dose but not the image quality and can
filter with thin (1-2mm) plate of material (Al or Cu).
56
28
6/18/21
Added filtering
X-ray tube
29
6/18/21
59
600 mA
300 mA
30
10 20 30 40 50 60 70 80 90 100
Energy (keV)
Increased kVp
6/18/21
Number of photons
Factors affecting X-ray emission 100 kVp
50 kVp
The tube current (mA) - equal to the number of electrons flowing from the
cathode10 to20the 30anode
40 per
50 unit
60 time.
70 80 The
90 exposure
100 of the beam for a given kVp
Energy (keV)
and filtration is proportional to the tube current.
FIGURE 2.6 Effect of increasing tube voltage on X-ray output.
is an increase in the
quantity of the electrons as
well as the X-ray photons
produced. The X-ray
600 mA
300 mA quality remains essentially
the same, i.e. the maximum
photon energy and the
10 20 30 40 50 60 70 80 90 100
Energy (keV) effective photon energy
FIGURE 2.7 Effect of increasing tube current on X-ray output.
remains unchanged.
61
62
31
6/18/21
63
Radiographic Grid
32
6/18/21
Radiographic Grid
Radiographic Grid
Image contrast:
Difference in optical
density between
adjacent structures.
33