USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000)

PRODUCT : ELECTROMAGNETIC FLOWMETER


AP1013 – AP1014 – AP1017 – AP1018 – AP1016/DIN – AP1116

: CONVERTER AP501 – AP501/B

TECHNICAL NOTE : NT593E


VERSION : 1.0

Revision 0 date 1 date 2 date


Prepared 23/03/2000 15/05/2000 28/08/2000
Approved 23/03/2000 15/05/2000 28/08/2000
IN ACCORDANCE WITH THE EUROPEAN STANDARD EN ISO 9001
Automation Progetti
TITLE Page 2 / 54

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000)

INDEX
1. INTRODUCTION 5
1.1 GENERAL INFORMATION................................................................................................................................................. 5
1.2 - CHARACTERISTICS.......................................................................................................................................................... 5
1.3 - SPECIFICATIONS .............................................................................................................................................................. 7
1.4 - DESCRIPTION .................................................................................................................................................................... 9
1.5 - CONNECTION CABLE .................................................................................................................................................... 10

MODULAR CONVERTER FOR ELECTROMAGNETIC FLOWMETER 11

OPTIONAL AP501 / B o HI-TECH 12

ELECTROMAGNETIC FLOWMETER 13

ELECTROMAGNETIC FLOWMETER 14

ELECTROMAGNETIC FLOWMETER 15

ELECTROMAGNETIC FLOWMETER 16

2 - PRINCIPLE OF OPERATING 17

3 - INSTALLATION 18
3.1 - PRECAUTION FOR THE CHOICE OF THE PLACE ..................................................................................................... 18
3.2 - INSTALLATION OF THE FIELD VERSION CONVERTERS ...................................................................................... 19
3.3 - DIMENSIONE AND TEMPLATE OF THE RACK VERSIONS .................................................................................... 20
3.4 - INSTALLATION OF THE CABLE BETWEEN FLOWMETER AND CONVERTER.................................................. 21
3.5 - INSTALLATION CONDITIONS...................................................................................................................................... 22
3.6 - HOW TO CONNECT CABLES ........................................................................................................................................ 24

4 - AP1116 MOUNTING 25

5 - GROUNDINGS OF THE FLOWMETERS AP1013-AP1014-AP1017-AP1018 26

6 - GROUNDINGS OF THE FLOWMETER AP1116 27

7 - START UP 28
7.1 - STARTING CHECKS........................................................................................................................................................ 28
7.2 - MEASUREMENT STARTING ......................................................................................................................................... 28
7.2.1 - CHECK OF ZERO POINT ............................................................................................................................................. 28
7.2.2 - CHECK OF FLOW DIRECTION................................................................................................................................... 28
7.3 - MEASUREMENT.............................................................................................................................................................. 28

8 - PROGRAMMING OF THE CONVERTER AP501 – AP501/B WITHOUT OPTIONAL D (see


optionals) 28

9. - PROGRAMMING OF THE CONVERTER AP501 – AP501/B WITH OPTIONAL D (see


optionals) 29
Automation Progetti
TITLE Page 3

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.1 -ACCESS KEYWORD INSERTION................................................................................................................................... 29
9.2 -COUPLING OF CONVERTER TO ND OF PIPING......................................................................................................... 29
9.3 -PROGRAMMING OF THE TECHNICAL UNITS OF THE INSTANTANEOUS FLOW .............................................. 29
9.4 -PROGRAMMING OF THE TECHNICAL UNITS OF TOTALIZERS ............................................................................ 30
9.5 -PROGRAMMING OF THE FIRST LINE OF LCD........................................................................................................... 31
9.6 -PROGRAMMING OF THE SECOND LINE OF LCD...................................................................................................... 31
9.7 -PROGRAMMING OF THE DIGITAL OUTPUT 1 ........................................................................................................... 31
9.8 -PROGRAMMING OF THE DIGITAL OUTPUTS 2 , 3 , 4 .............................................................................................. 31
9.9 -PROGRAMMING OF THE LENGHT OF THE TOTALIZATION PULSES OF THE DIGITAL OUTPUT 1............... 32
9.10 -PROGRAMMING OF THE LENGHT OF THE TOTALIZATION PULSES OF THE DIGITAL OUTPUTS 2, 3, 4... 32
9.11 -PROGRAMMING OF THE KIND OF ANALOG OUTPUT .......................................................................................... 32
9.12 -PROGRAMMING OF THE FULL SCALE OF THE ANALOG OUTPUT.................................................................... 33
9.13 -PROGRAMMING OF THE FILTER TIME..................................................................................................................... 33
9.14 -PROGRAMMING OF THE MINIMUM SET.................................................................................................................. 33
9.15 -PROGRAMMING OF THE MAXIMUM SET ................................................................................................................ 33
9.16 -PROGRAMMING OF PRESETTER................................................................................................................................ 34
9.17 -PROGRAMMING OF THE KIND OF RESET IN PRESETTER.................................................................................... 34
9.18- MANUAL RESET OF PRESETTER ............................................................................................................................... 34
9.19 -PROGRAMMING OF THE ACCESS KEYWORD ........................................................................................................ 34
9.20 -PROGRAMMING OF THE SPECIFIC WEIGHT ........................................................................................................... 35
9.21 -PROGRAMMING OF TIME............................................................................................................................................ 35
9.22 -PROGRAMMING OF DATE ........................................................................................................................................... 35
9.23 -PROGRAMMING OF CUT-OFF ..................................................................................................................................... 36
9.24 -PROGRAMMING OF THE SERIAL ADDRESS............................................................................................................ 36
9.25 -PROGRAMMING OF THE SERIAL BAUD-RATE....................................................................................................... 36
9.26 -ZERO ADJUSTMENT ..................................................................................................................................................... 37
9.27 -PROGRAMMING OF LANGUAGE ............................................................................................................................... 38
9.28 -CONFIGURATION OF THE DIGITAL INPUT ............................................................................................................. 38
9.29 -DISPLAY OF THE ISTANTANEOUS FLOW................................................................................................................ 38
9.30 -DISPLAY OF THE ABSOLUTE TOTALIZER OF THE DIRECT FLOW .................................................................... 38
9.31 -DISPLAY OF THE ABSOLUTE TOTALIZER OF THE INVERSE FLOW.................................................................. 39
9.32 -DISPLAY OF THE RESETTABLE TOTALIZER OF THE DIRECT FLOW ................................................................ 39
9.33 -DISPLAY OF THE RESETTABLE TOTALIZER OF THE INVERSE FLOW ............................................................. 39
9.34 -RESET OF THE RESETTABLE TOTALIZERS ............................................................................................................. 39
9.35 -DISPLAY OF THE TIME COUNTER............................................................................................................................. 39
9.36 -DISPLAY OF THE PROGRAMMED ND ....................................................................................................................... 40
9.37 -DISPLAY OF THE FILTER TIME .................................................................................................................................. 40
9.38 -DISPLAY OF THE TECHNICAL UNIT OF THE ISTANTANEOUS FLOW............................................................... 40
9.39 -DISPLAY OF THE TECHNICAL UNIT OF TOTALIZERS .......................................................................................... 40
9.40 -DISPLAY OF THE MESSAGE WRITTEN ON THE FIRST LINE OF LCD ................................................................ 40
9.41 -DISPLAY OF THE MESSAGE WRITTEN ON THE SECOND LINE OF LCD ........................................................... 41
9.42 -DISPLAY OF THE MINIMUM SET ............................................................................................................................... 41
9.43 -DISPLAY OF THE MAXIMUM SET.............................................................................................................................. 41
9.44 -PRESETTER DISPLAY ................................................................................................................................................... 41
9.45 -DISPLAY OF THE KIND OF RESET ............................................................................................................................. 41
9.46 -DISPLAY OF THE DIGITAL OUTPUT 1 AND RESPECTIVE LENGHT OF THE TOTALIZATION PULSES....... 42
9.47 -DISPLAY OF THE DIGITAL OUTPUT 2, 3, 4, AND RESPECTIVE LENGHT OF THE TOTALIZATION PULSES42
9.48 -DISPLAY OF THE ANALOG OUTPUT......................................................................................................................... 42
9.49 -CONFIRM OF CONFIGURATION AFTER THE MESSAGE “ERROR IN DATA/EEPROM MEMORY” ................ 42
9.50 -DISPLAY OF THE VERSION OF SOFTWARE............................................................................................................. 43
9.51 -DISPLAY OF THE SERIAL NUMBER .......................................................................................................................... 43
9.52 -ERROR MESSAGES ........................................................................................................................................................ 43
9.53 -SERVICE MESSAGES..................................................................................................................................................... 43

10 - ELECTRIC CONNECTIONS OF ELECTROMAGNETIC FLOWMETER AP1013 - AP1014 WITH


CONVERTER AP501 WITHOUT OPT. 18 (WITHOUT CONNECTOR) 44
Automation Progetti
TITLE Page 4

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
11 - ELECTRIC CONNECTIONS OF ELECTROMAGNETIC FLOWMETER AP1013 - AP1014 WITH
CONVERTER AP501 WITH OPT. 18 (WITH CONNECTOR) 45

12 - ELECTRIC CONNECTIONS OF ELECTROMAGNETIC FLOWMETER AP1017 - AP1018


AP1016/DIN – AP1116 WITH CONVERTER AP501 WITHOUT OPT. 18 (WITHOUT CONNECTOR)
46

13 - ELECTRIC CONNECTIONS OF ELECTROMAGNETIC FLOWMETER AP1017 - AP1018


AP1016/DIN – AP1116 WITH CONVERTER AP501 WITH OPT. 18 (WITH CONNECTOR) 47

14 - ELECTRIC CONNECTIONS OF THE TWO-WIRE 4÷20mA OUTPUT 48

15 - ELECTRIC CONNECTIONS OF THE ACTIVE 4÷20mA OUTPUT 49

16 - ELECTRIC CONNECTIONS OF THE PROGRAMMABLE DIGITAL OUTPUTS 50

17 - ELECTRIC CONNECTION OF THE RS232 OUTPUT 51

18 - ELECTRIC CONNECTIONS DIGITAL INPUT 52

19 - CONNECTION DIAGRAMM BETWEEN RS232 AND P.C. 53

20 - DIMENSIONS AND WEIGHTS OF FLANGES 54


Automation Progetti
TITLE Page 5

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
1. INTRODUCTION

1.1 GENERAL INFORMATION


The electromagnetic flowmeter uses the Faraday’s electromagnetic induction law to measure the flow of conductive
liquids.
This device was engineered to be compact and light. It is composed by two parts separately installed:
- Nr 1 insertion electromagnetic flowmeter; it must be installed on piping where the process liquid flows
(AP1013–AP1014-AP1017-AP1018-AP1016/DIN-AP1116).
- Nr 1 converter; where the electric signal coming from meter is amplified and processed (AP501 – AP501/B).
They are connected by only one cable.
In the IP68 version this cable is sealed and it is coming out from meter.
Another particular characteristic of this instruments is the utilization of a square wave excitation current. This allows
to obtain a good zero stability (autozero).

1.2 - CHARACTERISTICS
The insertion electromagnetic flowmeters and the converter AP501 – AP501/B present the following characteristics:
1- compact, light, easy installation and low cost.
2- high accuracy with short upstream and downstream rectilinear tract. It is possible to obtain accurate
measurements also with uneven flow.
3- square wave excitation; zero stability.
4- the utilization of integrated circuit with C-MOS technology. The utilization of a microprocessor reduces the
complexity of the electronic part and increases reliability.
5- limited power consumption. Only 9 Watts (with 24 Vdc power supply)
6- separated or joined installation of converter on meter. If circumstances don't allow the installation of remote
converter, we can supply the compact version.
7- it doesn't need a previous choice of the fluid type. If it is conductive and non-magnetic, it is possible to obtain
stable measures of flow without errors.
8- it allows accurate measurements of the flow of sludge. The non-uniform distribution of velocity, present in the
flow of sludge, doesn't influence measurement, allowing an high accuracy.
9- negligible pressure loss. The pressure losses aren't appreciable, because the flow of liquid doesn't find any
obstacles.
10- linear relation between flow and output signal. It can execute accurate measurements on a large range of flows.
11- the parameters of fluids don't influence measurement. The temperature, the pressure, the density, the
conductivity and the type of fluid don't influence the accuracy of measurement.
12- the presence of deposits on electrodes doesn't influence the measurement. Flowmeter will not work any more
only when electrodes will be completely insulated.
Automation Progetti
TITLE Page 6

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
Automation Progetti
TITLE Page 7

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
1.3 - SPECIFICATIONS
Range of measurement :from 0,1 m/s to 6 or 12m/s (velocity of flow)
Accuracy : see fig. 1
Automation Progetti
TITLE Page 8

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54

FIG. 1

Minimum conductivity: 5 µS/cm.


Pressure of fluid ( Kg/cm2 ) : from 0 to the nominal pressure of flanges for flowmeters with Teflon lining;
from –1 bar to the nominal pressure of flanges for flowmeters with hard rubber
lining.
Automation Progetti
TITLE Page 9

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
1.4 - DESCRIPTION
It is constituted by an AISI 304 pipe, that has, at its own ends, two carbon steel flanges on the versions AP1018 -
AP1017 - AP1013 - AP1014 or two AISI 304 unions type DIN 11851 on the versions AP1016/DIN. An insulator
material ( teflon or hard rubber ) lines it too. This lining material is present also on flanges.
On demand, you can have the meter with stainless steel flanges.
In the version AP1116 the lining ( only teflon ) is used also as principal structure.
The windings, to produce the magnetic field, are in the external side of the pipe. A steel cover and a polyurethane
resin casting protect them.
The two electrodes are in the middle of meter on two points that are diametrically opposed.
The connector is in the external side of the meter and is equipped with a protective cap. To avoid the entry of dirt,
don’t ever leave it open!
To avoid the loss of cap, a plastic wire connects it to meter.
In the IP 68 version, cable is resinated inside a junction box to assure the protection. You have to declare the lenght
of this cable when order the flowmeter.
A matching card (AP550) completes the supply of meter. This card has the same serial number of flowmeter and
must be inserted on the suitable connector inside converter AP501 – AP501/B ( see drawing 5582-5583-5584-5585).
If the converter is comprised in the supply, we already insert the card AP550. In case of replacement of the converter
AP501 – AP501/B, you must insert the card AP550 inside the new converter.

MEASUREMENT BLOCKAGE
Converter may have the measurement blockage device.
You have to use this device when you don’t desire to count a liquid flowing inside flowmeter.
Making contact between clamps 12 and 13, you block measurement. It will start again when you will open on the above-
mentioned contact (see drawing 5590). See “Configuration of the digital input” for the right programming.
Automation Progetti
TITLE Page 10

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54

1.5 - CONNECTION CABLE


In version with remote converter, supply includes a connection cable between flowmeter and converter.
The standard lenghts of cable are 5 and 10 meters.
There are four kind of cable to connect flowmeter to converter.

CABLE AP216-0
It’s a cable without connectors.
Flowmeter and converter have a cabling terminal block.
You have to use to connect flowmeters mod. AP1013, AP1014, AP1017(IP68) and AP1018(IP68) with converters
AP501 and AP501/B without optional 18 (see optionals).

CABLE AP216-1C
It’s a cable with only one connector at the converter end.
Flowmeter has a cabling terminal block.
You have to use to connect flowmeters mod. AP1013, AP1014, AP1017(IP68) and AP1018(IP68) with converters
AP501 and AP501/B with optional 18 (see optionals).

CABLE AP216-1M
It’s a cable with only one connector at the flowmeter end.
Converter has a cabling terminal block.
You have to use to connect flowmeters mod. AP1017, AP1018 AP1016/DIN and AP1116 with converters AP501
and AP501/B without optional 18 (see optionals).

CABLE AP216-2C
It’s a cable with two connectors.
You have to use to connect flowmeters mod. AP1017, AP1018 AP1016/DIN and AP1116 with converters AP501
and AP501/B with optional 18 (see optionals).
Connector to insert in flowmeter has a reference band (see drawing 5584-5585).
The maximum lenght is 50 meters.
The extenal diameter of connector is 29mm about.
CREATE A BEND SO CONDENSATE OR RAIN DON’T REACH CONNECTOR (SEE DRAWING 5142).
DON’T MAKE THIS CABLE PASS WITH POWER CABLES TO AVOID NOISES ON MEASUREMENT.
IF POSSIBLE PROTECT IT BY A STEEL PIPE.
Automation Progetti
TITLE Page 11

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
MODULAR CONVERTER FOR ELECTROMAGNETIC FLOWMETER
AP501 – version BASE (low-cost) / AP501 – version HI-TECH
Table of working/ranges

DN Min. Min. F.S. Max. F.S. Max. F.S.


measurement (6m/s) (12 m/s)
(mm) (mc/h) (mc/h) (mc/h) (mc/h)
3 0,0025 0,015 0,15 0,30
6 0,01 0,06 0,6 1,2
8 0,017 0,1 1 2
10 0,034 0,2 2 4
BLIND BLIND 15 0,067 0,4 4 8
FOR FIELD MOUNTING FOR RACK MOUNTING 20 0,14 0,8 8 16
25 0,17 1 10 20
32 0,34 2 20 40
40 0,5 3 30 60
50 0,67 4 40 80
65 1 6 60 120
80 1,7 10 100 200
100 3,4 20 200 400
125 5 30 300 600
150 6,7 40 400 800
175 8,4 50 500 1000
WITH DISPLAY WITH DISPLAY 200 14 80 800 1600
FOR FIELD MOUNTING FOR RACK MOUNTING 250 17 100 1000 2000
BASE VERSION BASE VERSION 300 25 150 1500 3000
350 34 200 2000 4000
400 50 300 3000 6000
450 60 360 3600 7200
500 67 400 4000 8000
600 100 600 6000 12000
700 134 800 8000 16000
800 167 1000 10000 20000
900 200 1200 12000 24000
1000 250 1500 15000 30000
1100 280 1680 16800 33600
WITH DISPLAY WITH DISPLAY 1200 334 2000 20000 40000
FOR FIELD MOUNTING FOR RACK MOUNTING
HI-TECH VERSION HI-TECH VERSION

GENERAL TECHNICAL DATA


Version HT : standard blind microprocessor based (C-MOS); with non-volatile memory (EEPROM); i.c.s. with sockets
Version B : standard blind microprocessor based (C-MOS); with non-volatile memory (EEPROM); welded i.c.s.
Mounting : standard field (see optional bull. 204)
Casing : in plastic material (field version)
Compensation : automatic of zero (autozero)
Range : 0÷6/12 m/s
F.S. : programmable (see table)
Autorange : automatic
Input : from electromagnetic flowmeter
DC output : galvanically insulated monodirectional 4÷20mA:
passive (10 ÷ 30 Vdc) ; active (0 ÷ 600 Ω) (see optional bull. 204)
Interchangeability : total for any flowmeter
Interchangeab.accuracy : without adjustment: B vers. ±0,15% R.V. – HT vers. ±0,1% R.V.
Connections : with terminal block and PG11/PG13,5 fittings cable
Electric protection : line filters, fuses, varistors on power supply (CE)
Mec. Protection : IP65 CEI EN 60529 (see optional bull. 204)
Anti – tampering sealing : field version
Ambient temp. : B vers. –10 ÷ 50 °C - HT vers. –20 ÷ 60 °C
Humidity : 4÷90% non condensing
Supply voltage : std 220/110 Vac ±10% (3W) (see optional bull. 204)
Dimensions : W240 H210 D110 mm (field version)
: L213 H128 P185 mm (rack version) (standard rack 3U) Bull. 203
Weight : 2Kg Date January 2000
Burn-in : B vers. 40 hours - HT vers. 120 hours Italian version
MTBF : B vers. 50'000 hours - HT vers. 100'000 hours English version X
Optionals : see bull. 204
Accuracy of system (meter + conv.) after adjustment with absolute mode
(not by comparison ) : see flowmeters
Automation Progetti
TITLE Page 12

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
OPTIONAL AP501 / B o HI-TECH

DESCRIPTION TAG
Rack mounting 3U 19” ½ rack standard (213x128x200 mm) (prot. IP55) Q
3U 19” rack mounting frame for two equipments B
3U 19” ½ rack closing panel for frame C
Keyboard and display Clear 16 keys anti-dust keyboard with mnemonic symbols and access keyword
16 characters 2 lines LCD with Italian / English messages D
Visualizations in volume and weight engineering units
Istantaneous flow, resettable and absolute totalization, autotest, programming
Bach-lit LCD - it needs of the opt. D E
Range 6 m/s 6
12 m/s 12
Totalizers Nr 1 resettable - it needs of the opt. D G
Nr 1 absolute (fiscal) - it needs of the opt. D H
Nr 2 resettable + Nr 2 absolute (for bidirectional flow) - it needs of the opt. D Y
+ Nr 1 presetter for batch
Analog output Mono / bidirectional 4÷20 mA with galvanic separation P
Digital outputs Nr 1 relay (Max 50Vdc 0,2A) for totalization - non compatible with opt. R M
Nr 4 relay (Max 50Vdc 0,2A) programmable as: - it needs of the opt. D
- direct totalizer; - non compatible with opt. M
- inverse totalizer;
- minimum and maximum flow alarm; R
- malfunction – empty pipe alarm;
- flow direction;
- presetter (batch)
Digital input Nr 1 passive contact programmable as: - non compatible with opt. 3
- partial totalizers reset; - it needs of the opt. G or Y
- zero blockatge of measure (for washing) T
- presetter reset - it needs of the opt. Y
Serial output RS485 – interface for P.C. 4
RS232 – interface for P.C. / modem / radio link 3
Special function Calendar clock; - it needs of the opt. D
Time counter for fiscal mesurement; U
Others on specific requirements
Cut-off Programmable - it needs of the opt. D K
Empty pipe device - Zero blockage of measure
- Signalintg on display - it needs of the opt. D W
- Alarm on digital output - it needs of the opt. R
Electrodes auto-cleaning 16
Connections to flowmeter Industrial connector 18
Power supply 220/230Vac ±10% 2
110/115Vac ±10% 1
24Vac ±10% N
24Vdc ±10% Z
Altre a richiesta S

Bull. 204
Date January 2000
Italian version
English version X
Automation Progetti
TITLE Page 13

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
ELECTROMAGNETIC FLOWMETER
AP1116 - HI-TECH version

Overal dimensions(mm)
DN DE Weight (Kg)
3-6-8-10-15-
105 2
20-25-32
40 - 50 105 3
65 127 3
80 142 4
100 162 4

TECHNICAL DATA
Version : industrial HI-TECH
Mounting : wafer
Nominal size : from ND3 to ND100 mm
Line pressure : NP10 Bar
Structure : stainless steel (AISI304-1.4301 - AISI316-1.4401) repairable (not welded)
Lining : TEFLON
Process temperature : -40 ÷ 120 °C
Anticondensate sealing : polyurethane resin
Measure electrodes : AISI316L-1.4404 (see optional)
Measure range : 0,1 ÷ 12 m/s
Interchangeability : total for any converter
Interchangeability accuracy : without adjustment, ±1% R.V.
Output : to converter
Electrical connections : with industrial connector
Insulation thermic class : H (180 °C)
Ambient temperature : -30 ÷ 80 °C
Protection : IP66 CEI EN 60529 (see optional)
Humidity : 100% (no air inside)
Supply : from converter (24Vdc)
Power consumption : 4W
Dimensions : see table
Burn-in : 120 hours
MTBF : 130'000 hours
Optionals : - nominal pressure NP16 - NP40
: - Hastelloy B or C – Titanium electrodes
: - IP68-10 w.c.m. protection (cable connected and resinated by manufacturer)
: - lenght on request
: - special executions on request
Accuracy of system (meter + conv.) after adjustment with absolute mode (not by comparison )
Standard : ±1% R.V.
On request : ±0,2% R.V. - ±0,5% R.V.
For other data : see converters AP219 - AP501

NOTE: Teflon is registered trademark of DuPont Dow Elastomers


Bull. 207
Date January 2000
Italian version
English version X
Automation Progetti
TITLE Page 14

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
ELECTROMAGNETIC FLOWMETER
AP1016 DIN - HI-TECH version

Overal dimensions (mm)


ND ED Weight (Kg)
25 105 6
32 105 6
40 105 6
50 105 7
65 127 7
80 142 8
100 162 9
150 220 11

TECHNICAL DATA
Version : industrial for alimentary liquids HI-TECH
Mounting : DIN 11851
Nominal size : from ND25 to ND150 mm
Line pressure : NP10 Bar
Structure : stainless steel (AISI304-1.4301 – AISI316-1.4401)
Structure : repairable (not welded)
Lining : TEFLON
Process temperature : -40 ÷ 150 °C
Anticondensate sealing : polyurethane resin
Measure electrodes : AISI316L-1.4404 (see optional)
Measure range : 0,1 ÷ 12 m/s
Interchangeability : total for any converter
Interchangeability accuracy : without adjustment ±0,2% R.V.
Output : to converter
Electrical connections : with industrial connector
Insulation thermic class : H (180 °C)
Ambient temperature : -30 ÷ 80 °C
Protection : IP66 (see optional)
Humidity : 100% (no air inside)
Supply : from converter (24Vdc)
Power consumption : 4W
Dimensions : see table
Burn-in : 120 hours
MTBF : 130'000 hours
Optional : - nominal pressure NP16
: - Hastelloy B or C - Titanium electrodes
: - IP68-10 w.c.m. protection (cable connected and resinated by manufacturer)
: - lenght on request
: - special executions on request
Accuracy of system (meter + conv.) after adjustment with absolute mode (not by comparison )
Standard : ±0,2% R.V.
For other data : see converters AP219 - AP501
Bull. 208
NOTE: Teflon is registered trademark of DuPont Dow Elastomers Date January 2000
Italian version
English version X
Automation Progetti
TITLE Page 15

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
ELECTROMAGNETIC FLOWMETER
AP1017 – AP1018 - HI-TECH version
Overall dimensions (mm) with flanges UNI 2277 PN10 (DIN 2576).
ND A L Weight (Kg)
∅ Standard ISO Standard ISO
25 152 250 200 7 5
32 178 250 200 8 6
40 178 250 200 9 7
50 178 250 200 10 8
65 219 250 200 12 9
80 219 250 200 13 11
100 235 250 200 16 12
125 273 250 250 19 16
150 278 300 300 23 19
200 327 300 350 30 23
250 387 300 450 38 32
300 425 300 500 50 42
350 490 300 550 72 54
400 527 300 600 96 77
450 590 300 102 112
500 628 300 120
600 730 400 150
700 834 400 195
800 940 400 250
900 1038 400 290
1000 1138 400 360
1100 1234 600 450
1200 1341 600 540

TECHNICAL DATA
Version : industrial HI-TECH
Mounting : flange UNI – ANSI – DIN – ISO
Nominal size : AP1017 from ND25 to ND400 mm ; AP1018 from ND25 to ND1200 mm
Nominal pressure : standard NP10 (see optional)
Measure piping : AISI316
Structure, flange : painted carbon steel (see optional)
Structure : repairable (not welded)
Lining : AP1017 TEFLON; AP1018 alimentary hard rubber
Process temperature : AP1017 -40 ÷ 150 °C ; AP1018 -25 ÷ 110 °C
Anticondensate sealing : polyurethane resin
Measure electrodes : AISI316L-1.4404 (see optional)
Measure range : 0,1 ÷ 12 m/s
Interchangeability : total for any converter
Interchangeability accuracy : without adjustment,
ND25÷300 ±0,2% R.V. ; ND350÷500 ±0,5% R.V. ; ND600÷1000 ±1% R.V. ; ND1200 ±2% R.V.
Output : to converter
Electrical connections : with industrial connector
Insulation thermic class : H (180 °C)
Ambient temperature : -30 ÷ 80 °C
Protection : IP 66 CEI EN 60529 (see optional)
Humidity : 100% (no air inside)
Supply : from converter (24Vdc)
Power consumption : 6W
Dimensions : see table
Lenght : L standard (see optional)
Burn-in : 120 hours
MTBF : 130'000 hours
Optionals : - nominal pressure NP16 – NP25 – NP40 – NP64
: - Stainless steel construction (AISI304-1.4301 - 316-1.4401)
: - Hastelloy B or C – Titanium electrodes
: - IP67 protection
: - IP68-10 w.c.m. protection (cable connected and resinated by manufacturer)
: - lenght ISO 13359
: - lenght on request
: - special executions on request
Accuracy of system (meter + conv.) after adjustment with absolute mode (not by comparison)
Standard : ±0,2% R.V. - ±0,5% R.V. (function of diameters)
±1% R.V. - ±2% R.V. (function of diameters) Bull. 206
For other data : see converters AP219 - AP501 Date January 2000
Italian version
NOTE: Teflon is registered trademark of DuPont Dow Elastomers English version X
Automation Progetti
TITLE Page 16

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
ELECTROMAGNETIC FLOWMETER
AP1013 / AP1014 – BASE version (low-cost)
Overall dimensions (mm) with flanges UNI 2277 PN10 (DIN 2576 – ISO2501).

ND A L Weight (Kg)
∅ Standard ISO Standard ISO
20 152 250 200 6 5
25 152 250 200 7 6
32 178 250 200 8 7
40 178 250 200 9 8
50 178 250 200 10 9
65 219 250 200 12 11
80 219 250 200 13 12
100 235 250 250 16 16
125 273 250 250 19 19
150 278 300 300 23 23
200 327 300 350 30 32
250 387 300 450 38 42
300 425 300 500 50 54
350 490 300 550 72 77
400 527 300 600 96 112
450 590 300 102
500 628 300 120
600 730 400 150

TECHNICAL DATA
Version : base – LOW COST
Mounting : flange UNI – ANSI – DIN – ISO
Nominal size : AP1013 from ND20 to ND400 mm ; AP1014 from ND25 to ND600 mm
Nominal pressure : standard NP10 (see optional)
Measure piping : AISI316
Structure, flange : painted carbon steel
Structure : repairable (not welded)
Lining : AP1013 PTFE ; AP1014 hard rubber
Process temperature : AP1013 -30 ÷ 130 °C (150 °C 1 hour max) ; AP1014 -10 ÷ 100 °C
Anticondensate sealing : polyurethane resin
Measure electrodes : AISI316L – 1.4404 (see optional)
Measure range : 0,1 ÷ 12 m/s
Interchangeability : total for any converter
Interchangeability accuracy : without adjustment
ND20 ÷ 300 ±0,5% R.V. ; ND350 ÷ 600 ±1% R.V.
(see optional)
Output : to converter
Electrical connections : with terminal block and PG13,5 fitting cable
Insulation thermic class : F (155 °C)
Ambient temperature : -30 ÷ 70 °C
Protection : IP67 CEI EN 60529 (see optional)
Humidity : 4 ÷ 95 % non condensing (see optional) (air inside)
Supply : from converter (24Vdc)
Power consumption : 6W
Dimensions : see table
Lenght : L standard (see optional)
Burn-in : 40 hours
MTBF : 60'000 hours
Optionals : - nominal pressure NP16 – NP25 – NP40
: - Hastelloy B or C - Titanium electrodes
: - IP68/10 w.c.m. protection
100% humidity (no air inside) cable connected and resinated by manufactrurer
: - lenght ISO 13359
: - lenght on request
: - special executions on request
Accuracy of system (meter + conv.) after adjustment with absolute mode (not by comparison )
Standard : ±0,5% R.V. - ±1% R.V. (function of diamenter)
On request : ±0,2% R.V. - ±0,5% R.V. (function of diameter)
Bull. 205
For other data : see converters AP219 – AP501
Date January 2000
Italian version
English version X
Automation Progetti
TITLE Page 17

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54

2 - PRINCIPLE OF OPERATING
The magnetic flowmeter is an instrument measuring the flow of a fluid using the Faraday's electromagnetic induction
law.
As illustrated in picture Nr 2, a magnetic flux with density B (perpendicular to the direction of flow) crosses an
insulated pipe where fluid flows.
Flux flowing through the pipe must be electrically conductive. In this case this magnetic flux induces a potential
difference E that’s detectable by two electrodes (perpendicular to the direction of flow and to the magnetic field).
This potential is proportional to the velocity of fluid V in the measurement point.
E = K*B*D*V (V) (formula 1)

E : voltage induced to electrodes (V)


K : dimensionless constant
B : magnetic flux density
D : distance between the two electrodes
V : velocity of fluid
The flow of the fluid Q (m3/s) is given by the following formula:
Q = PI*D*D*V/4 (formula 2)
The combination of both formulas gives as result the following formula:
E = K*B*D*Q*4 / PI*D*D = 4 K*B*Q / PI*D
The signal E, multiplied for a suitable coefficent due to installation, is a signal proportional to flow.

Fig. 2 - Principle of operating

The electromagnetic flowmeters may have d.c. or a.c. excitations. When the systems are excited in a.c., electrostatic
or electromagnetic noises may influence measurement.
Otherwise the d.c. excited systems are less sensible to noise produced by the electromagnetic induction, because
voltage inducted to electrodes is a d.c. However they may have errors of measurement because of the electrochemical
polarization between electrodes and fluid.
The electromagnetic flowmeters excited by a square wave exclude the handicaps of both d.c. and a.c. systems and
have the advantages of both ones.
First of all a timer circuit with about 8 pulses/sec frequency controls the polarity of the square wave current exciting
flowmeter that gives an electromotive force with a wave form like the one of the excitation current.
The input stage of converter, an high impedance differential amplifier, receives and amplifies this electromotive
force.
The next stage, a variable gain amplifier automatically controlled by microprocessor, establishes the velocity range
of fluid. Now the A/D converter module receives this signal. This stage integrates signal for an established period,
then microprocessor impresses to the input of integrator a reference voltage that inverts the slope of the ramp of
integrator. As result we have a pulse with a length that is proportional to the amplitude of the input signal and
conseguently proportional to the velocity of fluid. Microprocessor receives, measures width with resolution higher
than one microsecond, converts in technical units and shows on LCD the results of processing.
Automation Progetti
TITLE Page 18

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
3 - INSTALLATION

3.1 - PRECAUTION FOR THE CHOICE OF THE PLACE


MECHANICAL PRECAUTION
• Choose a place where there are no flow pulsations.
• Avoid to install flowmeter where pipe presents vacuum.
• Avoid to install equipments where there are vibrations.
• Avoid the exposure at the sun-light or use a protective screen (for converter only).
• Avoid to install equipments inside place with dampness or corrosive environment.
• Avoid to install equipments too high or inside a close space.
• Choose an airy place where you can easily work.
• Don’t install flowmeter on piping with cathodic protection without the respective insulating flanges.
• Avoid condensate reaches connector (see drawing 5142).
• Inject other liquids downstream flowmeter.
• Respect the flow direction indicated by arrow imprinted on flowmeter (see drawing 5604).

ELECTRIC PRECAUTION
Inputs / outputs :
• Execute connections with 2 x 1 mm2 shielded cable.
Power supply :
• Execute connections with 3 x 1 mm2 cable.
• Public main : 220/110 Vac ±10% 50/60 Hz (instrumental line without noises).
• 24Vdc converters connect the (-) of the power supply to ground. So, to avoid “ground loops”, this power supply
has not to feed other instruments connecting (-) to ground.
• Be sure to have a good earthing.
• Avoid to install the equipment near electrical devices that can cause noises in measurement (like motors,
transformers, radio transmitters and equipments generating electromagnetic or electrostatic noises). In case of
doubt, make cables pass inside to at least 2mm thick metallic tubes and ground them.
• In case of particulary hampered ambient, as the heavy industrial ambient, you have to tube cables inside conduits
and ground them. You have to re-establish the electromagnetic ambient of the light industry.
• Keep plugged the unused cable fittings.
Automation Progetti
TITLE Page 19

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
3.2 - INSTALLATION OF THE FIELD VERSION CONVERTERS
Automation Progetti
TITLE Page 20

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
3.3 - DIMENSIONE AND TEMPLATE OF THE RACK VERSIONS
Automation Progetti
TITLE Page 21

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
3.4 - INSTALLATION OF THE CABLE BETWEEN FLOWMETER AND CONVERTER
Automation Progetti
TITLE Page 22

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
3.5 - INSTALLATION CONDITIONS
A) install two straight piping tracts upstream and downstream of flowmeter without disturbing elements (see fig.3).
The inner diameter of piping must be equal or upper (3% max) then the inner diameter of flowmeter.

Notes: fig.3; D is the diameter of piping, L is the lenght of the straight piping upstream flowmeter, L1 is the
lenght of the straight piping downstream flowmeter.
A diameter variation of pipe must be carried out by cones with enlargement / narrowing angle ≤ 7,5°.

Fig.3. Minimum lenght of straight pipings to install between the noise elements and flowmeter.
Automation Progetti
TITLE Page 23

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54

B) Flowmeter can be installed in horizontal, vertical or sloping pipings.


If you install it in an horizontal piping, it must always work completely full of liquid. Flowmeter must be installed
at one side of piping (no on top or down).
If you install it in a vertical piping, you must take care to make the fluid run upwards (see fig. 4).

Fig.4. Installation position of meter


In every condition, electrodes must be in horizontal position respect to the ground plane (see fig.5).

Fig. 5

C) The direction of flow is indicated on meter by an arrow.


D) If flowmeter will frequently need a check up or a cleaning, it must be installed with a by-pass .
E) To easy installation, big diameter flowmeters have two eye-bolt.

IMPORTANT:
BEFORE TO INSTALL IT, CLEAN PERFECTLY ELECTRODES WITH TRICHLOROETHYLENE OR
DEGREASING SUBSTANCES.
Automation Progetti
TITLE Page 24

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
3.6 - HOW TO CONNECT CABLES
The outputs of cables may have IP66 or IP67 waterproof cable fittings (see fig.6) or industrial connectors.
a) - Use of the IP66 cable fittings:

b) - Use of the IP67 cable fittings (for converters with optional 7 only):

Fig.6

To connect a cable to converter take down the cable fitting (see fig.6). Put out the blind cap and put cable inside the
various components. Connect wires to terminals through cable terminals. Be sure to have the right lenght of cable,
fix nut to realize airproof. Be sure to have the boot of cable in perfect condition in the contact zone.
Never leave unused cable fittings without the blind cap.
To connect equipments use only the supplied cable.
c) – Use of connectors:
Before to engage connector, remove the protection sheath.
Connector is polarized, so it’s no possible to make connection errors.
Cable has a reference band at an end. This end must be connected to meter.
IMPORTANT :
Never leave the unused connectors of equipments without the relevant protections.
Possible dirt or humidity can distort measurement.
Automation Progetti
TITLE Page 25

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
4 - AP1116 MOUNTING
Automation Progetti
TITLE Page 26

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
5 - GROUNDINGS OF THE FLOWMETERS AP1013-AP1014-AP1017-AP1018
Automation Progetti
TITLE Page 27

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
6 - GROUNDINGS OF THE FLOWMETER AP1116
Automation Progetti
TITLE Page 28

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54

7 - START UP

7.1 - STARTING CHECKS


After installation and connections of the electromagnetic flowmeter, check the following points :
A) verify the serial number of the matching card AP550. It must be the same of flowmeter.
B) verify the power supply wrotten on label. It has to have the same value and tollerance of your main.
C) verify exactitude of the electric connections.
D) verify tightening of the terminal blocks screws.
E) verify the perfect watertight of the cable fittings and gaskets.
F) check the flow direction. It must be the same of arrow on flowmeter.
G) verify to have a good grounding of flowmeter in according with the UNI standard.

7.2 - MEASUREMENT STARTING


If all controls result positive, check the following points :

7.2.1 - CHECK OF ZERO POINT


A) be sure to have flowmeter full of motionless water
B) power on the converter and wait for at least 10 minutes
C) when you power on converter, it carries out an autotest. Verify it doesn’t show an error message. In case, see
“ERROR MESSAGES”
D) display must show 0.0

7.2.2 - CHECK OF FLOW DIRECTION


A) make liquid flow
B) check signal shown on display increases. If it becomes negative, verify the flow direction is correct or verify
connections inside converter and flowmeter (in instruments without connector only).

7.3 - MEASUREMENT
After the above-mentioned checks, flowmeter is ready for use. Now you can program converter using the following
instructions.

8 - PROGRAMMING OF THE CONVERTER AP501 – AP501/B WITHOUT OPTIONAL D


(see optionals)

This kind of converter is supplied with software in accordance with data of customer.
This because it has no keyboard.
Automation Progetti
TITLE Page 29

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9. - PROGRAMMING OF THE CONVERTER AP501 – AP501/B WITH OPTIONAL D
(see optionals)
The converters AP501 - AP501/B have a keyboard with 16 membrane keys. By this keyboard you can completetly
program converter (ND, outputs and visualizations). Symbols you can see on each key are simple to remember. This
to help your work.
Words inserted between the square brackets are what display will show; the ones between the round brackets are a
comment.
Converter has an access keyword to block the programming functions. This device is necessary to avoid
programming alterations by not qualified operators.
AUTOMATION PROGETTI supplies converter with the access keyword = 0. Qualified operators can change this
value.

9.1 -ACCESS KEYWORD INSERTION

• Press [ ACCESS KEYWORD ]


• Write the key number ( Now 0 ) [ ∗ ]

ENTER
• Press [ CORRECT ]

9.2 -COUPLING OF CONVERTER TO ND OF PIPING


• Insert the access keyword
DN
• Press 4 [ DIAMETER ND ]

Press more times or till it shows the desired ND [ 3, 6, 8, 10, 15, 20, 25, 32, 40, 50, 65. 80,
100, 125, 150, 175, 200, 250, 300, 350, 400, 450, 500, 600, 700, 800, 900, 1000, 1200, 1400].

ENTER
• Press

9.3 -PROGRAMMING OF THE TECHNICAL UNITS OF THE INSTANTANEOUS FLOW


• Insert the access keyword

l Kg
• Press 6 [ MEASURE UNITS / FLOW ]

Press more time or till it shows the desired technical unit


(Only with opt. 6)
[ l/h , l/m , cc/s , %V , kg/h , kg/m , g/s , m/s for ND from 3 to 10 ;
mc/h , l/m , l/s , %V , t/h , kg/m , kg/s , m/s for ND from 15 to 80 ;
mc/h , mc/m , l/s , %V , t/h , t/m , kg/s , m/s for ND from 100 to 700 ;
mc/h , mc/m , mc/s , %V , t/h , t/m , t/s, m/s for ND from 800 to 1400 ]
Automation Progetti
TITLE Page 30

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54

(Only with opt. 12)


[ l/h, l/m, cc/s, %V, kg/h, kg/m, g/s, m/s, for DN from 3 to 8;
mc/h, l/m, l/s, %/V, t/h, kg/m, kg/s, m/s, for DN from 10 to 80;
mc/h, mc/m, l/s, %V, t/h, t/m, kg/s, m/s, for DN from 100 to 500;
mc/h, mc/m, mc/s, %V, t/h, t/m, t/s, m/s, for DN from 600 to 1400 ]

ENTER
• Press

9.4 -PROGRAMMING OF THE TECHNICAL UNITS OF TOTALIZERS


(With opt. G, H or Y)
• Insert the access keyword

l Kg
• Press two times 6 [ MEASURE UNITS / PULSES ]

Press more times or till it shows the desired technical unit

(Only with opt. 6)


[ cc , cl , dl , l , dal , g , dg , hg , kg , dkg , for ND 3 ;
cl , dl , l , dal , hl , dg , hg , kg , dkg , qli for ND from 6 and 10 ;
dl , l , dal , hl , mc , hg , kg , dkg , qli , t for ND from 15 to 32 ;
l , dal , hl , mc , dmc , kg , dkG , qli , t , dt for ND from 40 to 100 ;
dal , hl , mc , dmc , hmc , dkg , qli , t , dt , ht for ND from 125 to 350 ;
hl , mc , dmc , hmc . kmc , qli , t , dt , ht , kt for ND from 400 to 1400 ]

(Only with opt. 12)


[ cc, cl, dl, l, dal, g, dg, hg, kg, dkg, for DN 3;
cl, dl, l, dal, hl, dg, hg, kg, dkg, q, for DN from 6 to 8;
dl, l, dal, hl, mc, hg, kg, dkg, q, t, for DN from 10 to 25;
l, dal, hl, mc, dmc, kg, dkg, q, t, dt, for DN from 32 to 80;
dal, hl, mc, dmc, hmc, dkg, q, t, dt, ht, for DN from 100 to 300;
hl, mc, dmc, hmc, kmc, q, t, dt, ht, kt, for DN from 350 to 1000;
mc, Dmc, hmc, kmc, Dkm, t, Dt, ht, kt, Dkt, for DN from 1100 to 1400

( cl = 0.01l , dl = 0.1 l , dal = 10 l , hl = 100 l , dg = 10 g , hg = 100 g , dkg = 10 kg , dmc = 10 mc ,


dt=10t , hmc=100 mc , ht=100t , kmc=1000mc , kt=1000t )

ENTER
• Press
NOTE :
Changing the technical unit , the pulses number showed on display doesn’t change .
Automation Progetti
TITLE Page 31

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.5 -PROGRAMMING OF THE FIRST LINE OF LCD
• Insert the access keyword

• Press DISPLAY 1 [ LINE 1 FORMAT ]

• Press more time or till it shows the desired message


[ Flw (Istantaneous flow), Tot+ (Absolute totalizer of the direct flow) (with opt. H or Y), Trs+ (Resettable totalizer
of the direct flow) (with opt. G or Y), Tot- (Absolute totalizer of the inverse flow) (with opt. Y), Trs- (Resettable
totalizer of the inverse flow) (with opt. Y), Smin (Minimun Set) (with opt. R), Smax (Maximum Set) (with opt.
R), Pres (Presetting) (with opt. Y), CHor (Time counter) (with opt. U), Time (with opt. U), Date (with opt. U),
AP501v1.0 CYCLE (It shows at 5 seconds intervals the following messages: Flw, Tot+, Tot-, Smin, Smax, Pres,
CHor) ]

ENTER
• Press

9.6 -PROGRAMMING OF THE SECOND LINE OF LCD

• Insert the access keyword

• Press two times DISPLAY 1 [ LINE 2 FORMAT ]


• Act as the programming of the first line

9.7 -PROGRAMMING OF THE DIGITAL OUTPUT 1


(With opt. R)
• Insert the access keyword

D.O.
• Press 0 [DIGITAL OUT # 1]

• Press more times till it shows the desired message


[Dir flow pul (Direct Flow), Inv flow pul (Inverse Flow) (with opt. Y), Presetter (Count end pulse) (with opt. Y),
Empty Pipe Al. (Empty pipe alarm) (with opt. W), Min flow alarm (Minimum flow alarm), Max flow alarm (
Maximum flow alarm), Flow direction (with opt. P), Data failure]

ENTER
• Press

9.8 -PROGRAMMING OF THE DIGITAL OUTPUTS 2 , 3 , 4


(With opt. R)
• Insert the access keyword

D.O.
• Press two , three , four times 0 [DIGITAL OUT # 2 , 3 , 4 ]

• Act as the programming of the digital output 1.


Automation Progetti
TITLE Page 32

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.9 -PROGRAMMING OF THE LENGHT OF THE TOTALIZATION PULSES OF THE DIGITAL OUTPUT 1
(With opt. R)
• Insert the access keyword

D.O.
• Press 0 [ 5 ( At the end of the second line of LCD ) ] .
(5 is the lenght of pulses in msec)

• Press more times till it shows the desired lenght of pulses.


[ 5 , 10 , 15 , 20 , 25 , 35 , 50 , 75 , 100 , 150 , 200 , 250 , 350 , 500 , 750 , 999 (1sec) ]

ENTER
• Press
NOTES :
On both alarm and count end outputs this programming is no possible, because they will be on as long as alarm
persists or till the reset of presetter .
The lenght of pulse must be compatible with the kind of totalizer you want use and with the chosen totalization.
With opt. M lenght is 20msec.

9.10 -PROGRAMMING OF THE LENGHT OF THE TOTALIZATION PULSES OF THE DIGITAL OUTPUTS
2, 3, 4
(With opt. R)
• Insert the access keyword

D.O.
• Press two , three , four times 0 [ DIGITAL OUTPUT # 2 , 3 , 4 ]

• Act as the programming of the lenght of the totalization pulses of the digital output 1 .

9.11 -PROGRAMMING OF THE KIND OF ANALOG OUTPUT


(With opt. P)
• Insert the access keyword

A.O. .
• Press [ FLOW ]

A.O.
.
• Press more times till it shows the desired kind of output .
[ Direct Flow ( 0 ÷ +100% → 4 ÷ 20 mA ), Inverse Flow ( 0 ÷ -100% → 4 ÷ 20 mA ), Bidirect Flow (-
100÷0÷+100% → 4÷12÷20 mA ), Abolute Flow ( 0÷+100 % → 4÷20 mA ; 0÷-100% → 4÷20 mA )]

ENTER
• Press
NOTES :
Without the opt. P, the analog output is proportional to the direct flow.
Automation Progetti
TITLE Page 33

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.12 -PROGRAMMING OF THE FULL SCALE OF THE ANALOG OUTPUT
• Insert the access keyword

A.O. .
• Press [ FS ( Full Scale ) ]

• Press or till it shows the desired full scale .

ENTER
• Press
NOTE :
If display isn’t in percent , at every single press of key you haven’t a changing of the displayed full scale . So you
have to stop display just when it changes from the lower value to the desired one .

9.13 -PROGRAMMING OF THE FILTER TIME


• Insert the access keyword
FILTER

• Press 5 [ FILTERING ]

• Press more times or till it show the desired filter time [ 1 , 2 , 4 , 8 ]


(Converter gives the average of the taken measures in 1 , 2 , 4 , 8 seconds).

ENTER
• Press

9.14 -PROGRAMMING OF THE MINIMUM SET


(With opt. R)
• Insert the access keyword
MAX

MIN

• Press 2 [ MIN FLOW SET / Smin ]

• Press or till it shows the desired set.

ENTER
• Press

9.15 -PROGRAMMING OF THE MAXIMUM SET


(With opt. R)

• Insert the access keyword


MAX

MIN

• Press two times 2 [ MAX FLOW SET / Smax ]

• Act as the programming of the minimum set .


Automation Progetti
TITLE Page 34

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.16 -PROGRAMMING OF PRESETTER
(With opt. Y)

• Press BATCH 3 [ PRESET VALUE ]

• Press or till it shows the desired value.


[ 0 ÷ 30000 ]

ENTER
• Press

9.17 -PROGRAMMING OF THE KIND OF RESET IN PRESETTER


(With opt. Y)

• Press two times BATCH 3 [ RESET KIND ]

• Press or to choice the manual [ MAN ] or automatic [ AUT ] reset.

The manual mode allows to reset presetter through the key ;in automatic mode it resets itself after 1
second from the count end .

ENTER
• Press

9.18- MANUAL RESET OF PRESETTER


(With opt. Y)

• Press two times BATCH 3 [ RESET BY KEY ]

• Press

ENTER
• Press

9.19 -PROGRAMMING OF THE ACCESS KEYWORD


• Insert the current access keyword

F.N.
• Press DISPLAY 1

ENTER
• Press [ ACCESS KEYWORD ]
• Write the new numerical code. Its range is from 0 to 99·999·999.

ENTER
• Press
Automation Progetti
TITLE Page 35

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.20 -PROGRAMMING OF THE SPECIFIC WEIGHT
• Insert the access keyword
MAX

F.N. MIN

• Press 2

ENTER
• Press [ SPECIFIC WEIGHT ]
• Write the specific weight value .
Range must be included between 0.0000 e 1.9999 .
Value must always be written with four decimal figures .

ENTER
• Press

9.21 -PROGRAMMING OF TIME


(With opt. U)

• Insert the access keyword

F.N.
• Press BATCH 3

ENTER
• Press [ TIME ]
• Write the 6 figures of time .

ENTER
• Press to make clock start .

ENTER
• Press

9.22 -PROGRAMMING OF DATE


(With opt. U)

• Insert the access keyword

F.N. DN
• Press 4

ENTER
• Press [ DATE ]
• Write the 6 figures of date

ENTER
• Press two times
Automation Progetti
TITLE Page 36

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.23 -PROGRAMMING OF CUT-OFF
(With opt. K)

• Insert the access keyword

F.N. F
• Press 7

ENTER
• Press [ CUT-OFF SET /COFF ]

• Keep pressed or till it shows the desired value.


(Range : 1.5 ÷ 20% for converter with opt. 6 – 0.75 ÷ 20% for converter with opt. 12).

ENTER
• Press

9.24 -PROGRAMMING OF THE SERIAL ADDRESS


(With opt. 3 or 4)

• Insert the access keyword

F.N. D.O.
• Press DISPLAY 1 0

ENTER
• Press [ SERIAL ADDRESS ]

• Press more times or till it shows the desired value .


Range must be included between 1 and 255 .

ENTER
• Press

9.25 -PROGRAMMING OF THE SERIAL BAUD-RATE


(With opt. 3 or 4)

• Insert the access keyword

F.N.
• Press DISPLAY 1 DISPLAY 1

ENTER
• Press [ BAUD-RATE = ]

• Press more times or till it shows the desired value .


[ 300 , 600 , 1200 , 2400 , 4800 , 9600 , 19200 , 38400 ]

ENTER
• Press
Automation Progetti
TITLE Page 37

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.26 -ZERO ADJUSTMENT
• Insert the access keyword

F.N.
• Press DISPLAY 1 BATCH 3

ENTER
• Press [ MEASURE RESET / CONFIRM BY KEY ]

• Press [ WAIT ]
After about 15 seconds converter will show again the programmed visualization
NOTE:
Use this function only when you are sure to have piping full of still liquid.
If this condition is not observed, you will put a zero error.
Normally you have not to use this function because Automation Progetti makes this operation in factory. Anyway it
makes a perfect adjustment between flowmeter and converter.
Automation Progetti
TITLE Page 38

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54

9.27 -PROGRAMMING OF LANGUAGE


• Insert the access keyword

F.N. F
• Press DISPLAY 1 7

ENTER
• Press [ LANGUAGE SELECT./ENGLISH;
SELEZIONE LINGUA / ITALIANO ]

Press more times or to choice the desired language.

ENTER
• Press

9.28 -CONFIGURATION OF THE DIGITAL INPUT


(With opt. T)

• Insert the access keyword


• Press “FN””2””0”
• Press “ENTER” [DIG. INP. CONF.]
• Press more times “freccia su” or “freccia giu’” till it shows the desired configuration.
[Measure lock, Res. Part. Tot. (Reset partial totalizer) (with opt. G or Y), Presetter reload (with opt. Y)]
• Press “ENTER”

9.29 -DISPLAY OF THE ISTANTANEOUS FLOW

F
• Press 7 [ INSTANT FLOW / FLOW ]

ENTER
• Press

9.30 -DISPLAY OF THE ABSOLUTE TOTALIZER OF THE DIRECT FLOW


(With opt. H or Y)

• Press [ ABSOLUTE TOTAL . / Tot+ ]

ENTER
• Press
Automation Progetti
TITLE Page 39

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.31 -DISPLAY OF THE ABSOLUTE TOTALIZER OF THE INVERSE FLOW
(With opt. Y)

• Press two times [ ABSOLUTE TOTAL . / Tot- ]

ENTER
• Press

9.32 -DISPLAY OF THE RESETTABLE TOTALIZER OF THE DIRECT FLOW


(With opt. G or Y)

• Press three times [ RESETTABLE TOTAL / Trs+ ]

ENTER
• Press

9.33 -DISPLAY OF THE RESETTABLE TOTALIZER OF THE INVERSE FLOW


(With opt. Y)

• Press four times [ RESETTABLE TOTAL / Trs- ]

ENTER
• Press

9.34 -RESET OF THE RESETTABLE TOTALIZERS


(With opt. G or Y)

• Insert the access keyword

• Press five times [ TOTALIZ . RESET / CONFIRM BY KEY ]

• Press [ TOTALIZER RESET / Trs+ ]

ENTER
• Press

9.35 -DISPLAY OF THE TIME COUNTER


(With opt. U)

00:00

• Press 9 [ TIME COUNTER ]

ENTER
• Press
Automation Progetti
TITLE Page 40

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.36 -DISPLAY OF THE PROGRAMMED ND

DN
• Press 4 [ DIAMETER ND ]

ENTER
• Press

9.37 -DISPLAY OF THE FILTER TIME


FILTER

• Press 5 [ FILTERING ]

ENTER
• Press

9.38 -DISPLAY OF THE TECHNICAL UNIT OF THE ISTANTANEOUS FLOW

l Kg
• Press 6 [ MEASURE UNITS / FLOW ]

ENTER
• Press

9.39 -DISPLAY OF THE TECHNICAL UNIT OF TOTALIZERS


(With opt. G, H or Y)

l Kg
• Press two times 6 [ MEASURE UNITS / PULSES ]

ENTER
• Press

9.40 -DISPLAY OF THE MESSAGE WRITTEN ON THE FIRST LINE OF LCD

• Press DISPLAY 1 [ LINE 1 FORMAT ]

ENTER
• Press
Automation Progetti
TITLE Page 41

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
9.41 -DISPLAY OF THE MESSAGE WRITTEN ON THE SECOND LINE OF LCD

• Press two times DISPLAY 1 [ LINE 2 FORMAT ]

ENTER
• Press

9.42 -DISPLAY OF THE MINIMUM SET


(With opt. R)
MAX

MIN

• Press 2 [ MIN FLOW SET / Smin ]

ENTER
• Press

9.43 -DISPLAY OF THE MAXIMUM SET


(With opt. R)

MAX

MIN

• Press two times 2 [ MAX FLOW SET / S max ]

ENTER
• Press

9.44 -PRESETTER DISPLAY


(With opt. Y)

• Press BATCH 3 [ PRESET VALUE ]

ENTER
• Press

9.45 -DISPLAY OF THE KIND OF RESET


(With opt. Y)

• Press two times BATCH 3 [ KEY TO RESET / RESET KIND ]

ENTER
• Press
Automation Progetti
TITLE Page 42

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54

9.46 -DISPLAY OF THE DIGITAL OUTPUT 1 AND RESPECTIVE LENGHT OF THE TOTALIZATION
PULSES
(With opt. M or R)

D.O.
• Press 0 [ DIGITAL OUTPUT # 1 ]

ENTER
• Press

9.47 -DISPLAY OF THE DIGITAL OUTPUT 2, 3, 4, AND RESPECTIVE LENGHT


OF THE TOTALIZATION PULSES
(With opt. R)

D.O.
• Press two , three , four times 0 [ DIGITAL OUT # 2 , 3 , 4 ]

ENTER
• Press

9.48 -DISPLAY OF THE ANALOG OUTPUT

A.O. .
• Press [ FLOW ]

ENTER
• Press

9.49 -CONFIRM OF CONFIGURATION AFTER THE MESSAGE “ERROR IN DATA/EEPROM MEMORY”


• Verify the programmed parameters.
• Insert the access keyword

F.N. l Kg
• Press 6 [ CKS CALCULATION / CONFIRM BY KEY ]

• Press

ENTER
• Press
Automation Progetti
TITLE Page 43

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54

9.50 -DISPLAY OF THE VERSION OF SOFTWARE


• Insert the access keyword

F.N. l Kg
• Press DISPLAY 1 6 [ AP501 v1.0 XXX ]
( The number 1.0 after the letter “v” identifies the version of the software installed inside converter and the version
of the user’s manual )

9.51 -DISPLAY OF THE SERIAL NUMBER


• Insert the access keyword

F.N.
• Press DISPLAY 1

ENTER
• Press [SERIAL NUMBER]

ENTER
• Press

9.52 -ERROR MESSAGES


• ERROR IN DATA / EEPROM MEMORY (see F.Nr 6)
• WARNING! / WRONG CHECKSUM!
• NOT IN USE / EQUIPMENT
• ERROR IN DATA / USER PARAMETERS
• CHECK : ND , FILTER / L - Kg , DIG. OUT
• ERROR IN DATA / STARTING CONFIG
• ADC CONVERTER / NOT RELIABLE .

9.53 -SERVICE MESSAGES


• AUTOMATION / PROGETTI ( trade name of manufacturer)
• AP501 -v1.0 ( type and version of software )
• AUTO TEST ( starting check )
• WARNING! / AP550 ABSENCE ( it points out the absence of the matching card AP550 )
• WARNING! / WRONG CONNECTION (it points out a wrong connection between flowmeter and converter )
• WARNING! / EMPTY PIPE ( it points out flowmeter is empty )
• NOT QUALIFIED / OPERATOR (it points out the need of the access keyword)
Automation Progetti
TITLE Page 44

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
10 - ELECTRIC CONNECTIONS OF ELECTROMAGNETIC FLOWMETER AP1013 - AP1014
WITH CONVERTER AP501 WITHOUT OPT. 18 (WITHOUT CONNECTOR)
Automation Progetti
TITLE Page 45

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
11 - ELECTRIC CONNECTIONS OF ELECTROMAGNETIC FLOWMETER AP1013 - AP1014
WITH CONVERTER AP501 WITH OPT. 18 (WITH CONNECTOR)
Automation Progetti
TITLE Page 46

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
12 - ELECTRIC CONNECTIONS OF ELECTROMAGNETIC FLOWMETER AP1017 - AP1018
AP1016/DIN – AP1116 WITH CONVERTER AP501 WITHOUT OPT. 18
(WITHOUT CONNECTOR)
Automation Progetti
TITLE Page 47

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
13 - ELECTRIC CONNECTIONS OF ELECTROMAGNETIC FLOWMETER AP1017 - AP1018
AP1016/DIN – AP1116 WITH CONVERTER AP501 WITH OPT. 18
(WITH CONNECTOR)
Automation Progetti
TITLE Page 48

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
14 - ELECTRIC CONNECTIONS OF THE TWO-WIRE 4÷20mA OUTPUT
Automation Progetti
TITLE Page 49

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
15 - ELECTRIC CONNECTIONS OF THE ACTIVE 4÷20mA OUTPUT
Automation Progetti
TITLE Page 50

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
16 - ELECTRIC CONNECTIONS OF THE PROGRAMMABLE DIGITAL OUTPUTS
Automation Progetti
TITLE Page 51

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
17 - ELECTRIC CONNECTION OF THE RS232 OUTPUT
Automation Progetti
TITLE Page 52

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
18 - ELECTRIC CONNECTIONS DIGITAL INPUT
Automation Progetti
TITLE Page 53

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
19 - CONNECTION DIAGRAMM BETWEEN RS232 AND P.C.
Automation Progetti
TITLE Page 54

USER’S MANUAL
N.T.593E CONVERTER AP501 - V 1.0 – (23/03/2000) of pages 54
20 - DIMENSIONS AND WEIGHTS OF FLANGES