EZ7300 Series – ATP
1. Legal information
2. Analytical specifications
Please refer also to the respective technical datasheet at Hach Support Online.
3. Analysis method
Summary
ATP (Adenosine Triphosphate) is used by all organisms, from the simplest bacteria to
human being, as their primary energy currency. ATP transports chemical energy within cells
for metabolism. The EZ-ATP® on-line analyzer quantifies the ATP concentration in the
sample by counting the light units that are produced through reaction with luciferin and
luciferase. The higher the light output, the higher the ATP content and the higher the
bacterial load of the sample.
3
EZ7300 Series – ATP
The ATP measurement consists of two parts, a free and a total ATP measurement. In a
water sample, ATP can be located freely in the water mass, coming from disrupted cells.
This is called extracellular ATP or free ATP. Another possibility is the presence of ATP
inside bacteria and other cells. This is called intercellular ATP. In order to have a clear
picture of the biological load, it is important to differentiate between these two types of ATP.
The portion of intracellular ATP correlates with the living bacterial biomass and the portion of
free ATP can indicate previous cell growth. The total amount of ATP is the sum of free and
intracellular ATP.
Analysis steps
Calibration
The calibration procedure measures a REF1 solution (channel 9, REF1 valve) and a REF2
solution (channel 10, REF2 valve) to adapt the slope and offset factors by means of a two-
point calibration.
Cleaning
4
EZ7300 Series – ATP
4. Reagents
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Read the
safety data sheet from the supplier before bottles are filled or reagents are prepared. For
laboratory use only. Make the hazard information known in accordance with the local
regulations of the user.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
In the tables below, the products that are needed to prepare the reagents are listed. The
product name, the formula, the molecular weight, the CAS No. and the amount needed to
prepare 1 liter of the reagents is given. Check the consumption of the reagents (28 days) to
adapt the volumes needed.
5
EZ7300 Series – ATP
Quality of chemicals
All chemicals should be of ACS grade or better. We recommend the use of pro analysis
chemicals.
Quality of water
Reagent grade, de-ionized water must be used to prepare the chemical solutions and for
rinse purposes.
Storage of Reagents
While operating the instrument, keep in mind the ambient temperature conditions as stated
in the data sheet of the instrument.
Store the reagents cold; Store the reagents in the dark; Refresh the reagents after one
month (unless stated differently in the chapters below).
Storage
Store the detection reagent always in the foil pouch in which the reagent arrives to prevent
degradation due to light.
For long term storage, store the reagent at -20°C. Please note the expiration date(s) of the
packaging and use in appropriate order if there are different expiration dates to guarantee
optimal functionality.
Before use, store the reagent at 4°C and make sure that the reagent is completely thawed
before use in the analyser.
When new detection reagent is needed, transfer a new reagent bottle to the cooling unit
inside the analyzer and connect the pump tubing. The priming function can be used to flush
the pump tubing with fresh reagent. It is highly recommended to perform a cleaning cycle
after the priming function and before starting a calibration or analysis.
Storage
For long term storage, store the calibration kit at -20°C. Please note the expiration date(s) of
the packaging and use in appropriate order if there are different expiration dates to
guarantee optimal functionality.
Before use, store the reagent at 4°C and make sure that the reagent is completely thawed
before use in the analyser.
6
EZ7300 Series – ATP
US
Customers need to go directly in contact with:
Promega Corporation
2800 Woods Hollow Road
Madison, WI 53711 USA
COD@promega.com
Germany:
France:
Italy:
Netherlands:
Spain:
Sweden:
7
EZ7300 Series – ATP
Switzerland:
UK:
Preparation
Preparation
Take 83 mL of concentrated hydrochloric acid (HCl, 32%) and dilute to 1 litre with de-ionized
water.
8
EZ7300 Series – ATP
The cleaning procedure should prevent any build-up of biological growth in the flowcell and
tubing. The 3-step cleaning procedure includes a HCl flushing step, followed by a NaOH and
di-ionized water flush. Total cleaning time is 15 minutes.