33K6-4-3011-1
TECHNICAL MANUAL
CALIBRATION PROCEDURE
FOR
BALANCE/ANALYZER
8500C
(CHADWICK-HELMUTH)
Distribution Statement C - Distribution authorized to U. S. Government agencies and their contractors for
official use or for administrative or operational purposes only, 30 December 1996. Requests for this
document shall be referred to 562 CBSG/GBHA, 813 Irving-Wick Dr W, Heath, OH 43056-1199.
Destruction Notice - For unclassified, limited documents, destroy by any method that will prevent
disclosure of the contents or reconstruction of the document.
30 JULY 2004
CHANGE 1 - 30 APRIL 2007
T.O. 33K6-4-3011-1
INSERT LATEST CHANGED PAGES. DESTROY SUPERSEDED PAGES.
LIST OF EFFECTIVE PAGES NOTE: The portion of the text and illustrations affected by the changes is indicated by a vertical
line in the outer margins of the page.
Title ...................................1
A ........................................1
1 - 12..................................0
13 - 14................................1
15 - 20................................0
21 .......................................1
22 Blank.............................0
A-1.....................................0
A-2 Blank ..........................0
A Change 1 USAF
T.O. 33K6-4-3011-1
BALANCER/ANALYZER
8500C
(CHADWICH-HELMUTH)
1 CALIBRATION DESCRIPTION:
Table 1.
Balance Measurement
Frequency Range: 3.3 Hz to 10 kHz Apply a known Signal as
(200 to 600,000 rpm) monitored on a Dynamic
Signal Analyzer and monitor
Accuracy: ±0.1 Hz TI display
Anti-alias Filtering
500 Hz Range: 0.2 Hz to 10 kHz
Spectrum
High Frequency Range: 149 Hz to 10 kHz
Accuracy: ±0.2 dB
1
T.O. 33K6-4-3011-1
Table 1. (Cont.)
Vibration Meter
(9660-1 and 9660-2)
2 EQUIPMENT REQUIREMENTS:
Range: Squarewave,
140 mV rms, 194.9 Hz
Accuracy:
Frequency, ±0.15% of rdg;
Voltage, N/A
2
T.O. 33K6-4-3011-1
3 PRELIMINARY OPERATIONS:
3.1 Review and become familiar with entire procedure before beginning Calibration Process.
WARNING
3.2 Connect test equipment to appropriate power source. Set all POWER switches to On and allow warm-up as
required by manufacturer.
3.3 Throughout this procedure set the TI functions as follows: Press TI right or left arrow key to move cursor on
display, press CHANGE key to set TI for selected function and then press ENTER key.
3
T.O. 33K6-4-3011-1
3.5 Perform only that portion of the Calibration Process applicable to TI being calibrated.
3.6 Make copies of Table 3 to use as worksheets.
4 CALIBRATION PROCESS:
NOTE
Unless otherwise specified, verify the results of each test and take corrective
action whenever the test requirement is not met, before proceeding.
CAUTION
Before hooking up the Input Adapter Cable P/N 8998, verify that its polarity is
A=LOW(-) and B=HI(+). Some P/N 8998 cables may be reversed. If hooked up
backwards, and the 28 VDC Signal is applied, the screen on the TI will go dim.
The technician will have about 5 seconds to disconnect the power before damage
is done to the TI. Switch polarity for correct operation and apply power.
4.1.1 Connect equipment as shown in Figure 1.
DUAL DC
POWER SUPPLY
TI 28 VDC
SIGNAL
- +
A=LOW (-)
B=HIGH (+)
MAG PICKUP
INPUT (4 PIN)
SYNTHESIZED FUNCTION GENERATOR
D=HIGH (+) VELOCIMETER INPUT
B=LOW (-) (6 PIN)
SIGNAL SYNC
OUT OUT
D=HIGH (+)
B=LOW (-)
FEEDTHROUGH
TERMINATION
INPUT
Figure 1.
4.1.2 On the TI, hold down the START key and then set the Dual DC Power Supply for 28 VDC. Wait until Power-Up
Self-Test display appears and then release START key. This will erase all memory. The TI will automatically run a
Power-Up Self-Test when power is applied to the TI.
4.1.3 Set the Dynamic Signal Analyzer as follows:
FORMAT SINGLE
4
T.O. 33K6-4-3011-1
MKR MKR ON
FUNCTION SINEWAVE
FREQUENCY 100 Hz
4.1.7 Press TI ENTER key and wait 10 seconds for the TI gain to settle.
4.1.9 Verify the TI indicates within the values listed in the Limits column of Table 2.
Table 2.
4.2.1 Disconnect Velocimeter and Magnetic Pickup cables from Input Adapter Cable.
5
T.O. 33K6-4-3011-1
SEEK <OFF>
4.2.6 On the TI display scale, each dot on the display is equal to 0.001 ips. If any noise signals appear, use the
left-right arrows to move cursor (dotted line) to noise peak. The ips level and frequency will readout at top of display.
4.2.7 Verify any Noise or Spurious Signal noise present on the TI display is ≤0.005 ips.
4.3.1.2 Set the Dynamic Signal Analyzer CENTER FREQUENCY to 1,722 Hz.
4.3.1.3 Set the Synthesized Function Generator FREQUENCY to 1,722 Hz and AMPLITUDE for 13.44 mV rms
(sinewave) as verified on the Dynamic Signal Analyzer.
4.3.1.8 Verify any Signal Aliasing present on the TI display is ≤0.005 ips.
4.3.2.2 Set Synthesized Function Generator FREQUENCY to 15,600 Hz and AMPLITUDE for 13.44 mV rms
(sinewave) as verified on the Dynamic Signal Analyzer.
4.3.2.7 Verify any Signal Aliasing present on the TI display is ≤0.005 ips.
6
T.O. 33K6-4-3011-1
4.3.2.8 Set Synthesized Function Generator and Dual DC Power Supply outputs to minimum and disconnect test setup.
DUAL DC
POWER SUPPLY TI
28 VDC
SIGNAL
- + A=HIGH (+)
B=LOW (-)
DC POWER (CONNECT
DIRECTLY TO TI)
OUTPUT/POWER SIGNAL
VELOCIMETER VIBRATION CHANNELS TO TI SELECTOR CABLE
CABLE
1 4 7 10 SYNTHESIZED FUNCTION GENERATOR
2 5 8 11
3 6 9 12
SIGNAL SYNC
MAGNETIC PICKUP/
PHOTOCELL CHANNELS
OUT OUT
A B C D
SIGNAL SELECTOR
INPUT
FEEDTHROUGH
TERMINATION
Figure 2.
4.4.1.2 Set Dynamic Signal Analyzer CENTER FREQUENCY to 195 Hz and DEFINE FULL SCL (vert scale) to
150 mV rms signal.
4.4.1.3 Set the Synthesized Function Generator FREQUENCY to 194.9 Hz and AMPLITUDE for 134.4 mV rms
(squarewave) as verified on the Dynamic Signal Analyzer.
4.4.1.4 On the TI, hold down the START key and then set the Dual DC Power Supply for 28 VDC. Wait until
Power-Up Self-Test display appears and then release START key. This will erase all memory.
7
T.O. 33K6-4-3011-1
SEEK <OFF>
4.4.1.9 The TI will display a range of odd harmonic peaks from the 1st (fundamental) to the 51st. For each odd
harmonic, record the peak of the fundamental and the 2 peaks to the left and the 2 peaks to the right of the fundamental.
4.4.1.10 Use the TI arrow keys to move the cursor to peak of the 1st Harmonic (fundamental).
4.4.1.11 Ensure the TI display frequency readout indicates the frequency in Column C of Table 3.
NOTE
The TI must indicate the exact frequency as listed in Column C of Table 3. The
TI must indicate the frequency as listed in column A, B, D, E ±0.1 Hz.
4.4.1.12 Record the TI Amplitude Readout (in ips) at the top of the display in column C of the copy of Table 3.
4.4.1.13 Press the TI left arrow key once to move the cursor to the first peak left of the fundamental. Record the
amplitude (in ips) in column B of the copy of Table 3.
4.4.1.14 Press the TI left arrow key once to move the cursor to the second peak left of the fundamental. Record the
amplitude (in ips) in column A of the copy of Table 3.
4.4.1.15 Press the TI right arrow key to move the cursor to the first peak to the right of the fundamental. Record the
amplitude (in ips) in column D of the copy of Table 3.
4.4.1.16 Press the TI right arrow key to move the cursor to the second peak to the right of the fundamental. Record the
amplitude (in ips) in column E of the copy of Table 3.
4.4.1.17 Repeat steps 4.4.1.8 through 4.4.1.16 for the remaining values listed in the Harmonic Peak column of Table 3.
4.4.1.18 Calculate and record the True Amplitude of each harmonic peak in the copy of Table 3 using the following
formula:
4.4.1.19 Verify the True Amplitude of each Table 3 harmonic peak is within the values listed in the True Amp ips
column of Table 3.
8
T.O. 33K6-4-3011-1
Table 3.
NOTE
A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips
9
T.O. 33K6-4-3011-1
Table 3. (Cont.)
NOTE
A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips
10
T.O. 33K6-4-3011-1
Table 3. (Cont.)
NOTE
A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips
4.4.2.1 Set the Dynamic Signal Analyzer CENTER FREQUENCY to 490 Hz.
4.4.2.2 Set the Synthesized Function Generator FREQUENCY to 490 Hz and AMPLITUDE for 134.4 mV rms
(squarewave) as verified by the Dynamic Signal Analyzer.
NOTE
You cannot proceed until the Analog Filter test passes. Occasionally the
Analog Filter test may fail with nothing permanently wrong with the TI. If
it does fail, press OK key and then EXECUTE key to run again. If it
continues to fail, refer to the commercial data.
4.4.2.9 After the TI display indicates Analog filter test successful, press OK key. The TI Self-Test menu will return.
11
T.O. 33K6-4-3011-1
NOTE
4.4.2.12 On TI, Select <Hz> using arrow keys and press CHANGE key to change <Hz> to <RPM>.
SEEK ON
4.4.2.16 Verify the 134.4 mV rms (squarewave) output of the Synthesized Function Generator on the Dynamic Signal
Analyzer.
4.4.2.17 Use the TI arrow keys to move the cursor to the fundamental peak. Verify the TI readout is between 9.5 and
10.4 ips (±0.2 dB).
4.4.2.18 Repeat steps 4.4.2.1, 4.4.2.2 and 4.4.2.15 through 4.4.2.17 for the following TI frequencies: 406, 308, 210,
154 and 98 Hz.
4.4.2.20 Set TI MAXIMUM FREQUENCY to 16596 <RPM>, and press TI ENTER key and wait 10 seconds.
4.4.2.21 Repeat steps 4.4.2.1, 4.4.2.2 and 4.4.2.15 through 4.4.2.17 for the following TI frequencies: 51, 21, 9, 6 and 3 Hz.
4.4.2.22 Set the Synthesized Function Generator and Dual DC Power Supply outputs to minimum and disconnect test
setup.
4.5.1 Connect Synthesized Function Generator (#1) through Feedthrough Termination (#1) to the Digital Multimeter HI
and LO INPUT terminals.
4.5.3 Set Synthesized Function Generator (#1) FREQUENCY to 100 Hz and AMPLITUDE for 100 mV rms
(sinewave) signal. Adjust Synthesized Function Generator (#1) for a Digital Multimeter indication of 100.00 mV rms.
4.5.4 Disconnect Synthesized Function Generator ( #1) and Feedthrough Termination (#1) from the Digital Multimeter
INPUT terminals.
12
T.O. 33K6-4-3011-1
4.5.5 Connect Synthesized Function Generator (#2) through Feedthrough Termination (#2) to the Digital Multimeter
INPUT terminals, observing polarity.
4.5.6 Set Synthesized Function Generator (#2) FREQUENCY to 100 Hz and AMPLITUDE for 100 mV rms
(sinewave) signal. Adjust Synthesized Function Generator (#2) for a Digital Multimeter indication of 100.00 mV rms.
4.5.7 Disconnect Synthesized Function Generator (#2) and Feedthrough Termination (#2) from the Digital Multimeter
INPUT terminals.
4.5.8 Connect RLC Digibridge to TI Charge Amplifier A and B Input terminals. Set the RLC Digibridge to measure the
input capacitance on the Charge Amplifier A and B and record the value as Cin.
4.5.10 Remove Charge Amplifier cover and set DIP switches for the sensitivity closest to that of the intended
accelerometer. Replace Charge Amplifier cover after setting DIP switches.
A C
OUTPUT/POWER
DIGITAL MULTIMETER B
VIBRATION CHANNELS TO TI
CHARGE AMPLIFIER 1 4 7 10
(INPUT SIDE) 2 5 8 11
3 6 9 12
SIGNAL SELECTOR CABLE
VIEW A MAGNETIC PICKUP/
PHOTOCELL CHANNELS
A B C D
SIGNAL SELECTOR
TO DIGITAL A C
MULTIMETER
B
CHARGE AMPLIFIER
(INPUT SIDE)
VIEW A
Figure 3.
Change 1 13
T.O. 33K6-4-3011-1
4.5.12 Fine tune Synthesized Function Generator (#1) AMPLITUDE for 100.00 mV rms as indicated on Digital
Multimeter.
⎛
⎜ 61.45 * (Cin + Cacc ) ⎞⎟
ips = 2 * Vin ⎜ ⎟
⎜ f * SQ ⎟
⎝ ⎠
f = 100 Hz
Example:
4.5.18 Set Synthisized Function Generator outputs to minimum and disconnect test setup.
4.6.1.2 Set Dual DC Power Supply output as needed for -15 VDC at TI pins A - B and +15 VDC at TI pins C - B as
shown in Figure 4.
14 Change 1
T.O. 33K6-4-3011-1
4.6.1.4 Set TI FILTER switch to the first position listed in the Filter column of Table 4.
METER CALIBRATOR
CH 1 CH 2 CH 3 CH 4
MILS D.A.
CHANNEL
2 3
1 4
DUAL DC
FILTER TEST
POWER SUPPLY IN
OUT
TI TI cable and
+ - + -
connector.
A C
Shorting Bar B
Figure 4.
4.6.1.5 Set Meter Calibrator to the first value listed in the Applied column of Table 4.
4.6.1.7 The TI display must indicate within the corresponding values listed in the Limits column of Table 4.
4.6.1.9 Repeat steps 4.6.1.4 through 4.6.1.8 for each remaining value listed in Table 4.
4.6.1.10 Repeat steps 4.6.1.1 and 4.6.1.3 through 4.6.1.9 for TI CHANNEL 2, CHANNEL 3 and CHANNEL 4
INPUT.
Table 4.
15
T.O. 33K6-4-3011-1
Table 4. (Cont.)
4.6.2.1 Connect TI CH 4 input connector pin D (HI) and pin B (LO) to Dynamic Signal Analyzer INPUT connector,
observing polarity.
4.6.2.2 Set Dual DC Power Supply as needed for -15 VDC at TI pins A - B and +15 VDC at TI pins C - B as shown in
Figure 4.
4.6.2.4 From step 4.6.2.5 through step 4.6.2.14 press TI TEST pushbutton as needed to apply TI TEST signal to
Dynamic Signal Analyzer INPUT connector.
4.6.2.5 Ensure the Dynamic Signal Analyzer MARKER is at the peak of the TI 64 Hz signal and record the
MARKER indication.
4.6.2.6 Disconnect TI CH 4 input from Dynamic Signal Analyzer INPUT connector and connect Meter Calibrator
OUTPUT terminals to Dynamic Signal Analyzer Input connector, observing polarity.
4.6.2.7 Set Meter Calibrator for 27.4 mV rms/64 Hz and set OPR/STBY switch to OPR.
4.6.2.8 Edit Meter Calibrator display as necessary for a Dynamic Signal Analyzer MARKER indication equal to the
value recorded in step 4.6.2.5.
4.6.2.9 The Meter Calibrator must indicate between 27.126 to 27.674 mV rms.
4.6.2.11 Disconnect Meter Calibrator OUTPUT terminals from Dynamic Signal Analyzer INPUT connector.
4.6.2.12 Connect TI CH 4 input connector pin D (HI) and pin B (LO) to Dynamic Signal Analyzer INPUT
connector,observing polarity.
16
T.O. 33K6-4-3011-1
4.6.2.14 Ensure the Dynamic Signal Analyzer MARKER is at the peak of the TI 256 Hz signal and record the
MARKER indication.
4.6.2.15 Disconnect TI CH 4 Input from Dynamic Signal Analyzer INPUT connector and connect Meter Calibrator
Output terminals to Dynamic Signal Analyzer INPUT connector, observing polarity.
4.6.2.16 Set Meter Calibrator for 27 mV rms/256 Hz and set OPR/STBY switch to OPR.
4.6.2.17 Edit Meter Calibrator display as necessary for a Dynamic Signal Analyzer MARKER indication equal to the
value recorded in step 4.6.2.14.
4.6.2.18 The Meter Calibrator must indicate between 27.126 to 27.674 mV rms.
4.6.2.19 Set all POWER switches to OFF/STBY, disconnect and secure all equipment.
Applied Limits
Applied Limits
17
T.O. 33K6-4-3011-1
High Frequency:
Note
A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips
18
T.O. 33K6-4-3011-1
High Frequency:
Note
A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips
19
T.O. 33K6-4-3011-1
High Frequency:
Note
A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips
Low Frequency:
20
T.O. 33K6-4-3011-1
Range Limits
Frequency Response:
APPENDIX A
SHIELD
D
B
A NC
SHIELD
C NC
POWER/SIGNAL
BLK
INPUTS C A +V 12-15 VDC
GRN
R B GND
P
RED
C ±28 VDC MAG/PHOTOCELL
WHT
D D SIGNAL
E
BLK
J A -V
GRN
S B GND VELOCIMETER
RED
T C +V
WHT
H BRAID
D SIGNAL
G N/C E
N/C F
B N/C
K P5-1 FROM DIGITAL BOARD
U P5-3 FROM DIGITAL BOARD
F N/C
A-1/(A-2 Blank)