You are on page 1of 25

T.O.

33K6-4-3011-1

TECHNICAL MANUAL

CALIBRATION PROCEDURE

FOR

BALANCE/ANALYZER

8500C
(CHADWICK-HELMUTH)

Distribution Statement C - Distribution authorized to U. S. Government agencies and their contractors for
official use or for administrative or operational purposes only, 30 December 1996. Requests for this
document shall be referred to 562 CBSG/GBHA, 813 Irving-Wick Dr W, Heath, OH 43056-1199.

Destruction Notice - For unclassified, limited documents, destroy by any method that will prevent
disclosure of the contents or reconstruction of the document.

Published under Authority of the Secretary of the Air Force

30 JULY 2004
CHANGE 1 - 30 APRIL 2007
T.O. 33K6-4-3011-1
INSERT LATEST CHANGED PAGES. DESTROY SUPERSEDED PAGES.
LIST OF EFFECTIVE PAGES NOTE: The portion of the text and illustrations affected by the changes is indicated by a vertical
line in the outer margins of the page.

Date of issue for original and changed pages are:

Original..................0 ............................... 30 July 2004


Change ...................1 ..............................30 April 2007

TOTAL NUMBER OF PAGES IN THIS PUBLICATION IS 26, CONSISTING OF THE FOLLOWING:

Page * Change Page * Change Page * Change


No. No. No. No. No. No.

Title ...................................1
A ........................................1
1 - 12..................................0
13 - 14................................1
15 - 20................................0
21 .......................................1
22 Blank.............................0
A-1.....................................0
A-2 Blank ..........................0

*Zero in this column indicates an original page

A Change 1 USAF
T.O. 33K6-4-3011-1

BALANCER/ANALYZER

8500C

(CHADWICH-HELMUTH)

1 CALIBRATION DESCRIPTION:

Table 1.

Test Instrument (TI) Performance Test


Characteristics Specifications Method

Balance Measurement
Frequency Range: 3.3 Hz to 10 kHz Apply a known Signal as
(200 to 600,000 rpm) monitored on a Dynamic
Signal Analyzer and monitor
Accuracy: ±0.1 Hz TI display

Phase Range: 0 to 360°

Accuracy: ±2.0° (electrical)

Amplitude Range: 0.01 to 10.0 ips

Accuracy: ±3% of indication

Noise and Range: minimum


Spurious Signals
Accuracy: ≤0.005 ips

Anti-alias Filtering
500 Hz Range: 0.2 Hz to 10 kHz

Accuracy: Flatness, ≤0.005 ips, 0.2 to 500 Hz

10 kHz Range: 0.2 Hz to 10 kHz

Accuracy: Flatness, ≤0.005 ips, 0.5 to 10 kHz

Spectrum
High Frequency Range: 149 Hz to 10 kHz

Accuracy: as listed in Table 3

Low Frequency Range: 3 to 490 Hz

Accuracy: ±0.2 dB

Charge Amplifier Range: 100 ips @ 100 Hz

Accuracy: ±1% of ips rdg

1
T.O. 33K6-4-3011-1

Table 1. (Cont.)

Test Instrument (TI) Performance Test


Characteristics Specifications Method

Vibration Meter
(9660-1 and 9660-2)

Frequency Response Range: 8.5 Hz to 5 kHz Compared to a Meter


Calibrator
Accuracy: Filter IN, 150 to 1000 Hz at
-3 dB points; Filter OUT, 8.5 to 5000 Hz at
-3 dB points

Internal Calibration Source Range: Compared to a Meter


256 Hz tone, 2.5 mil (2.04 ips, 27.4 mV rms); Calibrator signal with the
64 Hz tone, 10 mil (2.04 ips, 27.4 mV rms) Dynamic Signal Analyzer

Accuracy: Amplitude, ±1% of setting

2 EQUIPMENT REQUIREMENTS:

Minimum Use Calibration Sub-


Noun Specifications Equipment Item

2.1 SYNTHESIZED Range: Sinewave, Hewlett-Packard


FUNCTION 3 to 15,600 Hz, 10 to 140 mV rms 3325B
GENERATOR (2 EA)
Accuracy:
Frequency, ±0.025% of rdg;
Voltage, N/A

Range: Squarewave,
140 mV rms, 194.9 Hz

Accuracy:
Frequency, ±0.15% of rdg;
Voltage, N/A

2.2 RLC DIGIBRIDGE Range: 0.1 mF Genrad


1689M
Accuracy: ±0.15% of rdg

2.3 DIGITAL Range: 100 mV rms, 100 Hz Fluke


MULTIMETER 8840A/AF
Accuracy: ±0.15% of rdg

2.4 DYNAMIC SIGNAL Range: 3 to 15,600 Hz Hewlett-Packard


ANALYZER 3561A
(P/O PHASE NOISE Accuracy: ±0.2 dBV
MEASUREMENT
SYSTEM)

2.5 FEEDTHROUGH Range: 50 Ω Hewlett-Packard


TERMINATION 11048C
(2 EA) Accuracy: ±0.1% of nominal

2
T.O. 33K6-4-3011-1

Minimum Use Calibration Sub-


Noun Specifications Equipment Item

2.6 DUAL DC POWER Range: Single output 28 VDC; As Available


SUPPLY Dual Output, -15 and +15 VDC;
5A

Accuracy: ±0.5 VDC

2.7 INPUT ADAPTER Range: N/A Chadwick-Helmuth


CABLE 8998
Accuracy: N/A

2.8 VELOCIMETER Range: N/A Chadwick-Helmuth


CABLE 8278
Accuracy: N/A

2.9 MAGNETIC PICKUP Range: N/A Chadwick-Helmuth


CABLE 10502
Accuracy: N/A

2.10 SIGNAL SELECTOR Range: N/A Chadwick-Helmuth


CABLE 10811
Accuracy: N/A

2.11 SIGNAL SELECTOR Range: N/A Chadwick-Helmuth Chadwick-Helmuth


8520C 8520C-36
Accuracy: N/A

2.12 METER Range: 27.126 to 27.674 VAC, Fluke


CALIBRATOR 64 to 256 Hz 5700A

Accuracy: VAC, ±0.25% of setting;


Frequency, N/A

3 PRELIMINARY OPERATIONS:

3.1 Review and become familiar with entire procedure before beginning Calibration Process.

WARNING

Unless otherwise designated, and prior to beginning the Calibration Process,


ensure that all test equipment voltage and/or current outputs are set to zero (0)
or turned off, where applicable. Ensure that all equipment switches are set to
the proper position before making connections or applying power.

3.2 Connect test equipment to appropriate power source. Set all POWER switches to On and allow warm-up as
required by manufacturer.

3.3 Throughout this procedure set the TI functions as follows: Press TI right or left arrow key to move cursor on
display, press CHANGE key to set TI for selected function and then press ENTER key.

3.4 Refer to Appendix A for construction of Interconnect Cables if required.

3
T.O. 33K6-4-3011-1

3.5 Perform only that portion of the Calibration Process applicable to TI being calibrated.
3.6 Make copies of Table 3 to use as worksheets.

4 CALIBRATION PROCESS:

NOTE
Unless otherwise specified, verify the results of each test and take corrective
action whenever the test requirement is not met, before proceeding.

4.1 BALANCE MEASUREMENT CALIBRATION:

CAUTION

Before hooking up the Input Adapter Cable P/N 8998, verify that its polarity is
A=LOW(-) and B=HI(+). Some P/N 8998 cables may be reversed. If hooked up
backwards, and the 28 VDC Signal is applied, the screen on the TI will go dim.
The technician will have about 5 seconds to disconnect the power before damage
is done to the TI. Switch polarity for correct operation and apply power.
4.1.1 Connect equipment as shown in Figure 1.
DUAL DC
POWER SUPPLY
TI 28 VDC
SIGNAL

- +

A=LOW (-)
B=HIGH (+)

INPUT ADAPTER CABLE

MAG PICKUP
INPUT (4 PIN)
SYNTHESIZED FUNCTION GENERATOR
D=HIGH (+) VELOCIMETER INPUT
B=LOW (-) (6 PIN)
SIGNAL SYNC
OUT OUT
D=HIGH (+)
B=LOW (-)

MAGNETIC PICKUP CABLE


VELOCIMETER CABLE

DYNAMIC SIGNAL ANALYZER COAX CABLE

FEEDTHROUGH
TERMINATION
INPUT

Figure 1.
4.1.2 On the TI, hold down the START key and then set the Dual DC Power Supply for 28 VDC. Wait until Power-Up
Self-Test display appears and then release START key. This will erase all memory. The TI will automatically run a
Power-Up Self-Test when power is applied to the TI.
4.1.3 Set the Dynamic Signal Analyzer as follows:
FORMAT SINGLE

VERT SCALE LINEAR

4
T.O. 33K6-4-3011-1

VERT SCALE DEFINE FULL SCALE (SCL), 20 mV rms

WINDOW FLAT TOP (HI ACC)

FREQ DEFINE SPAN, 100 Hz

FREQ DEFINE CENTER, 100 Hz

MKR MKR ON

MKR PEAK TRK ON

4.1.4 Set the Synthesized Function Generator as follows:

FUNCTION SINEWAVE

FREQUENCY 100 Hz

AMPLITUDE 13.44 mV rms (sinewave) (as indicated


on the Dynamic Signal Analyzer)

4.1.5 Press TI BALANCE key.

4.1.6 Press TI SETUP key and set the TI as follows:

BALANCE FREQUENCY 100.0 <Hz> X 1

BALANCE TYPE <MAG PICKUP> <MAN STOP>

ANGLE UNITS <DEGREES>

4.1.7 Press TI ENTER key and wait 10 seconds for the TI gain to settle.

4.1.8 Press TI START key.

4.1.9 Verify the TI indicates within the values listed in the Limits column of Table 2.

Table 2.

Function Nominal Limits

Frequency 100 Hz 99.9 to 100.1 Hz

VIB Level 1.000 ips 0.960 to 1.040 ips

Phase Range 180° 177 to 183°

4.1.10 Press TI STOP key.

4.1.11 Set the Synthesized Function Generator output to minimum.

4.2 NOISE AND SPURIOUS SIGNALS CALIBRATION:

4.2.1 Disconnect Velocimeter and Magnetic Pickup cables from Input Adapter Cable.

4.2.2 Press TI SPECTRUM key.

5
T.O. 33K6-4-3011-1

4.2.3 Press TI SETUP key and set controls as follows:

MAXIMUM FREQUENCY 1,000 <Hz>

SEEK AMPLITUDE 0.010 ips

REPETITION MODE <SINGLE>

SEEK <OFF>

4.2.4 Press TI ENTER key and wait 10 seconds.

4.2.5 Press TI START key.

4.2.6 On the TI display scale, each dot on the display is equal to 0.001 ips. If any noise signals appear, use the
left-right arrows to move cursor (dotted line) to noise peak. The ips level and frequency will readout at top of display.

4.2.7 Verify any Noise or Spurious Signal noise present on the TI display is ≤0.005 ips.

4.3 ANTI-ALIAS FILTERING CALIBRATION:

4.3.1 500 Hz FILTER RESPONSE:

4.3.1.1 Connect the Velocimeter Cable to Input Adapter Cable.

4.3.1.2 Set the Dynamic Signal Analyzer CENTER FREQUENCY to 1,722 Hz.

4.3.1.3 Set the Synthesized Function Generator FREQUENCY to 1,722 Hz and AMPLITUDE for 13.44 mV rms
(sinewave) as verified on the Dynamic Signal Analyzer.

4.3.1.4 Press TI SETUP key.

4.3.1.5 Set TI MAXIMUM FREQUENCY to 500 Hz.

4.3.1.6 Press TI ENTER key and wait 10 seconds.

4.3.1.7 Press TI START key.

4.3.1.8 Verify any Signal Aliasing present on the TI display is ≤0.005 ips.

4.3.2 10 kHz FILTER RESPONSE:

4.3.2.1 Set Dynamic Signal Analyzer CENTER FREQUENCY to 15,600 Hz.

4.3.2.2 Set Synthesized Function Generator FREQUENCY to 15,600 Hz and AMPLITUDE for 13.44 mV rms
(sinewave) as verified on the Dynamic Signal Analyzer.

4.3.2.3 Press TI SETUP key.

4.3.2.4 Set TI MAXIMUM FREQUENCY to 1,000 Hz.

4.3.2.5 Press TI ENTER key and wait 10 seconds.

4.3.2.6 Press TI START key.

4.3.2.7 Verify any Signal Aliasing present on the TI display is ≤0.005 ips.

6
T.O. 33K6-4-3011-1

4.3.2.8 Set Synthesized Function Generator and Dual DC Power Supply outputs to minimum and disconnect test setup.

4.4 SPECTRUM CALIBRATION:

4.4.1 HIGH FREQUENCY:

4.4.1.1 Connect equipment as shown in Figure 2.

DUAL DC
POWER SUPPLY TI
28 VDC
SIGNAL

- + A=HIGH (+)
B=LOW (-)

DC POWER (CONNECT
DIRECTLY TO TI)
OUTPUT/POWER SIGNAL
VELOCIMETER VIBRATION CHANNELS TO TI SELECTOR CABLE
CABLE
1 4 7 10 SYNTHESIZED FUNCTION GENERATOR
2 5 8 11
3 6 9 12
SIGNAL SYNC
MAGNETIC PICKUP/
PHOTOCELL CHANNELS
OUT OUT
A B C D

SIGNAL SELECTOR

DYNAMIC SIGNAL ANALYZER

INPUT

FEEDTHROUGH
TERMINATION

Figure 2.

4.4.1.2 Set Dynamic Signal Analyzer CENTER FREQUENCY to 195 Hz and DEFINE FULL SCL (vert scale) to
150 mV rms signal.

4.4.1.3 Set the Synthesized Function Generator FREQUENCY to 194.9 Hz and AMPLITUDE for 134.4 mV rms
(squarewave) as verified on the Dynamic Signal Analyzer.

4.4.1.4 On the TI, hold down the START key and then set the Dual DC Power Supply for 28 VDC. Wait until
Power-Up Self-Test display appears and then release START key. This will erase all memory.

4.4.1.5 Press TI SPECTRUM key.

4.4.1.6 Press TI SETUP key and set controls as follows:

MAXIMUM FREQUENCY 9989.8 <Hz>

SEEK AMPLITUDE 0.010 ips

REPETITION MODE <SINGLE>

7
T.O. 33K6-4-3011-1

INPUT CHANNEL 1<ONLY>

SEEK <OFF>

4.4.1.7 Press TI ENTER key and wait 10 seconds.

4.4.1.8 Press TI START key.

4.4.1.9 The TI will display a range of odd harmonic peaks from the 1st (fundamental) to the 51st. For each odd
harmonic, record the peak of the fundamental and the 2 peaks to the left and the 2 peaks to the right of the fundamental.

4.4.1.10 Use the TI arrow keys to move the cursor to peak of the 1st Harmonic (fundamental).

4.4.1.11 Ensure the TI display frequency readout indicates the frequency in Column C of Table 3.

NOTE

The TI must indicate the exact frequency as listed in Column C of Table 3. The
TI must indicate the frequency as listed in column A, B, D, E ±0.1 Hz.

4.4.1.12 Record the TI Amplitude Readout (in ips) at the top of the display in column C of the copy of Table 3.

4.4.1.13 Press the TI left arrow key once to move the cursor to the first peak left of the fundamental. Record the
amplitude (in ips) in column B of the copy of Table 3.

4.4.1.14 Press the TI left arrow key once to move the cursor to the second peak left of the fundamental. Record the
amplitude (in ips) in column A of the copy of Table 3.

4.4.1.15 Press the TI right arrow key to move the cursor to the first peak to the right of the fundamental. Record the
amplitude (in ips) in column D of the copy of Table 3.

4.4.1.16 Press the TI right arrow key to move the cursor to the second peak to the right of the fundamental. Record the
amplitude (in ips) in column E of the copy of Table 3.

4.4.1.17 Repeat steps 4.4.1.8 through 4.4.1.16 for the remaining values listed in the Harmonic Peak column of Table 3.

4.4.1.18 Calculate and record the True Amplitude of each harmonic peak in the copy of Table 3 using the following
formula:

True Amp = 0.7983431 × ( Amp A) 2 + ( Amp B) 2 + ( Amp C) 2 + ( Amp D) 2 + ( Amp E ) 2

4.4.1.19 Verify the True Amplitude of each Table 3 harmonic peak is within the values listed in the True Amp ips
column of Table 3.

8
T.O. 33K6-4-3011-1

Table 3.

NOTE

These frequencies are based on a 194.9 Hz squarewave input.

A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips

149.5 Hz l74.4 Hz 199.3 Hz 224.2 Hz 249.1 Hz Max: 11.411


1st ips ips ips ips ips Actual:
Min: 8.761

523.2 Hz 548.1 Hz 573.0 Hz 597.9 Hz 622.8 Hz Max: 3.804


3rd ips ips ips ips ips Actual:
Min: 2.920

921.8 Hz 946.7 Hz 971.6 Hz 996.5 Hz 1021.4 Hz Max: 2.282


5th ips ips ips ips ips Actual:
Min: 1.752

1320.4 Hz 1345.3 Hz 1370.2 Hz 1395.1 Hz 1420.0 Hz Max: 1.630


7th ips ips ips ips ips Actual:
Min: 1.252

1694.0 Hz 1718.9 Hz 1743.9 Hz 1768.8 Hz 1793.7 Hz Max: 1.268


9th ips ips ips ips ips Actual:
Min: 0.973

2092.6 Hz 2117.5 Hz 2142.5 Hz 2167.4 Hz 2192.3 Hz Max: 1.037


11th ips ips ips ips ips Actual:
Min: 0.796

2491.2 Hz 2516.1 Hz 2541.1 Hz 2566.0 Hz 2590.9 Hz Max: 0.878


13th ips ips ips ips ips Actual:
Min: 0.674

2864.9 Hz 2889.8 Hz 2914.7 Hz 2939.7 Hz 2964.6 Hz Max: 0.761


15th ips ips ips ips ips Actual:
Min: 0.584

3263.5 Hz 3288.4 Hz 3313.3 Hz 3338.2 Hz 3363.2 Hz Max: 0.671


17th ips ips ips ips ips Actual:
Min: 0.515

3662.1 Hz 3687.0 Hz 3711.9 Hz 3736.8 Hz 3761.8 Hz Max: 0.601


19th ips ips ips ips ips Actual:
Min: 0.461

4035.8 Hz 4060.7 Hz 4085.6 Hz 4110.5 Hz 4135.4 Hz Max: 0.543


21st ips ips ips ips ips Actual:
Min: 0.417

9
T.O. 33K6-4-3011-1

Table 3. (Cont.)

NOTE

These frequencies are based on a 194.9 Hz squarewave input.

A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips

4434.4 Hz 4459.3 Hz 4484.2 Hz 4509.1 Hz 4534.0 Hz Max: 0.496


23rd ips ips ips ips ips Actual:
Min: 0.381

4833.0 Hz 4857.9 Hz 4882.8 Hz 4907.7 Hz 4932.6 Hz Max: 0.456


25th ips ips ips ips ips Actual:
Min: 0.350

5206.7 Hz 5231.6 Hz 5256.5 Hz 5281.4 Hz 5306.3 Hz Max: 0.423


27th ips ips ips ips ips Actual:
Min: 0.324

5605.3 Hz 5630.2 Hz 5655.1 Hz 5680.0 Hz 5704.9 Hz Max: 0.393


29th ips ips ips ips ips Actual:
Min: 0.302

6003.9 Hz 6028.8 Hz 6053.7 Hz 6078.6 Hz 6103.5 Hz Max: 0.368


31st ips ips ips ips ips Actual:
Min: 0.283

6377.6 Hz 6402.5 Hz 6427.4 Hz 6452.3 Hz 6477.2 Hz Max: 0.346


33rd ips ips ips ips ips Actual:
Min: 0.265

6776.1 Hz 6801.1 Hz 6826.0 Hz 6850.9 Hz 6875.8 Hz Max: 0.326


35th ips ips ips ips ips Actual:
Min: 0.250

7149.8 Hz 7174.7 Hz 7199.7 Hz 7224.6 Hz 7249.5 Hz Max: 0.308


37th ips ips ips ips ips Actual:
Min: 0.237

7548.4 Hz 7573.3 Hz 7598.3 Hz 7623.2 Hz 7648.1 Hz Max: 0.293


39th ips ips ips ips ips Actual:
Min: 0.225

7947.0 Hz 7971.9 Hz 7996.9 Hz 8021.8 Hz 8046.7 Hz Max: 0.312


41st ips ips ips ips ips Actual:
Min: 0.190

8320.7 Hz 8345.6 Hz 8370.5 Hz 8395.4 Hz 8420.4 Hz Max: 0.298


43rd ips ips ips ips ips Actual:
Min: 0.182

10
T.O. 33K6-4-3011-1

Table 3. (Cont.)

NOTE

These frequencies are based on a 194.9 Hz squarewave input.

A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips

8719.3 Hz 8744.2 Hz 8769.1 Hz 8794.0 Hz 8819.0 Hz Max: 0.285


45th ips ips ips ips ips Actual:
Min: 0.174

9117.9 Hz 9142.8 Hz 9167.7 Hz 9192.6 Hz 9217.6 Hz Max: 0.272


47th ips ips ips ips ips Actual:
Min: 0.166

9491.6 Hz 9516.5 Hz 9541.4 Hz 9566.3 Hz 9591.2 Hz Max: 0.261


49th ips ips ips ips ips Actual:
Min: 0.159

9890.2 Hz 9915.1 Hz 9940.0 Hz 9964.9 Hz 9989.8 Hz Max: 0.251


51st ips ips ips ips ips Actual:
Min: 0.153

4.4.2 LOW FREQUENCY:

4.4.2.1 Set the Dynamic Signal Analyzer CENTER FREQUENCY to 490 Hz.

4.4.2.2 Set the Synthesized Function Generator FREQUENCY to 490 Hz and AMPLITUDE for 134.4 mV rms
(squarewave) as verified by the Dynamic Signal Analyzer.

4.4.2.3 Press TI STATUS key.

4.4.2.4 Press TI MORE KEYS key.

4.4.2.5 Press TI SELF-TEST key.

4.4.2.6 Press TI HW TEST key. The TI Self-Test menu appears.

4.4.2.7 Select TI ANALOG FILTER TEST with the arrow keys.

4.4.2.8 Press TI EXECUTE key.

NOTE

You cannot proceed until the Analog Filter test passes. Occasionally the
Analog Filter test may fail with nothing permanently wrong with the TI. If
it does fail, press OK key and then EXECUTE key to run again. If it
continues to fail, refer to the commercial data.

4.4.2.9 After the TI display indicates Analog filter test successful, press OK key. The TI Self-Test menu will return.

11
T.O. 33K6-4-3011-1

4.4.2.10 Press TI QUIT key to exit Self-Test menu.

4.4.2.11 Press TI SETUP key.

NOTE

Setting the TI units to <RPM> will provide finer resolution.

4.4.2.12 On TI, Select <Hz> using arrow keys and press CHANGE key to change <Hz> to <RPM>.

4.4.2.13 Set TI controls as follows:

MAXIMUM FREQUENCY 29788 <RPM>

SEEK AMPLITUDE 0.010 ips

RECEPTION MODE <SINGLE>

SEEK ON

INPUT CHANNEL 1 <ONLY>

4.4.2.14 Press TI ENTER key and wait 10 seconds.

4.4.2.15 Press TI START key.

4.4.2.16 Verify the 134.4 mV rms (squarewave) output of the Synthesized Function Generator on the Dynamic Signal
Analyzer.

4.4.2.17 Use the TI arrow keys to move the cursor to the fundamental peak. Verify the TI readout is between 9.5 and
10.4 ips (±0.2 dB).

4.4.2.18 Repeat steps 4.4.2.1, 4.4.2.2 and 4.4.2.15 through 4.4.2.17 for the following TI frequencies: 406, 308, 210,
154 and 98 Hz.

4.4.2.19 Press TI SETUP key.

4.4.2.20 Set TI MAXIMUM FREQUENCY to 16596 <RPM>, and press TI ENTER key and wait 10 seconds.

4.4.2.21 Repeat steps 4.4.2.1, 4.4.2.2 and 4.4.2.15 through 4.4.2.17 for the following TI frequencies: 51, 21, 9, 6 and 3 Hz.

4.4.2.22 Set the Synthesized Function Generator and Dual DC Power Supply outputs to minimum and disconnect test
setup.

4.5 CHARGE AMPLIFIER CALIBRATION:

4.5.1 Connect Synthesized Function Generator (#1) through Feedthrough Termination (#1) to the Digital Multimeter HI
and LO INPUT terminals.

4.5.2 Set Digital Multimeter for a VAC measurement at 100 mV rms.

4.5.3 Set Synthesized Function Generator (#1) FREQUENCY to 100 Hz and AMPLITUDE for 100 mV rms
(sinewave) signal. Adjust Synthesized Function Generator (#1) for a Digital Multimeter indication of 100.00 mV rms.

4.5.4 Disconnect Synthesized Function Generator ( #1) and Feedthrough Termination (#1) from the Digital Multimeter
INPUT terminals.

12
T.O. 33K6-4-3011-1

4.5.5 Connect Synthesized Function Generator (#2) through Feedthrough Termination (#2) to the Digital Multimeter
INPUT terminals, observing polarity.

4.5.6 Set Synthesized Function Generator (#2) FREQUENCY to 100 Hz and AMPLITUDE for 100 mV rms
(sinewave) signal. Adjust Synthesized Function Generator (#2) for a Digital Multimeter indication of 100.00 mV rms.

4.5.7 Disconnect Synthesized Function Generator (#2) and Feedthrough Termination (#2) from the Digital Multimeter
INPUT terminals.

4.5.8 Connect RLC Digibridge to TI Charge Amplifier A and B Input terminals. Set the RLC Digibridge to measure the
input capacitance on the Charge Amplifier A and B and record the value as Cin.

4.5.9 Disconnect test setup.

4.5.10 Remove Charge Amplifier cover and set DIP switches for the sensitivity closest to that of the intended
accelerometer. Replace Charge Amplifier cover after setting DIP switches.

4.5.11 Connect equipment as shown in Figure 3.

SYNTHESIZED FUNCTION SYNTHESIZED FUNCTION


GENERATOR (#1) * GENERATOR (#2) *

SIGNAL SIGNAL Connect Synthesized Function


FREQUENCY * Generator (#1) and (#2) to 115V/60 Hz
OUT EXT REF OUT
line input using a ground buster.
FEEDTHROUGH
Connect the Synthesized TERMINATION (#1)
Function Generator (#2)
HI Output to Synthesized FEEDTHROUGH
Function Generator (#1) TERMINATION (#2)
LO Output for a differential TI 28 VDC
output. SIGNAL
CHARGE AMPLIFIER
(OUTPUT SIDE)

CHARGE AMPLIFIER CABLE

A C
OUTPUT/POWER
DIGITAL MULTIMETER B
VIBRATION CHANNELS TO TI
CHARGE AMPLIFIER 1 4 7 10
(INPUT SIDE) 2 5 8 11
3 6 9 12
SIGNAL SELECTOR CABLE
VIEW A MAGNETIC PICKUP/
PHOTOCELL CHANNELS
A B C D

SIGNAL SELECTOR

Connect the Synthesized Connect the Synthesized


Function Generator (#1) Function Generator (#2)
through Feedthrough through Feedthrough
Termination (#1). Termination (#2).

Connect the Synthesized


Function Generator (#2)
HI Output to Synthesized
Function Generator (#1)
LO Output for a differential HI HI
output.

TO DIGITAL A C
MULTIMETER
B

CHARGE AMPLIFIER
(INPUT SIDE)

VIEW A

Figure 3.

Change 1 13
T.O. 33K6-4-3011-1

4.5.12 Fine tune Synthesized Function Generator (#1) AMPLITUDE for 100.00 mV rms as indicated on Digital
Multimeter.

4.5.13 Press TI BALANCE key.

4.5.14 Press TI SETUP key and set the TI as follows:

BALANCE FREQUENCY 100.0 <Hz> X 1

BALANCE TYPE <STROBEX> <MAN STOP>

4.5.15 Press TI START key (ignore TI STROBEX instructions).

4.5.16 Calculate required TI indication using the following formula:


⎜ 61.45 * (Cin + Cacc ) ⎞⎟
ips = 2 * Vin ⎜ ⎟
⎜ f * SQ ⎟
⎝ ⎠

Vin = 100 mV rms

Cin = recorded value from step 4.5.8 in pF

Cacc = 12,200 pF (For accelerometer #6222S-100A), 1350 pF (For


accelerometer #6233C-50)

f = 100 Hz

SQ = 100 pC/g (For accelerometer #6222S-100A), 50 pC/g (For


accelerometer #6233C-50)

Example:

Measured input capacitance from step 4.5.8: Cin=100400 pF

Amplitude and frequency from


Synthesized Function Generator: Vin =100 mV rms, f =100 Hz

SQ = 100 pC/g (For accelerometer #6222S-100A), 50 pC/g (For


accelerometer #6233C-50)

⎛ 61.45 * ( 100400 + 12200) ⎞⎟


ips = 2 * 0.100⎜⎜ ⎟ = 97.85 ips
⎝ 100 * 100 ⎠

4.5.17 Verify TI indication is within ±1% of value calculated in step 4.5.16.

4.5.18 Set Synthisized Function Generator outputs to minimum and disconnect test setup.

4.6 VIBRATION METER CALIBRATION:

4.6.1 FREQUENCY RESPONSE:

4.6.1.1 Connect equipment as shown in Figure 4.

4.6.1.2 Set Dual DC Power Supply output as needed for -15 VDC at TI pins A - B and +15 VDC at TI pins C - B as
shown in Figure 4.

14 Change 1
T.O. 33K6-4-3011-1

4.6.1.3 Set TI CHANNEL switch to the CHANNEL 1 position.

4.6.1.4 Set TI FILTER switch to the first position listed in the Filter column of Table 4.

METER CALIBRATOR

Connect Meter Calibrator


OUTPUT SENSE
HI terminal to TI CH1
connector pin D (HI)
and Meter Calibrator LO
terminal to TI CH1
connector pin B (GND).

CH 1 CH 2 CH 3 CH 4

MILS D.A.

CHANNEL
2 3
1 4

DUAL DC
FILTER TEST
POWER SUPPLY IN

OUT

TI TI cable and
+ - + -
connector.

A C
Shorting Bar B

Figure 4.

4.6.1.5 Set Meter Calibrator to the first value listed in the Applied column of Table 4.

4.6.1.6 Set Meter Calibrator OPR/STBY switch to OPR.

4.6.1.7 The TI display must indicate within the corresponding values listed in the Limits column of Table 4.

4.6.1.8 Set Meter Calibrator OPR/STBY switch to STBY.

4.6.1.9 Repeat steps 4.6.1.4 through 4.6.1.8 for each remaining value listed in Table 4.

4.6.1.10 Repeat steps 4.6.1.1 and 4.6.1.3 through 4.6.1.9 for TI CHANNEL 2, CHANNEL 3 and CHANNEL 4
INPUT.

Table 4.

Applied (mV rms) Filter Nominal (ips) Limits (ips)

636 @ 150 Hz IN 70.65 56.5 to 84.8

63.6 @ 15 Hz OUT 100 90 to 110

318 @ 75 Hz OUT 100 90 to 110

318 @ 75 Hz IN 6.2 4.1 to 8.3

212 @ 500 Hz OUT 10.0 9.0 to 11.0

15
T.O. 33K6-4-3011-1

Table 4. (Cont.)

Applied (mV rms) Filter Nominal (ips) Limits (ips)

424 @ 1000 Hz OUT 10.0 9.0 to 11.0

424 @ 1000 Hz IN 7.05 5.6 to 8.5

678 @ 2000 Hz OUT 8.0 7.2 to 8.8

678 @ 2000 Hz IN 0.4 0.0 to 0.8

4.6.1.11 Set Dual DC Power Supply outputs to minimum.

4.6.1.12 Disconnect Meter Calibrator from test setup.

4.6.2 INTERNAL CALIBRATION SOURCE:

4.6.2.1 Connect TI CH 4 input connector pin D (HI) and pin B (LO) to Dynamic Signal Analyzer INPUT connector,
observing polarity.

4.6.2.2 Set Dual DC Power Supply as needed for -15 VDC at TI pins A - B and +15 VDC at TI pins C - B as shown in
Figure 4.

4.6.2.3 Set Dynamic Signal Analyzer as follows:

Vert Scale Define Full Scale (SCL) 30 mV rms

Frequency Define Span 100 Hz

Frequency Define Center 64 Hz

4.6.2.4 From step 4.6.2.5 through step 4.6.2.14 press TI TEST pushbutton as needed to apply TI TEST signal to
Dynamic Signal Analyzer INPUT connector.

4.6.2.5 Ensure the Dynamic Signal Analyzer MARKER is at the peak of the TI 64 Hz signal and record the
MARKER indication.

4.6.2.6 Disconnect TI CH 4 input from Dynamic Signal Analyzer INPUT connector and connect Meter Calibrator
OUTPUT terminals to Dynamic Signal Analyzer Input connector, observing polarity.

4.6.2.7 Set Meter Calibrator for 27.4 mV rms/64 Hz and set OPR/STBY switch to OPR.

4.6.2.8 Edit Meter Calibrator display as necessary for a Dynamic Signal Analyzer MARKER indication equal to the
value recorded in step 4.6.2.5.

4.6.2.9 The Meter Calibrator must indicate between 27.126 to 27.674 mV rms.

4.6.2.10 Set Meter Calibrator OPR/STBY switch to STBY.

4.6.2.11 Disconnect Meter Calibrator OUTPUT terminals from Dynamic Signal Analyzer INPUT connector.

4.6.2.12 Connect TI CH 4 input connector pin D (HI) and pin B (LO) to Dynamic Signal Analyzer INPUT
connector,observing polarity.

16
T.O. 33K6-4-3011-1

4.6.2.13 Set Dynamic Signal Analyzer as follows:

Frequency Define Center 256 Hz

4.6.2.14 Ensure the Dynamic Signal Analyzer MARKER is at the peak of the TI 256 Hz signal and record the
MARKER indication.

4.6.2.15 Disconnect TI CH 4 Input from Dynamic Signal Analyzer INPUT connector and connect Meter Calibrator
Output terminals to Dynamic Signal Analyzer INPUT connector, observing polarity.

4.6.2.16 Set Meter Calibrator for 27 mV rms/256 Hz and set OPR/STBY switch to OPR.

4.6.2.17 Edit Meter Calibrator display as necessary for a Dynamic Signal Analyzer MARKER indication equal to the
value recorded in step 4.6.2.14.

4.6.2.18 The Meter Calibrator must indicate between 27.126 to 27.674 mV rms.

4.6.2.19 Set all POWER switches to OFF/STBY, disconnect and secure all equipment.

CALIBRATION PERFORMANCE TABLE

4.1 BALANCE MEASUREMENT CALIBRATION:

Applied Limits

100 Hz 99.9 to 100.1 Hz

1.000 ips 0.960 to 1.040 ips

180° 177° to 183°

4.2 NOISE AND SPURIOUS SIGNAL CALIBRATION:

Applied Limits

1,000 Hz ≤0.005 ips

4.3 ANTI-ALIAS FILTER RESPONSE CALIBRATION:

Range Applied Limits

500 Hz 1,722 Hz @ 13.44 mV ≤0.005 ips

10 kHz 15,600 Hz @ 13.44 mV ≤0.005 ips

17
T.O. 33K6-4-3011-1

CALIBRATION PERFORMANCE TABLE (Cont.)

4.4 SPECTRUM ACCURACY CALIBRATION:

High Frequency:

Note

These frequencies are based on a 194.9 Hz squarewave input.

A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips

149.5 Hz l74.4 Hz 199.3 Hz 224.2 Hz 249.1 Hz Max: 11.411


1st ips ips ips ips ips Actual:
Min: 8.761

523.2 Hz 548.1 Hz 573.0 Hz 597.9 Hz 622.8 Hz Max: 3.804


3rd ips ips ips ips ips Actual:
Min: 2.920

921.8 Hz 946.7 Hz 971.6 Hz 996.5 Hz 1021.4 Hz Max: 2.282


5th ips ips ips ips ips Actual:
Min: 1.752

1320.4 Hz 1345.3 Hz 1370.2 Hz 1395.1 Hz 1420.0 Hz Max: 1.630


7th ips ips ips ips ips Actual:
Min: 1.252

1694.0 Hz 1718.9 Hz 1743.9 Hz 1768.8 Hz 1793.7 Hz Max: 1.268


9th ips ips ips ips ips Actual:
Min: 0.973

2092.6 Hz 2117.5 Hz 2142.5 Hz 2167.4 Hz 2192.3 Hz Max: 1.037


11th ips ips ips ips ips Actual:
Min: 0.796

2491.2 Hz 2516.1 Hz 2541.1 Hz 2566.0 Hz 2590.9 Hz Max: 0.878


13th ips ips ips ips ips Actual:
Min: 0.674

2864.9 Hz 2889.8 Hz 2914.7 Hz 2939.7 Hz 2964.6 Hz Max: 0.761


15th ips ips ips ips ips Actual:
Min: 0.584

3263.5 Hz 3288.4 Hz 3313.3 Hz 3338.2 Hz 3363.2 Hz Max: 0.671


17th ips ips ips ips ips Actual:
Min: 0.515

3662.1 Hz 3687.0 Hz 3711.9 Hz 3736.8 Hz 3761.8 Hz Max: 0.601


19th ips ips ips ips ips Actual:
Min: 0.461

4035.8 Hz 4060.7 Hz 4085.6 Hz 4110.5 Hz 4135.4 Hz Max: 0.543


21st ips ips ips ips ips Actual:
Min: 0.417

18
T.O. 33K6-4-3011-1

CALIBRATION PERFORMANCE TABLE (Cont.)

4.4 SPECTRUM ACCURACY CALIBRATION: (Cont.)

High Frequency:

Note

These frequencies are based on a 194.9 Hz squarewave input.

A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips

4434.4 Hz 4459.3 Hz 4484.2 Hz 4509.1 Hz 4534.0 Hz Max: 0.496


23rd ips ips ips ips ips Actual:
Min: 0.381

4833.0 Hz 4857.9 Hz 4882.8 Hz 4907.7 Hz 4932.6 Hz Max: 0.456


25th ips ips ips ips ips Actual:
Min: 0.350

5206.7 Hz 5231.6 Hz 5256.5 Hz 5281.4 Hz 5306.3 Hz Max: 0.423


27th ips ips ips ips ips Actual:
Min: 0.324

5605.3 Hz 5630.2 Hz 5655.1 Hz 5680.0 Hz 5704.9 Hz Max: 0.393


29th ips ips ips ips ips Actual:
Min: 0.302

6003.9 Hz 6028.8 Hz 6053.7 Hz 6078.6 Hz 6103.5 Hz Max: 0.368


31st ips ips ips ips ips Actual:
Min: 0.283

6377.6 Hz 6402.5 Hz 6427.4 Hz 6452.3 Hz 6477.2 Hz Max: 0.346


33rd ips ips ips ips ips Actual:
Min: 0.265

6776.1 Hz 6801.1 Hz 6826.0 Hz 6850.9 Hz 6875.8 Hz Max: 0.326


35th ips ips ips ips ips Actual:
Min: 0.250

7149.8 Hz 7174.7 Hz 7199.7 Hz 7224.6 Hz 7249.5 Hz Max: 0.308


37th ips ips ips ips ips Actual:
Min: 0.237

7548.4 Hz 7573.3 Hz 7598.3 Hz 7623.2 Hz 7648.1 Hz Max: 0.293


39th ips ips ips ips ips Actual:
Min: 0.225

7947.0 Hz 7971.9 Hz 7996.9 Hz 8021.8 Hz 8046.7 Hz Max: 0.312


41st ips ips ips ips ips Actual:
Min: 0.190

8320.7 Hz 8345.6 Hz 8370.5 Hz 8395.4 Hz 8420.4 Hz Max: 0.298


43rd ips ips ips ips ips Actual:
Min: 0.182

19
T.O. 33K6-4-3011-1

CALIBRATION PERFORMANCE TABLE (Cont.)

4.4 SPECTRUM ACCURACY CALIBRATION: (Cont.)

High Frequency:

Note

These frequencies are based on a 194.9 Hz squarewave input.

A B C D E
2nd Amp 1st Amp Harmonic 1st Amp 2nd Amp True Amp
Harmonic Left Left Peak Right Right ips

8719.3 Hz 8744.2 Hz 8769.1 Hz 8794.0 Hz 8819.0 Hz Max: 0.285


45th ips ips ips ips ips Actual:
Min: 0.174

9117.9 Hz 9142.8 Hz 9167.7 Hz 9192.6 Hz 9217.6 Hz Max: 0.272


47th ips ips ips ips ips Actual:
Min: 0.166

9491.6 Hz 9516.5 Hz 9541.4 Hz 9566.3 Hz 9591.2 Hz Max: 0.261


49th ips ips ips ips ips Actual:
Min: 0.159

9890.2 Hz 9915.1 Hz 9940.0 Hz 9964.9 Hz 9989.8 Hz Max: 0.251


51st ips ips ips ips ips Actual:
Min: 0.153

Low Frequency:

Frequency Amplitude Limits

490 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

406 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

308 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

210 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

154 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

98 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

51 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

21 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

9 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

6 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

3 Hz 134.4 mV rms 9.5 to 10.4 ips (±0.2 dB)

20
T.O. 33K6-4-3011-1

CALIBRATION PERFORMANCE TABLE (Cont.)

4.5 CHARGE AMPLIFIER CALIBRATION:

Range Limits

100 Hz @ 100 mV rms (sinewave) Within ±1% of the calculated


value of step 4.5.16

4.6 VIBRATION METER CALIBRATION:

Frequency Response:

Applied (mV rms) Filter Nominal (ips) Limits (ips)

636 @ 150 Hz IN 70.65 56.5 to 84.8

63.6 @ 15 Hz OUT 100 90 to 110

318 @ 75 Hz OUT 100 90 to 110

318 @ 75 Hz IN 6.2 4.1 to 8.3

212 @ 500 Hz OUT 10.0 9.0 to 11.0

424 @ 1000 Hz OUT 10.0 9.0 to 11.0

424 @ 1000 Hz IN 7.05 5.6 to 8.5

678 @ 2000 Hz OUT 8.0 7.2 to 8.8

678 @ 2000 Hz IN 0.4 0.0 to 0.8

Internal Calibration Source:

Output Limits (mV rms)

27.4 mV rms @ 64 Hz 27.126 to 27.674

27.4 mV rms @ 256 Hz 27.126 to 27.674

Change 1 21/(22 Blank)


T.O. 33K6-4-3011-1

APPENDIX A

C-H #402-1061E (crimped pins)


MS3476L10-6P (crimped pins) BNC
ITT Cannon #PV76L10-6P (crimped pins) CONNECTOR

RG-58 COAX CABLE

SHIELD

TEST CABLE #8278 VELOCIMETER-TO-BNC


C-H #402-0841B (soldered pins)
MS3116F8-4P (soldered pins)
MS3126F8-4P (crimped pins)
ITT Cannon #KPSE06F8-4P (crimped pins) BNC
ITT Cannon #KPT06F8-4P (soldered pins) CONNECTOR

RG-58 COAX CABLE

D
B
A NC
SHIELD
C NC

TEST CABLE #10502 MAGNETIC PICK-UP-TO-BNC


P10
BLK
A
GRN A PWR GND
N PWR
WHT
L B +28 VDC PWR
RED
M

POWER/SIGNAL
BLK
INPUTS C A +V 12-15 VDC
GRN
R B GND
P
RED
C ±28 VDC MAG/PHOTOCELL
WHT
D D SIGNAL
E
BLK
J A -V
GRN
S B GND VELOCIMETER
RED
T C +V
WHT
H BRAID
D SIGNAL
G N/C E
N/C F
B N/C
K P5-1 FROM DIGITAL BOARD
U P5-3 FROM DIGITAL BOARD
F N/C

TEST CABLE #8998 INPUT ADAPTER CABLE

A-1/(A-2 Blank)