You are on page 1of 10

Colloidal Dynamics

leaders in colloid measurement

HOW DO THE ZETAPROBE AND ACOUSTOSIZER II


MEASURE ZETA POTENTIAL AND PARTICLE SIZE?
The ZetaProbe and the AcoustoSizer II measure the particle size and zeta
potential in concentrated colloids. Most of the other commercial devices need to
dilute the colloid to make these measurements. The dilution takes time, and it
often alters the very things you are trying to measure. In this article we will
describe how our products actually determine size and zeta in concentrates,
and we will present some applications.

Introduction
Background: What is the Zeta
Potential?
The ZetaProbe and AcoustoSizer II have
Colloidal particles are usually
become routine instruments for the
electrically charged. They are
measurement of particle size and zeta
surrounded, in suspension by a
potential in colloidal systems. One of the
cloud of ions that carry an equal and
great advantages of these devices is that
opposite charge. The zeta potential
they can measure size and zeta without
is the voltage difference between
dilution.
the particle surface and the liquid
beyond the charge cloud. In colloids
The traditional devices for measuring
that are electrostatically stabilized,
size and zeta use optical techniques such
zeta provides a measure of the
as light scattering. The samples need to
electrical repulsive force between
be quite transparent to use these
the particles. For other systems
products, and this usually means they
where the stability comes from
must be diluted down to concentrations
steric components, the zeta
of approximately 10 mg per litre. Even
potential can be used as a measure
the more recent optical devices that
of the state of the surface. It can be
claim to be able to measure zeta
used to monitor optimum levels of
potential in ‘concentrated’ suspensions
dispersant or other agents to be
can only do so at concentrations of 0.5
added to the particles.
% w/v under favourable conditions of
low opacity. This is still one to two
Thus zeta potential can be used for
orders of magnitude more dilute than
monitoring and controlling colloid
most colloids.
stability and as an indicator of the
This dilution is time consuming, an particle surface chemistry.
important consideration in QC labs
where sample throughput is vital. And
the dilution can lead to large errors in light, and so it is not necessary to see
the zeta potential if there is a change in through the suspension to measure it.
the concentration of the background
electrolyte. Furthermore, the diluted Two measurement techniques are
sample has such a small particle surface employed – electroacoustics and
area that the zeta can be altered by ultrasonic attenuation. The first
trace amounts of surface-active technique is used in both products, while
impurities in the sample. the attenuation method is only used in
the AcoustoSizer II, for getting particle
How can the ZetaProbe and AcoustoSizer size.
II measure in concentrates? The reason
is that they measure sound rather than
We will begin our description with the up by a piezoelectric transducer shown
electroacoustic method, since this is on the left-hand side of the diagram.
common to both products.
The ESA depends on the particle
THE ESA MEASUREMENT velocity, which in turn depends on the
particle charge. The ESA also depends
on the particle size, for bigger particles
In this technique we measure ultrasound
generated by the particles. A MHz have more inertia, and this causes them
voltage is applied across a pair of to lag behind the electric field. Thus
there will be a time delay between the
electrodes in contact with the
field changing direction and the particle
suspension. Colloidal particles are
velocity changing direction. Hence the
electrically charged, so they move back
and forth, first towards one electrode particle size affects the phase of the
and then the other. It is this particle ESA, and so by measuring the
magnitude and phase of the ESA sound
motion that generates the ultrasound, in
wave we can determine both particle
the same way that the back and forth
charge (or zeta potential) and size.
motion of a loudspeaker generates
sound. The sound wave from each
particle is very tiny, but the particles The dynamic mobility
move in phase and this cooperative
motion results in a measurable beam of The ESA depends on particle size and
ultrasound. zeta, but it also depends on the
geometry of the measuring cell, the
This effect, of sound waves generated by strength of the applied field, the material
an applied electric field, is called the the cell is made of, the type of
Electrokinetic Sonic Amplitude Effect, or transducer used and so on. The first step
ESA for short. in the process of getting size and zeta is
to extract, from the measurement, a
quantity that only depends on the colloid
and not on the device used for making
the measurement.

The colloidal property that we extract


from the ESA measurement is called the
dynamic mobility, denoted by the
symbol µD. This is like the normal
electrophoretic mobility, but it is defined
for a sinusoidal electric field instead of a
steady field. In this section we will give a
precise definition of this quantity.

A sinusoidal electric field of the form

E cos ω t (1)
Figure 1 The AcoustoSizer II ESA measurement
cell will cause the colloidal particles to move
with a sinusoidal velocity
In figure 1 we show a schematic
diagram of the AcoustoSizer V cos (ω t − θ ) (2)
measurement cell. The colloid flows up
between a pair of parallel plate where t is time and ω is the angular
electrodes that are connected to a MHz frequency. The phase angle θ has been
voltage source. The ESA sound wave
introduced to allow for the phase lag
passes out of the suspension and travels
caused by particle inertia.
through a glass block where it is picked

Colloidal Dynamics www.colloidal-dynamics.com 2


The particle velocity and the applied field frequency increases, and the inertia
are represented by the curves in figure becomes significant, a phase lag will
2, for the case of a 300 phase lag. develop and the dynamic mobility will
have a negative argument. The inertia
The standard electrophoretic mobility is forces only usually become significant
defined by the single quantity V/E, but in for colloidal particles at frequencies in
the case of a sinusoidal field, two the MHz range. It would be impossible to
quantities are required to define the measure the dynamic mobility by optical
mobility, viz the ratio V/E and the phase techniques at such high frequencies, but
angle θ. the ESA effect provides a way of
determining this quantity.
1.5
Before moving on to this next step of
1
getting the mobility from the ESA, we
0.5
note that if the sign of the zeta potential
is reversed- something that can be done
E(t)
0
V(t)
for oxide particles by altering the pH-
-0.5
0 20 40 60 80 100
the argument of the mobility will swing
through 180 degrees. This is because
-1 the particle velocity will now be in the
opposite direction. Hence there will be a
-1.5
minus sign out the front of the formula
(2), which is the same as adding 180
Figure 2 The sinusoidal particle velocity V degrees to the argument in that formula.
and electric field E
This is how we determine the isolectric
point (iep) of a colloid- that is, the pH
These two parameters are combined in at which the particle has zero zeta
the definition of the dynamic mobility, potential. As we change the pH from one
which is a complex number with a side of the iep to the other, the
magnitude and argument are given by argument of the mobility swings through
180 degrees. This allows us to determine
V the iep very accurately.
Mag ( µ D ) = , and Arg ( µ D ) = −θ (3)
E

Readers who are not familiar with the Getting the dynamic mobility
terms Mag and Arg may recall that a from the ESA
complex number z can be represented as
a point on the x-y plane, where the x To find the mathematical relation
coordinate is the real part and the y between the measured ESA and the
coordinate is the imaginary part of the dynamic mobility, it is necessary to solve
complex number. The magnitude (or the equations of motion for a colloid
Mag) of z is the distance from the origin subjected to an applied electric field.
to the point (x,y), and the argument The mathematical details are set out in
(Arg) is the angle that the line from the the Appendix of the 2003 paper by
origin to the point makes with the x axis. O’Brien, Jones and Rowlands (see the
Thus the magnitude of the dynamic references for details), where it is shown
mobility is defined in the same way as that
the electrophoretic mobility, and its
argument is minus the phase lag. zs zg ρp − ρ
ESA = A(ω ) φ µD (4)
zs + z g ρ
At low frequencies, the inertia forces are
very small and the particles will move in
phase with the electric field. In this limit There are a number of new terms in this
the dynamic mobility is the same as the formula. The quantity φ is the particle
electrophoretic mobility. But as the volume fraction, that is, the fraction of

Colloidal Dynamics www.colloidal-dynamics.com 3


the total suspension volume occupied by
the particles. ρ p is the density of the
particles and ρ is the solvent density. A
is an instrument factor; this depends on
things like the transducer dimensions,
Calibration
the applied field, and the instrument
To determine the dynamic mobility from
geometry but it does not depend on the
the ESA using the formula (4), it is
colloid.
necessary to know the instrument factor
A(ω). We determine this by measuring
Finally, the quantities zs and zg are the
the ESA from a solution with a known
acoustic impedance of the suspension
dynamic mobility factor.
and the AZR2 glass block respectively.
The acoustic impedance is a measure of
This step, of determining the instrument
how much force is required to generate
factor by measuring the ESA from a
an ultrasound beam in a material. It is
reference solution is known as
defined as follows: suppose that a
calibration.
material- the suspension say- is
bounded by a flat solid surface which is
In the other commercial electroacoustic
acted on by a sinusoidal pressure. This
devices, the calibration is carried out by
will cause the surface to move back and
making measurements on a colloid with
forward with a sinusoidal velocity of
a “known” zeta potential, that is, one for
amplitude V. The pressure P on the
which zeta has been determined on one
material will be proportional to the
of the standard optical instruments. But
velocity. The acoustic impedance z of the
there are problems with this approach:
material is defined by as the constant of
first, the colloid has to be diluted for
proportionality between the pressure
measurement in the optical device and
and the velocity, that is
this can alter zeta, and secondly the zeta
potential of a colloid often changes with
P time, so a colloid that has been left
z= (5)
V standing in a bottle will have a different
In the case of a homogeneous material, zeta from a fresh sample.
such as the glass block, the acoustic
impedance is equal to the speed of At Colloidal Dynamics we have overcome
sound times the density of the material, these problems by using an electrolyte
but for a suspension, it depends on the as a calibrant. The ESA in this case is
properties of the particles and the generated by the ions, rather than
solvent. We determine the acoustic colloidal sized particles, but the formula
impedance by measuring the reflection (4), still applies; the only difference is
of various waves from the suspension that the term
interface.
ρp − ρ
When the ESA sound wave strikes the φ µD (6)
transducer at the far end of the glass
ρ
block in figure 1, it generates a
sinusoidal voltage signal. The quantity is replaced by a sum over the individual
on the left hand side of (4) is a complex ionic species. A solution of potassium
number with a magnitude equal to the silico tungstate (KSiW) ions has been
amplitude of that sinusoidal voltage, and chosen as the calibrant because it gives
an argument equal to the phase a large, stable ESA signal. In the 1995
difference between this voltage and the paper by O’Brien, Cannon and Rowlands,
driving voltage that is applied to the it is shown that the quantity (6) for this
colloid to generate the ESA signal. Thus electrolyte is equal to –3.02×10- 9 K,
the ESA in this formula is a complex where K is the conductivity of the
number, like the dynamic mobility, with solution in SI units.
a magnitude and an argument.

Colloidal Dynamics www.colloidal-dynamics.com 4


Thus the calibration involves the From the data in Figure 3 it can be seen
measurement of the ESA and the that the two nanoparticle suspensions
conductivity from a KSiW solution. The have quite flat magnitude spectra, and
instrument factor is determined from the small arguments. This is because the
measured ESA using a formula obtained inertia forces are very small compared to
by solving equation (4) for A(ω ). the viscous forces for these particles.
These particles are moving with almost
Dynamic mobility the same mobility as for a steady field,
even at our highest measurement
measurements frequency of 18 MHz. For the Klebersol,
the inertia forces are so small that the
In figure 3 we show AcoustoSizer II particles cannot be sized by the ESA
measurements of the dynamic mobility method over this frequency range. We
spectra of four silica samples: can still determine its zeta potential of
− A one micron spherical course, but to size these particles we
colloidal silica sol from need to use the ultrasonic attenuation
Geltech, method, which will be discussed shortly.
− An 0.3 micron diameter
spherical colloidal silica from At very high frequencies, the inertia
Nissan Chemical, forces are so large that the particles
− Klebesol, 30N50 colloidal barely move in the electric field, and the
silica with median diameter ESA is then too small to measure. This is
around 70nm, and what happens for particles larger than
− Cabot SCE fumed silica, with about 10 microns in the AcoustoSizer II.
median diameter of about 50 Even at our lowest measurement
nm. frequency of 1 MHz, the ESA signal is
too small to allow particle size
The units for mobility magnitudes in the determination for particles much larger
left hand graph are 10-8 SI units. than 10-15 microns.

From these measurements it can be Thus, in the AcoustoSizer II, the ESA
seen that the mobility magnitudes drop method can be used for getting particle
off more rapidly with frequency for size in the range of about 70 nm to 10
larger particles, and the phase lags microns diameter. For smaller particles
increase more rapidly as the particle size the ESA technique can give the zeta
goes up. Thus the size can be potential, but not the particle size.
determined from the shape of the
dynamic mobility spectrum. And once Our ZetaProbe zeta potential analyzer
the size has been calculated the zeta can also measures the Dynamic Mobility
be determined from the absolute Spectrum but uses a different frequency
mobility magnitudes. range than the AcoustoSizer II. The
ZetaProbe measures the Dynamic

Mobility Magnitude Spectra Mobility Argument Spectra


Compare several silica slurries Compare several silica slurries
5
7
0
Mobility Mag. [ m^2.v^-1.s^-1.10e-

6 -5
Mobility Arg. [degrees]

-10
5
-15
4
-20
8]

3 -25

2 -30
-35
1
-40
0 -45
0.0 5.0 10.0 15.0 20.0 0.0 5.0 10.0 15.0 20.0
Frequency [MHz]
Frequency [MHz]

Nissan 0.3 Micron Cabot SCE Nissan 0.3 Micron Geltech 1.0 Micron
Geltech 1.0 Micron Klebesol 30N50 Cabot SCE Klebesol 30N50

Colloidal Dynamics www.colloidal-dynamics.com 5


Figure 3 The measured dynamic mobility magnitudes for four silica
samples
Mobility at 7 different frequencies over a getting size and zeta. There is much
frequency range of 300 kHz to 3 MHz. more that can be said on this topic, but
By operating at lower frequencies, the these details, will be deferred for other
ZetaProbe can accurately fit the Mobility articles. The main aim of this document
Spectrum for average particle sizes up to is to describe the measurement
about 30 microns for the particle d50 techniques used in the AcoustoSizer II
size. In this lower frequency range, the and the ZetaProbe, so we will now leave
Mobility Spectrum will be flat for particle the ESA method and turn to the second
sizes less than about 0.5 microns. For measurement technique, ultrasonic
this reason, the ZetaProbe is primarily a attenuation.
zeta potential analyzer, not a particle
sizing device. The instrument THE ULTRASONIC
automatically compensates for the effect ATTENUATION METHOD
of particle inertia for large particles and
has no lower limit in particle size. Both
In the ESA method, the particles create
the ZetaProbe and the AcoustoSizer II
the ultrasound, but in the ultrasonic
can measure the zeta potential of
attenuation measurements the
nanoparticles approaching the ultrasound is generated by an external
dimensions of macro ions. transducer. This beam of ultrasound is
then measured on a second transducer
Getting size and zeta from the after passing through the colloid. The
dynamic mobility spectrum colloid attenuates the ultrasound by an
amount that depends on the particle size
To determine a particle size distribution and concentration. By measuring the
and a zeta potential from the measured attenuation as a function of frequency
dynamic mobility spectrum, it is we can therefore determine a particle
necessary to have a theoretical formula size distribution.
that relates dynamic mobility to particle
size and zeta. Such a formula is now The measurement is carried out in the
available for concentrated colloids (in AcoustoSizer cell shown in figure 4. This
O’Brien et al’s 2003 paper), so the is the same cell used for the ESA
measured mobility spectra can now be measurement, but now we have included
used for obtaining the zeta and size the extra glass block and transducer on
distribution in colloids of arbitrary the right hand side of the diagram.
concentration. Our software routines do These were not shown in figure 1
this by adjusting the size distribution because they are not used in the ESA
and zeta potential to give the best fit measurement.
between the measured and theoretical
dynamic mobility spectra. In the attenuation measurement a large
sinusoidal voltage pulse is applied to the
Note that we use the entire mobility transducer on the right hand side of the
spectrum in the determination of size cell in figure 4. This causes the
and zeta from the ESA. This method of transducer to expand and contract in a
using the mobility measurements over a sinusoidal fashion, thereby generating a
range of frequencies is part of our beam of ultrasound. This beam travels
patented technology. The other through the glass block on the right,
commercial electroacoustic devices are through the suspension and then
limited, because of our patents, to through the glass block on the left hand
measurements at a single frequency. As side, where it is picked up by the
a result they cannot get size and zeta transducer attached to the back of that
from their ESA measurements. The best block.
they can do is to determine zeta for
colloids with a known particle size. As the sound wave passes through the
suspension it moves the particles back
This completes our introductory and forth. If the density of the particles
description of the ESA method for is different from the solvent, the
different inertia forces will cause the

Colloidal Dynamics www.colloidal-dynamics.com 6


particles to move relative to the solvent. When a locally planar sound wave
This relative motion generates a local passes through a suspension, its
flow around each particle. Energy is amplitude drops off like exp (-αx ),
dissipated in this local flow by the where x is distance into the suspension.
frictional forces in the liquid. Hence
some of the sound wave energy is The voltage signal in the receiving
converted into heat and thus the transducer of the AcoustoSizer II is
soundwave is attenuated as it passes obviously related to this attenuation
through the colloid. factor α, for as the attention increases
the signal will decrease. The precise
This dissipation mechanism, caused by mathematical form of this relation can
be shown to be

zs zg
S = B (ω ) exp ( −α L ) (7)
( zs + zg )
2

where B is an instrument factor and L is


distance across the suspension.

Figure 4 A sketch of the AZRII The instrument factor B is determined by


measurement cell measuring the signal S with water in the
cell, using the known attenuation factor
the relative motion between the particles for water.
and the liquid, is usually the dominant
cause of attenuation in colloids. It is The acoustic impedance of the
often referred to as viscous suspension is measured by comparing
attenuation because it arises from the the signal that has passed once through
frictional forces in the liquid. the suspension with the signal that has
passed through the suspension three
times, having been reflected internally at
each of the glass/suspension interfaces.

With the instrument factor and zs known,


The attenuation coefficient the formula (7) can be used for

Attenuation Coefficient Spectrum Size Distribution - Differential


for several silica slurries with different particle size Compare several silica slurries

0.5 0.12
0.45
Mean Attenuation Coefficient

0.1
Mean Differential Fraction

0.4
0.35 0.08
0.3
0.25 0.06

0.2
0.04
0.15
0.1 0.02
0.05
0
0
0.001 0.010 0.100 1.000 10.000
0.0 5.0 10.0 15.0 20.0
Diameter [µm] (Logarithmic scale)
Frequency [MHz]

Nissan 0.3 Micron Cabot SCE Nissan 0.3 Micron Geltech 1.0 Micron

Geltech 1.0 micron Klebesol 30N50 Cabot SCE Klebesol 30N50

Figure 5 The measured attenuation coefficient spectra, and the size distributions obtained from
those spectra for the four silica suspensions of table 1.

Colloidal Dynamics www.colloidal-dynamics.com 7


determining the attenuation factor α For the Klebersol and Cabot colloids, the
from the measured S value. particle radius is small and thus the
maximum lies far to the right on the
In the AcoustoSizer II the attenuation graph, outside the measurement range.
measurements are reported in terms of And for the 0.3 micron Nissan the
the non-dimensional attenuation maximum occurs at a frequency that is
coefficient χ , where about nine times higher than that for the
1 micron Geltech sample. With this rule
αλ of thumb you can get a rough idea of the
χ= (8) particle size just by looking at the
φ attenuation coefficient spectrum. Of
course to get accurate size estimates it
Here λ is the wavelength of sound in the is necessary to use software to fit the
solvent, and as before φ is the particle theoretical and measured spectra. Again,
volume fraction. the mathematical process is almost
identical to that used with the dynamic
In left hand graph in figure 5 we show mobility spectra.
the measured attenuation coefficient
spectrum for the silica suspensions used In the right hand graph in figure 5 we
in the dynamic mobility measurements show the size distributions obtained for
of figure 3. these silica sols from the attenuation
measurements.

The trends in these graphs are not as These sizes for these suspensions are
clear as those in the dynamic mobility consistent with those from those
graphs of figure 3. For the very small obtained from electron micrographs.
Klebesol and Cabot particles, the
attenuation coefficient is small and rises Measurements on
monotonically with frequency. For the concentrated colloids
0.3 micron Nissan suspension the
attenuation is still increasing
A number of independent studies have
monotonically with frequency, but it been undertaken to determine the
seems to be reaching a maximum at the accuracy of our ESA and attenuation size
highest frequency, and for the 1 micron
and zeta estimates in concentrated
Geltech silica is reaches a maximum and
colloids. To carry out these tests it is
then decreases with increasing
necessary to compare the concentrated
frequency. measurements with measurements on
dilute samples. Since dilution can easily
The problem of determining the
alter zeta, and sometimes alter size, the
theoretical link between the attenuation
authors in these studies took great care
coefficient and the particle size is very
to dilute the colloid using the same
similar to the dynamic mobility problem, background electrolyte as that in the
and the same methods can be used to
concentrated suspensions. For
obtain the theoretical formula.
suspensions where the particles readily
sediment, this involves decanting the
We will not go into the theory here, but supernatant after the particles have
one useful feature of the theory that settled out and using it to dilute a small
deserves mention is that the maximum
sample of the concentrated suspension.
in the attenuation coefficient as a
For smaller particles it is necessary to
function of frequency occurs at
separate out the supernatant by
filtration or centrifugation. Of course,
ω = O (ν a 2 ) (9) this is only necessary for carrying out
where a is the particle radius and ν is the comparisons with dilute measurements.
To make ZetaProbe and AcoustoSizer II
kinematic viscosity of the liquid.
measurements in practice there is no

Colloidal Dynamics www.colloidal-dynamics.com 8


need to obtain supernatant; you just solution of the same composition as the
measure the suspension as is. water phase.

In figure 6 we show ESA zeta potential From this figure it can be seen that the
measurements obtained by Johson at al ESA yields a consistent size over this
(1998) on a series of alumina slurries. concentration range.
The first measurements were made on a
sample with a particle volume fraction of
30%. This sample was then diluted with
1

supernatant in a series of steps, down to 0.9

3% by volume. At each stage the 0.8

suspension was titrated and the zeta 0.7

determined by ESA measurements. From 0.6

this figure it can be seen that the zeta 0.5

potentials for each concentration are 0.4


consistent, as is expected in systems 0.3
with the same background electrolyte. 0.2
The solid curve in the graph represents 0.1
the zeta values obtained from 0
electrophoresis studies on an extremely 2% 5% 10% 20% 30% 40% 50%
dilute sample. Thus the ESA zeta in this
case is consistent with the
electrophoretic mobility zeta. This is a Figure 7 Median droplet diameter, in
good demonstration of the validity of microns vs droplet volume fraction for an
the electrolyte calibration procedure oil in water emulsion
used in our ESA measurements.

Comparison with other sizing


methods

Our products can only measure down to


a minimum particle concentration of
about 0.5 wt%, so to compare with
other techniques, which require
extremely dilute samples, it is necessary
to dilute in a way that does not alter
particle size. In figures 8 and 9 we show
the comparison of the AcoustoSizer II
size measurements with measurements
on a Malvern PCS device and the Matec
CHDF device; the former product uses
light scattering and the latter uses the
Capillary Hydrodynamic Fractionation
method to get particle size.
Figure 6 Zeta potential curves for an
alumina concentration series.
The results are presented in terms of
three diameters:
In figure 7 we show results from another
careful dilution sequence, this time from q D50-the median size (in
the 2001 paper by Kong et al. The graph microns)
represents the median diameter, in q D15- the diameter that is
microns, of droplets in a sunflower-oil- smaller than 85% of the
in-water emulsion. The droplets were population, and
stabilised with SDS surfactant and then q D85 – the diameter that is
carefully diluted with a surfactant larger than 85% of the
population.

Colloidal Dynamics www.colloidal-dynamics.com 9


measurements, but the variation in
These three parameters give a measure these parameter from batch to batch. It
of the spread of the particle size is vital that changes in the reported
particle size and zeta represent real
160 changes in the suspension and not
variability due to instrument artefacts.
Thus measurement precision is a key
120
property.

80 The ZetaProbe and AcoustoSizer II have


been designed to give excellent
40 measurement precision, and furthermore
they can be used for on-line monitoring
0
of the production stream, since they do
Matec Malvern CD not require sample dilution. Thus these
products have found a place, not only in
R&D and QC labs, but also on the factory
Figure 8 Comparing size distributions for floor where they are used to keep the
DuPont DP106 Silica colloidal product within specifications.
distribution.

The vertical bars in figure 8 represent


the three diameters. These
measurements were carried out on a
suspension of spheroidal silica particles
manufactured by DuPont. The vertical
axis gives the particle size in nm. From
this graph it can be seen that the
AcoustoSizer II size distribution is
similar to that of the Malvern device in
this case, while the Matec gives a bigger
size.

The electron micrograph of this sample


is shown in figure 9.

The “correct” size distribution for this


colloid depends on how the particles are
weighted; for example a size distribution Figure 9 TEM of the DuPont silica
based on particle number will be
different to one that is based on the References
particle mass. And the reported
distribution from the various instruments
1995 O'Brien R. W., Cannon
also depends on the degree of
D.W.Rowlands W.N. J. Colloid Interface
flocculation of the particles. Sci. 173 , 406-418.

The different measuring techniques all


1998 Johnson S.B. Russell A.S., Scales
give the same size when used on
P.J. Colloids and Surfaces A 141, 119-
monodisperse spherical particles, but
130.
this is the ideal case. When the different
techniques are applied to real
suspensions as shown above, they can 2001 Kong, L., Beattie, J.K., Hunter, R.J.
yield different sizes. For engineers who J. Colloid Interface Sci. 238, 70-79.
are controlling particle size and to give
the suspension desired properties, such 2003 O'Brien, R.W., Jones, A., Rowlands,
as colour and rheology, the important W.N. Colloids And Surfaces A 218 89-101
thing is not the absolute value of the

Colloidal Dynamics www.colloidal-dynamics.com 10

You might also like