4000TOC Sensor
M300 TOC Transmitter
4000TOC Sensor
M300 TOC Transmitter
58 130 196 Rev B
4000TOC Sensor 2
Operation Manual
4000TOC Sensor
M300 TOC Transmitter
Content
1 Introduction_______________________________________________________________________________________________ 9
2 Safety instructions________________________________________________________________________________________ 10
2.1 Definition of equipment and documentation symbols and designations___________________________________________ 10
2.2 Correct disposal of the unit _____________________________________________________________________________ 11
3 Unit overview____________________________________________________________________________________________ 12
3.1 Overview, 4000TOC Sensor_____________________________________________________________________________ 12
3.1.1 Sensor Details_________________________________________________________________________________ 12
3.1.2 High Pressure Applications_______________________________________________________________________ 13
3.1.3 Potential Condensation Applications _______________________________________________________________ 13
3.1.4 High Temperature Applications____________________________________________________________________ 13
3.2 Overview, M300 TOC__________________________________________________________________________________ 13
3.2.1 1/4DIN______________________________________________________________________________________ 13
3.2.2 1/2DIN______________________________________________________________________________________ 14
3.3 Control/Navigation Keys________________________________________________________________________________ 14
3.3.1 Menu Structure________________________________________________________________________________ 14
3.3.2 Navigation keys________________________________________________________________________________ 15
3.3.2.1 Navigating the menu tree________________________________________________________________ 15
3.3.2.2 Escape______________________________________________________________________________ 15
3.3.2.3 Enter________________________________________________________________________________ 15
3.3.2.4 Menu________________________________________________________________________________ 15
3.3.2.5 Calibration mode______________________________________________________________________ 15
3.3.2.6 Info mode____________________________________________________________________________ 15
3.3.3 Navigation of data entry fields_____________________________________________________________________ 15
3.3.4 Entry of data values, selection of data entry options____________________________________________________ 15
3.3.5 Navigation with u in Display______________________________________________________________________ 16
3.3.6 ”Save changes” dialog__________________________________________________________________________ 16
3.3.7 Security Passwords_____________________________________________________________________________ 16
3.4 Display_____________________________________________________________________________________________ 16
4 M300 TOC Installation instruction____________________________________________________________________________ 17
4.1 Unpacking and inspection of the M300 TOC________________________________________________________________ 17
4.1.1 Panel cutout dimensional information –
1/4DIN models________________________________________________________________________________ 17
4.1.2 Installation procedure – 1/4DIN models_____________________________________________________________ 18
4.1.3 Panel cutout dimensional information –
1/2DIN models________________________________________________________________________________ 19
4.1.4 Installation procedure – 1/2DIN models_____________________________________________________________ 20
4.1.5 Assembly – 1/2DIN version_______________________________________________________________________ 20
4.1.6 1/2DIN version – Dimension drawings______________________________________________________________ 21
4.1.7 1/2DIN version – Pipe mounting___________________________________________________________________ 21
4.1.8 1/4DIN version – Dimension drawings______________________________________________________________ 22
4.2 Connection of power supply_____________________________________________________________________________ 23
4.2.1 1/4DIN housing (panel mount)____________________________________________________________________ 23
4.2.2 1/2DIN housing (wall mount)_____________________________________________________________________ 24
4.3 Connector PIN definition________________________________________________________________________________ 24
4.3.1 TB1 and TB2 for 1/2DIN and 1/4DIN versions________________________________________________________ 25
4.3.2 TB3 for 1/2DIN and 1/4DIN versions _______________________________________________________________ 25
4.4 Connection of TOC Sensor______________________________________________________________________________ 25
5 4000TOC Installation Instructions____________________________________________________________________________ 26
5.1 Unpacking and inspection of the 4000TOC_________________________________________________________________ 26
5.2 Installation of the 4000TOC_____________________________________________________________________________ 26
5.3 Sample Tubing Connections_____________________________________________________________________________ 27
5.4 Sample Conditioning Coil_______________________________________________________________________________ 28
5.5 AC Power Connection__________________________________________________________________________________ 29
5.6 Sensor Connection____________________________________________________________________________________ 30
6 Placing transmitter in, or out, of service ______________________________________________________________________ 31
6.1 Placing transmitter in service____________________________________________________________________________ 31
6.2 Placing transmitter out of service ________________________________________________________________________ 31
7 4000TOC Sensor Operation_________________________________________________________________________________ 32
7.1 Initial Startup________________________________________________________________________________________ 32
7.2 Setting Sample Flow Rate_______________________________________________________________________________ 32
7.3 Start TOC Measurement________________________________________________________________________________ 33
7.4 Normal Operation_____________________________________________________________________________________ 33
7.5 System Suitability Testing_______________________________________________________________________________ 34
7.6 Calibration__________________________________________________________________________________________ 34
7.6.1 TOC Calibration Methods_________________________________________________________________________ 34
8 Quick Setup______________________________________________________________________________________________ 36
9 Sensor Calibration________________________________________________________________________________________ 37
9.1 Enter Calibration Mode_________________________________________________________________________________ 37
10 Configuration_____________________________________________________________________________________________ 38
10.1 Enter Configuration Mode_______________________________________________________________________________ 38
10.2 Measurement________________________________________________________________________________________ 38
10.2.1 Channel Setup_________________________________________________________________________________ 38
10.2.2 Conductivity/Resistivity Temperature Compensation____________________________________________________ 39
10.2.3 Set Averaging_________________________________________________________________________________ 39
10.3 TOC Setup___________________________________________________________________________________________ 40
10.3.1 TOC General Parameters_________________________________________________________________________ 40
10.3.2 Lamp Control__________________________________________________________________________________ 41
10.3.3 Autobalance__________________________________________________________________________________ 41
10.4 Analog Outputs_______________________________________________________________________________________ 42
10.5 _____________________________________________________________________________________________________
Setpoints ___________________________________________________________________________________________ 43
10.6 Alarm______________________________________________________________________________________________ 46
10.7 Display_____________________________________________________________________________________________ 47
10.7.1 Measurement__________________________________________________________________________________ 47
10.7.2 Resolution____________________________________________________________________________________ 48
10.7.3 Backlight_____________________________________________________________________________________ 48
10.7.4 Name_______________________________________________________________________________________ 48
10.8 Hold Analog Outputs__________________________________________________________________________________ 49
11 System__________________________________________________________________________________________________ 50
11.1 Set Language________________________________________________________________________________________ 50
11.2 USB_______________________________________________________________________________________________ 50
11.3 Passwords__________________________________________________________________________________________ 51
11.3.1 Changing Passwords___________________________________________________________________________ 51
11.3.2 Configuring Menu Access for Operator______________________________________________________________ 51
11.4 Set/Clear Lockout_____________________________________________________________________________________ 51
11.5 Reset______________________________________________________________________________________________ 52
11.5.1 Reset System__________________________________________________________________________________ 52
11.5.2 Reset Analog Calibration_________________________________________________________________________ 52
11.5.3 Reset Sensor Calibration Data to Factory Settings______________________________________________________ 53
11.5.4 Reset Calibration Data of Sensor Electronics to Factory Settings___________________________________________ 53
12 4000TOC Sensor Service Requirements_______________________________________________________________________ 54
12.1 UV Lamp Replacement_________________________________________________________________________________ 54
12.2 High Capacity Inlet Filter Replacement_____________________________________________________________________ 55
13 M300 TOC Service________________________________________________________________________________________ 56
13.1 Enter Service Menu____________________________________________________________________________________ 56
13.2 Diagnostics_________________________________________________________________________________________ 56
13.2.1 Model/Software Revision_________________________________________________________________________ 56
13.2.2 Digital Input___________________________________________________________________________________ 57
13.2.3 Display______________________________________________________________________________________ 57
13.2.4 Keypad______________________________________________________________________________________ 57
13.2.5 Memory______________________________________________________________________________________ 57
13.2.6 Set Relay_____________________________________________________________________________________ 58
13.2.7 Read Relays__________________________________________________________________________________ 58
13.2.8 Set Analog Outputs_____________________________________________________________________________ 58
13.2.9 Read Analog Outputs____________________________________________________________________________ 58
13.3 Calibrate____________________________________________________________________________________________ 59
13.3.1 Verify Conductivity Electronics_____________________________________________________________________ 59
13.3.2 Calibrate Analog_______________________________________________________________________________ 59
13.3.3 Calibrate Unlock_______________________________________________________________________________ 60
13.4 Tech Service_________________________________________________________________________________________ 60
14 Info_____________________________________________________________________________________________________ 61
14.1 Info Menu___________________________________________________________________________________________ 61
14.2 Messages___________________________________________________________________________________________ 61
14.3 Calibration Data______________________________________________________________________________________ 61
14.4 Model/Software Revision_______________________________________________________________________________ 62
14.5 TOC Sensor Info______________________________________________________________________________________ 62
15 Maintenance_____________________________________________________________________________________________ 63
15.1 For Technical Support__________________________________________________________________________________ 63
15.2 Front Panel Cleaning__________________________________________________________________________________ 63
1 Introduction
Statement of intended use – The 4000TOC sensor is a Total Organic Carbon sensor designed to
measure the concentration of organic substances in pure and ultra pure water applications. The
M300 TOC is a single channel transmitter with a large four line backlit Liquid Crystal Display
which conveys measuring and setup information for the 4000TOC sensor.
When used in combination, the 4000TOC sensor and M300 TOC transmitter, form a TOC mea-
surement system which provides the design flexibility for full integration into water systems. The
separate components allow for close positioning of the sensor to the sample point for fastest re-
sponse, while allowing integration of the transmitter into the control system.
This manual applies to the 4000TOC sensor, and M300 TOC transmitter:
4000TOC
Designation Order no.
4000TOC Sensor, 110VAC, 50/60 Hz 58 036 021
4000TOC Sensor, 220VAC, 50/60 Hz 58 036 022
M300 TOC
Designation Order no.
M300 TOC 1/4DIN 58 006 302
M300 TOC 1/2DIN 58 006 312
Throughout this manual, the terms ‘M300’ and ‘M300 TOC’ are used. Within the context of this
document, both terms should be taken to refer to the M300 TOC transmitter.
The print screen images in this manual have a general explaining character and can differ from
the real display in your transmitter.
This description corresponds to the firmware release, version 1.0 of the M300 TOC transmitter.
Changes are taking place constantly, without prior notification
2 Safety instructions
This manual includes safety information with the following designations and formats.
a On the transmitter or in this manual text indicates: Caution and/or other possible hazard in-
cluding risk of electric shock (refer to accompanying documents).
The following is a list of general safety instructions and warnings. Failure to adhere to these in-
structions can result in damage to the equipment and/or personal injury to the operator.
• The M300 Transmitter should be installed and operated only by personnel familiar with
the transmitter and who are qualified for such work.
• The M300 Transmitter must only be operated under the specified operating conditions
(see section 19).
• Repair of the M300 Transmitter must be performed by authorized, trained personnel only.
• With the exception of routine maintenance, cleaning procedures or fuse replacement, as
described in this manual, the M300 Transmitter must not be tampered with or altered in
any manner.
• Mettler-Toledo accepts no responsibility for damage caused by unauthorized modifica-
tions
to the transmitter.
• Follow all warnings, cautions, and instructions indicated on and supplied with this prod-
uct.
• Install equipment as specified in this instruction manual. Follow appropriate local and
national codes.
• Protective covers must be in place at all times during normal operation.
• If this equipment is used in a manner not specified by the manufacturer, the protection
provided by it against hazards may be impaired.
WARNINGS:
a • Installation of cable connections and servicing of this product require access to shock
hazard voltage levels.
• Main power and relay contacts wired to separate power source must be disconnected be-
fore servicing.
• Switch or circuit breaker shall be in close proximity to the equipment and within easy
reach of the OPERATOR; it shall be marked as the disconnecting device for the equip-
ment.
• Main power must employ a switch or circuit breaker as the disconnecting device for the
equipment.
• Electrical installation must be in accordance with the National Electrical Code and/or any
other applicable national or local codes.
• Safety and performance require that this instrument by connected and properly grounded
through a three-wire power source.
h NOTE! RELAY CONTROL ACTION: the M300 transmitter relays will always de-energize on loss of
power, equivalent to normal state, regardless of relay state setting for powered operation. Config-
ure any control system using these relays with fail-safe logic accordingly.
h NOTE! PROCESS UPSETS: Because process and safety conditions may depend on consistent
operation of this transmitter, provide appropriate means to maintain operation during sensor
cleaning, replacement or sensor or instrument calibration.
3 Unit overview
M300 models are available in both a 1/4DIN and 1/2DIN case size. The 1/4DIN is a panel-
mount only design and the 1/2DIN models provide an integral IP65 housing for wall-, or pipe-
mount.
The 4000TOC Sensor and M300 TOC transmitter are configured with factory default settings to
minimize measurement setup time. These default settings allow for sensor operation with mini-
mal keystrokes. The 4000TOC Sensor has more features than other sensors and may require
additional configuration dependant upon the operational setup. Once these parameters are set,
the TOC sensor will function automatically, and will require user interface only for periodic ser-
vice and maintenance. Measurements available from the 4000TOC Sensor include TOC, Con-
ductivity/Resistivity (compensated and uncompensated) and Temperature.
The 4000TOC Sensor is designed with four local LED indicators and a UV Lamp control key on
the front panel, as shown in the illustration below. The LEDs are designed to provide local indi-
cation of sensor status. The operation of the LEDs is synchronized with the M300TOC display
status messages that appear in the 6-digit measurement name location.
The UV Lamp control key is designed to provide a local On/Off control for the UV lamp to allow
for quick lamp change-out and assist the operator during troubleshooting, if necessary.
h NOTE: If the UV Lamp Key is pressed to turn the UV Lamp off, the UV Lamp cannot be turned on
from the M300 TOC, it can only be turned on by pressing the sensor key again.
The front cover of the sensor is hinged on the left side. Located at the top and bottom right cor-
ners of the sensor enclosure are two triangle-shaped door fasteners. The installation kit provid-
ed with the Sensor includes the special tool needed to loosen these fasteners and open the front
door to the enclosure. Periodic access will be required to perform routine service and mainte-
nance. Additional front cover tools are available from Mettler-Toledo Thornton, Inc. See
Section 18 for part numbers and descriptions.
3.2.1 1/4DIN
4.01
[102]
1
8 M300
METTLER TOLEDO
4
3
4.01
5 7 [102]
6
ESC
3.2.2 1/2DIN
5.90
[150]
1
M300
5
3
5.90
[150]
ESC
6 4
Menu Cal Info Enter
2 7 8
Measurement
Mode M300
Set/Clear
Set Flow Alarm/Clean
Lockout
Hold Outputs
h NOTE: In order to back up one menu page, without escaping to the measurement mode,
move the cursor under the UP Arrow character (u) at the bottom right of the display screen and
press [Enter].
3.3.2.2 Escape
Press the and key simultaneously (escape) to return to the Measurement mode.
3.3.2.3 Enter
Use the e key to confirm action or selections.
3.3.2.4 Menu
Press the key to access the main Menu.
Use the key to increase or the key to decrease a digit. Use the same keys to navigate with-
in a selection of values or options of a data entry field.
h NOTE: Some screens require configuring multiple values via the same data field (ex: configuring
multiple setpoints). Be sure to use the or key to return to the primary field and the or
key to toggle between all configuration options before entering to the next display screen.
3.4 Display
h NOTE: In the event of an alarm or other error condition the M300 Transmitter will display a flash-
ing a in the upper right corner of the display. This symbol will remain until the condition that
caused it has been cleared.
h NOTE: During calibrations, clean, Digital In with Analog Output/Relay/USB in Hold state,
a flashing H will appear in the upper left corner of the display. This symbol will remain for
20 seconds until after the calibration or clean is completed. This symbol will also disappear
when Digital In is deactivated.
If there is no apparent damage, unpack the container. Be sure all items shown on the packing
list are present.
If items are missing, notify Mettler-Toledo immediately.
3.62 +0.02
–0
[92 +0.5
–0 ]
3.62 +0.02
–0
[92 +0.5
–0 ]
PANEL CUT-OUT
MOUNTING SCREW
4.01 [90]
[102]
2 PLACES
3.55
MOUNTING SCREW
9
[90] 8
7
6
5
4
3
2
4.01 1
2 PLACES
[90] 1 2 3 4 5 6 7
6
5
4
3
2
1
USB 1 2 3 4 5 6 7
4.01
[102]
MAX. 0.118
[3]
MOUNTING 4.01 FLAT GASKET
BRACKET [102]
2 PLACES
4.96
[126] 4.92
[125]
MAX. 0.118
[3]
NG FLAT GASKET
ET
ES
4.96
1.77
[126] 4.92 [45]
[125] 4.33
[110] 3.54
[90]
1.77
[45]
4.33
[110] 3.54
[90]
Below are cut-out dimensions required by the 1/2DIN models when mounted within a flat panel
or on a flat enclosure door. This surface must be flat and smooth. Textured or rough surfaces are
not recommended and may limit the effectiveness of the gasket seal provided.
5.39+0.02
–0.00
[137 +0.5
–0.0]
5.39+0.02
–0.00
PANEL CUT-OUT
[137 +0.5
–0.0]
Optional hardware accessories are available that allow for panel- or pipe-mount.
Refer to Section 18 “Accessories and Spare Parts” for ordering information.
2
3
136 mm / 5.35"
150 mm / 5.9"
29 mm
6.5 mm
42 mm
0.256"
1.65" 1.14"
84 mm
3.31"
35 mm / 90 mm / 3.54"
1.38" 80 mm / 3.15"
75 mm / 2.95"
138 mm ± 0.5 mm
5.43" ± 0.02"
90 mm / 3.54"
0.236"
6 mm /
138 mm ± 0.5 mm
5.43" ± 0.02"
TB4 TB3
92 mm ± 0.5 mm
Fuse
102 mm / 4.02"
3.62" ± 0.02"
12
Power TB1A
92 mm ± 0.5 mm TB1B
3.62" ± 0.02" USB TB2
102 mm / 4.02"
126 mm / 4.96"
125 mm / 4.92"
a Be sure power to all wires is turned off before proceeding with the installation. High voltage may
be present on the input power wires and relay wires.
A two-terminal connector on the rear panel of all M300 models is provided for power connec-
tion. All M300 models are designed to operate from a 20–30 VDC or a 100 to 240 VAC power
source. Refer to specifications for power requirements and ratings and size power wiring accord-
ingly.
The terminal block for power connections is labeled ”Power” on the rear panel of the transmitter.
One terminal is labeled – N for the Neutral wire and the other + L for the Line (or Load) wire. The
terminals are suitable for single wires and flexible leads up to 2.5 mm2 (AWG 14). There is no
earth ground terminal on the transmitter. For this reason the internal power wiring within the
transmitter is double insulated and the product label designates this using the d symbol.
Power connections are labeled – N for Neutral and +L for Line, for 100 to 240 VAC or 20 – 30 VDC.
⁄4 DIN
1
TB2 for 1⁄4 DIN TB3 ⁄2 DIN
1
TB2 for 1⁄2 DIN
1 AO1+ 9 1 AO1+
2 AO1–/AO2– 2 AO1–/AO2–
3 AO2+ 3 AO2+
1
4 - 4 -
5 - 9 5 -
6 - TB1A 6 - TB1 1 14
1 7
7 - 7 -
8 - TB1B 1 8 - 1 9 1 9 1 9
9 - 1 7 TB2 9 - TB2 TB3 TB4
TB1A for 1⁄4 DIN TB1B for 1⁄4 DIN TB1 for 1⁄2 DIN
1 NO2 1 NO1 1 NO1 8 -
2 COM2 2 COM1 2 COM1 9 -
3 NC2 3 NC1 3 NC1 10 -
4 - 4 - 4 NO2 11 NO3
5 - 5 - 5 COM2 12 COM3
4 NO4 4 NO3 6 NC2 13 NO4
5 COM4 5 COM3 7 - 14 COM4
• 4000TOC Sensor
• 4000TOC Sensor Instruction Manual CD
• 4000TOC Start-up Booklet (Included with CD
• Certificate of Calibration
• Installation Kit includes:
• One - 6 foot (2m) length of PTFE tubing, 0.125 inch (3mm) O.D.
• One - 5 foot (1.5m) length of tubing, 0.31 inch (8mm) O.D.
• One stainless steel drain tube
• One - Plastic 30cc Syringe
• One - Tool, TOC Cover
• One - High Capacity Inlet Filter Assembly
• One - Adapter, 0.25 inch (6mm) O.D. to 0.125 inch (3mm) O.D. Tubing Reducer
• One – Sample Conditioning Coil
Mount the 4000TOC Sensor as close to the sample point of your system as possible. Shorter
sample tubing length between the sampling point and the 4000TOC Sensor will provide faster
response time. As shown above, the 4000TOC Sensor is supplied with wall-mounting tabs for
wall or panel mount. The 4000TOC Sensor can be mounted to a pipe using a pipe-mount ac-
cessory (See Section 18).
When pipe-mounting this sensor, there is risk of sensor damage due to sudden shock or exces-
sive vibration. Be sure the pipe is securely fastened.
The 4000TOC Sensor can also be placed on a bench-top or table. In this type of installation,
the mounting tabs must be removed from the bottom of the sensor enclosure in order for the
sensor to stand on its own. The bottom safety drain fitting should also be removed for bench
top installation, leaving the small opening in the bottom of the enclosure open.
Some of the internal components for the sensor are made of fragile materials; therefore precau-
tions should be taken to avoid damage due to improper handling. The 4000TOC Sensor re-
quires routine maintenance; therefore it is beneficial to mount the sensor in an easily accessible
location.
The following items are necessary to ensure the proper installation of the sensor:
• Sample isolation valve (not provided with sensor).
• 0.125 inch compression tube fitting for sample point (common tube fitting
adapters are available from Thornton; see Section 18.2.1 for details
h NOTE: The sample point should contain a shut-off valve to isolate the sensor when necessary.
Proper installation guidelines should be followed when installing this valve to reduce the possi-
bility of air entrapment or sediment in the sample line.
• Remove the protective covers from the sample connections on the sensor.
• Attached the open tube end of the PTFE tubing to the sample isolation valve. Cut excess
tubing to minimize sample tubing length. Be sure all fittings are properly fastened to
avoid leaks and the possibility of air ingress.
Flush the sample inlet tubing to remove any particles that may be in the line or fittings before
connecting it to the 4000TOC Sensor, or prior to initial sensor startup.
• Connect the stainless steel drain tube to the Sample Outlet connection. The fitting is at-
tached to the tube; therefore thread the fitting into the Sample Outlet connection on the
sensor. Do not over tighten.
• Locate the High Capacity Inlet Filter Assembly provided in the sensor installation kit. The
male-threaded end of the assembly is screwed into the female-threaded sample inlet fit-
ting. Secure the assembly with a wrench to ensure the connection is tight. Do not over
tighten.
• Connect the 0.125 inch (3mm) PTFE tubing to the high capacity inlet filter by securing
the tubing in the compression fitting on the assembly. Do not over tighten.
• Run the safety drain tubing to an atmospheric drain located close to the sensor. See the
illustration in section 5.2, for dimensional detail of the drain location. Take note of the 6”
(15cm) to 36” (90cm) drain requirement.
a Important! The stainless steel drain tube must be fed to a nearby standpipe or drain sump to
create an air gap and prevent siphoning. Installation must adhere to the dimensions shown in
section 5.2 for proper sensor operation.
• Once all tubing connections are complete and the sample inlet tubing has been flushed
to drain, sample water can be introduced to the sensor. The inlet needle valve is used to
vary flowrate and may need adjustment to establish flow. Refer to section 7.2 First, turn
the adjustment knob fully clockwise to a closed position. Then, turn adjustment knob
counter-clockwise three turns. Slowly open the sample point isolation valve until flow is
observed at the sample outlet drain. Once there is flow, check to be sure there are no
leaks inside the enclosure. Allow the sensor to initially flush with sample water (4-24
hrs is recommended). For more details on inlet needle valve operation, see Section 7.2.
Shut off the sample flow to the sensor and connect AC power to the 4000TOC Sensor.
High Capacity
7.5"
Filter REF.
1. Remove compression fitting from filter assembly. As shown, insert port connector into com-
pression fitting that was removed from filter. Reconnect compression fitting to the filter as-
sembly and tighten to crimp. Do not over tighten.
2. Connect Fitting A (1/8”-1/8") of Sample Conditioning Coil to the port connector and tighten
to crimp. Do not over tighten.
3. Attach Fitting B (1/8" X 1/4” NPT) of Sample Conditioning Coil to an appropriate connection
on the process pipe.
4. Ensure that excessive force is not applied to the inlet fittings. If necessary, support the Sam-
ple Conditioning Coil in an appropriate manner.
h NOTE: The Sample Conditioning Coil can be extended in length to 10’ (3m) maximum.
Main power must employ a switch or circuit breaker as the disconnecting device for the equip-
ment. The breaker should be properly labeled and easily accessible for system isolation by ser-
vice personnel. The circuit breaker shall have a maximum of 20 amp rating.
The 'Line' and 'Neutral' input terminal connections for AC power are located on the printed circuit
board mounted to the front cover of the sensor, as shown in the illustration provided below. The
Earth/Ground connection is on the grounding plate mounted at the top of the inside front-cover.
A factory-installed ground wire connects the ground/earth connection at the AC power terminal to
the grounding plate. To ensure safe operation, this wire must not be removed or altered in any
way.
For safe operation, the AC power cable used should meet the following minimum specifications:
VW-1/FT-1; min 300V; 105°C, Minimum 18 AWG (0.821mm2). Cord diameter shall be
Min/Max 0.16”(4.06mm) to 0.31” (7.87mm). The free length of the jacketed cable for
the AC line cord within the TOC sensor enclosure shall be at least 6 inches (150mm).
The free length of the line and neutral leads should be at least 1 inch (25mm) in length
with a recommended strip-length of ¼" (7 mm). The free length of the earth/ground lead
shall be at least 3.5 inches (90 mm), and must be connected to #10 stud located on
the grounding plate.
Keep AC power separated from all other internal wiring. Use the fasteners provided as part of the
AC terminal and grounding plate to secure the power leads. Ensure there is adequate cable
length to avoid putting mechanical stress on the wiring when the front door is fully opened. The
diagram below shows the terminal connections for AC power. Be sure the line fuse is properly
installed when making electrical connections. A spare fuse kit is available from Thornton. See
the spares parts list at the end of this manual for more details.
Once both electrical connections are secured, the sample shut-off valve can be opened and the
sensor flushed with sample water.
This manual defines the TOC functions available when using the 4000TOC Sensor in conjunc-
tion with the M300 TOC transmitter. Read this section carefully prior to operating the
4000TOC Sensor.
a After connecting the transmitter to power supply circuit, it will be active as soon as the circuit is
powered.
To set the flow rate, enter the measurement menu for the measurement displaying the TOC value.
Press Page Down until ’Set Flow Rate’ menu is displayed. Change the ‘No’ to ‘Yes’ to immedi-
ately initiate the ‘Set Flow Rate’ mode at the sensor. It will stay in this mode until the ‘Yes’ is
changed back to ‘No’.
In this mode, the sensor LED lights on the front cover act as a bar-graph type indicator as de-
scribed below. The flow rate is set by adjusting the inlet needle valve installed on the side of the
4000TOC Sensor (see the illustration below.). Turning the adjustment knob counter-clockwise
will increase the flow rate while turning it clockwise will decrease the flow rate. Initially, start
adjustment with the inlet needle valve adjustment knob turned clockwise until there is no flow
observed. Once this is completed, begin turning the adjustment knob counter-clockwise slowly
to gradually increase flow.
The flow rate value is indicated by the four LEDs on the front of the 4000TOC Sensor. When the
flow rate is set to 20 mL/min (±5%), all four LEDs will be on as shown in the table below. If
the flow rate is too high, only the top LEDs will be on (turn the adjustment knob clockwise).
If the flow rate is too low, the bottom LEDs will be on (turn the adjustment knob counterclock-
wise). If the flow rate is far from 20 mL/min, only the top (Fault) LED or the bottom (UV Lamp)
LED will be on. As the flow rate approaches the correct value more LEDs will turn on. Turning
on 3 or 4 LEDs is acceptable. Refer to the table below.
Low Low Low Near Set Set Near Set High High High
Not Illuminated
Illuminated
Return the Set Flow Rate option to ‘No’ in order to leave this mode and this menu. The flow rate
setting operation is now complete.
At this time, the Key on the sensor must be pressed before the UV Lamp will turn on and a TOC
measurement will begin. This process is designed for safety, to insure that an operator is not
working inside the sensor when someone at the M300 TOC instrument is trying to turn on the UV
Lamp.
During the setup of the TOC sensor, if the Auto Start function in the TOC setup menus is changed
from ‘No’ to ‘Yes’, the M300 TOC will still prompt the operator to press the key at the sensor dur-
ing this initial start-up. However after this startup, the Auto Start function takes priority and the
sensor will automatically start.
UV OFF Flashes alternately with measurement name when UV lamp is turned off.
RINSE Flashes alternately with the measurement name when sensor is in Rinse mode.
AUTBAL Flashes alternately with the measurement name when sensor is in the Autobal-
ance mode.
ERROR Flashes alternately with measurement name when an Error exists with the
TOC sensor.
FAULT Continuously displayed in place of measurement name when Fault
condition exists.
If an ERROR or FAULT condition is displayed on the measurement screen, the user can go to the
Messages screen to retrieve more information regarding the cause of the condition. Refer to Sec-
tion 14 of this manual for more information. If the Auto-balance function is set to “yes”, the
4000TOC sensor will perform an auto-balance based on the interval set within the M300 TOC
TOC sensor setup menus.
7.6 Calibration
A Combination System Suitability Test, TOC and Conductivity Calibration Test Kit (p/n 58 091
534) is available for the 4000TOC Sensor. If TOC calibration only is needed, the System Suit-
ability Test and TOC Calibration Kit (p/n 58 091 525) is sufficient to perform this function. The
4000TOC Sensor TOC Calibration SOP (p/n 58 130 095) and 4000TOC Sensor Full Calibration
SOP (p/n 58 130 094) describe these calibrations in detail. Calibration standards (p/n 58 091
529) are also available. Contact Mettler-Toledo Thornton Customer Service for ordering informa-
tion. Note: It is recommended that a TOC calibration be performed after every UV lamp change,
and not to exceed 1 year. A full calibration is recommended annually.
Multi-Linear: For most applications that operate over a wide dynamic range, the recommended
TOC calibration type is Multi-Linear, referred to as “multi” in the M300 TOC. A multi-linear cali-
bration requires two TOC calibration solutions, typically at 250 ppb and 500 ppb. Two calibra-
tion multipliers and adders are determined in the range of 0 250 ppb TOC and 250 -500 ppb
TOC. The two lines intersect at 250 ppb. This calibration features provides an improved system
accuracy. This calibration type is the default calibration method with which the 4000TOC is sup-
plied, and is recommended when low and high TOC may be encountered from either the water
purification process or the calibration/SST solutions.
2-point: A 2-point calibration is intended when both a low and high TOC standard solution is re-
quired for calibration. Selecting a “2-point” calibration will deactivate all other active calibra-
tions.
1-point slope: A 1-point slope calibration method will adjust the TOC calibration multiplier. If a
multi-linear calibration is being used with your 4000TOC and the “1 point slope” is performed, it
will over-write the “multi” calibration.
1-point offset: This method will adjust the the TOC calibration adder. If a multi-linear calibration
exists and the “1 point offset” is performed, it will modify the adder on the 0 – 250 ppb range.
The TOC calibration standard concentration used for the “1-point offset” must be lower than the
midpoint of the “multi” calibration or 50 ppb, whichever is lower. This method is recommended
for applications where low TOC is predominantly expected.
8 Quick Setup
(PATH: Menu/Quick Setup)
Select Quick Setup and press the [ENTER] key. Enter the security code if necessary
(see section 11 “Passwords”).
h NOTE: Pbooklet
lease find the complete description of the Quick Setup routine described in the separate
”Quick Setup Guide for Transmitter M300” enclosed in the box.
h NOTE: Refer to section 3.3 “Control/Navigation Keys” for information on menu navigation.
9 Sensor Calibration
(PATH: Cal)
The calibration key allows the user one-touch access to Sensor calibration and verification
features and Analog Output calibration see chapter 13.3.
h NOTE: During Calibration, a flashing ”H” in the upper left corner of the display indicates a
calibration is in process with a Hold condition active. (The hold output function need to be
activated.) See also chapter 3.4 “Display”
Press the or key to select the type of calibration desired. The choices for each sensor type
are: TOC, C1, C2, T1, T2, Flow, Edit Press [ENTER].
For a detailed explanation on how to perform a calibration in the 4000TOC please refer to the
Standard Operating Proceedure, 58130188.
10 Configuration
(PATH: Menu/Configure)
Configure
Measurement Analog Outputs Set Points Alarm Cal Contact Display Hold Outputs
10.2 Measurement
(PATH: Menu/Configure/Measurement)
Press the [ENTER] key to select this Menu. The following sub menus can now be selected:
Channel Setup, conductivity, and Set Averaging.
Press the [ENTER] key to select the ”Channel Setup” Menu. To select the parameter that will be
displayed on each line.
The options include: TOC in ppb, g C/L, Ω·cm, S/cm, S/m, °C, °F, Lamp Remaining, Nothing.
The 4 lines of the display can now be configured for each line of the display, selecting measure-
ments and unit multipliers. Pressing the [ENTER] key will display the selection for lines c and d.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will
discard the entered values and return to the measurement display screen, selecting Yes will
save changes made.
h
Select Resistivity and press [ENTER].
The temperature compensation mode for any of the four measurement lines configured to display
conductivity can be selected. Choices are ”None”, and ”Standard”.
Standard compensation includes compensation for non-linear high purity effects as well as con-
ventional neutral salt impurities and conforms to ASTM standards D1125 and D5391.
Press the [ENTER] key to select this Menu. The averaging method (noise filter) for each mea-
surement line can now be selected. The options are Special (Default), None, Low, Medium and
High:
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard
the entered values and return to the measurement display screen, selecting Yes will save chang-
es made.
The 'TOC General Para' menu allows the user to configure several basic operational characteris-
tics of the 4000TOC sensor. These functions are explained in detail below.
Auto Start
Auto Start is a feature that allows the 4000TOC Sensor to automatically start making TOC mea-
surements when it is connected to an M300 TOC or when the M300 TOC or 4000TOC restarts
after a power failure. The factory default setting for the Auto Start function is ‘No’. Therefore dur-
ing initial start-up, the M300 TOC will identify the sensor as a 4000TOC Sensor, yet will not start
a TOC measurement. To start a TOC measurement, the user must go to the ‘UV Lamp: On/Off’
screen and set this mode to ‘On’ to turn on the UV lamp and start a TOC measurement. The user
may also push the UV lamp button at the Sensor to start a measurement. Once the ‘Auto Start’
mode is set to ‘Yes’, it is no longer necessary to turn the ‘UV Lamp’ option to ‘On’ as the ‘Auto
Start’ function will override the UV Lamp: On/Off function.
To enable Auto Start, enter the measurement menu for the measurement displaying the TOC val-
ue. Press Page Down until “Push 5 for TOC menu” is displayed. Push 5 and then press Page
Down until the Auto Start function is displayed. Select ‘Yes’ if you want to enable the Auto Start
feature. Select ‘No’ to disable this feature. In the event that direct human interaction is desired
to re-start measurements, this feature should be set to ‘No’.
Conductivity Limit
The Conductivity Limit allows configuring the sensor so that exceeding the sepcified conductivity
limit will alert the operator when the limit is exceeded by giving an error indication and mes-
sage. The limits are selectable, and must be set above 0.0
Over-ride Limit
Related to the Conductivity Limit setting, an error message will occur when the Conductivity Limit
is exceeded when ‘Yes’ is selected. The Conductivity Limit is ignored when ‘No’ is selected.
1. UV Lamp: The 'UV Lamp' function allows the user to turn on or off the UV lamp. For ini-
tial start-up, the sensor key “UV Lamp On” must be pressed.
2. Lamp Limit: The UV lamp in the 4000TOC Sensor is rated for 4500 hours of normal us-
age. The M300 TOC will display an error message when the lamp operating time has
exceeded the lamp limit, which is set to 4500 hours by default. The user can adjust the
Lamp Limit to any value from 400 to 9999 hours. It is the user’s responsibility to assure
by calibration or other means that the lamp is outputting sufficient UV light for the sensor
to make accurate TOC measurements. Remaining lamp life as Indicated by the lamp life
counter can be viewed by pressing the 'Info' key. Please refer to Section 14 for more in-
formation.
3. Lamp Time Reset: During normal operation, the 4000TOC sensor will automatically track
the operating time of the UV Lamp. Once the set lamp limit is exceeded, an error indica-
tion will be given to alert the user that the lamp should be replaced. Following lamp re-
placement, the lamp life counter must be re-set by the user. To re-set the lamp life coun-
ter, enter the Lamp Control menu, select 'Lamp Time Reset = Yes and press the enter key.
This will reset the counter to the Lamp Limit.
10.3.3 Autobalance
(Path: Menu/Configure/TOC Setup/Autobalance)
The 4000TOC Sensor is capable of automatically balancing the two conductivity sensors. This
step is performed to account for small differences in conductivity measurement between the two
conductivity sensors. The default factory condition of the AutoBalance feature of the 4000TOC
Sensor is No. This will result in no AutoBalance activity being performed automatically. The user
may change this attribute to ‘Yes’ at anytime. For applications with very low TOC (<5 ppb) AND
high resistivity (>15 Mohm-cm), it is advised to set this feature to ‘YES’ for improved accuracy.
For applications that do not meet both criteria, such as USP water production, the user may elect
to remain with ‘No’ AutoBalance with little observable effect.
If power to the sensor is interrupted and restored, or if the patch cable is disconnected and then
re-connected, an Autobalance will occur if the Autobalance function is enabled. The user also
has the ability to set or change settings for several functions associated with Autobalance.
These additional functions will not be accessible unless Autobalance is enabled.
3. Autobalance Hold: Set a hold ‘last measurement’ so the analog outputs and relays are
held in their current state while an Autobalance occurs
4. Cycle Time: Set a time interval for the Autobalance to occur automatically. This value is
selectable from 24 to 4500 hours, with a factory default setting of 4500 hours. Note
that the time remaining until the next autobalance occurs can be viewed by pressing the
Info key and selecting 'TOC Sensor Info'. Please refer to Section 14 for additional infor-
mation.
5. Autobal Limit: Set the tolerance for balancing the sensors (limit in %). This is the per-
centage difference allowed between the two conductivity cell readings during an Autobal-
ance cycle when the lamp is off.
6. Rinse Time: Set the length of time that the 4000TOC sensor will have sample water flow-
ing through it before performing the Autobalance function. The Rinse Time can be set
from 1 to 999 minutes. During the rinse and autobalance process, the UV lamp will be
turned off, and the sensor will not provide a TOC indication.
Enter configuration mode as described in Section 10.1 ”Enter Configuration Mode” and navigate
to the menu “Analog Outputs” by using the or . key.
Press the [ENTER] key to select this Menu, which lets you configure the 2 Analog Outputs.
Once analog outputs have been selected, use the and buttons to navigate between config-
urable parameters. Once a parameter is selected, its setting can be selected per the following ta-
ble:
When an Alarm Value is selected, the analog output will go to this value if any alarm condition
occurs.
The Aout type can be Normal, Bi-Linear, Auto-Range or Logarithmic. The range can be 4–20mA
or 0–20mA. Normal provides linear scaling between the minimum and maximum scaling l
imits and is the default setting. Bi-Linear will also prompt for a scaling value for the mid-point
of the signal and allows two different linear segments between the minimum and maximum
scaling limits.
If Auto-range was selected then Aout max1 can be configured. Aout max1 is the maximum
value for the first range on Auto-Range. The maximum value for the second range on Auto-
Range was set in the previous menu. If Logarithmic Range was selected, it will also prompt
for the number of decades as ”Aout1 # of Decades =2”.
The value for the Hold mode can be configured to hold the Last value or can be set to a Fixed
value.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will
discard the entered values and return to the measurement display screen, selecting Yes will
save changes made.
10.5 Setpoints
(PATH: Menu/Configure/Setpoints)
4 Setpoints can be configured on any of the measurements (a thru d). The possible Setpoint
types are Off, High, Low, Outside and Between. Options are also provided for measurement,
%USP, %EP PW, %EP WFI and JP Cond for configuration with the conductivity measurement.
An ”Outside” Setpoint will cause an alarm condition whenever the measurement goes above its
high limit or below its low limit. A ”Between” Setpoint will cause an alarm condition to occur
whenever the measurement is between its high and low limits.
USP and EP Setpoints on Thornton models provide a high alarm used for pharmaceutical water
monitoring with non-temperature compensated conductivity measurements. USP (United States
Pharmacopoeia) General Chapter <645> and European Pharmacopoeia require that non-temper-
ature compensated conductivity of pharmaceutical waters must be below a limit from tables
based
on the temperature of the sample. In other words, pharmaceutical requirements temperature-
compensate the limit rather than the measurement.
The Mettler Toledo Thornton M300 has these pharmaceutical limit tables in memory and auto-
matically determines the conductivity limit based on the measured temperature. USP and EPWFI
(Water for Injection) setpoints use Table 10.1. The limit is the conductivity value corresponding
to the 5 °C temperature step immediately below or equal to the measured temperature value. EP
Highly Purified Water limits are identical to EP WFI limits.
EP PW (Purified Water) setpoints use Table 10.2. The limit in this case is the conductivity value
interpolated for the measured temperature. The M300 takes care of this automatically.
The pharmaceutical setpoint value entered into the M300 is the percentage safety margin
below the limits to activate the setpoint. For example, the USP table conductivity limit at 15 °C
is 1.0 µS/cm. If the setpoint value is set at 40% then the setpoint will activate whenever the con-
ductivity goes above 0.6 µS/cm at 15 °C.
Table 10.1: USP General Chapter <645> Stage 1, EP WFI (Water for Injection), EP Highly Purified
Water, and JP Conductivity Limits as a Function of Temperature.
Enter the desired value(s) for the Setpoint and press [ENTER]
This screen provides the option to configure a setpoint to be active on an over range condition.
Select the setpoint and ”Yes” or ”No”. Select the desired relay that will activate when the setpoint
alarm condition is reached.
Out of Range
Once configured, the selected relay will be activated if a sensor Out of Range condition is detect-
ed on the assigned input channel.
Delay
Enter the delay time in seconds. A time delay requires the setpoint to be exceeded continuously
for the specified length of time before activating the relay. If the condition disappears before the
delay period is over, the relay will not be activated.
Hysteresis
Enter the hysteresis as a percentage-value. A hysteresis value requires the measurement to re-
turn within the setpoint value by a specified percentage before the relay is deactivated.
For a high setpoint, the measurement must decrease more than the indicated percentage
below the setpoint value before the relay is deactivated. With a low setpoint, the measurement
must rise at least this percentage above the setpoint value before the relay is deactivated. For
example, with a high setpoint of 100, when this value is exceeded, the measurement must fall
below 90 before the relay is deactivated.
Hold
Enter the Relay Hold Status of ”Last”, ”On” or ”Off”. This is the state the Relay will go to during a
Hold status.
State
Relay contacts are in normal state until the associated setpoint is exceeded, then the relay is ac-
tivated and the contact states change.
Select ”Inverted” to reverse the normal operating state of the relay (i.e. Normally open contacts
are in a closed state, and normally closed contacts are in an open state, until the setpoint is ex-
ceeded). ”Inverted” relay operation is functional when power is applied to the M300 transmitter.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will dis-
card the entered values and return to the measurement display screen, selecting Yes will
save changes made.
10.6 Alarm
(PATH: Menu/Configure/Alarm)
Using the and buttons, navigate to ”Use Relay #”. Using the or keys, select a relay
to be used for the Alarm and press [ENTER].
If any of these criterias are set to Yes and the conditions for an alarm are given, the flashing
symbol a will be shown in the display, an alarm message will be recorded (see also chapter
14.1 ”Messages”; PATH: Info/Messages) and the selected relay will be activated. Furthermore an
alarm can be indicated by the current output if this has been parameterized (see chapter 10.4
”Analog Outputs”; PATH: Menu/Configure/Analog Outputs)
For 1 and 2 the alarm indicator will be turned off when the alarm message is cleared. It will re-
appear if the power is constantly cycling or if the watchdog is repeatedly resetting the system.
Each Alarm Relay can be configured in either a Normal or Inverted state. Select ”Inverted” to re-
verse the normal operating state of the relay (i.e. Normally open contacts are in a closed state,
and normally closed contacts are in an open state, until an alarm occurs). ”Inverted” relay oper-
ation is functional when power is applied to the M300 transmitter.
In addition, a Delay for the activation can be set. Enter the delay time in seconds. A time delay
requires the alarm to occur continuously for the specified length of time before activating the re-
lay. If the alarm disappears before the delay period is over, the relay will not be activated.
If power failure is turned on, only inverted state is possible and cannot be changed.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will
discard the entered values, selecting Yes will make the entered values the current ones.
h Please note, that there are additional alarms, which will be indicated in the display. See there-
fore in chapter 16 ”Troubleshooting” the different warning- and alarm lists.
10.7 Display
(PATH: Menu/Configure/Display)
This Menu allows for the configuration of the values to be displayed and also the configuration
of the Display itself.
10.7.1 Measurement
The Display has 4 lines. Line 1 on top and Line 4 on the bottom.
Select the ”Error Display” mode. If this is set to ”On” when an alarm has occurred, the message
”Failure – Press Enter” will be displayed on Line 4 when an alarm occurs in the normal Mea-
surement mode.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will
discard the entered values, selecting Yes will make the entered values the current ones.
10.7.2 Resolution
This menu allows the setting of the resolution of each displayed value.
Pressing the [ENTER] key will bring up the Save Changes dialog.
10.7.3 Backlight
This Menu allows the setting of the back light options of the display.
Possible settings are On, On 50% or Auto Off 50%. If Auto Off 50% is selected then the back-
light will go to 50% after 4 minutes with no keypad activity. The backlight will automatically
come back on if a key is pressed.
Pressing the [ENTER] key Will bring up the Save Changes dialog.
10.7.4 Name
This menu allows for the configuration of an alpha-numeric name which is displayed in the first
9 characters on Lines 3 and 4 of the Display. The default is nothing (blank).
If a name is entered on line 3 and/or 4 a measurement can be still displayed on the same line.
Use the and keys to navigate between digits to be altered. Using the and keys to
change the character to be displayed. Once all digits of both display channels have been en-
tered, press [ENTER] to bring up the Save Changes dialog.
The resulting display in the measurement mode appears on Lines 3 and 4 ahead of the mea-
surements.
The ”Hold outputs” function applies during the calibration process. If set ”Hold outputs” to Yes,
during calibration process the analog output, the output relay and USB ouptut will be at hold
state. The hold state depends on the setting. For the possible hold settings, see the list below.
The following options are possible:
The ”DigitalIn” function applies all the time. As soon as a signal is active on the digital input
the transmitter goes to hold mode and the values on the analog output, the output relays and the
USB output will be at hold state.
h NOTE: DDigitalIn2 is
igitalIn1 is to hold channel A
to hold channel B*
11 System
(PATH: Menu/System)
System
While in Measurement mode press the key. Press the or key to navigate to ”System” –
Menu and press [ENTER].
Pressing the [ENTER] key will bring up the Save Changes dialog.
11.2 USB
(PATH: Menu/System/USB)
USB Hold may be set to either Off or Last Values. An external host device may poll the M300
for data. If the USB Hold is set to Off, current values are returned. If the USB Hold is set to Last
Values, the values present at the time the hold condition was established are returned.
11.3 Passwords
(PATH: Menu/System/Passwords)
This Menu allows for the configuration of Operator and Administrator Passwords, as well as set-
ting up a List of allowed Menus for the Operator. The Administrator has rights to access all
Menus. All default passwords for new transmitters are ”00000”.
The Passwords Menu is protected: Enter the Administrator Password to enter the Menu.
Press the [ENTER] key and confirm the new password. Press [ENTER] again to bring up the
Save Changed dialog.
Choose either Yes or No to give/deny access to the above Menus and press [ENTER] to advance
to the next items. Pressing the [ENTER] key after configuring all menus will bring up the Save
Changes dialog. Selecting No will discard the entered values, selecting Yes will make the en-
tered values the current ones.
This menu enables/disables the Lockout functionality of the transmitter. The user will be asked
for a password before being allowed into any menus if the Lockout functionality is enabled.
The Lockout – Menu is protected: Enter the Administrator Password and select YES to enable or
NO to disable the Lockout functionality. Pressing the [ENTER] key after the selection will bring
up the Save Changes dialog. Selecting No will discard the entered value, selecting Yes will make
the entered value the current one.
11.5 Reset
(PATH: Menu/System/Reset)
Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No
will return the user to the Measurement mode with no changes. Selecting Yes will reset
the meter.
Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No
will return the user to the Measurement mode with no changes. Selecting Yes will reset the Ana-
log Output calibration.
Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No
will return the user to the Measurement mode with no changes. Selecting Yes will reset the cali-
bration data of the sensor to factory settings.
h NOTE: To ensure best measuring results, a new calibration of the sensor is recommended after a
reset of the calibration data to factory settings.
Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No
will return the user to the Measurement mode with no changes. Selecting Yes will reset the cali-
Thornton recommends replacement of the UV lamp inside the 4000TOC Sensor after 4500
hours of operation, or 6 months of continuous use. This is a simple procedure that requires only
a few minutes to complete. The following steps explain the proper procedure for the UV lamp
change-out. Refer to the illustration below..
a CAUTION: Use of a UV lamp other than those provided by Mettler-Toledo Thornton specifically for
use with the 4000TOC Sensor will affect performance and void the warranty of this product.
1. At the sensor, turn off the UV lamp by pressing the lamp control button (UV lamp ON LED
will turn off). If the LED does not turn off, check that the Sensor Key Lock is in the off po-
sition in the M300 TOC. Refer to Section 10.3.1 Sensor Key Lock function in this manu-
al.
2. Once power to the UV lamp is off, open the front cover of the sensor enclosure with the
front cover tool.
3. Remove the side cover labeled ‘UV LAMP REPLACEMENT COVER’ on the left side of the
sensor enclosure. Use a large-bladed flat-head screwdriver and turn the cover counter
clockwise to loosen and unscrew the cover.
4. Disconnect the power cable to the UV lamp. This connector is located on the backside of
the front cover, above the circuit board.
5. Loosen the UV lamp holding screw located on the left side of the oxidation chamber.
6. Slide the cable of the UV lamp through the side opening of the enclosure and gently slide
the UV lamp out of the oxidation chamber assembly (stainless steel cylinder). Be careful
not to let the UV lamp hit the quartz glass tube inside the chamber.
7. Use the gloves supplied with each replacement bulb. Hold the new lamp from the ends of
the lamp. Do not touch the bulb. Slide the new UV lamp into the side opening of the en-
closure and into the oxidation chamber opening until it stops. Do not use excessive
force to insert the UV lamp as this may cause damage to the lamp or the internal compo-
nents of the oxidation chamber.
9. Feed the power cable through the side opening of the enclosure. Re-connect it to the
power connector on the front door.
10. Close front cover of the sensor and secure fasteners with the front cover tool.
11. Install the UV Lamp replacement cover on the opening on the side of the enclosure.
12. At the M300 TOC, press the ‘Menu’ key, and select Path: Menu/Configure/TOC Setup/
Lamp Control. Press the ‘Enter’ key until ‘Lamp Time Reset’ is displayed. Select ‘Yes’,
then press enter twice to confirm your selection and save the changes.
13. After the lamp replacement has been completed, a TOC calibration must be performed.
In addition, a System Suitability Test is recommended in applications where compliance
with Pharmacopeia regulations is necessary.
Service
13.2 Diagnostics
(PATH: Menu/Service/Diagnostics)
Enter Service Menu as described in section 13.1 ”Enter Service Menu” and press [ENTER] .
This Menu is a valuable tool for troubleshooting and provides diagnostic functionality for the fol-
lowing items: Model/Software Revision, Digital Input, Display, Keypad, Memory, Set Relays,
Read Relays, Set Analog Outputs, Read Analog Outputs.
13.2.3 Display
All pixels of the display will be lit for 15 seconds to allow troubleshooting of the display. After
15 seconds the transmitter will return to the normal Measuring mode or press [ENTER] to exit
sooner.
13.2.4 Keypad
For keypad diagnostics, the display will indicate which key is pressed. Pressing [ENTER] will re-
turn the transmitter to the normal Measuring mode.
13.2.5 Memory
If Memory is selected then the transmitter will perform a RAM and ROM memory test. Test pat-
terns will be written to and read from all RAM memory locations. The ROM checksum will
be recalculated and compared to the value stored in the ROM.
0 = Normal
1 = Inverted.
13.3 Calibrate
(PATH: Menu/Service/Calibrate)
Enter Service Menu as described in section 13.1 ”Enter Service Menu”, select Calibrate, and
press [ENTER].
This menu has the options to calibrate the transmitter and the analog outputs and also allows
the unlocking of calibration functionality.
h NOTE: Fmeasurements,
or detailed instructions on calibration of the flow, temperature, conductivity and TOC
please refer to Standard Operating Procedure XX XXX XXX
Connect an accurate milliamp meter to the Analog output terminals and then adjust the five
digit number in the display until the milliamp meter reads 4.00 mA and repeat for 20.00 mA.
As the five digit number is increased the output current increases and as the number is de-
creased the output current decreases. Thus coarse changes in the output current can be
made by changing the thousands or hundreds digits and fine changes can be made by chang-
ing the tens or ones digits.
Pressing the [ENTER] key after entering both values will bring up a confirmation screen.
Selecting No will discard the entered values, selecting Yes will make the entered values the
current ones.
Selecting Yes means that the conductivity verification menu (see chapter 13.3.1 ”Verify Conduc-
tivity Electronics”) and Analog Output calibration menu (see chapter 13.3.2 ”Calibrate Analog”)
will be selectable under the CAL Menu. Selecting No means that only the Sensor calibration is
available under the CAL Menu. Press [ENTER] after the selection to display a confirmation
screen.
h NOTE: This Menu is for Mettler Toledo Service personnel use only.
14 Info
(PATH: Info)
Info
Pressing the key will display the Info Menu with the options Messages, Calibration Data and
Model/Software Revision.
14.2 Messages
(PATH: Info/Messages)
Enter Info Menu as described in section 14.1 ”Info Menu” and press [ENTER].
The most recent message is displayed. The up and down arrow keys allow scrolling through
the last four messages that have occurred.
Clear Messages clears all the messages. Messages are added to the message list when the con-
dition that generates the message first occurs. If all messages are cleared and a message con-
dition still exists and started before the clear then it will not appear in the list. For this message
to re-occur in the list the condition must go away and then reappear.
Enter Info Menu as described in section 14.1 ”Info Menu”, select Calibration Data, and press
[ENTER].
The menu displays the calibration constants for each parameter within the TOC sensor. Use the
up and down arrow keys to select TOC, Conductivity, Temperature or Flow
1. TOC: Will display sensor type, last TOC calibration date as entered during the calibration,
sensor firmware version, sensor Serial Number and sensor Part number
3. Autobalance: Displays time remaining until next autobalance, and the current autobal-
ance offset value. If the Autobalance function is turned off, the time remaining will be
displayed as 'N/A'
4. Flow: Displays current measured flow rate through the sensor will be displayed in ml/
min, and the last flow sensor calibration date, as entered during the calibration.
5. C1 (C2): Displays sensor serial number and last C1 (C2) calibration date, as entered
during the calibration.
Enter Info Menu as described in section 14.1 ”Info Menu”, select ISM Sensor Info, and press
[ENTER].
After plugging in a TOC sensor, information about the various portions of the TOC sensor will be
shown in this menu. Use up and down arrows to scroll through the menu. To select TOC, C1,
C2, Flow, auto balance or UV Lamp Time Remaining.
Once selected press the {ENTER} to view additional details pertaining to the selected item.
15 Maintenance
If the power consumption of the M300 transmitter is too high or a malfunction leads to a
short circuit the fuse will blow. In this case remove the fuse and replace it with one specified
in Section 18 “Accessories and Spare Parts”.
Furthermore the detection of some alarms can be activated or deactivated (see chapter 8.5
“Alarm”; PATH: Menu/Configure/Alarm) for an indication on the display. If one of these alarms
occurs and the detection has been activated, the flashing symbol a will be shown on the dis-
play. The message will be recorded through the menu Messages
(see chapter 14.1 “Messages”; PATH: Info/Messages).
Alarms, that are caused by a violation of the limitation of a setpoint or the range (see chapter
8.4 “Setpoints”; PATH: Menu/Configure/Setpoint) will be shown by a flashing symbol a and re-
corded through the menu Messages (see chapter 14.1 “Messages”; PATH: Info/Messages).
According to the parameterisation of the transmitter the hint ”Failure – Press Enter” will be shown
at line 4 of the display if a warning or alarm has occurred (see chapter 10.7 “Display”;
PATH: Menu/Configure/Display/Measurement).
No Flow from sensor sample outlet • High capacity inlet filter clogged.
• Inlet needle valve not adjusted properly.
Erratic flow through the sensor • Sample outlet tubing not installed per manufacturer’s
instructions.
• Inlet needle valve failure or not adjusted properly.
• Erratic water system/sample feed pressure
An Error is generated by a condition that may cause a problem with the proper operation of the
sensor. Under an error condition, the sensor will continue to make measurements but the mea-
sured value may be in error depending upon the cause of the error message.
Fuse, 1.25A, Sensor PCB (For use on both 110VAC and 220 VAC Models) 58 091 520
System Suitability Standards (For use with SST KIT #58 091 559) 58 091 522
Solutions, Calibration 4000TOC / 5000TOC (For use with SST KIT #58 091 58 091 523
Combined Calibration and SST Solutions (For use with SST KIT #58 091 58 091 524
19 Specifications
19.1 4000TOC
Measurement Range 0.05 - 1000 μgC/L (ppbC)
Accuracy ± 0.1 ppb C for TOC < 2.0 ppb (for water quality > 15 MΩ-cm)
± 0.2 ppb C for TOC > 2.0 ppb and < 10.0 ppb (for water quality >
15 MΩ-cm)
± 5% of measurement for TOC > 10.0 ppb (for water quality 0.5 to
18.2 MΩ-cm)
Conductivity Sensor
Conductivity Accuracy ±1%, 0.02 to 20 μS/cm; ±3%, 20-100 μS/cm
Constant Accuracy 2%
Temperature Sensor Pt1000 RTD, Class A
Temperature Accuracy ± 0.25°C
General Specifications
Overall Dimensions 11" [280mm] W x 8.8" [188mm) H x 5.25" [133mm) D
Sample Connections
Inlet 0.125" [3mm] O.D. (6' [2m] FDA compliant PTFE tubing supplied)
Outlet Stainless Steel drain tube
(5' [1.5m] flexible tubing provided)
Inlet Filter 316SS, inline 60 micron
Weight 5.0 lb. [2.3 kg]
Enclosure material Polycarbonate plastic, flame retardant, UV and chemical resistant
UL # E75645, Vol.1, Set 2, CSA #LR 49336
Wetted parts 316SS/Quartz/PEEK/Titanium/PTFE/Silicone/FFKM
Ambient Temperature/ 5 to 50°C / 5 to 80% Humidity, non-condensing
Humidity rating
Power requirements 100 - 130VAC or 200 - 240VAC, 50/60 Hz, 25W Maximum
Wall Mount Standard, mounting tabs provided
Pipe Mount Optional, with pipe-mount bracket accessory
(for nominal pipe sizes 1" [2.4 cm] to 4" [10 cm])
Maximum Sensor Distance 300ft [91m]
Local Indicators Four LED lights for Fault, Error, Sensor Status and UV Lamp ON
Ratings/Approvals CE Compliant, UL and cUL (CSA Standards) listed.
Conductivity and temperature sensors traceable to NIST and
ASTM D1125 and D5391
Meets ASTM D5173 Standard Test Method for On-Line Monitoring of
Carbon Compounds in Water by UV Light Oxidation
* Temperature above 70°C requires Sample Conditioning Coil p/n 58 079 511 (included).
** For power plant cycle chemistry samples, pH may be adjusted by measurement after cation exchange.
*** Process pressure above 85 psig (5.9 bar) requires optional High Pressure Regulator p/n 58 091 552.
20 Default tables
21 Certificate
Mettler-Toledo Thornton, Inc., 900 Middlesex Turnpike, Building 8, Billerica, MA 01821, USA
has obtained Underwriters Laboratories’ listing for M300 Model Transmitters. They bear the
cULus Listed mark, signifying that the products have been evaluated to the applicable ANSI/UL
and CSA Standards for use in the U.S. and Canada
Declaration of Conformity
We,
Mettler-Toledo Thornton, Inc.
900 Middlesex Turnpike, Building 8
Billerica, MA 01821, USA
Declare Under our sole responsibility that the product:
4000TOC Sensor
to which this declaration relates, in conformity with the following European, harmonized and
published standards at the date of this declaration:
EMC Emissions and Immunity: EN 61326-1 2006, Measurement Control and Laboratory equip-
ment EMC requirements.
UL LISTING
22 Warranty
METTLER TOLEDO warrants this product to be free from significant deviations in material and
workmanship for a period of one year from the date of purchase. If repair is necessary and not
the result of abuse or misuse within the warranty period, please return by freight pre-paid and
amendment will be made without any charge. METTLER TOLEDO”s Customer Service Dept. will
determine if the product problem is due to deviations or customer abuse. Out-of-warranty prod-
ucts will be repaired on an exchange basis at cost.
The above warranty is the only warranty made by METTLER TOLEDO and is lieu of all other war-
ranties, expressed or implied, including, without limitation, implied warranties of merchantability
and fitness for a particular purpose. METTLER TOLEDO shall not be liable for any loss, claim,
expense or damage caused by, contributed to or arising out of the acts or omissions of the Buy-
er or Third Parties, whether negligent or otherwise. In no event shall METTLER TOLEDO’s liability
for any cause of action whatsoever exceed the cost of the item giving rise to the claim, whether
based in contract, warranty, indemnity, or tort (including negligence).
ISO ISO
9001 14001
certified certified
www.mt.com
Designed, produced
and controlled according to
ISO 9001 / ISO14001
Subject to technical changes. Mettler-Toledo Thornton, Inc.
© Mettler-Toledo Thornton, Inc. 900 Middlesex Turnpike
05/14 Printed in USA. Billerica MA, 01821
58 130 196 Rev B Tel. +1-781-301-8600, Fax +1-781-301-8701