1 Contents
2 Overview _____________________________________________________________________________ 5
3 Safety and Warranty Registration __________________________________________________________ 6
3.1 Safety Instructions _________________________________________________________________ 6
3.2 Warranty Registration _______________________________________________________________ 6
4 Technical Specifications _________________________________________________________________ 7
5 Installation ____________________________________________________________________________ 8
5.1 Site Assessment ___________________________________________________________________ 8
5.2 Unpacking and Inspection ___________________________________________________________ 9
5.3 Assembly _______________________________________________________________________ 10
5.3.1 Installing the Rotating Arm ______________________________________________________ 10
5.3.2 Installing the Ribbon Cable ______________________________________________________ 10
5.4 Mounting ________________________________________________________________________ 11
5.5 Plumbing ________________________________________________________________________ 11
5.6 Electrical Connections _____________________________________________________________ 12
5.7 4-20 mA Connections and Dry Contacts _______________________________________________ 12
6 Start-Up_____________________________________________________________________________ 13
6.1 Start-up Instructions _______________________________________________________________ 13
7 Operator Interface _____________________________________________________________________ 13
8 Zeroing (Calibration) ___________________________________________________________________ 14
8.1 Using RO or DI Water Method _______________________________________________________ 14
8.2 Using Known Water Method _________________________________________________________ 15
9 Menu Structure _______________________________________________________________________ 16
10 4-20 mA Setup _______________________________________________________________________ 20
11 RS-232 Data Logging __________________________________________________________________ 21
12 Slope Correlation Method _______________________________________________________________ 23
13 Dehumidifier for Humidity Control _________________________________________________________ 24
14 Maintenance _________________________________________________________________________ 25
14.1.1 Cleaning Frequency ___________________________________________________________ 25
Cleaning Fluid ________________________________________________________________________ 26
14.2 Cleaning Quartz Flow Cell __________________________________________________________ 26
14.2.1 In-situ Quartz Flow Cell Cleaning Method __________________________________________ 27
14.2.2 Manual Quartz Flow Cell Cleaning Method and Quartz Flow Cell Removal ________________ 28
14.3 Ribbon Cable Replacement _________________________________________________________ 29
15 Dual Feed ___________________________________________________________________________ 30
List of Tables
List of Illustrations
2 Overview
Thank you for purchasing from Real Tech Inc.
Real Tech’s Colour C4100 analyzer works on the principle of visible light absorbance at a 390 or 450nm
wavelength. Colour found in water and wastewater absorb light at this frequency, which makes their
identification possible. Compounds that may interfere with colour measurement, such as particles (turbidity), are
compensated for using an additional reference wavelength at 870 nm when needed.
Serial Number*
Please read and follow all safety instructions outlined in this owner’s manual prior to
installation and/ or operation.
To activate your standard two-year warranty for your product, the end user of the product is responsible for
filling out the warranty registration information for their company. Failure to do so could void the warranty.
See the warranty section for full details on Real Tech’s two-year warranty program.
4 Technical Specifications
The Real Colour C4100’s Split-Sense Pro technology provides affordability, increased accuracy, and reduced
maintenance.
5 Installation
5.1 Site Assessment
Prior to installation, the site should be assessed to collect all relevant information that may affect the installation
or operation of the system.
Real Tech has provided a checklist that can be used to ensure any installation or operational problems can be
addressed, resulting in the most efficient and effective installation and startup of the system.
CHECK TIP
1 System installed outdoors? Avoid areas with high exposure to sun and weather if possible.
CHECK TIP
If necessary, use a needle valve to lower flow and maintain less than 20
Pressurized system able to
10 PSI in analyzer. Splice connection from the side of the pressurized pipe, to
supply 300-1000 ml/min flow?
avoid air bubbles from top and sediment on bottom.
Select Dual Feed System to switch from process flow to a second stream
13 Is process flow intermittent?
such as tap water.
ITEM QUANTITY
Real Colour C4100 Analyzer 1
Rotating Arm Assembly 1
1L Calibration Bottle with ¼” Flex Tubing 1
24 VDC Power Adapter and Cable 1
Dehumidifier 1
Please ensure all items are unpacked and accounted for before moving on to assembly.
5.3 Assembly
2. Press the connector down until you feel it click into place
and then push in the hold down clamps in towards the
connector.
3. Connect the other end of the cable to the slot on the main Figure 5-3: Tighten set screw
control board and then secure the top and bottom hold
down clamps against the connector.
Press clamp in
Press clamp in
5.4 Mounting
Use screws and washers that are sturdy enough to support the
cabinet weight of approximately 20 lbs.
Note: If installing with a Real Clean System, Pump Clean System, Pump System, Dual Feed or Debubbler refer
to the accessory product owner's manual for mounting location in relation to the analyzer.
5.5 Plumbing
Dual Feed Connection – For the dual feed option, a 7 Clean Control Out
second pair of 4-20 mA output connections are from 8 Clean Control Out
negative terminal #1 and positive terminal #2. This 9 24VDC Out – (Black) (for use with Real Clean)
output signal is for the water stream 2.
10 24VDC Out + (Red) (for use with Real Clean)
Automatic Chemical Clean Connection - Power to 11 24VDC In - (Black)
the optional Automatic chemical clean system is
12 24VDC In + (Red)
provided at connector terminals #9 and #10, which run
to the left terminal block on the circuit board in the
chemical clean cabinet. Matching the correct polarity
Note: Terminal numbering from left to right
is important. The chemical clean logic is through connector
terminals #7 & #8, which run to the right terminal block on
the circuit board in the chemical clean cabinet. Polarity is not
important.
6 Start-Up
After the analyzer has been installed as described in the Installation section, operation of the analyzer is very
straight-forward.
1. Inspect the inside of the analyzer to ensure all packing material (which may restrict moving parts inside
the analyzer), has been removed.
2. Ensure the 24 VDC power wires are securely attached to the correct corresponding terminals. Do not
plug in the analyzer at this time.
3. After making sure the flow rate has been set to 0.3 – 1.0 L/min and the drain is unrestricted and not
kinked, open the inlet valve to allow test water to flow through the analyzer and check for any leaks in
the flow cell, piping, and fittings for at least 10 minutes before proceeding to the next step. Ensure there
are no bubbles flowing through the flow cell and no water droplets are on the inside of the analyzer.
4. Plug in the power supply. Display screen will read “Initializing”.
5. Calibrate your analyzer. Measurements are not assumed to be accurate until the analyzer is calibrated.
See Section 8, Zeroing (Calibration).
6. Configure the analyzer to your application i.e. enter warning and alarm setpoints, dry contact, 4-20mA
set-up, data logging. See Section 9, Menu Structure.
IMPORTANT: In areas of high humidity, the analyzer door must remain closed for a minimum of three
hours to allow for the dehumidifier to dry the air inside the enclosure. (Refer to Section 13). If you
receive a Humidity alarm while calibrating, the door may not have been closed long enough. Wait until
the alarm is removed, and then start calibration again.
7 Operator Interface
8 Zeroing (Calibration)
The analyzer must be calibrated before accurate colour data can be read. Two methods for calibration can be
used:
• Zero analyzer to DI or RO water
• Zero analyzer to a known water value
The first method is using DI or RO water as the standard by which the analyzer is zeroed, whereas the second
method uses a known concentration. Both are accurate, but the second method allows for greater accuracy
because the expected measured values are presumably clustered around the known value instead of falling
within a wider range in the measurement scale.
3. A default value is displayed in the ‘New PCU:’ field (in this example, 75.0).
4. Press the up arrow to increase the number above 75.0 or the down arrow to decrease the number
below 75.0. When you have keyed in the number of known water transparency, press Enter.
9 Menu Structure
MENU ITEM DESCRIPTION
AVERAGING
On: 1 Off: 0
Average Press ‘Enter’ then up/down to select running average for last 10 readings (Default = 0: Off)
%T/Abs MODE
0 : %T
1 : Abs
2 : Colour
Select Press ‘Enter’ then up/down to select %T, Abs, or Correlation mode (Default = 0: %T)
SETPOINTS
%T Warning Press ‘Enter’ then up/down to select %T warning set point (Default = 75)
%T Alarm Press ‘Enter’ then up/down to select %T alarm set point (Default = 50)
Abs Warning Press ‘Enter’ then up/down to select Abs warning set point (Default = 0.20)
Abs Alarm Press ‘Enter’ then up/down to select Abs alarm set point (Default = 0.30)
Colour Warning Press ‘Enter’ then up/down to select Colour warning set point (Default = 160)
Colour Alarm Press ‘Enter’ then up/down to select Colour alarm set point (Default = 200)
4-20mA SETUP
4mA %T Press ‘Enter’ then up/down to select 4mA %T reading for scaling (Default = 0)
20mA %T Press ‘Enter’ then up/down to select 20mA %T reading for scaling (Default = 100)
4mA Abs Press ‘Enter’ then up/down to select 4mA Abs reading for scaling (Default = 0)
20mA Abs Press ‘Enter’ then up/down to select 20mA Abs reading for scaling (Default = 1)
4mA Colour Press ‘Enter’ then up/down to select 4mA Colour reading for scaling (Default = 0)
20mA Colour Press ‘Enter’ then up/down to select 20mA Colour reading for scaling (Default = 200)
SECONDARY DATA
Lamp Displays the direct lamp output
Water Displays the water output after it passes through the flow cell
VIS Lamp Displays VIS direct output
VIS Water Displays VIS output after passing through flow cell
Temperature Displays temperature inside cabinet – Celsius
Humidity Displays the relative humidity inside cabinet – RH
Status Code Displays status code for system diagnostics
DRY CONTACT SETUP
See Set Alarms
0: Nothing
1: Alarm
Select Press ‘Enter’ then up/down to select 0 or 1 to turn on/off dry contact (Default = 0: Nothing)
* Factory setting - unit specific value set at factory. Value can be located on the label on the inside of the cabinet
door.
Datalog Frequency
0: No Data Log
1: 10 seconds
2: 1 minute
3: 10 minutes
4: 1 hour
Select Press ‘Enter’ then up/down 1 - 4 to select logging frequency (Default = 1: 10 seconds)
ALARMS
Humidity Alarm status (0 = no alarm 1 = alarm)
Temperature Alarm status (0 = no alarm 1 = alarm)
Leak Detection Alarm status (0 = no alarm 1 = alarm)
%T Warning Alarm status (0 = no alarm 1 = alarm)
%T Alarm Alarm status (0 = no alarm 1 = alarm)
Abs Warning Alarm status (0 = no alarm 1 = alarm)
Abs Alarm Alarm status (0 = no alarm 1 = alarm)
Colour Warning Alarm status (0 = no alarm 1 = alarm)
Colour Alarm Alarm status (0 = no alarm 1 = alarm)
Lamp Failure Alarm status (0 = no alarm 1 = alarm)
Lamp Low Alarm status (0 = no alarm 1 = alarm)
Lamp High Alarm status (0 = no alarm 1 = alarm)
Cleaning Alarm status (0 = no alarm 1 = alarm)
Clean Fld Empty Alarm status (0 = no alarm 1 = alarm)
Clean Fluid Low Alarm status (0 = no alarm 1 = alarm)
System Fault Alarm status (0 = no alarm 1 = alarm)
SET ALARMS
See Dry Contact
0: Disabled
1: Alarm
Humidity Press ‘Enter’ then up/down to select high cabinet humidity alarm (Default = 1)
Temperature Press ‘Enter’ then up/down to select high cabinet temperature alarm (Default = 1)
Leak Detection Press ‘Enter’ then up/down to select leak inside cabinet alarm (Default = 1)
%T Warning Press ‘Enter’ then up/down to select low %T warning point as alarm (Default = 0)
%T Alarm Press ‘Enter’ then up/down to select low %T alarm point as alarm (Default = 0)
Abs Warning Press ‘Enter’ then up/down to select high Abs warning point as alarm (Default = 0)
Abs Alarm Press ‘Enter’ then up/down to select high Abs alarm point as alarm (Default = 0)
Colour Warning Press ‘Enter’ then up/down to select high Colour warning point as alarm (Default = 0)
Colour Alarm Press ‘Enter’ then up/down to select high Colour alarm point as alarm (Default = 0)
Lamp Failure Press ‘Enter’ then up/down to select lamp failure alarm (Default = 1)
Lamp Low Press ‘Enter’ then up/down to select low lamp output alarm (Default = 1)
Lamp High Press ‘Enter’ then up/down to select high lamp output alarm (Default = 1)
Cleaning Press ‘Enter’ then up/down to select when cleaning cycle active (Default = 0)
Clean Fld Empty Press ‘Enter’ then up/down to select when cleaning solution empty alarm (Default = 0)
Clean Fluid Low Press ‘Enter’ then up/down to select when cleaning solution low alarm (Default = 0)
System Fault Press ‘Enter’ then up/down to select internal system error alarm (Default = 1)
MUTE
Alarms Press ‘Enter’ then up/down (0 or 1) to select audible alarms (Default = 0)
DUAL FEED (Only used with Dual Feed units)
Enable Press ‘Enter’ then up/down to select 0 to disable or 1 to enable (Default = 0)
Press ‘Enter’ then up/down to select number of minutes to wait while switching between sample
Idle Time lines
(Default = 2)
Press ‘Enter’ then up/down to select minutes for total dual feed cycle time (15 - 60 minutes)
Cycle Time
(Default = 30)
DARK CURRENT *Factory setting
Set Value Press ‘Enter’ then up/down to adjust the dark current
SET POTENTIOMETERS *Factory setting
Set Water Press ‘Enter’ then up/down to adjust the water potentiometer
Set Lamp Press ‘Enter’ then up/down to adjust the lamp potentiometer
CORRELATION
(Default = 100)
CALCULATION
Press ‘Enter’ then up/down to adjust the linear correlation factor for the correlated measurement
Slope 1
– single feed
Press ‘Enter’ then up/down to adjust the linear correlation factor for the correlated measurement
Slope 2
– dual feed
TURBIDITY COMP.
Enable Press ‘Enter’ then up/down to select 0 or 1 to enable
Weighting ** (Default = 100)
Dark Current 0
Pot Value 245
* Factory setting - unit specific value set at factory. Value can be located on the label on the inside of the cabinet
door.
10 4-20 mA Setup
The analyzer is equipped with a 4-20 mA output capability.
The 4-20 mA can be set to produce a user selectable PCU
value for the low (4 mA) and a separate value for high (20
mA) up to 100 PCU.
Figure 10-1 below is an example of a 4-20 mA setup for a
customer who wanted to log colour values between 25 and 55
PCU.
Refer to 4-20 mA setup in the menu system.
9-pin serial
Note that the communication protocol for RS232 is ASCII. An example of one data stack is shown here based
on the use of optional turbidity compensation:
The order of the data stack in ASCII (from left to right) is: %T (or Abs) | Lamp | Water | VIS Lamp | VIS Water | |
Temperature | Humidity | Status. The output value can be %T, Abs, or Colour mode, which is selected via the
Abs/%T Mode in the Analyzer menu in Section 9.
Table 5 itemizes each parameter and its corresponding value. Without turbidity compensation, no VIS Lamp or
Water values would be listed.
Table 5: Data Logging - ASCII string description for Analyzer
PARAMETER VALUE
%T 91.2
LAMP 27650
WATER 25300
TEMPERATURE 31.0
HUMIDITY 43.3
STATUS 150
If your system has a Dual Feed accessory (and optional turbidity compensation is still in effect), the ASCII string
would be slightly different. %T is replaced by %T 1 and 2, which correspond to Stream 1 and 2, and %
Removal. An example of one row of characters is shown below:
The order of the data stack in ASCII (from left to right) is: %T1 | %T2 | % Removal | Lamp | Water | VIS Lamp|
VIS Water | Temperature | Humidity | Status
Table 6 itemizes each parameter and its corresponding value. Without optional turbidity compensation, no VIS
Lamp or Water values would be listed.
Table 6: Data Logging - ASCII string description for Analyzer with Dual Feed
PARAMETER VALUE
%T 1 (Stream 1) 91.2
%T 2 (Stream 2) 95.0
% REMOVAL 3.8
LAMP 27650
WATER 25300
TEMPERATURE 31.0
HUMIDITY 43.3
STATUS 150
1. Make sure the main display screen shows the absorbance (Abs) value of the water currently flowing
through the analyzer. Record the Abs value (Abs Value).
3. Using a lab or a portable test instrument, determine the colour concentration for the water sample (Lab
Value).
5. Alternatively, the slope can be determined via multiple test points over the expected testing range of the
analyzer. In this case, the slope can be calculated simply using a spreadsheet package (such as Microsoft
Excel) to plot a best fit line through several test points.
Note: Analyzers with Dual Feed option use two correlation factors (Slope 1 for stream 1 and Slope 2 for stream
2).
14 Maintenance
A standard recommended maintenance schedule is as follows:
FREQUENCY TASK
Once in operation, observe the flow cell for signs of fouling prior to the first cleaning cycle. If fouling is present,
increase the cleaning frequency accordingly. The measured data can also be observed for signs of a “sawtooth”
effect in Abs mode as shown in Figure 14-1. If the sawtooth effect in Abs mode shows a upward trend as
displayed in Figure 14-2, the sensor is drifting, and cleaning frequency must be increased.
Figure 14-1: Absorbance Mode – Effective cleaning Figure 14-2: Absorbance Mode – Ineffective cleaning
Cleaning Fluid
Selecting an effective cleaning solution is essential for optimal performance. The following commercial cleaning
solutions and acids are readily available to remove common fouling agents.
Lactic Acid <20% solution (CLR®) Removes lime, calcium, rust, magnesium, and other dissolved minerals
Sulfamic Acid <10% solution (Lime Away®) Removes lime, calcium, rust, magnesium, and other dissolved minerals
Phosphoric Acid <30% solution (Rust Off,
Removes lime, calcium, rust and color staining
Hagasen Blue)
Citric Acid <20% solution Removes mineral scaling
The optimal cleaning method for the analyzer is an automatic scheduled clean with the Real Clean System II.
Without the Real Clean System II, the cleaning process is similar to the calibration methods for the analyzer.
Cleaning of the quartz flow cell can be performed using either the in-situ method or the manual method. The in-
situ method is recommended since it is much faster and easier and just as effective.
1. Shut off the water supply to the analyzer if not already done.
3. Fill the supplied wash bottle with cleaning fluid and connect
the ¼” bottle tubing to the inlet connection on the analyzer.
4. Hold a finger over the small vent hole in the wash bottle cap,
as shown in Figure 14-3.
5. Squeeze the bottle to get the cleaning fluid to flow through the
analyzer.
6. Once the fluid starts to flow, remove finger from the vent hole
and place bottle on top of the analyzer, as shown in Figure
14-4. The fluid should continue to flow through the analyzer. If
it stops go back to step 4. Figure 14-3: Setting up flow
7. Let all the fluid run through the quartz flow cell until the bottle
is empty, then disconnect the hose and bottle from the fitting.
8. Let the fluid remain in the quartz flow cell for 5-10 minutes
until the quartz flow cell is clean.
9. Connect the supply line to the unit and open the valve to start
the flow running.
10. If absorbance values do not come back in line with the known
absorbance value, it is recommended to re-zero the analyzer.
14.2.2 Manual Quartz Flow Cell Cleaning Method and Quartz Flow Cell Removal
1. Shut off the power to the analyzer and close the water supply.
2. Open the door of the analyzer and disconnect the two black
hoses from the flow cell tube as shown in Figure 14-5.
3. Slowly pull the flow cell tube from the two spring clips straight
out.
6. Place the quartz windows back in the nut (ensuring that there
are no fingerprints on both quartz windows where the light path
comes through).
7. Reassemble the flow cell tube with windows, O-ring, and nuts.
8. Push flow cell tube back into the spring clicks and reconnect
the black tubing.
Note: A spare cable is included with every new analyzer and is attached to the inside of the cabinet door. To
order more cables please contact Real Tech.
15 Dual Feed
15.1 Dual Feed Setup
The wall mountable dual feed system is shown in Figure 15-1 with
the various plumbing connections shown in Figure 15-2 and
communication connection to the analyzer shown in Figure 15-3.
15.1.1 Enable
This allows a unit to switch between dual feed and single feed
modes. Note that under single feed mode, the feed must be
connected to the influent port on the outside of the analyzer.
16 Troubleshooting Guide
An operator has three sources of information for system diagnostics:
1. LEDs on the main circuit board
2. System alarms displayed on the operator interface panel
3. Raw sensor data displayed via the “Secondary Data” menu on the operator interface panel
If the system ever “freezes”, open the door and check the state of the LEDs on the Main Control Board.
Normally, the pattern of lighting from the LEDs change in specific sequence to indicate different states of
operation. If the LEDs are not changing after one minute, make a note of which ones are on or off, and
compare that pattern to the table below. In the table, a zero means an LED is OFF, while 1 means an LED is
ON.
Warning: The UV light may be left on when the door is opened. Make sure you have appropriate eye
protection or restart with the door open.
Note: Each analyzer built at the Real Tech factory is calibrated individually to the components and the finished
construction of the unit. Information for the calibration of a unit is stored on the main circuit board. Therefore,
swapping boards means the calibration for a given unit is no longer correct as it based on the precise
relationships between the components. When communication issues arise that could be an indication of circuit
board failure or damage, contact a Real Tech technician for assistance.
Alarms and warnings related to lamp operation and analyzer status are listed under the “Alarms” menu.
Alarms and warnings need to be enabled under the “Set Alarms” menu in order to be displayed.
The lamp output is low but not too It is nearly time for the lamp to be replaced, however, the
Lamp Low low for the analyzer to provide lamp output is still high enough for the analyzer to produce
accurate measurements. accurate measurements.
Shut off the power to the analyzer after shutting off the water
flow.
Assess the damage inside the analyzer and call Real Tech
A leak has been detected inside
Leak Detection for further instructions.
the analyzer.
If the Real Clean system is connected to the analyzer it will
have been signaled to shut of the water flow assuming no
electrical short has occurred. Note that the Real Clean
system requires power and signals from the analyzer to shut
off the water flow.
The “Secondary Data” menu gives raw data on four variables: Lamp, Water, Temperature, and Humidity.
ROTATION
MENU ITEM WATER PATHLENGTH IDEAL VALUE
POSITION
Rotation position 1 means the lamp is oriented to the reference position i.e. lamp is aligned with a sensor to
read the lamp value. Rotation position 2 means the lamp has rotated 90 degrees so that light is directed
through sample water within the flow cell to the sensor.
If all four numbers display 0, this indicates the system did not start up correctly (refer to the diagnostic LED’s). If
Lamp and Water numbers are 0 but humidity and temperature data appears to be correct then this indicates a
communication problem with the sensor board located on the rotating assembly.
Current lamp may be brighter than Change lamp or wait for a 14-day burn-in period to
40,000 and higher
previous one. see if lamp value falls below 40,000.
Humidity should be kept as low as possible especially when the sample water is cold as condensation can form
on the flow cell which can severely affect accuracy. Please refer to the Dehumidifier section of this manual.
Standard Power Supply 24 VDC Power adapter and cable for C series analyzers UVT-076112
Quartz windows Quartz windows and seals for 4000 series models. UV254-5H25
ACCESSORY PRODUCT
DESCRIPTION PART NUMBER
NAME
Real Clean System II Automatic chemical cleaning system for continuous analyzers UVT-118010
19 Warranty
Real Tech Inc. (Real Tech) warrants to the original Real Tech assumes no liability for indirect or
purchaser of this instrument that it will be free of consequential damage, losses, injury or costs of
defects in material and workmanship, in normal use any kind, and the original purchaser, by placement
and service, for a period of two years from date of of any order for the instrument, or parts thereof,
delivery to the original purchaser. Real Tech's shall be deemed liable for any and all damages
obligation under this warranty is limited to replacing incurred by the use or misuse of the instruments, or
or repairing the instrument, at Real Tech’s option, at
parts thereof, by the purchaser, its employees, or
Real Tech's factory. Excluded from this limited
others, following receipt thereof.
warranty are accessories and supplies.
The original purchaser is responsible for return of Carefully inspect this product for shipping damage, if
the instruments, or parts thereof, to Real Tech's damaged, immediately notify the shipping company
and arrange an on-site inspection. Real Tech cannot
factory. This includes all freight and insurance
be responsible for damage in shipment and cannot
charges incurred in shipping to and from Real
assist with claims without an on-site inspection of the
Tech's factory. damage.
Real Tech is not responsible for any personal injury This warranty is given in lieu of and excludes all other
or damage to the instrument, or parts thereof, warranties, conditions and guarantees, expressed or
resulting from any accident, improper use of the implied, statutory or otherwise including those of
instrument, negligence or willful neglect on the part merchantability and fitness for use. No agent or
of the user of the instrument. Real Tech is not representative is authorized to assume any liability
responsible for any personal injury or defects to the for Real Tech.
instrument resulting from repairs or alterations
made by any person not authorized by Real Tech.