Professional Documents
Culture Documents
1700/1900
Service Manual
©2010 Perten Instruments AB
Perten Instruments AB
PO Box 5101
SE-141 05 Kungens Kurva
SWEDEN
Tel +46 8 880 990
Fax +46 8 881 210
E-mail info@perten.com
Web www.perten.com
Contents
Part 1— Getting started __________________________________________ 1
1.1 Safety instructions _________________________________________________2
Symbols _______________________________________________________ 2
1.2 Introduction ______________________________________________________3
1.2.1 Recommended tools _________________________________________ 3
1.2.2 Spare parts ________________________________________________ 3
1.2.3 Operation _________________________________________________ 3
1.2.4 Differences between FN 1700 and FN 1900 _______________________ 3
1.2.5 Changing the operating voltage ________________________________ 4
1.2.6 RS-232 communication _______________________________________ 4
1.2.7 Block diagram ______________________________________________ 5
Contents i
Falling Number 1700/1900—Service Manual v6
ii Contents
Falling Number 1700/1900—Service Manual v6
Symbols
1 On (power) switch
position * caution—hot surface
~ alternating current
/ caution—risk of electric
shock
1.2 Introduction
This service manual covers the Falling Number 1700 (FN 1700) and Falling Number 1900 (FN
1900) instruments, manufactured by Perten Instruments. Both instruments are based on the
same platform. Except where noted, all parts of this manual apply to both instruments.
1.2.3 Operation
For details on the installation and operation of the FN 1700 or FN 1900, please refer to the
operation manual for the instrument.
' Notes
The part numbers for fuses are T5AL 250V (220–240 Vac,
5 × 20 mm) and T10AL 250V (100–120 Vac, ceramic
6.3 × 32 mm).
No changes are required when switching between 50 and
60 Hz.
remote
WAND LCD display
J21 controller
connector
board
display driver
board
STIRRER
J20 keyboard
connector
16
start
opto-
arm right stop
switches J01 8 J11 J10 10 J25 PC J27 2 sensor
board
board
stirrer
motor 4 J14
optical
J01 3 cassette
sensor
J05 J17 J04 J13 J01 J02 J12
RS-232 NC
connector 3
2 4 2 6
test
button 2 water bath
temperature level/
regulation mains 5 temp.
solid-state relay transformer sensor vertical
(FN 1900 only) heating
FN1700: jumper element
mounted on J21
2 4 water bath
connector 2
J21 J17 J18
mains power
inlet and 4 relay board J20
J19 4 horizontal
voltage 2
heating element
selector 1
No. Function
1 Auto-start disabled. The stirrer arm and counting will not start until the visor is lowered.
2 Water bath disabled. The heating elements are switched off and the water temperature
and water level sensors are disabled. The instrument will allow tests to be run with no
water or with cold water. The instrument can be activated by lowering the visor.
3 Auto-start and water bath disabled. Combination of test functions 1 and 2. The
instrument can be activated by lowering the visor.
4 Continuous self-test mode. A test is automatically started two minutes after the
previous test cycle is finished. The instrument can be activated by lowering the visor. This
function will run with cold water; i.e. the heating elements will not be switched on.
5 Language. After lowering the visor, the language can be set. Press the test button until
the required language number appears on the display. Exit by switching off the
instrument.
0 = English 3 = Spanish 6 = Italian 9 = German 12 = Russian
1 = Swedish 4 = Turkish 7 = Norwegian 10 = Finnish 13 = Danish
2 = Czech 5 = French 8 = Portuguese 11 = Polish 14 = Slovenian
6 Stirrer motor on/off. The stirrer motor can be started by lowering the visor and stopped
by raising the visor.
Before using this function, use test function 9 to ensure the start arm is in the forward
position. Otherwise, the crankshaft may hit one of the link arms for the start arm.
7 Stirrer motor to top position. When the visor is lowered, the stirrer motor starts and
then stops at the top dead end position.
8 Stirrer motor to stirrer release position. When the visor is lowered, the stirrer motor
starts and then stops at the stirrer release position. The claw solenoid is then activated
while the stirrer motor reverses a few steps to the top dead end position.
9 Start arm forward. When the visor is lowered, the start arm moves to the forward stop
position.
No. Function
10 Start arm back. When the visor is lowered, the start arm moves to the back stop
position.
11 Claw activated 2 s. When the visor is lowered, the claw solenoid is activated for
2 seconds.
12 Display/LED test. When the visor is raised, the display shows tES12 and the LED on the
start arm should be off.
Every second time the visor is lowered, all segments on the display should be lit,
including the left/right indicators, and the LED on the start arm should be red.
Every other time the visor is lowered, all segments on the display should be unlit and the
LED on the start arm should be green.
13 Vertical element. When the visor is lowered, the vertical element in the water bath is
switched on. When the visor is raised, the element is switched off.
14 Horizontal element. When the visor is lowered, the horizontal element in the water bath
is switched on. When the visor is raised, the element is switched off.
15 Beeper. When the visor is lowered, the beeper is switched on. When the visor is raised,
the beeper is switched off.
16 Calibration of stirrer top dead end. When the visor is lowered, the stirrer motor (stepper
motor) is switched on for one revolution. During this revolution the number of motor steps
with the top position sensor active is measured and displayed. This number divided by
two defines the number of motor steps between triggering of the top position sensor and
the actual top dead end position. This value should be between 20 and 30. If the number
is outside these limits, the vertical position of the top sensor board may need to be
adjusted.
Before using this function, use test function 9 to ensure the start arm is in the forward
position.
This test function must be run twice consecutively for the correct value to be stored
correctly in the internal memory.
17 Stirrer stop position. Prior to using this function, place two stirrers and test tubes in a
cassette and place the cassette in the water bath. When the visor is lowered, the stirrer
motor will lower the stirrers to the position where the stirrer sensor should be triggered.
This test function is commonly used together with test function 18 to adjust the stirrer
sensor (see 3.5.2, page 28).
18 Stirrer sensors status. This function is mainly used to adjust the vertical trigger point of
the stirrer sensors. Each display shows 1 when the corresponding sensor is activated
and 0 when it is not activated. The upper display shows the left sensor and the lower
display shows the right sensor. Activate and deactivate this test function by lowering and
then raising the visor.
Before adjusting the stirrer sensors, the stirrers must be positioned exactly at the trigger
point using test function 17. To adjust, move the sensors upwards until the sensor status
switches from 0 to 1. Use test functions 20 and 21 to check adjustments.
No. Function
19 Micro-controller input states. When the visor is lowered, the logic input states to the
micro-controller are shown on the display. Raise the visor to exit this test mode.
Each segment represents one input signal. The respective segment is lit when the
respective input to the micro-controller (or IC17 latch circuit) is high. The diagram below
shows what each display segment represents.
A = OFF—water bath level high
888
B = OFF—start arm in forward position A B C
C = OFF—left stirrer stop sensor activated
D = OFF—water bath disconnected
E = OFF—start arm in back position
F = OFF—right stirrer stop sensor activated D E F
G = OFF—stirrer module top sensor activated
888
H = OFF—visor in down position G H I
I = OFF—test button pressed
J = not used
K = OFF—cassette with stirrer(s) present
J K L
L = not used.
20 Left stirrer falling distance. When the visor is lowered, the number of stirrer motor steps
between the release position and sensor trigger point for the left stirrer will be counted.
Use this test function to verify that the stirrer sensors are correctly adjusted. The
displayed value should be in the range 142–146. This value may change slightly after
putting back the start arm cover.
Before using this test function, use test function 9 to ensure the start arm is in the front
position.
21 Right stirrer falling distance. Same as for test function 20 but applied to the right stirrer.
Before using this test function, use test function 9 to ensure the start arm is in the front
position.
22 Temperature and ADC value. The water bath temperature, recalculated from the
temperature sensor ADC (Analog to Digital Converter) value, is shown on the upper
display. The ADC value for the temperature measurement is shown on the lower display.
PW Functions
91 Demonstration mode and production tests—used for demonstrations and long-term tests.
The test functions correspond to test functions 1–4 on the FN 1700.
92 Functions for checking operation and facilitating adjustments and repair. The test
functions correspond to test functions 6–22 on the FN 1700.
93 Temperature calibration—used when the temperature sensor needs to be re-calibrated,
for example following exchange of the water temperature sensor or controller board.
94 Service interval counter—shows number of analyses and date for last service.
95 Actual temperature—shows actual calculated water temperature in degrees Celsius.
' Notes
1 Loosen the two screws that secure the guide plate (a). a
1 Tip the instrument onto its side and locate the three
vertical positioning bolts for the water bath (a).
2 Using a spanner and Allen key, loosen the nuts securing b
the positioning bolts and turn the positioning bolts to the
required position (b).
3 Tip the instrument upright, attach the water bath and lid
and insert a cassette loaded with tubs and stirrers.
4 Check that the stirrer collars touch the start arm. If
required, insert different stirrers and cassettes to verify
that the adjustment is correct.
5 Tip the instrument back onto its side and tighten the nuts
to secure the vertical positioning bolts.
' Notes
' Notes
' Notes
4 Undo the screw that secures the display unit (FN 1700)
or connector unit (FN 1900). Pull the connector straight
out to prevent it from being damaged and remove the unit
(c).
5 Undo the eight screws holding the main cover. Pull the c
cover forward until the cover tabs are released from the
front chassis and lift off the cover (d).
6 Disconnect the grounding cable between the main body
and the cover (e) before removing the cover entirely.
d
' Notes
' Notes
' Notes
When mounting the motor, the key end of the screws holding
the motor should be aligned with the outer surface of the
locking nuts for the screws.
The stopper screws need to act as an axle for the motor. If
these are screwed in too far, the arms will become splayed
and may jam in the slots on the front panel when the arm
moves forward.
' Note
The forward and back stop positions will affect each other and
must be adjusted simultaneously.
' Note a
' Note
• 8-bit CMOS microcontroller (IC08). This circuit also holds the internal program in a built-in
PROM.
• 8 k serial EEPROM for adjustable settings. For details on making changes to the
EEPROM, see 5.2.5, page 53.
• CAN interface (IC20), used for communicating with the display (FN 1700) or the external
control box (FN 1900).
• Reset generator (IC06).
• RS-232–TTL signal level converter (IC05), used for external communication to a PC or
optional printer.
• Latch circuits for input/output signals to and from the microprocessor (IC10, IC17, IC18
and IC19)
• Adjustable signal amplifier for the stirrer cassette sensor (IC11A).
• Non-adjustable preamplifier for the water bath temperature sensor (IC11B).
• Logic and drivers for the stepper type stirrer motor (IC12, IC13 and IC14).
• Logic and drivers for the stepper type start arm motor (IC15 and IC16).
• Power supply for 8 Vdc stabilized (IC01) and 2 × 5 Vdc stabilized (IC02 and IC04).
3.6.2 Jumpers
There are two jumpers on the board:
• J07 should always be left open. This is only used for program development when running
from external memory.
• J20 should have a jumper. Removing the jumper allows the buzzer to be switched off
temporarily.
TP Description
TP1 Cassette sensor signal. On newer instruments with an optical cassette sensor, this
voltage should be steady. On older instruments with a magnetic cassette sensor,
the voltage should be 0–300 mVdc with no stirrer present. The signal may fluctuate
within the allowed range because of variations in the ambient magnetic field. This
is the processed analog signal from the magnetic field sensor mounted on the start
arm board. The offset voltage is set using potentiometer R37. This offset defines
the toggle level for the sensor system. See 3.4, page 25 for details on adjusting the
toggle level. This signal is monitored by LED H09 on the controller board.
TP2 Water bath temperature sensor signal. The signal is linearly proportional to the
water temperature and should be approximately 1 Vdc at 25°C and 4 Vdc at 100°C.
There are no adjustments for this signal. The temperature is instead recalculated
from this input voltage using constants in cells 16, 18, 20 and 22 in the EEPROM
(see 3.8, page 38).
TP3–4 Digital signals for switching the horizontal (TP3) and vertical (TP4) heating
elements on and off. These should be 0 V when off and 7–8 Vdc when on.
TP5–10 A 4-bit binary stepper motor code signal to rotate the stirrer motor on TP5, TP6,
TP8 and TP9. The motor current is switched on when the signal on TP7 and TP10
is high.
TP11 Motor current setting signal for stirrer motor.
TP12–13 Current sense feedback for internal switching regulator in stepper motor drivers for
limiting the current to the motor.
TP14–19 Start arm stepper motor signals. Similar function as for TP5–10.
TP20–21 Start arm stepper motor signals. Similar function as for TP12–13.
TP22 Ground connection. The ground is connected to the chassis. The chassis can
therefore normally be used as a reference for signal measurements. If using the
chassis, connect as close as possible to one of the four screws holding the
controller board.
TP23 Buzzer signal. Should be 0 V when the buzzer is off and 4–5 Vdc when the buzzer
is on.
TP24 +8 Vdc stabilized voltage. Tolerance ±5%.
TP25 +5Vdc stabilized voltage for most circuits on the controller board. Tolerance ±5%.
TP26 +5Vdc stabilized voltage for some peripheral circuits such as the display (FN 1700),
the control box (FN 1900) and circuits in the water bath. Tolerance ±5%.
TP27 Claw solenoid on/off signal. Should be 0 V when the solenoid is not activated and
8 Vdc when it is activated.
TP28 Water temperature regulation signal. Only used in the FN 1900. The water
temperature can be pulse-regulated with a variable duty cycle. The pulses are sent
to a solid state relay that directly controls the horizontal heating element.
1 Before removing the old board, save all values from the
EEPROM (see 5.2.5, page 53).
2 Dismantle the main body of the instrument (see 3.3.1,
page 23).
a
3 Remove the four screws holding the controller board (a).
4 Disconnect the cables and remove the old board.
5 Insert the new board and connect the cables.
6 Secure the new board using the four screws.
7 Check that the start arm flag moves freely within the
photo interrupters H12 and H13 for the start arm end
positions.
8 Adjust the start arm end positions (see 3.5.2, page 28).
9 Enter the EEPROM values from the old board into the
EEPROM cells on the new board (see following note for
exceptions).
10 Recalibrate the temperature sensor (see 3.8, page 38).
11 Adjust the sensitivity of the cassette sensor (see 3.4,
page 25).
' Notes
3.7.2 Jumpers
J07 and J10 are used only for development purposes and should always be left open.
J08 is for a jumper that connects the battery for the real time clock to the clock circuit.
J02 needs to be unlocked by lifting the collar before pulling out the foil cable.
3.7.5 Miscellaneous
The WAND connection currently has no function.
The communication between the stirrer unit and the control box uses a CAN (controller-area
network) protocol.
6 The display will show LO TEMP and the heating T. 23.8 AD 227
LO TEMP 23.2`C>
elements will be switched off to allow calibration of the
low temperature point. The display will also show the
temperature in degrees Celsius calculated from the
existing temperature constants (e.g. 23.8) followed by the
value from the Analog to Digital Converter (ADC)
(e.g. 227).
7 Measure the water temperature using a precision
T. 23.8 AD 227
thermometer. Once the measured temperature value is LO TEMP 23.7`C>
stable, type the value using the keypad and press
ENTER.
' Notes
' Notes
If the level/temperature sensor or controller board is exchanged, the temperature sensor must
be recalibrated (see 3.8, page 38).
If the controller board is exchanged, check if the grounding system for preventing static
discharge needs to be updated. Older instruments may require additional grounding (see 3.9,
page 40).
4.2 Troubleshooting
Symptom Description
Blinking display This indicates a low water level was detected during the test. Possible
(FN 1700) causes include:
• faulty level sensor
• faulty connection between the water bath and main unit, and
• fault in the controller board
Only one display This indicates that only one of the stirrers was detected when starting the
shows falling test. Possible causes include:
number value • using only one stirrer for the test (normal operation)
(FN 1700)
• faulty reed switch or reed switch requires adjustment
• flour not mixed in bottom of viscometer tube, causing the stirrer to sit
too high and preventing the reed switch from detecting the magnet.
• fault in the cable between the start arm sensor board and the controller
board, and
• fault in the controller board.
Stirrer not This problem can be caused by the claws that hold the stirrers not being
released during fully held backwards by the spring at the top claw solenoid. This is a
test particular problem with older instruments employing a brass spring. See
3.5.4, page 31.
Some instruments employ a top claw solenoid with a 100% duty cycle.
These solenoids are not as strong as those normally used and may cause
problems with releasing the claws after stirring.
The problem can also be caused by the claws being mechanically tight and
not moving freely to the back position.
Sudden stop Internal static discharge is a potential problem that may cause the
during test instrument to stop randomly or even damage circuits on the controller
board. This problem was detected and the design changed to include an
additional grounding system to reduce the risk of static discharge. Older
instruments that are not equipped with the separate grounding system
should be updated to eliminate the risk of this problem. See 3.9, page 40.
Problems with This problem may be caused by the top sensor board being accidentally
sensors for stirrer mounted on the upper side of the console. To rectify this problem it should
top dead end and be removed and mounted on the underside of the console. After moving
for visor the boards, check that the indicator flags do not hit the sensors.
The problem can also be cause by a bad connection in one of the photo
interrupters. In this case, replace the top sensor board.
Start arm locks If there is an attempt to run a test with a cassette in place but no stirrers or
cassette in place test tubes, the start arm will move out and lock the cassette in place. To
when no tubes or remove the cassette, switch off the instrument.
stirrers in place
' Note
If the level/temperature sensor or controller board is exchanged, the temperature sensor must
be recalibrated (see 3.8, page 38).
If the top sensor board is moved, the stirrer top dead end must be recalibrated (see 3.5.2,
page 28).
Part 5—Appendices
Part 5—Appendices 45
Falling Number 1700/1900—Service Manual v6
5.1.1 Introduction
The FN1700 has a single RS-232 serial port located on the
stirrer unit. The FN1900 has two RS-232 serial ports, one on
the stirrer unit and one on the control box.
The port located on the stirrer unit (Port 1) is used for reading
from and writing to the EEPROM on the controller board. The
port can also be used to connect an optional printer. The port
located on the control box of the FN1900 (Port 2) can be used
either for entering a sample ID, from a device such as a
barcode reader or keyboard, or for installing language files
using Perten Language Editor (see 5.3, page 57).
FN1700/FN1900 Computer
Signal Pin Pin
Tx 2 2
Rx 3 3
5 (9 pin)
Gnd 5
7 (25 pin)
FN1700/F1900 Printer
Signal Pin Pin
Tx 2 2
Rx 3 3
5 (9 pin)
Gnd 5
7 (25 pin)
46 Part 5—Appendices
Falling Number 1700/1900—Service Manual v6
Parameter Settings
Bit rate 9600
Data length 8
Parity none
Stop bits 1
ETX/ACK off
option description
0 serial two-way communication
1 FN1700 standard printout
2 extended printout
Part 5—Appendices 47
Falling Number 1700/1900—Service Manual v6
Command Reply
Description (from PC) (from stirrer unit) Notes
Instrument SP7?E7<LF> ok<CR><LF> 1,2
status SP7=X
Error status SP6?E8<LF> ok<CR><LF> 1,3
SP6=Y
Get analysis PRC,EF<LF> printout of 1,3
results analysis result
Send left ID IDL,"Z",CS<LF> ok<CR><LF> 1,4,5
' Notes
48 Part 5—Appendices
Falling Number 1700/1900—Service Manual v6
PC operator selects a
sample for analysis
yes
status = 0 send ID
no
no
yes
status = 2
no
no
status = 0
yes
Part 5—Appendices 49
Falling Number 1700/1900—Service Manual v6
Following are example printouts (in English) both without and with altitude correction. All
characters are noted by their hexadecimal codes.
You can capture the results as supplied, allowing you to operate in parallel with a printer by
only connecting the PC RxD pin.
50 Part 5—Appendices
Falling Number 1700/1900—Service Manual v6
5.2 InstTest
The InstTest communication program can be used to check
and change the contents of the memory on both the FN 1700
and FN 1900.
5.2.1 Installation
To install InstTest
' Notes
Part 5—Appendices 51
Falling Number 1700/1900—Service Manual v6
5.2.4 Calibrating
All instruments have individually calibrated water bath
temperature sensors. This section of the program is used to
calibrate these sensors. See 3.8, page 38 for detailed
calibration procedures.
52 Part 5—Appendices
Falling Number 1700/1900—Service Manual v6
5.2.5 Eeprom
To view the contents of the EEPROM, click Read. All
information in the EEPROM memory will be displayed along
with all information that can be read and changed from the
Date, Serial no, Type no and Calibrating windows.
The cells are organized in pairs. Each cell is two bytes and
each cell pair four bytes.
To change the content of a box, enter the new value in the
Amend field and click Write.
Unused controller boards will show default values according to
the following table.
Part 5—Appendices 53
Falling Number 1700/1900—Service Manual v6
54 Part 5—Appendices
Falling Number 1700/1900—Service Manual v6
Part 5—Appendices 55
Falling Number 1700/1900—Service Manual v6
5.2.7 Troubleshooting
Error message: Debug Assertion Failed!.
Likely cause: The file attribute for the file Comms.txt is set to
read only.
Solution:
1 Close InstTest and open Windows Explorer.
2 Locate the file Comms.txt. The file will be in the same
folder where InstTest has been installed.
3 Right-click the file and select Properties.
4 Clear the check box labeled Read-only and click OK.
5 Restart InstTest.
56 Part 5—Appendices
Falling Number 1700/1900—Service Manual v6
To install PLE
Part 5—Appendices 57
Falling Number 1700/1900—Service Manual v6
' Note
58 Part 5—Appendices
Falling Number 1700/1900—Service Manual v6
Sample particle size If testing flour, ensure that a representative sample is taken. Use
a sample divider. If testing grain, grind at least 300 g and ensure
that the laboratory mill used produces a sample with a correct
particle size distribution. Suitable mills include Perten Instruments
LM 3100 and LM 120 laboratory mills. ICC 107/1 describes the
correct particle size distribution to use for Falling Number
analysis.
Sample weight Check that the balance used for weighing samples is accurate
using known standard weights.
Moisture correction When preparing a sample, correct for moisture content (see the
installation and operation manual for details). This ensures that
the relative proportions of starch and water in each test sample
are consistent. ICC 107/1 requires correction to a 14% moisture
basis. AACC methodology uses either a weight adjustment or a
recalculation to a fixed (14%) moisture content. Check that
comparisons are being made using the same moisture correction
methods.
Water volume The volume of water should be 25.0 ± 0.2 mL. Check that the
pipette used is accurate by weighing the volume of water. The
weight of 25.0 mL of water is 24.96 g at 20°C or 24.93 g at 24°C.
Water quality Use distilled water or water of equivalent purity. The pH of the
water will affect the measurement. If water quality is suspect,
compare by using water from other source. (Pharmacies often
supply distilled water, which could be used for comparison.)
Dispenser water The temperature of the dispensed water should be 22 ± 2°C.
temperature Large variations in water temperature can adversely affect
accuracy. The viscometer tubes should not be hot (e.g. after
being dried in an oven) as this will affect results.
Water bath On the FN 1700 and FN 1900, ensure that the water bath is
pressed firmly towards the tower until it clicks and comes to the
rear stop.
Part 5—Appendices 59
Falling Number 1700/1900—Service Manual v6
60 Part 5—Appendices