You are on page 1of 8

Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250

A hemispherical high-pressure xenon gamma radiation


spectrometer
Royal Kessick, Gary Tepper*
Department of Chemical Engineering, Virginia Commonwealth University, 601 West Main Street, 23284-3028 Richmond,
VA 23284, USA

Received 21 November 2001; received in revised form 20 March 2002; accepted 27 March 2002

Abstract

A prototype hemispherical high-pressure xenon gamma radiation spectrometer was designed, constructed and tested.
The detector consists of a pair of concentric hemispherical electrodes contained inside a thin-walled stainless steel
pressure dome. Detector performance parameters such as energy resolution, linearity and vibration sensitivity were
determined and compared to previous cylindrical and planar designs. Without a Frisch grid, the hemispherical detector
provides a total room temperature energy resolution of 6% @ 662 keV and is relatively insensitive to acoustic
interference. r 2002 Elsevier Science B.V. All rights reserved.

PACS: 29.40.Cs; 29.40.Kv

Keywords: High-pressure xenon; Hemispherical; Radiation detection

1. Introduction oped in both planar and cylindrical geometries by


several groups [1–4]. High-pressure xenon (HPXe)
A variety of field applications including envir- detectors offer several attractive features including
onmental characterization, treaty verification and a combination of relatively high-energy resolution,
geological exploration require gamma radiation large volume and thermal stability as well as
spectrometers that can provide stable operation insensitivity to radiation damage that make
over a range of ambient conditions. Scintillation them particularly attractive for field applications.
detectors generally provide inadequate energy Table 1 lists some of the important properties of
resolution for these applications while high purity HPXe spectrometers. HPXe detectors are usually
germanium detectors must be maintained at operated at a xenon density near 0.6 g/cm3. At
cryogenic temperatures, which is inconvenient in higher densities, the energy resolution is degraded
the field. Radiation detectors based on highly due to the local density fluctuations normally
purified and compressed xenon have been devel- observed in fluids near the critical point [5].
The best energy resolution reported to date for
*Corresponding author. Tel.: +1-804-827-4079; fax: +1-
HPXe detectors is for a cylindrical detector with a
804-828-4269. shielding mesh [4]. Bolotnikov and Ramsey [4]
E-mail address: gctepper@saturn.vcu.edu (G. Tepper). demonstrated that, in a cylindrical geometry, it is
0168-9002/02/$ - see front matter r 2002 Elsevier Science B.V. All rights reserved.
PII: S 0 1 6 8 - 9 0 0 2 ( 0 2 ) 0 1 0 0 7 - 0
244 R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250

Table 1 factors suggests that a HPXe detector could be


Properties of high-pressure xenon developed in which most of the induced charge on
Atomic number Z ¼ 54 the anode is generated in the anode proximity
Critical point 58 atm, 290 K, 1.1 g/cm3 where the electric field magnitude is the strongest.
W value 21.9 eVa Thus, even without a Frisch grid structure,
Electron drift velocity B105 cm/s (for EX2 kV/cm) the effect of charge (electron and ion) motion in
Fano factor 0.17a
Intrinsic energy resolution 0.56% @ 662 keV
the weak-field region near the cathode may be
reduced. This effect was previously investigated in
a
Density dependent. a cylindrical HPXe spectrometer without a shield-
ing mesh [8]. However, very high cathode voltages
(ca. 15 kV) were required before any significant
possible to compensate for the inherent grid improvement in energy resolution could be rea-
shielding inefficiency in order to obtain very lized. The larger electric field gradient provided by
high-energy resolution (o2%). However, these the hemispherical geometry may provide a similar
detectors, like the previous planar designs, are benefit, but without the need for extremely high
sensitive to acoustic interference associated with voltages.
mechanical vibrations of the Frisch grid electrode.
A variety of grid designs have been tested and,
while some incremental improvements have been 2. Experimental
realized by improving the grid rigidity and support
structure, the problem remains [6]. Fig. 1a is a photograph of the prototype
The function of the grid electrode is to shield the hemispherical HPXe spectrometer and Fig. 1b is
anode from the effect of charge motion in the a schematic diagram illustrating the main compo-
ionization region. The rise time of the anode pulses nents of the detector. The pressure vessel consists
is, therefore, determined by the electron drift time of a stainless steel hemisphere with a wall thickness
between the grid and anode electrodes rather than of approximately 0.0600 welded to a standard 600
the location of the ionization event and the effect diameter conflat flange. The electrical feed-
of positive ion motion is essentially eliminated. In throughs and vacuum/fill port are welded into a
the absence of a grid electrode, the anode pulse rise second 600 flange and a third 600 double-sided flange
time and amplitude will depend on the ionization is used as a spacer and to support the internal
coordinate. Rise time compensation has been electrode structure. The anode is a solid, 1.500
investigated as a possible approach to electro- diameter stainless steel hemisphere and is sup-
nically perform the function of the Frisch grid ported by a 400 diameter, 0.37500 thick alumina
electrode [4,7]. However, this approach introduces ceramic disc. Holes were machined into the
additional pulse processing requirements and at ceramic anode support disc for the anode and
this time has not been successfully applied to cathode electrical feedthroughs as well as a pump
HPXe detectors. out port to evacuate the hemispherical ionization
In this paper we investigate another possible region. The cathode electrode consists of a silver
approach toward the development of a practical coating deposited onto the inside surface of an
field-portable HPXe spectrometer. The hemisphe- alumina ceramic hemisphere with an inner dia-
rical detector geometry offers some unique perfor- meter of 3.500 . The ceramic hemisphere is seen
mance advantages. The thin-walled hemispherical prominently in the photograph of Fig. 1a.
pressure dome provides a wide field of view. For The requirements and methodologies for achiev-
concentric hemispherical electrodes, the electric ing and maintaining the necessary xenon purity
field magnitude drops off quickly as one over the have been addressed elsewhere and are only briefly
square of the distance from the anode electrode. In summarized here [9]. An ultra-high vacuum system
addition, the anode electrode can, in principle, be containing both getters and a titanium spark
made arbitrarily small. The combination of these purification chamber was used to purify the xenon
R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250 245

Fig. 1. (a) Photograph of the hemispherical HPXe spectrometer, (b) schematic diagram of the detector.

and to prepare and fill the detector. The lower height analysis. The high voltage was supplied
limit of the electron lifetime in xenon was using a low-noise Bertan power supply. A low pass
established by monitoring the duration of pulses filter with a time constant of about 2 s was
produced by cosmic muons using an ionization installed directly at the cathode feedthrough for
chamber located inside the spark chamber. The additional power supply noise suppression.
xenon purity was considered acceptable when the
lower limit of the electron lifetime exceeded 3 ms at
a density of approximately 0.5 g/cm3. The detector 3. Results
was baked at 2001C for approximately 1 week
under ultra-high vacuum conditions. The detector Fig. 2 is a typical pulse height spectrum
was then filled with the purified xenon to a density obtained with the detector exposed to a 137Cs
of approximately 0.3 g/cm3 as determined by calibration source at a cathode voltage of 8 kV.
weighing before and after. The maximum xenon As stated previously, the xenon density was lower
density in this prototype detector was limited by than optimal resulting in the relatively large
the pressure vessel design. The double-sided Compton shoulder visible in Fig. 2. The total
conflat flange configuration was not able to FWHM energy resolution for the peak of Fig. 2
reliably support internal pressures greater than was about 6% @ 662 keV at a shaping time of
about 700 Psi. The flanges were used to provide 5 ms. The peak is not symmetric and the low energy
convenient access to the detector electrodes during tailing is responsible for most of the peak width.
prototyping, but will be eliminated in future For example, if the energy resolution is calculated
detector generations. The effects of xenon density based on two times the high-energy half width, the
on detector energy resolution have been reported resolution drops to 4%. A test pulse generator was
previously and will not be discussed here [10]. used to establish the base electronics noise. The
The data were obtained using an AmpTek 250 FWHM of the pulser peak was 2.2% for the
preamp operated at room temperature and operating conditions described above.
mounted directly to the anode. The anode Fig. 3a is a pulse height spectrum for the
capacitance was 22 pF. The output of the pre- detector exposed to a 133Ba source and Fig. 3b is
amplifier was fed to an Ortec shaping amplifier a pulse height spectrum for a 22Na source under
and then into a multi-channel analyzer for pulse the same detector conditions as Fig. 2. The
246 R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250

1400

1200

1000

800
Counts

600

400

200

0
100 150 200 250 300 350 400 450 500 550 600
Channel Number
137
Fig. 2. Pulse height spectrum of a Cs calibration source at a cathode voltage of 8 kV.

detector energy resolution was sufficient to clearly detector output pulse height is linearly dependent
resolve the two low energy peaks in the 133Ba on energy as is expected.
spectrum. However, due to the relatively low The effect of the electric field strength on the
xenon density, the high-energy peak in the 22Na charge collection was also investigated. Fig. 5 is a
spectrum is not clearly resolved. plot of the 137Cs peak channel number versus
Fig. 4 is a plot of gamma energy versus channel cathode voltage from approximately 5 to 9 kV
number. The data were obtained during a single at a shaping time of 5 ms. As can be seen in the
run with the detector simultaneously exposed to figure, the peak channel number (which is a
137
Cs, 22Na and 133Ba sources and a cathode measure of the total collected charge) increases
voltage of 8 kV. This plot demonstrates that the linearly with applied voltage and appears to begin
R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250 247

20000 6000

18000
5000
16000

14000
4000
12000

Counts
Counts

10000 3000

8000

2000
6000

4000
1000

2000

0 0
100 150 200 250 300 350 100 300 500 700
(a) Channel Number (b) Channel Number
133 22
Fig. 3. Pulse height spectrum of (a) Ba and (b) Na.

to saturate at a cathode voltage of 9 kV. It was a room temperature energy resolution of approxi-
not possible to probe higher voltages because of mately 6% @ 662 keV. This resolution is about a
limitations in the current feedthrough design. factor of three worse than the best energy
Fig. 6 is a plot of the electric field magnitude resolution reported for HPXe detectors incorpor-
versus radius for several applied cathode voltages. ating a shielding mesh, and is more typical of the
At a cathode voltage of 8 kV, the electric field best results reported for cylindrical detectors
magnitude ranges from a high of about 8 kV/cm at without a shielding mesh [4,7,8].
the anode to a low of about 1.5 kV/cm at the Therefore, although the hemispherical geometry
cathode electrode. appears promising, the resolution of our prototype
is still about a factor of 2 or 3 worser than is
necessary to offer a competitive advantage. How-
4. Discussion ever, we believe that there is significant room for
further optimization. Specifically, at this time, only
The principal goal of this work was to investi- one detector geometry (anode and cathode dia-
gate the possibility of developing a high resolution, meter) was tested and the xenon density was less
room temperature radiation spectrometer for field than optimal. Furthermore, the prototype detector
applications based on hemispherical HPXe detec- was only designed to operate reliably at cathode
tors. The hemispherical geometry provides a very voltages less than or equal to 8 kV and, based on
wide field of view as well as an excellent structural the results of Fig. 5, there is good reason to believe
design for containing the relatively high internal that the energy resolution could be improved by
xenon pressures. Without a Frisch grid the operating at higher cathode voltages. In the
prototype hemispherical HPXe detector provides hemispherical geometry, the electric field drops
248 R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250

1400

1200

1000
Gamma Energy (keV)

800

600

400

200

0
0 100 200 300 400 500 600 700
Channel Number
Fig. 4. Gamma energy versus channel number.

364

362

360

358
Channel Number

356

354

352

350

348

346

344

342
4000 5000 6000 7000 8000 9000 10000
Volts
Fig. 5. Peak channel number versus cathode voltage.
R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250 249

10.000

9.000
10kV

8.000

7.000
Electric Field Magnitude (kV/cm)

8kV

6.000

5.000
6kV
4.000

3.000

2.000

1.000

0.000
1.5 2 2.5 3 3.5 4 4.5 5
Radius (cm)
Fig. 6. The electric field distribution for four different cathode voltages. The operational voltage (8 kV) is indicated as the solid line.

off rapidly, as 1=r2 ; resulting in the need for anode diameter. However, the effect of the anode
relatively high cathode voltages in order to sustain diameter on the energy resolution is not clear.
an acceptable field magnitude at all locations Increasing the anode diameter decreases the effect
within the ionization region. However, at the of the 1=r2 electric field concentration and, there-
same time, the distance between the anode and fore, may be problematic in terms of the detector
cathode electrodes must be large enough to energy resolution. Further research is required in
provide adequate detection efficiency for gamma order to optimize the detector geometry to provide
radiation. One solution to this problem is to the best combination of energy resolution, effi-
increase the anode diameter [3]. Because the ciency and voltage requirements.
volume of a hemisphere is proportional to r3 ; the The electron drift velocity in HPXe saturates at
anode electrode occupies a negligible portion of approximately 105 cm/s for electric fields greater
the total ionization volume, even at relatively large than 2 kV/cm [11,12]. According to Fig. 6, at a
diameters. However, for a given cathode diameter, cathode voltage of 8 kV, the electric field
the minimum voltage required between the anode magnitude is greater than 2 kV/cm over more than
and cathode electrodes decreases with increasing 90% of the drift region. Therefore, despite the fact
anode diameter. Therefore, in a hemispherical that the electric field magnitude is increasing
geometry, it is possible to reduce the required rapidly as the electrons approach the anode, the
detector voltage without significantly decreasing electron drift velocity will be relatively constant
the detector volume simply by increasing the and equal to the saturation value of 105 cm/s.
250 R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250

A shaping time of 5 ms provided the best energy 5. Conclusions


resolution for the data presented in Figs. 2 and 3.
However, electrons drift a distance of about 0.5 cm A new hemispherical high-pressure xenon gam-
in 5 ms, and this is only about 15 of the cathode-to- ma radiation spectrometer was designed, built and
anode distance. Normally, it is preferable to use an tested. Without a shielding grid or pulse shape
integration time that is long compared to the analysis the detector provided a total room
maximum drift time of the charge carriers. Shorter temperature energy resolution of approximately
integration times lead to ballistic deficit and a 6% @ 662 keV. The elimination of the vibration-
reduced energy resolution particularly in detectors sensitive grid electrode resulted in a significant
with a variation in the charge collection time. Our improvement in the detector performance during
results indicate that, for this particular detector the kind of acoustic disturbances normally en-
geometry, the best energy resolution is achieved by countered in field applications. Our results suggest
using a shaping time much less than would be that, with additional optimization, hemispherical
expected based on the maximum electron drift high-pressure xenon detectors may find use in a
time. variety of field applications requiring rugged,
The primary problem with previous HPXe portable, instruments that operate at room tem-
detectors containing a shielding grid is acoustic perature and without the need for sophisticated
sensitivity caused by mechanical vibrations of the post-data processing.
Frisch grid structure. This vibration sensitivity is
particularly problematic in cylindrical detectors
because of the need to extend the grid along the References
entire length of the cylinder. We qualitatively
[1] G.J. Mahler, B. Yu, G.C. Smith, W.R. Kane, J.R. Lemley,
compared the acoustic sensitivity of the hemisphe- IEEE Trans. Nucl. Sci. NS-45 (3) (1998) 1029.
rical design to that of our previous planar and [2] G. Tepper, J. Losee, Nucl. Instr. and Meth. A 356 (1995) 339.
cylindrical detectors containing a shielding grid. [3] G. Tepper, J. Losee, R. Palmer, Nucl. Instr. and Meth. A
The preamplifier output was monitored during 413 (1998) 467.
[4] A. Bolotnikov, B. Ramsey, IEEE Trans. Nucl. Sci. NS-44
exposure to common sources of acoustic inter-
(3) (1997) 1006.
ference including foot traffic, voices and physical [5] J.M.H. Levelt Sengers, Critical behavior of fluids: concepts
handling of the detector and, as expected, due to and applications, in: E. Kiran, J.M.H. Levelt Sengers
the elimination of the Frisch grid electrode, the (Eds.), Supercritical Fluids, 1994, pp. 3–38.
new hemispherical detector exhibited superior [6] V.V. Dmitrenko, I.V. Chernysheva, V.M. Gratchev, O.N.
vibration resistance. Kondakova, K.V. Krivova, S.E. Ulin, Z.M. Uteshev, K.F.
Vlasik, IEEE Trans. Nucl. Sci. NS-47 (2000) 939.
Finally, because in a hemispherical geometry the [7] G. Troyer, B. Keele, G. Tepper, J. Radioanal. Nucl. Chem.
anode electrode can, in principle, be made 248 (2) (2001) 267.
arbitrarily small, it should also be possible to [8] V.V. Dmitrenko, A.E. Bolotnikov, A.M. Galper, V.M.
develop a hemispherical detector containing a very Gratchev, S.E. Ulin, O.M. Kondakova, V.B. Komarov,
small shielding mesh. A small hemispherical mesh S.R. Krivov, S.I. Suchkov, Z.M. Uteshev, Yu.T. Yurkin,
K.F. Vlasik, SPIE 1734 (1992) 295.
is expected to be far less susceptible to the kind of [9] A. Bolotnikov, B. Ramsey, Nucl. Instr. and Meth. A 383
mechanical vibrations that have plagued the (1996) 619.
elongated mesh designs of cylindrical HPXe [10] A.E. Bolotnikov, V.V. Dmitrenko, A.S. Romanyuk, S.I.
spectrometers. We are currently modifying our Suchkov, Z.M. Uteshev, Pribory I Tekhnika Eksperimenta
4 (1986) 42.
prototype hemispherical detector to include alter-
[11] V.V. Dmitrenko, A.S. Romanyuk, S.I. Suchkov, Z.M.
native anode diameters as well as a shielding Uteshev, Sov. Phys. Tech. Phys. 28 (12) (1983) 1440.
mesh and will report these results in a future [12] S.S.S. Huang, G.R. Freeman, J. Chem. Phys. 68 (4) (1978)
publication. 1355.

You might also like