Professional Documents
Culture Documents
Received 21 November 2001; received in revised form 20 March 2002; accepted 27 March 2002
Abstract
A prototype hemispherical high-pressure xenon gamma radiation spectrometer was designed, constructed and tested.
The detector consists of a pair of concentric hemispherical electrodes contained inside a thin-walled stainless steel
pressure dome. Detector performance parameters such as energy resolution, linearity and vibration sensitivity were
determined and compared to previous cylindrical and planar designs. Without a Frisch grid, the hemispherical detector
provides a total room temperature energy resolution of 6% @ 662 keV and is relatively insensitive to acoustic
interference. r 2002 Elsevier Science B.V. All rights reserved.
Fig. 1. (a) Photograph of the hemispherical HPXe spectrometer, (b) schematic diagram of the detector.
and to prepare and fill the detector. The lower height analysis. The high voltage was supplied
limit of the electron lifetime in xenon was using a low-noise Bertan power supply. A low pass
established by monitoring the duration of pulses filter with a time constant of about 2 s was
produced by cosmic muons using an ionization installed directly at the cathode feedthrough for
chamber located inside the spark chamber. The additional power supply noise suppression.
xenon purity was considered acceptable when the
lower limit of the electron lifetime exceeded 3 ms at
a density of approximately 0.5 g/cm3. The detector 3. Results
was baked at 2001C for approximately 1 week
under ultra-high vacuum conditions. The detector Fig. 2 is a typical pulse height spectrum
was then filled with the purified xenon to a density obtained with the detector exposed to a 137Cs
of approximately 0.3 g/cm3 as determined by calibration source at a cathode voltage of 8 kV.
weighing before and after. The maximum xenon As stated previously, the xenon density was lower
density in this prototype detector was limited by than optimal resulting in the relatively large
the pressure vessel design. The double-sided Compton shoulder visible in Fig. 2. The total
conflat flange configuration was not able to FWHM energy resolution for the peak of Fig. 2
reliably support internal pressures greater than was about 6% @ 662 keV at a shaping time of
about 700 Psi. The flanges were used to provide 5 ms. The peak is not symmetric and the low energy
convenient access to the detector electrodes during tailing is responsible for most of the peak width.
prototyping, but will be eliminated in future For example, if the energy resolution is calculated
detector generations. The effects of xenon density based on two times the high-energy half width, the
on detector energy resolution have been reported resolution drops to 4%. A test pulse generator was
previously and will not be discussed here [10]. used to establish the base electronics noise. The
The data were obtained using an AmpTek 250 FWHM of the pulser peak was 2.2% for the
preamp operated at room temperature and operating conditions described above.
mounted directly to the anode. The anode Fig. 3a is a pulse height spectrum for the
capacitance was 22 pF. The output of the pre- detector exposed to a 133Ba source and Fig. 3b is
amplifier was fed to an Ortec shaping amplifier a pulse height spectrum for a 22Na source under
and then into a multi-channel analyzer for pulse the same detector conditions as Fig. 2. The
246 R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250
1400
1200
1000
800
Counts
600
400
200
0
100 150 200 250 300 350 400 450 500 550 600
Channel Number
137
Fig. 2. Pulse height spectrum of a Cs calibration source at a cathode voltage of 8 kV.
detector energy resolution was sufficient to clearly detector output pulse height is linearly dependent
resolve the two low energy peaks in the 133Ba on energy as is expected.
spectrum. However, due to the relatively low The effect of the electric field strength on the
xenon density, the high-energy peak in the 22Na charge collection was also investigated. Fig. 5 is a
spectrum is not clearly resolved. plot of the 137Cs peak channel number versus
Fig. 4 is a plot of gamma energy versus channel cathode voltage from approximately 5 to 9 kV
number. The data were obtained during a single at a shaping time of 5 ms. As can be seen in the
run with the detector simultaneously exposed to figure, the peak channel number (which is a
137
Cs, 22Na and 133Ba sources and a cathode measure of the total collected charge) increases
voltage of 8 kV. This plot demonstrates that the linearly with applied voltage and appears to begin
R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250 247
20000 6000
18000
5000
16000
14000
4000
12000
Counts
Counts
10000 3000
8000
2000
6000
4000
1000
2000
0 0
100 150 200 250 300 350 100 300 500 700
(a) Channel Number (b) Channel Number
133 22
Fig. 3. Pulse height spectrum of (a) Ba and (b) Na.
to saturate at a cathode voltage of 9 kV. It was a room temperature energy resolution of approxi-
not possible to probe higher voltages because of mately 6% @ 662 keV. This resolution is about a
limitations in the current feedthrough design. factor of three worse than the best energy
Fig. 6 is a plot of the electric field magnitude resolution reported for HPXe detectors incorpor-
versus radius for several applied cathode voltages. ating a shielding mesh, and is more typical of the
At a cathode voltage of 8 kV, the electric field best results reported for cylindrical detectors
magnitude ranges from a high of about 8 kV/cm at without a shielding mesh [4,7,8].
the anode to a low of about 1.5 kV/cm at the Therefore, although the hemispherical geometry
cathode electrode. appears promising, the resolution of our prototype
is still about a factor of 2 or 3 worser than is
necessary to offer a competitive advantage. How-
4. Discussion ever, we believe that there is significant room for
further optimization. Specifically, at this time, only
The principal goal of this work was to investi- one detector geometry (anode and cathode dia-
gate the possibility of developing a high resolution, meter) was tested and the xenon density was less
room temperature radiation spectrometer for field than optimal. Furthermore, the prototype detector
applications based on hemispherical HPXe detec- was only designed to operate reliably at cathode
tors. The hemispherical geometry provides a very voltages less than or equal to 8 kV and, based on
wide field of view as well as an excellent structural the results of Fig. 5, there is good reason to believe
design for containing the relatively high internal that the energy resolution could be improved by
xenon pressures. Without a Frisch grid the operating at higher cathode voltages. In the
prototype hemispherical HPXe detector provides hemispherical geometry, the electric field drops
248 R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250
1400
1200
1000
Gamma Energy (keV)
800
600
400
200
0
0 100 200 300 400 500 600 700
Channel Number
Fig. 4. Gamma energy versus channel number.
364
362
360
358
Channel Number
356
354
352
350
348
346
344
342
4000 5000 6000 7000 8000 9000 10000
Volts
Fig. 5. Peak channel number versus cathode voltage.
R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250 249
10.000
9.000
10kV
8.000
7.000
Electric Field Magnitude (kV/cm)
8kV
6.000
5.000
6kV
4.000
3.000
2.000
1.000
0.000
1.5 2 2.5 3 3.5 4 4.5 5
Radius (cm)
Fig. 6. The electric field distribution for four different cathode voltages. The operational voltage (8 kV) is indicated as the solid line.
off rapidly, as 1=r2 ; resulting in the need for anode diameter. However, the effect of the anode
relatively high cathode voltages in order to sustain diameter on the energy resolution is not clear.
an acceptable field magnitude at all locations Increasing the anode diameter decreases the effect
within the ionization region. However, at the of the 1=r2 electric field concentration and, there-
same time, the distance between the anode and fore, may be problematic in terms of the detector
cathode electrodes must be large enough to energy resolution. Further research is required in
provide adequate detection efficiency for gamma order to optimize the detector geometry to provide
radiation. One solution to this problem is to the best combination of energy resolution, effi-
increase the anode diameter [3]. Because the ciency and voltage requirements.
volume of a hemisphere is proportional to r3 ; the The electron drift velocity in HPXe saturates at
anode electrode occupies a negligible portion of approximately 105 cm/s for electric fields greater
the total ionization volume, even at relatively large than 2 kV/cm [11,12]. According to Fig. 6, at a
diameters. However, for a given cathode diameter, cathode voltage of 8 kV, the electric field
the minimum voltage required between the anode magnitude is greater than 2 kV/cm over more than
and cathode electrodes decreases with increasing 90% of the drift region. Therefore, despite the fact
anode diameter. Therefore, in a hemispherical that the electric field magnitude is increasing
geometry, it is possible to reduce the required rapidly as the electrons approach the anode, the
detector voltage without significantly decreasing electron drift velocity will be relatively constant
the detector volume simply by increasing the and equal to the saturation value of 105 cm/s.
250 R. Kessick, G. Tepper / Nuclear Instruments and Methods in Physics Research A 490 (2002) 243–250