You are on page 1of 9

bs_bs_banner

© 2011, Copyright the Authors


Artificial Organs © 2011, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Validation of an Axial Flow Blood Pump: Computational


Fluid Dynamics Results Using Particle Image Velocimetry

*Boyang Su, †Leok Poh Chua, and ‡Xikun Wang

*DHI-NTU Water & Environment Centre and Education Hub, Nanyang Technological University; †School of Mechanical
and Aerospace Engineering, Nanyang Technological University; and ‡Maritime Research Centre, Nanyang Technological
University, Singapore

Abstract: A magnetically suspended axial flow blood transparent model, the blood analog fluid with the refrac-
pump is studied experimentally in this article. The pump tive index close to that of acrylic was used to avoid
casing enclosed a three-blade straightener, a two-blade refraction. According to the CFD results, the axial flow
impeller shrouded by a permanent magnet-embedded cyl- blood pump could generate adequate pressure head at the
inder, and a three-blade diffuser. The internal flow fields rotating speed of 9500 rpm and flow rate of 5 L/min, and
were simulated earlier using computational fluid dyna- the same flow condition was applied during the PIV
mics (CFD), and the pump characteristic curves were measurement. Through the comparisons, it was found that
determined.The simulation results showed that the internal the experimental results were close to those obtained by
flow field was basically streamlined, except the diffuser CFD and had thus validated the CFD model, which could
region. Particle image velocimetry (PIV) measurement of complement the limitation of the measurement in assessing
the 1:1 pump model was conducted to validate the CFD the more detailed flow fields of the axial flow pump. Key
result. In order to ensure the optical access, an acrylic pro- Words: Computational fluid dynamics—Axial flow blood
totype was fabricated with the impeller driven by a servo- pump—Particle image velocimetry.
motor instead, as the magnet is opaque. In addition to the

Each year, approximately 30 000 patients at risk of 6000~12 000 rpm (8–11). The hemolysis induced by a
dying from severe heart failure are awaiting heart rotary blood pump is the main concern besides the
transplant, but only around 3500 cases are performed hydraulic performance (12). It is found that the
(1) mainly constrained by the limited heart donors hemolysis is significantly related to the flow behavior,
available. Ventricular assist devices (VADs) have and the scalar shear stress and exposure time are two
been developed for decades as an alternative main parameters used to estimate the hemolysis (13).
therapeutic modality and can be produced in the Therefore, the rotary blood pump poses a challenge,
quantity needed. VADs are used for bridge-to- especially for axial flow blood pump with higher
transplantation, bridge-to-recovery, or even long- rotating speed.
term cardiac support (2,3). The continuous-flow Computational fluid dynamics (CFD) has been
VADs usually include a rotating impeller to generate widely used as a powerful tool to investigate many
adequate flow rate and pressure head, and so are also blood-contacting medical devices including blood
named as rotary blood pumps. Typically, the rotating pumps (14,15). With the aid of CFD, the design phase
speed of a centrifugal blood pump is 2000~4000 can be shortened significantly from several years to
rpm (4–7), and that of an axial blood pump is several months (16). It can also support and explain
experimental results. Compared with the traditional
design, CFD can predict the detailed flow fields,
doi:10.1111/j.1525-1594.2011.01362.x pump characteristics curves, and other parameters
Received November 2010; revised July 2011. such as hemolysis (12,17). However, the numerical
Address correspondence and reprint requests to Mr. Boyang Su, results still need validating using the nonintrusive
School of Mechanical and Aerospace Engineering, Nanyang Tech-
nological University, Nanyang Drive, Singapore 639798, Singapore.
measurements, although they are usually difficult and
E-mail: bysu@ntu.edu.sg limited by the optical access. For the whole flow field,

Artificial Organs 2012, 36(4):359–367

aor_1362 359..367
360 B. SU ET AL.

particle image velocimetry (PIV) technique is widely


adopted (18–20), as compared with the point-wise
hot wire (21) or laser Doppler anemometer (22,23).
The internal flow seeded with particles is exposed by
double-frame laser sheets, and the velocity vector
map is calculated using a cross-correlation technique
in PIV measurement.
Due to the complexity of the structure of an axial
flow blood pump, which usually includes a straight-
ener, impeller, and diffuser, and its high rotating
speed, there are only few studies using PIV to
measure its internal flow (14,20) as compared with
the centrifugal blood pump (18,19,24–26). Therefore,
FIG. 2. Grid independence test.
in this study, the axial flow blood pump was analyzed
by both CFD and PIV techniques, and both numeri-
cal and experimental results were compared for
validation. fire 4.0 (Parametric Technology Corporation,
Needham, MA, USA), and was then exported into
commercial grid generator software ICEM (ANSYS,
MATERIALS AND METHODS Inc., Canonsburg, PA, USA). The numerical model
consisted of three domains, that is, straightener,
Physical model impeller, and diffuser, with both straightener and dif-
The axial blood pump comprises of three compo- fuser prolonged to make the flow develop. The grid
nents: a three-blade straightener, an enclosed two- independence test was conducted based on the pres-
blade impeller, and a three-blade diffuser as shown in sure head across the blood pump as shown in Fig. 2. It
Fig. 1. Its overall axial length and outer diameter of can be found that the pressure head difference with
the inner components are 60 and 19 mm, respectively. respect to the finest model was less than 1% as the
The impeller was the only rotating component grid number of the numerical model increased to 1.3
enclosed inside a cylinder, which was driven by the million. Therefore, the model with 1.3 million grids
Lorentz force. As a result, the impeller and the cylin- was selected as a compromise between the accuracy
der rotated simultaneously as a single component and computational resource. In between the clear-
(27). As the magnet was installed in the rotor cylin- ance gap between the pump casing and the impeller
der, there was a clearance gap between the rotor shroud, 10 layers of grid cells were generated in the
cylinder and the pump casing. The current diffuser selected model. The skew angles of these grids were
has only three blades for the ease of fabrication and between 20 and 160° with the y+ value of grids at
PIV measurement. near-wall region maintained around 1, in order to
ensure the quality of grids. Figure 3a shows the
CFD method surface grids on the inner components and the longi-
The axial flow blood pump was generated using tudinal cross section with the complete computa-
computer aided design software Pro/Engineer Wild- tional domain is shown in Fig. 3b.
The numerical simulation was conducted using the
finite-volume CFD code FLUENT 6.3.26. The blood
was assumed to be Newtonian with a viscosity of
0.0035 Pa·s and a density of 1050 kg/m3 (28). At the
rotating speed of 9500 rpm, the Reynolds number
based on the impeller tip was about 3.4 ¥ 104, so the
internal flow was turbulent.The shear stress transport
(SST) k-w turbulence model developed by Menter
(29) was selected because it combines the advantage
of k-w and k-e models in the near-wall regions and
outside the boundary layers, respectively (26,30,31).
In addition, it predicts better results in adverse pres-
sure gradients and separating flow (32). The sliding
FIG. 1. The axial flow VAD. mesh model was applied to accurately predict the

Artif Organs, Vol. 36, No. 4, 2012


VALIDATION OF BLOOD PUMP USING PARTICLE IMAGE VELOCIMETRY 361

FIG. 3. Surface grids on the axial flow


(a) Detailed surface grids blood pump. (a) Detailed surface grids, (b)
complete numerical domain.

(b) Complete numerical domain

rotor–stator interaction with 300 time steps in one The test model is partially fabricated from acrylic for
period (33). Velocity inlet and pressure outlet were the illumination of the laser sheet and image pairs
applied at the pump inlet and outlet, respectively. The capturing as demonstrated in Fig. 4b. In order to
static pressure head across the numerical model was minimize the vibration of the operating impeller, a
used as the macro parameter to monitor the conver- bearing holder, which is a hollow step cylinder, is
gence, and the iteration was stopped when the con- designed as shown in Fig. 4c. Its front end is designed
vergence criterion of 10-4 was satisfied (34). to place the hollow diffuser in position with the diam-
eter of the diffuser hub equal to that of the bearing
Experimental method holder at Section B. Note that the influences of the
Figure 4a shows the schematic drawing of the con- vibrations of the inner components such as impeller
structed experimental test rig. The pump model con- and diffuser are not considered in this study. The
structed is the exact size of the prototype in order to straightener, impeller, and diffuser are all made of
avoid any scaling problem that alters the results. acrylic as well as the impeller shroud that is glued
When the pump operates, the flow rate is controlled together with the impeller. The impeller is driven by
by the throttle valve located downstream of the flow- a steel shaft, which is inserted through the bearings
meter, and the pressure head across the pump model and bearing holder. These components are enclosed
and the flow rate is obtained from the pressure trans- in the chamber, which is capped by two components
ducer and the flowmeter, respectively. connected to the test rig.
In order to reduce the possibility of hemolysis, The PIV system used in the present study was the
magnetic suspension technique is widely adopted in Dantec FlowMap PIV system (Dantec Dynamics
rotary blood pumps, including the present study. A/S, Skovlunde, Denmark). The Gemini PIV 200-15
However, it is impossible to perform PIV measure- (New Wave Research, Fremont, CA, USA), a double-
ments due to the opaque magnets. Therefore, the cavity pulsed neodymium yttrium aluminum garnet
impeller is driven by a servomotor through a shaft. (Nd:YAG) laser system, was used as the illumination

Artif Organs, Vol. 36, No. 4, 2012


362 B. SU ET AL.

(a)

(b)

FIG. 4. Schematic view of (a) the test rig


and (b) test model mounted on the
traverse stage in PIV measurement.
(c) CAD drawing of the bearing holder.

(c)

Artif Organs, Vol. 36, No. 4, 2012


VALIDATION OF BLOOD PUMP USING PARTICLE IMAGE VELOCIMETRY 363

source. The two pulsing beams from a double-cavity volume (35). Its refractive index was 1.488 to match
Q-switched Nd:YAG laser are combined and formed with that of the acrylic model with a density of
into pulsing light sheets using optics. Q-switched 1155 kg/m3 and a dynamic viscosity of 3.8 mPa·s. The
pulse can effectively freeze the motion of particle, kinetic viscosities of the working fluid and blood
and two cavities ensure the widest possible ranges of were 3.29 and 3.33 mm2/s, respectively, so as to main-
time between pulses. The CCD camera used in tain Reynolds number similarity. The solution was
the FlowMap PIV system is HiSense MkII seeded with neutral buoyant hollow glass spheres
camera (Hamamatsu C8484-05 digital CCD chip with the mean diameter of 10 mm. In the present PIV
(Hamamatsu City, Shizuoka, Japan); resolution measurement, 100 image pairs were captured at each
1280 ¥ 1024) equipped with Nikon AF Micro-Nikkor measured area. The measurement area was 10.76 ¥
60 mm f/2.8D (Tokyo, Japan). FlowManager software 8.26 mm2 with the scale factor, S, of 1.25. The record-
(Dantec Dynamics A/S) controlled the synchroniza- ing aperture was f/5.6~8, and thus, the camera image
tion sequence to acquire data, and stored the record- of the seeding particle had a diameter 2~3 pixels
ing data. The “once-per-revolution” signal from the under the wavelength of 532 nm. The interrogation
shaft encoder was sent to the processor, which then area of 32 ¥ 32 pixels was used with 50% overlap. The
fires the laser and acquires the images from the PIV pulse duration of this PIV is 10 ns, so the particle
camera as shown in Fig. 3a, so images could be taken movement during this time duration was negligible.
at a certain angular position of the impeller. More- In the straightener, the pulse separation varied
over, the timing signals are used to control the 10~25 ms at different measurement areas according to
camera to make sure that the CCD sensor is only the local flow velocity. In order to minimize the
exposed to the light from the illumination system, as effects of the out-of-plane velocity component, the
the ambient light will reduce the signal : noise ratio. thickness of the laser sheet was 0.5 mm, and the time
In order to measure the flow field at a different between light pulses was 10 ms so as to satisfy the
angular position, of the impeller a delay was equation below (36):
set between the external signal and the image
acquisition. Two successive images are then trans- vz ⋅ Δt
≤ 0.25
ferred to the FlowMap processor and stored finally. Zl
As a result, a pair of seeding particle images with
time delay was captured at a certain impeller angular where vz is the out-of-plane velocity, Dt is the time
position, and the velocity vector map was derived between pulses and Zl is the thickness of the laser
from the movement of particles during the time sheet. Postprocessing validation methods including
delay. peak height validation and velocity range validation
In order to avoid refraction, the working fluid were applied on the raw vector. Finally, the mean
involved mixing 40% of glycerin and 60% of aqueous phase-averaged velocity profiles over 100 recordings
ammonium thiocyanate (NH4SCN) solution by were obtained in each region.

FIG. 5. Experimental and numerical


pump characteristic curves.

Artif Organs, Vol. 36, No. 4, 2012


364 B. SU ET AL.

RESULTS AND DISCUSSION

Figure 5 compares the pump performance curves


of both CFD and experimental results at different
rotating speeds. Note that each point corresponds to
a specific flow rate and rotating speed under steady
flow condition. Both the experimental and CFD
simulation results show that the static pressure head
decreases with the increasing flow rate. It can be
found that CFD simulations overestimate the perfor-
mance, mainly due to the obstruction of the bars as
indicated by the black color in Fig. 4c. However, they (a) straightener
follow the trends of the experimental results. In order
to match the operating conditions used in the
numerical simulation, the impeller rotating speed and
the pump flow rate were 9500 rpm and 5 L/min,
respectively, in the PIV experiment.
In order to show the flow field between the hub
(R = 5.5 mm) and pump casing (R = 7.5 mm), Fig. 6
shows the longitudinal cross-sectional planes through
the Z-axis in each region. Figure 7 shows the com-
parisons of velocity profiles between the CFD and
PIV results in the straightener, impeller, and diffuser
regions, respectively. In these figures, the solid line
and dotted line denote PIV and CFD results,
(b) impeller
respectively. In Fig. 7a, the flow in the straightener
region is smooth in the positive axial direction. At the
region close to the walls of the hub and casing of the
pump, the fluid has lower speed due to the boundary
layer effect as compared to that at the middle of the
channel. The axial velocity profile is almost parabolic
in shape. Due to the reduced cross-sectional area,
the fluid is accelerated as it flows through the
straightener.
As the fluid flows into the impeller channel, the
dominant flow is in the positive Z direction, and
the high axial velocity is located slightly close to the
impeller hub (R = 5.5 mm), especially further down- (c) diffuser
stream until Z = 18 mm, as shown in Fig. 7b. As a
result, the fluid velocity close to the impeller shroud FIG. 6. Demonstration of the cross-sectional planes in each
region. (a) Straightener, (b) impeller, (c) diffuser.
(R = 7.5 mm) is much smaller than that close to the
hub (R = 5.5 mm). From Z = 18 mm onward, it can be
observed that the forward velocity at the impeller
shroud has increased in magnitude, while near diffuser is the most irregular with strong reverse flow
the impeller exit, reverse flow is detected close to the close to the diffuser hub (R = 5.5 mm). It is because
impeller hub. Because the diffuser placed at the the fluid there has lower velocity and cannot over-
impeller downstream induces an adverse static come the adverse pressure gradient. However, the
pressure gradient at the impeller exit, the fluid near forward flow becomes more dominant from Z =
the rotating impeller hub, which has lower momen- 26.5 mm onward where the blade ends. A similar flow
tum than that near the impeller shroud, cannot pattern is predicted by the numerical simulation.
overcome it and flow is reversed. A qualitative agreement between the numerical
Figure 7c shows the velocity vector distributions in and experimental results can be observed, especially
the diffuser. Compared with the flow patterns in both in the straightener followed by the impeller and the
straightener and the impeller, the flow pattern in the diffuser. The discrepancy becomes relatively obvious

Artif Organs, Vol. 36, No. 4, 2012


VALIDATION OF BLOOD PUMP USING PARTICLE IMAGE VELOCIMETRY 365

(a) Straightener

FIG. 7. Comparison of the velocity pro-


files obtained from PIV (solid line) and
CFD (dotted line) results in (a) straight-
ener, (b) impeller, and (c) diffuser.

(b) Impeller

(c) Diffuser

at the regions close to blades, hub, and inner casing laser thickness and the time between pulses as men-
surfaces due to a number of reasons. First, the tioned earlier in the experimental method section.
trapped seeding particles on surfaces could induce The L2 relative error norm of the differences in axial
relatively strong reflection, resulting in error in PIV velocity between CFD and PIV with respect to the
measurements. Second, the fluid close to the inter- PIV in the straightener, impeller, and diffuser is 11,
faces usually has higher velocity gradient in radial 19, and 27%, respectively. Note that the L2 relative
direction, and it could cause error due to the difficulty error norm is defined as
in PIV measurements. In order to calculate the esti-
N
mated velocity accurately, the seeding particles
within an interrogation area, where the camera image ∑ ( fi E − fi N )2
i
is usually divided into 32 pixels by 32 pixels in size, N

should be homogenous, and have exactly the same ∑ ( fi E )2


i
displacement. However, the displacement of indi-
E N
vidual particles is always different from each other where fi and fi represent the values at node
under high velocity gradient. Third, the out-of-plane i for PIV experiment and numerical simulation,
motion of the flow field will introduce signal dropout. respectively. These values are consistent with the
However, it has been minimized by selecting proper observation earlier.

Artif Organs, Vol. 36, No. 4, 2012


366 B. SU ET AL.

CONCLUSION ventricular assist device for infants and children. ASAIO J


2007;53:754–61.
Computational fluid dynamics was initially used to 11. Zhang Y, Zhan Z, Gui X, et al. Design optimization of an axial
blood pump with computational fluid dynamics. ASAIO J
study the internal flow field of the axial flow blood 2008;54:150–5.
pump.The experimental test showed that the numeri- 12. Arvand A, Hormes M, Reul H. A validated computational
cal simulation could predict the pump characteristic fluid dynamics model to estimate hemolysis in a rotary blood
pump. Artif Organs 2005;29:531–40.
curves with similar trends. Furthermore, the particle 13. Paul R, Apel J, Schugner S, Shwindke P, Reul H. Shear stress
image velocimetry measurement was conducted for related blood damage in laminar couette flow. Artif Organs
the validation at the flow conditions of 9500 rpm 2003;27:517–29.
14. Triep M, Brucker C, Schroder W, Siess T. Computational fluid
rotating speed and 5 L/min flow rate. Generally, the dynamics and digital particle image velocimetry study of the
CFD results agree fairly well with the PIV results, flow through an optimized micro-axial blood pump. Artif
especially in the straightener region. It is also found Organs 2006;30:384–91.
15. Zhang J-M, Chua LP, Ghista DN, Zhou T-M, Tan YS. Valida-
that the matching between numerical and measured tion of numerical simulation with PIV measurements for two
results is related to the flow pattern, that is, a better anastomosis models. Med Eng Phys 2008;30:226–47.
agreement is always obtained in a smooth flow field 16. Burgreen GW, Antaki JF, Wu ZJ, Holmes AJ. Computational
fluid dynamics as a development tool for rotary blood pumps.
such as in the straightener region. The major differ- Artif Organs 2001;25:336–40.
ences between the experimental and simulation 17. Goubergrits L, Affeld K. Numerical estimation of blood
results are usually found near interfaces, due to the damage in artificial organs. Artif Organs 2004;28:499–
507.
high velocity gradients, strong reflections from the 18. Sankovic JM, Kadambi JR, Mehta M, Smith WA, Wernet MP.
seeding particles trapped there, and the high out- PIV investigations of the flow field in the volute of a rotary
of-plane velocity component. Basically, a fair agree- blood pump. J Fluids Eng 2004;126:730–4.
19. Day SW, McDaniel JC. PIV measurements of flow in a cen-
ment of the simulations and measurements can be trifugal blood pump: steady flow. J Biomech Eng 2005;127:244–
observed, and it has reasonably validated the simula- 53.
tion results. Therefore, SST k-w turbulence model is 20. Triep M, Brucker C, Kerkhoffs W, Schumacher O, Marseille O.
Investigation of the washout effect in a magnetically driven
able to predict the flow phenomenon in the axial flow axial blood pump. Artif Organs 2008;32:778–84.
blood pump. 21. Chua LP, Ong KS, Yu CMS, Chan WK, Wong YW. Gap veloc-
ity measurements of a blood pump model. Artif Organs
2002;26:682–94.
22. Chua LP, Ong KS, Song G. Study of velocity and shear stress
REFERENCES distributions in the impeller passages and the volute of a bio-
centrifugal ventricular assist device. Artif Organs 2008;32:376–
1. Mielniczuk L, Mussivand T, Davies R, et al. Patient selection 87.
for left ventricular assist devices. Artif Organs 2004;28: 23. Chua LP, Ong KS, Song G, Ji W. Measurements by laser
152–7. doppler velocimetry in the casing/impeller clearance gap of a
2. Borovetz HS, Ramasamy N, Zerbe TR, Portner PM. Evalua- biocentrifugal ventricular assist device model. Artif Organs
tion of an implantable ventricular assist system for humans 2009;33:360–72.
with chronic refractory heart failure. ASAIO J 1995;41:42– 24. Day SW, McDaniel JC, Wood HG, Allaire PE, Landrot N,
8. Curtas A. Particle image velocimetry measurements of blood
3. Tayama E, Olsen DB, Ohashi Y, et al. The DeBakey ventricu- velocity in a continuous flow ventricular assist device. ASAIO
lar assist device: current status in 1997. Artif Organs 1999;23: J 2001;47:406–11.
1113–6. 25. Day SW, McDaniel JC, Wood HG, et al. A prototype
4. Anderson JB, Wood HG, Allaire PE, Olsen DB. Numerical HeartQuest ventricular assist device for particle image veloci-
analysis of blood flow in the clearance regions of a continuous metry measurements. Artif Organs 2002;26:1002–5.
flow artificial heart pump. Artif Organs 2000;24:492–500. 26. Song X, Wood HG, Day SW, Olsen DB. Studies of turbulence
5. Song X, Throckmorton AL, Wood HG, Antaki JF, Olsen DB. models in a computational fluid dynamics model of a blood
Quantitative evaluation of blood damage in a centrifugal pump. Artif Organs 2003;27:935–7.
VAD by computational fluid dynamics. J Fluids Eng 2004;126: 27. Lim TM, Zhang D. Development of Lorentz force-type self-
410–8. bearing motor for an alternative axial flow blood pump design.
6. Chua LP, Song G, Yu SCM, Lim TM. Computational fluid Artif Organs 2006;30:347–53.
dynamics of gap flow in a biocentrifugal blood pump. Artif 28. Chua LP, Su BY, Lim TM, Zhou TM. Numerical simulation of
Organs 2005;29:620–8. an axial blood pump. Artif Organs 2007;31:560–70.
7. Legendre D, Antunes P, Bock E, Andrade A, Biscegli JF, Ortiz 29. Menter FR. Two-equation eddy-viscosity turbulence
JP. Computational fluid dynamics investigation of a centrifu- models for engineering applications. AIAA J 1994;32:1598–
gal blood pump. Artif Organs 2008;32:342–8. 605.
8. Mitamura Y, Nakamura H, Okamoto E, Yozu R, Kawada S, 30. Zhu L, Zhang X, Yao Z. Shape optimization of the diffuser
Kim D-W. Development of the Valvo pump: an axial flow blade of an axial blood pump by computational fluid dynamics.
pump implanted at the heart valve position. Artif Organs Artif Organs 2010;34:185–92.
1999;23:566–71. 31. Kido K, Hoshi H, Watanabe N, et al. Computational fluid
9. Zhu X, Zhang M, Zhang G, Liu H. Numerical investigation on dynamics analysis of the pediatric tiny centrifugal blood pump
hydrodynamics and biocompatibility of a magnetically sus- (TinyPump). Artif Organs 2006;30:392–9.
pended axial blood pump. ASAIO J 2006;52:624–9. 32. Fraser KH, Taskin ME, Griffith BP, Wu ZJ. The use of com-
10. Throckmorton AL, Untaroiu A, Allaire PE, et al. Numerical putational fluid dynamics in the development of ventricular
design and experimental hydraulic testing of an axial flow assist devices. Med Eng Phys 2011;33:263–80.

Artif Organs, Vol. 36, No. 4, 2012


VALIDATION OF BLOOD PUMP USING PARTICLE IMAGE VELOCIMETRY 367

33. Chua LP, Song GL, Zhou TM. Numerical analysis of the inner 35. Doyle MG, Vergniaud JB, Tavoularis S, Bourgault Y.
flow field of a biocentrifugal blood pump. Artif Organs Numerical simulations of blood flow in artificial and natural
2006;30:467–77. hearts with fluid–structure interaction. Artif Organs 2008;
34. Throckmorton AL, Untaroiu A, Allaire PE, et al. Computa- 32:870–9.
tional analysis of an axial flow pediatric ventricular assist 36. Dantec-Dynamics. Flowmanager software and introduction to
device. Artif Organs 2004;28:881–91. PIV instrumentation (software user’s guide). 2002.

Artif Organs, Vol. 36, No. 4, 2012

You might also like