RTFI Q&E (CTMS 308196) PARTICIPANT INFORMATION
INTRODUCTION
Radiography is a non-destructive testing (NDT) method that is used for
volumetric testing of components. Radiography uses x-rays or gamma rays to
produce a latent image of a test specimen on a radiographic film. Prior to
radiography, each component should be visually inspected to determine if it
meets the requirements of the applicable code or standard.
A radiographic film interpreter is an experienced and knowledgeable
individual who can judge the quality of a radiograph, identify discontinuities
on a radiograph and evaluate if these discontinuities meet the applicable
code or standard. This module will describe the terms and applications that
will enable the interpreter to evaluate film quality and component integrity.RTFI Q&E (CTMS 308196) PARTICIPANT INFORMATION
RADIATION SOURCES
The two types of radiation sources are x-rays and gamma rays. They are similar
in characteristics and differ in their source of origin
X-rays are produced in a vacuum tube when high energy electrons produced
by a cathode filament interact with an anode target that emits x-rays. The
production of x-rays is stopped as soon as the x-ray machine is switched off
Figure 1 shows an x-ray tube
CATHODE FOCUSING
FILAMENT
; \
a ee
|
(GLASS ENVELOPE:
Figure 1. Basic x-ray tubeRTFI Q&E (CTMS 308196) PARTICIPANT INFORMATION
RADIATION SOURCES (CONT’D)
Gamma rays ate produced by the nuclei of radioisotopes which are
undergoing disintegration because of their instability. The two types of gamma
fay sources used in Saudi Aramco are Iridium-192 and Cobalt-60. Ir-192 is used
for radiography of steel and similar materials of up to 3 inches in thickness and
Co-60 has a thickness range of up to 8 inches. A gamma ray source is
constantly emitting radiation and cannot be turned off. When the source is
stored inside the exposure device, the surrounding areas are shielded from
the source. Figure 2 shows a gamma ray source exposure device.
a SHIELD PIG
"STORED"
‘SOURCE STORED IN SHIELDED PIG
SOURCE CRANKED PARTLY OUT.
SOURCE NO LONGER SHIELDED.
wy.
i
oT
SOURCE ALL THE WAY OUT TO
EXPOSURE POSITION. "ON"
SITCHIN TIP TS ACTIVATED.
Figure 2.RTFI Q&E (CTMS 308196) PARTICIPANT INFORMATION
RADIOGRAPHIC TECHNIQUES
Source Positioning
The radiation source should be positioned where the radiation beam is as
close to being perpendicular to the test specimen and film being
radiographed. See Figure 3 below.
specimen
Figure 3,
Location Markers
Location markers (normally lead letters) are used to identify the location of the
film positioning on the test specimen. They shall be placed on the test
specimen and not on the film cassette. The location of the markers shall be
permanently identified on the test specimen with a permanent marker or low
stress metal stamp. The required number of markers varies with the exposure
arrangement and test specimen size. The requirements for location marker
placement is listed in SAEP-1143 par. 7.3 and also covered later in this course
in the section on interpretation.
Film Identification
Each radiograph shall be permanently identified by lead letters or with a light
flash technique. The requirements for film identification is listed in SAEP-1143.
par. 7.4 and also covered later in this course in the section on interpretation.RTFI Q&E (CTMS 308196) PARTICIPANT INFORMATION
RADIOGRAPHIC TECHNIQUES (CONT’D)
Image Quality Indicator (IQI) Placement
A source side IQI shall be used at all times unless it is not physically possible
to place the IQI on the source side, in which case a film side [QI shall be used.
When a film side IQ! is used, a lead letter ‘F’ at least as high as the IQ!
identification numbers shall be placed adjacent to a block-type IQI and on a
wire-type IQI.
When using a wire type IQ], it shall be placed perpendicular across the weld.
When using a block-type 1Q|, a shim of similar material approximately as thick
as the weld crown shall be placed under the IQI. The block-type IQI shall be
placed adjacent to the weld.
Each radiograph shall have an IQI, except when using exposure arrangement
‘A’ (panoramic) in which case there shall be four IQIs placed equally around
the circumference of the weld.
Film Placement
The radiographic film shall be placed as close as possible to the back of the
test specimen and as close to perpendicular to the radiation source.
Detection of Backscatter
Each film cassette shall have a lead letter ‘B’ which is 1/2” in height and 1/16”
in thickness attached to the back of the cassette to detect any backscatter.