ELIMIN-OX™

DR 900 Analytical Procedure AP-60/900

SPECIFICATIONS
Method: Colorimetric (DR 900)
Range: 180 to 6000 ± 100 ppb as ELIMIN-OX
Testing Time: 15 minutes

DESCRIPTION
The portable DR 900 colorimeter, an accurate,
simple method of colorimetric analysis, is utilized
to quickly analyze ELIMIN-OX concentration for
monitoring proper application levels. The test is
based on iron reduction chemistry and is simple to
perform. The ELIMIN-OX in the sample reduces
ferric iron to ferrous iron in direct proportion to
the ELIMIN-OX concentration. This solution then
reacts with Reagent 1 to form a purple color with
the ferrous iron.
Nalco DR 900 Colorimeter
ORDERING INFORMATION
Order all replacement parts and reagents by their
part numbers. To be sure this product is compatible
with your treatment program, contact your local
Description Part No.
Nalco representative. Items marked with an “*”
ELIMIN-OX , Reagent 1, 100/pkg 460-S0195.87
are included with the kit or set. To place your
ELIMIN-OX , Reagent 2 ,
order, please contact your local customer service
3.4 oz./100 mL 460-S0196.72
department.
Nalco DR 900 Colorimeter 400-P0900.88
Thermometer, lab, 0 to 230°F 500-P2028.88
REPLACEMENT PARTS AND REAGENTS Pipet, disposable, 1 mL 500-P2107.88
Sample cells, 10-20-25 mL w/cap,
Required Parts and Reagents
6/pkg 500-P2555.88
Description Part No. Flask, Erlenmeyer, glass, 250 mL 500-P1501.88
ELIMIN-OX 25 mL Reagent Set, Cap, plastic (for 500-P1501.88) P500-3037.88
100 Tests 460-P3115.88
Includes:
ELIMIN-OX Reagent 1* 100/pkg (2 included)
ELIMIN-OX Reagent 2* 3.4 oz/100 mL
Optional Parts and Reagents 2. Reaction glassware should be scrupulously clean.
For best results, isolate this glassware to be used
Description Part No.
only for this test. If glassware is con-taminated,
Thermometer, -40 to 450°F, digital 500-P2098.88
it can be cleaned by rinsing with 1:1 HCl (460-
Thermometer, 50 to 300°F, dial 500-P2019.88
S0726) followed by several iron-free distilled
Thermometer, 0 to 230°F, glass 500-P2028.88
water rinses. Hinckley and Schmitt distilled water
Pocket pH Meter 400-P1326A.88
is acceptable for use.
pH (3-color) test strips (0-14 pH)
(100/box) 500-P3204.88 3. Periodically clean the sample cells with a warm
pH (3-color) test strips (5–10 pH) solution of distilled water containing 1.0 mL of
(100/box) 500-P3207.88 Fe-4 (460-S0734) per 25 mL water. After soaking,
Sample Cooler, Threaded rinse the clean cells with a few mL of 1:1 HCl
(Inconel/SS) 500-P4625.88 (460-S0726) followed by at least 3 rinses with
Sample Cooler II (316 SS/MS) 500-P4627.88 iron-free distilled water.
Powder Pillow Snippers 500-P1605.88
INTERFERENCES
Description Size Part No.
Substances that reduce ferric iron (such as other
1:1 HCl 1.7 oz./50 mL 460-S0726.71
oxygen scavengers) can interfere. Substances that
1:1 HCl 3.4 oz./100 mL 460-S0726.72
strongly complex iron may also interfere. The fol-
1:1 HCl 16.9 oz./500 mL 460-S0726.74
lowing substances may interfere when present
1:1 HCl 34.8 oz./1 L 460-S0726.75
in concentrations exceeding those listed below.
1:1 HCl 1.06 gal./4 L 460-S0726.77
Fe-4 1.7 oz./50 mL 460-S0734.71
Fe-4 16.9 oz./500 mL 460-S0734.74 1 mg/L = 1 ppm
Ammonia 100 ppm
SAFETY PRECAUTIONS Borate (as Na2B4O7) 500 ppm
Cobalt 25 ppb
Refer to Material Safety Data Sheet (MSDS) for
specific reagent personal safety and disposal Copper 50 ppb
information. Cyclohexylamine 3 ppm
Hardness (as CaCO3) 1000 ppm
Hydrazine 250 ppb
SUPPORT
Iron (Ferrous) *See Note
If you have any questions regarding this procedure, Lignosulfonates 50 ppb
please contact your Nalco representative. In North
Manganese 0.8 ppm
America you can also contact the Equipment Help
Molybdenum 80 ppm
Desk at 630-305-CHEM.
Morpholine 50 ppm
N7200, N7205 25 ppm
SAMPLING Nickel 0.8 ppm
1. This is a low level test and is very susceptible to Phosphate, phosphonates 10 ppm
contamination. Care must be taken to eliminate Sulfate 1000 ppm
sources of air dissolution and other interfering
Zinc 50 ppm
species. A cooler equipped with a copper
cooling coil should NOT be used. Trace levels of Light can also interfere with color development. Keep
copper catalyze the reaction of ELIMIN-OX and reacted samples in a dark area during the waiting step.
oxygen. As a result of copper contamination, Note: Any Ferrous iron will cause a positive interference and can be
any ELIMIN-OX present in the system may be blanked out by performing the procedure on the sample omitting
consumed prior to analysis. Step #6. Subtract this result from the sample’s EOx result to deter-
mine the sample’s true ELIMIN-OX concentration. Do not use this
iron blanking procedure to quantify Fe+2.
ELIMIN-OX DR 900 PROCEDURE
Special Notes: Analyze sample as soon as it is collected. The temperature of the sample should be 72-82°F (22-28°C).
Do not use a copper-based sample cooler as copper interferes with this reaction. If the temperature of the sample is not in
the proper range, correct the result as follows:
EOx – [9 X (C*1.8 - 45)] = EOx TC OR EOx – [9 X (°F - 77)] = EOx TC

Start

1 2 3 4
Start Program Collect a feedwater sample Fill one sample cell to the Add the contents of one (1)
after the ELIMIN-OX feed 25 mL mark with sample. ELIMIN-OX Reagent 1 pillow
732 ELIMIN-OX point. Stopper the flask with Fill a second cell with to each sample cell.
a plastic stopper (500- demineralized or distilled
P3037.88). water. This second cell will Swirl the cells to dissolve.
be the blank.
Notes: Do not use a Note: For accurate results
rubber stopper. Note: The sample use pillows from the same
temperature should be manufacturing lot.
Sampling technique is 72-82°F (22-28°C).
extremely important. Refer
to Sampling Section of this
procedure.

Zero

5 6 7 8
Add exactly 0.5 mL ELIMIN- Immediately after the When the timer beeps, the Press: ZERO
OX Reagent 2 to each cell. addition of the ELIMIN-OX display will show:
Mix. Place both cells in the Reagent 2, Start TIMER 1 The cursor will move to the
dark. and a 10 minute reaction µg/L EL-OX right, then the display will
period will begin. show:
After wiping any moisture or
Note: A purple color will fingerprints from the outside 0 µg/L EL-OX
slowly develop if ELIMIN- Notes: Keep both cells of the cell, place the blank
OX is present. in the dark during this into the cell holder and
period. tightly cover the sample cell
with the instrument cap.
Temperature and reaction
time affect the results.
Be sure these factors are
controlled as described.
Read

9 10
Immediately remove the Press: READ
blank cell and insert the
The cursor will move to
sample cell. Tightly cover
the right, then the result
the sample cell with the
in µg/L ELIMIN-OX will be
instrument cap.
displayed.

Nalco, an Ecolab Company


North America: Headquarters – 1601 West Diehl Road • Naperville, Illinois 60563 • USA
Nalco Champion – 7705 Highway 90-A • Sugar Land, Texas 77487 • USA
Europe: Richtistrasse 7 • 8304 Wallisellen • Switzerland
Asia Pacific: 2 International Business Park • #02-20 The Strategy Tower 2 • Singapore 609930
Latin America: Av. das Nações Unidas 17.891 • 6° Andar 04795-100 • São Paulo • SP • Brazil
www.nalco.com

Ecolab, ELIMIN-OX, NALCO and the logos are trademarks of Ecolab USA Inc.
©2013 Ecolab USA Inc. All Rights Reserved 8/13