You are on page 1of 3

MAGLUMI D-dimer(CLIA)

1. INTENDED USE
The kit has been designed for the quantitative determination of D-dimer in
human serum or plasma.
130206008M 100 The method can be used for samples over the range of 0-10 μg/ml.
The test has to be performed on the Maglumi fully auto analyzer (Including
Maglumi 1000, Maglumi 2000, Maglumi 2000 plus).

2. SUMMARY AND EXPLANATION OF THE TEST


D-dimer is a fibrin degradation product (or FDP), a small protein fragment
present in the blood after a blood clot is degraded by fibrinolysis. It is so
Shenzhen New Industries Lotus Global Co., Ltd named because it contains two crosslinked D fragments of the fibrinogen
Biomedical Engineering Co., Ltd 15 Alexandra Road protein.[1]
4F,Wearnes Tech Bldg,
London UK D-dimer concentration may be determined by a blood test to help diagnose
Science&Industry Park,
NW8 0DP thrombosis. Since its introduction in the 1990s, it has become an important
Nanshan,Shenzhen, China 518057
Tel. + 44-20-75868010 test performed in patients suspected of thrombotic disorders. While a
Tel. + 86 755 26508518
Fax.+ 44-20-79006187 negative result practically rules out thrombosis, a positive result can indicate
Fax.+ 86 755 26508339
thrombosis but does not rule out other potential causes. Its main use,
therefore, is to exclude thromboembolic disease where the probability is low.
In addition, it is used in the diagnosis of the blood disorder disseminated
intravascular coagulation.
D-dimers are not normally present in human blood plasma, except when the
coagulation system has been activated, for instance because of the presence
FOR PROFESSIONAL USE ONLY of thrombosis or disseminated intravascular coagulation. The D-dimer assay
Store at 2...8 °C depends on the binding of a monoclonal antibody to a particular epitope on
the D-dimer fragment. Several detection kits are commercially available; all of
them rely on a different monoclonal antibody against D-dimer. Of some of
COMPLETELY READ THE INSTRUCTIONS these it is known to which area on the D-dimer the antibody binds. The
BEFORE PROCEEDING
binding of the antibody is then measured quantitatively by one of various
laboratory methods
D-dimer testing is of clinical use when there is a suspicion of deep venous
en
thrombosis (DVT) or pulmonary embolism (PE). In patients suspected of
disseminated intravascular coagulation (DIC), D-dimers may aid in the
SYMBOLS USED ON LABELS diagnosis.[citation needed]
For DVT and PE, there are various scoring systems that are used to
Authorized Representative in Europe determine the a priori clinical probability of these diseases; the best-known
were introduced by Wells et al.
Manufacturer
3. PRINCIPLE OF THE TEST
Sandwich immunoluminometric assay;
Use an anti-D-dimer monoclonal antibody to label ABEI, and use another
Attention. See Instructions For Use monoclonal antibody to label FITC. Sample, Calibrator or Control, ABEI Label,
FITC Label and magnetic microbeads coated with anti-FITC are mixed
thoroughly and incubated at 37 ℃, forming a sandwich; after sediment in a
Contents of kit magnetic field, decant the supernatant, then cycle washing for 1 time
Subsequently, the starter reagents are added and a flash chemiluminescent
reaction is initiated. The light signal is measured by a photomultiplier as RLU
within 3 seconds and is proportional to the concentration of D-DIMER present
In vitro diagnostic medical device
in controls or samples.
(In vitro diagnostic use)
4. KIT COMPONENTS
Lot number
4.1 Material supplies

Catalogue Code Reagent Integral for 100 determinations


Nano magnetic microbeads: TRIS buffer, 1.2%(W/V),
0.2%NaN33, coated with sheep anti- FITC polyclonal 2.5ml
Expiry date (Use by…) antibody.
Calibrator low 2.5ml
Calibrator high 2.5ml
ABEI Label: anti-D-dimer monoclonal antibody labeled
Temperature limitation 6.5ml
( store at 2...8 °C) ABEI, contains BSA, 0.2%NaN3.
FITC Label: anti-D-dimer monoclonal antibody labeled
6.5ml
FITC, contains BSA, 0.2%NaN3.
Number of tests
All reagents are provided ready-to-use.

Keep away from sunlight Accessories required but not provided


Maglumi Reaction module
Maglumi Starter kit 1+2

Biological risks Maglumi Light check


Maglumi Wash /System Liquid

4.2 Preparation of the Reagent Integral


130206008M-v1.0-EN 1/3
Before the sealing is removed, gentle and careful horizontal shaking of the instructions of the Maglumi Fully Auto Analyzer. Each test parameter is
Reagent Integral is essential (avoid foam formation!) Remove the sealing and identified via a barcode on the Reagent Integral. For further information
turn the small wheel of the magnetic microbeads compartment to and fro, please refer to the Maglumi Fully Auto Operator’s Manual.
20μl Sample, calibrator or controls
until the colour of the suspension has changed into brown. Place the Integral
+40μl ABEI Label
into the reagent area and let it stand there for 30 mins. During this time, the
+40μl FITC Label
magnetic microbeads are automatically agitated and completely
+20μl Nano magnetic microbeads
resuspended. 10 min Incubation
Do not interchange nano magnetic microbeads from different reagents. 400μl each time Cycle washing
4.3 Storage of the Reagents Integral 3 sec Measurement
z Sealed:Stored at 2-8℃ until the expiry date.
z Opened : Stable for 4 weeks. After this period, it is still possible to
keep on using the Reagent Integral provided that the controls are 10. Quality Control
found within the expected ranges. z Observe quality control guidelines for medical laboratories.
z Keep upright for storage. z Use suitable controls for in-house quality control.
z Keep away from direct sunlight.
11 Results
5. Origin of Calibrators.
11.1 Calculation of Results
Calibrators in the Reagent Kit are from Fitzgerald.
Biological origin: Recombinant protein. No HBsAg, anti-HCV, and anti-HIV is The analyzer automatically calculates the D-DIMER concentration in
found. each sample by means of a calibration curve which is generated by a
2-point calibration master curve procedure. The results are expressed in
6. Calibration IU/ml. For further information please refer to the Maglumi Fully Auto
6.1 2 point recalibration
Operator’s Manual.
Via the measurement of calibrators, the predefined master curve is adjusted
(recalibrated) to a new, instrument-specific measurement level with each
11.2 Interpretation of Results
calibration. z Reference values: < 0.5 μg/ml
6.2 Frequency of Recalibration z Results may differ between laboratories due to variations in
z After each exchange of lot (Reagent Integral or Starter Reagents). population and test method. Each laboratory should establish its own
z Every week and/or each time a new Integral is used reference range.
(recommendation).
12. Limitations of the procedure
z After each servicing of the Maglumi Fully Auto analyzer.
z If controls are beyond the expected range. 12.1
Assay results should be utilized in conjunction with other clinical and
7. Sample Collection, Material and Storage
laboratory data to assist the clinician in making individual patient
z Collect samples using standard procedures.
management decisions.
z Sample material: serum or plasma.
z Store at 2-8℃: 24 hours. A skillful technique and strict adherence to the instructions are necessary to
z For longer storage periods: freeze to below - 20℃. obtain reliable results.
Procedural directions must be followed exactly and careful technique must be
z Avoid repeated freezing and thawing cycles.
used to obtain valid results. Any modification of the procedure is likely to alter
z Stored samples should be thoroughly mixed prior to use (Vortex
the results.
mixer).
Bacterial contamination or repeated freeze-thaw cycles may affect the test
Serum
results.
z elbow vein blood 5ml in the tube, centrifugation at room temperature,
serum was separated and stored at 2 ℃ - 8 ℃
12.2 HAMA
z Serum samples were stable for 12 hours at 2-8 ℃. If preserved more
Patient samples containing human anti-mouse antibodies (HAMA) may give
than 12 hours, please packed, -20 ℃ can be stored for 30 days,
falsely elevated or decreased values. Although HAMA-neutralising agents
z to avoid repeated freezing and thawing
are added, extremely high HAMA plasma concentrations may occasionally
Plasma:
influence results.
z elbow vein blood 5ml in the tube, then add EDTA anticoagulant,
centrifuged and separated plasma ,stored at2-8 ℃.
12.3 High-Dose Hook
z Plasma was stable at 2-8 ℃ for 12 hours. If preserved more than 12
No high-dose hook effect was seen for D-DIMER concentrations up to 50
hours, please packed, -20 ℃ can be stored for 30 days.
μg/ml.
z to avoid repeated freezing and thawing.

8. WARNING AND PRECAUTIONS FOR USERS 13. Performance Characteristics


z For use in IN-VITRO diagnostic procedures only.
13.1 Accuracy
z Do not interchange reagents from different lots. Do not use kit
Consider calibrator high of known concentration as a sample, dilute it by 1:2
components beyond their labeled expiry date.
ratio with diluents, and measure its diluted concentration for 10 times. Then
z All samples, biological reagents and materials used in the assay must
calculate the recovery of measured concentration and expected
be considered potentially able to transmit infectious agents. They
concentration. The recovery should be within 90% -110%.
should therefore be disposed of in accordance with the prevailing
regulations and guidelines of the agencies holding jurisdiction over
13.2 Precision
the laboratory, and the regulations of each country. Disposable
Intra-assay coefficient of variation was evaluated on Calibrator High
materials must be incinerated; liquid waste must be decontaminated
repeatedly measured 10 times in the same assay, calculating their coefficient
with sodium hypochlorite at a final concentration of 5% for at least half
of variation, the results should ≤10%.
an hour. Any materials to be reused must be autoclaved using an
Inter-assay coefficient of variation was evaluated on three batches of kit,
overkill approach(USP 24,2000,p.2143). A minimum of one hour at
repeatedly measured 10 times of Calibrator High, calculating three batches of
121℃ is usually considered adequate, though the users must check
kit for Calibrator High between the measured values of the coefficients of
the effectiveness of their decontamination cycle by initially validating it
variation, the results should ≤15%.
and routinely using biological indicators.
z The calibrators in this kit are prepared from bovine serum products.
13.3 Sensitivity
However, because no test method can offer complete assurance that
The sensitivity of the assay defined as the concentration of D-DIMER
HIV, Hepatitis B Virus or other infectious agents are absent, these
equivalent to the mean RLU of 20 replicates of the zero standard plus two
reagents should be considered a potential biohazard and handled
standard deviations corresponding to the concentration from the standard
with the same precautions as applied to any serum or plasma
curve. The sensitivity is typically less than 0.025 μg/ml.
specimen.
13.4 Specificity
9. Test Procedure
The specificity of the D-DIMER assay system was assessed by measuring
To ensure proper test performance, strictly adhere to the operating
the apparent response of the assay to various potentially cross reactive
130206008M-v1.0-EN 2/3
analytes.
When Troponin I=50μg/ml, the detection results of D-DIMER < 0.025μg/ml.

13.5 Linearity
Conduct a logarithmic transform to the RLU value and concentration value of
each standard. After a double logarithmic fitting, the absolute value of its
linearity should exceed 0.9800.

14. References
1.FreyburgerG.TrillaudH.LabroucheS.etal . Ddimer .
strategyinthrombosisexclusionaglodstandardstudyin100patientssuspectedof
deepveinthrombosisorpulmonaryembolism; 8DDmethodscompared [ J ] .
ThrombHaemost,1998,79:3237.
2.Lee AJ,Fowkes GR,Lowe GD,et al. Determinants of fibrin D-dimer in the
Edinburgh Artery Study[J]. Arteriosder Thromb Vasc Biol,1995,15(2):1094.
3. TIMI Ⅲ B investigators. Effects of tissue plasminogen activators and a
comparison of early invasive and conservative strategy in unstable angina
and non-Q wave myocardial infarction:Results of the TIMI Ⅲ B trial [ J ] .
Circulation,1994,89(8):1545.
4.CarrJM.MckinneyM.McDonaghJ.etal.Diagnosisofdisseminatedintravascula
rcoagulation:roleofD-dimer[J].AmJClinPathol,1989,91:280.

130206008M-v1.0-EN 3/3

You might also like