You are on page 1of 8

Article

pubs.acs.org/IECR

The Effect of Column Diameter and Packing Height on the Pressure


Drop and on the HETP of Structured Packings
L. Valenz, J. Haidl, and V. Linek*
Department of Chemical Engineering, Institute of Chemical Technology Prague, CZ-166 28 Prague 6 Czech Republic

ABSTRACT: The standardization of the measuring method for the determination of packing separation efficiency should start
with a full understanding of all the conditions that affect its measurements in order to maintain them at the identical level during
the measurement of different packings with the aim of comparing or evaluating the effects of the geometric modification of newly
See https://pubs.acs.org/sharingguidelines for options on how to legitimately share published articles.

developed/designed packings. The effect of the column diameter, the bottom concentration of the distillation mixture, and the
Downloaded via UNIVERSIDADE DO PORTO 01100 on August 17, 2021 at 16:58:35 (UTC).

packing height on the HETP and the pressure drop are evaluated on the basis of the data measured in columns with diameters of
0.15, 0.43, and 1.22 m packed with RSP 250, Mellapak 250Y, or Montz BSH 250 structured packing, and utilizing the
cyclohexane/n-heptane distillation system. It was found that below the loading point, the pressure drop was only slightly affected
by the column diameter. The difference between the pressure drops of the 0.15 and 1.22 m diameter columns was within the
experimental data scatter. The effect of the concentration range of distillation mixture on the separation efficiency is more
pronounced than is the effect of the column diameter. HETP increases with increasing bottom concentration and packing height,
and the increase found in this paper reached 50%. The effect of the column diameter onto the packing separation efficiency is not
significant. The HETP, measured in columns with widely different diameters using a distillation mixture of approximately the
same concentration range, did not demonstrate any significant effect of the diameter. The presented data do not support the
recommendation of Ottenbacher et al. (Chem. Eng. Res. Des., 2011, 8, 1427) to consider as relevant only the packing efficiencies
measured on columns with a diameter of at least 0.4 m because the data from the smaller diameter columns are dominated by the
wall effects.

1. INTRODUCTION bulk and this results in a lower local pressure drop. In small
In the article “Structured Packing EfficiencyVital Information columns, there is a comparatively higher proportion of gas or
for The Chemical Industry”, Ottenbacher et al.1 define the vapor channels along the walls which results in a lower pressure
practices that should serve as an open standard for both the drop in comparison to columns with a wider diameter.
measurement and the interpretation of the separation efficiency The effect of the column diameter on pressure drops of
of structured packing. The standardization of the measuring structured packings was studied by Olujic.3 His results,
method is the basis for the comparison of the packings and for presented here in Figure 1, show an opposite behavior: in
the evaluation of the effect of geometric modifications of newly columns with a smaller diameter the Montz B1−250 structured
designed packings. The current state of the art is such that the packing has shown substantially higher pressure drops in
scatter of HETP measured by different teams on the same comparison with the drops measured in columns of a larger
packing can hide the separation efficiency improvements of diameter. For example, the pressure drops in a column with a
newly developed packings.
Ottenbacher et al.1 have recommended many improvements.
We consider as the most relevant the use of the CB/EB
(chlorbenzene/ethylbenzene) system instead of the C6/C7
(cyclohexane/n-heptane) system, since the HETP measured
using the C6/C7 system depends to a greater extent on the
concentration range in comparison with that measured using
the CB/EB system, due to the larger variation of the relative
volatility over the concentration range. The authors published a
comprehensive list of physical and thermodynamic properties
of the CB/EB system. For obtaining reliable efficiency values
the authors also recommend the use of columns with a
diameter of at least 0.4 m because the results from columns
with smaller diameters are dominated by their wall effects. Figure 1. Effect of column diameter on pressure drops in the air−
water system. Data of Montz B1 250 are redrawn from Olujic.3
There are only a few studies in the literature that deal with
the effects of the column diameter on pressure drop and the
HETP. Random packings in columns with a smaller diameter Received: September 5, 2012
achieve lesser pressure drops (Billet and Mackowiak2). The Revised: March 21, 2013
packing density close to the column wall (the number of Accepted: April 4, 2013
packing pieces per unit volume) is lower in comparison to the Published: April 4, 2013

© 2013 American Chemical Society 5967 dx.doi.org/10.1021/ie302397q | Ind. Eng. Chem. Res. 2013, 52, 5967−5974
Industrial & Engineering Chemistry Research Article

diameter of 0.2 m are 3-times greater in comparison to those dependence of the pressure drop on the column diameter
measured in a column with a diameter of 0.8 m. Olujic3 disappears. This limiting diameter is equal to 0.21 m for
concluded that “A practical consequence of this observation is Mellapak 250Y. In accordance with this model,6 the pressure
that predictions based on academic scale data can lead to grave drop of dry packing in a column with a diameter 0.15 m is 15%
overestimates in the pressure drop and consequently column greater than that in a column with a limiting diameter of 0.21
diameter, when applied to industrial scale columns.” The m.
recommendation to use columns with a diameter of at least 0.4 The independence of pressure drops and of the HETP of the
m for obtaining reliable data would make the equipment and its column diameter is well documented by the data measured by
operation more expensive and might discriminate against Billet7 and Meier et al.8 using the CB/EB distillation system
academic research which is usually conducted in columns and Mellapak 250Y. The data measured7 in the small column
with smaller diameters. Nevertheless, the following survey of with a diameter of 0.22 m coincided with those measured in the
the literature shows that there is not another work which large column with a diameter of 1.0 m in regard to the range of
confirms such a profound effect of the column diameter onto the vapor capacity factor of 0.4−2.8 Pa0.5. These data are
the pressure drop or HETP. presented in Figure 8.17 of Billet’s book.7 A similar degree of
The pressure drops of the Montz B1 250 packing presented consistency between the HETP measured in columns with
by Olujic3 are compared with those of the very similar Mellapak diameters of 0.16, 0.25, and 1 m was found by Meier et al.8
250Y packing in Figure 1. These packings are of the same size The data in the literature led us to the conclusion that the
and wettability of their surfaces by water and have the same recommendation to use the smallest 0.4 m diameter columns in
corrugation angle. The diameter effect on the pressure drop of order to obtain reliable data in regard to efficiency is not
the Mellapak is described in the Sulcol software package generally valid and should be analyzed in greater detail. First,
provided by the supplier Sulzer for its own packings. It was the situations should be differentiated for when data are being
deduced from experiments performed in columns of different measured for the design of large scale columns and when they
diameters. Olujic’s data show three-times greater pressure drops are being measured for comparing the separation efficiency of
in the column with a diameter of 0.2 m at Fv = 1 Pa0.5 than the packings or for evaluating the effect of the geometric
those measured in a 0.8 m diameter column, while the increase modifications on newly developed packings. The measurement
evaluated from the Sulcol is only 50% of that for the similar of data using the large-scale columns is financially more
Mellapak 250Y packing. demanding, and the data are more prone to error because of
At low liquid flow rates (B < 10 m/h), well below the loading misdistribution and the nonuniformity of the flow profile
point, when the upwardly flowing gas does not hinder the through the packing. Either of these phenomena may cause a
downward flow of the liquid, the dry and wet pressure drops of severe reduction in the separation efficiency and should be
the packing differ only slightly. For example, based on the data prevented during the packing test experiments. On the other
measured by Olujic,3 it makes only a maximum 15% difference hand the data from small scale columns may be dominated by
in a column with d = 0.2 m and is negligible in a column with d the wall effects. The hydraulic wall effects are demonstrated by
= 0.8 m. Olujic’s data in regard to dry and wet packing are pressure drops. Olujic paper 3 is the only systematic
presented in Figure 1. The data for dry Mellapak 250Y experimental study dealing with the effect of the column
measured by Stichlmair et al.4 in a column with a diameter of diameter on pressure drops in structured packings. However
0.3 m are compatible with those of the Sulcol software for d = the marked diameter effect revealed3 was not observed by other
0.4 m. We present this data to illustrate the accuracy and authors4,7 nor predicted by the available correlations5,6 (see also
reliability of this software. Sulcol). A similar systematic study dealing with the effect of the
The models described in the literature also provide column diameter and concentration of the distillation mixture
unambiguous results of the effect of the column diameter on we did not find in literature at all.
the pressure drop of the structured packings. The models that The aim of this study is to evaluate the effect of column
are based2 on the difference between the packing density close diameter on pressure drops and on separation efficiency and
the column wall and in comparison with the bulk also take into the dependence of the effect on the concentration range of the
account the wall factor K, defined as follows5 distillation mixture used during the measuring.
2
Δp a F 2 1 de
= λ 3 v K; K=1+ ; 2. EXPERIMENTAL SECTION
H ε 2 31−ε d
2.1. Columns and the Data Utilized. Apart from our own
1−ε
de = 6 measured data in Column 4, the data used are taken from the
ap (1) following reports: The pressure drop and the HETP measured
on the Raschig Super-Pak 250 (RSP250), the Mellapak 250Y,
On the basis of the large quantity of data in regard to various and the Montz BSH 250 packings, employing the C6/C7
types of packings, Mackowiak5 recommends the value of K = 1 distillation system in columns of different sizes are presented.
for structured packings, meaning that no effect of the column Within these reports the columns and the experimental and
diameter is expected. The model created by Brunazzi and evaluation procedures are described in detail. We provide here
Paglianti,6 based on the losses resulting from changes in the gas the following basic information concerning the columns and
flow direction that occur at the transitions between the packing the data utilized for this work.
elements, predicts some increase in the pressure drop in
columns with smaller diameters operating at the same specific Column 1 FRI: Report “Test of RSP250” by Cai T.J.,
gas and liquid flow rates. This increase is proportional to the King D.W., June 15th, 2009. Report made available for
number of bends per unit of the packing height. If the column the purposes of this study by Raschig. The data utilized:
diameter is greater than the limiting diameter d = Hp/tg θ, the Fv, the pressure drop, xbottom, HETP. Conditions: p =
number of bends depends only on the element height and the 1.62 b; d = 1.22 m; H = 3.5 m.
5968 dx.doi.org/10.1021/ie302397q | Ind. Eng. Chem. Res. 2013, 52, 5967−5974
Industrial & Engineering Chemistry Research Article

Column 2 FRI: Topical Report 108, 1990. Report made


available for the purposes of this study by Montz. The
data utilized: Fv, the pressure drop, xbottom, HETP, liquid
concentration profile measured at positions z = 0, 0.81,
1.42, 2.03, 2.64, 3.25, 4.01 m. Conditions: p = 1.62 b; d =
1.22 m.
Column 3 SRP: Report “Distillation Characteristics of
the Raschig-Jaeger RSP 250 Structured Packing” by
Seibert F., Perry M., Briggs S., Fair J., December 14,
2009. Report made available for the purposes of this
study by Raschig. The data utilized: Fv, the pressure drop,
xbottom, HETP. Conditions: p = 1.65 b; d = 0.43 m; H =
3.0 m.
Column 4 our laboratory. p = 1.0 b; d = 0.15 m; H = 2.1
m. Sketch of the column and one of the internals are
presented in Rejl et al.9 A description of this column
follows.
Column 4 was packed with 9 (RSP 250) or 10 (Mellapak Figure 2. Pressure drop and HETP of Raschig RSP 250 packing
250Y) pcs of structured packing to a height of 2.1 m. Each of measured in C6/C7 distillation system. Filled and empty symbols are
the packing units was equipped with two wall wipers that pressure drops of the packings with and without holes drilled for
redirected the wall flow back into the packing. The small sampling of phases.
diameter of the column and the large number of wipers justify
our assumption that the higher liquid flow close to the column
wall is noticeably suppressed. An electrically heated reboiler the C6/C7 distillation system were taken from Onken and
equipped with an electric power regulator maintained constant Arlt11 and the HETP was calculated from the relation
reboiler duty for the batches of different composition. The H
volumetric reflux flow rate was measured to an accuracy of HETP =
Ntheor (2)
0.001 L/min. A shower distributor with 25 openings (1472
holes/m2) was utilized. The column and the reboiler were
thermally insulated. The heat-loss, evaluated from the temper- 3. RESULTS
ature decrease of the hot water (90 °C) flowing through the 3.1. Effect of the Column Diameter on Pressure Drop
column represented only 3% of the lowest reboiler duty utilized and HETP in C6/C7 Distillation System. 3.1.1. Raschig
(8 kW). Super-Pak RSP 250. The pressure drop and the HETP
The column was provided with seven openings for placing measured using the C6/C7 distillation system in the columns
thermistor and sampling devices at regular distances along the of the diameters 0.15, 0.43, and 1.22 m and at the pressures of
column for the withdrawal of liquid and vapor samples. The 0.33, 1.0, and 1.65 bar are plotted in Figure 2 as a function of
holes for this sampling were drilled through the structured the gas capacity factor Fv. The pressure drops measured below
packing using a water jet. We believe that the sampling points the loading points in columns of different diameters and at
do not significantly affect the flow of phases in the packing different pressures vary only within the scatter of the
since the volume of the holes affects only 3.3% of the packing experimental points. As can be expected, the highest value of
volume and the pressure drop also differs undetectably between the loading point, recognized as the point at which the slope of
the drilled and undrilled packings (compare the empty and the the pressure dependence on the Fv starts to increase, was found
full symbols, shown in Figure 2), for the undrilled and the at the lowest pressure (0.33 b) and is equal to approximately
drilled packings, respectively. 3.2 Pa0.5. The 2.5 Pa0.5 loading point found in the columns with
2.2. Pressure Drop Measuring Methods. Great attention larger diameters (0.43 and 1.22 m) at an elevated pressure of
has been given to the avoidance of all the possible causes of 1.65 b is consistent with the value at which the dependence
incorrect pressure drop measurements that have been curve of pressure drops starts to bend when measured in the
thoroughly discussed by Cai and Resetarits.10 In the distillation small Column 4 of 0.15 m diameter at atmospheric pressure.
column, the pressure tap consists of small holes with an inner The measuring of higher Fv-factors in Column 4 was restricted
diameter of 6 mm that are drilled into the column wall below by the maximum output of the reflux pump. Therefore it is not
and above the packing and are connected to a pressure possible to estimate a reliable value of the loading point in this
transducer via independent tubes. To ensure that no vapor column.
condensation occurs inside the lines, nitrogen was blown The HETP measured in the different columns differ
through both the lines connected to the pressure drop significantly, see Figure 2. The data measured in the same
measurement device at extremely low flow rates eliminating column with the same packing also differs. This is caused by a
any pressure drop in the lines. HETP dependence on the concentration range in which the
2.3. HETP Measuring Methods. The HETP were measurement was made due to the significant variation of the
measured under total reflux and evaluated on the basis of a relative volatility over the concentration range. In Figure 2, the
number of theoretical stages Ntheor between the mole fractions concentration range is characterized in accordance with the
of the light component in the liquid phase at the top and the bottom molar fraction of C6 and thereby the data measured at
bottom of the packed bed section H. Rigorous (stage-to-stage) the same xbottom may still differ in accordance with whether they
calculation was used for this purpose. The equilibrium data of are determined in a short or a long column. The HETP
5969 dx.doi.org/10.1021/ie302397q | Ind. Eng. Chem. Res. 2013, 52, 5967−5974
Industrial & Engineering Chemistry Research Article

measured in lean cyclohexane mixtures are much lower than the bottom concentration from 0.20 to 0.75. The HETP
those measured in rich mixtures. In the short (H = 1.8 m) and dependence on xbottom ceases with concentrations higher than
the small diameter (d = 0.15 m) Column 4, the HETP increases 0.5 as is shown by the data of the Montz packing. The HETP
sharply from 0.28 to 0.38 m with the bottom C6-concentration measured in these rich C6 mixtures fluctuates at around 0.46 m
from 0.03 to 0.10. The HETP dependence on x bottom in the whole range of the Fv-factor utilized (from 0.5 to 2.1
diminishes/ceases at concentrations higher than 0.32 in Pa0.5).
which the HETP values fluctuate at ca. 0.41 m in the whole The effect of the packing height on the HETP can be
range of the Fv-factor utilized (from 0.7 to 2.5 Pa0.5). assessed from the data provided by the Montz BSH 250
The data measured in FRI Column 1 with H = 3.5 m and d = packing, measured at the same bottom concentration of 0.20−
1.22 m at low bottom concentrations of 0.10 to 0.12 lie in the 0.26 using different packing heights, specifically 4 and 1.8 m.
0.32−0.39 m interval and are somewhat higher (by 0.03 m The HETP measured at the low packing height is ca. 0.34 m
approximately) than those measured in SRP Column 3 with a d and is much lower than the HETP = 0.42 m measured at the
= 0.43 m diameter. This small difference of the HETP may be higher packing height. Fitz et al.12 measured a somewhat lower
the result of the lower packing height H = 3 m used in Column HETP = 0.39 m for the Mellapak 250Y at the lower packing
3 in comparison with the 3.5 m used in FRI Column 1. height of 3.5 m.
3.1.2. Mellapak 250Y and Montz BSH 250. Both these 3.1.3. Concluding Remarks Concerning the Column
packings are of the same size and wettability of their surface. Diameter Effect on Pressure Drops. The pressure drops
The pressure drop and the HETP measured in columns with presented here in regard to the RSP 250, Mellapak 250Y, and
diameters of 0.15 and 1.22 m at the pressures of 0.35, 1.0, and Montz BSH 250 packing and the C6/C7 distillation system
1.65 bar are compared in Figure 3. Similarly as with RSP measured at flow rates below the loading point in columns of
different diameters (0.15−1.22 m) at total pressures (0.31−
1.65 b) show only small differences, far less than 15%, which
indicate zero or an insignificant column diameter effect.
3.1.4. Concluding Remarks Concerning the Column
Diameter Effect on the HETP. The effect of the column
diameter on the HETP is difficult to evaluate from experiments
carried out in columns with different packing heights and
distillation mixtures with different concentrations, because the
impact of both these conditions is more pronounced than is the
potential effect of the column diameter. Nevertheless, the
HETP measured in columns of widely different diameters,
using distillation mixtures of approximately the same bottom
concentration, did not show any significant column diameter
effect that would support the recommendation1 to discount the
data measured in the columns with diameters smaller than 0.4
m. For example, the HETP of the same Mellapak 250Y and
Montz BSH 250 corrugated packings measured in Columns 4
and 2 of very different 1.22 and 0.15 m diameters differ only
slightly (0.36 and 0.34 m, respectively) provided that the same
low packing height 1.80 m and the bottom C6-concentration of
Figure 3. Pressure drop and HETP of geometrically similar packings 0.2 are used, as is shown in the data in Figure 3. A similar
measured using the C6/C7 distillation system. concordance exists between the data measured in Columns
with xbottom higher than 0.5, in which the HETP dependence on
xbottom has ceased (see Figure 4).
packing, the measured pressure drops below the loading points Thanks to lower dependence of the HETP on the
in the columns of the different diameters and at different concentration range, as measured by the CB/EB distillation
pressures differ only in the scatter of the experimental points. system, in comparison with the C6/C7 system (due to the
Also the 2.4 Pa0.5 loading point identified in the column of
greater diameter (1.22 m) is consistent with the value at which
the pressure drop in the small column of 0.15 m diameter also
starts to increase. The pressure drops measured below the
loading points in these packings do not confirm the significant
effect of the column diameter reported by Olujic.3
The HETP values measured on short packing depth (1.80 or
1.83 m) in columns of different diameters packed with
Mellapak 250Y (d = 0.15 m) or Montz BSH 250 (d = 1.22
m) increase with bottom concentration in a similar manner as
was revealed for the RSP in the short Column 4 (H = 1.8 m
and d = 0.15 m). The HETP of the Mellapak increases from 0.3
to 0.43 m with an increase of the bottom C6-concentration
from 0.03 to 0.68. Similarly, the HETP of the Montz evaluated Figure 4. HETP evaluated from short column sections heights of 1.9
for the short packing sections, in which ΔH = 1.83 m in ± 0.1 m plotted against C6-concentration at the bottom of the
Column 2, increases from 0.33 to 0.47 m with an increase of sections.

5970 dx.doi.org/10.1021/ie302397q | Ind. Eng. Chem. Res. 2013, 52, 5967−5974


Industrial & Engineering Chemistry Research Article

lower variation of the relative volatility over the concentration values of the xmean and the Ntheor,mean presented in the figure
range), the results produced by the CB/EB system are more represent the arithmetic mean of the C6-concentrations and of
definitive and therefore more convincing. For example, the the number of theoretical stages, or are measured at these
independence of the HETP of Mellapak 250Y from the column points in all the experiments undertaken at the various Fv
diameter, using this system, was identified by Billet7 and Meier factors. The data show a steady increase in the HETP from 0.24
et al.8 The HETP measured by Billet7 in the 0.22 and 1 m to 0.42 m with an increase in the packing depth from 0.76 to
diameter columns did not differ and was equal to 0.32 ± 0.01 m 4.01 m, and they have a tendency to increase further.
in the 0.4 to 2.8 Pa0.5 range of the vapor capacity factor. Meier If the HETPs were evaluated using the concentrations
et al.8 found a similar concordance between the HETP measured at the first sampling point located 0.76 m above the
measured in columns with diameters of 0.16, 0.25, and 1 m packing bottom (i.e., not from the sampling point situated
at various pressures between 50 and 100 mbar and between 400 below the packing), the resulting dependence of the HETP on
mbar and the value of the atmospheric pressure. This data did the packing depth would have changed considerably as is
not depend on the packing heights that were used, that is, 1.4, shown in the upper part of Figure 6. Instead of a steady
4.4, and 6.6 m.
In contrast to the insignificant and rather incomprehensible
effect of the column diameter, the effect of the concentration
range in which the measurement was conducted is considerable.
The HETP increases considerably with the increasing bottom
concentration and the packing height, and this increase reaches
up to 50% of the HETP value measured. For this reason, the
effect of the concentration range is analyzed in more detail in
the next section.
3.2. Analysis of the HETP Dependence on the
Concentration Range Used. 3.2.1. The HETP Dependence
on Bottom Concentration. The HETP evaluated for the short
packing sections ΔH (Column 2, ΔH = 1.83−1.98 m; Column
4, ΔH = 1.80 m) is plotted in Figure 4 against the C6-
concentration at the bottom of the sections. In the small Figure 6. HETP calculated from concentration profiles plotted as a
Column 4 (d = 0.15 m), the variation of xbottom was achieved function of packing height taken from the first sampling points from
utilizing different concentrations of the distillation mixture in the bottom (z = 3.25 m) and the top (z = 0.81 m) of the packing.
the reboiler. In the large Column 2 (d = 1.22 m), the Symbols are the same as in Figure 5.
concentration profile was measured along the column and the
HETP was evaluated between sample points remote from each increase, the HETPs are independent of the packing height.
other by 1.98, 1.83, and 1.83 m. The HETP increases linearly Similar behavior is shown by the HETP calculated at the first
from 0.28 to 0.48 m with the increase of the xbottom from 0.02 to sampling point situated 0.81 m below the top of the packing;
0.75. The wide scatter of the data is caused by the inclusion of see the data plotted in the lower part of Figure 6. Only the
all the HETPs measured at various values of the Fv capacity HETPs calculated from the concentrations measured just below
factor. the packing are significantly lower. It seems that these
3.2.2. The HETP Dependence on the Packing Height. The concentrations are significantly undervalued. If these data are
HETPs calculated from the concentration profiles measured in excluded from the evaluation, the HETP values fluctuate
Column 2 are plotted in Figure 5 as a function of the packing around a constant value. Similar behavior is shown by the
HETP data presented in Figure 3 for Mellapak 250 Y as
measured at a similarly high bottom concentration of 0.6.
Apparently, the concentrations in the liquid that falls out of the
packing may differ from those in the liquid at the bottom of the
packing and their difference may create a significant part of an
equilibrium stage. The evidence that these data are mistaken or
are influenced by the end-effects is also supported by our
further finding that the profiled method was not convergent,
when these concentrations were taken into account for the
evaluation of the concentration profiles.
The end-effects may cause an erroneous appearance of an
improvement in the performance of the packing. Therefore it is
necessary to discount both the top and the bottom samples
Figure 5. HETP calculated for the sections ΔH of the packing taken from the packing to obtain the correct HETP values reflecting
from the bottom of the packing.
only the improvements in the performance of the packing.
3.2.3. The Suitable Packing Height for the Measurement
height. The heights were measured from the bottom of the of the Packing Efficiency. Ottenbacher et al.1 recommend
packing to sampling points located at the heights of H = 0.76, performing the measurement of the packing efficiency in a
1.36, 1.98, 2.59, 3.20, and 4.01 m. The bottom concentrations “large column comprising 15 to 20 theoretical stages (which
were measured at a sampling point just below the packing, the industrial applications usually do contain); this is equivalent to
head concentrations are those measured at the reflux and the a bed height of 4−6 m”. A packing height of 4 m represents
remaining samples were withdrawn from the packing. The approximately 10 theoretical stages of the C6/C7 distillation
5971 dx.doi.org/10.1021/ie302397q | Ind. Eng. Chem. Res. 2013, 52, 5967−5974
Industrial & Engineering Chemistry Research Article

system, and the further increase in the packing height system with the lowest concentration dependence of slope of
dramatically augments the demands on the accuracy of the equilibrium line m, for example, using the CB/EB distillation
analytical method. For example, starting with a bottom C6- system rather than the C6/C7 system as recommended by
concentration of 0.2 (the same concentration that was used in Ottenbacher et al.,1 the second issue is hardly controllable. As
the testing of the Montz BSH 250 in Column 2), the shown in the relation 4 deduced from the basic principles of the
concentration at the top of the 4 m packing, representing 10 interfacial mass transfer
stages, equals 0.985. In the next two stages, the C6-
⎛ 1 m ⎞ ln(m(uL /uV ))
concentrations would be 0.991 and 0.995. At such high HETP = uL⎜ + ⎟
concentrations, the experimental error of the C6-molar fraction ⎝ kLa k Va ⎠ m(uL /uV ) − 1 (4)
0.003 corresponds to one theoretical stage. In this case, the
application of a packing height of more than 10 stages might the HETP depends on the relative resistances concentrated in
lead to an unacceptable error of the HETP (>10%). A suitable the individual phases. The local slope of equilibrium line m may
packing height should be chosen in accordance with the affect considerably the role of resistances of individual phases.
distillation system utilized and with the accuracy of the As a result, the HETPs measured by various distillation systems
analytical method used. are variously sensitive/receptive to identifying an improvement
3.2.4. Comparison of the Shortcut and the Rigorous of the mass transfer rate developed in one of the phases. For
Evaluation Methods. The shortcut methods currently used for example, the improvement of the mass transfer coefficient
the evaluation of the HETP are based on the Fenske equation achieved solely in the liquid phase is not detectable as an HETP
deduced for total reflux, a constant relative volatility α, and a value when measured using the distillation system with a
constant molar overflow as shown in the formula controlling mass transfer resistance in the vapor phase. In
addition, this situation is complicated by the fact that the

Ntheor =
(
ln
xhead 1 − x bottom
1 − xhead x bottom ); αave = αheadαbottom
relative resistance of the phases varies along the column height.
This problem might be solved/avoided by comparison not only
ln αave of the HETP values but also of the local volumetric mass
(3) transfer coefficients measured directly in the distillation
The number of theoretical stages Ntheor evaluated using the columns at the same concentration and flow rates of the
shortcut method and the rigorous stage-to-stage method for the phases. For the determination of the coefficients, the profile
C6/C7 and CB/EB distillation systems are compared in Figure method13−15 was recently developed. The method evaluates the
7, in which their relative differences, calculated for several coefficients from the concentration profiles measured along the
distillation column. Presently, the physical consistency/
relevance of the evaluated coefficients is verified by comparing
absorption and distillation mass transfer coefficients measured
in a wetted-wall column.
3.4. Effect of the Concentration Range on the HETP
Deduced from the Concentration Profiles. The concen-
tration range in which the measurement was carried out is
determined by the two following process parameters: the xbottom
bottom concentration of the distillation mixture and the height
H of the packing used. Their effects on the HETP measured in
columns with different bottom concentrations and packing
heights are quantified below for the data for Mellapak 250Y
Figure 7. Relative difference between number of theoretical stages measured in Column 4 with the C6/C7 distillation system at
calculated between xbottom and xtop by the Fenske eq 3 and by the reboiler duty of 16 kW which provides Fv = 2.05 Pa0.5 at a
rigorous stage-to-stage method.
bottom concentration of 0.0428. The concentration profile
shown in Figure 8 was obtained by the profile method applied
values of xbottom, are plotted against xhead. The equilibrium data
were taken from Onken and Arlt.11 The differences in terms of
the CB/EB system are negligible (less than 0.03%) and, in
terms of the C6/C7 system, are less than 2%. The linear
interpolation used between zero and one equilibrium stage for
an unlikely separation corresponding to less than one stage is
really obscure and is probable the cause of the sinusoidal
highlighted difference between the Fenske equation and the
stage-to-stage calculation. The comparison shows that the
shortcut method does not lead to any noticeable differences in
the theoretical stages of either distillation system.
3.3. The Suitability of the HETP for the Objective
Evaluation of Packing Efficiency. The two properties that
preclude the HETP from providing a comprehensive
comparison between different packings are (i) dependence on
the concentration range within which the measurement was
carried out and (ii) inability to identify and to distinguish mass
transfer improvements achieved in the liquid and vapor phases. Figure 8. Concentration profile obtained by the profile method at Fv =
While the first issue can be minimized by using the distillation 2.05 Pa0.5 for various bottom concentrations of the distillation mixture.

5972 dx.doi.org/10.1021/ie302397q | Ind. Eng. Chem. Res. 2013, 52, 5967−5974


Industrial & Engineering Chemistry Research Article

onto the experimental data measured at four different bottom values, that is, 15, 13, and 10 cm, with xbottom = 0.3. (iii) The
concentrations ranging from 0.0428 to 0.7088 at the same conditions in which the effect of the concentration range and of
reboiler duty 16 kW. The method starts with the choice of the an experimental concentration error on the measured value of
mass transfer correlation describing the dependence of the the HETP is minimal are as follows: a packing height of 3 m
volumetric mass transfer coefficients on physical properties and and a bottom concentration of 0.2 C6-mol fraction.
phases velocities. Then, the front factors of the correlations are
varied in a systematic way such that the predictions from the 4. CONCLUSIONS
rate based simulation model are matched to experimental The pressure drops of the RSP 250, Mellapak 250Y, and Montz
concentration profiles as closely as possible. The plug flow of BSH 250 packings, measured in columns of different diameters
both phases in the column is considered and the film model of (0.15, 0.43, and 1.22 m), using the cyclohexane/n-heptane
the interfacial mass transfer is used in which the convective flow distillation system below the loading points do not show any
is taken into account. The profile method is described in detail differences that would indicate any significant column diameter
by Linek et al.,13 Rejl et al.,9,14 and Valenz et al.15 For the sake effect. In contrast to the insignificant effect of the column
of completeness, we have used two different mass transfer diameter, the effect of the concentration range in which the
correlations, namely the RBF model published by Rocha et measurement of the HETP was carried out is considerable. The
al.16,17 and the Delft model published by Olujic et al.,18 to show HETP increases with an increase of the bottom concentration
the effect of the choice of the mass transfer correlations on the and the packing height, and this increase can reach up to 50%.
results. The profiles obtained using the RBF and Delft model The HETP of the different packings can be compared only if
practically coincide, and their difference is not detectable in the concentration range along the packings to be compared is
Figure 8 in which the calculated profiles are plotted together the same. This requirement is hard to fulfill but the HETP
with the experimental data. The mean relative deviation measured in approximately the same concentration range did
between the calculated and the experimental concentrations is not show any significant column diameter effect.
3.9% for x-profiles and 2.4% for y-profiles. The results presented do not support the recommendation of
Using this profile, the dependence of the HETP on the Ottenbacher et al.1 to use columns with a diameter of at least
bottom concentration and the packing height can be evaluated 0.4 m for obtaining significant efficiency values, since the use of
from eq 2 for H = z2 − z1 and the corresponding theoretical smaller column diameters will result in measurements that are
stages Ntheor are defined between the mole fractions xC6(z1) and strongly influenced by wall effects.


xC6(z2) using rigorous stage-to-stage calculation. The depend-
ence of the HETP on xbottom calculated for three packing AUTHOR INFORMATION
heights H = 1.8, 3, and 4 m are presented in Figure 9. The
Corresponding Author
*E-mail: linekv@vscht.cz.
Notes
The authors declare no competing financial interest.

■ ACKNOWLEDGMENTS
The supports from RASCHIG GmbH and Grant Agency of
Czech Republic through the project no. 13-01251S are
gratefully acknowledged.

■ NOMENCLATURE
a = effective mass transfer area, 1/m
Figure 9. Dependences of HETP onto the bottom concentration
ap = packing geometrical area, 1/m
evaluated for three packing heights.
B = superficial liquid velocity, m/h
c = molar concentration of more volatile component in the
dependences are terminated when the value of xhead reaches bulk, kmol/m3
0.985 that is, the value above which an experimental error of de = equivalent diameter, m
the C6-molar fraction might lead to an unacceptable error in d = column diameter, m
the HETP as discussed in section 3.2.3. The line calculated for FV = uV √ρV gas capacity factor, Pa1/2
the height of H = 1.8 m corresponds well with the experimental H = packing height, m
data presented in Figure 3 for Fv = 2 ± 0.1 Pa0.5. Hp = height of packing element, m
The results presented in Figure 9 illustrate the following ΔH = height of packing section, m
findings: (i) At the same bottom concentration of the HETP = height equivalent to a theoretical plate calculated
distillation mixture, the HETPs measured in the lower columns from eq 2, m
are higher than those measured in the higher columns. For K = wall factor given by eq 1, -
example, the HETP values measured at xbottom = 0.1 in the kLa = liquid-side volumetric mass transfer coefficient, 1/s
columns with packing heights of 1.8, 3, and 4 m increase to kVa = vapor-side volumetric mass transfer coefficient, 1/s
0.32, 0.35, and 0.37 m. (ii) The sensitivity of the HETP values m = slope of the equilibrium line, -
to changes of the bottom concentration is greater at lower Ntheor = number of theoretical stages
concentrations and at lower packing heights. This sensitivity is p = pressure, Pa
defined by the local slope of the curves shown in Figure 9. The Δp/H = pressure drop, Pa
slopes equal to 33, 31, and 25 cm for the packing heights of 1.8, u = superficial velocity, m/s
3, and 4 m with xbottom = 0.1 and are approximately half of these x = liquid phase molar fraction of more volatile component
5973 dx.doi.org/10.1021/ie302397q | Ind. Eng. Chem. Res. 2013, 52, 5967−5974
Industrial & Engineering Chemistry Research Article

z = column height coordinate taken from the top of the (18) (5) Olujič, Ž .; Kamerbeek, A. B.; de Graauw, J. A. Corrugation
packing (z = 0), m geometry based model for efficiency of structured distillation packing.
Chem. Eng. Process. 1999, 38, 683.
Greek Symbols
α = relative volatility, -
ε = packing porosity, -
λ = resistance coefficient, -
θ = corrugation angle, deg
Indexes
bottom = bottom of packing
C6 = cyclohexane
head = top of packing
L = liquid phase
V = vapor phase

■ REFERENCES
(1) Ottenbacher, M.; Olujic, Z.; Adrian, T.; Jödecke, M.; Großmann,
C. Structured packing efficiencyVital information for the chemical
industry. Chem. Eng. Res. Des. 2011, 8, 1427.
(2) Billet, R.; Mackowiak, J. How to use the absorption data for
design and scale-up of packed columns. Fette Seifen Anstrichmittel
1984, 86 (9), 349.
(3) Olujic, Z. Effect of column diameter on pressure drop of
corrugated sheet structured packing. Chem. Eng. Res. Des. 1999, 77,
505.
(4) Stichlmair, J.; Bravo, J. L.; Fair, J. R. General model for prediction
of pressure drop and capacity of countercurrent gas/liquid packed
columns. Gas. Sep. Purif. 1989, 3, 19.
(5) Mackowiak, J. Pressure drop in irrigated packed columns. Chem.
Eng. Process 1991, 29, 93.
(6) Brunazzi, E.; Paglianti, A. Mechanistic pressure drop model for
columns containing structured packings. AIChE J. 1997, 43, 317.
(7) Billet, R. Packed Towers; VCH Verlagsgesellschaft mbH, D-
69451: Weinheim,New York, Basel, Cambridge, Tokyo, 1995.
(8) Meier, W.; Hunkeler, R.; Stocker, W. D. Sulzer Mellapak, Eine
Neue, Geordnete Packung fur Stoffaustausch-Apparate. Chem. Ing.
Technol. 1979, 51, 119.
(9) Rejl, F. J.; Valenz, L.; Linek, V. “Profile Method” for the
measurement of kLa and kVa in distillation columns. Validation of rate-
based distillation models using concentration profiles measured along
the column. Ind. Eng. Chem. Res. 2010, 49, 4383.
(10) Cai, T. J.; Resetarits, M. R. Pressure drop measurements on
distillation columns. Chin. J. Chem. Eng. 2011, 19 (5), 779.
(11) Onken, U.; Arlt, W. Recommended Test Mixtures for Distillation
Columns; The Institution of Chemical Engineers: Rugby, England,
1990.
(12) Fitz, C. W.; Kunesh, J. G.; Shariat, A. Performance of structured
packing in a commercial-scale column at pressures of 0.02−27.6 bar.
Ind. Eng. Chem. Res. 1999, 38, 512.
(13) Linek, V.; Moucha, T.; Rejl, F. J.; Prokopová, E. Simultaneous
determination of vapor- and liquid-side volumetric mass transfer
coefficients in distillation column. Chem. Eng. Res. Des. 2005, 83, 979.
(14) Rejl, F. J.; Linek, V.; Moucha, T.; Prokopová, E.; Valenz, L.;
Hovorka, F. Vapor- and liquid-side volumetric mass transfer
coefficients measured in distillation column. Chem. Eng. Sci. 2006,
61, 6096.
(15) Valenz, L.; Rejl, F. J.; Linek, V. Effect of gas and liquid axial
mixing on the rate of mass transfer in a distillation column packed with
Mellapak 250Y. Ind. Eng. Chem. Res. 2011, 50, 2262.
(16) Rocha, J. A.; Bravo, J. L.; Fair, J. R. Distillation columns
containing structured packings: A comprehensive model for their
performance. 1. Hydraulic models. Ind. Eng. Chem. Res. 1993, 32, 641.
(17) Rocha, J. A.; Bravo, J. L.; Fair, J. R. Distillation columns
containing structured packings: A comprehensive model for their
performance. 2. Mass-transfer model. Ind. Eng. Chem. Res. 1996, 35,
1660.

5974 dx.doi.org/10.1021/ie302397q | Ind. Eng. Chem. Res. 2013, 52, 5967−5974

You might also like