Mars 1417V3 Installation Guide
Page 1/40

Mars1417V3 Installation Guide

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 2/40

Content
Content ................................................................................................................................ 2
1. Production Description ................................................................................................ 3
2. Installation Requirement ............................................................................................. 3
2.1 Environment ................................................................................................................... 3
2.2 Workstation configuration requirements ........................................................................ 3
3. Package Check .............................................................................................................. 4
3.1 Package instruction ........................................................................................................ 4
3.2 Product list...................................................................................................................... 4
3.3 Optional product component .......................................................................................... 6
3.4 Component description .................................................................................................. 7
3.4.1 Detector ............................................................................................................... 7
3.4.2 Battery ................................................................................................................. 8
3.4.3 Battery Charger ................................................................................................... 9
4. Installation .................................................................................................................. 10
4.1 Panel Installation .......................................................................................................... 10
4.1.1 Attach Battery Pack ........................................................................................... 10
4.1.2 Attach DC Power ................................................................................................11
4.1.3 DC power interface ........................................................................................... 12
4.2 Battery Charger Installation ......................................................................................... 12
4.2.1 Battery activation .............................................................................................. 13
4.3 System requirement ...................................................................................................... 14
4.3.1 Environment setup............................................................................................. 15
4.4 Connection Mode ......................................................................................................... 15
4.4.1 Wireless Client Mode ........................................................................................ 15
4.4.2 Wireless AP Mode ............................................................................................. 21
4.5. Trigger mode switching ............................................................................................... 26
5. Correction and Calibration Template Generation .................................................. 27
5.1 Pre-offset Template Generation .................................................................................... 27
5.2 Gain Calibration Template Generation......................................................................... 28
5.3 Defect Correction Template Generation ....................................................................... 31
5.4 Image Check and upload .............................................................................................. 33

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 3/40

5.4.1 Local Image Check............................................................................................ 33


5.4.2 Panel Image Upload .......................................................................................... 33
5.5 Defect Template Check and Modification .................................................................... 34
5.5.1 Defect Template Check ..................................................................................... 35
5.5.2 Defect Template Modification........................................................................... 36
6. Correction and Calibration Management ................................................................ 37
6.1 Correction and Calibration template synchronization .................................................. 37
6.2 Correction and Calibration management...................................................................... 39

1. Production Description

The Mars1417V is a wireless X-ray flat panel detector based on amorphous silicon thin
film transistor technology. It provides high quality radiographic images with 2304 x 2800
active pixel matrices and a pixel size of 150um. The Mars1417V's biggest improvement is
the ability to support wireless communication between the detector and the workstation, as
well as battery-powered, making it truly portable.

2. Installation Requirement

2.1 Environment
This product needs to be used within the specified temperature and humidity range:
Operating temperature Operating humidity Operating pressure range
10℃~30℃ 10%~75% 700~1060hPa
If the product is not used for a long time, please re-seal the product in an anti-static bag
and store it in the specified storage environment.
Storage temperature Storage humidity Storage pressure range
-20℃~55℃ 5%~93% RH 700~1060hPa
Additional requirements for the installation environment can be found in the product
instruction manual.
2.2 Workstation configuration requirements
Name Recommend

System Windows 7/10 64bit

CPU Intel Core i7 3.6G

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 4/40

Memory 4G or more
Storage 640G or more
network adapter Intel Pro EXP9301CT PRO

3. Package Check

Before unpacking, please confirm whether the current environment meets the
requirements in 2.1.
3.1 Package instruction
The product box contains two separate boxes.

Detector box(Yellow)

Two Accessory boxes(Yellow)

3.2 Product list


Items Description

Mars1417V3
1 main unit
Detector

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 5/40

1 pcs
Medical Adapter
DC:24V

Battery 2 pcs

1pcs
Ethernet cable
3.5 m

1pcs
Gigabit cable
3m

AC power cable 1pcs

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 6/40

DC power 1pcs
cable 3.5 m

Charger 1pcs

Gain template
Defect template
CD Software
development kit
User manual

3.3 Optional product component


Name Description

Wireless Router 1 pcs

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 7/40

3.4 Component description


3.4.1 Detector

AB CDE F

Item Name Description


A DC Jack 24V DC input
B Network port Gigabit network interface
C Detector Indicator Detector indicator of control panel
D Power Button Power button of control panel

E Reserved
F Antenna Antenna

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 8/40

3.4.2 Battery

Item Name Description


A Battery Label /
B Battery Interface 8 Pin Battery connector
C Pilot Pin /
D Indicator Installation direction indicator

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 9/40

3.4.3 Battery Charger

C
E G

Item Name Description


A Battery Interface 8 Pin Battery connector
B Battery Interface 5 Pin Battery connector
C Battery Interface 5 Pin Battery connector
C Power Indicator The indicator definition is as follow
D Anti-shedding lock /
E The limit ball plug /
F AC power interface 220V AC input
The battery charging capacity indicator definition:
X Group Indicator Lighting Status Operating Status

Complete extinction No external power input

A light is always on External power input

A light is always on

B and C lights flash Battery insertion self test

alternately 2 times

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 10/40

A and B lights are


Charging batteries
always on

A and C lights are


The battery is full, stop charging
always on

A light is always on

B and C lights Battery charging is abnormal

alternately flash

Note: It is forbidden to insert two or more batteries at the same time, otherwise the charger
will stop working automatically.

4. Installation

4.1 Panel Installation


4.1.1 Attach Battery Pack
The Mars1417V can be powered by DC power or battery. When the battery is plugged in
or the DC power is plugged in, the detector will start immediately. If neither the battery
nor the DC power is plugged in, the Mars1417V will shut down. See the following for
battery installation of the detector.
 Make sure that the connectors on the
battery package are pointed to the cave in
battery compartment.

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 11/40

 Slide battery package into battery

compartment(Make sure battery capacity

overpass 10%).

 Slide the battery lock lever.

4.1.2 Attach DC Power


Please see below for DC power installation.
 Connect one end of DC Power Cable to the
Medical Adapter

 As figure is power interface Ethernet


interface

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 12/40

 Connect another end of DC Power Cable to


the DC input of the detector.

4.1.3 DC power interface


The Mars1717V3 has two DC power connectors, as shown in the following figure:
 3PIN DC power
interface

 4PIN DC power
interface

4.2 Battery Charger Installation


Operation Figure

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 13/40

 Unload Battery from battery charger.

 Insert battery into battery charger.


 Note the interface position as figure.

 Press the battery to the bottom of battery


compartment.

4.2.1 Battery activation


To ensure the safety of the battery during transportation or storage, the battery can be set
to a transport mode in which the battery is locked without voltage output. Therefore, you
need to exit the transport mode before using the battery for the first time.
 Two ways to enter the transport mode

Entry mode Step

Factory /

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 14/40

configuration

 Power the detector with an adapter, the detector is turned on, and

connected to the PC;

Web write  Insert the battery into the battery compartment;

 Type “http://192.168.8.8/cgi-bin/shipmode.cgi?action=on” into the

web to put the battery into transport mode.

 Three ways to exit the transport mode

Unlock mode Step

Web unlock  Power the detector with an adapter, the detector is turned on, and

connected to the PC;

 Insert the battery into the battery compartment;

 Type “http://192.168.8.8/cgi-bin/shipmode.cgi?action=off” on the

web to bring the battery out of transport mode.

Adapter +  Power the detector with an adapter and turn the detector on;

FPD  Insert the battery into the battery compartment;

 Press the power button for about 4 seconds to turn off the detector;

 In the off state, press the power button 3 times to activate the battery.

battery  Power on the charger;

charger  Insert the battery into the charger and complete the battery activation

in about 3~5 seconds.

4.3 System requirement


iDetector is developed and deployed on Windows Operation System, it can be run on
Windows XP/Windows 7/Windows 8/Windows 10, OS should install latest service pack.
And requires computer memory 4 GB minimum. Firewall should be shut down to avoid

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDCare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 15/40

commuication issue.
4.3.1 Environment setup
Setup files and download url are included in Software Development Kit(hereafter refers as
SDK) directory: Tools\env_setup.
1. Please install Microsoft .NET Framework 4.5(Windows XP only can install V4.0 ).
Download from Microsoft web site, please.
2. Visual C++ redistributed package need to be installed: vcredist_x86_2013(or
vcredist_x64_vs2013).
3. For Windows XP, full path should be used in file “bind.txt”.

4.4 Connection Mode


Mars1417V supports two connection modes as follows, the IP address and other
information mentioned below is as the example, user should configure the connection with
the specific requirement.
1) Wireless Client Mode

AP

2) Wireless AP Mode
AP

To establish a connection between the workstation and the tablet, the user should follow
the steps below.
4.4.1 Wireless Client Mode
To complete Wireless Client mode configuration, user has to finish actions listed below.
Configuration of External wireless AP

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDCare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDCare.
Doc ID 085-500-05
JUDCare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 16/40

Connect one end of Gigabit /


Ethernet Cable to Workstation,
Connect another end to LAN
port of External wireless AP
Open local network
management interface

Open local network


configuration

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDCare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDCare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 17/40

open IPV4 setting

IP and network mask setting Select “Obtain an IP address automatically”


Open browser and type
192.168.1.1
Log into external wireless AP

Wireless setup

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 18/40

Configure 2.4GHz wireless SSID: NETGEAR_BIG_24


network Security: WPA2-PSK
Password: 12345678
Channel: [Please check the current Wi-Fi environment, and
choose a relatively clean channel]
Configure 5GHz wireless SSID: NETGEAR_BIG_50
network Security: WPA2-PSK
Password: 12345678
Channel: [Please check the current Wi-Fi environment, and
choose a relatively clean channel]
LAN setup

Configure LAN IP address IP address: 192.168.8.1


Subnet Mask: 255.255.255.0
External wireless AP reboot Apply above settings and reboot your wireless router.

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDCare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDCare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 19/40

Recover local network IPv4


setting

IP setting IP address: 192.168.8.188


Network mask setting Subnet mask: 255.255.255.0

Configuration of detector

Either Wired Cable can be used to configure detector in wireless client mode. The wired connection should be used
by the service operator only.

To start configuration with wired cable. It is necessary to finish the configuration of external wireless AP, then
proceed to the steps below.

Connect panel to Workstation with


Ethernet Cable like 3.4.1

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDCare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDCare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 20/40

Click “Detector”

Click ”Wifi”

Click ”Read Config”

Choose “Client” mode

Click “Add”

Type SSID and Password

Click “Apply”

Choose SSID and select(There will be


√ occurred)”

Click “write config” to save


parameters.

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDCare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDCare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 21/40

Turn on wireless router.

Make sure there are wired connection


between router and work station and IP
192.168.8.188.

Click “Read wifi Status” to check


wireless transmission status, numericl
value occurred means the link is up
and available.

Since we have chosen default SSID and password, it would connect to wireless AP
immediately after powered on next time.

4.4.2 Wireless AP Mode


To complete wired connection configuration, user has to finish actions listed below.
Configuration of detector
Either Wired cable can be used to configure panel wireless AP mode. The wired
connection should be used by the service operator only.
To start wired cable configuration, users should finish the configuration of external
wireless AP, then proceed to the steps below.
Connect panel to Workstation
with Ethernet Cable like 3.4.1

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 22/40

Click “Detector”
Select “wifi”
Choose AP mode

Click “Read Config” to get


default setting. Change SSID and
password setting , make sure
SSID is different from other
already exist;

Change channels and


frequency setting

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 23/40

Click “Country” and choose a


country setting

Click ”Channel” and choose a


clean frequency and channel

Click ”write config”

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 24/40

Do not remove wired cable until


FPD status from Busy become
Ready

Since we have chosen default SSID and password, it would connect to wireless AP
immediately after powered on next time.
Configuration of external wireless card
Open local wireless signal list

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 25/40

Select SSID which belongs to


detectors;
Input password and log into
system

Open wireless card configuration

open IPV4 setting

IP setting IP address: 192.168.8.188

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 26/40

Network mask setting Subnet mask: 255.255.255.0


Open SDK and choose product
start connection

4.5. Trigger mode switching


The Mars1417V is mainly used to acquire X-ray images. The acquisition of the detector
should be synchronized with the X-ray generator. The Mars1417V itself has three ways to
get X-ray images: software mode, inner2 mode and FreeSync mode.
The customer can switch the trigger mode according to the requirements. When switching
the mode, the corresponding parameters must be modified to ensure that the tablet can
work normally. The parameters that need to be modified corresponding to different trigger
modes are as follows:
Trigger Mode Soft Inner2 FreeSync

Pre CapMode Clear Acq Cycle Acq Clear Acq

Self CapEnable off off off

Self Clear Enable on on on


For example:
If the customer needs to switch the trigger mode to FreeSync, please follow the steps
below (see the figure below):
1. Connect the flat panel detector according to the flat panel connection mode described in
section 4.4, and go to the Acquire interface to remove all calibration modes.
2. Enter the Detector interface of the IDetector software and select the Parameters
interface under the Detector interface to modify the related parameters.
3. Modify the relevant parameters according to the above table;

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 27/40

4. Click the "write" button on the right to save the modified parameters.

5. Correction and Calibration Template Generation

After the installation is complete or the DR system is re-adjusted, it is recommended that


the user perform a calibration template generation. In addition, we recommend that users
update the calibration template every six months.
5.1 Pre-offset Template Generation
The detector is configured for pre-correction mode and the pre-correction template must
be generated.
Select “Calibrate”
Click “Start Generate
Templates”

Click “Create Offset”

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 28/40

Click “Start create offset


template file”

It will show ”Offset Map


Generating”

After “generating”, it
would show “Offset
MAP Genetated”

5.2 Gain Calibration Template Generation


Before Gain template generating, make sure SID1.2m, no copper is required, the GUI of
the software maybe different with the below figures.

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 29/40

On Gain template
generating page, there
are five images that
need to be got

Click “start” button

Click “PREP”, and start


exposure

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 30/40

After exposure, click


“Acquire” to get the
light image

If the value meet the


expected value, click
“Accept”, then get the
other four images.
If the value does not
meet the expected value,
please do not click the
“Accept”, and adjust the
exposure dose ,then
click the “PREP” to get
light image again
After getting five
images, click
“Generate” to generate
gain template

Notes:
1. Please use software post offset correction.
2. Make sure your X-ray dose is correct. If your dose is out of range, iDetector will remind

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 31/40

you to adjust the dose. Then you can click to adjust the dose and shoot again.

5.3 Defect Correction Template Generation


Before Defect template generating, make sure SID1.2m, no copper is required, the GUI of
software maybe different with the below figures.
On the “Defect
Calibration” page, start
exposure, there are 8
images need to be
captured.

Click “Start” button

Click “PREP”, and start


exposure

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 32/40

After exposure, click


“Acquire” to get the
light image

If the value meet the


expected value, click
“Accept”, then get the
other 7 images.
If the value does not
meet the expected value,
please do not click the
“Accept”, and adjust the
exposure dose ,then
click the “PREP” to get
light image again
After getting 8 images,
click “Generate” to
generate gain template

Note: 1 please use software post offset mode.


2. Make sure your x ray dose is right, if your dose is out of the range, iDetector will
remind you to adjust the dose. Then you can click “start creating” and try again.
3. If users operate with two panels, SDK has a probability of quit automatically.

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 33/40

5.4 Image Check and upload


“OPEN” provides two features for image check and uploading. Local Image Check, Panel
Image Upload. Local Image Check defines function to check image saved in Workstation.
Panel Image Upload defines function to upload images stored in panel.

5.4.1 Local Image Check


Click “Local File” button
in “Local File” UI, choose
the specified file

Choose images stored in /


Workstation, images
would be shown on screen

5.4.2 Panel Image Upload


Make sure firewall is closed

Panel Image is uploaded as following.

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 34/40

Click “Image” page in “Detector”


UI

Click “Query Images” button and


the list will be showed which are
stored in the detector

Click “Upload Images”: choose


the specified image, and click
“OK”, after the state changed to
“success” which means that the
image has been uploaded

User can cancel the process of


uploading if click “Stop Upload”

The uploaded images are saved


under the path of the detector SN

5.5 Defect Template Check and Modification


iDetector provides function to check defect template. If defect template has updates, user
could add and delete defect pixel or defect lines by modifying defect template opened.

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 35/40

5.5.1 Defect Template Check


Click “LoadFile” on
“LocalFile” page

Choose the specified


defect template, and click
“Open”

The defect template will


showed on the UI

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 36/40

5.5.2 Defect Template Modification


Open the specified defect
template

The defect management


dialog box will be
showed

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 37/40

Find the pixel that needs


to be managed, type the
coordinate of the pixel
and click “Add”, the
information will be added
to the template
If click “Delete”, the
information will be
deleted
Click “Save”

It is similar to manage the


defect pixel, If user need
to add the defect line,
type the coordinate of the
line and click “Add”
If the information needs
to be deleted, click
“Delete”

6. Correction and Calibration Management

6.1 Correction and Calibration template synchronization


Panel supports correction and calibration template storage. So template in panel could be
uploaded to Workstation, and template in Workstation could also be downloaded to panel.

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 38/40

After generating the

offset ,gain and defect

templates, select the

templates and click

“Download to FPD”

Click “DownLoad”

Click “Read Status”

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 39/40

If the activity shows

disenable, please select

the one and click “Active”

to enable it

6.2 Correction and Calibration management


Panel supports two ways to do correction and calibration. Software Correction and
Calibration defines the scenario that Workstation completes all correction and calibration.
If panel complete all correction and calibration by itself, it is named as Hardware
Correction and Calibration.
User can set the calibration

method on “Acquire” page

文件包含巨鼎公司机密资料,非经巨鼎正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of JUDcare, any reproduction, disclosure, or use in whole or in part is
strictlyprohibited, except as may be specifically authorized by prior written agreement or permission of JUDcare.
Doc ID 085-500-05
JUDcare Version A0
Revision Date 2019.9.20
Mars 1417V3 Installation Guide
Page 40/40

Choose “HWPostOffset”,

“HWGain”, “HWDefect”,

the hardware-based

calibration is on

Choose “SWPostOffset”,

“SWGain”, “SWDefect”,

the software-based

calibration is on