You are on page 1of 12

Manufacturer ECRI'S Draeger Inc Draeger Inc

RECOMMENDED Infinity CentralStation Vista 120 Central


SPECIFICATIONS Monitoring System
Basic Central
Monitoring Stations
WHERE MARKETED Worldwide Worldwide, except
Canada, Norway and
USA.
FDA CLEARANCE Yes No
CE MARK (MDD) Yes Yes
MONITORED PATIENTS
Max no./display 16 on a single ≤32 on a single
display, ≤32 on a display, ≤64 on a
dual display dual display
Max no./central station ≥16 64 (32 on main 64 on dual screen
screen, 32 in
surveillance mode)
CENTRAL DISPLAY
Type Widescreen TFT Widescreen TFT
active matrix LCD, active matrix LCD
hard-coated, optional
resistive touch
technology
Single or dual Single or optional Single or dual
dual
Size, cm (in) ≥38.1 (15), larger 54.6 (21.5) 48.3 (19) or 55.9 (22)
preferred
REMOTE CENTRAL DISPLAY Patient Watch/Infinity Vista 120
Symphony

Noninteractive (slave) Yes Yes


Interactive No No
REMOTE VIEWING (outside Optional via Infinity Yes, connected to
hospital) Gateway Patient same IP address
Watch for bed view segments with the
and Infinity central station
Symphony for trends,
full and event
disclosure

USER INTERFACE Keyboard, mouse, Keyboard, mouse


optional touchscreen

PARAMETERS (NUMERICS) Required Yes Yes


Max no. main display (patient ≤15, selectable ≤4, selectable
sector)
Max no. individual patient ≤24, selectable ≤15, selectable
display
WAVEFORMS Required Yes Yes
Max no./patient sector (main 4 4
display)
Max no. (individual patient ECG: 12; all others: 8 ECG: 7; all others: 18
display)
TRENDING Yes Yes
Graphical Yes Yes
Tabular Yes Yes
Length of time, hr 72, optional ≤120 240

ARRHYTHMIA ANALYSIS Required Yes Yes


Bedside monitors Required Yes Yes
Telemetry Required Yes No
No. of leads analyzed 1 or 2, user 1
simultaneously selectable
Multi-lead view Yes Yes
ST ANALYSIS Preferred Yes Yes
Bedside monitors Preferred Optional Yes
Telemetry Preferred Optional No
No. of leads analyzed ≤12 bedside and 1
telemetry
EVENTS Required Yes Yes
Bedside monitors/telemetry Yes/yes Yes/no
Type Arrhythmia and alarm Arrhythmia and alarm
events, manual events

Storage capacity
Max no. of events 1,000/patient 20,000

Length of time, hr 120 240


Other User annotation; User annotation;
arrhythmia reports
reclassification,
calipers feature,
reports
FULL DISCLOSURE Preferred Yes Yes
Length of time, hr ≥24 ≤120 ≤96
No. of parameters stored All Determined by Determined by
source patient source patient
monitor monitor
No. of waveforms stored All ≤16 ≤18
ALARM DEFAULT SETTINGS
Systolic BP - low, mm Hg Determined by Determined by
source patient source patient
monitor monitor
Systolic BP - high, mm Hg Determined by Determined by
source patient source patient
monitor monitor
Diastolic BP - low, mm Hg Determined by Determined by
source patient source patient
monitor monitor
Diastolic BP - high, mm Hg Determined by Determined by
source patient source patient
monitor monitor
Mean BP - low, mm Hg Determined by Determined by
source patient source patient
monitor monitor
Heart rate - low, bpm Determined by Determined by
source patient source patient
monitor monitor
Heart rate - high, bpm Determined by Determined by
source patient source patient
monitor monitor
Pulse oximetry - low Determined by Determined by
source patient source patient
monitor monitor
Respiratory rate - low, bpm Determined by Determined by
source patient source patient
monitor monitor
Respiratory rate - high, bpm Determined by Determined by
source patient source patient
monitor monitor
Volume settings 10, 20, 30, 40, 50 0-100%, user
(default), 60, 70, 80, selectable
90, 100%
ALARMS Required Yes Yes
Type Audible and visual Audible and visual

Priority/levels Life-threatening 3 levels


(high), serious
(medium), advisory
(low)
User defined For arrhythmia For all
Silence active alarm Yes Yes
Silence duration, sec 120, new alarms 60, 120, 180,
breakthrough permanent
All-alarm suspend For telemetry Yes
Suspend duration, sec 60-300, configurable 6, 120, 180,
for telemetry permanent

Signal loss (bedside and/or Required Yes Yes


telemetry stop communicating
with
Alarmcentral)
latching Preferred Yes No
Alarm audit logs Via Infinity Gateway Not specified
Alarm History
Database
Other Choice of Draeger Choice of different
Infinity, IEC fast alarm audios
(default), or IEC slow
(IEC tones with
selectable cardiac or
alternate tone) audio
alarm patterns

REPORTS Preferred Yes Yes


Data type Simultaneous ECG, Trend, full disclosure,
full disclosure (strip, events, single-bed
1-hour, 24-hour, shift, view
selected strip,
graphical trend,
patient status, ECG
caliper report), event
disclosure (strip shift,
graphical trend,
selected event, ECG
caliper), rest ECG

SYSTEM CONFIGURATION
Hardware platform Proprietary PC Proprietary PC;
workstation optional software only

Operating system Linux Windows

SYSTEM SECURITY Required Clinical and biomed Clinical and biomed


password controls, password controls
hard drives can be
locked with key

HIPAA COMPLIANT Required Yes NA


NETWORKING Yes Yes
Communication between Preferred Yes No
multiple
Max central
no. of stations
central stations on 256 NA
single
Max network
no. of devices on single 1,024 64
network
Total (bedside and 1,024 64
telemetry)
Bedside monitors 1,024 64
Telemetry devices 1,024 No
Bedside monitor-to-bedside Yes Yes
monitor
Alarmscommunication Preferred Configurable in Configurable in
groups groups
Remote viewing Preferred Yes Yes
Central station required for No, bedside patient No
networking monitoring; yes,
telemetry
INTERFACES Based on hospital Yes Yes
needs
HL7 outbound (vital signs) Via optional Infinity XML format
Gateway
CIS/bedside information Via optional Infinity XML format
system Gateway
EMR Via optional Infinity XML format
Gateway
ECG data management Via optional Infinity NA
system Gateway
HL7 inbound (ADT) Yes No
Ancillary alarm notification Via optional Infinity Not specified
systems Gateway
STRIP CHART RECORDER Yes Yes
No. of channels 2 3
PRINTER Dual-laser printer One laser printer
support support
LINE POWER, VAC As required 115-230, 2-4 A 115-230, 2-4 A
UPS Preferred Optional Optional
SUPPORT
Response time, hr <24 <2 phone, on-site On-site next business
next business day day
Phone support Yes No
Local Yes Yes
Remote diagnostics Yes No
PURCHASE INFORMATION
List price $49,050<1> $15,077
(€13,398)<1>
Warranty 1 year 1 year

Training On-site training On-site training


customized based on customized based on
staff size and staff size and
monitoring solution monitoring solution
installed installed; typically 1
hr end-user and 2 hr
super-user

Delivery time, ARO 4-6 weeks 4-6 weeks


Fiscal year January to December January to December

OTHER SPECIFICATIONS Optional RAID for Optional RAID.


operating system and
RAID for database.

UMDNS CODE(S) 20179 20179 20179


LAST UPDATED June 2020 June 2020
Supplier Footnotes These
recommendations are
the opinions of
ECRI's technology
experts. ECRI
assumes no liability
for decisions made
based on this data.

Model Footnotes
Data Footnotes <1>Pricing <1>$5,711 (€5,075)
information derived for software solution
from ECRI Institute's only.
PricePaid database.
Includes base
system, displays and
optional software.
Nihon Kohden Corp Nihon Kohden Corp
CNS-6201A CNS-6801A

Worldwide, except USA


USA

Yes Yes
Yes Yes

16 16

32 32

LCD LCD

Single and dual Single and dual

61 (24) 61 (24)

Yes Yes

Yes Yes
Yes Yes
Yes Yes

Keyboard, mouse, Keyboard, mouse,


touchscreen touchscreen, wireless
remote controller

Yes Yes
8 8
75 75
Yes Yes
8 8
16 16

Yes Yes
Yes Yes
Yes Yes
120, minute by 120, minute by
minute minute
Yes Yes
Yes Yes
Yes Yes
≤12, varies ≤12, varies

Yes Yes
Yes Yes
Yes Yes
Yes Yes
≤12, varies ≤12, varies

Yes Yes
Yes/yes Yes/yes
Alarm history, ST Alarm history, ST
recall, recall,
hemodynamics, hemodynamics,
arrhythmia recall, arrhythmia recall,
interpretive 12-lead interpretive 12-lead
ECG ECG

10,000 files event 10,000 files event


history; 120 hours ST history; 120 hours ST
recall; 256 files recall; 256 files
hemodynamics; hemodynamics;
1,500 files arrhythmia 1,500 files arrhythmia
recall; 200 files recall; 200 files
interpretive 12-lead interpretive 12-lead
ECG, per patient ECG, per patient

120 120
None specified None specified

Yes Yes
120 120
All All

16 16 per bed

80 (user- 80 (user-
programmable) programmable)

Off (user- Off (user-


programmable) programmable)

Off (user- Off (user-


programmable) programmable)
Off (user- Off (user-
programmable) programmable)

60 (user- 60 (user-
programmable) programmable)

50 (user- 50 (user-
programmable) programmable)

140 (user- 140 (user-


programmable) programmable)

90 (user- 90 (user-
programmable) programmable)

Off (user- Off (user-


programmable) programmable)

Off (user- Off (user-


programmable) programmable)

3 (out of 7), user- 3 (out of 7), user-


programmable programmable

Yes Yes
High/low, arrhythmia, High/low, arrhythmia,
technical, crisis, technical, crisis,
warning, advisory, warning, advisory,
information information

Yes Yes

Yes Yes
Yes Yes
User configurable User configurable

No No
NA NA

Telemetry and Telemetry and


bedside bedside
For alarm duration For alarm duration
Not specified Not specified
None specified None specified

Yes Yes
Various Various

PC PC

Embedded Windows Windows 10


2009
Multilevel user ID and Multilevel user ID and
password, user levels password, user levels

Yes Yes
Yes Yes
Yes Yes
Unlimited Unlimited
2,000 2,000
2,000 2,000
≤1,000 ≤1,000
≤1,000 ≤1,000
Yes Yes
Yes Yes

Yes Yes
No No

HIS, CIS HIS, CIS

Yes Yes
Yes Yes

Yes Yes

Yes Yes

Yes Yes
Not specified Not specified

Yes Yes
3 3
Optional Optional

By country By country
Yes Yes

<1 <1

Yes Yes
Yes Yes
No No

$33,080 $36,388

2 years, parts and 2 years, parts and


labor labor
Included Included

30 days 30 days
April to March April to March

None specified. Send to EMR


functionality; remote
client also available
(RNS-6803/RNS-
6804).
20179 20179
June 2020 June 2020