You are on page 1of 1

-|Transparent setting guide|-

DRW170305AD

Ordering Information Specifications


TPS30 G 2 9 V G8 00 (0 to 0.5MPa)
Non-indicating Pressure Transmitters
Series TPS30
※1
① ② ③ ④ ⑤ ⑥ ⑦ ⑧ Pressure type Gauge pressure, absolute pressure Sealed gauge pressure Gauge pressure
Rated pressure range (MPa) 0 to 0.1 0 to 0.2 0 to 0.7 0 to 1 0 to 2
-0.1 to 0 -0.1 to 0.1 -0.1 to 0.7 -0.1 to 1 -0.1 to 2 0 to 3.5 0 to 5 0 to 10 0 to 20 0 to 40 0 to 50 0 to 60

TPS30 SERIES
Description
①Item TPS30 Pressure Transmitter Expanded analog output -0.1 to -0.1 to -0.1 to -0.1 to -0.1 to
0 to 0.11 0 to 0.22 0 to 0.77 0 to 1.1 0 to 2.2 0 to 3.85 0 to 5.5 0 to 11 0 to 22 0 to 44 0 to 55 0 to 66
②Measurement G Gauge pressure, sealed gauge pressure※1 range (MPa) 0.01 0.12 0.78 1.11 2.21
Max. pressure range (MPa) 0.6 0.6 3 3 3 0.6 0.6 3 3 3 10 20 50 80 120 120 120
INSTRUCTION MANUAL pressure A
1
Absolute pressure
Head type Burst pressure (MPa) 0.6 0.6 3 3 3 0.6 0.6 3 3 3 15 30 75 120 160 160 160
2 DIN43650-A connector type Measured materials Liquid, gas, oil
③Cable 3 M12 connector type Power supply Voltage output type: 8-36VDCᜡ (ripple P-P: max. 10%) Current output type: 11-36VDCᜡ (ripple P-P: max. 10%)
4 DT04-3P connector type Permissible voltage range 90 to 110% of rated voltage
5 Cable type Current consumption Voltage output type: max. 20mA Current output type: max. 30mA
Gauge pressure Absolute pressure Response time Max. 1ms
3 0 to 0.1MPa 0 to 0.1MPa Protection circuit Reverse polarity protection circuit
4 0 to 0.2MPa 0 to 0.2MPa Output type Voltage output type: 1-5VDCᜡ Current output type: DC4-20mA
5 0 to 0.7MPa 0 to 0.7MPa Compensation temperature -10 to 80℃ 0 to 80℃
6 0 to 1MPa 0 to 1MPa Accuracy Max. ±0.5% F.S. (including linearity, hysteresis, reproducibility)
7 0 to 2MPa 0 to 2MPa Linearity Max. ±0.2% F.S.
8 ※2
0 to 3.5MPa - Hysteresis Max. ±0.2% F.S.
9※2 0 to 5MPa - Temp. Zero Shift Max. ±0.1% F.S./10℃ (standard), max. ±0.25% F.S./10℃ (max.)
A※2 0 to 10MPa - Temp. Span Shift Max. ±0.1% F.S./10℃ (standard), max. ±0.25% F.S./10℃ (max.)
B※2 0 to 20MPa - Temperature characteristics - -25 to 100℃: max. ±1.5% F.S. / -40 to 125℃: max. ±2.5% F.S.
④Pressure range Load resistance Current output type: max. 700Ω (supplying 24VDC)
C※2 0 to 40MPa -
D※2 0 to 50MPa - Dielectric strength 500VAC 50/60Hz for 1 minute
E ※2
0 to 60MPa - Insulation resistance Over 100MΩ (at 500VDC megger)
Cable type DIN43650-A DT04-3P M12 Head type Sealed gauge pressure
※1 Ambient Voltage output Head type, DIN43650-A connector type, M12 connector type, DT04-3P connector type: -40 to 125℃, storage: -40 to 125℃ Cable type: -40 to 80℃, storage: -40 to 80℃

Environ-
Connector type Connector type Connector type temp. Current output Head type, DIN43650-A connector type, M12 connector type, DT04-3P connector type: -40 to 85℃, storage: -40 to 125℃ Cable type: -40 to 80℃, storage: -40 to 80℃

ment
F -0.1 to 0MPa
G -0.1 to 0.1MPa Ambient humidity 35 to 85%RH
Thank you for choosing our Autonics product. Fluid temp. -40 to 125℃
H -0.1 to 0.7MPa
Please read the following safety considerations before use. J -0.1 to 1MPa Vibration 10g, 20 to 2,000Hz 20g, 20 to 2,000Hz
K -0.1 to 2MPa Shock 100g/6ms 500g/1ms
Safety Considerations Z Others Tightening torque Max. 10N m.
V Voltage (1-5VDC) output Protection structure Head type, M12 connector type, DT04-3P connector type, cable type: IP67 (IEC standard) DIN43650-A connector type: IP65 (IEC standard)
※Please observe all safety considerations for safe and proper product operation to avoid ⑤Output type Material Stainless steel 316L (head component of head type: aluminium diecasting), connector: Polybutylene terephthalate G30, water-proof rubber: Silicon
A Current (DC4-20mA) output
hazards. Connection Voltage output type: +, -, Vout Current output type: +, -
G8 G3/8 (PF) (EN837)
※ symbol represents caution due to special circumstances in which hazards may occur. G4 G1/4 (PF) (EN837) Approval
⑥Pressure port R2 R1/2 (PT) (DIN3852) Head type: approx. 330g (approx. 250g)
Warning Failure to follow these instructions may result in serious injury or death. N4 NPT1/4 (DIN3852)
Weight
※2
DIN43650-A connector type, M12 connector type, DT04-3P connector type: approx. 130g (approx. 50g) Cable type: approx. 200g (approx. 120g)
Caution Failure to follow these instructions may result in personal injury or product damage. ZZ ※3
Others (option) ※1: The sensor is sealed structure. It is based on atmospheric pressure 101.3kPa (1.013bar). ※2: The weight includes packaging. The weight in parentheses is for unit only.
00 Not used ※Environment resistance is rated at no freezing or condensation.
Warning ⑦Option
2I "I" type 2m
1. Fail-safe device must be installed when using the unit with machinery that may (connector cable)
※4 2L
5I
"L" type 2m
"I" type 5m Connector Connection Cable (sold separately) (unit: mm)
cause serious injury or substantial economic loss. (e.g. nuclear power control, ● Voltage output type ● CID3-2/CID3-5 ● CLD3-2/CLD3-5
5L "L" type 5m
medical equipment, ships, vehicles, railways, aircraft, combustion apparatus, safety DIN43650-A M12 connector DT04-3P
⑧User pressure range User pressure range※5 40 Ø14.8
equipment, crime/disaster prevention devices, etc.) Head type Cable type
※1: The pressure is sealed gauge pressure. The unit is sealed structure. It is based on atmospheric connector type type connector type M12 M12
Failure to follow this instruction may result in personal injury, fire or economic loss. pressure 101.3kPa (1.013bar). Ø5 Ø5

27.5
2. Do not use the unit in the place where flammable/explosive/corrosive gas, high ※2: G1/4 is the standard pressure port. Pin
humidity, direct sunlight, radiant heat, vibration, impact, or salinity may be present. For the other pressure ranges, G3/8, R1/2 are standard pressure ports. type Brown
Failure to follow this instruction may result in explosion or fire. ※3: The option ports are sold separately. In case of large amount ordering, contact the Autonics for Black
3. Do not disassemble or modify the unit. manufacturing the requested pressure port. Blue Ø14.8 32
Failure to follow this instruction may result in fire or electric shock." ※4: Only for M12 connector type. L (m) L (m)
※5: Write the desired pressure range and it is the default of user pressure range.
(select "Z" at ④Pressure range) Function Pin
Caution
+ + 1 1 A Brown Model L(m) Meterial
1. Do not apply beyond the rated pressure. Dimensions (unit: mm) - - 3 C Blue C D3-2 2
Failure to follow this instruction may result in product damage. ● Head type 115
Vout Vout 2 4 B Black PVC
2. Use the unit within the rated specifications. 19 14.2 82 N.C - 3 2 - - C D3-5 5
Failure to follow this instruction may result in fire or product damage.
5.6 ● Current output type
3. Fix the cable through the cable connection part and do not turn the cable of the ※ Only for M12 connector.
unit. DIN43650-A M12 connector DT04-3P
Head type Cable type
22

Failure to follow cause instruction may result in product damage. connector type type connector type
4. Keep metal chip, dust, and wire residue from flowing into the unit.
Failure to follow this instruction may result in fire or product damage. Pin Cautions during Use

90
5. Check the polarity of the contact before wiring the unit. type Brown
1. Follow instructions in 'Cautions during Use'. Otherwise, It may cause unexpected accidents.

Failure to follow this instruction may result in product damage by a fire. G3/8 1 Blue 2. 8-36VDC, 11-36VDC power supply should be insulated and limited voltage/current or Class
6. This product is designed to detect the pressure of noncorrosive fluid. Do not use (EN837)
2, SELV power supply device.
for corrosive fluid.
Failure to follow this instruction may result in product damage. Function Pin 3. When installing the unit on pipe line, use the hexagon part of connections not to turn the
7. Use a dry cloth to clean the unit, and do not use water or organic solvent. ● DIN43650-A connector type + + 1 1 A Brown unit with a pipe wrench. Do not use the unit with strong vibrations.
Failure to follow this instruction may result in fire. 63.2 □27 4. Store the unit at the place without moisture, dust, and vibration.
- - 3 C Blue
Ø24 19 5.6 24.4 33.2 N.C Vout 2, 3 2, 4 B - 5. This product which does not have drive part at sensing part does not need to repair it. Even
Connections ※In case of head type, remove the top cover. though inside of pressure pipe is normally clean, it needs to take maintenance once a year
Voltage (1-5VDC) output type Current (DC4-20mA) output type as below instructions.
22

① Check the broken status of outside.


(brown) VCC (brown) VCC ② Check the pressure slot, cleanliness inside, and corrosion state.
46.5

LOAD LOAD ※1
G3/8 ③ Short each terminal and check the insulation resistance between the case and power.
Main Circuit

Main Circuit

(black) Voltage
(1-5VDC) Output + Current + (EN837)
8-36VDC amplifier 11-36VDC 6. When removing a sensor for maintenance, follow the below instructions.
DIN43650-A connector 46.8
-
(blue) GND
-
● DT04-3P connector type ①Replace an O-ring which is used once.
Ø24
(blue) GND
19 5.6 18.8 19 3.4 ②Be sure that diaphragm part is not damaged.
● M12 connector type
7. Switch or circuit breaker for suppling or cutting off the power should be installed nearby
35.9
Ø24 users for convenient control.
※Cable color is only for cable type.
19 5.6 20.3 10 8. The unit cannot be repaired due to disassembled structure.
Troubleshooting

Ø21
22

9. The unit is fixed with bolt and nut at the both sides of case.
Error Troubleshooting Do not press excessive load (approx. 300kg/cm²), or it may cause damage to the unit.
Check the power supply. ※1 Connector: CS-DT3P, 10. This unit may be used in the following environments.
Ø21

G3/8
22

No outputs Check the polarity (+, -) when wiring the cable. DT04-3P connector sold separately
(EN837) ① Indoor / Outdoor (in the environment condition rated in 'Specifications')
Check the connection part.
② Altitude max. 2,000 m
Abnormally fluctuating Check the power supply. ※1 ● Cable type 47.6
Check the supplied pressure. G3/8 M12 connector Ø24 ③ Pollution Degree 2
output (EN837)
Check the pressure line. 19 5.6 20.3 21.7 ④ Installation Category II
※1: Pressure port
Check the power supply. 15 14.3 12
Check the load resistive value of current output type for a
Out of zero point output receiver is over 700Ω. (when supplying 24VDC)
value Check the measuring point and transmission distance.

Ø21
22

Check the line resistance is below 700Ω.


※The above specifications are subject to change and some models may be discontinued
without notice.
G1/4 R1/2 NPT1/4 ※1 Ø5, 2m 18, Bansong-ro 513Beon-gil, Haeundae-gu, Busan, Republic of Korea, 48002
※Be sure to follow cautions written in the instruction manual and the technical descriptions G3/8
(catalog, homepage). www.autonics.com | +82-51-519-3232 | sales@autonics.com
(EN837) (DIN3852) (DIN3852) (EN837)