23.8.2018 How to calibrate pressure gauges - 20 things you should consider
ENG
1. Accuracy classes
2. Pressure media
3. Contamination
4. Height difference
https://blog.beamex.com/how-to-calibrate-pressure-gauges 2/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
What is pressure?
Before we discuss each one of the things to consider when calibrating pressure gauges, let’s take a
quick look into a few more basic concepts.
What is pressure? Pressure is the force that is perpendicular to the surface divided by the area it is
effecting. So pressure equals force per area, or p = F / A.
There are a large number of different pressure units used around the world and this can be
sometimes very confusing. The engineering unit for pressure, according to SI system, is Pascal
(Pa), being a force of one Newton per one square meter area, 1 Pa = 1 N / m2. Since Pascal is a
very small unit, it is most often used with coefficients, such as hecto, kilo and mega. A large
number of different pressure unitgs are being used around the world. For more information on
pressure and different pressure units and their background, please see the blog post Pressure units
and pressure unit conversion.
For an on-line pressure unit conversion tool, please visit the page Pressure unit converter.
Pressure types
https://blog.beamex.com/how-to-calibrate-pressure-gauges 3/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Several different pressure types exist including gauge, absolute, vacuum, differential and barometric.
The main difference ofENG types is the reference point against which the measured
thesepressure
pressure is being compared. Pressure gauges also are available for all of these pressure types. Also,
compound gauges are available, including a combined scale for both positive gauge pressure and
vacuum (negative gauge) pressure.
For more detailed information on different pressure types, please see post Pressure calibration
basics – pressure types.
Pressure gauges
When talking about pressure gauges, it is normal to refer to analog pressure indicators which are
provided with a pointer needle and a pressure scale. These are normally manufactured according to
EN 837 or ASME B40.100 standard.
Often these kind of analog pressure gauges are built with a Bourdon tube, diaphragm or capsule.
There is a mechanical structure that moves the pointer as pressure increases causing the pointer
move across the scale.
Pressure gauges are divided into different accuracy classes that specify the accuracy of the gauge
as well as other attributes. Available pressure ranges are typically divided in steps with coefficients
1, 1.6, 2.5, 4, 6 continuing into the next decade (10, 16, 25, 40, 60) and so on. The different gauge
diameters (of scales) are typically 40, 50, 63, 80, 100, 115, 160 and 250 mm (1 ½, 2, 2 ½, 4, 4 ½,
and 6 inches). More accurate gauges typically have a bigger diameter.
Pressure connectors are normally parallel pipe threads (G) according to ISO 228-1, or taper pipe
threads (NPT) according to ANSI/ASME B1.20.1.
There are also digital pressure gauges that have a numeric pressure indication instead of an analog
pointer. This article focuses on analog gauges, but most of the principles are valid for both.
Pressure gauges are commonly used in all industries and are a very common instrument to be
calibrated. As with any process measurement device, it should be calibrated at regular intervals to
assure that it is measuring correctly. Gauges being mechanical instruments, adds the risk for them to
drift due to mechanical stress.
For more information on why you should calibrate instruments, please see the blog post Why
calibrate?
For more information on how often instruments should be calibrated, please see post How often
should instruments be calibrated?
https://blog.beamex.com/how-to-calibrate-pressure-gauges 4/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
1 - Accuracy classes
Pressure gauges are available in many different accuracy classes. Accuracy classes are specified in
ASME B40.100 (accuracy classes from 0.1 to 5 % range) as well as in EN 837 (accuracy classes
from 0.1 to 4 % range) standards. The accuracy class specification most often being “% of range”
means that if the accuracy class is 1% and if the scale range is zero to 100 psi, the accuracy is ± 1
psi.
Make sure you know the accuracy class of the gauge you are going to calibrate, as this will naturally
specify the acceptable accuracy level, but it will also have other effects on the calibration
procedure.
2 - Pressure media
When calibrating pressure gauges, the most common pressure media are gas or liquid. Gas is most
often regular air, but in some applications, it can also be different gases, such as nitrogen. Most
commonly, the liquid is water or oil. The pressure media during the calibration depends on the
media that is used in the process that the gauge is connected to. Media also depends on the
pressure range. Low pressure gauges are practical to calibrate with air/gas, but as the pressure
range gets higher it is more practical and also safer to use liquid as the media.
https://blog.beamex.com/how-to-calibrate-pressure-gauges 5/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
3 - Contamination ENG
While installed in a process, the pressure gauge uses a certain type of pressure media, this should
be taken into account when selecting the media for the calibration. You should not use a media
during the calibration that could cause problems when gauge is installed back to process. Also, the
other way around, sometimes the process media could be harmful to your calibration equipment.
There can be dirt inside the gauge that can get into the calibration equipment and cause harm.
With a gas operated gauges, you can use a dirt/moisture trap, but for a liquid operated gauge, you
should clean the gauge prior to calibration.
One of the most extreme process situations is if the gauge is used in to measure the pressure of
oxygen. If any grease goes into a high pressure oxygen system during the calibration of the gauge,
it can be very dangerous and could cause an explosion.
4 - Height difference
If the calibration equipment and the gauge to be calibrated are at a different height, the
hydrostatic pressure of the pressure media in the piping can cause errors. This normally is not an
issue when gas is used as the media, as gas is light compared to liquid. But when liquid is used as
media, the liquid in the piping will have a weight due hydrostatic pressure and can cause errors. The
magnitude of the error depends on the density of the liquid and the difference in height, as the
gravity is pulling the liquid inside the tubing. If it is not possible to have the calibrator and gauge at
the same height, then the effect of height difference should be calculated and taken into account
during the calibration.
An example of effect of hydrostatic pressure:
https://blog.beamex.com/how-to-calibrate-pressure-gauges 6/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
As on example: if water is the media (density 997.56 kg/m3), local gravity is 9.8 m/s2 and there is
a 1 meter (3.3 feet) difference
ENG between the DUT and the reference equipment, this will cause an
6 - Adiabatic effect
In a closed system with gas as the pressure media, the temperature of the gas effects the volume
of the gas, which has an effect to the pressure.
When pressure is increased quickly, the temperature of the gas will rise, and this higher
temperature makes the gas to expand, thus having a bigger volume and higher pressure. When the
temperature starts to cool down, the volume of the gas becomes smaller and this will cause the
pressure to drop. This pressure drop may seem like a leak in the system, but it is actually caused by the
adiabatic effect due to change in the gas temperature. The faster the pressure is changed, the
bigger is the effect is. The pressure change caused by this effect will gradually get smaller as the
temperature stabilizes.
So, if you change the pressure quickly, make sure you let it stabilize for a while before judging that
there is a leak in the system.
7 – Torque force
https://blog.beamex.com/how-to-calibrate-pressure-gauges 7/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Especially for torque sensitive gauges, don’t use excessive force when connecting pressure
connectors to the gauge,
ENGas it
may damage the gauge. Follow manufacturer’s instructions for the
allowed torque force. Take the time to use proper tools, appropriate adapters and seals.
9 - Generating pressure
To calibrate a pressure gauge, you need to source the pressure applied to the gauge.
There are different ways to do that: you can use a pressure hand pump, a pressure regulator with a
bottle or even a dead weight tester. A dead weight tester will provide a very accurate pressure and
you don’t need a separate calibrator to measure the pressure, but dead weight tester is expensive,
not very mobile, requires a lot of attention to use and it is sensitive to dirt.
It is more common to use a pressure calibration hand pump to generate pressure and an accurate
pressure measurement device (calibrator) to measure the pressure. A pressure controller can also
be used to supply the pressure.
https://blog.beamex.com/how-to-calibrate-pressure-gauges 8/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Picture: Left gauge in the below picture is difficult to read accurately as the indicator is between
scale marks, while the right one is easy to read since the applied pressure is adjusted so that pointer
is exactly on scale mark:
Picture: Many high accuracy pressure gauges are provided with a mirror along the scale, helping to
view the gauge perpendicular, as the mirror image of the pointer is hidden behind the pointer, or
with the help of the reflection of the pointer:
https://blog.beamex.com/how-to-calibrate-pressure-gauges 9/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
ENG
Remaining topics
To prevent this blog post coming way too long, please download the white paper and read all the 20
topics from that.
The remaining topics not covered here include:
https://blog.beamex.com/how-to-calibrate-pressure-gauges 10/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
ENG
Related resources
Beamex products suitable for pressure calibration, including pressure gauge calibration:
https://www.beamex.com/calibrators/pressure-calibrators/
An online tool for pressure unit conversion on Beamex web site: Pressure unit converter
Back to top
https://blog.beamex.com/how-to-calibrate-pressure-gauges 11/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Dear Julian,
We have translating some material into Spanish and plan to do that more in the future.
Unfortunately this one is not yet translated.
Thanks for the comment,
Heikki
Dear Jose,
Thanks for your kind comments.
As you see my earlier comment, we are working to get more Spanish material in the future.
Thanks,
Heikki
https://blog.beamex.com/how-to-calibrate-pressure-gauges 12/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Hi
thanks for your beautiful explaination .
I have a question . for a analog gage what's the relationship between accuracy and
resolution?
which of them is smaller?i think normally it would be correct that accuracy must be smaller
than the distance between two minor sign in a analog gage . Is that correct?
Dear Sir,
I can see that I did not cover that topic enough in the blog post.
There are several variables that will affect the relationship between gauge accuracy and
minor scale division (resolution). This depends on the: accuracy class of the gauge, the
nominal size (diameter) of the gauge and the pressure range of the gauge.
In most cases, it is like you suggested, i.e. the resolution (minor scale division) is actually
larger that the accuracy of the gauge. But that is not always the case.
• If we look at the accuracy classes 1, 1.6, 2.5 and 4, there are different situations. All of
these accuracy ENGclasses have
the common specifications for the scale numbering, although
the accuracies are different. The resolution of these varies between 1.6% to 5%. So we can
see that if you have accuracy class 1, the resolution is always larger than accuracy. But if
you have accuracy class 4 gauge, then the resolution is in most cases better than the
accuracy class.
Generally, we can say that if the gauge diameter is larger, then there are more scale
markings and the scale division is smaller and it offers better resolution compared to
accuracy class.
In practice, you can often read the indication of an analog gauge more accurately than the
minor scale division is. I mean that when the pointer is in between of scale marks, you can
estimate the indicated value between scale marks.
Thanks,
Heikki
I would like to clarify a doubt regarding the part of "Height difference" where it explains the
effect that the hydrostatic pressure has on the system. Specifically my doubt is about the
formula you are using for that calculation where you are relating the gravitational
acceleration (g = local gravity (m / s2) and I do not see how to get to the 9.8 kPa result
when we start canceling units. As my best knowledge, the formula in the International
Measurement System (S.I.) for determining the hydrostatic pressure in kPa created by a
liquid head is as follows:
Ph = SG x h (kPa /0.102 m)
Where:
Ph = hydrostatic pressure (kPa, gauge)
SG = specific gravity of liquid (dimensionless)
h = liquid level (m)
As can be seen, only the specific gravity of the liquid (also known as relative density) and
https://blog.beamex.com/how-to-calibrate-pressure-gauges 14/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
the height is needed to determine the pressure exerted by the liquid column. Based on your
be:
example this wouldENG
I would like to clarify a doubt regarding the part of "Height difference" where it explains the
effect that the hydrostatic pressure has on the system. Specifically my doubt is about the
formula you are using for that calculation where you are relating the gravitational
acceleration (g = local gravity (m / s2) and I do not see how to get to the 9.8 kPa result
when we start canceling units. As my best knowledge, the formula in the International
Measurement System (S.I.) for determining the hydrostatic pressure in kPa created by a
liquid head is as follows:
Ph = SG x h (kPa /0.102 m)
Where:
Ph = hydrostatic pressure (kPa, gauge)
SG = specific gravity of liquid (dimensionless)
h = liquid level (m)
As can be seen, only the specific gravity of the liquid (also known as relative density) and
the height is needed to determine the pressure exerted by the liquid column. Based on your
example this would be:
https://blog.beamex.com/how-to-calibrate-pressure-gauges 15/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Dear Ricardo,
ENG
Thanks for the comments.
I must admit that I am not very familiar with the use the "relative density".
But as you did show it will get to the same result and is maybe a little bit easier calculation.
As Pascal is N/m^2, and Newton is kg m /s^2, that result kg/m*s^2 (or kg*m-1*s-2 or
kg/(m*s^2)) is Pascal.
I'm sure you know that Pascal is one of the “derived” units in the SI system, being derived
from SI base units kilogram, meter and second.
Kind regards,
Heikki
Interesante informacion
https://blog.beamex.com/how-to-calibrate-pressure-gauges 16/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
There was a lot of pressure gauges that I wasn't aware of. I really liked that this simplified the
principle of it to its ENG
minimum.
Understanding the basics tends to make other aspects of this
type of equipment easier to understand.
Great information here. Thanks for that. The problem is that calibration of pressure gauges
is an often-overlooked process, which means that pressure readings could get
compromised. If a plant is working with pressure gauges that are out of calibration, they are
not only putting product quality at risk, they are also putting safety at risk.
https://blog.beamex.com/how-to-calibrate-pressure-gauges 17/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Hi John,
Thank you for your feedback and very good comments.
You are absolutely right, if pressure gauges (or any other important process
instrumentation) are not calibrated properly, they can cause product quality issues and also
safety issues. And these issues will cause even bigger issues for the company in the long
run.
Heikki
Hi, Dear
Dear Yury,
We do offer many kinds of training, the availability depends on the country you are located
in.
https://blog.beamex.com/how-to-calibrate-pressure-gauges 18/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Have a look at Beamex representative in your country in the below link and contact them
for training availability
ENG in your country:
https://www.beamex.com/beamex-worldwide/
Thanks,
Heikki
Dear Jose,
We do offer many kinds of training, the availability depends on the country you are located
in.
Have a look at Beamex representative in your country in the below link and contact them
for training availability in your country:
https://www.beamex.com/beamex-worldwide/
Thanks,
Heikki
how we identify the classes of the gauge ? can you plz explain briefly ....
https://blog.beamex.com/how-to-calibrate-pressure-gauges 19/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Dear Mohamed,
Typically the pressure
ENG gauge has the accuracy class indicated in it's front cover. It maybe
If the gauge have no accuracy class indicated and the manufacturer data is
not available, it gets a bit difficult.
You can guess something based on the diameter of the gauge, the bigger
ones are more accurate. Generally diameter 40 - 63 mm are 1.6 - 2.5 % and
100 - 160 mm are 1 %.
Also, the more accurate ones have more scale marks (better resolution).
Thank you so much ....for your answer .....your doing great job
https://blog.beamex.com/how-to-calibrate-pressure-gauges 20/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
ENG
Heikki Laurila 12.12.2017, 11:43:15
Dear Mario,
Thank you for your interest on the article, I am sorry that we have not yet translated this
article into Spanish.
Please note that we have many articles translated into Spanish and these can be found in
our Spanish web site in following link:
https://www.beamex.com/es/white-papers-ebooks/
Thanks,
Heikki
https://blog.beamex.com/how-to-calibrate-pressure-gauges 21/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Hello Adebesin,
Thanks for your interest!
If you fill in your email address to the "Subscribe to Email Updates" field on the right hand
side of blog, or in blog footer, you will be notified by short email when a new blog post is
available. Normally that is once or twice in a month.
Or alternatively, you can fill in your email to the "Get calibration tips right to your inbox:"
that can be found in the footer of all our web pages.
Thanks,
Heikki
Dear Sir,
Best Regards
https://blog.beamex.com/how-to-calibrate-pressure-gauges 22/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Dear Kemal,
The standards don't
ENG mention
"verification", they are talking about the "calibration".
So it is not perfectly clear to me what you mean with this verification, but I assume it is not
a full calibration, but a more simple "check" of the accuracy of the gauge in between actual
periodical calibrations.
Anyhow, I am not aware of any official recommendation of the tolerance for this kind of
verification.
But if you make this verification every month, I think the tolerance should be tighter than
the tolerance for the calibration that is done for example once per year. I mean if your one-
year calibration tolerance is for example 1%, it should not drift more than a fraction of that
in one month, especially in the beginning of that one year period, but can be closer to 1 year
tolerance when the time is getting closer to 1 year.
Feel free to send me email to discuss more.
Thanks,
Dear Sir,
Thank you very much for your clarfying informations. It is very impressive to see you
helping many people all over the world...
Best Regards
https://blog.beamex.com/how-to-calibrate-pressure-gauges 23/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
I am a calibration technician in Lagos, Nigeria. I read one of Beamex books, which was of
great help to me. Thanks so much.
https://blog.beamex.com/how-to-calibrate-pressure-gauges 24/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Heikki Laurila
ENG
23.2.2018, 8:21:33
Hi Adewale,
Thanks for your comments!
I'm glad the book was helpful.
All the best to Lagos.
Thanks,
I just read ur blog and iam very interested with ur blog. I have a problem to make a decision
of acceptence criteria with differential pressure calibration. I was read at some literature
that shown the acceptance criteria of pressure calibration is 1,5% of full scale. But the
others is shown 10% of full scale. May u giving me any ideas to make sure which is better?
Notes = my result of calibration is always shown above of 1,5%. The range is 3 - 6 %.
Hi Bobby,
Thanks for your kind comments and also for your questions.
The suitable acceptance criteria depend on many things. It depends on the accuracy of the
instrument you are calibrating, on the accuracy need of your process using that meter, on
the criticality of that measurement the instrument is being used, on the regulatory
requirements, on the customer requirements and also on the uncertainty of your
calibration equipment.
So it is impossible for me to suggest a suitable acceptance criteria without knowing all the
details.
https://blog.beamex.com/how-to-calibrate-pressure-gauges 25/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Thanks,
Heikki
Thanks for the answers given and I concluded to use the accuracy of the instrument
I calibrated. Your answer really helped my problem. For the second question, I want
to ask how the temperature influence on pressure reading? the example of the case
study is I want to calibrate the differential pressure with a capacity of 0 - 60
Pascals. In the calibration process I also read thermohygrometer to read
temperature and RH environment. At point 0, I read the temperature of 25.1
degrees Celsius. point to 10 i read 25.6 degrees celcius. But the numbers do not
show any instability. Do I still have to write the temperature readings at each point?
or just enough once when I am in the process of taking data?
Best Regards,
Bobby Rahadian
https://blog.beamex.com/how-to-calibrate-pressure-gauges 26/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Heikki
ENG
Laurila 9.4.2018, 10:13:29
Hi Bobby,
But this depends very much on the instrument to be calibrated, and how
sensitive that is for temperature changes. If e.g. a pressure gauge is
calibrated and it is not sensitive for temperature, it is not that critical and one
temperature is enough to record.
But if the instrument to be calibrated is sensitive for temperature changes,
then it gets more important to record temperature, even in every calibration
point.
Most instruments have a temperature coefficient specified, so you can see
how sensitive your device is.
https://blog.beamex.com/how-to-calibrate-pressure-gauges 27/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
values. When the pointer falls between graduation, do i record the closest graduation value
or an estimate decimal value in addition to the upper graduation.
Dear Adewale,
Generally, you can read the analog gauge with higher precision than the scale marks
(graduation).
In our lab we normally use the reading accuracy of 1/5 of the scale marks, i.e. the space
between the scale marks is divided in 5 steps. The lab guys said that this 1/5 is based on a
standard.
Thanks for that 1/5 division of the space between scale marks info. I would love to know the
name of the standard. The book i read by Beamex is "Ultimate Calibration" Thank you
Dear Adewale,
The main source for this specification is in the standard EN 837-3; "Pressure gauges - Part
3: Diaphragm and capsule pressure gauges - Dimensions, metrology, requirements and
testing".
In the section 10.2 (Accuracy and hysteresis) The standard states following: "Pressure
readings shall be interpolated between 1/4 and 1/10 of the distance between two scale
marks."
So depending on how accurate scales marks there are in the pressure gauge, you can divide
https://blog.beamex.com/how-to-calibrate-pressure-gauges 28/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
the distance between the scale marks with a number from 4 to 10.
So the readability
ENGis between
1/4 to 1/10 of the scale marks.
The standard does not specify how to select the divider.
Thanks for your help. Trivially, our lab practice here is to wrap the thread(Teflon)
tape clockwise around pressure gauges threading. I would like to know if this is a
generally workshop best practice.
Hi Adewale,
The use of teflon tape depends on the threads on the pressure gauge, or in
any pressure connectors.
In case you have a "tapered" thread in the pressure connector (NPT), you will
need to use teflon tape to prevent leakage.
But if you have a parallel thread (BSP or G), you should never use tape, but
use a sealing washer instead.
Great! Your reply has made me make more research into different threadings. I would still
like to know what direction should the Teflon tape be wrapped around an everyday
thread(righthanded). Clockwise or anticlockwise?
https://blog.beamex.com/how-to-calibrate-pressure-gauges 29/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Hi Adewale,
The tape should be put so that when you tighten up the connector, the tape goes tighter.
If you put the tape wrong way, it starts to come loose when you tighten the connector.
Thanks,
Very useful and so peaceful blog for us. Such a good work and so good write up.
Kindly add more comments on the adjustments of pressure gauge readings deviation.
Pointer extraction and/or pointer gears realignment.
Dear Adewale,
Thanks for the good suggestion.
https://blog.beamex.com/how-to-calibrate-pressure-gauges 30/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Thanks for mentioning how a thread gage should be good at calibrating. I also like how you
said that they should control pressure as well too. My husband and I are looking for a thread
gage; thanks for the post. http://westportcorp.com/gages/thread.html
You know when it comes to high pressure more knowledge is better. Like your Blog and
your ideas and i think its beneficial for us and i keep to visit your blog regularly because i got
lot of information through you blog. Thank your for this info.
Ammonia Loss Reporting
Hi Angela,
Thanks for the comments, you are welcome to come back at anytime... ;-)
By the way, if you subscribe for blog notifications in the blog, you will be automatically
notified by email when a new blog post is published.
Leave a Reply
First Name*
https://blog.beamex.com/how-to-calibrate-pressure-gauges 31/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Last Name
ENG
Email*
Comment*
Submit Comment
Disclaimer
*
E-mail address...
OK
https://blog.beamex.com/how-to-calibrate-pressure-gauges 32/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
ENG
Calibrators
Software
Services
About Beamex
Our story
Testimonials
Case stories
Customer Service & Support
Worldwide contacts
Support contacts
Terms & Conditions
Learn more
https://blog.beamex.com/how-to-calibrate-pressure-gauges 34/35
23.8.2018 How to calibrate pressure gauges - 20 things you should consider
Beamex blog
Explore better ways to calibrate
ENG
Calibration tips
Manuals and brochures
E-mail address...
OK
https://blog.beamex.com/how-to-calibrate-pressure-gauges 35/35