SIRECON
Temperature Monitoring with the SIRECON Unit and OPTILUX Tube Units
2 Document Version / Disclaimer / Copyright
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens office to order a current version or refer to our website
http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is” without the assumption of any liability under
any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.
Copyright
“© Siemens, 2001” refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany
1 Introduction 5
2 Temperature Indicators 7
1.1.1 Indicators
Blue adhesive stickers with a temperature printed on them. The basic color of the inside
square is white (temperature not exceeded). If the admissible temperature is exceeded,
that inside square turns black.
Fig. 1:
37 C 37 C 40 C 40 C
The temperature indicators on the SIRECON housing are located in the vicinity of the
identification labels. These indicators must be read prior to beginning installation.
If one or both indicators on the I.I. housing show that the admissible temperature has
been exceeded by a color change, the 37˚C indicator on the OPTILUX tube unit absolutely
must be read. Access to these indicators is described on the following pages.
The I.I. models used in these systems have 37˚C and 40˚C temperature indicators only
on the outside, in the vicinity of the identification labels. For the procedure, see ( SIRE‐
MOBIL Compact / L, SIREMOBIL Iso-C, POWERMOBIL, ARCOSKOP, ARCADIS Varic and AR‐
CADIS Orbic / Page 7).
If exceeding the admissible temperature of 37˚C is also found for the I.I. tube, the
OPTILUX tube unit must also be replaced.
Exceptions: SIRECON 40-4 HDR, SIRECON 33-4 HDR-WA, SIRECON 33-4 HDR/W, SIRE‐
CON 33-4 HDR-WD, SIRECON 23-3 HDR-DA, SIRECON 23-3 HDR-DQ, SIRE‐
CON 23-3 HDR-D, SIRECON 23-3 HDR-DL, SIRECON 23-3 HDR-DX, SIRE‐
CON 17-2 HDR-L (LITHOSTAR 2), SIRECON 23 ISX, SIRECON 33 MD and
SIRECON 33 are replaced completely.
If no change in color is found, remove the temperature indicators on the SIRECON hous‐
ing prior to turning the unit over to the customer.
Return of the SIRECON unit or the OPTILUX tube unit to the factory must be made with
the completed image intensifier-tube unit questionnaire.
2.1.5.1 Procedure:
The I.I. housing types used on these systems only have externally readable temperature
indicators for 37C and 40C, which are located near the identification labels.
Read the two temperature indicators and evaluate them per the following description.
If there are differences to the values that were determined (within the listed tolerances)
prior to replacing the I.I., make sure that there are no other errors in the system (e.g. in‐
correct adjustment of the optical focus, electrical focus...).
If there are no longer any differences after making the repair, remove the two tempera‐
ture indicators and continue with startup. Otherwise, replace the I.I.
2.2.7 SIRECON 33 MD
(77 21 819 G5379) serial numbers 01001 and higher
2.2.8 SIRECON 33 M
(88 70 201 G5379) serial numbers 01001 and higher
1. With this Sirecon, the housing is not a bonded housing, but because of its internal design, the input
screen and the MU metal can be damaged.
2.2.9.1 Procedure:
The I.I. housing types used on these systems only have externally readable temperature
indicators for 37C and 40C, which are located near the identification labels.
Read the two temperature indicators and evaluate them per the following description.
n Take note of the surface of the image intensifier housing ( 2/Fig. 7 Page 14).
n Remove the appropriate cover panel.
n Read the temperature indicators as in ( 1/Fig. 7 Page 14).
Fig. 8: Location of the 33 HDR-WA temperature indicator Fig. 9: 33 HDR-WA temperature indicator
n Remove the cover and power supply and read the temperature indicators as in
( 1/Fig. 8 Page 15) and ( 1/Fig. 9 Page 15).
Fig. 10: SIRECON 33 ISX temp. indicator removed Fig. 11: SIRECON 33 ISX temp. indicator
(1) Location of the temp. indicator (1) Location of the temperature indicator
n The inside indicator can be seen only if the SIRECON is removed from the C-Arm
( Removing the 22 ISX / AXA3-500.841.01). See AXA3-500.841.01. ... CB-DOC Artis
U, Replacement of Parts.
Fig. 15: Removing the power supply Fig. 16: Temperature indicator
n Lift the power supply until the temperature indicators to the right of the cable become
visible ( 1/Fig. 16 Page 18) and read the temperature indicators.
healthcare.siemens.com/services
siemens.com/healthineers