You are on page 1of 4

Kingdom of Saudi Arabia ΔϳΩϮόδϟ΍ΔϴΑήόϟ΍ΔϜϠϤϤϟ΍

Ministry Of Health ΔΤμϟ΍Γέ΍ίϭ


General Directorate of Equipment Supplies ΔϴΒτϟ΍Ε΍ΰϴϬΠΘϟ΍Γέ΍Ω·
Department Of Specs. & Development
ήϳϮτΘϟ΍ϭΕΎϔλ΍ϮϤϟ΍Γέ΍Ω·

TECHNICAL & PERFORMANCE SPECIFICATION FOR MEDICAL EQUIPMENTS

ΔϳΑρϟ΍Ε΍ίϳϬΟΗϟ΍
ΔόηϷ΍ϡγϗΓίϬΟ΃

Item Code 4455200030001 GROUP # MI072

ITEM DESCRIPTION LITHOTRIPSY ϰλΣϟ΍ΕϳΗϔΗίΎϬΟ Ω˰˰ϧ˰˰Α˰˰ϟ΍ϡ˰˰˰˰˰γ΍

MODEL OFFERED MANUFACTURER COUNTRY OF ORIGIN

# TECHNICAL PARAMETERS SPECIFIED RESULT


1 UNIVERSAL ESWL SYSTEM FOR THERAPHY PROCEDURE WITH A SINGLE SHOCK WAVE
HEAD WITH FLOUROSCOPIC AND US LOCALIZATION.
2 FDA/CE CLEARANCE Yes
3 SHOCK-WAVE GENERATOR
4 Focal Dimensions, mm Specify
5 Type Specify
6 Focusing technique Specify
7 X- ray Locating Specify : ( Inline Motorized
Localization or Motorized C-arm
& Automatic Stone Positioning )
8 Number of shock wave heads One
9 Triggering ECG & Automatic
10 Physiological monitoring for ECG triggering. Required with NIBP and SPO2
included
11 Pulse frequency / min (Specify the Value)
12 Penetration depth > 160 mm approx.
13 Pressure at focus, MPa (Specify the Value)
14 Dynamic Range (Min- Max), MPa (Specify the Value)
15 Coupling method Dry or Water cushion (specify)
16 Coupling Cushion Diameter, mm Specify
17 Pressure Level Specify
18 Therapy Head Aperture Angle specify
19 Guaranteed life time for shock wave Coil or Electroconductive Electrode > 2.000.000 shocks approx..
20 Table side SW control Specify
21 Screen Operation in Treatment Room Specify
22 Control console SW control Yes
23 Localization Mode Auto positioning (Localization)
24 Ultrasound locating system Yes, (specify) - inline Motorized
Localization or Free Line
Localization
25 IMAGING fluoroscopy digital system
26 Generator High frequency
27 Nominal Voltage •  .9DSSUR[
28 Radiation dose reduction measures Yes needed
29 output , KW •  .:DSSUR[6SHFLI\
30 Kvp output 40 To 120 Kvp
31 Cine memory Up to 1300 images or more

16/Nov/2017 01:16 PM Page 1 of 4


Kingdom of Saudi Arabia ΔϳΩϮόδϟ΍ΔϴΑήόϟ΍ΔϜϠϤϤϟ΍
Ministry Of Health ΔΤμϟ΍Γέ΍ίϭ
General Directorate of Equipment Supplies ΔϴΒτϟ΍Ε΍ΰϴϬΠΘϟ΍Γέ΍Ω·
Department Of Specs. & Development
ήϳϮτΘϟ΍ϭΕΎϔλ΍ϮϤϟ΍Γέ΍Ω·

TECHNICAL & PERFORMANCE SPECIFICATION FOR MEDICAL EQUIPMENTS

# TECHNICAL PARAMETERS SPECIFIED RESULT


32 mA Range Up to 630 mA
33 Digital fluoroscopy Yes
34 Digital Radiography Yes
35 Pulse Fluoroscopy specify
36 AEC Yes
37 X-Ray Tube
38 Dual Focus Yes
39 Focal spot size, mm Specify
40 Anode heat storage capacity •  .+8$SSUR[
41 Multileaf collimator Yes
42 large Image intensifier Yes , Specify
43 Grid Specify
44 Fluoroscopy function Yes
45 Reference image display Yes
46 Image Matrix 1024 x 1024
47 Volatile image memory LIH & 100 images / storage on
USB stick or image archiving
system should be possible
48 Ultrasound
49 Transducer Yes, Micro convex
50 Central Frequency Mhz. 3.5 approx.
51 Modes B, M, B/M, B/B
52 Monitor size > 12 Inch or more
53 US Transducer Movement Yes, (Specify) - Inside (Inline) /
Outside (Outline) therapy head or
Freeline
54 US Transducer can get in contact with patient body Yes
55 Built in Storage 999 images in compact Flash
56 Cine Memory 1300 images or more
57 Beam Formation 12 bit Digital system
58 PATIENT TABLE
59 Automated Movement Yes , Electro Hydraulic- other
specify
60 Vertical movement, cm. Specify
61 Longitudinal movement, cm Specify
62 Lateral movement, cm Specify
63 Trendelenburg Yes
64 Treatment of Pediatric Patients Yes
65 Max patient weight in Kg. •  
66 Table Cut-Off Specify
67 CONTROL CONSOLE
68 Number of monitors 2 TFT/LCD
69 Reporting functionality yes
70 Monitor size, diagonal, cm 45 or more
71 Touch Screen Control Yes

16/Nov/2017 01:16 PM Page 2 of 4


Kingdom of Saudi Arabia ΔϳΩϮόδϟ΍ΔϴΑήόϟ΍ΔϜϠϤϤϟ΍
Ministry Of Health ΔΤμϟ΍Γέ΍ίϭ
General Directorate of Equipment Supplies ΔϴΒτϟ΍Ε΍ΰϴϬΠΘϟ΍Γέ΍Ω·
Department Of Specs. & Development
ήϳϮτΘϟ΍ϭΕΎϔλ΍ϮϤϟ΍Γέ΍Ω·

TECHNICAL & PERFORMANCE SPECIFICATION FOR MEDICAL EQUIPMENTS

# TECHNICAL PARAMETERS SPECIFIED RESULT


72 Patient monitoring features ECG, patient ID, patient
Demographics, Specify others if
available
73 Image storage Size, pixels x bits 1024 x 1024 x 10 or more
74 Number of images Specify
75 Number of images storage on HD >100.000 in 1024 x 1024
76 Display of X-ray & endoscopic Specify
77 Image post-processing Specify
78 Integrated Lithotripsy Database Specify
79 APPLICATIONS
80 ESWL Yes
81 ESWT Yes
82 Urological diagnosis and endourology Specify
83 Prostatic therapy Yes
84 Cystoscopy Yes
85 DICOM 3.0 Yes, full DICOM 3.0 Please
supply your DICOM
Conformance Statement
86 DICOM Print Yes
87 DICOM Send Yes
88 DICOM query/retrieve Yes
89 DICOM Worklist Yes
90 DOCUMENTATION Yes
91 CD writer or USB Key Yes
92 B&W Thermal Printer For US Doc
93 Network Connection Available
94 Laser paper printer For report documentation
95 ACCESSORIES
96 Endoscopy interface integrated in the unit Specify
97 Head wedge for patient Yes
98 Compression belt Yes
99 Patient leg knee and foot support Yes, for adult
100 Patient arm rests Yes, for both arms
101 Patient shoulder supports Yes
102 Patient wrist restraints Yes
103 Patient elbow supports Yes
104 Endoscopy shelf Specify
105 foot switches Yes, for fluoroscopy /
radiography / table movement.
106 18 TFT color monitors or x-ray diagnostic monitors suspended from the unit Yes
107 Drain pan for perineal table end Yes
108 Holder for plastic drain bag for the perineal table end Yes
109 Infusion bottle holder Yes,
110 OTHER SPECIFICATIONS
111 All access. needed
112 X-ray protection apron with wall stands. Yes, 4

16/Nov/2017 01:16 PM Page 3 of 4


Kingdom of Saudi Arabia ΔϳΩϮόδϟ΍ΔϴΑήόϟ΍ΔϜϠϤϤϟ΍
Ministry Of Health ΔΤμϟ΍Γέ΍ίϭ
General Directorate of Equipment Supplies ΔϴΒτϟ΍Ε΍ΰϴϬΠΘϟ΍Γέ΍Ω·
Department Of Specs. & Development
ήϳϮτΘϟ΍ϭΕΎϔλ΍ϮϤϟ΍Γέ΍Ω·

TECHNICAL & PERFORMANCE SPECIFICATION FOR MEDICAL EQUIPMENTS

# TECHNICAL PARAMETERS SPECIFIED RESULT


113 PRE-INSTALLATION WORK Required
114 Mild site preparation Site visit is recommended
115 WARRANTY
116 SW Coil Or Electroconductive Electrode Replacement Free of charge during Warranty
period
117 After Warranty Service Contract Not more than 7% annually of the
system value including spare
parts and to be valid for at least 3
years.
118 TRAINING Yes, Required
119 VISUAL THERAPY (Pictures should be choices by hospital administrator ) One ceiling mural approx (2-3) m
/ Two well mural approx (1.5-2) m
120 HIPPA Compliance Specify
121 5 years warranty and operational maintenance include both the update and upgrade (includes Yes
HW/SW)
122 Maintenance and Service Contract Value with Spare Parts for the First Year after Warranty Specify the Value
123 Maintenance and Service Contract Value with Spare Parts for the Second Year after Warranty Specify the Value
124 Maintenance and Service Contract Value with Spare Parts for the Third Year after Warranty Specify the Value
125 Maintenance and Service Contract Value with Spare Parts for the Fourth Year after Warranty Specify the Value
126 Maintenance and Service Contract Value with Spare Parts for the Fifth Year after Warranty Specify the Value

16/Nov/2017 01:16 PM Page 4 of 4