You are on page 1of 5

Kingdom of Saudi Arabia ΔϳΩϮόδϟ΍ΔϴΑήόϟ΍ΔϜϠϤϤϟ΍

Ministry Of Health ΔΤμϟ΍Γέ΍ίϭ


General Directorate of Equipment Supplies ΔϴΒτϟ΍Ε΍ΰϴϬΠΘϟ΍Γέ΍Ω·
Department Of Specs. & Development
ήϳϮτΘϟ΍ϭΕΎϔλ΍ϮϤϟ΍Γέ΍Ω·

TECHNICAL & PERFORMANCE SPECIFICATION FOR MEDICAL EQUIPMENTS

ΔϳΑρϟ΍Ε΍ίϳϬΟΗϟ΍
ΔόηϷ΍ϡγϗΓίϬΟ΃

Item Code 4455200020003 GROUP # MI016

ITEM DESCRIPTION MRI 3 T ϼγΗ3ϲγϳρΎϧϐϣϥϳϧέίΎϬΟ Ω˰˰ϧ˰˰Α˰˰ϟ΍ϡ˰˰˰˰˰γ΍

MODEL OFFERED MANUFACTURER COUNTRY OF ORIGIN

# TECHNICAL PARAMETERS SPECIFIED RESULT


1 MODEL:
2 MANUFACTURER/MODEL
3 CLINICAL APPLICATION Whole body
4 Field strength 3 Tesla
5 Configuration Short bore& Wide bore
6 Magnet type Superconductive
7 Radial (x.y) 0.5mT Fringe Field m < or = 3.2
8 Axial (z) 0.5m T Fringe Field m < or = 5.2
9 Magnet shimming Superconductive
10 Bore &Patient couch:
11 Aperture Width cm > or =70
12 Table Type Motorized Fixed /Dockable, (If
fixed,extra mobile MRI
combatable height adjustable
Trolley should be supplied or
sliding table top .
13 Laser lights Axial, Sagittal, Coronal reference
planes
14 Body mass limit kg >200 with movement
15 Magnet specification
16 Shielding Active
17 Cooling system Liquid Helium
18 Boil-off rate l/h
19 Helium refill >5years(refill is included in the
warranty period if needed)
20 RF system specification
21 RF Amplifier Technology specify
22 RF transmitter Power > or =35 KW
23 Number of independent RF receiver channels > or =32
24 Bandwidth of each independent RF receiver channel MHz >1
25 Gradient system
26 Shielding Active
27 Gradient system (specify name)
28 Maximum amplitude mT/m >45
29 Slew rate T/m/s >200
30 Resolution parameters
31 Minimum FOV mm < 10

16/Nov/2017 01:15 PM Page 1 of 5


Kingdom of Saudi Arabia ΔϳΩϮόδϟ΍ΔϴΑήόϟ΍ΔϜϠϤϤϟ΍
Ministry Of Health ΔΤμϟ΍Γέ΍ίϭ
General Directorate of Equipment Supplies ΔϴΒτϟ΍Ε΍ΰϴϬΠΘϟ΍Γέ΍Ω·
Department Of Specs. & Development
ήϳϮτΘϟ΍ϭΕΎϔλ΍ϮϤϟ΍Γέ΍Ω·

TECHNICAL & PERFORMANCE SPECIFICATION FOR MEDICAL EQUIPMENTS

# TECHNICAL PARAMETERS SPECIFIED RESULT


32 Imaging matrix 1024X1024
33 Minimum 2D slice thickness mm < 0.5
34 Minimum 3D slice thickness mm < 0.1
35 Sequence packages:
36 standard packages yes
37 Liver & MRCP imaging yes
38 angiography and cardiac imaging yes
39 perfusion imaging EPI
40 diffusion imaging (ADC) maps./Single & high
resolution EPI , / Diffusion Tensor
Imaging Package (DTI), Multi-
directional
41 functional imaging fMRI/ Advanced /BOLD (Blood Oxygen Level Dependent) Imaging: yes
42 fMRI stimulation device including:- Audio stimulation, -Video stimulation yes
43 spectroscopy Single Voxel Spectroscopy
(SVS), Chemical Shift Imaging
(CSI), Protocols for prostate
spectroscopy/software package
with sequences and protocols for
proton spectroscopy.
44 Main computer system ? architecture
45 Operating system Windows / Linux
46 CPU speed Intel Xeon Dual-core (2x2.6GHz)
or more
47 Image capacity256x256 images > 200,000
48 Archive drive CDR/RW &DVD or MOD
49 Main computer image display monitor > 18 LCD color (1280X1024) ,>
16 bit
50 Recon time Not slower than 12000
images/sec for 2D FFT in256 x
256 matrix
51 Radio-frequency (RF) coils Higher channels & Elementsare
preferable. (higher channels&
Elements should be submitted by
the vendor if available)
52 1.Built-in body coil T/R Q,> or =32rungs
53 2.Head coil Quadrature Head, (stander)
54 3.Head coil 32 Element (>10 channel should
be submitted if available)
55 4.Neuro Vascular Array coil > or = 13 Element(> or = 10
channels)
56 5.CTL Spine coil or equivelent > or = 24 Element
57 6.Cardiovascular coil Cardiac/ ,> or = 30 channels
58 7.Peripheral vascular coil Peripheral Vascular/,> or = 12
channels
59 8.Breast imaging coil Bilateral > or = 8 channels
60 9.Breast imaging, Biopsy& spectroscopycoil if available should be supplied
61 10.Knee coil > or =15channels
62 11.TMJ coil Paired 2channels
63 12.Shoulder coil /Large > or =16 channels
64 13.Shoulder coil /small > or = 16 channels
65 14.Lower extremity coil s/r coil

16/Nov/2017 01:15 PM Page 2 of 5


Kingdom of Saudi Arabia ΔϳΩϮόδϟ΍ΔϴΑήόϟ΍ΔϜϠϤϤϟ΍
Ministry Of Health ΔΤμϟ΍Γέ΍ίϭ
General Directorate of Equipment Supplies ΔϴΒτϟ΍Ε΍ΰϴϬΠΘϟ΍Γέ΍Ω·
Department Of Specs. & Development
ήϳϮτΘϟ΍ϭΕΎϔλ΍ϮϤϟ΍Γέ΍Ω·

TECHNICAL & PERFORMANCE SPECIFICATION FOR MEDICAL EQUIPMENTS

# TECHNICAL PARAMETERS SPECIFIED RESULT


66 15.General purpose coils Set of 6 flex of different size
67 16.Abdomen coil > or = 32 Elements
68 17.Hand / Wrist coil > or =16 channel
69 18.Foot / ankle coil > or = 16 channel
70 Other specifications
71 Automatic examination protocols for the requested clinical applications with graphic assisted yes
instructions.
72 Noise reduction hard & software measures. yes
73 SAR Reduction Technologies yes
74 Parallel imaging Yes
75 Stepping capability for peripheral Angiography. Yes
76 Standard pulse sequences, Inversion Recovery including FLAIR , T1 ? FLAIR, Fast SE 2D and 3D, Yes
Single shot FSE for MRCP, Fast Gradient Echo with Echo Planer read out, Dual Echo Fast Gradient
Echo Single shot Echo Planer Imaging with 255 ETL, Fat-Water Sepa
77 ECG gating / triggering including one set of patient lead wires, one carton of adult-size electrodes Standard with Multi-slice single-
and one fiber optic peripheral gating probe. phase, Multi-slice multi-phase,
Single-slice multi-phase
78 Peripheral trigger Yes
79 Pulse / Respiratory trigger Yes
80 Comprehensive MR Angio Fluoro Triggered MRA, Time resolved MRA with high temporal resolution Yes
and high spatial resolution.
81 Peripheral Angiography using multi-step table movements and automatic bolus chase and post- Yes
processing.
82 Cardiac Package for advance anatomical and functional imaging with appropriate gating Yes
83 Coronary Artery Imaging package. Yes
84 2D/3D Delayed Enhancement package Yes
85 Myocardial Perfusion Yes
86 Liver Imaging in short breath-hold Technique that minimize patient motion, Diffusion weighted Yes
imaging using single-shot EPI, Diffusion tensor imaging, Diffusion tractography, MRCP
87 Dynamic susceptibility contrast perfusion technique. Perfusion imaging (brain) Whole body T1, STIR Yes
and T2 imaging
88 Whole Body screening with T1sg, T1 Coronal, T2 Coronal, STIR and DWI. Yes
89 Image Motion artifact correction application Yes
90 Functional imaging Yes , all software and hardware
needed should be supplied
91 Pediatric Package yes
92 Clinical Application Packages Advanced Angiographic 2D/3D
(ToF) & (PC) and Contrast &
Non-Contrast MRA and Dynamic
4D MRA and (MTC) packages.
Advanced Automated
Angiographic evaluation,
analysis, Subtraction, (MIP),
(SSD) and reporting
packages.Advanced 2D/3D and
Real-Time Cine with
93 (Software & Hardware) ,8 concurrent licenses on a high end Rack or Blade server with a minimum of Advanced Automated anatomical
64 GB RAM for access from any workstation in the radiology department. This is other than the cardiac evaluation, analysis and
software &licenses needed for the Main console for production reporting packages. Advanced all
functional & flow, viability and
stress cardiac packages.
Advanced Automated all
functional & flow, viability and
stress cardiac evaluation,
analysis and repor

16/Nov/2017 01:15 PM Page 3 of 5


Kingdom of Saudi Arabia ΔϳΩϮόδϟ΍ΔϴΑήόϟ΍ΔϜϠϤϤϟ΍
Ministry Of Health ΔΤμϟ΍Γέ΍ίϭ
General Directorate of Equipment Supplies ΔϴΒτϟ΍Ε΍ΰϴϬΠΘϟ΍Γέ΍Ω·
Department Of Specs. & Development
ήϳϮτΘϟ΍ϭΕΎϔλ΍ϮϤϟ΍Γέ΍Ω·

TECHNICAL & PERFORMANCE SPECIFICATION FOR MEDICAL EQUIPMENTS

# TECHNICAL PARAMETERS SPECIFIED RESULT


94 DICOM CD/DVD of real pathological cases as complete study done on similar machine should be 1)Coronary, 2)Cardiac perfusion,
supplied after removing demographic patient DATA. 3)Cardiac anomalies, 4)Brain
perfusion, 5)Body perfusion, 6)
Spectroscopy, 7)Function
imaging, 8)Brain, 9)Vascular, 10)
MRCP, 11)Abdomen, 12)Pelvis,
13)Orthopedic, 14)Breast, 15)
Pediatric
95 MR Contrast Media Injector and injector accessories. 1) MRI contrast Media Injector: a
WIRELESS, battery operated,
fully digital, user programmable
double-head injector which can
be remotely controlled and used
for the administration of contrast
agents and physiological saline
solution (NaCl) inside the MRI
96 Portable Creatinine Handheld Analyzer Yes for instant creatinine
analysis. 2000 creatinine strips.
97 MRI Infusion Pump Yes 3 units
98 Drug Administration System with three syringe pumps for the application of drugs during MRI Yes, including Cart and IV stand
examinations Also can be operated with re-
chargeable battery pack
99 MRI Sphygmomanometer Yes, 2 units
100 Non-Magnetic Stethoscope Yes, 2 units
101 Non-Magnetic Laryngoscope Yes, 2 complete sets
102 MRI compatible Vital signs monitor including pulse oximeter Yes 2 units, wireless with
additional remote monitor in
control room
103 MRI Anesthesia Cart yes
104 MRI Surgical Lights yes
105 MRI compatible chair Yes, 4 units
106 Non-Magnetic Neonatal Noise Guards yes
107 Non-Magnetic Disposable Ear Plugs 5000 pairs
108 Non-Magnetic MRI Prism Glasses yes
109 MRI ECG electrodes 1000 set
110 Non Ferrous MRI Wheelchair yes
111 Non Ferrous MRI Stretcher Yes, 2 units
112 Ferromagnetic Detection Systems frame The FerrAlert Halo System or
equivalent: 1) FerrAlert Halo
PreScreen ,2) FerrAlert Halo
Entry, Unlike conventional metal
detectors which detect many
different metals, FERRALERT?
alarms ONLY upon detection of
ferromagnetic metals, and helps
to pinpoint the
113 Suitable UPS for the system Yes whole system 100 kva or
more
114 Suitable chiller Yes
115 RF shielding, Magnetroom (with RF shielding) andOxygen monitor and Exhaustion fan, Noise Full RF cabin with interior design.
(sound) reduction(for the room) Sound and noise reduction as
enclosure specify the dbi, and
percent of noise reduction.
Medical gases and suction in the
wall
116 Control room Control room window with
dimming facility, High quality
desk for the MR console,
monitors and RIS terminal, 4 high
quality chairs, 4 employee
lockers, large storage locker.

16/Nov/2017 01:15 PM Page 4 of 5


Kingdom of Saudi Arabia ΔϳΩϮόδϟ΍ΔϴΑήόϟ΍ΔϜϠϤϤϟ΍
Ministry Of Health ΔΤμϟ΍Γέ΍ίϭ
General Directorate of Equipment Supplies ΔϴΒτϟ΍Ε΍ΰϴϬΠΘϟ΍Γέ΍Ω·
Department Of Specs. & Development
ήϳϮτΘϟ΍ϭΕΎϔλ΍ϮϤϟ΍Γέ΍Ω·

TECHNICAL & PERFORMANCE SPECIFICATION FOR MEDICAL EQUIPMENTS

# TECHNICAL PARAMETERS SPECIFIED RESULT


117 Equipments room Examination room computer
controlled visual atmosphere
facility. Motorized sliding
examination room door
118 Doctor room with work station
119 2 Dressing Rooms 2 patient lockers with
nonmagnetic keys
120 Recovery area Medical gases and suction in the
wall. Large Storage lockers.
121 Access control system for the whole MRI Unit, with motorized glass door for the unit entry with RF yes
control system
122 Coil storage (high quality cart or shelves), 2 Fire extinguisher,fully equipped Crash cart with ECG
and defibrillator, 2 I.V stand, Nonmagnetic stool (2 steps).patient transportation board, (software
/hardware music system or audio system, complete MRI c
123 MR Compatible patient vital signs monitor SPECIFY with pulse oximeter
,and second slave monitor in
control room
124 MR Compatible, ANESTHESIA machine. Ventilator , vital signs monitors Suctions machine Yes, Compact anaesthesia
machines .The anaesthesia
machines must be suitable for
anaesthesia in a rebreathing
system and provided with an
anaesthetic ventilation facility for
the ventilation of paediatric and
adult patients. The anaesthesia
machines must
125 Remote Diagnostics Yes
126 Independent workstation Workstation with 2 high
resolution >18 ?LCD monitors,
with DICOM viewer with 3D
functionality with access to the
application Server. Height
adjustable high quality desk for
the workstation, 2 high quality
chairs
127 DICOM3.0 Print, Worklist, Store, Sent/
Receive / retrieve /query /storage
commitment, MPPS, Connection
to PACS and RIS free of charge
128 HIPAA security compliance Yes, protection against
unauthorized access
129 Measures for Data protection against deletion and export Yes, different privilege levels
130 Video Camera for monitoring the patient along the tunnel Yes 2 with 2 monitors in control
room
131 VISUAL THERAPY Lighting (Pictures should be choices by hospital administrator ) One ceiling mural approx (2-3) m
/ Two well mural approx (1.5-2) m
132 MRI compatible containers Yes
133 MRI compatible cleaning tools yes
134 MRI compatible hardware tools yes
135 Preinstallation work Site visit is required
136 Warranty 5 years full warranty
137 Maintenance and Service Contract Value with Spare Parts for the First Year after Warranty Specify
138 Maintenance and Service Contract Value with Spare Parts for the Second Year after Warranty Specify
139 Maintenance and Service Contract Value with Spare Parts for the Third Year after Warranty Specify
140 Maintenance and Service Contract Value with Spare Parts for the Fourth Year after Warranty Specify
141 Maintenance and Service Contract Value with Spare Parts for the Fifth Year after Warranty Specify

16/Nov/2017 01:15 PM Page 5 of 5