You are on page 1of 8

Application Bulletin 48/4 e

Check of Silver, Platinum and Gold electrodes

Branch
Check of Pt and Au
General analytical laboratories
electrodes with redox
Keywords standard solution
Branch 1; check; test; silver electrode; gold electrode, Instruments
platinum electrode; ring electrode; Titrode; rod electrode;
• pH meter
metal electrode: potential measurement; titration;
• Stirrer
bivoltammetric titration; redox standard solution; ready-to-
use redox standard; 6.2306.020 • Water bath

Summary Electrodes

It is essential to know before starting the sample analysis if Combined Pt-ring electrode 6.0451.100
the electrode is in a good state or not. A well working 6.0451.300
electrode will increase the quality of your results, as the 6.00403.300
accuracy and precision will be increased. Furthermore, Combined Au-ring electrode 6.0452.100
tedious error tracking can be omitted and no sample is
Separate Pt-ring electrode* 6.0351.100
wasted due to a defect or old electrode.
Separate Au-ring electrode* 6.0352.100
There exist several ways how to check metal electrodes, e.g.,
Pt rod electrode* 6.0331.000
measurement of redox potentials, potentiometric titration or
bivoltammetric titration. This bulletin describes the best 6.1247.010 + 6.1241.040
methods for the various by Metrohm available metal Au rod electrode* 6.1248.030 + 6.1241.040
electrodes. *for a separate metal electrode a reference electrode (e.g.,
6.0750.100) with the reference electrolyte c(KCl) = 3 mol/L must be
used.

Reagents
• Electrolyte solution c(KCl) = 3 mol/L: Metrohm No.:
6.2308.020
• Redox standard U = + 250 ± 5 mV: Metrohm No.:
6.2306.020
Or
• Ultra pure quinhydrone
• Buffer solution pH = 7.00: Metrohm No.: 6.2307.110
• and/or Buffer solution pH = 4.00: Metrohm No.
6.2307.100

Page 1 of 8
Application Bulletin 48/4 e
Check of Silver, Platinum and Gold electrodes

Solutions 25 + 266.1
Quinhydrone redox Buffer solution (pH 4.00 or 7.00) is 30 + 252.0
standard mixed with ultrapure quinhydrone pH 7.00 (6.2307.110) 5 + 102.7
until a saturated buffer solution is 15 + 90.5
obtained. Some excess of 20 + 84.6
undissolved quinhydrone must 25 + 78.6
remain. 30 + 72.2
* the potentials given here are only correct for a reference system
which is filled with c(KCl) = 3 mol/L. Otherwise the potential of the
Sample preparation reference system has to be taken into account.

No sample preparation is required.

Parameters
Analysis Mode MEAS U

Check with ready-to-use redox standard solution (e.g., Signal drift 2 mV/min
6.2306.020) Min. waiting time 0s
The electrode is checked with a ready-to-use, certified redox Max. waiting time 52 s
standard solution (e.g., Metrohm No. 6.2306.020). For
example, the Metrohm redox standard has a redox potential
Example determination
of U = +250 mV ± 5 mV at 20 °C against the reference
Potential measurement 1 - MEAS U 1
system Ag/AgCl/c(KCl) = 3 mol/L. mV 255.0 mV/min
2
To perform the measurement, the redox standard is first 254.8 0
tempered to 20 °C using a water bath. When 20 °C has been
Potential

254.6 -2
reached, the electrode to be tested is immersed in the

dU/dt
-4
solution and the potential of the solution is measured while 254.4

gently stirring. The measured value should not deviate from 254.2
-6

the certified value of the standard by more than 5 mV. For -8


254.0
example, the measured value of a certified redox standard
0 2 4 6 8 10
with a potential of 253.5 mV should range between 248.5 and Zeit s

258.5 mV. Fig. 1: Potential measurement performed with a Pt ring electrode


using a standard redox solution.
Check with self-made redox standard solution
A saturated solution of quinhydrone in acidic or neutral (but
not alkaline) buffer solution represents a well-defined redox Comments
system. • Only Pt- and Au-electrodes can be tested with the
The electrode to be tested is immersed in the solution and the redox standard solutions due to oxidation of less noble
potential of the solution is measured while gently stirring. The metals such as silver.
obtained potential is then compared to the calculated • Only combined metal ring-electrodes or separate metal
theoretical values in Table 1. The deviation of the measured electrodes can be tested with the redox standard
value to the theoretical value should not exceed 5 mV. solution as a stable potential of an Ag/AgCl/c(KCl)=
3 mol/L reference is required.
Table 1: Theoretical potential of a Pt- or Au-electrode in a buffer • If deviations from the theoretical value exceed 5 mV,
solution saturated with quinhydrone. the potential of the reference electrode used, must be
Metrohm pH buffer, T / °C U / mV* measured against a second reference electrode known
saturated with to be reliable. The reference electrode being tested is in
quinhydrone a good shape, if the measured potential is zero.
pH 4.00 (6.2307.100) 5 + 273.2 • A metal electrode that measures an incorrect potential
15 + 264.9 is either defective or contaminated in some way.
20 + 260.8 Directions for cleaning such electrodes are given in the
corresponding leaflets or monographs.

Page 2 of 8
Application Bulletin 48/4 e
Check of Silver, Platinum and Gold electrodes

Check of Au electrodes by Parameters


Mode MET U
titration Signal drift 50 mV/min
Instruments Min. waiting time 0s
• Titrator with MET mode Max. waiting time 26 s
• 10 mL burette Volume increment 0.1 mL
• Stirrer Dosing rate Maximum
Stop volume 3 mL
Stop U Off
Electrodes
Stop EP 9
Combined Au-ring electrode 6.0452.100
EP recognition All
Micro Au Titrode 6.0435.110
6.00435.120
Separate Au-ring electrode* 6.0352.100 Evaluation

Au rod electrode* 6.1248.030 + 6.1241.040 There are several parameters that are evaluated for the
*for a separate metal electrode a reference electrode (e.g., electrode performance. Titration duration and titrant
6.0750.100) with the reference electrolyte c(KCl) = 3 mol/L must be consumption until the equivalence point, reproducibility and
used. the potential jump.
The potential jump is calculated as follows using fixed point
evaluation:
Reagents
FP{1} = VEP1 x 0.9
• Sodium thiosulfate, Na2S2O3, p.a.
FP{2} = VEP1 x 1.1
• Sulfuric acid c(H2SO4) = 0.1 mol/L ∆U90-110% = FP{2} – FP{1}

Solutions VEP1: Titrant consumption of sample determination


until first equivalence point in mL
Iodine solution c(I2) = 0.05 mol/L FP{1}: Potential in mV at 90% of the consumed titrant
If possible this solution should be volume
bought from a supplier. FP{2}: Potential in mV at 110% of the consumed
titrant volume
Titrant c(Na2S2O3) = 0.1 mol/L
∆U90-110%: Potential difference between fix point 1 and 2
15.8 g sodium thiosulfate is
weighed into a 1000 mL
volumetric flask. 400 mL deion. The following criteria must be met for an electrode passing
water is added and the sodium the test:
thiosulfate is dissolved. Afterwards • Only one EP should be found per titration.
the volumetric flask is filled up to • The relative standard deviation of the three repetitions
the mark with deion. water. of the titration should be less than 0.5% (n = 3).
• The titrant consumption until the equivalence point (EP)
Sample preparation should be between 1.95 and 2.05 mL.
No sample preparation is required. • The titration duration should be no longer than 160 s.
• The potential jump ∆U90-110% should be ≥70 mV.

Analysis
A 100 mL beaker is filled with 50 mL deion. water. 1 mL
sulfuric acid is added and then 2 mL c(I2) = 0.05 mol/L is
pipetted to the solution. The sample is titrated while stirring
with c(Na2S2O3) = 0.1 mol/L until after the equivalence point.
Afterwards, the titration equipment is rinsed well with deion.
water. It is recommended to run each test in triplicate.

Page 3 of 8
Application Bulletin 48/4 e
Check of Silver, Platinum and Gold electrodes

Example determination
Check of Pt electrodes by
titration
Instruments
• Titrator with DET mode
• 10 mL burette
• Stirrer

Fig. 2: Titration curve obtained for the check of an Au ring Electrodes


electrode.
Pt Titrode 6.0431.100
6.0471.300
Comments 6.00401.300
• An electrode that measures an incorrect value is Combined Pt ring electrode 6.0451.100
defective or contaminated in some way. Directions for 6.0451.300
cleaning of such electrodes are given in the 6.00403.300
corresponding instructions for use and monographs. Micro Pt Titrode 6.0431.100
• If the titrant consumption until the equivalence point Separate Pt ring electrode* 6.0351.100
(EP) is not between 1.95 and 2.05 mL, please check
Pt rod electrode* 6.1248.000 + 6.1241.040
first the buret if it is correctly dosing.
6.0331.000
*for a separate metal electrode a reference electrode (e.g.,
6.0750.100) with the reference electrolyte c(KCl) = 3 mol/L must be
used.

Reagents
• Sodium thiosulfate, Na2S2O3
• Sulfuric acid c(H2SO4) = 0.1 mol/L

Solutions
Iodine solution c(I2) = 0.05 mol/L
If possible this solution should be
bought from a supplier.
Titrant c(Na2S2O3) = 0.1 mol/L
15.8 g sodium thiosulfate is
weighed into a 1000 mL
volumetric flask. 400 mL deion.
water is added and the sodium
thiosulfate is dissolved. Afterwards
the volumetric flask is filled up to
the mark with deion. water.

Sample preparation
No sample preparation is required.

Page 4 of 8
Application Bulletin 48/4 e
Check of Silver, Platinum and Gold electrodes

Analysis • The titration duration should be no longer than 150 s.


A 100 mL beaker is filled with 50 mL deion. water. 1 mL • The potential jump ∆U90-110% should be ≥70 mV.
sulfuric acid is added and then 2 mL c(I2) = 0.05 mol/L is
pipetted to the solution. The sample is titrated while stirring
with c(Na2S2O3) = 0.1 mol/L until after the equivalence point. Example determinations
Afterwards, the titration equipment is rinsed well with deion.
water. It is recommended to run each test in triplicate.

Parameters
Mode DET U
Signal drift 50 mV/min
Min. waiting time 0s
Max. waiting time 26 s Fig. 3: Titration curve obtained for the check of a Pt ring
Meas. point density 4 electrode.

Min. increment 10 µL
Dosing rate Maximum
Stop volume 3 mL
Stop U Off
Stop EP 9
EP recognition All

Evaluation
There are several parameters that are evaluated for the Fig. 4: Titration curve obtained for the check of a Pt Titrode.
electrode performance. Titration duration and titrant
consumption until the equivalence point, reproducibility and
Comments
the potential jump.
• An electrode that measures an incorrect value is
The potential jump is calculated as follows using fixed point
defective or contaminated in some way. Directions for
evaluation:
cleaning of such electrodes are given in the
FP{1} = VEP1 x 0.9
corresponding instructions for use and monographs.
FP{2} = VEP1 x 1.1
• If the titrant consumption until the equivalence point
∆U90-110% = FP{2} – FP{1}
(EP) is not between 1.95 and 2.05 mL, please check
first the buret if it is correctly dosing.
VEP1: Titrant consumption of sample determination
until first equivalence point in mL
FP{1}: Potential in mV at 90% of the consumed titrant
volume
FP{2}: Potential in mV at 110% of the consumed
titrant volume
∆U90-110%: Potential difference between fix point 1 and 2

The following criteria must be met for an electrode passing


the test:
• Only one EP should be found per titration.
• The relative standard deviation of the three repetitions
of the titration should be less than 0.5% (n = 3).
• The titrant consumption until the equivalence point (EP)
should be between 1.95 and 2.05 mL.

Page 5 of 8
Application Bulletin 48/4 e
Check of Silver, Platinum and Gold electrodes

Check of Ag electrodes by Analysis


A 100 mL beaker is filled with 50 mL deion. water. 2 mL
titration c(HCl) = 0.1 mol/L is pipetted to the water. The sample is
titrated while stirring with c(AgNO3) = 0.1 mol/L until after the
Instruments
equivalence point. Afterwards, the titration equipment is
• Titrator with DET mode rinsed well with deion. water. It is recommended to run each
• 10 mL burette test in triplicate.
• Stirrer

Parameters
Electrodes Mode DET U
Ag Titrode 6.0430.100(S, Br, Cl) Signal drift 50 mV/min
6.00430.100(S, Br, Cl) Min. waiting time 0s
6.0470.300(S, Br) Max. waiting time 26 s
6.00470.300(S, Br)
Meas. point density 4
6.00400.300
Min. increment 10 µL
6.00404.300
Dosing rate Maximum
Combined Ag ring electrode 6.0450.100(S, Br, Cl)
Stop volume 3 mL
6.00450.100(S, Br, Cl)
Stop U Off
6.0450.300
6.00450.300 Stop EP 9

6.00402.300 EP recognition All


Micro Ag Titrode 6.0433.100
Separate Ag ring electrode* 6.0350.100(S) Evaluation
6.00350.100(S) There are several parameters that are evaluated for the
Ag rod electrode* 6.1248.010(S) + electrode performance. Titration duration and titrant
6.1241.040 consumption until the equivalence point, reproducibility and
6.0331.010(S) the potential jump.
*for a separate metal electrode a reference electrode (e.g., The potential jump is calculated as follows using fixed point
6.0750.100) with the reference electrolyte c(KNO3) = 1 mol/L must evaluation:
be used.
FP{1} = VEP1 x 0.9
FP{2} = VEP1 x 1.1
Reagents ∆U90-110% = FP{2} – FP{1}
• Hydrochloric acid c(HCl) = 0.1 mol/L
VEP1: Titrant consumption of sample determination
until first equivalence point in mL
Solutions FP{1}: Potential in mV at 90% of the consumed titrant
volume
Silver nitrate c(AgNO3) = 0.1 mol/L
FP{2}: Potential in mV at 110% of the consumed
c(AgNO3) = 0.1 mol/L If possible, this solution should titrant volume
be bought from a supplier. ∆U90-110%: Potential difference between fix point 1 and 2

Sample preparation The following criteria must be met for an electrode passing
the test:
No sample preparation is required.
• Only one EP should be found per titration.
• The relative standard deviation of the three repetitions
of the titration should be less than 0.5% (n = 3).
• The titrant consumption until the equivalence point (EP)
should be between 1.95 and 2.05.

Page 6 of 8
Application Bulletin 48/4 e
Check of Silver, Platinum and Gold electrodes

• The titration duration should be no longer than 150 s.


Check of metal electrodes by
• The potential jump ∆U90-110% should be ≥70 mV.
bivoltammetric titration
Example determinations Instruments
• Titrator with MET Ipol mode
• 10 mL burette
• Stirrer

Electrodes
Double Pt sheet electrode 6.0309.100
Double Pt wire electrode 6.0338.100
Fig. 5: Titration curve obtained for the check of an Ag ring 6.0340.000
electrode. 6.0341.100
Double Au ring electrode 6.00353.100

Reagents
• Sodium thiosulfate, Na2S2O3
• Sulfuric acid c(H2SO4) = 0.1 mol/L

Solutions
Iodine solution c(I2) = 0.05 mol/L
Fig. 6: Titration curve obtained for the check of an Ag Titrode. If possible this solution should be
bought from a supplier.
Sodium thiosulfate c(Na2S2O3) = 0.1 mol/L
Comments
solution 15.8 g sodium thiosulfate is
• An electrode that measures an incorrect value is weighed into a 1000 mL
defective or contaminated in some way. Directions for volumetric flask. 400 mL deion.
cleaning of such electrodes are given in the water is added and the sodium
corresponding instructions for use and monographs. thiosulfate is dissolved. Afterwards
• If the titrant consumption until the equivalence point the volumetric flask is filled up to
(EP) is not between 1.95 and 2.05 mL, please check the mark with deion. water.
first the buret if it is correctly dosing.

Sample preparation
No sample preparation is required.

Analysis
A 100 mL beaker is filled with 50 mL deion. water. 1 mL
sulfuric acid is added and then 2 mL of c(I2) = 0.05 mol/L is
pipetted to the solution. The sample is titrated while stirring
with c(Na2S2O3) = 0.1 mol/L until after the equivalence point.
Afterwards, the titration equipment is rinsed well with deion.
water. It is recommended to run each test in triplicate.

Page 7 of 8
Application Bulletin 48/4 e
Check of Silver, Platinum and Gold electrodes

Parameters Example determination


Mode MET Ipol
I(pol) 1.0 µA
Signal drift 50 mV/min
Min. waiting time 0s
Max. waiting time 26 s
Volume increment 0.1 mL
Dosing rate Maximum
Stop volume 3 mL Fig. 7: Titration curve obtained for the check of a Pt wire
Stop U Off electrode.
Stop EP 9
EP recognition All
Comments
• An electrode that measures an incorrect value is
Evaluation defective or contaminated in some way. Directions for
There are several parameters that are evaluated for the cleaning of such electrodes are given in the
electrode performance. Titration duration and titrant corresponding instructions for use and monographs.
consumption until the equivalence point, reproducibility and • If the titrant consumption until the equivalence point
the potential jump. (EP) is not between 1.95 and 2.05 mL, please check
The potential jump is calculated as follows using fixed point first the buret if it is correctly dosing.
evaluation:
FP{1} = VEP1 x 0.9
References
FP{2} = VEP1 x 1.1
∆U90-110% = FP{2} – FP{1} • Leaflet for metal electrodes (8.109.8047)
• Monograph: Electrodes in Potentiometry (8.105.5013)
VEP1: Titrant consumption of sample determination • Monograph: Practical Titration (8.029.5003)
until first equivalence point in mL
• Metrohm Application Bulletin AB-025
FP{1}: Potential in mV at 90% of the consumed titrant
volume Coatings on silver electrode
FP{2}: Potential in mV at 110% of the consumed
titrant volume
Date
∆U90-110%: Potential difference between fix point 1 and 2
December 2018
The following criteria must be met for an electrode passing
the test: Author
• Only one EP should be found per titration. Competence Center Titration
• The relative standard deviation of the three repetitions Metrohm International Headquarters
of the titration should be less than 0.5% (n = 3).
• The titrant consumption until the equivalence point (EP)
should be between 1.90 and 2.10.
• The titration duration should be no longer than 200 s.
• The potential jump ∆U90-110% should be ≥100 mV.

Page 8 of 8

You might also like