Has electrical safety

1 year

Mechanical inspection (96102) – Check that the displays and all optional equipment are clean.

Mechanical inspection (96102) – Check that the displays are securely mounted, if applicable.

Mechanical inspection (96102) – Check that the equipment cases are not damaged.

Mechanical inspection (96102) – Check that there are no loose or damaged cables.

Mechanical inspection (96102) – Check that the equipment is kept in a cool and dry location.

Mechanical inspection (96102) – Check that the Xhibit Central chassis fan vents are not obstructed.

Functional tests (96102) – Verify that the Action Bar (at the bottom of the screen) shows the correct
time and date.

Functional tests (96102) – Verify that all displays have the correct resolution (1680 x 1050).

Functional tests (96102) – Verify that the touchscreen response and all other UI controls function
smoothly.

Functional tests (96102) – Verify that the number of Total Zone Licenses in the Display Settings—
Configuration menu matches the total number of patient zone licenses purchased.

Functional tests (96102) – Verify that appropriate data shows for each active device assigned to a
patient zone.

Functional tests (96102) – Verify that in the Clinical—Audio Settings menu, the audio is properly
configured and audible for the reminder tone and all alarm priorities.

Functional tests (96102) – (If designated as a print server on the bedside monitor network) Verify that
the Print Server check box is selected in the Biomed—Network Setup menu.

Functional tests (96102) – (If configured for PDF printing, select the Verify button in the Biomed—
Report Setup menu) Verify connectivity to the network share PDF directory.

Functional tests (96102) – Verify that the IP addresses (set in the Biomed—Network Setup menu) for the
bedside monitor (UVSL) and telemetry (LAN2) networks are correct.

Functional tests (96102) – Verify the telemetry Farm Name (set in the Biomed—Network Setup menu) is
correct.

Functional tests (96102) – (If designated as the time master on the bedside monitor network) Verify that
the Time Master check box is selected in the Biomed—Network Setup menu.

Functional tests (96102) – (If set up with ICS) Verify that there is Clinical Access at the Xhibit Central
Station.
Functional tests (96102) – (If networked to an Xhibit Telemetry Receiver) Verify that tuning to telemetry
channel 1234 runs a telemetry demo.

Mechanical inspection (96501) – Check that the all-in-computer and all optional equipment are clean.

Mechanical inspection (96501) – Check that the equipment is securely mounted, if applicable.

Mechanical inspection (96501) – Check that the equipment case is not damaged.

Mechanical inspection (96501) – Check that there are no loose or damaged cables.

Mechanical inspection (96501) – Check that the equipment is kept in a cool and dry location.

Mechanical inspection (96501) – Check that the rear vent holes are not obstructed.

Functional tests (96501) – Verify that the Action Bar (at the bottom of the screen) shows the correct
time and date.

Functional tests (96501) – Verify that the display has the correct resolution (1680 x 1050).

Functional tests (96501) – Verify that the touchscreen response and all other UI controls function
smoothly.

Functional tests (96501) – Verify that the number of Total Zone Licenses in the Display Settings
Configuration menu is 4.

Functional tests (96501) – Verify that appropriate data shows for each active device assigned to a
patient zone.

Functional tests (96501) – Verify that in the Clinical—Audio Settings menu, the audio is properly
configured and audible for the reminder tone and all alarm priorities.

Functional tests (96501) – (If designated as a print server on the bedside monitor network) Verify that
the Print Server check box is selected in the Biomed—Network Setup menu.

Functional tests (96501) – (If configured for PDF printing, select the Verify button in the Biomed Report
Setup menu) Verify connectivity to the network share PDF directory.

Functional tests (96501) – Verify that the IP addresses (set in the Biomed—Network Setup menu) for the
bedside monitor (UVSL) and telemetry (LAN2) networks are correct.

Functional tests (96501) – Verify that the telemetry Farm Name (set in the Biomed—Network Setup
menu) is correct.

Functional tests (96501) – (If designated as the time master on the bedside monitor network) Verify that
the Time Master check box is selected in the Biomed—Network Setup menu.

Functional tests (96501) – (If networked to an Xhibit Telemetry Receiver) Verify that tuning to telemetry
channel 1234 runs a telemetry demo.