Has electrical safety

2 years

Temperature functions and associated probes calibration – Check that the accuracy is ±0.1 °C (±0.2 °F).

Perform the non-invasive blood pressure functionality calibration.

Visual inspection – Make sure that the housing is not cracked or broken and there are no signs of spilled
liquids or damage.

Visual inspection – Inspect all accessories (for example, sensors and cables).

Visual inspection – Make sure the backlight of the monitor is bright enough.

Visual inspection – Examine all system cables and power plugs for signs of damage.

Visual inspection – Inspect all patient cables, leads and strain reliefs for general condition. Make sure the
connectors are properly engaged at each end.

Scope of inspection/safety checks for the M540 – Verify the LEDs.

Scope of inspection/safety checks for the M540 – Perform a functional test of the internal battery.

Scope of inspection/safety checks for the M540 – Perform system tests (for example, communication
with the IACS, front buttons, alarm bar, and functional integrity of monitored parameters).

Scope of inspection/safety checks for the M540 – Check that the device combination is in good
condition: All labels are complete and legible.

Scope of inspection/safety checks for the M540 – Check that the device combination is in good
condition: There is no visible damage.

Scope of inspection/safety checks for the M540 – Check that the device combination is in good
condition: Fuses which are accessible from the outside are in compliance with the specified values.

Scope of inspection/safety checks for the M540 – Check the following safety features: The power LED
and the battery indicator LED function properly.

Scope of inspection/safety checks for the M540 – Check the following safety features: The C500/C700
are powered correctly.

Scope of inspection/safety checks for the M540 – Check the following safety features: Check the
functional integrity of the Infinity MCable – Nurse call.

Scope of inspection/safety checks for the M540 – Check the following safety features: Functional
integrity of the “alarm off” button located on the front of the device.

Scope of inspection/safety checks for the M540 – Check the following safety features: Functional
integrity of the non-invasive blood pressure overpressure sensor (including the valves and the pump.
Scope of inspection/safety checks for the M540 – Check the following safety features: Functional
integrity of the optical and acoustic alarm signals.

Replace the battery.

Replace the two non-invasive blood pressure air inlet filters of the M540.