Professional Documents
Culture Documents
CHAPTER-6
ANALYTICAL TECHNIQUES
6.1 INTRODUCTION
Wavelength ()
It is the distance between the two adjacent crests (C-C) or
troughs (T-T) in a particular wave. It is denoted by the letter
( lam da)
C C
Units used for wave length are :
Angstrom(A0) = 10-10m = 10-8 cms
Micrometer (m) = 10-6m
Nano meter (nm) = 10-9m
T T
Frequency ()
Frequency may be defined as the number of waves which
can pass through a point in one second (i.e. number of vibrations
in unit time) of an electromagnetic radiation.
Frequency () = Velocity of light / Wavelength
= c/
Velocity of light c = 3 1010 cm/sec
Wavelength in cm
6.2 Engineering Chemistry-II
The wave length, the frequency and the wave number are
related as follows :
1
= =
c
1) Lambert’s law
2) Beer’s law (Or) Beer - Lambert’s law
Lambert’s law
Lambert’s law states that “when a beam of
monochromatic radiation is passed through a homogeneous
absorbing medium, the rate of decrease of intensity of the
radiation ‘dI’ with thickness of absorbing medium ‘dx’ is
proportional to the intensity of the incident radiation ‘I’ ”.
dI
i.e, kI ……… (1)
dx
6.4 Engineering Chemistry-II
I x
dI
0
I
k dx
0
……… (2)
I
i.e., ln k x
I0
It is expressed as
dI
kIC …… (3)
dx
I x
dI
I0
I
k C dx
0
I
ln kC x
I0
I
(or) 2.303 log I k C x
0
I k
log 0 Cx
I 2.303
k
Where , = molar absorptivity coefficient
2.303
I0
log A Absorbance (or) optical density
I
As C s x
As
x ………. (1)
Cs
We know that
Au C u x
Au
x ……..….(2)
Cu
Analytical Techniques 6.7
Au
Cu Cs ………. (3)
As
We already know the concentration of standard solution
Cs and the absorbance Au and As were experimentally measured
and hence the unknown concentration Cu can be calculated from
the equation (3).
Solved Problems :
Solution
Transmittance T = 0.30
Absorbance is
1
A = log (or) log T
T
= – log 0.30
= 0.5229
6.8 Engineering Chemistry-II
Solution :
Given : % T = 25% (or) 0.25
C = 2.5 10-4M ; x = 1
i) Absorbance :
Absorbance, A log T
= –log 0.25
= 0.6021
Solution :
Given : A = 0.23 ; C = 1.2 10-4 ; x = 3.0 cm
We know that,
A
Molar absorptivity,
Cx
0.23
=
1.2 10 4 3.0
= 638.89 dm3 mol-1 cm-1.
Analytical Techniques 6.9
Solution :
Given : T = 20% (or) 0.20 ; x = 2.5 ;
12,000 dm3 mol-1 cm-1
A log T log 0.20
0.699
We know that,
A Cx
A 0.699
C = 12,000 2.5
x
= 2.33 10 5 mol 1 dm 3 .
Example 6 :
6.10 Engineering Chemistry-II
Solution:
I 1
Given : C = 0.06 M ; x = 8 cm I 4 , ?
0
I
log Cx
I0
log I I 0 log 1 4
Cx 0.06 8
( 0.6020)
0.06 8
= 1.254 dm3 mol-1 cm-1
________________________________________________
Atomic Spectroscopy
Interaction of electromagnetic radiation with atoms which are
most commonly in their lowest energy state (ground state)
resulting in electronic transitions is called atomic spectroscopy.
The spectrum obtained is a line spectrum.
6.12 Engineering Chemistry-II
Molecular spectroscopy
Interaction of electromagnetic radiation with molecules is
called Molecular spectroscopy. This may result in transitions
between rotational, vibrational energy levels and electronic
transitions. The spectrum obtained is a complicated spectrum.
Absorption spectroscopy
If electromagnetic radiations of certain wavelength range
are passed through the substance under analysis, radiations of
certain wavelengths are absorbed by the substance. The study of
this is called the absorption spectroscopy. The wavelength
absorbed characterizes some specific functional group of the
compound or the compound itself.
Higher energy level
h
Ground state
Absorption
Analytical Techniques 6.13
h
Ground state
Emission
Emission spectroscopy
If electromagnetic radiation is passed through a substance
or thermal energy is given to the substance under analysis, the
energy is absorbed by the atom. The electrons in the ground state
get excited to higher energy metastable states. These excited
electrons are short lived. So they emit energy to return to the
stable state. The study of this is called the emission spectroscopy.
The spectrum obtained is called the emission spectrum.
Fluorescence
The electron in the excited level return to it’s ground state
either directly or in steps with the emission of certain amount of
energy. When this emission of light is instantaneous the
phenomenon is known as fluorescence
Phosphorescence
When the electron in the excited level return to it’s
ground state with the emission of light after some time lag, it is
known as phosphorescence
Photochemical reaction
When the absorbed energy is stored by the atom or
molecule and used in producing some chemical reaction, the
resulting chemical reaction is called photochemical reaction.
6.14 Engineering Chemistry-II
6.4 CALORIMETRY
Introduction
Colorimetry is concerned with the visible region (400-
750nm) of the spectrum. The instrument, used for measuring
absorption of radiant energy in the visible region from the
substances is colorimeter.
Principle
Colorimetry is concerned with the determination of the
concentration of a substance by measurement of the relative
absorption of light in the visible region with respect to a known
concentration of substance. The intensity of the colour can be
easily measured by using a photoelectric colorimeter, from which
the concentration of coloured solution can be obtained by Beer –
Lamberts law.
I
A log T log 0 Cx
It
Current Light transmitted (1/ Concentration of solution)
Analytical Techniques 6.15
Working
In a colorimeter, a beam of light is directed through a
transparent cell containing a solution of the compound being
analysed. The compound absorbs light at a particular wavelength.
The unabsorbed radiation (transmitted light) is passed through the
detector where the light signals are converted into the electrical
signals which are recorded. The output graph obtained is a plot of
the absorbance of the light versus concentration.
A = log(I0/I)
Components
Radiation sources
Tungsten – filament lamp
Function:
Provides the wavelength range of the visible region
(400-750nm).
Sample chamber
Cuvet made of fused glass
Function:
A transparent cell made of fused glass holds the test
sample
Slits - Function:
Filter
Monochromator
6.16 Engineering Chemistry-II
Function:
Allows the light of the required wavelength to pass
through, but absorbs the light of other wavelengths.
Detector
Photosensitive devices are used to detect the radiations.
Function:
Converts the radiation signals received from the sample
into the electrical signals.
These detectors are capable of producing current, which is
directly proportional to the intensity of the radiation received.
Thus the radiant energy transmitted through the sample is
measured.
Recorder
Digital display
Function:
It displays the amount of radiation absorbed by the
sample
AMPLIFIER
RECORDER
Analytical Techniques 6.17
DETECTOR
Photo multiplier
tube
AMPLIFIER
RECORDER
Digital display
DETECTOR
Converts the light
signals into electrical
signals
AMPLIFIER
RECORDER
Displays the result
6.18 Engineering Chemistry-II
Absorbance, A = C x
Since, the path length (x) is fixed for a given cell, the absorbance
(A) is directly proportional to concentration (C). Similarly, to the
test solution same amount of NH4OH is added and the absorbance
of the test solution (unknown copper sulphate) is measured using
the same colorimeter. From the calibration curve, the
concentration of the unknown copper sulphate solution can be
evaluated.
Absorbance of
Unknown solution
Absorbance
Concentration of
unknown solution
Concentration
Calibration curve
Analytical Techniques 6.19
Absorbance, A = C x
Since, the path length (x) is fixed for a given cell, the
absorbance (A) is directly proportional to concentration (C).
Similarly, to the Fe3+ test solution, the same amount of thio
cyanate is added and the absorbance of the test solution is
measured using the same colorimeter. From the calibration curve,
the concentration of the unknown Fe3+ solution can be evaluated.
Absorbance of
Unknown solution
Absorbance
Concentration of
unknown solution
Concentration
Calibration curve
6.20 Engineering Chemistry-II
Applications of colorimetry
To determine the molar compositions of complexes.
To measure the dissociation constants of an acid-base
indicator.
To measure the molecular weight of a compound.
To determine the instability constants of complexes.
To determine the structure of inorganic compounds and
complexes.
Introduction
UV consists of broad absorption bands which have greater
utility for quantitative analysis. When polychromatic UV and
Visible radiation is passed through a medium containing
particles, they interact with the species and electronic transitions
occur by the excitation of electrons from the ground state to the
excited state. UV and visible radiation have sufficient energy to
ensure transitions of the outermost or bonding electrons only.
Principle
When UV-Visible radiations (wavelength range – 200-
800nm) are absorbed by atom / molecules, the transition of
valence electrons from lower electronic energy level to higher
electronic energy level occurs. The amount of energy required for
the transition depends on the difference in energy levels of
ground state and excited state.
E1 – E0 = h.
* * n *
* * n *
I0
A = log
I
Components :
The modern ultraviolet-visible spectrometers consist of
light source, sample chamber, Motor light attenuator, filter,
detector, amplifier and the recording devices.
Source of light :
>375 nm : Tungsten Filament lamp
< 375 nm : Hydrogen discharge lamp
Deuterium discharge lamp
Xenon discharge lamp
Analytical Techniques 6.23
Mercury arc
Function
Provides the wavelength region corresponding to UV
radiation.
Tungsten filament lamp is particularly rich in red
radiations,
Chopper
Circular disc with alternate quadrants removed.
Function
Rotates and transmits sample and reference beam
alternatively to grating
Filters :
Monochromators - It consists of an entrance slit, a
dispersing element (prism or grating) and an exit slit.
Grating made of quartz (OR) fused silica,
Function
Sends individual required frequencies to detector
Detectors :
Barrier layer cell
Photomultiplier tube
Photocell
Function
The detector converts the radiation signals into
current. The current is directly proportional to the
concentration of the solution.
6.24 Engineering Chemistry-II
LIGHT CHOPPER
SOURCE
REFERENCE
FILTER
DETECTOR
RECORDER
Analytical Techniques 6.25
Applications of UV Spectroscopy
Ultraviolet spectroscopy has been mainly applied for the
detection of functional groups (chromophore), the extent of
conjugation, detection of polynuclear compounds by comparison
etc. Some important applications of UV spectroscopy are as
follows :
2. Extent of conjugation :
The extent of conjugation in polyenes can be estimated.
6. Detection of impurities:
UV absorption spectroscopy is used for detecting impurities
in organic compounds, because
i) Bands due to impurities are very intense.
ii) Saturated compounds have little absorption band and
unsaturated compounds have strong absorption band.
I0
A log T log Cx
It
where,
= molar extinction coefficient
Analytical Techniques 6.27
C = concentration
x = length of the cell
Then the graph is plotted between absorbance vs concentration.
Thus we get the calibration curve in the form of a straight line.
From the graph, the concentration of unknown substance is found
out.
Introduction
The atoms in a molecule do not remain in fixed relative
positions but vibrate about some mean position. This fact is used
in the IR spectroscopy. The infrared region in the electromagnetic
spectrum extends from the red end of the visible spectrum to the
microwave region. The region includes radiation at wavelengths
between 0.7 and 500 —m or, in wave numbers, between 14000
and 20 cm-1. The spectral region is split into three regions
Principle
The IR spectra are given by diatomic molecules with
permanent dipole moments, i.e. heteronuclear atomic molecules
and polyatomic molecules. When IR radiations (wavelength
range –14000 – 20cm-1) are absorbed by atom / molecules, the
atoms/ molecules vibrate and undergo transition between
vibrational levels. The transitions in the vibrational energy levels
are also accompanied by transitions in the rotational energy
levels. (So IR is also called as Rotational – Vibrational spectra).
These results in the absorption bands that is characteristic to the
atoms / molecules.
Stretching vibration
The distance between two atoms changes, but the bond
angle is not altered.
Symmetric stretching
Either the movement of atoms occurs towards one
another or away from one another
Asymmetric stretching
One atom approaches the central atom while
the other departs from it .
Bending vibrations
The bond angle changes and the bond distance do not
change.
6.30 Engineering Chemistry-II
Working
In a IR spectrometer, a beam of light is split into two
halves. One half of the beam (the sample beam) is directed
through a transparent cell containing the compound being
analysed and one-half (the reference beam) is directed through an
identical cell that contains the solvent or blank. The intensities of
the two beams at each wavelength of the region is compared. As
the compound absorbs light at a particular wavelength, then
intensity of the sample beam will be less than that of the
reference beam. The motor light attenuator drives an optical edge
into the reference beam until the detector receives light of equal
intensity from sample and reference beam. The detector converts
IR radiation into electrical signals and sends it to recorder. The
output graph obtained is a plot of the wavelength of the entire
region versus the absorbance of the light at each wavelength.
A = log(I0/I)
The graph obtained is called as absorption spectrum
Components :
Analytical Techniques 6.31
Source of light :
Near IR - Tungsten filament
Mid IR - Nernst filament
– made of rare earth oxides
- Globar filament
– made of carborundum
Far IR - High pressure mercury arc lamp
Function
Provides the wavelength region corresponding to IR
radiation.
Optical systems
It is made of mineral salts (NaCl and KBr) transparent to
IR radiations
Function
The beam is guided and focused by mirrors silvered on
their surfaces.
Function
Used to hold the sample and reference materials.
Chopper
Circular disc with alternate quadrants removed.
Function
Rotates and transmits sample and reference beam
alternatively to grating
6.34 Engineering Chemistry-II
Filters :
Monochromators - It consists of an entrance slit, a dispersing
element (prism or grating) and an exit slit.
IR region – NaCl prism
Far IR region – KBr prism
Grating made of quartz or fused silica.
Function
Sends individual required frequencies to detector
Detectors :
Thermocouple (EMF)
Bolometer (ohms)
Golay pneumatic cell
Photo conductivity cell
Function
The detector converts the radiation signals into
current. The current is directly proportional to the concentration
of the solution.
Photometer
It is an instrument that furnishes the ratio or some
function of the ratio, of the radiant power of two electromagnetic
beams.
Spectrometer
An instrument with an entrance slit, a dispersing device
and one more exit slit with which the measurements are made at
selected wavelengths within the spectral range or by scanning
over the range. The quantity detected is a function of the radiant
power.
Spectrophotometer
It is a spectrometer with associated equipment, so that it
furnishes the ratio, or a function of the ratio, of the radiant power
of the two beams as a function of spectral wavelength. These two
beams are separated in time, space or both.
Principle
When a solution having a mixture of metallic species is
introduced into the flame, the solvent evaporates and vapours of
metallic species are obtained followed by the atomization. Some
of the metal atoms are excited sufficiently high to emit
characteristic radiation of metal i.e – a phenomenon used in
flame emission spectrometry.
COMPONENTS
NEBULISATION
Pneumatic nebulisers.
Function
Filaments of liquid are drawn out from the bulk of the
solution. Liquid sample is drawn through the capillary by the
pressure differential generated by the high velocity gas stream
which passes over the sample orifice. This liquid is set into
oscillation. The filaments collapse to form droplets. The cloud of
droplets strikes an obstruction in the spray chamber termed a
splinter or impact bead. This breaks the larger droplets into
smaller ones. The aerosol (a very fine fog) is mixed with the
oxidant - fuel mixture and carried into the burner. Droplets larger
than about 20 microm are trapped in the spray chamber and flow
to waste.
FLAME
BURNER
Turbulent flow or total consumption burner,
Laminar Premix burner
Non-flame atomizer.
Turbulent flow or total consumption burner:
Here the nebuliser and the burner are combined into a
single unit. The sample is drawn up the capillary and the fuel and
Analytical Techniques 6.37
Function
As soon as the aerosol produced by nebulisation in the spray
chamber is transported into the flame the following sequence of
events occurs in rapid succession.
FUEL GAS
Natural gas, Propane, Butane, Acetylene or Hydrogen
could be used.
OXIDANT
Air, Oxygen or Nitrous oxide could be used.
CHOPPER
Circular disc with alternate quadrants removed
Function
The intensity of the source is made to fluctuate at a
constant frequency. This process is called modulation. Thus
rotation of the disc at constant speed provides a beam that is
chopped to the desired frequency, thus modulating the radiation
from the source.
FILTER
Gratings are used.
Function
Minimize the flame background emission and resolve the
atomic emission lines from nearby lines and molecular fine
structure. Thus provides a narrow band width and isolate lines
that interfere with analysis.
SLITS
Adjustable slits
Analytical Techniques 6.39
6.40 Engineering Chemistry-II
INTRODUCTION
This technique involves the study of the absorption and
radiation by neutral atoms in the gaseous state. Thus in the AAS,
when a solution having a mixture of metallic species is
introduced into the flame, the solvent evaporates and vapours of
metallic species are obtained and then the absorption of radiation
by the atomic vapor is measured at a selected wavelength which
is characteristic of each individual element. The technique is also
called as absorption flame photometry, because all the analyte
applications of atomic absorption involve spraying the solution of
the sample into a flame. In AAS the unexcited atoms serve as
the basis for atomic absorption measurements.
Analytical Techniques 6.41
Principle
When a solution having a mixture of metallic species is
introduced into the flame, the solvent evaporates and vapours of
metallic species are obtained, followed by the atomization of the
vapor and then the absorption of radiation by the unexcited
atoms in the atomic vapor is measured at a selected wavelength
which is characteristic of each individual element. The flame
gases are treated as a medium containing free, unexcited atoms
capable of absorbing radiation from an external source. This
occurs when the radiation corresponds exactly to the energy
required for a transition of test element from ground electron
state to the excited state. The unabsorbed radiation passes
through a monochromator that isolates the exciting spectral lines
and into a photodetector. Absorption is measured by the
difference in transition signal in absence and presence of test
element.
I
log cx
I0
= absorption coefficient
x = path length
I = transmitted intensity
I0 = Incident intensity
c = concentration
RADIATION SOURCE
Hollow cathode lamp with tungsten anode and cylindrical
analyte metal cathode in a glass tube with filler gas as
Ne or Ar, is used
Function
When a potential is applied across the electrodes
ionisation of the gas occurs and the current flows as the ions
migrate to the electrodes. If potential is large the gaseous cation
acquire enough kinetic energy to dislodge some of the metal
atoms from the cathode surface and produce an atomic cloud
6.42 Engineering Chemistry-II
CHOPPER
Circular disc with alternate quadrants removed
Function
The intensity of the source is made to fluctuate at a
constant frequency. This process is called modulation. Thus
rotation of the disc at constant speed provides abeam that is
chopped to the desired frequency, thus modulating the radiation
from the source.
FUEL GAS
Natural gas, Propane, Butane, Acetylene or Hydrogen
could be used.
OXIDANT
Air, Oxygen or Nitrous oxide could be used.
BURNER
Turbulent flow or total consumption burner,
Laminar Premix burner
Non-flame atomizer is used.
Turbulent flow or total consumption burner:
Here the nebuliser and the burner are combined into a
single unit. The sample is drawn up the capillary and the fuel and
Analytical Techniques 6.43
6.44 Engineering Chemistry-II
FILTER
Glass filters can generally be used
Filters similar to the UV-Visible Monochromators could
be used.
Function
Provides a narrow band width and isolate lines that
interfere with analysis.