RMC-127C2

0
Content

1. Cautions ............................................................................ 3

2. Configuration .................................................................... 5

3. Device ............................................................................... 6

4. Installation......................................................................... 7

5. Operation .......................................................................... 8

6. Maintenance...................................................................... 9

7. Trouble Shooting ............................................................... 10

8. Specification ...................................................................... 11

9. Warranty ............................................................................ 12

2
1
Cautions
Please read this manual before using the RMC-127C2 oxygen concentrator.
Important information is highlighted throughout this manual with the following terms:

Notification of injuries and possible safety accidents


Warning
Information to prevent continuous damage to the product and
to urge the user's attention.
Caution

Additional important information regarding the device or procedure


Info.

◦ Use with the indicated product rating.


◦ Never open the outer case by anyone other than service personnel.
Contact with internal voltage can be very dangerous.
◦ Before operation, check the power and ground work.
◦ The oxygen concentrator inlet must be installed at least 20 seconds away from
the wall.
◦ Do not use devices that emit fire or harmful gases around the oxygen concentrator.
◦ Please clean the area around the oxygen concentrator and replace the intake filter
regularly. If the intake filter is clogged, it may shorten the life of the product or
cause a malfunction.
◦ In case of abnormal sound or other abnormal symptoms, do not disassemble
the product, turn off the power immediately and contact the place of purchase.
◦ Do not use near ignitable substances or open flames. Do not smoke nearby
when the oxygen concentrator is running. Do not use within 2 m of hot flammable
substances or exposed flame sources.
◦ Oxygen plays a role in promoting fire and spreading. Do not allow oxygen to be
supplied near bed covers and chair cushions.
◦ Do not immerse or leave the device in fluid, and do not expose it to dusty
conditions. Also, keep away from lint or other flammable materials that could
block the intake vents around the device.
◦ Do not use if there are signs of damage to the device or accessories.
◦ Do not install the device on an inclined floor.
◦ Do not use lubricants that have not been verified by the manufacturer on
devices or accessories.
3
◦ When not in use, the device should be kept in a clean, dry, and place where the
temperature is maintained between -30℃ and 60℃.
◦ Do not disassemble or modify devices or accessories other than those approved
by the manufacturer.
◦ Use only spare parts recommended by the manufacturer for proper functioning
of the device and to avoid the risk of fire and burns.
◦ Keep away from heat sources (stoves, heaters, etc.) that can raise the
temperature around the device to 40℃ or higher.
◦ Only accessories approved by the manufacturer should be used for the device.
The use of unapproved accessories may reduce the performance of the device.
◦ Do not place heavy objects on the device.
◦ Make sure that the intake and exhaust ports are not blocked while the machine
is operating. If the intake vents or exhaust vents are blocked, it may cause
internal overheating, downtime, or damage to the machine.
◦ Keep the device and its accessories out of reach of children, pets, and pests to
prevent damage or unintentional configuration changes.
◦ Be careful when touching the device in a high ambient temperature.
◦ Do not use if there are signs of damage to the device or accessories.
◦ There is a risk of tripping over the wires and piping or being wrapped around
the neck, and there is a risk of electric shock if damaged. Keep children and
pets out of reach.
◦ Always turn off the power before cleaning the device.
◦ The device concentrates and supplies oxygen from the surrounding air.
Therefore, if an oxygen concentrator is installed indoors, the indoor oxygen may
be unchanged or less.
◦ About 10 minutes or more after starting the operation of the machine is the
time to show the original performance of the machine.
◦ Oxygen concentration depends on the ventilation rate, occupants, and space
volume, depending on the ascent rate and maximum reach.
◦ If the device is used at high altitudes, the oxygen supply flow and concentration
performance may be negatively affected.
◦ If you feel sick or uncomfortable while using the device, stop using it and
consult a doctor or seek medical assistance.

This manual provides information on the RMC-127C2 oxygen


concentrator. Please read and understand the manual before using
the product. Important safety information is displayed throughout
the manual. Be sure to read all safety information. If you have any
questions, please contact your supplier.

4
2
Configuration

The product package of the oxygen concentrator RMC-127C2 includes accessories and
user manuals to improve usability. Please ask your supplier for a full list of available
accessories. Before using the device, you should check the device and its accessories for
damage.

Even if the box or packaging shows signs of damage, such as slight wear or dents, the
device can still be used without problems. However, if you see any signs of damage
to the device or accessory, you should contact the supplier for further information.
Info.

Products and Accessories List

◦ Oxygen Concentrator
◦ Humidifier Box (including 10Lpm Flow meter)
◦ 6mm tube
◦ Nasal Cannula

Flow Meter Cannula Port


(10Lpm)
There is a magnet on
the back of the box, so it
can be attached to the
device.
(Refer to 8 page)

Humidifier Box

5
3
Device

Oxygen Concentrator



① Mobile handle
Handle for lifting the product when moving the product
② Operation button
Button to operate the device
③ Intake door
Doors for service at the same time as the air inlet opening to the product
④ Humidifier Box
⑤ Rubber feet
Rubber feet for stably standing the product

6
4
Installation
When installing the oxygen generator, there are some things to consider about the power
supply and setting of the installation location.

Please refer to “1. Cautions”. Also, for product protection and safety from accidents, power
must be supplied through a circuit breaker.

Be sure to read the user manual and set up the device.


Warning
To prevent accidents such as electric shock.

Power connection

Connect the device's power cord to a 230V power source.


Warning

Oxygen connection

A tube with an outer diameter of 6 mm is connected at the device.


The tube connects to the flow meter and humidifier.
The cannula is used by connecting it to the humidifier.

7
5
Operation

The power cord must be connected to the power source prior to operation of the device.
The device is activated by pressing the operation button on the right side of device.
(Refer to 6page, Operation Button)

How to use Oxygen


Connect the 6mm tube enclosed as an accessory to the tube fitting at the bottom of the
oxygen concentrator.
Connect the 6mm tube to the right side of the flow meter Box, which is an enclosed
accessory.
Connects the nasal cannula to the humidifier.

Nasal Cannula
6mm tube

8
6
Maintenance

When washing the device, be sure to perform it while the


Warning
power of the device is turned off.

Contamination on the exterior of the device is not related to the occurrence of device
failure. However, dust in the surrounding space accelerates the filter replacement cycle
in the device, and contamination of the filter is closely related to the occurrence of
device failure. Also, maintain the cleanliness around the equipment so that the air flow
through the intake and exhaust ports is not blocked.

Filters

There are three types of filters in this device. Contact the purchaser or manufacturer
for the necessary service according to the filter cycle of each filter.

Filter Location Period Method

Dust Filter Intake Door Shake off the dust


Within 4 months
or 3000hr
Intake Filter In Intake Filter Replacement

Bacteria Filter Oxygen Discharge part every year Replacement

9
7
Trouble Shooting
In the event of a problem such as the symptoms below, check the confirmation in order
and check the confirmation. Please proceed with the action.

Symptom Check list Action


Immediately turn off the power to the
1.Does leakage current
device and contact the supplier or
occur?
manufacturer in charge.
Secure the intake and exhaust vents.
2. Are the intake or
For safety, stop the operation of the
exhaust ports of the
device for at least 30 minutes and then
device blocked?
Abnormal noise restart it.
is generated from For safety, stop the operation of the
the device. 3. Does the device
device for at least 30 minutes and
generate any specific
restart it. If symptoms recur, contact
heat?
the supplier or manufacturer in charge.
After stopping the operation of the
4. Are there any special device, restart it. If the increase is
signs found? repeated, please contact the supplier
or manufacturer in charge.
Secure the intake and exhaust vents.
1. Are the intake or
For safety, stop the operation of the
exhaust ports of the
device for at least 30 minutes and then
device blocked?
restart it.
Specifically
2. Is there a problem
Oxygen supply Please contact the supplier in charge
with the oxygen supply
Decreased. and inquire about reconstruction.
piping?
3. Do you think it is a
Please contact the supplier in charge
problem inside the
and inquire about the symptoms.
oxygen concentrator?

10
8
Specification

Oxygen Concentrator
Model: RMC-127C2

No. Item Specification

1 Discharged Oxygen Flow 8 ± 1 Lpm

2 Oxygen Purity 93 ± 3 %

3 Max. Discharged Pressure 1.2 bar

4 Input Voltage 230 VAC / 50 Hz

5 Power Consumption 450 W

6 Device Weight 37 Kg

7 Dimension (W x D x H) 579 x 265 x 593 mm

8 Noise 45 dB(A)

9 Operation Temperature 0 ~ 40 °C

10 Operation Humidity 15 ~ 95 %

The device is set to 12Lpm and is used by setting the prescribed flow rate with the
flow meter, which is a component.
The oxygen concentration at 12Lpm is about 70%.
11
Device Warranty

Name Oxygen Concentrator Model RAM-127C2

Purchase Date Serial No.

Seller Remarks

▣ When requesting a repair, please keep it safe as you will receive sufficient service only

after presenting this warranty with the date of purchase

▣ In the event of a device failure, please contact the marketing center where the device is

installed, and if you can't, please contact the customer support center of the

headquarters.

▣ Compensation and content notification will be resolved within 7 days of the requested

date and damage compensation within 14 days of the notification date.

▣ The warranty period for this device is one year or 7,000 hours from the date of

purchase, and the retention period is five years. According to the consumer damage

compensation regulations for each item, we guarantee the device as follows.