Professional Documents
Culture Documents
PICO
pH / Conductivity Meter (5P)
(with pH Data Logging & TDS – Optional)
Page 1 of 76
CONTENTS
Chapter 1 Safety
Safety Practices and Hazards 3
Electrical Hazards 4
Symbols 5
General hazards 6
Chapter 2 Introduction 7
pH Measurement Theory 8
Conductivity Measurement Theory 9
Features 12
Specification 13
Chapter 3 Setting Up
Unpacking 14
Installation 15
Panel details –
1) Front panel details 17
2) Key Functions 18
3) Back Panel Details 19
4) Electrode Stand and Holder Assembly 19
Chapter 4 Operation
Using the instrument 20
Part 1 – pH Measurement / Function Mode & Operation 22
Part 2 – Conductivity Measurement / Function Mode & 41
Operation
Chapter 5 Operation Guideline
Useful Hints for pH & mV measurement 60
Useful Hints for Conductivity & TDS 61
Chapter 6 Maintenance and Troubleshooting
Instrument Maintenance 62
Troubleshooting & Error Codes 65
Consumables and Accessories 66
Appendix A 67
Appendix B 68
Appendix C 69
Appendix D 70
Chapter 7 Service and Warranty
Terms of Warranty 75
Labindia Sales & Service Centers 76
THE CONTENTS OF THIS MANUAL ARE SUBJECT TO CHANGE WITHOUT NOTICE
Copyright 2019 LABINDIA ANALYTICAL INSTRUMENTS PVT.LTD.
Inst.Op_pH_Cond_5P_V2.23
Page 2 of 76
Chapter 1
Safety
Safety Practices and Hazards
PICO series pH, Conductivity meter has been carefully designed in accordance with the safety
standards. It is in the interest of the purchaser that he is aware of the resultant dangers of misuse
Information on safety practices is provided with your instrument and operation manuals. Before
using your instrument & accessories, you must thoroughly read these safety practices.
The following general safety precaution must be observed during all phases of operation, service
and repair of this instrument. Failure to comply with these precautions or with specific WARNIGS
elsewhere in this manual may impair the protection provided by the equipment. In addition it
violates safety standards of design, manufacture, and intended use of the instrument.
Page 3 of 76
Electrical Hazards:-
PICO pH, Conductivity meter contains electrical circuit, devices and components operate at very
Enclosure covers that are retain by fasteners, which require the use of the tool for removal may be
opened by Labindia trained, Labindia qualified or Labindia authorized service engineers. Personnel
without such training or persons currently undergoing a course of training require through
instructions. This operator’s manual should form the basis of these instructions.
Application of the wrong supply voltage connection of the instrument to an incorrectly wired supply
outlet or improper electrical grounding can create a fired hazards or a potentially serious shock
hazard and could seriously damage the instrument and any attached accessory.
DC Adapter – The meter uses detachable DC Adapter with earth connection. Failure to
ensure adequate earth grounding by not using this adapter may cause reading instability. If
the adapter becomes damaged, replace immediately with same rating and same
specifications.
Page 4 of 76
Symbols:-
Warning symbols:
Warning – Information about hazards which can be harmful to your health or lead to
instrument damage.
Warning!
This equipment contains DC Voltage of 30VDC maximum for its operation. Hence
less risk of electrical shock.
The product is marked with this symbol when it is necessary for the user to refer
to the instrument manual.
Caution Symbol:
A ‘caution’ message appears in the manual when failure to observe instructions could
result in damage to instrument.
Page 5 of 76
Use Of Instrument:-
The unit is design and built for use in laboratories. It is intended to be use to
determine the pH or Conductivity or Ionic Concentration of the liquids under
measurement and analysis. Use appropriate safety as per standards for handling of
chemicals during measurement.
Any use that does not confirm with technical data of the instrument is being to be a case of
misuse. The customer shall bear sole responsibility of any damage due to such misuse.
General Hazards:-
Page 6 of 76
Chapter 2
Introduction
LABINDIA herewith offers you the state-of-the-art micro-controller based pH / Conductivity / Ion
measurement system with user friendly dialog based easy to use operation, making error free and
precise readings. This bench-top pH/Conductivity/Ion meter is an ideal help in every laboratory for
routine measurement as well as R&D work.
The meter has Individual non volatile data memory storage block with memory scanning facility on
display, such as 40 pH Values, 40 mV Values, pH Data logging value and 40 Conductivity, TDS &
Salinity values and 40 Ion concentration values. With data logging function in pH, user can study
Kinetics of chemical reaction. The pH values recorded during Data Logging can be printed in the
graphical form with the limit marks.
The GLP Approach for Calibration and data presentation gives more authentic Reporting with its
print out on plain inexpensive paper. It has two Data report presentations –
(a) Standard : 40 data readings or (b) GLP : 40 data readings with alphanumeric Sample names
and ID numbers.
Page 7 of 76
pH Measurement Theory:
Measurement of pH of any Aqueous Sample solution gives the degree of acidity or alkalinity of the
solution. Mathematically pH is defined as the negative logarithm of the Hydrogen Ion
concentration. pH Electrode measures the pH of a Sample solution potentiometrically. When the
pH electrode is immersed in the sample solution the electrical signal gets developed across the
sensing membrane of indicator electrode and reference electrode. Output of the electrode varies
with change in pH value with a linear relation of 59.16mV / pH unit value. The electrode slope
changes with change in temperature, hence temperature compensation is required. With the PICO
pH meters this slope change can be compensated manually or automatically. For precise pH
measurement automatic temperature compensation (ATC) with separate temperature sensor is
needed.
+
PICO pH does Auto calibration with buffer recognition, measurement with ATC to read the pH
value @ 25°C. It also Stores, Protects, Scans & Prints the data with GLP concept.
Page 8 of 76
Conductivity Measurement Theory
Conductivity is the measurement of a materials ability to conduct electric current. The basic unit of
conductance "G" is the "siemens" (formerly called the Mho). Conductance is the reciprocal of
resistance (measured in ohms).
Conductivity can be determined from the voltage and current values. Because conductivity is
affected by cell geometry, specific conductivity should be used. The specific conductivity can be
calculated by multiplying the conductivity by the electrode’s cell constant. This constant is
determined by the formula, length / area. The unit S/cm is too large and usually the measurements
are done in the units mS/cm or uS/cm. The conductivity of the sample solution changes with
change in temperature, the relation is given as below,
Gt = Gtcal {1 + a(t - tcal)}
Where: Gt = Conductivity at any temperature in °C
Gtcal = Conductivity at calibration temperature in °C
a = Temperature coefficient of solution at tcal
tcal = Calibration temperature.
With the PICO pH/Conductivity meter’s this change can be compensated manually or
automatically by considering the Temperature Coefficients of the Sample. For precise Conductivity
measurement automatic temperature compensation (ATC) with separate temperature sensor is
needed and Sample and Standard should be at same temperature, preferably at the reference
calibration temperature 25°C
The Temperature coefficient mostly used is 2% per °C, since this approximates water and NaCl.
Conductivity is useful for measuring total dissolved solids (TDS) and salinity.
Page 9 of 76
Conductivity Cell
Table 1 Table 2
Some common conductivity values of typical Samples Some Typical Temperature Coefficients
0 0
Samples at 25 C Conductivity , uS/cm Percent/ C
Sample 0
Ultrapure Water 0.055 (at 25 C)
Distilled Water 0.5 Ultrapure Water 4.55
Power Plant Boiler Feed Water 1.0 Salt Solution (NaCl) 2.12
Drinking Water 50 5% NaOH 1.72
Potable Water 1055 Dilute Ammonia Solution 1.88
Sea/Ocean Water 53,000 10% HCl 1.32
5% NaOH 223,000 5% Sulfuric Acid 0.96
10% NaOH 355,000 98% Sulfuric Acid 2.84
10% HCl 700,000 Sugar Syrup 5.64
31.0% HNO3 865,000
Page 10 of 76
The pH values of some common industrial and household products/chemicals
are given as follows :
EXTREMELY ALKALINE
14.0
BOTTLE WASHING
13.0
COPPER PLATING BLEACH
12.0
MILK OF MAGNESIA
10.0
LIME - SODA
SOFTENING BORAX
9.0
BAKING SODA
AQUARIUM SALT WATER 8.0 SEA WATER
BREWING PROCESS
6.0
NICKEL PLATING
ORANGE JUICE
4.0
PICKLE PROCESSING
PHOTOENGRAVING
3.0 VINEGAR
LEMON JUICE
2.0
1.0
BATTERY ACID
EXTREMELY ACID
0.0
Page 11 of 76
FEATURES:
1) Advanced Micro-controller based user friendly state-of-the-art product design.
2) PICO+ pH/Conductivity Meter complies with USP and DIN Standards.
3) Sleek molded cabinet with Alphanumeric splash-waterproof polyester keyboard with soft keys.
4) User interactive software in dialog mode for ease of operation with protection against invalid
entries.
5) 20 X 2 Line Back lighted LCD Display.
6) Simultaneous Measurement of pH, mV and Temperature or Conductivity (S/mS)/ or TDS
(PPM/PPT), Salinity (Optional) and Temperature. SCAN Key for displaying the stored
values with Reading No. and Sample name.
7) Non volatile memory storage of,
a) 40 pH, 40 mV, 40 Conductivity (s/mS), 40 TDS (PPM/PPT) and 40 Salinity reading
(Optional) with temperature in pre-selected GLP or Standard format.
b) 24 data memory readings of pH, temperature and time in data logging mode.
c) Calibration data for pH & mV (relative).
8) MEM CLEAR Key to clear the memory buffer block within the selected measurement mode.
9) Calibration of instruments with User selected calibration standards. Calibration data of pH -
Slope and offset on display.
10) Password Protection for Calibration data and stored readings
11) Calibration data with date and time Stamp for authentication.
12) Calibration reminder alarm.
13) Alphanumeric entries of Sample Name, Identification Number for authentication.
14) 5 point Calibration with Auto Buffer Recognition & Slope is determined by Linear regression or
Segmentation (Point to Point) Interpolation Method.
15) Automatic Conductivity Range selection for faster measurement from 0.01S to 199.9mS and
TDS range 0.1PPM to 99.9PPT (G/L) with conductivity cells having constants 0.1, 1.0 & 10.0K.
16) Automatic temperature compensation by PT100 Temperature sensor Range 0 to 100°C.
For conductivity - User selectable temperature coefficient & Manual Compensation facility.
17) Input Stability Indicator for reproducible results.
18) Built-in Real Time Clock (RTC) for time display and on report printout.
19) Report Format:- Complying with GLP requirements.
20) Factory entered CUSTOMER NAME and INSTRUMENT SR. No. on report printouts for
authentication.
21) Probe Check Mode for checking Electrode Aging.
22) Error indication helps user to trace the problem.
23) Audible Beep indication during Valid Key operation.
Page 12 of 76
SPECIFICATIONS:
pH Millivolt
pH Range : -2.00 to 19.999 mV range : 0 to 1999.9 mV
Resolution : 0.001 pH Resolution : 0.1 mV
Accuracy : 0.002 pH Accuracy : 0.2 mV
Slope : 90% to 105%
Sensor : Combination pH electrode TDS
/ Two separate Electrode Range : 0.1PPM to 99.9PPT(G/L)
Calibration : Auto-calibration with buffer Resolution : 0.01 (Displayed).
recognition and printout of Accuracy : 0.5%.
Calibration report. TDS Factor : 0.1 – 0.9
(default = Last value, #0.5)
Conductivity Salinity
Range : 0.01uS/cm to 199.9mS/cm. Range : 0.1PPM to 99.9PPT(G/L)
(Depending on the Cell used 0.1 / 1.0 / 10.0 K). Resolution : 0.01 (Displayed).
Resolution : 0.001 (Displayed). Accuracy : 0.5%.
Accuracy : 0.5%. Salinity Factor : 0.1 – 0.9
Temperature Coeff.: 0.00 – 9.99 % /°C (default = Last value, #0.5)
(default = Last value, #1.92)
Temperature
Temperature range : 0 to 99 .9 Deg. C
Resolution : 0.1 Deg. C
Relative Accuracy : 0.2 Deg. C
Sensor : RTD PT 100.
Temp. Compensation : Automatic / Manual
Display : 20 x 2 line Back lighted LCD display.
Keyboard : Alphanumeric splash waterproof polyester soft keys.
Report format : (a) Report of selected pH/mV/Conductivity/Salinity readings
(b) Calibration Reports
(c) Data Table & Graph (Optional)
Input : 1 no. - BNC & 1no. - Banana for pH comb. & Ref. Electrode,
2 no. - Banana sockets for Conductivity Electrode,
1 no. - 5 pin Shell connector, ATC – PT100
Output : Printer (Parallel) : with selectivity for DMP or Deskjet,
Printer (Serial ) : 48 column Thermal Printer
PC–RS232(Optional)
Environmental Operating Conditions: -
a) Operation : Indoor
b) Temperature : Ambient to 45 °C
c) Relative Humidity : 5 to 90% non-condensing.
Power requirement : DC Adapter 90 – 270VAC / 9VDC 1A (10W Max)
Dimensions : 30 (L) x 25 (W) x 8 (H) cm
Shipping Weight : 5 Kg.
Page 13 of 76
Chapter 3
Setting Up
UNPACKING
Page 14 of 76
INSTALLATION
1. REQUISITES
Leveled Platform/Table with nearby 230Vac, 50Hz/ 5 Amp. Power socket. (Ensure the Input
Power supply is within +/- 10% tolerance).
Make arrangements for liberal amount of fresh distilled water and any five standard traceable
buffers (difference between consecutive buffer should not be greater than 4 pH unit) and
1~2gms. KCL or 0.1N HCL for pH Electrode activation.
Refer pH electrode maintenance section, Page no. 62.
Conductivity calibration solution 1.413 mS (0.01M KCL).
Clean & dry glass beakers for all sample measurements.
Tissue paper roll and Suitable container for Waste disposal.
Carefully remove the meter wrapping and the accessories from the packaging.
Place the meter on the table where it will be installed for measurement purpose.
Align the unit for complete view of LCD Display by opening the Tilt Legs.
For safety reasons and to ensure sufficient cooling in the electronic
compartment, the side walls of the instrument must be kept at least 20 cm
away from the walls or other objects. No containers, chemicals or other
appliances should be placed behind a instrument.
Get the Flat Base round plate & holder arm supplied with the unit.
The Electrode Stand is made by inserting the Holder arm bottom adapter onto the round base
plate pin firmly and tightens the wing nut on it for sturdy assembly.
Adjust the Electrode holder and stand angle as required and tighten the wing nut on the center
joint to lock the setting.
Carefully insert the Electrode from top and temperature probes in the holder.
Place the assembly on flat horizontal plane besides the meter.
Page 15 of 76
B) Serial Printer
Place the printer on the suitable side of the meter.
Connect the printer serial data cable to the RS232-2 port in 9 pin D-shell connector
provided on the back panel.
Ensure that printer is loaded with paper.
5. PC CONNECTIVITY
Connect the supplied serial cable to the 9 pin D Male connector at the rear side of the
instrument and other terminal to com port of the PC. A Hyper Terminal Program is used to
receive the data from the instrument.
Refer PC connectivity for more details.
Ensure proper Mains fuse has been installed in the instrument as per the
input voltage mentioned on the instrument label.
Connect the pH, Conductivity Electrode and Temperature Sensor to appropriate socket /
Connector to back panel of the meter.
Carefully remove the cap from end of the Electrode (This cap is used in shipping to protect the
electrode bulb and also to keep the electrode in moist condition until put into use)
Check the electrodes internal reference filling solution level, if necessary level it with
recommended solution (3M KCl).
Place the pH electrode into electrode storage solution, pH 4 buffer or saturated KCL solution.
Place the conductivity electrode into D. M. Water/ Distilled Water to remove any impurities
adhering to the probe body and the platinum pins.
Connect the DC adapter to a suitable 230 VAC/50 Hz stabilized power supply with proper earth
connection.
Switch ‘ON’ the meter. The 20 character, 2 line display will read as follows.
LABINDIA PICO Vx.xx LABINDIA PICO Vx.xx
pH Meter pH / CON Meter
--------------------------------------
Page 16 of 76
PANEL DETAILS
PICO – pH / CON meter has two functional Panels, for which user has to get familiar with.
Front Panel - Consisting of Name plate, 20 charactor X 2 line back lighted LCD display and
Splash waterproof polyester keyboard ( 20 keys ).
Back Panel - Consisting of BNC connection with 4mm banana socket for Combination / two
seperate pH Electrode, 4mm banana socket 2 no.s for Conductivity cell connection, 5-pin/M -
shell connector for RTD Temperature sensor, 25-pin/F - D type connector for Printer.
7 8 9 MEM
CLEAR RESET
STU VWX YZ -
4 5 6 FWD
CAL
JKL MNO PQR TEMPCO
1 2 3
SCAN PRINT
ABC DEF GHI
0 CE
MODE ENTER
- . / GLP BACK
Page 17 of 76
Key functions:
L.H.S. Keys: Alphanumeric Dual Function ( A~Z , 1~9 ), Zero (0) with Dash (-), Dot (.),Slash(/),
decimal point ( . ) and Clear Entry( CE ) Key.
FWD : This key has dual function, (a) To check the reading in scan mode. (b) To
TEMPCO select Auto / Manual temperature compensation.
SCAN : To SCAN / read the stored readings with the help of FWD/TEMCO key.
CE : To clear the wrong entry & to go backward for previous display screen.
BACK
: To Select the GLP or Standard Mode between sample measurement & storage.
GLP GLP mode :- In this selection, the Sample related detail such as
Sample Name and ID. NO. can be entered after each Measurement.
Standard mode:- In this selection, Direct Measurement can be done
with-out entering the Sample related details.
Page 18 of 76
2. Back Panel – details
BNC 5 Pin Shell
Connector for Connector for
pH Indicating PT100-RTD
/ Combination Temperature
Electrode probe
4mm X 2
Banana
DC Power socket for
Adapter Input Conductivity
Electrode
Page 19 of 76
Chapter 4
Operation
RESET key is used to initialize the Meter. By pressing the RESET key Meter gives Beep and
Display reads,
LABINDIA PICO Vx.xx
pH / CON Meter
To select PC/Printer,
Display reads,
LABINDIA PICO Vx.xx
pH / CON Meter
Press,
Enter Password Enter Password (Min. 4 digit & Max. 8 digit)
PRINT xxxx
Press,
Data Transfer
ENTER 1) Printer 2) PC
To transfer data to Printer, Press ‘1’ or to transfer data to PC, Press ‘2’.
Press,
1 Select Printer
ABC 1) DMP 2) Deskjet
To transfer data to DMP, Press ‘1’ or to transfer data to Deskjet, Press ‘2’.
Page 20 of 76
Press,
1 pH Mode
ABC Press ENTER or MODE
Press,
Select Measr. Mode
ENTER 1) pH 2) Cond.
Display reads,
LABINDIA PICO Vx.xx
pH / CON Meter
Press,
Enter Password Enter Password (Min. 4 digit & Max. 8 digit)
SCAN xxxx
Press,
pH Resolution
ENTER 1) 1 2) 2 3) 3
MODE Key is important key, as it allows the user to select the Operating and Measurement
Mode.
When first time the Mode key is pressed the instrument prompts for Measurement mode.
Display reads,
Select Measr. Mode:
1) pH 2) Conductivity
Display Display
Page 21 of 76
PART – 1,
pH MEASUREMENT / FUNCTIONAL MODES:-
The modes can be selected by pressing MODE key repetitively till desired mode/function is
displayed and later acknowledged by pressing ENTER key.
Please note the Second Line of the display shows the Active Keys. The very first Mode Displayed
is, pH Mode.
pH Mode
Press ENTER or MODE
ENTER A1 Refer page 23.
nd
2 press,
mV Mode
Press ENTER or MODE
ENTER B1 Refer page 32.
rd
3 press,
6th press,
Date & Time Mode Refer page 39.
Press ENTER or MODE
F1
th
7 press,
Password Mode Refer page 40
Press ENTER or MODE
G1
After desired Measurement / Functional Mode is displayed and then Enter Key is pressed to
confirm the selection .
Page 22 of 76
A. pH Measurement :-
Before starting the pH measurement, Ensure the Instrument is Setup properly as mentioned
on Page 15.
“Wait Stabilising” – indicates, the instrument is checking Electrode signal Stability and once
the signal is steady it shows the measured pH, mV and temperature.
Display meaning,
Operating Mode
Sample Temperature
mark - Standard
(ATC / Manual Comp.)
No mark – GLP
The function of the above keys are described at later in this manual.
Page 23 of 76
I. Selecting Temperature Compensation: TEMPCO
The user has two options for temperature selection at which the automatic Temperature
compensation is to be considered during Calibration & measurement.
ATC at : 1
1) 25°C 2) 20°C
For e.g - Compensation with respect to 25°C Press ‘1’ OR for 20°C Press ‘2’,
Press Display
1 ATC at : 1
ABC 1) 25°C 2) 20°C
For Manual Temp. Compensation, measure the Solutions Temperature with thermometer and
then enter the correct temperature value. say @ 25.0 °C user entered.
Press Display Press
2 Select TEMPCO Mode:2
ENTER
DEF 1) Auto 2) Manual
Display Press
Enter Temp ( °C ) = 2 5 0
Press DEF MNO GLP - . /
Page 24 of 76
II. Calibration of Instrument and pH Electrode: CAL
The PICO+ pH meter is available with two interpolation method for calibration.
1. Segmentation (default) or 2. Linear Regression (Optional) or Both.
In this manual both the calibration methods are explained for user information purpose.
User should consider the appropriate method as per instrument ordered & supplied or
applicable.
Press (1) Key for Calibration Linear regression interpolation or Press (2) key for Segmentation
interpolation method and refer to appropriate section for further operation. For more details about
the interpolation methods refer to Annexture – C.
Ensure the Electrode is connected to instrument and was properly Activated & stored.
Ensure the Temperature Sensor is connected to instrument.
Two Or more Buffers should be used for Calibration before making pH measurement.
NOTE : For proper conditioning and storage of electrode, refer Maintenance Section
Page 25 of 76
As a GLP Sequence the User should Perform the Standardisation first and then Calibrate using
either Acidic or Basic Buffer.
The CE key is used to Terminate / complete the calibration when using less than 5 buffers. Max 5
pH buffers can be used.
Enter the First Calibration Buffer Value,
Enter Buf1 = 4 0 0
JKL GLP - . / - . /
Note:- The next buffer Value should not be more than 4 pH unit, See Display msg. (##)
Enter Buf2 = 7 0 0
STU GLP - . / - . /
pH =+ A.BCD Enter.
ENTER
mV = - PQR.S 27.0°C
Page 26 of 76
Note : “L”in the RHS corner of Display indicates that the Linear Regression Interpolation
method is used for Calibration.
th
When the last (5 ) buffer is completed, instrument checks the Slope of the Electrode
obtained (Limit = 90 % ~ 105%) and
Page 27 of 76
III. Taking pH reading of Sample and Storing :- ENTER
Remove the pH electrode from the solution and rinse it with distilled water 2 ~ 3 times, soak
the excess water with tissue paper and insert the pH electrode and Temperature sensor
(RTD) in the Sample beaker.
Stir the Sample Solution keeping the pH Electrode and RTD immersed.
Incase of Standard Mode selected, pressing the ENTER key the readings are Stored in the
memory without Sample Name & ID.NO., else if GLP operating mode is selected at start
readings gets Stored in the memory with the Sample Name & ID. NO.
( GLP )
Sample Name
Enter the Sample name by using Alphanumeric Keys.
Refer Appendix-B.
ENTER
ENTER
Instrument also has a facility to Prints a Report of single Reading taken ONLINE if printer is
connected correctly (Printer is ON and Paper is loaded in the tray) else it will ignore printing
and continue the pH measurement.
Page 28 of 76
IV. Scanning Stored pH reading :- SCAN
PICO pH has memory storage capacity for 40 pH reading with / without Sample Name & ID.NO.
The user can select the Stored readings data for viewing by pressing SCAN key and can scroll the
readings further by pressing FWD/TEMCO key repeatedly.
pH =+ A.BCD R1 L
SCAN
mV = - PQR.S 27.0°C pH = + 07.003 No.1
FWD
TEMPCO pH = + 07.006 No.2
FWD
ENTER
TEMPCO pH = + 04.002 No.3
FWD
TEMPCO
FWD
TEMPCO pH = + 09.018 No.10
By pressing ENTER key anytime during scan, the Instrument Display returns to measurement
function.
Note :- If there is no pH reading stored in the instrument memory, it prompts the user by
displaying message,
No data for Scanning
Page 29 of 76
V. Printing the REPORT of pH readings :- PRINT
The PICO pH meter provides Report printing facility with an GLP approach. Comprising of
Customer Name, Instrument Sr. No., Date and Time, two types of report can be obtained,
1. Report with the user selected pH readings (by nos.)
2. Calibration Report. ( Last calibration Performed)
Display,
pH =+ A.BCD R10 L Print Report :
PRINT
mV = - PQR.S 27.0°C 1)Reading 2)Cal. Repo
After Printing is Over, automatically the Instrument Display returns to measurement function
Page No. 27 Q1
As now the contents of Storage memory are nil, now again new pH reading can be stored from
Reading no. 1 to 40.
Page 30 of 76
2. Clearing ALL data memory by MEM CLEAR Key. i.e a) Readings, b) Calibration data or c) All (
Both ) at one instant.
Display,
pH =+ A.BCD R10 I MEM Enter Password
mV = - PQR.S 27.0°C CLEAR _
Page 31 of 76
B. mV Measurement : B1 from page no.22
Using PICO pH meter to measure Absolute millivolt Electrode potential ( mV ) value or Relative
millivolt with consideration of OFFSET.
Ensure the Correct Electrode is connected to the BNC socket at the back of the instrument.
Clean the electrode with distilled water and sock the excess water with tissue paper.
Insert the electrode in the Standardisation solution.
Standardisation Over
ENTER
Offset =+ GHI.J
R1
Page 32 of 76
II. Taking mV reading of Sample and Storing :- ENTER
The Measurement is similar to pH part. Refer page no. – 23 for explanation.
The instrument reads the electrode potential and by pressing ENTER key it gets stored in the mV
memory block.
Maximum 40 mV readings can be stored.
A. When PICO mV memory storage capacity is Full ( > 40 pH reading ), the indication is
displayed by greater than sign ( > ) previous to Reading no. Further mV reading storage by
ENTER key is disabled and a Memory Full display appears enabling the user to Clear the
memory or Quit back to measurement function for taking printout.
Display,
mV Mode No.>40 s Memory full :
ENTER
mV = - PQR.S 27.0°C 1) Mem. Clear 2) Quit
Page No. 32 R1
As now the contents of Storage memory are nil, now again new mV reading can be stored from
Reading no. 1 up to 40.
Page 33 of 76
B. Clearing ALL data memory by MEM CLEAR Key. i.e a) Readings, b) Calibration data or c) All
(Both) at one instant.
Display,
mV Mode No.>40 s MEM Enter Password
mV = - PQR.S 27.0°C CLEAR _
Page 34 of 76
C. Data Logging of pH v/s Time :- C1 (Optional) from page no.22
Data Logging of pH v/s Time is an important facility for the study of Chemical Kinetics Reaction or
Enzymatic Kinetic Reaction. This can be also used for process monitoring in laboratory pilot plant
runs.
The instrument records the pH reading w.r.t the User entered Data Log. Intv. (1 ~ 60 min.) and
No. of Intervals ( no. of readings – 24 max). The pH Data can be logged upto 24hrs. duration
for Slow Kinetic reaction with the Log. Intv of 60 minutes each.
With pH Limit 1 & Limit 2 values for marks, the Graphical representation of the reaction trend
gives more meaning to the recorded values.
The Sample Name and ID. NO. adds GLP compliance and authentication.
Calibrate the Instrument and pH Electrode for 3 points by selecting pH Measurement function.
Press CAL
Key for performing Calibration. Refer Page no. 22 & 23.
Other Active Keys are PRINT for printing the Logged data report & graph and
GLP
Entering Data Logging Parameters & Start logging : Data Log. Mode : Press
ENTER/MODE/PRINT/GLP
Page 35 of 76
Enter the no. of Interval,
(say, 5 intervals )
5 No. Of Intervals : pH Limit 1 : _
ENTER
MNO ( 1 - 24 ) : 5 pH Limit 2 :
Enter the pH Limits ( multiple of 1 pH unit ) which will bracket the pH value being logged during
Kinetic reaction.
6 pH Limit 1 : 6 8
ENTER
PQR pH Limit 2 : _ VWX
Reading No: 2
Entered as + 6.856
pH =+ 7.025 No. 5 L
C1 D.Log (hh:mm) 27.0°C
The Instrument Logs the reading sequentially one after the other as per log interval (1min. in
above e.g.) till the final reading is recorded and Display returns to the Data Logging function.
To print the data and graph, press PRINT key after the data logging is over.
Ensure the Printer is ON, properly connected to the instrument and loaded with paper with Online
indication ON.
Data Log. Mode : Press Wait . . Xfering Data !
PRINT
ENTER/MODE/PRINT/GLP
Once the Printing data and graph is over the display returns to data Log function.
Page 36 of 76
D. Probe Check Mode :- D1 from page no.22
Probe check function is another helpful feature in PICO pH to establish the pH electrodes aging.
The probe Check function is performed with Buffer 7.00 pH to find out the Iso-potential point
offset.
Ensure all the precautions required for Electrode Cleaning.
Rinse the pH electrode with distilled water 2~3 times , remove the excess water by soaking it with
tissue paper and insert the pH electrode and RTD in 7.00 pH beaker.
Display
Probe Check Mode Dip Probe in 7 pH &
ENTER
Press ENTER or MODE Press ‘ENTER’
Wait ! Stabilising . .
ENTER
A1
ENTER
If the Probe ( pH Electrode ) is having Offset more than +30 mV or Less tan –30 mV , it can be
tried for reactivation several times and recheck the aging. If then instrument prompts as Probe
faulty, the Electrode should be thrown and another electrode should be used.
Calibration Reminder is another essential / helpful feature in PICO pH to assist the user for
complying with GLP / ISO regulatory needs.
The Calibration Reminder can be enabled by setting the next Probable DATE for performing
Calibration. PICO pH Meter’s internal Real Time Clock keeps the watch on it when powered ON
and the pH measurement function is selected on that date, the Buzzer will generate BEEP sound
and the Display prompts for re-calibration.
Page 37 of 76
Ensure the DATE and TIME of internal Real Time Clock is SET correctly. Refer page no. 39.
Display,
Calibration Reminder 1) Previous Calib :
E1 Press ENTER or MODE
ENTER
2) Set Reminder
Display,
Calibration Reminder Calibration Reminder
E Press ENTER or MODE
ENTER
1) ON 2) OFF :
To Enable / SET Calibration Reminder To Disable / SET Calibration Reminder
Select, 1) ON Select, 2) OFF
Press Press
1 Calibration Reminder 2 Calibration Reminder
ABC 1) ON 2) OFF : 1 DEF 1) ON 2) OFF : 2
ENTER ENTER
E1
Enter the Reminder Date, Month & Year and acknowledge it with ENTER key. The Cursor guides
the entry location.
Page 38 of 76
The reminder Date has to be at least next to Present ( today’s ) Date. For wrong
Date entry instrument displays it as INVALID ENTRY.
On the Reminder Date, when the instrument is Powered ON and pH measurement function is
selected,
Display reads, After 2~3 sec.
Instru. needs calib. Calibration :
1)Y 2)N
1 2
ABC DEF
Slope = 100.91%
ENTER
Offset = -05.55mV
pH =+ A.BCD R1 N
ENTER
mV = - PQR.S 27.0°C
Refer Page no. 25.
Enter password.
This mode is used to set the Real Time Clock built-in the PICO pH.
It Can also be used as the simple Clock when the PICO pH meter has no Sample pH / mV
measurement.
Display,
Date & Time Mode Enter Admin Password
ENTER
Press ENTER or MODE xxxx
Enter Password by using Alphanumeric Keys
ENTER-Edit / CE-Back
ENTER
dd/mm/yy hh:mm
Enter the Reminder Date, Month & Year sequentially and acknowledge it with ENTER key. The
Cursor guides the entry location.
For wrong Date & Time entry, instrument displays it as INVALID ENTRY.
Page 39 of 76
G. Password Setting :- G1 from page 22.
This mode is used to set the user password for protection against unauthorized access to
1)Calibration, 2) Readings in the memory, 3) Stored Calibration data and 4) Setting new Password
itself.
Password is in Only Numeric from, Entry required is min 4 nos. and max 8 nos.
Display
Password Mode 1) Change Password
ENTER
Press ENTER or MODE 2) Enable 3) Disable
To Change Password
Press
1 1) Admin 2) User
ABC
Press Press
1 Enter Old Password 2 Enter Admin Password
ABC ******** DEF ********
Press Press
Enter New Password Enter New Password
ENTER ENTER
******** ********
Press Press
Repeat New Password Repeat New Password
ENTER ENTER
******** ********
Press Press
New Password Stored New Password Stored
ENTER ENTER
Enable Password
Press
2 Enter Admin Password Password Mode
ENTER Press ENTER or MODE
DEF ********
Disable Password
Press
3 Enter Admin Password Password Mode
ENTER Press ENTER or MODE
GHI ********
---XXX---
Page 40 of 76
PART – 2, referred further from page no. 22,
Please note the Second Line of the display shows the Active Keys.
Display
Conductivity Mode Refer page 42.
Press ENTER or MODE
ENTER A2
nd
2 press,
##
TDS Mode
Press ENTER or MODE
ENTER B2 Refer page 52.
rd
3 press,
## Salinity Mode
ENTER C2 Refer page 53.
Press ENTER or MODE
th
4 press,
## Data Log Mode : Press
ENTER / MODE / GLP
ENTER D2 Refer page 54.
th
5 press,
Probe Check Mode
ENTER E2 Refer page 55.
Press ENTER or MODE
6th press,
Calibration Reminder
Press ENTER or MODE
ENTER F2 Refer page 56.
th
7 press,
Date & Time Mode ENTER G2
Press ENTER or MODE Refer page 58.
th
8 press,
Password Mode ENTER H2
Press ENTER or MODE Refer page 58.
After desired Measurement / Functional Mode is displayed and then press Enter Key to
confirm the selection . ## only If this mode is Available
Page 41 of 76
A. CONDUCITVITY Measurement :-
Before starting the Conductivity measurement, Ensure the Instrument is Setup properly as
mentioned on Page 15.
Select the Conductivity Mode, Refer page no.41. A2
The Instrument Display reads,
Enter Cell Const. :
1.000
Enter the Cell constant of the connected cell (electrode) -
ii) For same value of cell constant just press 'ENTER' key. The display reads,
Wait ! Stabilising . . Con = A.BCDS No.1 C
K = 1.000 27.0°C
“Wait ! Stabilising” – indicates, the instrument is checking Electrode signal Stability and
once the signal is steady it shows the measured Conductivity and temperature.
Display meaning,
Operating Mode
Sample Temperature
mark - Standard
(ATC / Manual Comp.)
No mark – GLP
Page 42 of 76
Before performing the Conductivity Measurement,
III. Enter the Temperature Coefficient &
FWD
Select the Temperature Compensation Type by pressing key.
TEMPCO
IV. Calibrate the Instrument with required CELL/Electrode press CAL key.
The function of the above keys are described at later in this manual.
Enter the required Temperature Coefficient value for ATC of Sample/STD Conductivity
Calibration & measurement, Refer Table 2, page no. 7 and given below for KCL standards.
If not known it can be entered as 2.0 % / °C, since for most application, this will provide good
results.
Conductivity @25.0°C Concentration of KCL, mol/lit. TEMCO @25°C.
12.880 mS 0.1 M + 1.90 %
1.413 mS 0.01 M + 1.94 %
147.0 S 0.001 M + 2.04 %
Please note :- If the instrument has been earlier entered with the TEMPCO value it will be
displayed for reference purpose, which user can acknowledge by pressing ENTER key.
For e.g. 2.0, Press
2 0 Enter Coefficient :
DEF GLP - . / 2.0
The user has two options for temperature selection at which the automatic Temperature
compensation is to be considered during Calibration & measurement.
ATC at : 1
1) 25°C 2) 20°C
For e.g - Compensation with respect to 25°C Press ‘1’ OR for 20°C Press ‘2’,
Page 43 of 76
Press Display
1 ATC at : 1
ABC 1) 25°C 2) 20°C
Press Display
1 Select TEMPCO Mode:1
ABC 1) Auto 2) Manual
Press Display
Con = A.BCDS No.1 C
ENTER
K = 1.000
Page 44 of 76
2) Calibration of Instrument and Electrode: CAL
Ensure the Correct Electrode is connected to instrument and was properly cleaned & stored.
Ensure the Temperature Sensor is connected to instrument.
Ensure the Correct Temperature Coefficient and Temperature compensation type is entered.
Rinse the Conductivity electrode with distilled water 2~3 times, remove the excess water by
soaking it with lint free tissue paper (Do not rub / scratch the platinum plates).
Press Display
Cell Const = 0.989
CAL Instrument will show earlier calibration.
Press Display
Enter Password
ENTER
_
Enter Password
Enter the Password Set by using the numeric keys
Press Display For Cell K=1, Press
Enter Cell Const. : 1 Enter Cell Const. :
ENTER
_ ABC 1
Press Display
Enter Soln Cond. :
ENTER
_ P2
a) The Sample Conductivity should be within the Range of the selected CELL to
b)
reduce the span errors.
The Conductivity Standard should bracket the Sample conductivity to reduce
the linearity errors.
Page 45 of 76
Below given is the table of suitable Calibration Standard/s to be used for calibration as per Cell
Constant chosen. Table 4.
Conductivity Cell Constant Conductivity standards. KCl Concentration
Approx. K S. @25.0°C. Mol/lit.
0.1 147.0 S 1) 0.001M
147.0S 1) 0.001M,
1.0 1.413mS 2) 0.01M,
12.880mS 3) 0.1M.
1.413mS 1) 0.01M,
10.0 12.880mS 2) 0.1M,
111.34mS 3) 1M.
Page 46 of 76
Press The Instrument checks the CAL Limits.
ENTER
Page 47 of 76
III. Taking Conductivity reading of Sample and Storing :- ENTER
Remove the Conductivity electrode and Temperature sensor from the solution and rinse them
with distilled water 2 ~ 3 times, soak the excess water with tissue paper.
Again rinse the Cell with the Sample solution, then insert the Cell and Temperature sensor
(RTD) in the another beaker containing same Sample.
Stir the Sample Solution keeping the Conductivity Cell and RTD immersed.
Incase of Standard Mode selected, pressing the ENTER key the readings are Stored in the
memory without Sample Name & ID.NO., else if GLP operating mode is selected at start
readings gets Stored in the memory with the Sample Name & ID. NO.
Sample Name
Enter the Sample name by using Alphanumeric Keys.
Refer Appendix – B.
ENTER
ENTER
# Instrument also has a facility to Print a Report of single Reading taken ONLINE, if printer is
connected correctly (Printer is ON and Paper is loaded in the tray). Else it will ignore printing
and continue the Conductivity measurement.
Page 48 of 76
IV. Scanning Stored Conductivity reading :- SCAN
PICO Conductivity meter has memory storage capacity for 40 reading with / without Sample Name
& ID.NO.
The user can select the Stored readings data for viewing by pressing SCAN key and can scroll the
readings further by pressing FWD/TEMCO key repeatedly.
FWD
TEMPCO Con= 1.414 mS No.2
ENTER FWD
TEMPCO Con= 100.4 mS No.3
FWD
TEMPCO
FWD
TEMPCO Con= 140.9 S No.10
By pressing ENTER key anytime during scan, the Instrument returns to measurement function.
Note : user
If there is no Conductivity reading stored in the instrument memory, it prompts the
by displaying message,
No data for Scanning
Page 49 of 76
V. Printing the REPORT of Conductivity readings :- PRINT
The PICO Conductivity meter provides Report printing facility with an GLP approach. Comprising
of Customer Name, Instrument Sr. No., Date and Time, two types of report can be obtained,
1. Report with the user selected conductivity readings (by nos.)
2. Calibration Report. (Last calibration Performed)
Display,
Con= A.BCDmS No.10 C Print Report :
PRINT
K = 0.996 27.0°C 1)Reading 2)Cal. Repo
After Printing is Over, automatically the Instrument Display returns to measurement function
If printer is not Connected or is OFFLINE due to No paper Display Printer not Ready!..
Page 50 of 76
2. Clearing ALL data memory by MEM CLEAR Key. i.e a) Readings, b) Calibration data or
c) All (Both) at one instant.
Display,
Con= A.BCDmS No.10 C MEM Enter Password
K = 0.996 27.0°C CLEAR _
Press Display
1 Memory Clear : 1 Memory is cleared.
ENTER
ABC 1) Rdgs. 2) Cal. 3) All
Press Display
2 Memory Clear : 2 Memory is cleared.
ENTER
DEF 1) Rdgs. 2) Cal. 3) All
Press Display
3 Memory Clear : 3 Memory is cleared.
ENTER
GHI 1) Rdgs. 2) Cal. 3) All
Page 51 of 76
B. TDS – Total Dissolved Solids Measurement (Optional):-
Before starting the TDS measurement, Ensure the Instrument is Setup properly as mentioned
on Page 15.
Select the TDS Mode, Refer page no.41 B2
The Instrument Display reads,
Enter TDS Factor :
0.500
Enter the TDS factor for KCL by using numeric keys.
The Calibration data is obtained from Conductivity mode. Calibration i.e. “CAL” key
is disabled in TDS mode. Ref page no.52.
So ensure the instrument is calibrated first in Conductivity mode with proper
conductivity standard solution and Cell/electrode.
Display reads,
TDS= xx.xx ppm No.1 N
T.F = 0.500 27.0°C
Now you can perform TDS measurement of samples. All operations are similar to Conductivity
measurement - Storing the reading in memory with/without GLP, Scanning the stored readings,
memory clearing and report printing, etc.
Display shows the TDS Value considering the New TDS factor entered.
Page 52 of 76
C. Salinity (Optional):-
Before starting the Salinity measurement, Ensure the Instrument is Setup properly as
mentioned on Page 15.
Select the Salinity Mode, Refer page no.41 C2
The Instrument Display reads,
Display Press
Enter Sal Factor : Enter Sal Factor :
ENTER
0.500 0.5
The Calibration data is obtained from Conductivity mode. Calibration i.e. “CAL” key
is disabled in Salinity mode. Ref page no.52.
So ensure the instrument is calibrated first in Conductivity mode with proper
conductivity standard solution and Cell/electrode.
Press Display
Sal = XXX.X ppm R1 C
ENTER SF = 0.50 *
24.7
Now you can perform Salinity for Salinity measurement of samples. All operations are similar
to Conductivity measurement - Storing the reading in memory with/without GLP, Scanning the
stored readings, memory clearing and report printing, etc.
Press Display
Enter Salinity Factor : Enter Salinity factor.
CE
0.5 Salinity Factor range :- 0.1 to 1.0
BACK
It will display the salinity value considering New Salinity factor entered.
Page 53 of 76
D. Data Logging:-
Select the Data Logging Mode, Refer page no.41 D2
To monitor the change in conductivity, over a period of time.
Say 1 min,
Press Display
1 Data Log. Intv.(min)
ABC (1 – 60) : 1
Press Display
To Start Logging
ENTER
Press ‘ENTER’
Display
100.1 ppm R1 C After each 1 min it will transfer data to printer or PC.
D.Log(00:00) * 24.8°
Page 54 of 76
E. Probe Check Mode :- E2 from page no.41.
Probe check function is another helpful feature in PICO Conductivity meter to establish the
Conductivity CELL aging.
The probe Check function is Similar to the Determination of Cell constant electrode/Calibration
with KCL standard solution.
Rinse the Conductivity Cell & RTD with distilled water 2~3 times , remove the excess water by
soaking it with tissue paper and insert the Conductivity Cell and RTD in Standard solution
beaker.
Press
Wait ! Stabilising . .
ENTER
If the Probe (Conductivity Cell/Electrode) Constant K is out off limit, it can be tried for reactivation
by re-platinizing several times and recheck the recovery. After rechecking if it prompts as Probe
faulty, the Electrode should be disposed off and another electrode should be used.
Labindia Analytical Instruments Pvt. Ltd. can Supply the Conductivity Cell Re-platinizing Kit
for Reactivation of the CELL. Please ask for details. - service.mfg@labindia.com
Page 55 of 76
F. Calibration Reminder :- F2 from page no.41
Calibration Reminder is another helpful feature in PICO Conductivity meter to assist the user for
complying with GLP / ISO regulatory needs.
The Calibration Reminder can be enabled by setting the next Probable DATE for performing
Calibration. PICO conductivity Meter’s internal Real Time Clock keeps the watch on it when
powered ON and the measurement function is selected on that date, the Buzzer will generate
BEEP sound and the Display prompts for re-calibration.
Ensure the DATE and TIME of internal Real Time Clock is SET correctly. Refer page no. 58.
Display,
Calibration Reminder 1) Previous Calib :
F2 Press ENTER or MODE
ENTER
2) Set Reminder
ENTER ENTER
Page 56 of 76
From previous page
Back to previous
screen . :1
1) Previous Calib
ENTER CE
BACK 2) Set Reminder
CE
BACK
2~3 sec
F2
Enter the New Reminder Date, Month & Year and acknowledge it with ENTER key. The Cursor
guides the entry location.
The reminder Date has to be at least next to Present ( today’s ) Date. For wrong
Date entry instrument displays it as INVALID ENTRY and returns back to the entry
screen.
On the Reminder Date , when the instrument is Powered ON and measurement function is
selected,
1 2
ABC DEF
ENTER ENTER
Calibrate : 1
1)Y 2)N
Q2
Refer Calibration – Page no. 42. Refer Page no. 47.
Page 57 of 76
G. Date & Time Mode :- G2 from page no.41.
This mode is used to set the Real Time Clock built-in the PICO conductivity meter.
It can also be used as the simple Clock when the PICO conductivity meter has no Sample
measurement.
Display,
Date & Time Mode Enter Admin Password
ENTER
Press ENTER or MODE xxxx
Enter Password by using Alphanumeric Keys
ENTER-Edit / CE-Back
ENTER
dd/mm/yy hh:mm
Enter the correct Date, Month & Year sequentially and acknowledge it with ENTER key. The
Cursor advances/guides the entry location.
For wrong Date & Time entry, instrument displays it as INVALID ENTRY.
This mode is used to set the user password for protection against unauthorized access to
1)Calibration, 2) Readings in the memory, 3) Stored Calibration data and 4) Setting new Password
itself.
Password is in Only Numeric from, Entry required is min 4 nos. and max 8 nos.
Display
Password Mode 1) Change Password
ENTER
Press ENTER or MODE 2) Enable 3) Disable
To Change Password
Press
1 1) Admin 2) User
ABC
Press Press
1 Enter Old Password 2 Enter Admin Password
ABC ******** DEF ********
Press Press
Enter New Password Enter New Password
ENTER ENTER
******** ********
Page 58 of 76
Press Press
Repeat New Password Repeat New Password
ENTER ENTER
******** ********
Press Press
New Password Stored New Password Stored
ENTER ENTER
Enable Password
Press
2 Enter Admin Password Password Mode
ENTER Press ENTER or MODE
DEF ********
Disable Password
Press
3 Enter Admin Password Password Mode
ENTER Press ENTER or MODE
GHI ********
---XXX---
Page 59 of 76
Chapter 5
Operation Guideline
Useful Hints for pH & mV measurement:-
New electrodes (or electrodes that have been in storage) should be conditioned in one inch
(25mm) of pH 7.0 or 4.0 buffer for several hours prior to their use. Make sure that the wetting
cap has been removed from the tip of the electrode.
Always remove the wetting cap and fill hole plug during calibration and measurements.
Replace the fill hole plug when done.
Calibrate the electrode daily. Though a 1-point calibration may be suitable for some
applications, we recommended that a 2-point calibration (that brackets the expected pH range)
be performed.
Be aware that if you are measuring hot or cold samples without the use of an ATC probe, the
values displayed are not accurate. You should be calibrating and measuring at room
temperature.
When transferring from one sample to another, always rinse the electrode with distilled water
and bolt dry. Handle the electrode carefully and do not use it as a stirring rod.
Do not use calibration buffers after the expiration date printed on the package. Keep the
bottles tightly capped and stored according to the manufacturer’s instructions. Never put used
buffer back into the bottle of new buffer.
For small sample volumes, make sure that both the pH bulb and the reference junction are in
contact with the sample.
Response time is a function of both the electrode and the solution. Some solutions have very
fast response while others, particularly those with low ionic strength, may take several minutes.
Samples must be in solution (water). You cannot measure the pH of a dry sample.
Page 60 of 76
Useful Hints for Conductivity & TDS
When measuring, make sure the solution is above the cell chamber slot.
Make sure the cell chamber is free of bubbles when measuring. To reduce air bubbles, stir
centrally and then leave to stabilize.
Allow sufficient time for the sensor to stabilize when measuring samples of different
temperatures.
Rinse sensor with next measurement sample between measurements, and ensure sensor is
dried when re-zeroing the instrument.
For greater accuracy, calibrate using a standard close in value to the sample.
Page 61 of 76
Chapter 6
Maintenance
And
Troubleshooting
Instrument Maintenance:-
The Labindia pH & Conductivity Meter needs no maintenance except for an occasional wipe with a
damp cloth. The front panel is made of ABS Resin and is not affected by the solvents. ABS Resin
is known to be affected by some organic solvents, do not use solvents like Chloroform, toluene,
xylene Acetone and methyl-ethyl-ketone. It is good laboratory practice to wipe away any spills as
soon as they occur.
Regular maintenance schedule and proper storage of pH electrode will maximize performances
and help to extend the life of electrode.
Weekly :-
1) Inspect electrode for scratches, cracks salt crystals built up on membrane / liquid junction
deposits.
2) The reference chamber of refillable electrode should be drained, flush with fresh filling solution
and refilled.
3) Use a dedicated pH electrode storage solution to keep electrode in proper working order.
4) Constant use of this solution will keep the electrode ready to use and provide fast stable
response.
Page 62 of 76
Short term storage :- Between measurement / up to one week.
Soak the electrode in pH storage solution.
pH ELECTRODE MAINTENANCE
Solutions used to clean electrodes must be handled with the care according to toxic
or corrosive substances. Please refer to relevant MSDS.
Do not allow the reference chamber of the electrode to dry out. Always keep it filled with the
proper fill solution.
Formation of KCl salt at the tip and side of the electrode is normal and should be rinsed off
with warm water.
Leave the tip of the electrode in an inch (25mm) of pH 7.0 or 4.0 buffer when it is not being
used. Do not leave it in distilled water. For long term storage follow the electrode
manufacturer’s instructions.
If the electrode slope is off or the electrode response has become sluggish or inaccurate,
change the fill solution. If that does not improve the response then the pH sending glass and
the junction have probably become coated with some of the samples being tested.
Test the junction for flow – rinse the tip of the electrode, blot dry and let the electrode hang in
the air for 15 minutes. A proper flowing junction should have KCl salt crystals forming on it. If
none appear, review the following suggestions for cleaning.
Protein contamination – Soak the tip of the electrode for 1 – 2 hours in a solution of
10% Pepsin and water with enough Hydrochloric acid added to bring the pH of the
solution to 1.0. Rinse the electrode and soak in pH 7.0 buffer until stable.
Oil contamination – Carefully clean the tip of the electrode using a cotton swab soaked
with alcohol or acetone. The tip of glass body electrodes can be put directly into organic
cleaners but do not put plastic or epoxy body electrodes into organic solvents. Rinse
the electrode and soak in pH 7.0 buffer until stable.
Hydrochloric acid for one hour. Rinse the electrode and soak in pH 7.0 buffer until
stable.
Page 63 of 76
pH Electrode cleaning
General
Soak the electrode in O.I.M HCL for 30 minutes
Removal of deposits
Protein :- Soak the electrode in 1% pepsin in 0.1 M HCL for 15 minutes.
Inorganic :- Soak the electrode in 0.1 M tetrasodium EDTA solution for 15
minutes
Grease & Oil :- Rinse the electrode with mild detergent of methanolic Solution.
Protect the conductivity system from dust and excessive heat and cold.
Keep all cable connector dry and free of dirt and extraneous matter.
Sensor cleaning
Conductivity sensors must be clean to produce accurate results; residue from previous
samples can coat surfaces of sensor and cause erroneous reading.
Clean sensors thoroughly with deionized water before and after making a measurement.
Remove oil residue or other chemical resistors (salts) with a detergent solution. Sensors
can soak in detergent solution for many hours without damage.
Page 64 of 76
Troubleshooting & Error Codes
Page 65 of 76
Consumables and Accessories
Sr.
Description Part No.
No.
1. pH Checker ( 4/7/10 pH) 50700002
Page 66 of 76
APPENDIX - A
SWITCH BANK 1.
SWITCH BANK 2.
NOTE :
After switches are correctly set turn the printer OFF and ON again to load the new
settings.
NOTE :
Please refer printer manual supplied along with the RP45 Shoppe Printer.
Using operation panel change the character pitch setting (CPI) to 20cpi for Esc P Emulation
Mode. (refer pg. no. 25 of TVSE-RP45 User Manual).
NOTE :
Please refer printer manual supplied along with the GP-80250IN Printer.
Page 67 of 76
APPENDIX – B
Alpha numeric Entries are required for Sample Name & ID.NO.
Sample Name :
Press Twice Press Thrice Press Thrice Press 4 times Press 4 times
4 1 4 1 1
JKL ABC JKL ABC ABC
Display reads,
Sample Name :
KCL1A
Page 68 of 76
APPENDIX – C
Linear Regression:
A single Best fit line is drawn for all the data points. Because a single line is
drawn to best fit all pH data points, the line may not go through all (any) of the
calibration points if there are more than two points.
The Slope and offset for this type of calibration are referred as best fit Slope
and best fit Offset. This type of calibration may be performed with
temperature compensation.
The Average of the Slopes and Offsets of the segments are reported in the
calibration report. This type of calibration may be performed with temperature
compensation. There must be at least three calibration points to apply this
method.
Page 69 of 76
APPENDIX – D
Installation/setup Instructions:
The instrument software has been incorporated with RS232C protocol interface for transferring the
printable data to PC having standard data terminal software Windows - Hyper terminal.
Setting HYPERACCESS NEW CONNECTION and COM port selection:- Enter Name PC-PICO,
Page 70 of 76
If the COM port chosen is not free or being used by another
application a pop up screen will be displayed with message
“Unable to Open COM_”.
Page 71 of 76
Now From the terminal window FILE menu select the Properties tool to set the Terminal Emulation
mode and Data Line
Once You click on the Properties menu -
Properties window will appear.
Page 72 of 76
Click the ASCII setup icon to enter the setting window.
To Generate the Icon for the PICO connectivity with PC, right click on the file name being in the
Hyper terminal Window – select Create Shortcut to desk top.
Click on the Icon and start the application. An Terminal Window will open automatically and now
the PC is ready to receive the data from PICO pH /Conductivity/Ion meters.
Page 73 of 76
Chapter 6
Service &
Warranty
Page 74 of 76
TERMS OF WARRANTY
LABINDIA ANALYTICAL INSTRUMENTS PVT LTD., hereinafter known as Manufacturer, warrants all their
products manufactured and sold by them or through their authorized distributors, to be free of any defect in
material and workmanship, subject to Manufacturer’s inspection by their authorized personnel within the
following terms and conditions: -
1) Period of Warranty: Free Warranty period is applicable only for those cases where full payment is
received as follows:
(1) 18 months if the full payment is received against supply and presentation of documents.
(2) 15 months if the full payment is received within 30 days of supply irrespective of installation.
(3) 12 months from the date of installation for the cases where full payment is not received till installation.
However, in this case, period between the date of installation and date of receipt of full payment will be
deducted from 12 months and reduced period of Warranty will only be applicable.
2) Scope: (a) Any part found defective by reason of faulty workmanship or defective material
(fair wear & tear expected) will be repaired or replaced free of charge.
(b) All pH/Ion/Conductivity instruments are to be returned freight to pay basis to us on following
address*. We after repair will send them to the customer freight paid. Since these are low
priced items, offering free service at buyer’s place by deputing our Engineer is not feasible.
You are therefore requested to co-operate with this request.
In case you need repair at your place we will depute our Engineer provided you agree to
st nd
pay 1 or 2 A/C return railway fare, hotel & incidental expenses, etc. of our Engineer. In
this case please note that we will not charge you for our Engineer’s time, etc. This payment
will be in advance and visit duration will be only for one day. On your request we will be
happy to furnish you our quotation for this repair.
Page 75 of 76
Sales & Service Centers
Head Office Delhi Office Gurgaon Office
201, Nand Chambers G-4, Pal Mohan Sadan Plot No. 372, Udyog Vihar Phase II
L. B. S. Marg, Nr. Vandana Cinema 26/32, East Patel Nagar Gurgaon – 122 015,
Thane (W) – 400 602 New Delhi – 110 008 Haryana
Page 76 of 76