Total Inhibitor Content (TBC) of Light Hydrocarbons: Standard Test Method For
Total Inhibitor Content (TBC) of Light Hydrocarbons: Standard Test Method For
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
D 1157 – 91 (2004)
swallowed. Avoid contact with eyes; may burn and impair and filter through a rapid, hardened, low-ash paper that has
vision. May be harmful to respiratory system. May produce previously been moistened. Repeat with two more 30-mL
quinones and flammable butylenes on decomposition. Use with portions of water. Combine all filtrates, add from a pipet 5.0 6
adequate ventilation. Store in flammable liquids storage area.) 0.1 mL of standard FeCl3 solution, dilute to 100 mL, and mix
Dissolve 0.63 g of p-tertiary-butylcatechol (95 % minimum well.
purity) in 10 mL of ethanol (95 %) and dilute with water to 100 9.2 Measurement of Sample—After the addition of the
mL in a volumetric flask. When used in place of 100 mL (63 g) FeCl3 reagent, allow the solution to stand for from 5 to 15 min,
of sample, consider 1.00 mL of this solution to be equivalent to then measure the absorbance of the solution by means of a
100 ppm of catechol. This solution is not stable and should be photoelectric photometer, using water as a reference standard.
prepared as needed. NOTE 2—Take care to have the comparison tubes clean and free from
fingerprints and spilled solution. From time to time, the two tubes should
7. Sampling be checked against each other with blank solution in both.
7.1 Supply samples to the laboratory in high-pressure 9.3 Measurement of Sample—Make a “blank” determina-
sample cylinders. Use the procedures described in Practice tion, following the same procedure, but omitting the addition of
D 1265 or similar methods. the sample. Subtract the absorbance of the “blank” from that of
the sample and record the difference as the “net” absorbance.
8. Calibration and Standardization
NOTE 3—A “blank” determination need be made only when first using
8.1 Preparation of Standard Solutions—Prepare a standard- a bottle of freshly prepared FeCl3 solution to ascertain that the reagent was
ization curve showing the relation between the absorbance and made properly.
the catechol content as follows: Make up solutions of known
catechol content by pipetting 0, 1, 2, 3, 4, and 5-mL portions of 10. Calculation
the standard catechol solution (1.00 mL = 100 mg/kg) into 10.1 By use of the calibration curve, convert the net
separate 100-mL volumetric flasks, or stoppered, graduated absorbance obtained to milligrams per kilograms of p-tertiary-
mixing cylinders. Then add enough water to each flask or butylcatechol.
cylinder to make a total volume of approximately 90 mL. Add
11. Precision and Bias
5.0 6 0.1 mL of standard FeCl3 solution, bring the total
volume to 100 mL and mix well. 11.1 Precision—The precision of this test method as ob-
8.2 Measurement of Standards—Five to fifteen minutes tained by statistical examination of interlaboratory test results
after the addition of the FeCl3 reagent, measure the absorbance is as follows:
of the solution by means of a photoelectric photometer, using 11.2 Repeatability—The difference between successive test
water as a reference standard and using light that is predomi- results obtained by the same operator with the same apparatus
nantly blue (425 nm). under constant operating conditions on identical test material
8.3 Preparation of Calibration Curve—Subtract the reading would, in the long run, in the normal and correct operation of
obtained for the zero catechol standard from each of the above the test method exceed the following values only one case in
readings, using the same volumes of FeCl3 solution and water twenty:
and observing the same time limits. Record the difference as p-tertiary butylcatechol
range, mg/kg Repeatability
the respective“ net” absorbance. Assuming 1.0 mL of the
standard catechol solution to be equal to 100 mg/kg of catechol 50 to 500 10
in the butadiene, plot the net absorbance against the amount of
11.3 Reproducibility—The difference between two single
added catechol in mg/kg.
and independent results obtained by different operators work-
NOTE 1—While this test method of plotting a curve is the recognized ing in different laboratories on identical test material would, in
method, it has been found that the blank or zero reading using the FeCl3 the long run, in the normal and correct operation of the test
reagent described in 6.3 does not change, thus enabling, for control work, method exceed the following values only one case in twenty:
the plotting of a curve that can be read directly.
p-tertiary butylcatechol
range, mg/kg Reproducibility
9. Procedure
9.1 Preparation of Sample—Measure 100 6 1 mL of liquid 50 to 500 20
sample (Warning—Extremely flammable gas under pressure. 11.4 Bias—Since there is no accepted reference material
May form explosive peroxides upon exposure to air. Harmful if suitable for determining the bias for the procedure in this test
inhaled. Irritating to eyes, skin, and mucous membranes.) into method for measuring TBC in hydrocarbons, no statement on
a graduate that has been cooled to below − 20°C. Pour the bias is being made.
sample into a 250-mL Erlenmeyer flask and allow the liquid to
evaporate at room temperature behind a shield in a well- 12. Keywords
ventilated hood. A steam bath may be used to complete the 12.1 light hydrocarbons; oxidation inhibitor; p-tertiary bu-
evaporation. Add 30 mL of water to the flask, stopper, shake, tylcatechol; photometer
2
D 1157 – 91 (2004)
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).