You are on page 1of 14

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: D8310 − 20

Standard Test Method for


Analysis of Target Phenols (TPs) in Soil by Multiple
Reaction Monitoring Liquid Chromatography/Mass
Spectrometry (LC/MS/MS)1
This standard is issued under the fixed designation D8310; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope D7858 Test Method for Determination of Bisphenol A in


1.1 This test method covers analysis of nonylphenol (NP), Soil, Sludge, and Biosolids by Pressurized Fluid Extrac-
nonylphenol monoethoxylate (NP1EO), nonylphenol diethoxy- tion and Analyzed by Liquid Chromatography/Tandem
late (NP2EO), octylphenol (OP), and bisphenol A (BPA), Mass Spectrometry
referred to collectively as target phenols (TPs), in soil, 2.2 Federal Standards:3
sediments, and biosolids by extraction with acetone, filtration, 29 CFR Part 1910 Occupational Safety and Health Standards
dilution with water, and analysis by liquid chromatography/ 40 CFR Part 136, Appendix B Definition and Procedure for
tandem mass spectrometry. The sample extracts are prepared in the Determination of the Method Detection Limit—
a solution of 75 % acetone and 25 % water because TPs have Revision 1.11
an affinity for surfaces and particles that is more pronounced at 3. Terminology
lower concentrations. The range of applicability of the test
3.1 Definitions of Terms Specific to This Standard:
method is shown in Table 1. The method may be extended
3.1.1 batch quality control, n—all the quality control (QC)
outside of these ranges depending on additional performance
samples and standards included in an analytical procedure.
studies not undertaken here.
3.1.2 bisphenol A, BPA—defined in Test Method D7858.
1.2 Units—The values stated in SI units are to be regarded
as the standard. No other units of measurement are included in 3.1.3 bisphenol A (propane-D6), BPA-D6—defined in Test
this standard. Method D7858.
1.3 This standard does not purport to address all of the 3.1.4 2-bromo-4-(1,1,3,3-tetramethylbutyl)phenol, Br-OP,
safety concerns, if any, associated with its use. It is the n—used in this test method as a surrogate.
responsibility of the user of this standard to establish appro- 3.1.4.1 Discussion—2-bromo-4-(1,1,3,3-tetramethylbutyl)
priate safety, health, and environmental practices and deter- phenol is not produced commercially and is not expected to be
mine the applicability of regulatory limitations prior to use. found in the environment. It was reported that compounds in
1.4 This international standard was developed in accor- highly chlorinated, bromide-rich wastewaters could potentially
dance with internationally recognized principles on standard- interfere with the Br-OP surrogate. If this interference is
ization established in the Decision on Principles for the encountered, n-nonylphenol is suggested as an alternative
Development of International Standards, Guides and Recom- surrogate.
mendations issued by the World Trade Organization Technical 3.1.5 nonylphenol, NP, n—mixture of branched
Barriers to Trade (TBT) Committee. p-nonylphenol isomers.
3.1.5.1 Discussion—Commercial NP is produced by the
2. Referenced Documents reaction of phenol with commercial nonene. Commercial
2.1 ASTM Standards:2 nonene is not simply a linear C9H18 alpha olefin; it is a
D1193 Specification for Reagent Water complex mixture of predominantly nine-carbon olefins, called
propylene trimer, containing no linear isomers. This synthesis
1
This test method is under the jurisdiction of ASTM Committee D34 on Waste
results in a mixture of various branched nonylphenol isomers
Management and is the direct responsibility of Subcommittee D34.01.06 on rather than a discrete chemical structure. The branched nonyl
Analytical Methods. group is positioned predominantly in the para position on the
Current edition approved Feb. 1, 2020. Published February 2020. DOI: 10.1520/ phenol ring.
D8310-20.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
3
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Available from U.S. Government Printing Office, Superintendent of
Standards volume information, refer to the standard’s Document Summary page on Documents, 732 N. Capitol St., NW, Washington, DC 20401-0001, http://
the ASTM website. www.access.gpo.gov.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D8310 − 20
TABLE 1 Tested Method Parameters of the Standard 3.2.19 MSP—method specific parameter
ESI MDL Reporting 3.2.20 NA—not available
Analyte
Mode (µg/kg)A Range (µg/kg)
Bisphenol A Negative 15.5 100–2500 3.2.21 NP1EO—nonylphenol monoethoxylate
Octylphenol Negative 44.2 200–5000
Nonylphenol Negative 30.4 100–2500 3.2.22 NP2EO—nonylphenol diethoxylate
Nonylphenol Monoethoxylate Positive 931.2 3000–45 000
Nonylphenol Diethoxylate Positive 7.4 100–2500
3.2.23 OP—octylphenol
A
MDL is calculated based upon nine spiked samples. 3.2.24 P&A—precision and accuracy
3.2.25 PPB—parts per billion
3.2.26 PPM—parts per million
3.1.6 nonylphenol diethoxylate, NP2EO, n—branched non- 3.2.27 PPT—parts per trillion
ylphenol diethoxylate. 3.2.28 PTFE—polytetrafluoroethylene
3.1.7 nonylphenol monoethoxylate, NP1EO, n—branched 3.2.29 PVDF—poly-vinylidene dichloride
nonylphenol monoethoxylate. 3.2.30 QA—quality assurance
3.1.8 normal nonylphenol, n-NP, n—normal straight chain 3.2.31 QC—quality control
nonylphenol. 3.2.32 QMP—quality management plan
3.1.8.1 Discussion—n-NP is used in this test method as a
surrogate. It is not produced commercially and is not expected 3.2.33 REC—percent recovery
to be found in the environment. 3.2.34 RL—reporting limit
3.1.9 normal nonylphenol diethoxylate, n-NP2EO, 3.2.35 RLCS—reporting limit check sample
n—normal straight chain nonylphenol diethoxylate. 3.2.36 RSD—relative standard deviation
3.1.9.1 Discussion—n-NP2EO is used in this test method as 3.2.37 RT—retention time
a surrogate. It is not produced commercially and is not
expected to be found in the environment. 3.2.38 RTS—retention time shift
3.1.10 octylphenol, OP, n—produced by the reaction of 3.2.39 SOP—standard operating procedure
phenol and diisobutylene to produce predominantly the 4-(1, 3.2.40 SRM—single reaction monitoring
1,3,3-tetramethylbutyl)phenol isomer. 3.2.41 SS—surrogate standard
3.1.11 reporting limit check sample, RLCS, n—this sample 3.2.42 TC—target compound
verifies that if the analyte was present at the reporting limit, it
3.2.43 TCL—target compound limit
would be confidently identified.
3.2.44 TP—target phenols
3.1.12 target phenols, TPs, n—in this test method, NP,
NP1EO, NP2EO, OP, and BPA, collectively. 3.2.45 UPLC—ultra performance liquid chromatography
3.2 Abbreviations: 3.2.46 VOA—volatile organic analysis
3.2.1 ADOC—analyst demonstration of capability
4. Summary of Test Method
3.2.2 BPA—bisphenol A
4.1 A sample (~2 g) is transferred to a VOA vial and spiked
3.2.3 Br-OP—2-bromo-4-(1,1,3,3-tetramethylbutyl)phenol
with surrogates (all samples) and TPs (laboratory control and
3.2.4 CAS—chemical abstract service matrix spike samples) and then extracted with 7.5 mL of
3.2.5 CCC—continuing calibration check acetone by tumbling on a rotator for 2 h. Any device may be
3.2.6 DL—detection limit used that mixes the sample; a rotator device was chosen in this
case because it inverts the sample, allowing the soil to be in a
3.2.7 EPA—U.S. Environmental Protection Agency
constant fluid motion and not clumped. The samples are
3.2.8 IC—initial calibration centrifuged at 1900 rpm for 10 min and then filtered through an
3.2.9 IDOC—initial demonstration of capability Acrodisc GxF/0.2 µm PVDF membrane syringe-driven filter
3.2.10 LC—liquid chromatography unit. Of Specification D1193 Type 1 water, 2.5 mL is added to
the filtered extract and then analyzed by LC/MS/MS. All
3.2.11 LCS—laboratory control sample
concentrations reported, only to the RL, using this method are
3.2.12 LCSD—laboratory control sample duplicate based upon a dry-weight basis.
3.2.13 LIMS—relational laboratory information manage- 4.2 The TCs are identified by comparing the SRM transition
ment system and RT. BPA has a confirmatory SRM transition also. The
3.2.14 MDL—method detection limit confirmatory SRM transition will be correlated to the known
3.2.15 MI—matrix interference standard SRM transition for identification of BPA (Table 2).
The RT for the analyte of interest shall also agree with the RT
3.2.16 MRM—multiple reaction monitoring
of the mid-level standard by 65 %. The TC is quantitated using
3.2.17 MS—mass spectrometry the SRM transition of the TC using external calibration. As an
3.2.18 MS/MSD—matrix spike/matrix spike duplicate additional QC measure, non-labeled surrogates (listed in 8.2)

2
D8310 − 20
TABLE 2 Variable Mass Spectrometer Parameters
Retention Time, SRM Mass Transition Cone Voltage, Collision Energy,
Analyte ESI Mode
min (Parent > Product) V eV
BPA negative 5.58 227.1 > 211.9 40 18
BPA confirmatoryA negative 5.58 227.1 > 132.7 40 25
OP negative 8.53 205.1 > 132.7 40 24
NP negative 9.70 219.3 > 133.1 40 28
NP1EO positive 9.71 282.3 > 126.9 20 9
NP2EO positive 9.65 326.3 > 182.9 25 11
BPA-D6 (surrogate) negative 5.57 233.3 > 214.9 40 19
BPA-D6 confirmatoryA (surrogate) negative 5.57 233.3 > 137.8 40 25
Br-OP (surrogate) negative 9.56 283.1 > 78.6 40 25
n-NP (surrogate) negative 10.71 219.3 > 105.6 40 20
n-NP2EO (surrogate) positive 10.69 326.4 > 88.8 25 15
A
Confirmatory transitions are optional but should be included for added qualitative information.

recoveries are monitored; the percent recovery of each should laboratory contribution to measurement uncertainty of this test
fall within the control limits of the method. The final report method for each analyte is listed in Table 3. These values are
issued for each sample lists concentration of TPs and derived from P&A samples from the initial IDOC study for this
surrogates, if detected, or non-detect at the RL if not detected, test method. The uncertainty will be greater near the RL and
in microgram/kilogram on a dry-weight basis. much greater near the DL. Also, uncertainty estimated based on
variability in LCS recovery is conservative because some
5. Significance and Use sources of variability are not included, such as subsample
5.1 This test method developed for the analysis of TPs in variability and matrix analyte recovery. This SOP covers
soil and sediment samples is based upon an LC/MS/MS multiple soil matrices and the uncertainty among the various
analysis. Any type of coupled liquid chromatography/mass matrices is variable.
spectrometry system may be used that meets the study objec-
tives of the individual project. These may include, but are not 6. Interferences
limited to: trap, single quadrupoles, time-of-flight, high 6.1 Method interferences may be caused by contaminants in
resolution, and others not mentioned here. solvents, reagents, glassware, LC vials/caps, disposable
5.2 The MDL and reporting range for TPs are listed in Table pipettes, and other apparatus that lead to discrete artifacts or
1. This SOP has been tested on Ottawa sand, four ASTM soil elevated baselines in the selected ion current profiles. The
types, biosolid sample, and one commercial soil. The P&A QC presence and magnitude of method interferences are deter-
acceptance criteria are listed in Table 3. Tables 4-17 display the mined by analysis of solvent and laboratory blanks.
TC and surrogate recoveries in the various soil types. 40 CFR 6.2 Matrix interferences may be caused by contaminants
Part 136, Appendix B was used as a guide to determine the from the sample, sampling devices, or storage containers. The
MDLs. The 40 CFR Part 136 MDL criteria were not met for extent of matrix interferences will vary considerably from
NP2EO; this does not affect the method because the SOP only sample source to sample source depending on variations of the
reports to the RL and is not a regulatory method. All site sample matrix. The analysis of matrix spikes is critical for
sample results are not reported below the RL using this determining the impact of matrix interferences.
method. RLCS concentrations may be reported below the RL 6.3 Warnings:
because they are spiked at or near the RL. 6.3.1 All reagents and solvents should be of pesticide
5.3 The RL for a specific soil sample may differ from that residue purity or higher to minimize interference problems,
listed depending on the nature of the interferences in the preferably LC/MS grade.
sample matrix. Variability in historical LCS spike recovery 6.3.2 Contaminants have been found in improperly cleaned
may be used to estimate uncertainty. The estimate of minimum glassware and glass syringes. TPs stick to surfaces if the

TABLE 3 QC Acceptance Criteria and UncertaintyA


Standard Deviation
Average Recovery, Number of Lower Control Limit Upper Control Limit Uncertainty (95 %
Parameter of Percentage
% Replicates, n (LCL), % (UCL), % Confidence Interval)
Recovery
BPA 92.6 4.5 6 70 130 4.7
OP 88.4 6.5 6 60 130 6.9
NP 92.9 3.4 6 70 130 3.6
NP1EO 98.4 7.0 6 70 130 7.3
NP2EO 96.1 4.5 6 70 130 4.7
BPA-D6 (surrogate) 91.1 3.2 8 70 130 2.7
Br-OP (surrogate) 87.9 4.7 8 70 130 3.9
n-NP (surrogate) 87.4 2.6 8 70 130 2.2
n-NP2EO (surrogate) 93.6 6.1 8 70 130 5.1
A
Uncertainty calculation based upon 95 % confidence interval and a two-tailed Student t distribution.
Uncertainty = Student t Value [(standard deviation) / (number of LCS)½].

3
D8310 − 20
TABLE 4 P&A Study for TPs in Ottawa Sand (Target Spike Recoveries)
P&A Data (625 µg/kg spike for BPA; 1250 µg/kg for OP, NP, and NP2EO; and 18 750 µg/kg for NP1EO)
Sample
BPA OP NP NP1EO NP2EO
MB1 <RL <RL <RL <RL <RL
MB2 <RL <RL <RL <RL <RL
P&A1 555.0 1020.0 1120.0 18 500.0 1150.0
P&A2 597.0 1070.0 1190.0 18 600.0 1160.0
P&A3 554.0 1160.0 1110.0 17 400.0 1160.0
P&A4 567.0 1010.0 1150.0 16 600.0 1200.0
P&A5 626.0 1180.0 1220.0 19 400.0 1280.0
P&A6 572.0 1190.0 1180.0 20 200.0 1260.0
Average Recovery, µg/kg 578.5 1105.0 1161.7 18 450.0 1201.7
% Average Recovery 92.6 88.4 92.9 98.4 96.1
Standard Deviation 28.0 81.7 42.6 1305.0 56.0
RSD, % 4.8 7.4 3.7 7.1 4.7

TABLE 5 P&A Study for TPs in Ottawa Sand (Surrogate Recoveries)


P&A Data (625 µg/kg Spike)
Sample
BPA-D6 Br-OP n-NP n-NP2EO
MB1 548 500 557 539
MB2 590 518 555 552
P&A1 538 567 569 564
P&A2 569 567 547 582
P&A3 579 556 515 566
P&A4 556 533 536 597
P&A5 583 581 553 637
P&A6 592 574 540 643
Average Recovery, µg/kg 569.4 549.5 546.5 585.0
% Average Recovery 91.1 87.9 87.4 93.6
Standard Deviation 20.1 29.2 16.4 38.2
RSD, % 3.5 5.3 3.0 6.5

TABLE 6 P&A Study in ASTM CL-1 Soil (Target Compound Recoveries)


P&A for ASTM CL-1 Soil (Lean Clay, Red Bucket) (625 µg/kg Spike for BPA; 1250 µg/kg for OP, NP, and NP2EO; and
Sample 18 750 µg/kg for NP1EO)
BPA OP NP NP1EO NP2EO
MB1 <RL <RL <RL <RL <RL
MB2 <RL <RL <RL <RL <RL
P&A1 553.0 1173.7 1231.1 17 936.0 1239.4
P&A2 586.3 1259.1 1300.3 19 901.6 1271.3
P&A3 564.7 1209.6 1188.7 18 313.8 1225.9
P&A4 587.3 1224.9 1226.7 17 870.8 1233.9
P&A5 609.7 1201.5 1239.1 17 757.6 1254.0
P&A6 553.6 1119.0 1179.7 18 965.0 1195.1
Mean Recovery (µg/kg dry weight) 575.8 1198.0 1227.6 18 457.5 1236.6
% Mean Recovery 92.1 95.8 98.2 98.4 98.9
Standard Deviation 22.5 47.8 43.0 833.0 25.9
RSD, % 3.9 4.0 3.5 4.5 2.1

TABLE 7 Surrogate Recoveries for P&A Study in ASTM CL-1 Soil


ASTM CL-1 Soil (Lean Clay, Red Bucket), 625 µg/kg Spike
Sample
BPA-D6 Br-OP n-NP n-NP2EO
MB1 627.5 571.5 602.3 603.2
MB2 599.6 535.9 618.2 592.2
P&A1 549.4 594.1 595.1 637.6
P&A2 628.8 608.5 579.3 669.0
P&A3 600.6 579.6 557.8 633.0
P&A4 605.5 597.4 569.7 645.5
P&A5 601.3 602.6 636.9 660.3
P&A6 589.8 601.7 589.8 622.5
Mean Recovery (µg/kg dry weight) 595.9 597.3 588.1 644.6
% Mean Recovery 95.3 95.6 94.1 103.1
Standard Deviation 26.2 10.0 27.4 17.4
RSD, % 4.4 1.7 4.7 2.7

glassware is not properly cleaned and rinsed with solvent, such trations affecting the analysis may be found. All of these
as acetone, 2-propanol, and acetonitrile, and low µg/L concen-

4
D8310 − 20
TABLE 8 P&A Study in ASTM CH-1 Soil (Target Compound Recoveries)
P&A Data for ASTM CH-1 Soil (Fat Clay, White Bucket) (625 µg/kg Spike for BPA; 1250 µg/kg for OP, NP, and NP2EO; and
Sample 18 750 µg/kg for NP1EO)
BPA OP NP NP1EO NP2EO
MB1 <RL <RL <RL <RL <RL
MB2 <RL <RL <RL <RL <RL
P&A1 622.3 1245.6 1220.7 19 408.3 1272.0
P&A2 639.1 1173.9 1192.9 19 910.7 1267.6
P&A3 666.6 1236.3 1260.1 20 325.8 1301.7
P&A4 604.4 1178.0 1200.8 19 328.7 1232.3
P&A5 644.1 1164.7 1327.7 19 230.7 1325.0
P&A6 664.5 1338.0 1268.8 19 386.1 1350.5
Mean Recovery (µg/kg dry weight) 640.2 1222.7 1245.1 19 598.4 1291.5
% Mean Recovery 102.4 97.8 99.6 104.5 103.3
Standard Deviation 24.1 65.9 50.8 428.0 42.8
RSD, % 3.8 5.4 4.1 2.2 3.3

TABLE 9 Surrogate Recoveries for P&A Study in ASTM CH-1 Soil


ASTM CH-1 Soil (Fat Clay, White Bucket), 625 µg/kg Spike
Sample
BPA-D6 Br-OP n-NP n-NP2EO
MB1 645.2 537.0 603.7 571.1
MB2 635.6 572.0 645.5 591.9
P&A1 634.1 602.6 553.9 660.0
P&A2 609.7 593.7 569.1 644.4
P&A3 632.4 650.5 628.1 686.3
P&A4 623.4 590.2 577.3 631.9
P&A5 661.0 645.6 659.3 660.9
P&A6 672.2 649.6 686.4 702.2
Mean Recovery (µg/kg dry weight) 638.8 622.0 612.3 664.3
% Mean Recovery 102.2 99.5 98.0 106.3
Standard Deviation 23.5 29.4 53.8 26.0
RSD, % 3.7 4.7 8.8 3.9

TABLE 10 P&A Study in ASTM SP-1 Soil (Target Compound Recoveries)


P&A Data (625 µg/kg Spike for BPA; 1250 µg/kg for OP, NP, and NP2EO; and 18 750 µg/kg for NP1EO)
Sample
BPA OP NP NP1EO NP2EO
MB1 <RL <RL <RL <RL <RL
MB2 <RL <RL <RL <RL <RL
P&A1 611.5 1231.4 1205.7 17 023.3 1205.9
P&A2 631.0 1234.1 1252.3 18 426.6 1229.1
P&A3 574.3 1067.6 1140.7 17 788.8 1153.3
P&A4 626.9 1262.1 1211.8 18 010.4 1232.4
P&A5 586.7 1213.2 1200.2 17 549.9 1230.1
P&A6 578.4 1230.4 1184.4 16 480.0 1113.1
Mean Recovery (µg/kg dry weight) 601.5 1206.5 1199.2 17 546.5 1194.0
% Mean Recovery 96.2 96.5 95.9 93.6 95.5
Standard Deviation 24.9 69.8 36.5 701.4 49.7
RSD, % 4.1 5.8 3.0 4.0 4.2

TABLE 11 Surrogate Recoveries for P&A Study in ASTM SP-1 Soil


ASTM SP-1 Soil (Sand, Yellow Bucket), 625 µg/kg Spike
Sample
BPA-D6 Br-OP n-NP n-NP2EO
MB1 568.1 518.1 605.1 533.2
MB2 610.1 548.3 545.4 553.0
P&A1 629.8 589.7 567.6 622.4
P&A2 608.8 588.7 596.6 601.7
P&A3 543.8 534.5 513.0 594.4
P&A4 576.1 589.2 593.4 601.3
P&A5 559.4 616.7 590.1 629.7
P&A6 552.5 532.5 548.5 551.9
Mean Recovery (µg/kg dry weight) 578.4 575.2 568.2 600.2
% Mean Recovery 92.5 92.0 90.9 96.0
Standard Deviation 34.1 34.0 32.7 27.3
RSD, % 5.9 5.9 5.8 4.6

materials are routinely demonstrated to be free from interfer- conditions as the samples. If found, measures should be taken
ences by analyzing laboratory reagent blanks under the same to remove the contamination or data should be qualified.

5
D8310 − 20
TABLE 12 P&A Study in ASTM ML-1 Soil (Target Compound Recoveries)
P&A Data for ASTM ML-1 Soil (Silt, Green Bucket) (625 µg/kg Spike for BPA; 1250 µg/kg for OP, NP, and NP2EO; and
Sample 18 750 µg/kg for NP1EO)
BPA OP NP NP1EO NP2EO
MB1 <RL <RL <RL <RL <RL
MB2 <RL <RL <RL <RL <RL
P&A1 559.5 1117.4 1076.1 16 702.3 1124.2
P&A2 547.9 1179.6 1091.9 17 457.8 1106.4
P&A3 573.0 1148.2 1146.7 17 540.1 1152.7
P&A4 550.3 1193.2 1108.3 16 291.8 1111.0
P&A5 549.0 1151.9 1121.4 16 871.6 1131.9
P&A6 568.6 1121.2 1102.0 16 855.4 1139.1
Mean Recovery (µg/kg dry weight) 558.0 1151.9 1107.7 16 953.2 1127.6
% Mean Recovery 89.3 92.2 88.6 90.4 90.2
Standard Deviation 10.8 30.4 24.4 472.4 17.4
RSD, % 1.9 2.6 2.2 2.8 1.5

TABLE 13 Surrogate Recoveries for P&A Study in ASTM ML-1 Soil


ASTM ML-1 Soil (Silt, Green Bucket), 625 µg/kg Spike
Sample
BPA-D6 Br-OP n-NP n-NP2EO
MB1 555.6 479.8 513.4 522.6
MB2 504.3 499.8 492.9 525.8
P&A1 530.0 522.7 512.2 571.7
P&A2 518.3 538.0 516.1 563.2
P&A3 553.1 551.1 512.9 580.9
P&A4 527.9 519.3 532.0 548.6
P&A5 527.2 537.0 554.7 551.0
P&A6 551.5 522.6 524.5 588.0
Mean Recovery (µg/kg dry weight) 534.7 531.8 525.4 567.2
% Mean Recovery 85.5 85.1 84.1 90.8
Standard Deviation 14.2 12.4 16.2 15.9
RSD, % 2.7 2.3 3.1 2.8

TABLE 14 P&A Study in Fairfield Biosolid Samples (Target Compound Recoveries)


P&A Data for Fairfield Biosolid Samples (625 µg/kg Spike for BPA; 1250 µg/kg for OP, NP, and NP2EO; and 18 750 µg/kg
Sample for NP1EO)
BPA OPA NPA NP1EO NP2EO
MB1 <RL 388.9 10 677.6 <RL <RL
MB2 <RL 224.7 6843.1 <RL <RL
P&A1 457.7 821.8 1727.7 13 795.6 931.3
P&A2 514.1 889.4 1329.6 15 307.5 996.4
P&A3 496.8 1002.2 2243.0 15 135.0 1007.5
P&A4 496.0 1007.7 2148.3 15 158.7 1005.0
P&A5 471.7 1114.0 6930.9 14 673.7 982.7
P&A6 456.5 1018.1 3984.5 14 310.2 966.2
Mean Recovery (µg/kg dry weight) 482.1 975.5 3060.7 14 730.1 981.5
% Mean Recovery 77.1 78.0 244.9 78.6 78.5
Standard Deviation 23.6 103.7 2102.7 588.8 29.0
RSD, % 4.9 10.6 68.7 4.0 3.0
A
P&A values are after subtraction of average of MB if $RL.

6.3.3 Glass syringes are used with this test method. A 7.1.1 Liquid Chromatograph (LC) System—An ultra-
thoroughly cleaned 10 or 20-mL hypodermic glass syringe performance LC (UPLC) with flow-through needle design.
with a PVDF filter unit is used to filter samples and has been 7.1.2 Analytical Column—An analytical column that will
shown to perform well. achieve adequate results that meet or exceed this test method.4
6.3.4 Nonylphenol was found in polypropylene. The filter 7.1.3 Mass Spectrometer (MS) System—A triple quadrupole
units shall be rinsed with at least 10 mL of acetonitrile and mass spectrometer is used. A mass spectrometer capable of
10 mL of methanol before use to remove nonylphenol. MRM analysis with fast enough cycle time to obtain at least ten
6.3.5 The procedure described in 7.3.8 should be followed scans over a peak is needed with adequate sensitivity.
to make glassware free from interferences. Alkylphenol deter- 7.1.4 Data Backup Device—A data archival unit to archive
gents shall not be used. Detergent may contain TPs, precursor, data.
or breakdown products and these shall be avoided.
4
7. Apparatus A Waters Acquity UPLC® BEH C18, 2.1 × 100 mm and 1.7 µm particle size
was used to develop this test method. Any column may be used that meets or
7.1 Equipment: exceeds the performance of this test method.

6
D8310 − 20
TABLE 15 Surrogate Recoveries for P&A Study in Fairfield Biosolid Samples
Fairfield Biosolid Samples, 625 µg/kg Spike
Sample
BPA-D6 Br-OP n-NP n-NP2EO
MB1 447.1 387.8 176.8 431.0
MB2 466.9 417.6 214.5 449.7
P&A1 413.3 391.4 179.6 450.8
P&A2 432.4 420.8 190.2 486.9
P&A3 456.1 419.7 177.0 496.5
P&A4 460.1 449.1 195.1 507.0
P&A5 433.6 408.5 151.4 468.0
P&A6 436.2 425.1 165.8 492.2
Mean Recovery (µg/kg dry weight) 438.6 419.1 176.5 483.5
% Mean Recovery 70.2 67.1 28.2 77.4
Standard Deviation 17.2 19.1 16.1 20.6
RSD, % 3.9 4.6 9.1 4.3

TABLE 16 P&A Study in Commercial Soil (Milorganite)


P&A Data for Milorganite (625 µg/kg Spike for BPA; 1250 µg/kg for OP, NP, and NP2EO; and 18 750 µg ⁄kg for NP1EO)
Sample
BPAA OPA NPA NP1EOA NP2EOA
MB1 109.7 194.2 3460.9 7875.8 4971.7
MB2 89.3 196.5 3512.1 7637.2 4587.9
P&A1 454.9 801.7 729.7 17 038.9 1832.0
P&A2 432.1 783.8 590.0 15 539.7 1452.2
P&A3 422.7 785.6 555.3 15 936.5 1453.7
P&A4 428.5 836.5 713.5 15 928.4 1550.2
P&A5 402.9 790.4 569.2 16 070.2 1306.2
P&A6 433.1 843.0 756.4 16 216.7 1669.0
Mean Recovery (µg/kg dry weight) 429.0 806.8 652.3 16 121.7 1543.9
% Mean Recovery 68.6 64.5 52.2 86.0 123.5
Standard Deviation 16.9 26.3 90.3 502.6 185.3
RSD, % 3.9 3.3 13.8 3.1 12.0
A
P&A values are after subtraction of average of MB if $RL.

TABLE 17 Surrogate Recoveries for Precision and Accuracy Study in Commercial Soil (Milorganite)
Milorganite, 625 µg/kg Spike
Sample
BPA-D6 Br-OP n-NP n-NP2EO
MB1 402.2 324.1 <RLA 508.3
MB2 425.1 335.5 <RL 512.7
P&A1 431.0 341.7 <RL 557.7
P&A2 363.5 325.9 <RL 548.5
P&A3 405.8 330.3 <RL 550.4
P&A4 354.4 312.1 <RL 545.0
P&A5 371.3 309.4 <RL 550.1
P&A6 404.9 329.1 <RL 564.9
Mean Recovery (µg/kg dry weight) 388.5 324.8 No recovery 552.8
% Mean Recovery 62.2 52.0 NA 88.4
Standard Deviation 29.8 12.1 NA 7.2
RSD, % 7.7 3.7 NA 1.3
A
<RL = below reporting limit.

7.1.5 Data System—Software shall be interfaced to the 7.3.3 Class A volumetric glassware.
LC/MS/MS that allows the continuous acquisition and storage 7.3.4 VOA Vials—20, 40, or 60 mL.
on machine-readable media of all mass spectra obtained 7.3.5 GxF/0.2 µm PVDF Membrane Syringe-Driven Filter
throughout the duration of the chromatographic program. Unit—The filters are washed with 10 mL acetonitrile followed
Software is used for all data quantitation. by 10 mL methanol before use.
7.2 Calibrated Support Equipment: 7.3.6 Syringe—Use 10 to 25-mL filter-adaptable glass sy-
7.2.1 Graduated Syringes—Use 10, 25, 50, 100, 250, 500, ringe with luer lock.
and 1000-µL graduated syringes. 7.3.7 Disposable glass pipettes. (Warning—Syringes shall
7.2.2 Analytical Balance (Certified Annually)—Use an ana- be cleaned with copious amounts of warm tap water, distilled
lytical balance accurate to 60.1 % of sample mass with Class water, and thoroughly rinsed with acetonitrile and methanol
3 or 4 weights to verify calibration. between uses.)
7.3 Glassware and Miscellaneous Supplies: 7.3.8 Glassware Cleaning Instructions—All glassware is
7.3.1 Vials—Use 2-mL autosampler vials with pre-slit washed in hot water (above 50 °C) with detergent, rinsed in hot
PTFE/silicone septa or equivalent. water, and rinsed with distilled water. The glassware is then
7.3.2 Gases—Ultra-pure argon and nitrogen. dried in an oven for up to 1 h (except volumetric glassware,

7
D8310 − 20
which is air dried). All glassware is subsequently rinsed with – 8.3.4.4. The calibration levels (prepared from the high-level
an organic solvent such as acetone, methanol, and acetonitrile. calibration stock), laboratory control samples, matrix spike
(Warning—Detergents containing alkylphenolic compounds samples, and duplicates are made for each batch of samples.
shall not be used.) They are usually prepared and analyzed immediately. A CCC
standard is also prepared for each batch of samples.
8. Reagents and Materials 8.3.4.1 TP Surrogate Spiking Solution—A surrogate stan-
8.1 Purity of Reagents—Reagent-grade chemicals shall be dard solution containing Br-OP, n-NP, BPA-D6, and n-NP2EO
used in all tests. Unless otherwise indicated, it is intended that is added to each 2-g soil sample. Br-OP and n-NP are used as
all reagents conform to the specifications of the Committee on the surrogates for NP and OP, BPA-D6 is used as a surrogate for
Analytical Reagents of the American Chemical Society, where BPA, and n-NP2EO is used as a surrogate for NP1EO and
such specifications are available.5 Other grades may be used, NP2EO. A stock surrogate spiking solution is prepared in
provided it is first ascertained that the reagent is of sufficiently methanol at 12.5 mg/L for Br-OP, n-NP, BPA-D6, and
high purity to permit its use without lessening the accuracy of n-NP2EO. The surrogates are added to each sample to achieve
the determination. a concentration of 625 µg/L (that is, 100 µL of a 12.5-mg/L
8.2 Solvents and Reagents: methanol solution containing Br-OP, n-NP, BPA-D6, and
8.2.1 Acetonitrile. n-NP2EO is added to a 2-g soil sample).
8.2.2 Water—HPLC mass spectrometry pesticide quality. 8.3.4.2 MS/MSD and LCS/LCSD Spiking Solution (Target
8.2.2.1 Purity of Water—Unless otherwise indicated, refer- Spike Solution)—Each ~2-g sample, approximately 2 g and
ences to water shall be understood to mean reagent water as weighed to the hundredths, MS/MSD or LCS/LCSD sample is
defined by Type I of Specification D1193. spiked with TPs containing BPA at 12.5 mg/L, NP, OP and
8.2.3 Methanol. NP2EO at 25 mg/L, and NP1EO at 375 mg/L to achieve a
8.2.4 Isopropyl alcohol. concentration of 1250 µg/L for OP, NP, and NP2EO, 625 µg/L
8.2.5 Ammonium acetate. for BPA, and 18 750 µg/L for NP1EO (that is, 100 µL of a
8.2.6 Acetone. methanol solution containing BPA at 12.5 mg/L, NP, OP, and
8.2.7 Nonylphenol monoethoxylate. NP2EO at 25 mg/L and NP1EO at 375 mg/L is added to a 2-g
8.2.8 Nonylphenol diethoxylate. soil sample).
8.2.9 Nonylphenol. 8.3.4.3 RL Check Spiking Solution—The RLCS is prepared
8.2.10 Octylphenol. by spiking the RLCS with the RL check standard near RL. A
8.2.11 n-Nonylphenol diethoxylate. 50-µL volume of a methanolic RL check standard containing
8.2.12 n-Nonylphenol. BPA at 2 mg/L; NP, OP, and NP2EO at 4 mg/L; and NP1EO at
8.2.13 2-bromo-4-(1,1,3,3-tetramethylbutyl)phenol. 60 mg/L is added to RLCS to achieve a concentration of 50
8.2.14 Bisphenol A. µg/L for BPA; 100 µg/L for OP, NP, and NP2EO; and 1500
8.2.15 Bisphenol A (propane-D6). µg/L for NP1EO.
8.3.4.4 Calibration Standards—Calibration stock standard
8.3 Reagents and Standard Preparation: Solution A is prepared from target spiking solutions containing
8.3.1 All calibration standard preparations shall be re- BPA at 12.5 mg/L; NP, OP, and NP2EO at 25 mg/L; and
corded. All calibration solutions are prepared with acetone:wa- NP1EO at 375 mg/L and surrogate spiking solutions containing
ter (75:25) unless otherwise stated. The concentrated stock Br-OP, n-NP, BPA-D6; and n-NP2EO at 12.5 mg/L. Add 1000
standard concentration can vary when preparing from neat µL of the target spike solution and surrogate spiking solution to
material, usually between 50 to 100 mg/L of each TP in a 25-mL volumetric flask and diluted to 25 mL volume with
methanol. 75:25 (acetone:water) (Solution B). Stock standard Solution A
8.3.2 Expiration time of a prepared standard is six months (Level 8, Tables 18 and 19) containing BPA and each surrogate
from the time prepared. The standards can be used for more at 500 µg ⁄L; OP, NP, and NP2EO at 1000 µg/L; and NP1EO at
than six months if they fall within 630 % from the calibration 15 000 µg ⁄L is diluted to prepare Levels 1 through 7 as shown
standard that is less than six months old. Presently, there is no in Tables 18 and 19. All calibration standards should contain
holding time study for this standard and the six months is an 75:25 acetone to water in 2-mL LC vials. The CCC is a
estimate with no laboratory data supporting this time period. mid-level calibration standard.
8.3.3 Label all standards and verify the correct grade of
solvents were used. Traceability of standards is established 9. Hazards
using the manufacturer’s specifications provided at the time of 9.1 Users of this test method should operate a formal safety
purchase. program.
8.3.4 The specific instructions for the preparation of stock
9.2 Warnings—Health Hazards—The toxicity and carcino-
standards, spiking solutions, and QC batch are listed in 8.3.4.1
genicity of each reagent used in this test method have not been
precisely defined; however, each chemical compound is treated
5
Reagent Chemicals, American Chemical Society Specifications, American as a health hazard. From this viewpoint, exposure to these
Chemical Society, Washington, DC. For suggestions on the testing of reagents not
listed by the American Chemical Society, see Analar Standards for Laboratory
chemicals should be reduced to the lowest possible level and
Chemicals, BDH Ltd., Poole, Dorset, U.K. and the United States Pharmacopeia and appropriate PPE used. Review SDSs for specific physical and
National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, MD. health and hazards, including appropriate PPE to be used.

8
D8310 − 20
TABLE 18 Concentrations of Calibration Standards, µg/L
Analyte/Surrogate LV 1 LV 2 LV 3 LV 4 LV 5 LV 6 LV 7 LV 8
BPAA 10 20 50 100 200 300 400 500
OPA 20 40 100 200 400 600 800 1000
NP 20 40 100 200 400 600 800 1000
NP1EOA 300 600 1500 3000 6000 9000 12 000 15 000
NP2EO 20 40 100 200 400 600 800 1000
BPA-D6 (surrogate) 10 20 50 100 200 300 400 500
Br-OP (surrogate) 10 20 50 100 200 300 400 500
n-NP (surrogate) 10 20 50 100 200 300 400 500
n-NP2EO (surrogate) 10 20 50 100 200 300 400 500
A
RL is set at Level 2 concentration; all others at Level 1 because of MDL determination. The Level 1 concentration did display adequate sensitivity for all analytes
(signal/noise ratio > 5).

TABLE 19 Preparation of Calibration Standards


SolutionA LV 1 LV 2 LV 3 LV 4 LV 5 LV 6 LV 7 LV 8
A 20 µL 40 µL 100 µL 200 µL 400 µL 600 µL 800 µL 1000 µL
B 980 µL 960 µL 900 µL 800 µL 600 µL 400 µL 200 µL 0
A
Solution A—Level 8 stock solution prepared according to Section 8 and at Table 18 concentrations.
Solution B—75 % acetone to 25 % water.

9.3 Waste Handling and Pollution Prevention: 11.1.2 Laboratory Control Sample/Laboratory Control
9.3.1 A tag identifying the contents of the carboy is placed Sample Duplicate—A ~2-g amount of Ottawa sand is added to
on the waste container. Attach the appropriate chemical waste each of two 40-mL VOA vials. The samples are spiked with
label with the date of beginning collection before using the 100 µL of a target spike solution containing BPA at 12.5 mg/L;
container. NP, OP, and NP2EO at 25 mg/L; and NP1EO at 375 mg/L and
9.3.2 Report all major spills according to your laboratory then taken through the sample preparation Step II in 11.2.
chemical hygiene plan. 11.1.3 RL Check—A ~2-g amount of Ottawa sand is added
9.3.3 All used vials shall be placed in the labeled carboy for to a 40-mL VOA vial. The sample is spiked with 50 µL of a RL
disposal. check solution containing BPA at 2 mg/L; NP, OP, and NP2EO
9.3.4 Analyses performed by this test method will generate at 4 mg/L; and NP1EO at 60 mg/L and then taken through the
solid and liquid wastes that are potentially hazardous. sample preparation Step II in 11.2.
11.1.4 Sample and Sample Duplicate—A ~2-g amount of
10. Sample Handling and Preservation sample is added to each of two 40-mL VOA vials. The samples
10.1 Sample Collection Criteria—Grab samples are col- are taken through the sample preparation Step II in 11.2. The
lected in glass containers with polytetrafluoroethylene (PTFE) sample is mixed in the laboratory with a clean spatula to try
lined caps. Surface binding may bias data. All samples are iced and have a homogeneous subsample taken for analysis.
or refrigerated at ≤6 °C from the time of collection until 11.1.5 Matrix Spike/Matrix Spike Sample Duplicate—A
extraction. ~2-g amount of sample is added to each of two 40-mL VOA
vials. The samples are spiked with 100 µL of a TP target spike
10.2 Sample Preservation and Storage—All samples are
solution containing BPA at 12.5 mg/L; NP, OP, and NP2EO at
iced or refrigerated at ≤6 °C from the time of collection until
25 mg/L; and NP1EO at 375 mg/L and then taken through the
sample extraction. At the laboratory, the samples and extracts
sample preparation Step II in 11.2.
are stored in the refrigerator at ≤6 °C while not being analyzed.
Holding times have not yet been established for these analytes 11.2 Sample Preparation: Step II:
in the various matrices or the extracts. The extracts are 11.2.1 A surrogate standard solution containing Br-OP,
normally analyzed the day of preparation or within seven days n-NP, BPA-D6, and n-NP2EO is added to each sample. Refer to
if multiple dilutions and analysis are required. 8.3.4.1.
11.2.2 After addition of the surrogate, 7.5 mL of acetone is
11. Sample Preparation and Analysis added, tumbled on a rotator for 2 h, and centrifuged at
11.1 Sample Preparation: Step I—Each batch of samples 1900 rpm for 10 min.
(20 or less) shall contain at least a method blank, laboratory 11.2.3 All the samples are filtered through a PVDF filter
control sample and laboratory control sample duplicate, matrix unit using the cleaned glass syringe.
spike and matrix spike duplicate (if enough sample is NOTE 1—It is important that this syringe is cleaned with water and
available), and an RLCS. The sample is mixed in the laboratory organic solvents as discussed in Section 7. The filters require rinsing with
with a clean spatula to have a homogeneous subsample. 10 mL acetonitrile followed by 10 mL methanol before use. These
chemicals are found in facilities that manufacture filter units. Nonylphenol
11.1.1 Method Blank—The method blank is prepared by is used in the polypropylene housing to extend shelf life and make the
measuring ~2 g of Ottawa sand in a 40-mL VOA vial and then plastic pliable and less brittle. It is important to wash the nonylphenol off
taken through the sample preparation Step II in 11.2. the filter unit.

9
D8310 − 20
11.2.4 To the extract, 2.5 mL of ASTM Type 1 water is 12.3 Calibration:
added and mixed. If precipitation is seen after adding water, the 12.3.1 Initial Calibration—The initial calibration contains
samples shall be filtered again through a PVDF filter unit. an eight-point curve. Depending on instrument type, the
11.2.5 An aliquot of that solution is transferred to an LC vial sensitivity and calibration curve responses may vary. At a
and a cap is applied. The final volume of the solution is 10 mL minimum, a five-point linear or a six-point quadratic calibra-
for quantitation purposes. tion curve will be used for all analytes. A calibration curve and
11.2.6 The analyte concentration for all samples will be an instrument blank will be analyzed at the beginning of each
reported in microgram/kilogram based on a dry-weight basis. run or daily to ensure instrument stability. A new curve will be
generated daily. The calibration method is saved and used to
12. Calibration and Standardization quantify all samples. Acceptance limit for calibration curve is
12.1 Calibration of Mass Spectrometer—The triple quadru- mentioned in 13.2.3 and 14.1.5.
pole mass spectrometer is calibrated monthly or when mass 12.3.2 CCC or End Calibration Check, or Both—The CCC
shifts of more than 0.2 Dalton are noticed by the analyst. The or end calibration check needs to be performed at the end of the
calibration file is saved in the software file folder. The batch and within 24 h of the initial calibration of that same
calibration solution normally used is a mixture of NaCsI. Other analytical batch. The CCC or end calibration check standard is
calibration solutions can also be used per manufacturer’s at or near the midpoint of the calibration curve. A QC
specifications. acceptance criterion for CCC is included in 14.2.7. A new
12.2 Instrument Operating Conditions—Analytical condi- calibration shall be generated every 24 h regardless of a
tions for LC and the mass spectrometer are: passing CCC.
12.2.1 Liquid Chromatographic Conditions: 12.4 Autosampler Schedule/Analytical Sequence—Prepare a
12.2.1.1 Analytical Column—UPLC BEH C18, 2.1 × sequence that includes all QC samples and field samples. The
100 mm, 1.7 µm particle size. first sample to be analyzed is a reagent blank sample. The
12.2.1.2 Injections of all standards and samples are nor- calibration standard’s levels are analyzed next. The next
mally made at a 10-µL volume. Other injection volumes may samples to be analyzed should be in the following sequence:
be used to optimize conditions. Standards and samples shall be reagent blank, method blank, RLCS, LCS/LCSD, diluted
in 75:25 acetone:water. In the case of extreme concentration samples, samples, duplicates, MS/MSD, and CCC.
differences among samples, it is wise to analyze a blank after
a concentrated sample and before a diluted sample to minimize 13. Procedure
carryover of analytes from injection to injection. However,
there should not be carryover between samples. The UPLC has 13.1 Sample Analysis Procedure:
a flow-through LC needle design. The gradient conditions for 13.1.1 Instrument conditions for LC/MS/MS are described
liquid chromatography are shown in Table 20. in 12.2. The target compound is identified by comparing the
12.2.2 Mass Spectrometer Conditions—To acquire the sample SRM transition to the known standard SRM transition.
maximum number of data points per SRM channel while Confirmatory transitions are available for BPA and BPA-D6
maintaining adequate sensitivity, the tune parameters shall be (Table 2 and Table 22). The RT for the analytes of interest shall
optimized according to the instrument. Each peak should have also agree with the RT of the mid-level standard by 65 %. The
at least ten scans per peak for adequate quantitation. This test target compound is quantitated using the SRM transition of the
method contains five target compounds and four surrogates that target compound using external calibration. The final report
are optimized in SRM experiment windows to acquire an issued for each sample lists total concentration of TPs, if
optimum number of scans and sensitivity. Variable parameters detected, or non-detect at the RL, if not detected, in
regarding retention times, SRM transitions, and cone and microgram/kilogram on a dry-weight basis.
collision energies are shown in Table 2. Mass spectrometer 13.1.2 If the absolute amount of a target compound in a
parameters used in the development of this test method are sample exceeds the working calibration range, the sample is
listed in Table 21. The instrument is set in the electrospray diluted and reanalyzed. This should be done by diluting the
ionization source setting. sample with 75:25 acetone:water.
13.2 Qualitative and Quantitative Analysis:
TABLE 20 Gradient Conditions for LC 13.2.1 The quantitation of the target analyte is accomplished
Percent Percent 100 mM
with the appropriate software. No internal standards are used.
Time, Flow, Percent
min µL/min CH3CN
95 % Water: NH4OAc in 95 % Refer to Table 22 for the SRM transitions and RTs. The SRM
5 % CH3CN Water: 5 % CH3CN analysis provides confirmation by isolating the precursor ion,
0 300 0 95 5
1 300 0 95 5 fragmenting it to the product ion, and also relating the
3 300 50 45 5 transition RT data. The software manual should be consulted to
4 300 60 35 5 use the software correctly. The quantitation method is set as an
6 300 70 25 5
7 300 70 25 5 external calibration using the peak areas in µg/L units.
9 300 95 0 5
13 300 95 0 5 NOTE 2—If suitable internal standards or isotopes of the native analytes
14 300 0 95 5 are identified, they can be used for isotope dilution or internal standard
16 300 0 95 5 quantitation. If these quantitation types are utilized, the recoveries must
meet the lower and upper control limits shown in Table 3. Any deviations

10
D8310 − 20
TABLE 21 MS Tune ParametersA
Electrospray Positive Mode Electrospray Negative Mode
Capillary Voltage: 3.5 kV Capillary Voltage: 1 kV
Cone: Variable depending on analyte (Table 1) Cone: Variable depending on analyte (Table 1)
Extractor: 2 V Extractor: 3 V
RF Lens: 0.1 V RF Lens: 0.1 V
Source Temperature: 150 °C Source Temperature: 150 °C
Desolvation Temperature: 450 °C Desolvation Temperature: 450 °C
Desolvation Gas Flow: 900 L/h Desolvation Gas Flow: 900 L/h
Cone Gas Flow: 200 L/h Cone Gas Flow: 300 L/h
Low Mass Resolution 1:9 Low Mass Resolution 1:8
High Mass Resolution 1:15 High Mass Resolution 1:14
Ion Energy 1: –0.1 V Ion Energy 1: 0.9 V
Entrance Energy: –1 V Entrance Energy: –1 V
Collision Energy: Variable depending on analyte (Table 1) Collision Energy: Variable depending on analyte (Table 1)
Exit Energy: 1 V Exit Energy: 0 V
Low Mass Resolution 2:9 Low Mass Resolution 2:7
High Mass resolution 2:15 High Mass resolution 2:13
Ion Energy 2: 1.0 V Ion Energy 2: 1.5 V
Multiplier: 700 V Multiplier: 700 V
Inter-Channel Delay: 0.02 s Inter-Channel Delay: 0.02 s
Inter-Scan Delay: 0.02 Inter-Scan Delay: 0.02
Repeats: 1 Repeats: 1
Span: 0.2 Dalton Span: 0.2 Dalton
Collision Gas Flow (mL/min): 0.25 Collision Gas Flow (mL/min): 0.35
Dwell: 0.1 to 0.105 s to optimize scans Dwell: 0.1 to 0.105 s to optimize scans
A
A Waters Quattro Premier XE mass spectrometer was used to develop this standard.

TABLE 22 RT and SRM Ions


Retention Time, SRM Mass Transition
Analyte ESI Mode
min (Parent > Product)
BPA negative 5.58 227.1 > 211.9
BPA confirmatoryA negative 5.58 227.1 > 132.7
OP negative 8.53 205.1 > 132.7
NP negative 9.70 219.3 > 133.1
NP1EO positive 9.71 282.3 > 126.9
NP2EO positive 9.65 326.3 > 182.9
BPA-D6 (surrogate) negative 5.57 233.3 > 214.9
BPA-D6 confirmatoryA (surrogate) negative 5.57 233.3 > 137.8
Br-OP (surrogate) negative 9.56 283.1 > 78.6
n-NP (surrogate) negative 10.71 219.3 > 105.6
n-NP2EO (surrogate) positive 10.69 326.4 > 88.8
A
Confirmatory transitions are optional but should be included for added qualitative information.

outside these limits must be fully documented and accompany the data curves, if one of the calibration standard injections, other than
generated. the high or low, skews the curve such that the r2 is
13.2.2 If there are two or more analyses for a particular unacceptable, this point shall be re-injected and replaced in the
sample because of sample dilution, the analyst shall determine calibration curve or a new calibration curve shall be made. If
which is best to report on the sample summary results sheet the low or high point is excluded, a six-point curve is
based on evaluation of all available data relevant to a given acceptable for quadratic fit and five-point for a linear fit, but the
sample. The target compound is identified by comparing the calibration range and reporting limits shall be modified to
sample primary SRM transition. Confirmatory transitions are reflect this change. Only the high or low point may be
available for BPA and BPA-D6 only (Table 2 and Table 22). If excluded, and internal calibration points may not be excluded.
a co-eluting peak is present that interferes with the primary
transition, a confirmatory transition may be used for quantita- 13.2.4 The RT window of the SRM transitions shall be
tion. This change shall be documented in the case narrative. within 5 % of the RT of the analyte in a Level 4 to 6 calibration
These changes should be rare exceptions. standard. If this is not true, the calibration curve needs to be
13.2.3 Regression fits should exclude the point of origin (X reanalyzed to see if there was a shift in RT during the analysis
= 0, Y = 0) and be weighted by 1/concentration to increase the and the sample needs to be re-injected. If the RT is still
accuracy of the curve at the lower concentrations. For linear incorrect in the sample, the analyte is referred to as an
regression to be used, the coefficient of determination, r2, unknown. If the RT is drifting more than 5 %, this will be
should be >0.98 for each analyte and, for quadratic regression, noticed in the end calibration check and shall be noted in the
the r2 should be >0.99. Upon inspection of the calibration case narrative.

11
D8310 − 20
14. Quality Control where:
14.1 Minimum Requirement: Vf = final volume,
14.1.1 An analyst shall have an approved analyst demon- Vi = initial volume (before dilution),
stration of capability (ADOC)6 before reporting data from a Cu = uncorrected concentration (before dilution), and
method. These studies evaluate whether the RLs and calibra- Cf = final concentration (corrected for dilution).
tion standard concentrations are appropriate. MDL studies shall 14.1.5 Calibration Range—The initial calibration contains
be performed annually for each method if values are reported an eight-point curve. Depending on instrument type, the
below the RL. An RL check standard shall be analyzed with sensitivity and calibration curve responses may vary. At a
each batch of 20 samples or less. minimum, a five-point linear or six-point quadratic calibration
14.1.2 Demonstration of Capability (DOC): curve will be used for all analytes. The coefficient of the
14.1.2.1 An initial demonstration of the laboratory capabil- determination (r2) of the linear fit shall be greater than or equal
ity to generate data of acceptable quality is made. A precision to 0.98. The r2 of the quadratic curve shall be greater than or
and accuracy (P&A) study shall be performed whenever a equal to 0.99. The calibration points used to generate the curve
major modification is made to this test method. The average should not be deviated more than 630 % from the curve. A
percent recovery (X), individual percent recovery, and the new calibration curve will be generated daily.
standard deviation (σ) of the recoveries are calculated for each 14.2 QC Acceptance Criteria—All control limits are pre-
analyte. QC confidence limits at 99.7 %, or three times the liminary because this is a new standard and collected data are
standard deviation of the replicates (3σ), are established. QC insufficient to determine historical acceptance criteria over the
acceptance criteria for the P&A study are shown in Table 3. course of a longer period. Preliminary QC acceptance criteria
14.1.2.2 For a P&A study, four samples containing will be addressed when the standard is renewed or when
1250 µg ⁄L for OP, NP, and NP2EO; 625 µg/L for BPA; and additional validation data become available.
18 750 µg/L for NP1EO shall be analyzed as replicates. These 14.2.1 Matrix Spike:
samples are then analyzed according to the method described 14.2.1.1 As part of a QC program, spike accuracy for each
in Section 11. matrix is monitored and updated regularly. Records are main-
14.1.3 MDL Test—The MDL test is performed using the tained of spiked matrix analyses, and the average percent
procedure outlined in 40 CFR Part 136, Appendix B as a guide. recovery (X) and the standard deviation of the percent recovery
MDL data for this test method are shown in Table 23. The are calculated. This procedure maintains a 99.7 % confidence
MDL criteria are not required to meet the criteria in 40 CFR interval from X 6 3σ control limits for spike compounds. The
Part 136 because this SOP only reports to the RL and this is not acceptance criteria that should be used are shown in Table 3
a regulatory method. MDL studies will be repeated at least until more matrix spike data are obtained.
yearly if any data are reported between the MDL and RL.
14.2.1.2 Calculate the percent recovery of the spike (P)
14.1.4 Example Calculation of Sample Concentration
using:
Reported—The concentration of sample is calculated using Eq
1 and Eq 2 if dilution is required. SSR 2 SR
P 5 100 3 (3)
SA
@ C i ~ µg/L! # 3 @ V s ~ L ! #
C s ~ µg/kg! 5 (1) where:
@ W d ~ kg! #
where: SSR = MS/MSD spike sample result,
SR = unspiked sample result,
Cs = concentration of target analyte in sample, SA = spike concentration, and
Ci = concentration of target analyte in sample from P = percent recovery.
instrument,
Vs = volume of sample extract, and 14.2.2 Surrogates—As part of a QC program, all samples
Wd = dry weight of the sample. are spiked with surrogate standard spiking solution as de-
scribed in Section 11. The percentage recovery limit for each
Vf surrogate compound is updated regularly based on historical
5 ~ C u! 5 C f (2)
Vi recovery data and is based on 99.7 % confidence interval from
X 6 3σ to establish control limits for surrogate compounds.
6
The acceptance criteria are shown in Table 3. These values are
An example of demonstration of capability is in the following link: https://
nelac-institute.org/docs/2003nelacstandard.pdf.
currently set to 70 to 130 % until more data are acquired.

TABLE 23 MDL Study


Spike Level, Number of Mean Recovery, Mean Recovery, Standard
Analyte Percent RSD MDL,A µg/kg
µg/kg Samples µg/kg % Deviation
BPA 50.0 9 36.9 73.9 14.5 5.3 15.5
OP 100.0 9 105.0 105.0 14.5 15.3 44.2
NP 100.0 9 108.0 108.0 9.7 10.5 30.4
NP1EO 1500.0 9 1472.2 98.1 21.8 321.6 931.2
NP2EO 100.0 9 90.1 90.1 2.8 2.6 7.4
A
MDL (µg/kg) = standard deviation × t (n – 1.1 – α = 0.99).

12
D8310 − 20
14.2.3 Duplicates—The relative percent difference (Eq 4) samples may be re-extracted into a new batch including all new
for the duplicate sample should be <30 % RPD. If greater, the batch QC. As stated in 14.2.7, if an end calibration check fails,
associated sample concentration is qualified, estimated, and the samples should be reanalyzed if possible. All reported
noted in the case narrative. exceedances of QC criteria need to be discussed in the case
14.2.4 Reagent Blank—A reagent blank is prepared with narrative.
75:25 % acetone and water for every 20 samples to investigate 14.3.1 LCS—As part of a QC program, spike accuracy for
for system/laboratory contamination. This solution is also used each matrix sample is monitored with each batch. Records are
to make the calibration curve standards and dilute samples, if maintained of spiked sample analyses, and the average percent
required. The concentration of the target analyte in the blank recovery and the standard deviation of the percent recovery are
shall be less than half the RL. calculated. This procedure maintains a 99.7 % confidence
14.2.5 Method Blank—A method blank for every 20 interval from X 6 3σ control limits for spike compounds.
samples will be prepared in Ottawa sand to investigate for Criteria for LCS/LCSD samples are shown in Table 3. All
contamination during sample preparation and extraction. The reported exceedances of QC criteria need to be discussed in the
concentration of target analytes in the blank shall be less than case narrative and data appropriately qualified according to
half the RL. your laboratory’s quality management plan requirements.
14.2.6 MS/MSD—An MS/MSD is extracted with each ma- 14.3.2 Matrix Spike Samples—As part of a QC program,
trix at a frequency of at least one MS/MSD pair for every 20 spike accuracy for each matrix sample is monitored with each
samples to investigate for matrix interferences. If the labora- batch. Records are maintained of spiked sample analyses, and
tory has not received MS/MSD samples for site-specific P&A the average percent recovery and the standard deviation of the
data, the site data quality will be evaluated solely on the LCS percent recovery are calculated. This procedure maintains a
criteria. Tables 4-17 display recoveries in different soils. 99.7 % confidence interval from X 6 3σ control limits for spike
14.2.7 CCC or End Calibration Check, or Both: compounds. Criteria for MS/MSD samples are shown in Table
14.2.7.1 Analyze a mid-level continuing calibration stan- 3 (LCS/LCSD criteria). All reported exceedances of QC
dard at the end of each batch. All analytes should be within criteria need to be discussed in the case narrative and data
630 % of their expected values. If not, a separately prepared appropriately qualified according to your laboratory’s quality
CCC may be analyzed. If the second CCC fails the criteria, management plan requirements. These criteria pertain only to
630 % of the expected analyte value, the data may be reported the native unspiked sample.
qualified as estimated or, if sample and time are available, be 14.3.3 Method Blank and Reagent Blank Samples—The
reanalyzed with a new calibration curve and end CCC check. If concentration of target analyte in the blank shall be less than
any data are reported with any failing QC, it shall be explained half the RL or the data shall be qualified that there is possible
in the case narrative and properly qualified according to your interference from the blank, or the RL for the associated
laboratory’s quality management plan. sample(s) shall be raised to at least three times above the blank
14.2.7.2 The percent difference (%D) between the calcu- contamination concentration. Since a quadratic fit is often used,
lated and expected concentration in the continuing calibration the concentrations below the RL are not accurate; therefore, the
verification standard is calculated using: response/area count of the blank should be less than half the
response/area count in the associated sample(s).
Calculated concentration 2 Expected concentration
%D 5 3 100 (4) 14.3.4 RLCS—The recovery criteria for the RLCS is 50 to
Expected concentration
150 %. If the analytes are not detected or biased low in the
14.2.8 LCS/LCSD—As part of requirement of a QC RLCS, then the data for all non-detects are qualified according
program, spike accuracy for water is monitored with each to your quality management plan and the QC failure is
batch. At least one LCS/LCSD pair for every 20 samples is explained in the narrative accompanying the data.
extracted and analyzed. The percent recovery limits for the 14.3.5 Duplicate Samples:
target compound are updated regularly based on historical 14.3.5.1 A duplicate sample is analyzed with every batch of
recovery data and are based on a 99.7 % confidence interval 20 samples. The relative percent difference in the duplicate
from X 6 3σ as control limits. The acceptance criteria are shall be less than 630 %. If not, the native sample is qualified
shown in Table 3. estimated. For problematic matrices, it is recommended that
14.2.9 RLCS—For each batch or within the 24-h analysis the frequency of duplicate analyses be increased above this
window, an RLCS shall be analyzed. The RLCS is processed criteria.
like an LCS and spiked at or near the lowest RL. This sample 14.3.5.2 Calculate the relative percent difference (RPD)
is to verify that, if the analytes were present at the RL, they using:
would be confidently identified. The recovery criteria for the
RLCS is 50 to 150 %. RPD 5
? @ S # 2 @ DS # ? 3 100 (5)
~ @ S # 1 @ DS # ! ⁄ 2
14.3 Immediate Corrective Actions—All control limits are
where:
preliminary because this is a new standard and collected data
are insufficient to determine acceptance criteria based on S = concentration of the sample, and
DS = concentration of the duplicate sample.
historical data. If the QC criteria are not met, the samples may
be reanalyzed to ensure the best possible results are reported. If 14.3.6 Surrogates—Br-OP represents OP and NP, BPA-D6
the QC samples are reanalyzed and continue to fail, the site represents BPA, and n-NP2EO represents NP1EO and NP2EO.

13
D8310 − 20
n-NP is not representative of NP or OP but is included as a maintenance, there is the potential for introducing new sub-
monitoring compound. n-NP is the straight chain analog that stances into the vacuum system. Some substances can be
does not mimic the branched. If the recovery for n-NP is low, readily removed by evacuation, while others require more
this does not indicate that the recovery for NP or OP is low cleaning or baking.
based on historical data. Qualification of data generated with 16.2.2.1 Solvents Used in the Cleaning Process—These may
this method should follow your laboratory’s quality manage- be present for a short time and should be pumped out to the
ment plan. vacuum system.
16.2.2.2 Water Absorbed on the Metal Surfaces While
15. Data and Records Management
Vented—This will pump out.
15.1 Calculations and Documentation:
16.3 Symptom—Mass Spectrometer Does Not Respond
15.1.1 Refer to Eq 1 on how the final concentration is
Probable Cause:
calculated.
15.1.2 Use of professional judgment in the data reduction 16.3.1 Mass Spectrometer Electronics Are Not On—Check
and verification process shall be documented. the switch.
15.1.3 All QA/QC data and final results are reported. 16.3.2 Secondary Fuse Blown—Check the secondary fuses
on the rear of the mass spectrometer and replace the faulty fuse
15.2 Reporting Result Procedure: or fuses.
15.2.1 The concentration in the sample is calculated using
16.3.3 Board failure.
the linear or quadratic calibration curve. The sample results are
reported to the RL in microgram per kilogram based on 16.3.4 Communication Loss—Reboot mass spectrometer,
dry-weight basis. These results are reported without any PC, and LC.
corrections for recovery data. All QC data obtained are
included in the data packages. 17. Preventive Maintenance
17.1 A logbook for recording preventive maintenance and
16. Troubleshooting repair work of LC/MS should be located near the instrument
16.1 Symptom—Inadequate Abundance or Sensitivity Prob- with dates of repair and initials. The reagents and standards are
able Causes: recorded in the analyst’s data logbook.
16.1.1 Dirty or contaminated ion source, electron multiplier,
17.2 The instrument manual should be located on the
or quadrupole rod surfaces.
instrument computer associated with the instrument software.
16.1.2 Potentials of ion source elements at wrong values
because of open or short circuits.
18. Precision and Bias
16.1.3 Faulty ion source electronics, detector electronics, or
power supply. 18.1 This standard has been tested on Ottawa sand, four
16.2 Symptom—High Background Probable Causes: ASTM soil types, biosolid sample, and one commercial soil
16.2.1 Carryover—There is the possibility that some previ- (results shown in Tables 4-17). The QC acceptance criteria are
ously injected sample can still be present in the vacuum system listed in Table 3.
long after it was thought to be evacuated. This phenomenon
depends on sample volatility, temperature, and so forth. 19. Keywords
16.2.2 Contamination in a Recently Cleaned Vacuum 19.1 Bisphenol A; LC/MS/MS; liquid chromatography;
System—After any venting of a vacuum system for mass spectrometry; nonylphenol; octylphenol; sediment; soil

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

14

You might also like