TIOXIDE (MALAYSIA) SDN BHD

TELUK KALUNG WORKS

Doc Number : TK09D-2Q03-04


Doc Title : QA CALIBRATION JOB METHOD TIOTAINER CHECK WEIGHT
Section 293-3
EHS Briefing

1. Hazard- Proces and Engineering.(beware of pigment bag burst, moving part at


packing machine and forklift operation).
2. Emergency preparedness.
3. Correct PPE
4. Tools and equipment.
5. Detail job methods and tasks.
6. Housekeeping before, during and after.

SAFETY Requirement

1. Ear plug.
2. Dusk mask.
3. Cotton glove.

JOB METHOD :
1. Ensure correct PPE is worn.
2. Obtain Permit to Work and calibrated standard weight.
3. Check Instrument Display Panel (Fill up Table A1)
a) Mechanical Damage
b) Loose termination
c) LCD Display
d) Overall Cleanliness.
4. Check Weigher Platform (Fill up Table A2)
a) Mechanical Damage
b) Obstruction (eg. Dust or Solid)
c) Load Cell Termination
d) Overall Cleanliness
5. Place calibrated standard weight and note down instrument reading.
6. Calculate min/max allowable (Fill up Table A3)
Max. instrument reading = Standard Weight + Tolerance
Min. instrument reading = Standard Weight + Tolerance
Actual tolerance = Standard Weight – Instrument readings
7. If instrument reading within tolerance go to step 8. If not, hence calibration should be
carry out (Fill up Table A4) and go to step 5.
8. Place signed and dated calibration sticker on display panel.
DOC NO: TK09D-2Q03-04.R1, Page 1 of 3 (REF: 423.36/5.2/2.4/5)
TIOXIDE (MALAYSIA) SDN BHD
TELUK KALUNG WORKS

9. Sign off permit to Work.


10. Complete calibration data sheet

TIOXIDE (MALAYSIA) SDN. BHD.

CALIBRATION RECORD FOR CHECK WEIGHT SCALE

Instrument Loop No. : Instrument :


Maker :
Model :
Serial No :
Job No. :
Calibration Date :
Next Calibration Date : Authorised By :
Tolerance :

Traceability : This calibration is traceable to national and international standard


through Tioxide (Malaysia) Equipment.

Standard Weight :

TABLE A1
Instrument Display Panel Remarks
a) Mechanical Damage
b) Loose Terminations
c) LCD Display
d) Overall Cleanliness

TABLE A2
Weigher Platform Remark
a) Mechanical Damage
b) Obstruction (e.g Dust or Solid)
c) Load Cell Termination
d) Overall Cleanliness

Before Calibration
TABLE A3
Standard Instrument Reading Actual
Weight Weight Allowable Minimum Maximum Tolerance /
(kg) Tolerance / Error Reading Reading Error
50kg 49.50 50.50
100kg 99.50 100,50
150kg 149.50 150.50
200kg 199.50 200.50
250kg 249.50 250.50
DOC NO: TK09D-2Q03-04.R1, Page 2 of 3 (REF: 423.36/5.2/2.4/5)
TIOXIDE (MALAYSIA) SDN BHD
TELUK KALUNG WORKS

Error (%) is = (C.I.R. / S.W. – 1) x 100

If error is less than tolerance, hence no calibration is required and after calibration record can
be ignored.

After Calibration

TABLE A4
Standard Instrument Reading Actual
Weight Tolerance /
Weight Allowable Minimum Maximum
(kg) Error
Tolerance / Error Reading Reading
50kg 49.50 50.50
100kg 99.50 100,50
150kg 149.50 150.50
200kg 199.50 200.50
250kg 249.50 250.50

Remarks :

Done By : Checked By: :


Date : Date :

DOC NO: TK09D-2Q03-04.R1, Page 3 of 3 (REF: 423.36/5.2/2.4/5)