You are on page 1of 2

Dynamic Test

Produce simulated vibration on machine and verify the response of equipment

To perform this test:

First determine the proper gap voltage settings by referring to the proximity probe spec sheet.

This can be done in two ways. The first is to use the recommended gap settings. The second is to
determine the midpoint of the linear range. Most probe and driver combinations are 200mv whereas
every 10 mills is equal to 2 volts

To determine the exact voltage at the center of the linear range use the formula: ((range / 2 + 10) *.2 ) –
1. This will equate the voltage at the center of the linear range. In the above case of 200mV over a range
of 80 mils, our result is -9 Volts DC. This calculation also matches the recommended gap setting in the
above example. Always remember that the linear range does not begin until 10 mils from the target.

Assemble proximity probe kit.

Locate proper size arm adapter

Mount probe tip to arm adapter.

Mount arm adapter to micrometer and secure probe in plate

Adjust leg height if necessary. Three height differences are included with the kit.
In the sensor setup screen locate the GAP V setting and adjust the gap voltage.

Use the micrometer to push the probe all the way down and contact the steel target. The DC value on
your calibrator should read less than -1 volt.

Then use the micrometer to slowly pull away until the DC voltage reads the previously calculated value.
(-9 Volts DC in our example of a 80 mil linear range.)

You are now ready to conduct your proximity probe test. In Automatic mode, the shaker will conduct
the test without the need for adjustments. In manual mode, adjust the speed to the same as the
equipment the probe will be used on. Then increase the amplitude over a range of 1 mil to 10 mils.

You might also like