You are on page 1of 6

But. J. Anaesth.

(1968), 40, 660

APPARATUS FOR CARBON DIOXIDE ABSORPTION


BY

A. BRACKEN AND L. A. Cox

Carbon dioxide absorption in anaesthesia became Clearly the choice of material will lie between
increasingly popular following the introduction of glass and one of the plastics. Glass has the advan-
cyclopropane and continues in wide use today tage of being dimensionally stable despite re-
largely because it conserves the new and relatively peated sterilization by autoclaving and as a raw
expensive volatile anaesthetic agents such as halo- material it is cheap. It is however, very fragile,
thane. Some of the arguments put forward in the and furthermore there are difficulties in manufac-
earlier days to justify the closed circuit have been turing a large tubular item to the reasonably close
shown to be largely fallacious, for example, the dimensional tolerances necessary, bearing in mind
claim that it conserves heat. This, and the chang- the need to effect a gastight seal with its mating
ing trends in anaesthetic techniques are good components. Some plastics, on the other hand,
reasons why the design specification for a canister can be readily moulded or otherwise fabricated
to operate with optimum efficiency should be and, provided the right material is used, not only
reviewed periodically. is the cost of the item more likely to be brought
The efficiency of carbon dioxide absorption is down to an acceptable level but there would be
governed not only by the absorbent and the de- the attendant advantage of less frequent replace-
sign of the soda-lime canister, but also by circuit ment due to breakages. There is an increasing
arrangements including the location of the spill range of plastics to choose from nowadays, includ-
valve and the directional valves, the rate of flow ing acrylic, polycarbonate, polystyrene, and cellu-
of fresh gases and where they enter the circuit, lose acetate, all of which are available in a dear
and the possible effects of a ventilator if employed. colourless form. All these materials are thermo-
Some of these topics are not necessarily pertinent plastic and until recendy there were limitations in
to the absorber itself, in which there is a close the mavimnm temperature to which they could be
interrelationship between size of canister and subjected; this precluded sterilization by auto-
nature of absorbent, and therefore they will not clave. Some of the more recently developed plas-
be discussed in detail here. tics, however, are better in this respect but
unfortunately are also considerably more expen-
sive.
THE CANISTER One fundamental difficulty in regard to the use
The most important change in apparatus design of transparent canisters arises from the fact that
in recent years has been the introduction of large the observer sees only the outside layer of the
transparent canisters (Hale, 1967). There are two absorbent In all the absorbers we have examined,
points in favour of making these transparent; first this layer does a disproportionate amount of the
and foremost is the obvious one that it permits work of absorption because the gases will prefer-
visual inspection of soda-lime throughout the entially pass between the granules and the smooth
absorption process, and in this respect has helped inner surface of the canister. One effect of this
to increase the popularity of colour indicators; channelling has been referred to by Robson and
secondly, it is not unreasonable to expect that a Pask (1954) who stated that when they devised a
large transparent canister could be constructed pot-scrubber compression device it stopped chan-
more cheaply than, say, a chromium-plated metal nelling, but the pressure drop increased from 42
one. There are, however, certain problems asso- mm H2O at a flow rate of 40 litres per minute,
ciated with transparent absorbers which need to to 7.8 mm HjO, because the gases passed through
be considered. the bed of soda-lime instead of partially bypass-
APPARATUS FOR CARBON DIOXIDE ABSORPTION 661

ing it. Some commercial designs embody a ring- the same weight of soda-lime. Absorbents such
shaped partition which partially divides the canis- as monoethanolamine, used in industry (Sharp,
ter into equal upper and lower compartments, and 1958), can readily be regenerated and so can
this is fairly effective in diverting the gas stream various types of ion exchange resins now avail-
into the soda-lime bed. able; molecular sieves can be good carbon dioxide
It is very necessary to take into account the absorbents and can also be regenerated. Inves-
hazard of static electrification. Large areas of tigations (e.g. Bracken, 1955) have shown that in
insulating material on which simple friction can some cases there are numerous technical objec-
generate a high charge of static electricity (and tions to their use in anaesthesia whereas in other
conceivably a spark) have in the past been con- cases the high cost is the dominating disadvan-
sidered dangerous. This danger is currently felt tage.
to be less serious than it was because increased This last point bears upon apparatus design
awareness of the explosion risks has led to better because following American work, notably by
precautions for their prevention, and because of Kappesser (1953), Ham (1958) and Brown, Seniff
the wider use of non-explosive agents like and Elam (1964), it has become usual to provide
halothane. Nevertheless, some basic objections large absorbers which are constructed in two
were raised following the introduction of the parts, the principle being that the exhausted half
transparent canister. It was claimed by some that can be refilled with fresh soda-lime and the
the change from a metal conductive item to a canister reversed so that the relatively fresh
non-metallic insulating one was a retrograde step second half makes first contact with the incoming
which could not be defended by instancing the gas. With this reverse flow principle, the soda-
use of glass containers for vaporizers of volatile lime is used right through to the end without en-
and flammable anaesthetics such as ether. After dangering the patient by carbon dioxide breaking
all, glass bottles have been used for anaesthetic through. However, it has been suggested to us
vaporizers since the earliest days, and any hazards that the apparatus is not always used in the
arising therefrom have always been present. Also, manner prescribed and it would seem that in
owing to its attraction of a thin moisture film in some hospitals attention to the absorption appara-
all except the driest atmospheres, glass has a tus rates very low in the list of duties carried out
surface resistivity of about 1010 ohms, which is by the theatre staff. The large canister concept
ostensibly less dangerous than most plastics whose should be a boon to busy theatre staff, as even
surface resistivities are usually around lO14 ohms. with continuous use one complete charge may last
Notwithstanding this, tests in our laboratory have two weeks or more before needing replacement.
shown that a canister filled with soda-lime (a Despite the popularity of indicator soda-lime,
relatively good conductor of electricity) cannot frequent analysis of the gas stream is the only
retain a dangerous static charge even on its outer really satisfactory way of detecting when the
surface since any high charge will leak away soda-lime is about to be exhausted and this should
through the soda-lime. be carried out as a regular routine.
The moisture content of soda-lime and its con-
sequent efficiency as an absorbent, are closely
THE ABSORBENT related to the canister size and even the choice
In considering apparatus used for carbon dioxide of a colour indicator can be affected by this rela-
absorption, reference must be made to the absor- tionship. The moisture content should be around
bent itself. In anaesthetic practice, only two ab- 15 per cent, the upper limit being dictated only
sorbents have been widely used—soda-lime and by considerations of waterlogging (Brown, Bakam-
barium hydroxide octahydrate (baryta-lime). jian and Seniff, 1959). In a 1 lb. canister, soda-
Others have been recommended from time to lime containing below about 10 per cent moisture
time; for example, sodium peroxide which like falls off sharply in its performance. On the other
some other substances will absorb carbon dioxide hand, if soda-lime with a low moisture content is
and at the same time liberate oxygen; lithium used in a large canister there is not the same fall-
hydroxide absorbs more carbon dioxide than does off. One reason is that it does not become as hot
662 BRITISH JOURNAL OF ANAESTHESIA

in use as a small canister. Therefore, water liber- (Nealon, Chase and Gibbon, 1958). When it is
ated in the reaction used in a 1-lb. canister the volume is such that the
CO, + 2NaOH=Na 2 CO 3 + H3O gas does not have time to react completely and
does not escape and there is a greater tendency in this case baryta-lime is inefficient.
Questions of mesh size and storage conditions
(a) for the drier soda-lime to approximate to an
of soda-lime have been well aired (e.g. Bracken
ideal moisture content;
and Sanderson, 1955). Most brands are made
(fc) for the soda-lime to become waterlogged either 4-8 mesh or, for greater efficiency, 5-10
if initially it has a high water content; and mesh. The slight increase in resistance to gas
(c) for the indicator to wash out if this is very flow in the smaller canister resulting from the use
water-soluble. of finer mesh soda-lime appears to be acceptable,
Further, the longer residence time of the gas in whilst in larger canisters the pressure drop is less
a large canister assists by giving a longer reaction anyway, because of their larger cross-sectional
time, and removes one of the objections to drier area. There appears to be no objection to combin-
soda-lime, i.e. "peaking" or "regeneration". This ing the above mesh sizes to produce a single stan-
was one of the problems of earlier years which is no dard of 4-10 mesh. The storage conditions of
longer considered of any great importance. It was soda-lime do not affect carbon dioxide absorption
the name given to the reactivation of apparently efficiency so much as they do moisture loss. In
spent soda-lime on standing for a few hours. some tests we carried out we have shown that
This was a nuisance because it was difficult to soda-lime stored in a badly sealed container may
decide whether to discard the soda-lime at the lose moisture but not gain very much carbon
end of the day. Regeneration was a failing of dry dioxide except in the surface layers, so that over-
soda-lime when used in small canisters (Adriani, all efficiency is little impaired.
1962). We have confirmed earlier reports (Lund, Some figures for the degree of exhaustion of
Andersen and Erikson, 1956) that a large canister soda-lime relating to size of canister are given by
rilled with drier soda-lime does not exhibit this Miles and Adriani (1959). They showed that 1
phenomenon because of self-moistening of the g of spent soda-nine from a standard 8 x 13 cm
charge. canister (holding 1 lb. of soda-lime) liberated
It is interesting to note that these earlier 100-125 ml of carbon dioxide, whereas with 1 g
workers have indicated that only 80-88 per cent from a large double canister (holding a total of
of the carbon dioxide is absorbed in small canis- 4 lb.) the average was 175 ml. Expressing this
ters. The question arises as to what is an accept- another way, the same authors stated that for
able carbon dioxide return to the patient from every extra 100 g of soda-lime in the canister
an anaesthetic apparatus. Obviously opinions will there was an extra hour of absorption capacity.
differ on this point but most writers who have One big advantage of the large canister stems
expressed themselves have followed the lead of from its large inter-granular air space. In anaes-
Woolmer and Iind (1954) that the figure should thetic practice large canisters should only be used
not be more than 0.1 or 02 per cent of carbon in a circle system as they are quite inappropriate
dioxide by volume. Using this criterion, 1-lb. in the Waters to-and-fro circuit. This is because
absorbers are satisfactory; larger absorbers are in a to-and-fro circuit the soda-lime nearest the
fully effective even when hah0 the soda-lime is patient would become exhausted first and the
exhausted. deadspace would increase until the patient's ex-
This aspect of canister size has a bearing on halation would eventually contact only exhausted
reports on the efficiency of baryta-lime. Some soda-lime. Then, because the fresh gas flow is
authors say it is inefficient, others claim it is satisfac- insufficient to drive all the exhaled gas to the far
tory. Some at least of the diverging views can be end of the canister, only about 1 lb. of a 4-lb.
explained by the fact that with a large canister the charge is likely to be effective and any attempt
gas stream stays in contact with the absorbent to utilize the remaining 3 lb. would be unavailing.
for a longer rime and under these conditions The patient, after a few hours, would re-inhale
baryta-lime can be as efficient as soda-lime exhalations not freed from carbon dioxide, as the
APPARATUS FOR CARBON DIOXIDE ABSORPTION 663

deadspace will have increased until it is equal to tidal volume, or a slow respiratory rate with a
the tidal volume. The situation would then be large tidal volume or whether there are variations
especially dangerous because a carbon dioxide in the ratios between inspiration, expiration and
analysis taken from the rebreathing bag might respiratory pause; nor does the input gas tempera-
show that all is well, yet the patient could be ture have any significant effect.
suffering from carbon dioxide accumulation.
CIRCUIT ARRANGEMENTS

TABLE I In assessing the overall performance of the equip-


Effect of canister size on carbon dioxide absorption ment, one must not confuse the increased size of
capacity using circle system. the absorber with the circuit arrangements pro-
Tidal volume 500 mL vided for venting excess gas during the patient's
Respiration rate 20 per minute. exhalations. Brown, Seniff and Flam (1964) re-
Carbon dioxide input 200 ml per minute. cord how they were surprised at the extremely
Minutes before Endurance factor long life of the canister being tested, this being
CO, in inspired (min of use due to blowing off excess gas as well as to extra
Canister ens exceeded per 1 oz. of soda-lime. They found that depending upon cir-
size 0.6% soda-lime)
cuit arrangements, a greater or lesser proportion
3 oz. 4 1.3 of <rhn1<-H gas is mixed with fresh gases and
5±oz. 74.5 13.6 blown off in part to atmosphere before reaching
1 lb. 322.5 20.2
4 1b. 1725 27.0 the soda-lime. Since the fully closed circuit is
6 1b. 2904 30.3 comparatively rarely used today, this is another
reason why large canisters, given the right circuit
Some data obtained in our laboratory relating arrangements, can be used for such very long
to size of canister and efficiency may be given periods of time.
here. The investigation covered canisters of
various sizes, set up under standard "circle" con- TABLE II
Effect of circuit arrangements on inhaled oxygen
ditions in the apparatus described by Bracken
percentage.
(1955) where pulses of carbon-dioxide-laden air Human subject (A.B.)
are pushed through the canister. The results are Minute volume 9 L
given in table I, and reveal that there is a further Respiratory rate 14 respirations per minute.
Volume vented 4-5 1. per minute.
advantage in using a large canister inasmuch as
Oxygen concentration, % v/v
4 lb. of soda-hme will last for a much longer
time than would be expected merely from the Point in circuit where In Inhaled by
increased weight of soda-lime. In fact the 4-lb. venting took place circuit subject
canister lasted more than five times as long as the On face mask 21 19-20
1-lb. canister. This applies to circle absorption 54 51.5-52.6
At fresh gas inlet, Le.
only and not to to-and-fro, where a small canister remote from face
suffers less by comparison because the gas passes mask 54 50-50.5
through the charge twice. Thus, with the very
small 3-oz. canister it took 4 minutes to reach Arising from this work and following the com-
0.6 per cent in a circle system but over 8 minutes munications by Fitton (1958, 1963) we considered
in a to-and-fro system. These are average figures it desirable to find out how much, if any, of the
the results being rather scattered because of patient's much-needed oxygen is vented at the
sampling problems arising from the rapid rise same time as exhaled gas. In order to test this,
in carbon dioxide concentration. subjects breathed in and out of the Boyle Mark
Confirming previous studies (Miles and Adri- HI absorber and oxygen measurements were taken
ani, 1959) we have found that it does not make at various parts of the circuit Even with input
any difference, especially in regard to large canis- gas containing as little as 21 per cent oxygen
ters, whether the pulses of air are put through to there was no significant fall in inspired oxygen
simulate a rapid respiratory rate with a small concentration (table LI). A recent paper by Eger
664 BRITISH JOURNAL OF ANAESTHESIA

and Ethans (1968) and an annotation by Rendell- ively on a number of patients without intermed-
Baker (1968) in the same publication refer to the iate sterilization. The evidence of cross-infection
effect of different circuit arrangements on system is not very conclusive but there is no doubt that
economy. a soda-lime canister is not a reliable bacterial
A great deal of thought has gone into the de- filter (Stark, Green and Pask, 1962). Naturally,
sign of large canisters and one author (Hampton, there will be some kind of valve system to control
1967) has commented that some U.S. manufac- the direction of gas flow and this should be de-
turers had not at his time of writing paid due signed with ease of cleaning and sterilizing in
attention to the design points raised by one of mind. However, there is the question of who
the pioneers, Elam (1958). In particular he re- will do the cleaning and, perhaps more import-
ferred to the absence of a small air space at the ant, of the safe and correct reassembly of the
top of the soda-lime where the patient's exhaled apparatus, and this raises difficulties. A circle
gas enters the canister. Such a space allows dif- absorber does not lend itself to being put into
fusion and helps to prevent gas channelling. an autoclave, and although other sterilization
Elam's article contains some useful quantitative methods can be used (such as ethylene oxide, for-
statements; for example, the service interval of 8 malin, combinations of heat and biodetergents,
hours of intermittent use, and the patient's res- and aerosol disinfection), these are not always
piratory parameters of 0.5 to 1.0 litres tidal applicable because of lack of apparatus and trained
volume, producing 12-18 litres per hour carbon staff. Bacterial filters (Bishop, Roper and Wil-
dioxide output. Thus, allowing for the fact that liams, 1963) are not yet widely used as a solution
100 g of soda-lime absorbs 15 litres of carbon to the problem of apparatus cross-infection, so to
dioxide before the exit gas contains 1 per cent some extent this problem of hygiene remains un-
carbon dioxide, he arrived at the figure of 870 g resolved.
of soda-lime for an 8-hour period, this quantity One final point concerns packing the canisters
occupying about 1 litre in bulk. About 47 per cent with soda-lime. If a canister is packed by putting
of this soda-lime would be void space which layers 1-2 inches deep and then tamping more or
would accommodate one average tidal volume, less vigorously, there is a risk that this will break
and it would increase at the rate of 60 ml per up some of the granules and form powder (Sam-
hour as absorption proceeded. This could be son, 1956). This risk is obviously greater with the
countered by a second chamber of a similar more friable, soft, porous types which incidentally
capacity. This followed Kappesser's suggestion have the most efficient absorptive properties. On
(1953) that the canister should be in two halves, the other hand, if soda-lime is produced with
each being filled separately and the canister re- harder granules (and this is within the power of
versed during use so that partly exhausted soda- the manufacturer to control by varying the con-
lime becomes the top half, meeting the patient's ditions of drying of the original cake in the oven)
exhaled gases, and the lower half containing fresh then it will be more resistant to breaking up and
soda-lime completes the carbon dioxide elimina- powdering. Unfortunately, there is the disadvan-
tion. Elam (1958) calculated the resistance to air tage that the calcium carbonate produced on the
flow as less than 1 cm HjO at a flow rate of 60 outer surface by the absorption process will make
L/minute if the length of the combined canisters the granules still harder and less permeable, thus
was less than 18 cm and the diameter greater than decreasing efficiency. It follows from this that
12 cm. To assist in eliminating channelling Elam there is less objection to using hard soda-
suggested the annular ring baffles which arc now lime in a large canister as it gives th: manufac-
so widely used. The foregoing considerations turer more tolerance in the direction of less rapid
amount to a design justification for a 4-lb. canis- absorption. On the other hand, a certain amount
ter, and have been widely accepted. of personal skill will always be required in pack-
The introduction of large and complicated ab- ing the canisters. It is known that a properly
sorbers would at first sight appear to increase the packed canister absorbs as efficiently in a hori-
hazard of cross-infection as compared with small zontal position as it does in a vertical position
absorbers, if only because they are used success- (Robson and Pask, 1954), rhannfiiing along the
APPARATUS FOR CARBON DIOXIDE ABSORPTION 665

top of a horizontal canister being serious only Fitton, E. P. ((1963). Theoretical investigation of oxy-
gen concentrations: correction and additional
with a loosely packed canister. observations. Brit. J. Anaesth., 35, 276.
Hale, D. E. (1967). The rise and fall of soda lime.
Curr. Res. Anesth., 46, 648.
REFERENCES
Hampton, L.' J. (1967). Absorption of carbon dioxide:
Adriani, J. (1962). The Chemistry and Physics of Anes- influence of canister design on performance charac-
thesia, 2nd ed., p. 173. Springfield, Illinois: teristics of commercial absorbers. Anesthesiology,
Thomas. 28, 255.
Bishop, C , Roper, W. A. G., and Williams, S. R. Kappesser, R. C (1953). Modification of double
(1963). The use of an absolute filter to sterilize the canister circle filter. Anesthesiology, 14, 415.
inspiratory air during intermittent positive pres- Lund, I., Andersen, K. L., and Erikson, H. (1956).
sure respiration. Brit. J. Anaesth., 35, 32. Efficiency of carbon dioxide absorption by soda
Bracken, A. (1955). Discussion on carbon dioxide lime in a closed system. Brit. J. Anaesth., 28, 13.
accumulations in anaesthetic circuit Proc. roy. Soc. Miles, G., and Adriani, J. (1959). Carbon dioxide
Med., 49, 215. absorption: a closer look. Curr. Res. Anesth., 38,
Sanderson, D. M. (1955). Some observations on 293.
anaesthetic soda lime. Brit. J. Anaesth., 27, 422. Nealon, T. F., Chase, H. P., and Gibbon, J. H. (1958).
Brown, E. S., Bakamjian, V., and Seniff, A. M. (1959). Factors influencing carbon dioxide absorption dur-
Performance of absorbents: effect of moisture. ing anesthesia. Anesthesiology, 19, 75.
Anesthesiology, 20, 613. Rendell-Baker, L. (1968). On the promise of economy
denied. Anesthesiology, 29, 5.
Seniff, A. M., and Elam, J. O. (1964). Carbon Robson, J. G., and Pask, E. A. (1954). Some data on
dioxide elimination in semi-closed systems. Anes- the performance of Waters canister. Brit. J.
thesiology, 25, 31. Anaesth., 26, 333.
Eger, E. I., and Ethans, C. T. (1968). The effects of Samson, H. H. (1956). An improved carbon dioxide
inflow, overflow and valve placement on economy absorber. Lancet, 2, 1196.
of the circle system. Anesthesiology, 29, 93. Sharp, T. W. (1958). Carbon dioxide removal and re-
Elam, J. O. (1958). The design of circle absorbers. covery. Power Gas Group Rev., 5, 8.
Curr. Res. Anesth., 19, 99. Stark, D. C. C , Green, C A., and Pask, E. A. (1962).
Fitton, E. P. (1958). A theoretical investigation of oxy- Anaesthetic machines and cross-infection. Anaes-
gen concentrations in gases inspired from various thesia, 17, 126.
semiclosed anaesthetic systems. Brit. J. Anaesth., Woolmer, R., and Lind, B. (1954). Rebreathing with a
30, 269. semiclosed system. Brit. J. Anaesth., 26, 316.

CORRESPONDENCE
FOURTH WORLD CONGRESS OF ANAESTHESIOLOGISTS

Sir,—We are concerned that the reasons for not issuing this letter anaesthetists from sixty-six countries have
day membership tickets for consultants at the Fourth signified their intention of being present and have paid
World Congress of Anaesthesiologists do not seem to their registration fees. It behoves us, therefore, to be
be understood. generous. Every member of the Organizing Committee
First, it must be realized that the costs of organizing has paid his full registration fee although it is unlikely
such a gathering must be shared by all who come to it that any will be able to attend the meetings.
and that these costs are the same whether attendance We hope that this will clear any misgivings on the
is for one day or for the whole period. Therefore the question of day registration for consultants.
registration fee must be for attending the whole Con- GEORGE ELLIS
gress and not fractions of it. The complications of Chairman of the Organizing Committee
introducing "day membership" for consultants or any C. B. LEWIS
variants on such a scheme would add enormously to Chairman of Registration Committee
the cost of the secretariat—already a major expense.
Secondly, it must be borne in mind that we are [This letter was received for publication too late for
hosts to the rest of the world and at the time of writing inclusion in the list of Contents on front cover.]

You might also like