EQUIPMENT LAYOUT ITEM DESCRIPTION

DIMENSIONS
LxWxH (mm)
WEIGHT
(kg)

1 5334 mm RAILS FOR 3 m BRIDGE 5334x3341x95 137.4


2 OTS WITH 3 m BRIDGE 3077x659x169 350
3 ELEVATING TABLE G2 2400x933x850 440
4 STANDARD WALL STAND G2 686x1344x2628 270
5 SYSTEM CABINET 707x662x1475 320
6 POWER DISTRIBUTION BOX (PDB) 835x635x300 40
2-153 7 TETHER INTERFACE BOX (TIB) 255x110x310 7
CAST ROOM
8 GRID HOLDER 512x202x505 13.8
9 DONGLE ASSEMBLY 84x155x83 0.8
10 OPERATOR CONSOLE 1200x750x1204 36
11 WALL BOX 370x170x112 5

WALL - ACCORDING TO RECEIVED DRAWING

7265
EXAM ROOM HEIGHT
FINISHED FLOOR TO SLAB HEIGHT -
FALSE CEILING HEIGHT 2.80 m
9
8 2-154

1065

1065
EXAM ROOM APPLICATIONS
29.26 m²
2-154A The chart shows the applications possible to perform with the present equipment positioning,
DRESSING ROOM however the sales contract may not include all of them.
2 1
Auto Image Pasting at Wall Stand YES
303
1850 Auto Image Pasting at Table NO
Note A Dual Energy Substruction YES
4591

3 1034 Horizontal at Extended Arm Wall stand NO

VolumeRAD
4 Vertical at Wall stand NO
2-154C
CONTROL Cross-Table at Wall stand NO
ROOM
Horizontal at Table NO
2720

3.46 m²
2-154B

2168
TECH

AL MOOSA SPECIALIST HOSPITAL


10
1400

AL-MUBARRAZ
11 SAUDI ARABIA
Imad Saleh
5 6 7
3200 GE Healthcare +966 500612562
5443 1650 Imad.saleh@ge.com

OPTIMA XR 646
PRELIMINARY STUDY
2-142 2-141B 2-155
ORTH PHYSICIAN ORTHO COORDINATOR PEDS CAST ROOM
EXAM ROOM

A 19/DEC/2019 First issue drawing (DC-119070)


REV DATE MODIFICATIONS
A mandatory component of this drawing set is the GE Healthcare Pre Installation manual.
Failure to reference the Pre Installation manual will result in incomplete documentation required for site design and preparation.
Pre Installation documents for GE Healthcare products can be accessed on the web at: www.gehealthcare.com/siteplanning
GE does not take responsibility for any damages resulting from changes on drawings made by others. Errors may occur by not referring to the complete set of
final issue drawing. GE cannot accept responsibility for any damage due to the partial use of GE final issue drawings, however caused.
Do not scale from printed pdf files. GE accepts no responsibility or liability for defective work due to scaling from these drawings.

Drawn by Verified by S.O. (GON) Concession PIM Manual Rev


Notes:
A) Control wall height: 2.20 m 2-141A E. Vékony V. Schäffer - - 5643854-1EN 9
ORTHO WORKROOM
Format Scale File Name Date Sheet
A3 1:50 RAD-B163865-PRE-00.DWG 19/DEC/2019 01/02
TEMPERATURE AND HUMIDITY SPECIFICATIONS POWER REQUIREMENTS

IN-USE CONDITIONS POWER SUPPLY


EXAM ROOM CONTROL ROOM FREQUENCIES
Min Max Min Max
Temperature POWER DEMAND 97kVA
15°C [59°F] 35°C [95°F] 15°C [59°F] 35°C [95°F]
MAXIMUM LINE RESISTANCE 380V : 0.118 / 400V : 0.131 / 415V : 0.138
Temperature gradient
PER 2 PHASES (Ohm) 440V : 0.154 / 480V : 0.185
Relative humidity (1) 30% to 60% 30% to 60%
Humidity gradient < 30%/h < 30%/h
Stand by Max Stand by Max · Power supply should come into a power distribution box (PDB) containing the protective units and controls.
System heat dissipation 0.896 kW 2.432 kW 0.176 kW 0.253 kW · The section of the supply cable should be calculated in accordance with its length and the maximum
3057 BTU/hr 8298 BTU/hr 601 BTU/hr 863 BTU/hr permissible voltage drops.
· There must be discrimination between supply cable protective material at the beginning of the installation
STORAGE CONDITIONS (main low-voltage transformer side) and the protective devices in the PDB.
Temperature -5°C [23°F] to +50°C [122°F]
SUPPLY CHARACTERISTICS
Temperature gradient · Power input must be separated from any others which may generate transients (elevators, air conditioning,
Relative humidity (1) 10% to 85% radiology rooms equipped with high speed film changers...)
Humidity gradient < 30%/h · All equipment (lighting, power outlets, etc...) installed with GE system components must be powered
Storage longer than 90 days is not recommended. separately.
(1) Non-condensing
AIR RENEWAL
According to local standards.

NOTE
In case of using air conditioning systems that have a risk of water leakage it is recommended not to install it above electric equipment or to
take measures to protect the equipment from dropping water.

DELIVERY DISCLAIMER

THE CUSTOMER/CONTRACTOR SHOULD: This drawing is a preliminary drawing. Site conditions and/or equipment configuration may have a significant
impact on room layout and site preparation. Final study must be done before installation of the GE equipment. GE
· Provide an area adjacent to the installation site for delivery and unloading of the GE equipment. cannot accept any responsibility for errors due to lack of information.
· Ensure that the dimensions of all doors, corridors, ceiling heights are sufficient to accommodate the
movement of GE equipment from the delivery area into the definitive installation room. The room dimensions used to create the equipment layout may originate from a previous layout and may not be
· Ensure that access routes for equipment will accommodate the weights of the equipment and any accurate as they may not have been verified on site. GE cannot take any responsibility for errors due to lack of
transportation, lifting and rigging equipment. information.
· Ensure that all necessary arrangements for stopping and unloading on public or private property not
belonging to the customer have been made. It is the responsibility of the customer to prepare the site in accordance with the specifications stated in the final
drawings. These drawings are not to be used for actual construction purposes. The company cannot take
responsibility for any damage resulting therefrom.

DIMENSIONS OF DELIVERY IN TRANSIT The customer must ensure the floor strength is sufficient to support the fixings as required. A qualified structural
engineer must be consulted and all work carried out according to his specifications.
EQUIPMENT DIMENSIONS WEIGHT
LENGTH 2111 mm 83.1 in Suitable radiological protection must be determined by a qualified radiological physicist in conformation with local
270 kg + 596 lbs + regulations. GE does not take responsibility for the specification or provision of radio-protection.
STANDARD WALLSTAND WIDTH 911 mm 35.9 in
fixture fixture
HEIGHT 1860 mm 73.2 in
LENGTH 2400 mm 94.5 in
G2 TABLE WIDTH 940 mm 37 in 440 kg + dolly 970 lbs + dolly
HEIGHT 800 mm 31.5 in THE UNDERSIGNED, HEREBY CERTIFIES THAT I HAVE READ AND APPROVED THE PLANS IN THIS DOCUMENT.

DATE NAME SIGNATURE


Pay attention to the lengths of the rails! They can also be 5.79 m [19 ft] and have a shipping dimension of 5.92 m x
178 mm x 76 mm [16'-10" x 7" x 3"].

AL MOOSA SPECIALIST HOSPITAL OPTIMA XR 646 RAD-B163865-PRE-00.DWG 1:50 Rev A Date 19/DEC/2019 Details 02/02