Professional Documents
Culture Documents
INSTALLATION / OPERATIONAL
QUALIFICATION PROTOCOL
MUFFLE FURNACE
AURIK PHARMACEUTICALS
NATIONAL INDUSTRIALZONE RAWAT ISLAMABAD .
Document Author
Name Position Signature Date
Approval
Name Position Signature Date
1. APPROVALS
DR.KHALID QCM
Comments
TABLE OF CONTENTS
Section Page
1. APPROVALS.............................................................................................................................................................2
2. DOCUMENT CONTROL..........................................................................................................................................3
3. INTRODUCTION & SCOPE.....................................................................................................................................4
4. DATA COLLECTION...............................................................................................................................................5
5. NON-CONFORMANCES..........................................................................................................................................6
6. PERSONS PERFORMING IOQ................................................................................................................................6
7. PRE-REQUISITES TO COMMENCE IQ SECTION...............................................................................................7
8. INSTALLATION CHECK (IQ).................................................................................................................................8
9. COMMISSIONING CHECK (IQ)...........................................................................................................................10
2. DOCUMENT CONTROL
3.1 Introduction
The purpose of this IOQ is to qualify the refrigerator (detailed below) to confirm that it has been installed according to
the design intent and operates in accordance with the manufacturer’s specifications, drawings and other relevant design
documents.
The IOQ will cover the following refrigerator:
Equipment Name Make & Model Identification No. Location
All personnel entering information in this protocol must be identified in the table below by name, affiliation, full
signature and initials.
Personnel must be fully aware of all the procedures to follow, specifically procedures for data collection and recording
discrepancies.
Name Company Signature Initials Date
6.1 Objective
To verify that the Furnace has been installed to a satisfactory standard and is safe to use.
To confirm there is adequate documentation to support the on-going operation and maintenance of the Furnace.
6.2 Pre-Requisites
Pre-Requisite Description Initials Comments
Table 2
Check Comment
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
6.6 Conclusion
This section of the protocol has been completed and is deemed to (initial appropriate box below):
Comments Non-conformance
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
7.1 Objective
To verify that the refrigerator has been satisfactorily commissioned and a commissioning report has been provided by
the supplier.
7.2 Pre Requisites
Pre-Requisite Description Initials Comments
Table 2
Check Comment
The commissioning report has been adequately
completed and is signed and dated
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
7.6 Conclusion
This section of the protocol has been completed and is deemed to (initial appropriate box below):
Comments Non-conformance
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
9.1 Objective
To verify that all Furnace components function correctly as specified for normal operation.
To verify that all alarms function as specified.
9.2 Pre Requisites
Pre-Requisite Description Initials Comments
Adjustments to environmental parameter operational set points must result in a corresponding response from the
Furnace control system.
Environmental parameter alarms must trigger in accordance with the set points above.
9.5 Results
Table 1
Check Comment
Record temperature set points (for both operating level Operating Set Point:
and the alarm limits) and also confirm these meet the Low Alarm Set Point:
acceptance criteria High Alarm Set Point:
Hold door open to confirm the alarm activates and record
the delay time. Close door and confirm alarm de-
activates
Increase the temperature set point by ≥3oC. Confirm that
the Furnace system responds to increase the temperature
and the alarm activates (visual and audible) when
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
Table 1
Check Comment
temperature exceeds alarm high limit
Ensure temperature set point is returned to the original
value on completion and confirm alarm de-activates
Decrease the temperature set point by ≥3oC. Confirm
that the Furnace system responds to decrease the
temperature and the alarm activates (visual and audible)
when temperature exceeds alarm low limit
Ensure temperature set point is returned to the original
value on completion and confirm alarm de-activates
9.6 Conclusion
This section of the protocol has been completed and is deemed to (initial appropriate box below):
Comments Non-conformance
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
10.1 Objective
To verify that the Furnace functions correctly in maintaining the correct temperature conditions throughout.
To determine how quickly temperature limits are exceeded when the door is opened or when there is a significant loss
of power, and how quickly temperature returns to within operating limits once the door is closed or and/or power
restored.
Note: By simulating a loaded refrigerator, this will provide data to support Performance Qualification requirements.
10.2 Pre Requisites
Pre-Requisite Description Initials Comments
Furnace is filled with simulated product to represent a
typical full load
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
10.8 Results
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
10.9 Conclusion
This section of the protocol has been completed and is deemed to (initial appropriate box below):
Comments Non-conformance
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
11.1 Objective
To verify that appropriate Standard Operating Procedures (SOP’s) have been identified to operate, clean and maintain
the refrigerator and these should be at least in draft format.
To confirm that a suitable level of training to operate, clean and maintain the Furnace has been identified.
11.2 Pre Requisites
Pre-Requisite Description Initials Comments
A system is in place to formally issue and control
operating procedures and instructions
There is a system for identifying and documenting
training
Table 2
Date
Personnel Identified Description Pass / Fail
Performed/Planned
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
Table 2
Date
Personnel Identified Description Pass / Fail
Performed/Planned
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
11.6 Conclusion
This section of the protocol has been completed and is deemed to (initial appropriate box below):
Comments Non-conformance
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
The following provides an outline summary of the test sections in this protocol.
Non
Check Pass / Pass with Comment / Fail
Conformance
IOQ-AURK-10-019UNITED TECHNOLOGIES
IQ-OQ AURIK PHARMACEUTICALS
SUPPLIER: UNITED TECHNOLOGIES
No Details
IV Additional Data
IOQ-AURK-10-019UNITED TECHNOLOGIES