You are on page 1of 4

Radiotherapy and Oncology 90 (2009) 106–109

Contents lists available at ScienceDirect

Radiotherapy and Oncology


journal homepage: www.thegreenjournal.com

Treatment verification

Use of skin markers and electronic portal imaging to improve verification


of tangential breast irradiation
Anke van der Salm a, Jennifer Strijbos a, Colette Dijcks a, Lars Murrer a,b,
Jacques Borger a,b, Liesbeth Boersma a,b,*
a
Maastro Clinic, Maastricht, The Netherlands
b
Department of Radiation Oncology (Maastro Clinic), GROW, University Hospital Maastricht, The Netherlands

a r t i c l e i n f o a b s t r a c t

Article history: We investigated the added value of skin markers in 566 electronic portal images (EPIs) in 48 breast can-
Received 11 April 2008 cer patients treated with tangential fields. EPIs were matched to the corresponding DRRs using skin
Accepted 3 May 2008 markers, anatomy, or a combination of both. Skin markers improved determination of setup errors in cra-
Available online 12 June 2008
nio-caudal direction.
Ó 2008 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 90 (2009) 106–109
Keywords:
Portal imaging
Breast cancer
Skin markers
Treatment verification

Current methods for treatment setup verification of breast irra- Patients and methods
diation are usually based on matching bony anatomical landmarks,
as visible on the Electronic Portal Images (EPIs), to the same ana- Patients
tomical landmarks visible in the digital reconstructed radiographs
(DRRs) [2–5,7–10]. Usually, portal images from the radiation fields Data were obtained in 48 breast cancer patients (Tis-4N0-2M0)
are taken [2–5,7–10], although sometimes separate orthogonal treated with post-operative radiotherapy. Thirty-seven patients
additional beams are used, designed to check the isocenter [1]. were treated with breast conserving therapy, and 11 patients were
However, verification of tangential breast irradiation using bony referred for post-mastectomy radiotherapy. All patients were trea-
anatomy may not be sufficiently adequate, since the target volume ted with tangential field irradiation.
(i.e. the breast) might move with respect to the bones. In addition
to this, some bony anatomy visible in the EPIs is located at some
distance of the target tissue (e.g. the spine). Recently, Bert et al. CT-simulation and planning
[1] and Spadea et al. [6] both described a method that showed
the feasibility of the use of skin alignment instead of bony anat- CT-simulation was performed using a standard protocol. In
omy. They found a significant improvement in patient setup using short, a planning CT was obtained with the patient in treatment
the skin alignment. Although the results of both studies were quite position. Three longitudinal laser lines and one transversal laser
good, these methods require installation of expensive new equip- line were positioned on the skin of the patient to ensure accurate
ment. Therefore, the aim of this study was to assess the feasibility and reproducible positioning. The palpable breast tissue, or the re-
and additional value of a relatively simple and more generally gion where palpable breast tissue was assumed to have been pres-
applicable method for portal imaging verification of patient setup ent (after mastectomy) (i.e. the Clinical Target Volume – CTV) was
using skin markers, for patients treated with tangential fields. marked with a copper wire prior to the CT-scan. In addition, four
We compared three methods of matching DRRs and EPIs: matching copper skin markers (Fig. 1A) were placed to mark the CTV at the
based on anatomy only (lung and breast contour), skin markers most cranial, caudal, medial, and lateral border of the palpable
only, or a combination of anatomy and skin markers. breast tissue (CTV) (Fig. 1B). Each skin marker contains three
straight or round markers, enabling recognition of lateral (straight)
and medial (round) markers on the EPI in case of overprojection.
CT-images were obtained at 3-mm slice thickness from the level
* Corresponding author. Maastro Clinic, Postbus 5800, 6202 AZ Maastricht, The
of the mandible down to the diaphragm. The position of the skin
Netherlands.
E-mail address: liesbeth.boersma@maastro.nl (L. Boersma). markers was indicated with ink to ensure reproducible placement

0167-8140/$34.00 Ó 2008 Elsevier Ireland Ltd. All rights reserved.


doi:10.1016/j.radonc.2008.05.014
Anke van der Salm et al. / Radiotherapy and Oncology 90 (2009) 106–109 107

Fig. 1. Example of the skin markers (A), how they are placed onto the skin (B), and how they project in the DRR (C) and EPI (D). Each skin marker contains three straight or
round markers, spaced 1 cm apart. They are made of 1 mm copper (CT) or 1 mm gold (linear accelerator). The two shapes allow distinction of the lateral and medial markers.
The lateral and medial markers were shifted slightly in cranio-caudal direction to minimize overprojection. (C) The markers in the DRR, and (D) in the EPI. Clearly visible are
lung contour, breast contour, and four markers, denoted by Mcran, Mcaud, Mlat and Mmed for the cranial, caudal, lateral and medial markers, respectively. By manual
matching of the delineated structures with the EPI, values of the setup error are obtained in x- and y-direction of the beam’s eye view.

in treatment position. A small piece of TegadermTM was stuck over Theraview-NT 2.4 (Cablon Medical) (Fig. 1D) [3]. The image regis-
these short lines, to prevent vanishing of the lines. tration was performed using three different methods:
Standard tangential fields were designed using virtual simula-
tion. The copper markers at the CTV borders were delineated using (1) Anatomy match: Only using anatomical landmarks, i.e. the
the treatment planning system (XiO, Computer Medical System, lung and breast contour, thereby ignoring the skin markers
Inc.) in order to be visualized on the digitally reconstructed radio- (these were not switched on in the reference contour set
graphs (Fig. 1C). during matching).
(2) Marker match: Only using the skin markers.
Radiation treatment and acquisition of EPIs (3) Clinical match: Using the skin markers with additional help
of the anatomical landmarks. We assume that a pure marker
Patients were placed in correct position using the longitudinal match will not be used in clinical practice, because using the
and transversal laser lines. The patient reference point was first additional information of the visible anatomy is judged to be
positioned to the machine isocenter and from there a table shift superior to the use of markers only. Therefore, we added the
was performed to move the patient to the correct position with clinical match, which will be commonly used in clinical
the isocenter in the breast. EPIs of the radiation fields were per- practice.
formed weekly with the four skin markers in place. At the linear
accelerator gold markers were used with the same geometry as Displacements were recorded in x- and y-direction in the
the copper markers, for better visibility with megavolt imaging. beam’s eye view of the separate fields. The y-direction represents
EPIs were made in both medio-lateral (ML) and latero-medial pure cranio-caudal direction, the x-direction represents a mixture
(LM) directions using a Theraview-NT system (Cablon Medical) of left–right direction and table height, depending on the gantry
[3] at a source detector distance of 150 cm. A total number of angle. Each type of match was performed by one of three observ-
566 EPIs was acquired, in a mean number of 5.9 fractions per pa- ers. To estimate the interobserver variation, all matches were per-
tient (range 4–9 fractions). formed by three observers, in a subset of ten patients with six
treatment fractions analyzed each. The interobserver variation for
Registration of the EPIs of the treatment fields with the DRRs one patient was calculated as the standard deviation of the three
match values per fraction, and was then averaged over all fractions
The displacement of the EPI with respect to the DRR was quan- analyzed. The estimate of the interobserver variation for this sub-
tified after manual registration with the corresponding DRR, using population (and hence for the entire population) was determined
108 Skin markers for EPI in breast cancer

by averaging the mean standard deviation over these 10 patients. accelerator. By using only beam angles identical to the treatment
The value of this mean standard deviation was small (1.1– fields, we were unable to correct for setup errors in the direction
1.6 mm) for all directions. of the beam axis. However, since a setup error in that direction
only results in a different source skin distance, we considered the
Analysis of the match-data effect of these setup errors on the dose-distribution to be of minor
importance. In patients in whom an additional boost to the tumor
First, the EPIs were analyzed qualitatively, to check whether the bed is given using an additional orthogonal beam, this simple EPI
skin markers (CTV) were located within the radiation portal. In pa- procedure obviously has to be extended with an additional orthog-
tients where the position of the markers changed >3 mm with re- onal EPI-beam.
spect to each other, only the cranial and medial markers were used,
since these were considered to be the most stable markers because Comparison of the different matches
they are located at a relatively stable anatomy.
Next, the average setup error with standard deviation was cal- Comparison of the clinical, anatomy, and marker matches
culated for each patient in two directions, for both the ML and showed no significant differences in average values of l0 , R0 , and
the LM fields. Although we did not measure full 3D setup errors, r0 . To analyze the data in more detail, we calculated the average
we analyzed our data in analogy to studies that applied 3D correc- difference in match value between the methods per patient, and
tions: (1) we calculated the systematic error (mean of means) for subsequently averaged these values over all patients (Table 1A).
each beam and each direction l0 , analogue to the l determined Although on average no significant differences were seen between
in full 3D; (2) we calculated the SD of l0 (R0 ), analogue to the SD either of the matches (Table 1A), the SD of the average differences
of the systematic error R in full 3D and (3) we calculated the aver- between matches that both used skin markers (clinical and marker
age of the SD per patient r0 , analogue to the SD of the random set- matches) tended to be smaller (range 1.3–1.9 mm) than the SD of
up error r in full 3D [11]. These three parameters were calculated the difference between matches that did use skin markers vs. with-
for the three different matches. Finally, the difference between the out skin markers (range 1.7–3.6 mm) (Table 1A). This suggests that
anatomy match, clinical match, and marker match was calculated the clinical matched is dominated by the presence of skin markers
for each EPI in each direction. Differences between matches and rather than by the anatomy.
between the x- and y-direction were tested for significance using Since we had the clinical impression that the markers were
the Students’ t-test. Assuming that a difference in match value especially helpful in cranio-caudal direction, we analyzed the var-
>5 mm is clinically relevant, we also evaluated the number of pa- iation of the match-differences in more detail. We counted the
tients with an average difference >5 mm for the different matches. number of patients that would be matched differently by a clini-
cally relevant distance of >5 mm. We found that >5 mm differences
Results and discussion in the y-direction (on average over the entire treatment) were ob-
served in about 6% of the patients comparing the anatomy match
Overall setup accuracy with the clinical match, whereas in x-direction all average differ-
ences were <5 mm (Table 1B). When comparing the anatomy
In all EPIs, the skin markers indicating the outline of the CTV on match with the marker match, this percentage was even higher
the skin were located within the radiation field, indicating that at (10%) (Table 1B). Although these percentages obviously only repre-
least large setup errors did not occur. This was confirmed when sent a minority of patients, this observation stresses the impor-
the match values were quantified. The mean of means l0 and R0 tance of the use of the skin markers for detecting outliers in the
for all three types of matches were <1.8 mm and <3.6 mm, respec- cranio-caudal direction. The additional value of the skin markers
tively, in both the x- and y-direction, in both LM and ML fields. in cranio-caudal direction is attributed to the lack of specificity
The SD of the random error r0 was <3.6 mm in all beams and direc- of the anatomical landmarks to correctly define the cranio-caudal
tions. The results of our ‘‘anatomy match” were comparable with position as also pointed out by van Tienhoven et al. [7].
the results of other studies, using anatomical landmarks for match-
ing. In our study, R0 of the anatomy match varied from 2.1 to 3.3 mm Feasibility of the use of skin markers
for the different beams and directions, whereas these figures varied
from 1.7 to 4.6 mm in other studies [2,4,5,7,9]. For r0 , these figures The position of the skin markers on the patient skin was marked
were 2.3–3.1 mm in our study, vs. 1.7–3.4 mm in other studies [4,5]. with ink as a short line. Tegaderm was used to cover the ink marks;
An obvious limitation of this study is that we did not measure however, when the Tegaderm loosened from the skin, the ink line
setup errors in full 3D. We deliberately did not choose to use an could disappear completely. Therefore, we are currently using
additional orthogonal EPI beam, to avoid additional radiation expo- Pointguards (www.beekley.com); this new material appears to
sure to the lungs, and to avoid increased workload at the linear yield much less irritation of the skin, and stays longer securely in

Table 1
Comparison of the different matches. Differences are given for the three different match procedures, for the ML beam and the LM beam, both in x- and y-direction of the beam’s
eye view

ML-x ML-y LM-x LM-y


(A) Average differences between each match ±SD (mm)
Anatomy – clinical match 0.1 ± 1.7 0.7 ± 2.8 0.1 ± 1.7 0.8 ± 3.0
Clinical – marker match 0.1 ± 1.3 0.2 ± 1.8 0 ± 1.6 0.3 ± 1.9
Anatomy – marker match 0 ± 2.1 0.9 ± 3.2 0.1 ± 2.1 1.0 ± 3.6
(B) Number of patients (%) with an average difference over the entire treatment >5 mm
Anatomy – clinical match 0 (0%) 3 (6%) 0 (0%) 3 (6%)
Clinical – marker match 0 (0%) 0 (0%) 0 (0%) 0 (0%)
Anatomy – marker match 0 (0%) 5 (10%) 0 (0%) 5 (10%)

The y-direction represents pure cranio-caudal direction, the x-direction represents a mixture of left–right direction and table height, depending on the gantry angle.
Anke van der Salm et al. / Radiotherapy and Oncology 90 (2009) 106–109 109

place. The whole skin marker method appeared to be a very easy a minimal additional burden on the daily workload. The clinical
and quick procedure, completely performed by the radiation tech- match is therefore now been employed routinely in our clinic.
nologists: 2 additional minutes at the CT simulator for placing the The data concerning systematic and random errors will allow fu-
skin markers into the skin, and for drawing the short inklines and ture implementation of automated off-line setup verification and
putting Tegaderm onto the skin, and 1 additional minute at the lin- correction protocols.
ear accelerator to place the skin markers when EPIs were made,
once a week. References
A discrepancy >3 mm was observed in the configuration of the
four skin markers in about 30% of the measurements. In these [1] Bert C, Metheany KG, Doppke KP, Taghian AG, Powell SN, Chen GT.
Clinical experience with a 3D surface patient setup system for alignment
cases, only the medial and cranial markers were used for matching, of partial-breast irradiation patients. Int J Radiat Oncol Biol Phys
since the position of these markers was considered to be most sta- 2006;64:1265–74.
ble. This may lead to the question whether all four markers are re- [2] Creutzberg CL, Althof VG, Huizenga H, Visser AG, Levendag PC. Quality
assurance using portal imaging: the accuracy of patient positioning in
quired for an adequate match. The reason for us to still stick to the irradiation of breast cancer. Int J Radiat Oncol Biol Phys 1993;25:529–39.
use of all four markers, is that in about 20% of the patients one of [3] Dirkx ML, de Boer JC, Heijmen BJ. Improvement of radiotherapy treatment
the ink marks is lost. By using a redundancy of short lines (four) we delivery accuracy using an electronic portal imaging device. Radiat Prot
Dosimetry 2006;121:70–9.
make sure that always at least two lines are preserved. In addition, [4] Kukolowicz PF, Debrowski A, Gut P, Chmielewski L, Wieczorek A, Kedzierawski
presence of the four markers also allows quick visual inspection, to P. Evaluation of set up deviations during the irradiation of patient suffering
check whether all markers, which give a rough indication of the from breast cancer treated with two different techniques. Radiother Oncol
2005;75:22–7.
CTV, fall within the radiation portal.
[5] Lirette A, Pouliot J, Aubin M, Larochelle M. The role of electronic portal imaging
The discrepancy between the markers may be caused by either in tangential breast irradiation: a prospective study. Radiother Oncol
a change in breast shape or an inaccurate placement of the skin 1995;37:241–5.
[6] Spadea MF, Baroni G, Riboldi M, et al. Patient set up verification by infrared
markers. A suspicion of persistent change in breast shape would
optical localization and body surface sensing in breast radiation therapy.
prompt a new CT-scan and treatment plan, especially of impor- Radiother Oncol 2006;79:170–8.
tance for the boost plan. However, in our patient group this was [7] Tienhoven van G, Lanson JH, Crabeels D, Heukelom S, Mijnheer BJ. Accuracy in
necessary in only two cases. tangential breast treatment set-up: a portal imaging study. Radiother Oncol
1991;22:317–22.
[8] Truong PT, Berthelet E, Patenaude V, et al. Short communication: setup
Conclusions variations in locoregional radiotherapy for breast cancer: an electronic portal
imaging study. Br J Radiol 2005;78:742–5.
[9] Valdagni R, Italia C. Early breast cancer irradiation after conservative surgery:
We showed that skin markers in combination with bony ana- quality control by portal localization films. Radiother Oncol 1991;22:311–3.
tomical landmarks (‘clinical match’) facilitate and improve the [10] Venables K, Miles EA, Hoskin PJ, Aird EG. START. Verification films: a study of
the daily and weekly reproducibility of breast patient set-up in the START trial.
determination of setup errors in cranio-caudal direction. Although
Clin Oncol (R Coll Radiol) 2005;17:337–42.
a clinically relevant improvement was seen only in a minority of [11] van Herk M. Errors and margins in radiotherapy. Semin Radiat Oncol
patients (6%), skin markers are very easy to apply, and cause only 2004;14:52–64.

You might also like