You are on page 1of 3

An American National Standard

Designation: E 1323 – 89 (Reapproved 2002)

Standard Guide for


Evaluating Laboratory Measurement Practices and the
Statistical Analysis of the Resulting Data1
This standard is issued under the fixed designation E 1323; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope ANSI Z34.1 American National Standard for


1.1 This guide covers key elements of an evaluation of a Certification—Third Party Certification Program4
laboratory’s measurement practices and the statistical analysis 2.3 ISO Standard:
of the resulting data. This guide addresses an evaluation that ISO 9000–9004 Quality Management and Quality Assur-
covers a broad range of in-house quality measurements, some ance Standards3
of which may be directly related to accreditation requirements. 2.4 Other Standard:
1.2 This guide describes what to look for as documentation ILAC/ISO Laboratory Accreditation—Principles and
in order to verify the operation of the practices, and what parts Practice—Collected Reports 1979–19834
of the data to test and interpret in order to verify the quality of 3. Terminology
data being generated by the laboratory.
1.3 This guide does not specify or provide guidance for the 3.1 Terms are defined in Terminology E 1187.
establishment or assessment of a quality program. 3.2 Definitions of Terms Specific to This Standard:

--````,,`,`,`,``,,`,,,,,`,`,,,`-`-`,,`,,`,`,,`---
3.2.1 duplicate—a separate specimen, taken from the same
NOTE 1—Generic guidelines dealing with elements of a quality pro- source as the first specimen, tested at the same time and in the
gram may be found in ANSI, A1, A3, Q1, Q2, Z34.1; ISO9000–9004; and
same manner as the first specimen. Duplicates can provide
ILAC/ISO Collected Reports 1979–1983.2
pooled precision data for a homogeneous specimen, the test
2. Referenced Documents method, and the test equipment.
2.1 ASTM Standards: 3.2.2 out of control—the condition that exists when a data
E 548 Practice for Preparation of Criteria for Use in the point falls outside the control limits and when retesting and
Evaluation of Testing Laboratories and Inspection Bodies3 further evaluation indicates that a problem exists in analyst
E 994 Guide for Laboratory Accreditation Systems3 performance, the method, equipment, standardization, or cali-
E 1187 Terminology Relating to Laboratory Accreditation3 bration.
2.2 ANSI Standards: 3.2.3 replicate analysis—the same specimen tested again,
ANSI/ASQC A1 Definitions, Symbols, Formulas and usually at a different time. Replicate data can be used to
Tables for Control Charts4 provide pooled precision of the test method, equipment, and
ANSI/ASQC A3 Quality Systems Terminology4 operator, providing the specimen is homogeneous in nature.
ANSI/ASQC Q1 Generic Guidelines for Auditing of Qual- 4. Significance and Use
ity Systems4
ANSI/ASQC Q2 Quality Management and Quality System 4.1 This guide is intended to provide guidance for an
Elements for Laboratories—Guidelines4 assessor to evaluate measurement practices of laboratories,
their protocol for statistically analyzing the resulting data from
these practices, and their statistical results from these practices.
1
This guide is under the jurisdiction of ASTM Committee E36 on Criteria for the 4.2 This guide is generic in the sense that it covers the entire
Evaluation of Testing and Inspection Agencies and is the direct responsibility of range of in-house quality measurement practices found in a
Subcommittee E36.60 on Laboratory Accreditation Operations. testing laboratory, but the results of the described evaluation
Current edition approved Dec. 29, 1989. Published February 1990.
2
ILAC/ISO Laboratory Accreditation-Principles and Practice-Collected Reports may be used by accrediting agencies if their requirements can
1979–1983. American Association for Laboratory Accreditation, 656 Quince Or- be satisfied through the laboratory’s existing quality data.
chard Rd. #704 Gaithersburg, MD 20878 4.3 It is not the intent of this guide to serve as sole criterion
3
Annual Book of ASTM Standards, Vol 14.02.
4
Available from American National Standards Institute, 1430 Broadway, New
for evaluating and accrediting laboratories. It is not intended to
York, NY 10018. cover the important generic guidelines for evaluating the

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

Copyright ASTM International 1


Provided by IHS under license with ASTM Licensee=Empresas Publicas de Medellin/5900316001
No reproduction or networking permitted without license from IHS Not for Resale, 04/21/2006 11:47:18 MDT
E 1323 – 89 (2002)
laboratory’s quality program which can be found in Practice 6.1.2.3 The time interval for testing, or a date for each data
E 548 and Guide E 994. point if a time interval is not practical, must be noted.
6.1.2.4 Directions on how to statistically compare the labo-
5. Purpose of Evaluating Measurement Practices and the ratory results of precision with a known statement of precision
Statistical Analysis of the Resulting Data for that method must be clearly documented. These directions
5.1 Data generated from the measurement practices of a must include the specific statistical test, the number of data
laboratory is evaluated to determine its capability to obtain points used for the test and the acceptable level of precision, be
accurate and precise data and to determine if the laboratory it known either from other studies on this specific method or as
correctly and efficiently analyzes and reacts to their own data. a limit determined by the laboratory itself.
6. Documentation of Measurement Practices and the 6.1.2.5 The method for determining if outliers exist must
Statistical Protocol for Analyzing the Resulting Data also be documented, stating when it is acceptable to ignore
such data points when computing control limits.
6.1 The following documentation shall contain the informa-
6.1.2.6 The method for determining if bias exists and the
tion necessary for someone to carry out the same steps
frequency for continued checks on the instrument or method
involved in generating, charting, and reporting the data.
having a bias must be recorded, including any adjustments
6.1.1 Documentation Relative to Calibration:
made to test data as a result of the bias determined from these
6.1.1.1 The material to be measured must be documented
measurements.
together with its source, expiration or shelf-life date, the
accuracy and its source, and any preparations or conditions 6.1.2.7 A table of actual values and the corresponding dates
required which are specific to this material before it can be must be kept.
utilized as a calibration material. Any additional components, 6.1.3 Documentation Relative to Operator Precision and
reagents, or physical sources used along with this material Accuracy:
which could potentially alter the reliability of the material must 6.1.3.1 The material, methods, and equipment used to de-
also be documented. termine levels of precision and accuracy of operators must be
6.1.1.2 The identification of the equipment used, together documented.
with the date and operator responsible for the run, and any 6.1.3.2 The source of stated accuracy of the material, the
preparations involved with the calibration run must be docu- current precision and accuracy of the equipment, and the
mented. current precision and bias of the method must be documented,
6.1.1.3 The type of data representation to be used, including together with the exact computations used to determine the
the exact number of data points to be used in the computation single or group operator precision, accuracy, or both.
of an average, standard deviation, or range, as well as how and 6.1.3.3 The number of data points generated per operator for
when these data points are to be generated must be docu- this comparison test as well as the type (duplicate, replicate or
mented. This requires information regarding testing of repli- single) and description of the type of runs being made. It must
cates, duplicates, or single runs tested on one day, a series of be clearly stated which runs are related to the accuracy and
days, or a specific time interval to be clearly stated for each set which are related to precision.
of data. 6.1.3.4 The limits of acceptability for both operator accu-
6.1.1.4 The mathematical formula for obtaining control racy and operator precision must be noted.
limits, the frequency of computing new limits together with 6.1.3.5 The records must show the frequency of obtaining
rules of acceptability of the new limits, must be documented operator accuracy and precision as well as the corrective action
whenever control limits are applied to a chart. taken whenever an operator fails to meet the limits of accept-
6.1.1.5 The corrective action taken whenever data points ability.
indicate that an out of control condition exists, or whenever
6.1.3.6 A table of actual values and corresponding dates for
trend analysis indicates a change or shift in the instrument
each operator must also be documented.
response must be recorded.
6.1.1.6 A table of actual values for each calibration or
7. Evaluation of the Laboratory’s Measurement Practices
calibration check and the corresponding date must be kept.
and of the Statistical Analysis from the Resulting Data
6.1.2 Documentation Relative to Method Precision and
Bias: 7.1 A general overview of the laboratory’s documentation
6.1.2.1 The reference of the specific method being followed of measurement practices and the statistical analysis of the data
for each set of data, as well as any changes to the method must should be made, verifying that all practices cited are actually in
be documented. If a method has not been published, then the use and that the rules stated by the laboratory for the generation
laboratory is responsible for preparing a detailed procedure. and use of the data are followed.
6.1.2.2 The type of run (duplicate, replicate, single) used to 7.2 A walk-through of the laboratory to verify the location
generate the data points, including specific directions on how of all data representation charts must be made as well as
to prepare and test a duplicate or replicate specimen, must be general discussions with laboratory personnel to verify their
recorded. knowledge of the measurement practices.

--````,,`,`,`,``,,`,,,,,`,`,,,`-`-`,,`,,`,`,,`---

Copyright ASTM International 2


Provided by IHS under license with ASTM Licensee=Empresas Publicas de Medellin/5900316001
No reproduction or networking permitted without license from IHS Not for Resale, 04/21/2006 11:47:18 MDT
E 1323 – 89 (2002)
7.3 The choice of representation of the data, as well as the 7.8 Raw data checks must also be made to verify the
generation of control chart limits must be reviewed to deter- frequency of testing for calibration. For example, a measure-
mine if appropriate models have been used for each type of ment practice that requires that a replicate be tested on 1 of
population. For further information, see ASTM STP 15D.5 every 10 specimens received, must show, through raw data,
7.4 The raw data from the calibration records must be that the appropriate number of replicate tests were performed,
compared to status indicated for that piece of equipment. Every reported, and transferred to the corresponding chart.
out of control or out of specification data point in the raw data 7.9 Statistical tests, such as t, f, chi square, ANOVA, or
must have a corresponding corrective action which resulted in other related tests as found in statistical analysis text books,
restoring the equipment to normal operation (status of in that have been performed by the laboratory, must be verified
calibration) or resulted in a recommendation for further work for accuracy. If none of these tests were performed then the
(status of out of calibration). Dates and results from the raw assessor could perform the tests for the laboratory, interpret the
data must agree with the status shown on that piece of results, and offer references to help prepare the laboratory to
equipment. perform these computations in the future.
7.5 All charts must be viewed for trend analysis. Checks of
8. Use of the Evaluation by Accrediting Bodies
trends or shifts must include the interpretation of such by
laboratory personnel as well as the corrective action taken. This 8.1 An accrediting body could use this evaluation to deter-
information does not have to be on the chart itself, but it must mine the extent of a laboratory’s participation in proficiency
be documented somewhere by the laboratory and available for testing programs, based on the credibility of the laboratory’s
review upon request by the assessor. statistical results of their measurement practices.
7.6 All data calculations for any kind of limits, be it for 8.2 An accrediting body may want to use the evaluation to
accuracy or precision of equipment, precision or bias of determine the extent and frequency of future on-site visits to
methods, or accuracy or precision of an operator, must be the laboratory. On site visits could possibly be replaced with
reviewed to determine if the appropriate type of chart has been correspondence by mail of the data from the measurement
used and the optimum number of data points have been used. practices in order to verify the laboratory’s continued state of
7.7 Spot checks of the raw data must also be made to verify statistical control.
that no data point is being disregarded unless an outlier test has 8.3 The accrediting body may want to use the evaluation to
indicated that the data point can be omitted from the calcula- determine the needs of the laboratory as far as education is
tion of control limits. The outlier is still documented as a check concerned. They could refer the laboratory to adequate sources
and listed in the table of data with corresponding dates. of reading material and short courses so that the laboratory
could make improvements in their overall quality program.
9. Keywords
5
Manual on Presentation of Data and Control Chart Analysis, ASTM STP15 D, 9.1 accreditation; laboratory accreditation; laboratory mea-
ASTM, 1982. surement practices; statistical analysis

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

--````,,`,`,`,``,,`,,,,,`,`,,,`-`-`,,`,,`,`,,`---
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

Copyright ASTM International 3


Provided by IHS under license with ASTM Licensee=Empresas Publicas de Medellin/5900316001
No reproduction or networking permitted without license from IHS Not for Resale, 04/21/2006 11:47:18 MDT

You might also like