Professional Documents
Culture Documents
ge
au
ge
au
au
G
1
ge
au
3
G
ge
au
ge
4. Wiring
The RS strain gauges are fitted with 30 mm leads to
enable the gauge to be soldered. The lead out wires
are fragile and should be handled with care.
2
232-5957
G2
G,, G, active gauges
R3
+ Input + Bridge Voltage 1 24 + Vs
R, R, 1kΩ 0.1%
Vout wire wound resistors N/C 2 23 N/C
– Input Compensation 3 22 -Vs
G1 N/C 4 21 N/C
R4 N/C 5 20 Bridge Ref Input
– Bridge Supply + Input 6 19 N/C
N/C 7 18 Feedback
N/C 8 17 N/C
N/C 9 16 Output
- Input 10 15 N/C
N/C 11 14 N/C
- Bridge Voltage 12 13 Zero Adjust
Figure 5 Quarter bridge (3 wire) Top view
3
232-5957
Figure 7 Basic circuit for printed circuit board RS stock no. 435-692 (gain approx. 1000)
+VS
+
R7
R9 C3
T1 R10
1 24 0V
+ Bridge Supply
Compensation 3 22
C6 10k
R6 20 SET BRIDGE
+ Input 6 SUPPLY
18
R5 R1 C1
C7 C5 16 OUTPUT
- Input 10
- Bridge Supply 13
12 R2
T2 R3
D1
10k
C2 R4 SET ZERO
0V
0V D2
T3
+
R8 R11
C4
-VS
R7 1k R10
T1 R9
1 24
+ Bridge Supply
Compensation IN827
3 22
C6 IN827
20
+ Input 65
C 10k SET BRIDGE SUPPLY
18
R1
C7 16 OUTPUT
- Input 10
- Bridge Supply 1k
12 13 R2
T2 R3
D1
R4 10k
SET ZERO
0V
D2
T3
R8 R11
-VS
4
232-5957
T1 and T2 provide bridge currents up to 60mA and gauges convert the strain into an electrical signal which
should be kept away from amplifier. can then be displayed, used as a control signal, etc.
T3 and T5 provide stability power supplies are being
used zero and bridge supply reference may be taken
direct from the power rails. Single point load cells
The high output of some semiconductor strain gauges This RS range of load cells comprise of centre type
may cause large amounts of asymmetry to the bridge. (2kg, 20k & 10kg oil damped) in which a double beam
In correcting for the common mode change, the is used. Also in this range is a 100kg low profile off-
negative bridge voltage will change, causing a span centred load cell designed for direct mounting of the
error. This may be calibrated out or the arrangement weighing platform. They are supplied complete with a
above used to eliminate the cause of the error. Some full bridge of four strain gauges fitted and calibrated
semiconductor strain gauge transducers are ready to connect to any suitable amplifier.
temperature compensated by the use of series arm Three sizes of standard units are available for weighing
compensation. Thus the common mode voltage up to 2kg, 20kg and 100kg. Although physically
changes the with temperature, and hence the different, the cells are 100kg similar in method of
arrangement above should be used. This operates by operation and construction.
referencing the positive bridge supply to the negative
supply, thus varying the common mode but not the
overall bridge supply.
Figure 9 Dimensions
Minimisation of noise
1. Inherent white/flicker noise in amplifier. 2kg load cell
118
To keep this to a minimum use high quality (metal film) 19 65
resistors and protect the amplifier from excessively
high temperatures. The inherent noise level may be
26.25
22.5
further reduced from its already low value by fitting C1
and C2 to reduce the operating bandwidth. MOUNTING HOLES M6 X 1.0
4 POSITIONS 81.5
48
then it is generally caused by the location of the supply
transformer, and/or the wiring. Relocate the supply
transformer, screen and input leads to the amplifier, 110.5 19
41.5
38.5
19
4. Electromagnetic interference
19
Underload screw
This may be picked up by input leads, output leads, 1 position M6 x 1.0 31
0.5mm gap for
supply leads or direct into the circuit. Minimisation overload protection
involves a combination of screening, decoupling and Limit of loading area
31 Mounting holes 6 positions
reducing operating bandwidth. Screening. The shield M6 x 1.0
should be connected to only one earth potential at the
receiving monitoring equipment end. Try not to earth
32
25
1.50
leads are more than 10m long and fit C6 if remote
sense is longer than 10m. Reduce the operating 40
Overload stop 2 places
bandwidth by fitting C1 and C2.
8 Mounting Holes
5/16" - 18 Unc x 15 Deep or
M8 - 1.25 x 15 Deep Ø 11.5
Load cells
Introduction
60 2 places
76.2
82
5
232-5957
When used in weighing scales a platform up to the recommended. An overload in excess of 150% (200%
maximum size given in the specification can be used for 100kg unit) will cause permanent damage to the
without loss of performance. cell.
Electrical connections An underload is simply a load which raises rather than
The cells use a six wire full bridge system for the most depresses the load face. The RS cells are capable of
accurate results. The lead to the cell is screened and measuring these types of load.
the cores are colour coded as shown in Figure 10. On the 2kg load cell both over and under load stops
The RS strain gauge amplifier RS stock no. 846-171 can are built into the device and therefore the cell will be
be used with these load cells. Use the circuit shown in protected if correctly mounted on a flat and rigid base.
Figure 7, connecting bridge supply to excitation and With the 20kg cell the base of the beam is machined
compensation to sense (Figure 10). In this circuit a five so that it will deflect and touch any flat base used at
wire system is employed so that the -ve sense wire rated load. Using a flat rigid base will, therefore,
shown in Figure 10 is not used and should be automatically provide overload protection.
connected to the -ve supply. An extra M6 x 1 tapped hole is provided in the base
Other amplifiers can be used but to achieve good for an underload stop. A mechanical stop should be
results an accurate low drift amplifier is required. provided with a no load clearance of 0.5mm so that the
load face of the load cell can be only be raised by
0.5mm which is equivalent to the full rated load of the
Figure 10 Internal wiring single point types
unit.
+ Sense (Blue)
+ Excitation (Green)
Single point - oil damped
6
232-5957
Responce
3 Vibration
Times
Seconds It is sometimes assumed that because the Load Cell is
2
damped it is impervious to external vibration. This is
1
not so. It is damped against its own natural vibration
Tare Weight (kg)
when loaded. If however, the Load Cell is oscillated by
0 external forces, such as adjacent vibrating machinery,
1 2 3 4 5 it may provide output signals corresponding to these
Results from 130gm weight dropped 100mm
forces because heavy structures tend to oscillate at
around 0.1 to 10Hz. It is impossible to damp the Load
As can be seen tare weight increases settling time and
Cell adequately to eliminate these effects and maintain
should be kept at a minimum.
coherent performance. The design aim must therefore
be to attach the unit to a firm flat, level base and to
Figure 12 Typical response times (damped) ensure that this base is free from vibration. The main
sources of vibration are likely to be rotating
60
Capacity 7kg machinery on the weigh structure, vibration from the
floor etc. Each of these must be nullified, preferably
50
by physical separation, but if that is not possible, by
40 shock absorbers, anti-vibration mounts or similar
Responce
Times
devices.
30
miliseconds
20
Applying the load
10
Tare Weight (kg)
The load must be applied via the bearing surface
0 which is uppermost on the load applicator (Figure 14).
0 2 4 6 8
Both holes must be used, evenly torqued to 7 Nm
(51b/fft) so that the load is evenly distributed.
Results from 130gm weight dropped 100mm
It is usual to use a flat bar or other load spreading
member between the applicator and the weigh
With the damped load cell it can be that the settling platform, table or live superstructure. The mating
time is drastically reduced from more than 1 second to section and the substructure, must be rigid, otherwise
less than 100ms. (Figure 12 and 13). the latter will oscillate and superimpose on the Load
Cell output, depending on the frequency and
Figure 13 Capacity 7Kg amplitude generated. The supporting member must
be flat. The load should be transported on to the weigh
Signal
Level platform in such a way that it creates the minimum
disturbance. If the load traverses across the platform,
it should, if possible, avoid knocking the platform
Results from 130gm weight
dropped 100mm
edge (i.e. no step). If the load is lowered on to the
platform it should be controlled placement, not a
dropped load, if possible. For optimum performance
the line of action of the applied load should act as near
as possible to the centre line of the Load Cell - in both
horizontal planes-to minimise eccentricity effects.
0 10 20 Time
(milliseconds)
Effect of temperature
Typical Profile
Variations in temperature will affect the viscosity of the
damping fluid and consequently the settling time of the
Mounting Load Cell. The standard unit is damped for ambient
The precision obtainable from the unit can only be working temperatures (around 20°C).
realised by careful attention to the mechanical It is recommended that the Load Cell temperature be
mounting. It will be appreciated that if the full scale within 10°C of the specified working temperature if
deflection of the Cell is 0.4mm and the scale is divided the settling time is to be within 50 milliseconds.
into 4000 divisions, one division on the scale is the
result of a deflection of 0.0001mm. So, any, force, from
whatever source, which brings about such a deflection
will introduce an error into the system.
It is for this reason that the baseplate is solid and has a
machined surface for mounting. Ensure, therefore that
the mounting support is correspondingly flat and rigid.
Holding down bolts must be equally torqued to 35-40
Nm (or 25-30 Ib ft). Also it is important that the Load
7
232-5957
Unstable reading Tension/compression load cells
If the repeated application of a load gives a steady but These general purpose load cells are for force
inconsistent reading, or the readings steady at around, measurement with loads up to 500kg. Mechanical
but not quite, zero between weighings, then the Load connections are by M12 x 1.75 threads in the body of
Cell is probably reflecting some form of mechanical the device (1/2 x 20 UNF on 500kg unit) and electrical
interference or "stiction". Check that the platter is not connections are via a 3m 5-core screened cable.
fouling and that dirt etc. has not accumulated These cells can be used for weighing but are ideally
somewhere in the scale mechanism. Remove the suited for the measurement of tensile, or compressive
source of interference and normal weighing should forces by using the cell to replace the structural
result. Inspect the Load Cell load applicator: this member under investigation. The 500kg unit is
should not be binding on the cover of the unit. Do not constructed using stainless steel & the 250kg utilises
unbolt the base unless absolutely necessary. If this has aluminium.
to be done, be sure to follow the bolting down
procedure described earlier. Other applications include, e.g. determining the
power output of a motor be replacing the mounting
with the cell and measuring the torque reaction
Outline dimensions produced.
Figure 14
220
200
Figure 17 250kg load cell dimensions
40
25
Sealing/Transportation Screw
58.5 Mounting Holes (2 Positions)
76mm
3.0in
Load Applicator
50 Dia.
21
Seal
2
Cable Length
115 (-0 +2)
3 Metres
Centre of Load
30
(1.03)
Flexible
_ Sense Yellow
(Brown)
BALANCED TEMPERATURE
COMPENSATION
_ Output White
8
232-5957
Shear beam load cells
Figure 23 Wiring schematic diagram
These shear beam load cells are available in stainless
steel (500kg & 1000kg) construction or in aluminium
+ve Input - (Red)
(500kg only) construction. All are low profile units
providing a high level of environmental protection +ve Sense - (Blue)
whilst featuring 6 wire output for temperature
compensation. Both types of unit are of rugged +ve Output - (Green)
construction but the stainless steel products are
hermetically sealed to enable the cells to function in
harsh environments whilst maintaining its operating _ve Input - (Black)
specifications. _ve Sense - (Brown)
Typical applications for these products would be in _ve Output - (White)
low profile platforms, pallet truck weighers, tanks and
silos etc.
Balanced temperature compensation
thread
M12 x 1.75
Ø 32
23
50.5
_ve Input
20
20
- (Black)
_ve Sense - (Grey)
_ve Output - (Red)
46 20.5
Balanced temperature compensation
130.00
+ve Output - (White)
- (Grey)
_ve Output - (Red)
Balanced temperature compensation
9
232-5957
The information provided in RS technical literature is believed to be accurate and reliable; however, RS Components assumes no responsibility for inaccuracies
or omissions, or for the use of this information, and all use of such information shall be entirely at the user’s own risk.
No responsibility is assumed by RS Components for any infringements of patents or other rights of third parties which may result from its use.
Specifications shown in RS Components technical literature are subject to change without notice.
RS Components, PO Box 99, Corby, Northants, NN17 9RS Telephone: 01536 201234
An Electrocomponents Company © RS Components 1997