Professional Documents
Culture Documents
OBJECTIVE
1. Understanding I/P and control valve principle and operation.
2. Compare signal trend for ATO & ATC control valve.
PROCEDURES:
Before starting the following experimental procedures, perform and review the “Start-up Check List
and Preliminary Exercises” and “Planning the Experiments” detailed in the earlier section for the
proper setup. Proceed with the experiment as follows:
1. Familiarize with the physical specifications and measuring principles of the Current-to
Pneumatic (I/P) converter for calibration check. Locate its “AIR SUPPLY” and “OUTPUT” ports.
Open its cover and locate its zero “-> 0 <-” and span “<-->” terminals. Locate its “+” and “-”
terminal block which has cable extensions labeled with “mA+” and “mA-”. Obtain the
specification by referring to the tag at the converter cover and record it into TABLE A.
Familiarize with the calibration terms used.
2. The I/P converter needs to be set up with the appropriate input and output calibration or test
equipment and proper connections.
3. Instrument Air Supply (IAS) for I/P converter
The I/P converter needs instrument air supply (IAS) to function. Connect the instrument air
supply using the flexible tubing to its supply “SUP” port of the I/P with 1.4 bar air supply (source)
from Model ICS911. Note that the same IAS shall be supplied for both I/P converter and air filter
regulator of the control valve.
4. Pressure Output (i.e. Pressure Output from I/P converter)
The I/P converter output is standard 3 to 15 psig signal. Decide the pressure output signal (in psi)
for the tested calibration range, to be read from the I/P converter. Two methods are provided
here:
1
Process Instrumentation & Control BTK3274
Recommended current
No %range Use current source
Read current at calibrator (Ametek ASC300)
1 0 4.0 mA
2 25 8.0 mA
3 50 12.0 mA
4 75 16.0 mA
5 100 20.0 mA
9. Calculate % accuracy and % deviation. Identify any zero drift, span error, combined zero drift
and span error. If adjustment is needed to eliminate errors, use the zero “-> 0 <-” and span “<--
>” terminals of the I/P converter.
10. Repeat the above steps 6-9 for RUN II. Record down the readings in TABLE B.
11. The pressure output from the calibrated I/P converter can then be connected to the positioner
of the pneumatic control valve for five-point visual stroke check.
2
Process Instrumentation & Control BTK3274
Note
A: Current injected to the converter using a signal calibrator for 4 to 20 mA
8 mA-4 mA
Example: x 100%=25%
20 mA-4 mA
B: Read from the signal calibrator for current injected into I/P converter
C: Read from the signal calibrator or test gauge and the measured output signal is compared to the theoretical output
signal
D: Theoretical output signal. Example for 25% calibration range, theoretical output of the 3-15 psig signal is
[(15-3)x0.25]+3 psig = 6 psig
G: Deviation provided by the manufacturer’s calibration certificate, if any.
Note
A: Current injected to the converter using a signal calibrator for 4 to 20 mA
8 mA-4 mA
Example: x 100%=25%
20 mA-4 mA
B: Read from the signal calibrator for current injected into I/P converter
C: Read from the signal calibrator or test gauge and the measured output signal is compared to the theoretical output
signal
D: Theoretical output signal. Example for 25% calibration range, theoretical output of the 3-15 psig signal is
[(15-3)x0.25]+3 psig = 6 psig
G: Deviation provided by the manufacturer’s calibration certificate, if any.
3
Process Instrumentation & Control BTK3274
Note
A: Current injected to the converter using a signal calibrator for 4 to 20 mA
8 mA-4 mA
Example: x 100%=25%
20 mA-4 mA
B: Read from the signal calibrator for current injected into I/P converter
C: Visual check at the control valve indicator
Note
A: Current injected to the converter using a signal calibrator for 4 to 20 mA
8 mA-4 mA
Example: x 100%=25%
20 mA-4 mA
B: Read from the signal calibrator for current injected into I/P converter
C: Visual check at the control valve indicator