Professional Documents
Culture Documents
1935 2015
SIGMA
MERCK
CHEMICAL
COMPANY
1954
2010
MILLIPORE
1668
MERCK
MILLIPORE
ANGEL PHARMACY founded
in Darmstadt, Germany by Friedrich Jacob Merck
2
What is “Troubleshooting”?
→ HPLC peak resolution and quantitative results are effected by solvents with lower quality
Compounds producing artefacts: Additives, Peroxides, Phtalates
NOTE: Use HPLC quality only! (Do not use old or unfiltered buffers!)
Solvent/additive quality
purest solvent and reagent (acids, bases, buffers) quality necessary
(e.g. LiChrosolv® UHPLC-MS, LiChrosolv® hypergrade for LC-MS, EMSURE®, Suprapur® and
Ultrapur solvents and reagents)
Reagent impurities and reagent contamination during handling might
increase background noise and hence decrease sensitivity (higher LOD)
increase complexity
cause “ghost peaks”
baseline drifts
cause signal suppression (MS)
cause adduct formation (MS)
Organic solvents for HPLC, e.g. LiChrosolv® acetonitrile and methanol, are available in four
qualities:
isocratic grade for liquid chromatography (for LC-UV isocratic runs)
gradient grade for liquid chromatography (for LC-UV gradient runs)
hypergrade for LC-MS (for LC-MS gradient runs)
LiChrosolv® UHPLC-MS (for UHPLC-MS gradient runs)
4.2E4 399.5
isocratic grade
4.2E4 4 gradient grade direct injection of two
443.5
broad peak range and broad peak range but different ACN qualities
high signal intensity low signal intensity
(„grades“) into the MS
MS intensity via syringe pump (flow
3
MS intensity
355.4
487.5
injection analysis, flow
2
2 12 mL/h)
531.5
413.4
lower signal intensity
→ better quality
1 443.5
1
305.3259.2 487.5
135.1
135.1
575.5 531.5
2172.161.2
217.1
0
0
0 200 400 600 800 1000 1200 1400 1600 1800 0 200 400 600 800 1000 1200 1400 1600 1800
m/z m/z
0 min 40 min
LC-MS requires
improved
procedures,
compared to HPLC
with UV detectors
HPLC separation of
methyl, ethyl and propyl
paraben
Spectrum Methyl
Ionization intensity is
paraben low, due to ion
suppression →
base peak intensity
< 4 x 102
Background mass signals
appear in the spectra,
making interpretation
more difficult
Solvents for Headspace GC and LC-MS
13
Preparation of Mobile phase
Day 1 Day 2
Bottle caps/adapters
professional equipment only, directly mounted to original bottle
(see image)
avoid decanting
no homemade solvent tubing solutions
→ solvent purity maintained, low background noise
Isocratic separations
isocratic grade solvent quality
Gradient elution
gradient grade solvent quality
no contaminant accumulation on column
(“ghost peaks”, see figure)
Contaminations in solvents
decrease in sensitivity
Might accumulate on stationary phase
Column Chromolith FastGradient RP-18 endcapped 50-2
elute as distinct peaks („ghost peaks“) in Mobile phase A: Acetonitrile + 0.1% formic acid, B:
water/acetonitrile 95/5 (v/v) + 0.1% formic acid
gradient runs Gradient 0 min 100% B, 3 min 5% B, 5 min 5% B
Flow rate 0.4 mL/min
Detection Pos. ESI-MS, TICs
Sample Plasticizers (in eluent B)
Retention Recovery
• Changing retention times*
• Poor sample recovery
• Decreasing retention times*
• Increasing retention times* Leaks
• Leak at column or fittings
Equilibration
• Leak at detector
• Slow column equilibration time
• Leak at injection valve
• Varying retention times
• Leak at pump
Peaks Selectivity
• Broad peaks*
• Differences in selectivity*
• Ghost peaks*
• Negative peaks* Baseline
• Peak doubling* • Disturbance at void time*
• Peak fronting* • Drifting baseline*
• Peak tailing* • Noise*
• Spikes Pressure
• No peaks • Decreasing pressure
• Peaks with shoulders, split • Fluctuating pressure
peaks* • High back pressure*
18
Retention Time
01 • Cause :
Change in mobile phase composition. (Small changes
can lead to large changes in retention times.)
• Solution :
Check make-up of mobile phase. If mobile phase is
machine mixed using proportioning values, hand mix
and supply from one reservoir.
02 • Cause :
Sample solvent incompatible with mobile phase.
• Solution :
Adjust solvent. Whenever possible, inject samples in
mobile phase.
Baseline Drift
• Cause :
Nonhomogeneous mobile phase. (Drift usually to higher
absorbance, rather than cyclic pattern from temperature
fluctuation.)
• Solution :
Use HPLC grade solvents, high purity salts, and additives.
Degas mobile phase before use, sparge with helium during
use.
Baseline Noise
www.SigmaAldrich.com www.MerckMillipore.com
22 Safety in Laboratory
Further Information?
More information
23 Safety in Laboratory
Further Information?
Contact us
Utama Putra
Channel Sales Team for LC & IA
Email : Utama.Putra@merckgroup.com
Phone : 0811 1807 216
Fitriah
Sales Manager Channel Sales for LC & IA
Email : fitriah.fitriah@merckgroup.com
Phone : 0811 1807 019