You are on page 1of 12

PRODUCT DATA

INNOVISION-EXII
(Ceiling Type/ ELIN-T5)
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)

CONFIGURATION
(1) Generator
(2) Tube
(3) Tube Support
(4) Wall Bucky
(5) Table
(6) Collimator
(7) Detector
(8) Workstation (Option)

SPECIFICATIONS
(1) Generator

HX-1000C(option) HX-630C
kW 80kW (Option) 50kW
kV range 40 - 150kV 40 - 150kV
500mA at 150kV 320mA at 150kV
Power Output Ratings
1000mA at 80kV 500mA at 100kV
sec range 1ms - 10sec 1ms - 10sec
mA range 10 - 1000mA 10 - 630mA
mAs range 0.1 - 1000mAs 0.1 - 630mAs
Anatomical Programing 1024 1024
Automatic Exposure Control
Optional Optional
(2 Chambers)
Bucky Signal Supply Standard Standard
380Vac, 50/60Hz 380Vac, 50/60Hz
Line Voltage
Three Phase Three Phase
Digital Interface Available Available
E7884X
Tube E7869X
RAD-14
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)
(2) Tube

E7869X E7884X (Option) Rad-14 (Option)


Maximum kVp rating 150 kVp 150kV 150 kVp
0.6mm-40kW 0.6mm-22kW 0.6mm-32kW
kW rating
1.2mm-100kW 1.2mm-54kW 1.2mm-77kW
Focal spot size 0.6 1.2mm 0.6 X 1.2 mm 0.6 1.2mm
Anode heat storage
600KHU 300kHU 300KHU
capacity
Target angle 12° 12° 12°
Weight 24 kg 16 kg 16 kg
3200 rpm -60Hz / 3,450 rpm- 60Hz /
Anode Rotation 3,200 rpm- 60Hz
9700rpm -180Hz 10,000 rpm – 180Hz

(3) Tube Support

ELIN-T5
Balance system Spring balance
Vertical Travel 1500mm
Longitudinal Travel Max. 3,668mm
Transverse Max. 2,077mm
Tube Rotation +/-180 deg
Tube Swivel +/-180 deg
Driven Type Motor & Manual Driven
THD* Tube angle & SID Display, LCD type

(4) Wall Stand

DS-100 DS-90 (Option)


Type Tilting Detector Stand Bucky Stand
Balance system Weight Balance Weight Balance
Cassette Tray type (Basic) / Cassette Tray type (Basic) /
Bucky type
Fixed Detector type (Option) Fixed Detector type (Option)
Vertical Travel 1,300mm 1,450mm
Height 2,034mm 2,026mm
Drive type
Manual & Motor Driven Manual
(Vertical)
Vertical lock Electro magnetic Electro magnetic
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)
(5) Table

TE-100 TE-90(Option)
Type 6 way Table 6 way Table
Dimension (H) Max. 816mm
Dimension (L) 2,300mm 2,000mm
Dimension (W) 817mm 800mm
Longitudinal Travel 820mm 630mm
Transverse Travel 250mm 300mm
Vertical Travel 310mm 290mm (566 ~ 856mm)
Bucky Movement Range 640mm 479mm
Vertical Movement Type Motor Driven Actuator type
Cassette Tray type (Basic) Cassette Tray type (Basic)
Bucky Type
Fixed Detector type (Option) Fixed Detector type (Option)
Acryl (Basic) Acryl (Basic) /
Tabletop Material
Melamine (Option)

(6) Collimator

R225ACS BL-50 (Option) BL-60(Option)


24 V AC/DC 2A
Electric Power 24VAC 7A, 150W 24VAC, 100W
50/60Hz
Auto Manual , Manual,
Type
Rectangular Rectangular Rectangular
kV Range 150kVp 150kVp
SID at 100cm less SID at 100cm less SID at 100cm less
Max Field X-ray Beam
than 47 x 47cm than 47 x 47cm than 47 × 47cm

SID at 100cm less SID at 100cm less SID at 100cm less


Min Field X-ray Beam
than 0 x 0cm than 5 x 5cm than 0 × 0cm
Lead Equivalent
3 mm pb
of Leaves
Filtration: Inherent
2.0mmAI. eq. 1.2mmAI. eq. 1.0mmAl
filtration
AC 24V, 1A /
Power Requirements
DC24V 1A
190 X 203 X 180
Dimensions 244 x 282 x 216mm 182 x 197 x 164mm
mm
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)
(7)Detector

FXRD-1717SA/SB
Detector type Indirect
A : CsI(Cesium Iodide)
X-ray Conversion
B : Gadox (Gadolinium Oxysulfide)
Wired/Wireless Wired
Panel size 17 x 17 inch
1717SA : 3,052 x 3,052 pixels
total Pixel Number
1717SB : 3,072 x 3,072 pixels
Image format 14bit
Spatial Resolution 4.0lp/mm
Pixel Pitch 140 micron
ELUI(DXIE) : 2.4 sec
Acq. time*
ELUI(XIPL) : 4.2 sec
AED Yes
Dimension 470 x 470 x 35mm
Weight 11kg

FXRD-1717NA/NB
Detector type Indirect
A : CsI(Cesium Iodide)
X-ray Conversion
B : Gadox
Wired/Wireless Wired
Panel size 17 x 17 inch
3,048(W) 3,048(H) pixel (FXRD-1717NA)
total Pixel Number
3,072(W) 3,072(H) pixel (FXRD-1717NB
Image format 16bit
Spatial Resolution 4.0lp/mm
Pixel Pitch 140 micron
ELUI(DXIE) : 2.7 sec
Acq. time*
ELUI(XIPL) : 4.5 sec
AED Yes
Dimension 460 x 460 x 15.5mm
Weight 4.2kg
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)

FXRD-1717VA/VB
Detector type Indirect
A : CsI(Cesium Iodide)
X-ray Conversion
B : Gadox (Gadolinium Oxysulfide)
Wired/Wireless Fixed
Panel size 17 x 17 inch
total Pixel Number 3,072 x 3,072
Image format 16bit
Spatial Resolution 4.0lp/mm
Pixel Pitch 140 micron
ELUI(DXIE) : 2.4 sec
Acq. time*
ELUI(XIPL) : 4.2 sec
AED Yes
Dimension 460 x 460 x 15mm
Weight 4.6kg

FXRD-1417SA/SB
Detector type Indirect
A : CsI(Cesium Iodide)
X-ray Conversion
B : Gadox (Gadolinium Oxysulfide)
Wired/Wireless Wired
Panel size 14 x 17 inch
total Pixel Number 2,560 x 3,072
Image format 14bit
Spatial Resolution 4.0lp/mm
Pixel Pitch 140 micron
ELUI(DXIE) : 2.4 sec
Acq. time*
ELUI(XIPL) : 4.2 sec
AED Yes
Dimension 444 x 460 x 15mm
Weight 3.6kg
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)

FXRD-1417WA/WB
Detector type Indirect
A : CsI(Cesium Iodide)
X-ray Conversion
B : Gadox (Gadolinium Oxysulfide)
Wired/Wireless Wireless
Panel size 14 x 17 inch
total Pixel Number 2,560 x 3,072mm
Image format 14bit
Spatial Resolution 4.0lp/mm
Pixel Pitch 140 micron
Wireless type Wired type
Acq. time* ELUI(DXIE) : 5.7 sec ELUI(DXIE) : 3.2 sec
ELUI(XIPL) : 7.5 sec ELUI(XIPL) : 5 sec
AED Yes
Dimension 384 x 460 x 15mm
Weight 3.4kg

FXRD-1717NAW/NBW
Detector type Indirect
A : CsI(Cesium Iodide)
X-ray Conversion
B : Gadox (Gadolinium Oxysulfide)
Wired/Wireless Wireless
Panel size 17 x 17 inch
total Pixel Number 3,060 x 3,060
Image format 16bit
Spatial Resolution 4.0lp/mm
Pixel Pitch 140 micron
Wireless type Wired type
Acq. time* ELUI(DXIE) : 4.2 sec ELUI(DXIE) : 2.7 sec
ELUI(XIPL) : 6 sec ELUI(XIPL) : 4.5 sec
AED Yes
Dimension 460 x 460 x 15.5mm
Weight 4.5kg
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)
(8) Workstation (Option)

Workstation (Option)
CPU i5 8500 3.0GHz
M/B GIGABYTE GA-H310M S2P
Memory DDR4 4G PC4-21300
Workstation HDD 1 SEGATE 1TB 7200RPM
HDD 2 Samsung 850 SSD 120G
POWER FSP ATX500-50HYN
OS WIN 10 PRO 64BIT
Monitor 21.5inch
Type LED
Brightness 250cd/m2
Pixel size 0.284 X 0.284
Monitor
Viewing angle 178(left/right) 178(upper/bottom)
Resolution 1920 X 1080
Contrast 3000:1
Screen ratio 16:09
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)

ROOM CONSIDERATIONS
Room Layout
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)
Elin-T5

DS-100
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)
TE-100

Generator
PRODUCT
DATA INNOVISION-EXII
(Ceiling type : ELIN-T5)

Công ty TNHH VIKO HEALTHCARE


Tầng 4, Tòa nhà HMS, 24A Phan Đang Lưu, Phường 6,
Quận Bình Thạnh, Hồ Chí Minh
Tel: 028-6271-9155
www.vikohealthcare.com
Remarks
Every value in this Product Data Sheet is a standard value, and it may vary a little from the actual at each site
The appearances and specifications are subject to change for reasons of improvement without notice.
Certain configurations may not be available pending regulatory clearance. Contact your DK Medical system representative
for information on specific configurations.

ISO 9001 : 2008


ISO 13482 : 2012
DK Medical System Co., Ltd
International Business Team
DK Bldg, 18, Baumoe-ro 7-gil, Seocho-gu, Seoul, 137-900, Korea
Certificate of GMP
Factory
52, Chupal-Ri, Paengsung-up, Pyeongtaek-si, Gyeonggi-do, 451-805,
Korea
Tel : 82-31-658-7231~7, FAX : 82-31-658-7239

http://www.dk.co.kr

You might also like