SHIMADZU CORPORATION 05/13/2023

SHIMADZU GE OEC Siemens Siemens Philips Philips Remarks

Model PLENO Flexiview8800 SIREMOBIL Iso-C SIREMOBIL BV Endura BV Libra


Compact

C-arm

Orbital rotation 120deg 135deg 190deg 125deg 115deg 115deg


Pivot rotation 300/120deg ±180deg ±190deg ±190deg ±180deg ±205deg
Panning motion ±12.5deg ±10deg ±12deg ±12.5deg ±10deg ±10deg
Free space I.I to tube 700mm 798mm 780mm 670mm 780mm 780mm
C-arm depth 660mm 660mm 660mm 660mm 610mm 610mm
S.I.D 900mm 990mm 1000mm 900mm 995m 995m
C-arm vertical movement 450mm 460mm(electric) 400mm(electric) 430mm(electric) 500mm(electric) 500mm(electric)
Horizontal travel 200mm 200mm 200mm 195mm 200mm 200mm
Locking system Electromagnetic lock Lever lock Electromagnetic lock Lever lock Lever lock Lever lock Fully balanced for smooth C-arm positioning(4 Electromagnetic
locking system).
Locking lever Not necessary Necessary Necessary Necessary Necessary Necessary
Locking release C-arm side Possible(op.) N.A N.A N.A N.A unknown
Cabling Inside C-arm Outside C-arm Outside C-arm Outside C-arm Outside C-arm Outside C-arm
Weight 269kg 275kg 250kg 300kg unknown
Generator

Inverter 50kHz 20kHz 20kHz 20kHz unknown unknown The x-ray generator incorporates a high frequency inverter that
generates a highly efficient output with reduced ripple
Norminal max.power 2kW 2.2kW 1.4kW 1.4kW 3.15kW 3.15kW
Fluoroscopy voltage 40~110kV 40-110kV 40-110kV 40-110kV 40-110kV 40-110kV
Continuos fluoroscopy max.110kV max.110kV 110kV 5mA 110kV 5mA 110kV 3mA 110kV 3mA
max.5mA max.8mA 70kV(8mA) 70kV(8mA)
Fluoroscopic mA 9mA 6mA 8.9mA 8.9mA max. 7.2mA max. 7.2mA
Fluoroscopyic timer 5minutes 5minutes unknown unknown unknown unknown
Pulsed fluoroscopy max.9mA 16mA 0.2-12.2mA 0.2-12.2mA 1 or 2fps. ?
Pulse rate 0.5-15fps.(EIA/ 1,2,4,8fps Variable Variable unknown unknown Three fluoroscopic mode, Normal / Low Dose / High Quality with
CCIR) pulsed fluoroscopy are available to provides the flexibility to handle
Radiography rating 60kV/200mAs unknown 110kV/100mAs 110kV/100mAs unknown unknown most situations.
110kV/100mAs
Radiography voltage 40-110kV 50-110kV 40-110kV 40-110kV 40-110kV 40-110kV
Radiographic current 20mA 20mA 6.9-20mA 6.9-20mA 20mA 20mA
Radiographic mAs 0.5-200mAs Up to 80mAs Up to 150mAs Up to 150mAs Up to 80mAs Up to 80mAs
X-ray tube
Anode Stationary Stationary Stationary Stationary Stationary Stationary
X-ray focal spot size(mm) 0.6 0.6/1.4 0.6 0.6 0.6/1.4mm 0.6
Thermal capacity(kHu) 100 74 50 50 51 51
Collimator position memory Iris Iris Iris Iris Iris Iris
Iris collimation+semitran collimation+compens collimation+compens collimation+asymmet collimation+asymmet
collimation+compens s-parent parallel ation ation ory shutter ory shutter
ation shutter (Parallel+rotation) (Parallel+rotation)
(Parallel+rotation) filter filter
filter
Beam hardening filter 3 type 1 type unknown 1 type 1type 1type BH filter selection is available for Shimadzu Opescope
Virtual collimation(from monitor side) Possible Possible(op.) Possible ? Possible ???/
Virtual collimation(from C-arm unit) Possible N.A N.A N.A N.A N.A
I.I & CCD camera CCD CCD CCD CCD CCD CCD

WHA-200_Comparison table
SHIMADZU CORPORATION 05/13/2023
I.I filed 9/6" 9/7"/4.5" 9/6inch 9/6inch, 7/4inch 9/7/5" 9/7/5"
6" 6"
Monitor size 18" color LCD 16" CRT 17"CRT 17"CRT 17"CRT 17 The conventional CRT monitor is replaced by flat 18inch LCD
monitors that have wide field of view. The LCD monitos save space
while displaying clear, easy to view images. Also LCD monitors have
vertical and tilt adjustment functions to permit positioning adjustment
of monitor to the operator's line of sight.

X-ray TV image rotation ±170 360 ±170 ±170 360 360


Last image hold Possible Possible Possible Possible Possible Possible
Operation panel(C-arm unit side) Color LCD touch Touch screen key Sheet key EL Sheet key
panel & sheet key board
Operation panel(monitor side) Mouse+Key board Sheet key Sheet key Sheet key
(touch panel:option)
Image acquisition
Acquisition mode Continuous Continuous Continuous Continuous Continuous Continuous Shimadzu Pleno offers RSM-DSA, Realtime DSA, Super Cine
fluoroscopy, fluoroscopy, fluoroscopy, fluoroscopy, fluoroscopy, fluoroscopy, Display
Pulsed Pulsed Pulsed Pulsed Pulsed Pulsed
fluorosocopy, Spott, fluorosocopy, Spott, fluorosocopy, Spott, fluorosocopy, Spott, fluorosocopy, Spott, fluorosocopy, Spott,
Serial, DSA, Map Serial, DSA, Map Serial, DSA, Map Serial, DSA, Map Serial, DSA, Map Serial, DSA, Map
Digital Image processing
Fluoroscopy 1024x1024, 12bit unknown unknown unknown unknown
max. 30fps.
Digital Radiography 1024x1024, 12bit 1000 x 1000, 8bit 512 x 512, 8bit 798 x 480, 8bit 798 x 480, 8bit
max. 7.5fps. 1/2/4/8fps. 6fps.(max) 5/8/30fps.
1 time exposure DSA Possible N.A N.A N.A N.A
Dynamic IBS Possible Possible N.A Possible Possible
Image memory(HD) 30,000frame (HD) 400 900 900 500/5000/10000(HD) 50/500/1000
Anatomical program Possible N.A N.A N.A N.A N.A Shimadzu Opescope Pleno offers APR (Anatomical programs) to
adjust imaging conditions such as X-ray condition, BH filter selection
and I.I field etc. during surgery

CCD rotation angle memory Possible no N.A N.A Possible Possible


Collimator position memory Possible no N.A N.A N.A N.A
Storage CD-R PC card reader MOD MOD VTR VTR
DICOM(Strage/Print) Option Option Option Option Option Option
Exposure worning lamp Possible N.A N.A N.A Possible Possible
Laser pointer Possible(op.) Possible(op.) Possible(op.) Possible(op.) Possible(op.) Possible(op.)
Power requirement Single Single Single Single Single phase100/240 Single phase100/240
phase100/200V,3kV phase110/230VAC, phase100/200V,20A phase100/200V,20A
A  15/8A