RF Puncture Generator†
Impedance (ohms)
Use pulse mode or constant mode, Pulse Mode: 300ms Pulsed RF delivery per second
and customize settings to your
preference. Improved cutting ability
Constant Mode: Continuous RF delivery
enables shorter RF activation time.
Intelligent Interface
Adjust settings in standby state. Automatic recognition of paired devices.
RF Puncture RF Ablation
To create a small opening in To create a lesion to destroy
Objective Objective
tissue electrically conductive tissue
Low power (5-25 watts) High power (30-50 watts)
Occurs under these Occurs under these
Short duration (1-3 seconds) Long duration (60-90 seconds)
conditions conditions
High voltage (270-400V) Low voltage (35-50V)
Impedance range 2000-6000 Ω Impedance range 150-300 Ω 1 STANDBY 2 READY 3 RF ON
Minimal collateral damage to surrounding tissue Thermal destruction of surrounding tissue Connect grounding pad and connector With all components connected, select Once RF ON button is pressed, generator
cable. Generator automatically desired Time and Cut settings. Using enters “RF ON” state. Elapsed time is
recognizes devices and makes available footswitch or RF ON button, initiate RF. displayed during RF delivery. Screen will
only appropriate modes. maintain until RF delivery concluded.
WARNING: The RFP-100A RF Puncture Generator is designed and intended for use with devices designed
by Baylis Medical Company
0.035” 0.024”
The NRG® Transseptal When the conventional The PowerWire® RF The Nykanen RF Wire The VersaCross® RF
Needle is uniquely solution may not be Guidewire is used to cross is designed to create a Transseptal Solution
designed to assist the optimal, use SupraCross® lesions in occluded blood controlled puncture in offers all-in-one versatility
physician in gaining RF Solution to gain vessels that are difficult tissue. for transseptal and
access to the left atrium. alternate access into the to cross with a standard beyond in a single device.
left atrium. guidewire.*