Professional Documents
Culture Documents
57 - Hill Smith - CO2
57 - Hill Smith - CO2
To cite this article: R. Jedd Hill & Philip A. Smith (2015) Exposure Assessment for
Carbon Dioxide Gas: Full Shift Average and Short-Term Measurement Approaches,
Journal of Occupational and Environmental Hygiene, 12:12, 819-828, DOI:
10.1080/15459624.2015.1053894
Article views: 43
centrations may elicit toxicity, even with oxygen concentrations Color versions of one or more of the figures in the article can be
that are not considered dangerous per se. Full-shift sampling found online at www.tandfonline.com/uoeh.
approaches to measure 8-hr time weighted average (TWA)
CO2 exposures are used in many facilities where CO2 gas
may be present. The need to assess rapidly fluctuating CO2
levels that may approach immediately dangerous to life or
health (IDLH) conditions should also be a concern, and sev- INTRODUCTION
eral methods for doing so using fast responding measurement
tools are discussed in this paper. Colorimetric detector tubes, Properties and Hazards of Carbon Dioxide (CO2 )
a non-dispersive infrared (NDIR) detector, and a portable At typical atmospheric temperature and pressure carbon
Fourier transform infrared (FTIR) spectroscopy instrument
were evaluated in a laboratory environment using a flow-
dioxide (CO2 ) is a gas, but it may be liquefied with increased
through standard generation system and were found to provide pressure. Its specific gravity is 1.52 relative to air,(1) and CO2
suitable accuracy and precision for assessing rapid fluctu- gas readily migrates to low elevations when not well-mixed.
ations in CO2 concentration, with a possible effect related It is colorless, odorless, and non-flammable, and solid CO2
to humidity noted only for the detector tubes. These tools (dry ice) sublimes directly to gas. Relatively pure CO2 is
were used in the field to select locations and times for grab
sampling and personal full-shift sampling, which provided
inexpensive; it is commonly transported in a compact liquid
laboratory analysis data to confirm IDLH conditions and 8- volume, and dry ice may be created from the liquid at a point
hr TWA exposure information. Fluctuating CO2 exposures are of use.
exemplified through field work results from several workplaces. Liquid CO2 and dry ice are widely used where rapid cooling
In a brewery, brief CO2 exposures above the IDLH value of materials is desired. For example, it is important to rapidly
occurred when large volumes of CO2 -containing liquid were
released for disposal, but 8-hr TWA exposures were not found
freeze products during large scale preparation of frozen food
to exceed the permissible level. In a frozen food production to eliminate conditions conducive to microbial growth (see
facility nearly constant exposure to CO2 concentrations above Figure 1). Also, sublimation of dry ice directly to gas allows
the permissible 8-hr TWA value were seen, as well as brief ex- for rapid cooling of product packed in boxes, without the
posures above the IDLH concentration which were associated potential for water damage to product or packaging.
with specific tasks where liquid CO2 was used. In a poultry
processing facility the use of dry ice to quickly freeze product
Ikeda et al. demonstrated that CO2 is not a simple asphyxi-
produced a nearly constant CO2 concentration that caused ant, as a mixture where CO2 replaced nitrogen in air (with 20%
exposures to approach the permissible 8-hr TWA exposure oxygen) was fatal to dogs within 1 minute.(2) Effects related
value. to CO2 exposure can range from relatively minor problems to
respiratory distress and death. Breathing rate is affected and
Keywords carbon dioxide, Fourier transform infrared analysis, often increases as airborne CO2 concentrations increase, and
immediately dangerous to life or health (IDLH), nondis-
persive infrared detection, time-weighted average sam-
headache, dizziness, and increased blood pressure and heart
pling, grab sampling rate may also occur.(3) A generalized weakness and feeling of
lethargy(4) is noted with CO2 intoxication, and a depressive
FIGURE 2. Personal TWA sampling for CO2 exposure using a Laboratory Analysis
low-flow pump with outlet hose barb (pump not visible) and a five-
layer aluminized bag.
Whole-air samples were shipped to the OSHA Salt Lake
Technical Center Industrial Hygiene Chemistry Division lab-
oratory, where a model 6890 GC instrument (Agilent Tech-
Another method for nearly instantaneous determination of nologies, Wilmington, DE) with a pulsed discharge detector
airborne CO2 concentration involves the molecular absorbance (model D-3, Valco Instrument Company, Houston, TX) was
of infrared (IR) energy. Absorbance at a correctly chosen IR used to complete analyses. Bag contents were pumped into
wavelength is mathematically related to the gas phase CO2 a 1.0 mL offline sample loop, and helium carrier gas flow
concentration when Beer-Lambert law conditions are met. A (99.999%, Praxair, Salt Lake City, UT) was aligned to inject
CO2 -specific detector that employs only a narrow wavelength sample loop contents. Injector temperature was 200◦ C, and a
band relevant to this gas (e.g., the asymmetrical stretch band 50:1 split ratio was used. The porous layer open tubular column
for CO2 occurring near 4.3 µm) is termed a non-dispersive (RT-QPLOT, Restek, Bellefonte, PA, 30 m × 0.32 mm i.d.)
IR (NDIR) detector, and several manufacturers offer handheld was maintained at 35◦ C throughout an analysis event, with
NDIR instruments for airborne CO2 measurement. carrier gas linear velocity of 50 cm/s.
Absorbance-based detection is also possible with a more
complicated gas-phase Fourier transform infrared (FTIR) Colorimetric CO2 Detector Tubes
spectroscopy instrument that uses an interferometer to rapidly Quantitation ranges for the CO2 detector tubes used were
scan a wide range of wavelengths. Quantitation of airborne 0.13–6% (2L,Gastec Corporation, Kanagawa, Japan) and
CO2 concentrations is possible with gas-phase FTIR 0.5–20% (2H, Gastec Corporation). Tubes were used to select
spectroscopy, and the simultaneous quantitation of a limited locations and individuals for grab sampling, but as a detector
number of other analytes is also possible with a modern field- tube requires preparation followed by operation of a hand
portable FTIR instrument. The operating principles of field- pump, this approach was used when the NDIR detector used
portable gas-phase FTIR spectroscopy instrumentation were (see below) was not available to provide continuous CO2
recently described by Heikkilä.(12) concentration information.
Detector tubes were evaluated in the laboratory with low
Objective (13.1 ± 1.8%) and high (80.1 ± 1.4%) humidity air at six
This article provides performance information for several CO2 concentrations (0.48–11.4%). Air was drawn through a
quantitative field detection methods based on laboratory work single tube at each concentration/humidity combination us-
completed using dynamically generated CO2 standards in dry ing the recommended hand pump, and manufacturer read-
and humid air. Approaches used to measure both full shift ing/correction instructions were followed. The 2L tubes have
Note: Readings were corrected for atmospheric pressure at the laboratory site following directions provided by the tube manufacturer. The percent differences
between the corrected tube readings and the CO2 concentrations sampled are provided in the third and fifth columns. AMeasured using a single full pump stroke
(100 mL); BMeasured using two full pump strokes (2 × 100 mL); CMeasured using one-half pump stroke (50 mL); DMeasurement not taken;EBeyond measurement
range.
increments are marked on a 2H tube (as is also the case for most the laboratory atmospheric pressure of 653 mm Hg (2.8% was
2L tube markings). Values shown in Table I were obtained by multiplied by 760 mmHg/653 mm Hg) to obtain the corrected
determining the color change front against the printed scale on reading of 3.3% CO2 . The detector tube manufacturer does
a tube to the nearest 1/10th marking interval, and this value was not list water vapor as an interference for either tube, and
then subjected to correction. For example, the reading of 3.3% no attempt was made to explore the apparent difference in
for the 2 L tube (dry air) from the measurement of 3.67% CO2 tube response between dry and humid airborne CO2 standards
was read as 1.4%, but this was doubled to 2.8% as only one- sampled.
half pump stroke was used. This value was then corrected for
NDIR Detector
Figure 4 shows that NDIR readings were highly correlated
with the CO2 concentrations sampled. The NDIR detector
readings obtained in the laboratory deviated by no more than
7.8% from the CO2 concentrations sampled.
FTIR Spectroscopy
Figure 4 shows that the readings obtained through FTIR
spectroscopy were highly correlated with the concentration of
CO2 sampled. The FTIR instrument produced readings in the
laboratory that deviated from the CO2 concentration sampled
by no more than 9.2%.
Examination of exhaled air by FTIR analysis showed CO2
content of about 4%, while the background CO2 concentration
in the room where the measurements were made was 0.06%.
With the FTIR inlet tubing placed only 5 cm away and directly
to the side of each author’s mouth the CO2 concentrations
FIGURE 3. Results from laboratory analyses using the col-
orimetric detector tubes described in the text for varied CO2
measured ranged between 0.10 and 0.11%. When the distance
concentrations created with the flow-through standard generator. between inlet and mouth was increased to 18 cm with the inlet
The dark circles and triangles, respectively, represent the 2L resting on a shoulder at a height just below the mouth (as was
and 2H detector tubes used with dry air standards, while the the case for full shift sampling of workers completed in the
white circles and triangles, respectively, represent the 2L and 2H field) the CO2 concentrations measured were identical to the
detector tubes used with humid air standards.
room air background CO2 conentration.
FIGURE 4. Results from laboratory analyses using the field- Note: Work included general maintenance, and tank emptying and cleanup.
AA single bag was used for both morning and afternoon sampling with
portable FTIR instrument (circles) and the handheld NDIR detector
(triangles) for both dry (darkened symbols) and humid (white pump and bag removed from employee and pump turned off with bag inlet
symbols) CO2 standards created with the flow-through standard valve closed during lunch break. Unsampled time at end of shift included
Downloaded by [152.180.134.66] at 07:45 04 November 2015
generator. duties without potential for elevated CO2 exposure. B1.0 = 0.5% CO2 as 8-hr
TWA; unsampled time counted as zero CO2 exposure. CThis sample included
exposure known to exceed 4% CO2 .
Field CO2 Measurements and Laboratory Analysis
Results for Whole-Air Samples
Brewery
handheld NDIR detector, and the IDLH conditions were con-
Cellars of the brewing facility were located within several
firmed through laboratory analysis of whole-air grab samples
multi-story buildings, and workers performed tasks there, and
(Table III).
also spent substantial time in a lounge area of a separate
Riley and Bromberger-Barnea reported on similar work in
building while awaiting assignments. After newly brewed beer
1979, and noted with regard to draining a tank that “. . .CO2
in a large tank (>105 L) was removed for further processing
which is heavier than air, pours out of the door almost like
it was necessary to empty liquid and solids remaining at the
a liquid. The CO2 concentration may thus be much higher
tank bottom. An access hatch was opened at the lowest level
closer to the floor than at normal breathing height.”(16) This
of a tank, allowing light brown opaque liquid and solids to
matches our observations, as CO2 concentrations just above
flow onto the floor in front of a tank, and then to a floor drain.
liquid being drained typically exceeded the IDLH value, but
Tank ventilation commenced after draining to allow entry for
on some occasions IDLH conditions at worker breathing zone
cleaning.
height were also identified.
During an initial survey in the cellars, CO2 levels near
2% were detected in some areas using the handheld NDIR Frozen Food Production Facility
instrument, and CO2 concentrations throughout the cellars Several kitchen areas were located in rooms adjoining a
were measured to be near or above the 0.5% OSHA PEL value. large open bay area within the frozen food production facility.
Consistent with the results of previous full shift TWA personal Food products were prepared separately in a kitchen, with
sampling completed by safety personnel at the brewery, 8-hr
TWA CO2 exposures did not provide evidence for overexpo-
sure expressed as exposure severity (Table II, 100% of 8-hr
TABLE III. Results for Laboratory Analysis of
TWA PEL = 1.0), presumably due to CO2 concentrations
Whole-Air Grab Samples Collected during Two Brew-
in the lounge area being much lower than the PEL value
ery Tank Opening Procedures
for CO2 .
However, tank draining was shown in several instances to Laboratory
be associated with airborne CO2 concentrations greater than Sample Height Analysis Result (% CO2 )
the IDLH value, and workers assigned to open tanks without
respiratory protection knew from experience to quickly leave Chest Height 8.0
the area in front of an access hatch after opening. Although Floor LevelA 10.7
a worker tended to not linger directly in front of a tank ASample obtained with inlet to the large syringe used to collect whole-air
opening, he or she would stay close enough to spray water samples just above liquid present on the brewery floor after being drained
at the hatch opening and the nearby floor. Airborne CO2 con- from a tank. While this grab sample was collected the CO2 concentration at
centrations which exceeded 4% at breathing zone level were chest height and breathing zone was indicated to be < 4% by measurement
identified in several instances using both detector tubes and the using the handheld NDIR detector.
hand from the bulk material into ball shapes with 5–6 cm
diameter, which were placed onto trays that were wheeled into nearly certain for packagers working a substantial portion of
a walk-in chiller adjacent to the kitchen. The chiller door was a full shift near the pneumatic machinery. After accounting
closed and liquid CO2 was sprayed inside to flash freeze the for sampling and analytical error, an 8-hr TWA overexposure
product. Local exhaust ventilation in the kitchens where liquid to CO2 for which a citation could be issued was narrowly
CO2 was used during mixing was limited to flexible hose avoided for a U.S. Department of Agriculture Food Safety
lines, as seen behind the liquid CO2 applicator in Figure 1. Inspection Service (FSIS) employee who only spent a portion
A duct to the exterior existed from the walk-in chiller, but of the day in the open bay area. Airborne CO2 concentrations
with no fan present to remove CO2 gas following use of the in the separate kitchen areas were mostly below the 8-hr TWA
chiller. PEL value (except when liquid CO2 was in use—see below),
Other products were combined in the open bay area after with workers preparing food in the kitchen areas generally
being cooked in a kitchen. Pneumatically operated machinery experiencing lower 8-hr TWA exposures.
dispensed liquid (e.g., stew liquid) onto solid items (e.g., meat, Two separate circumstances were observed to produce
vegetables, and rice) inside individual serving-sized bowls, and IDLH CO2 concentrations, with the first being the opening
moved the bowls through this process. Bowls containing solid of the walk-in chiller after use. When this occurred, airborne
items were passed under a nozzle for addition of liquid, then a CO2 was measured in the adjoining kitchen area at
TABLE V. Laboratory and Field FTIR Analysis Results for Personal 8-hr TWA Sampling of Cooks (Working in
Kitchens), Packagers (Working Near Pneumatic Machinery), and FSIS Inspector for CO2 Exposure
SampleA 8-hr TWA CO2 B Field FTIR Analysis Laboratory Analysis
Duration (min) Exposure Severity Result (% CO2 ) Result (% CO2 ) Primary Task
454 1.0 0.54 0.54 FSIS inspector
446 1.4C 0.74 0.73 Cook
440 0.7 0.44 0.36 Cook
402 0.9 0.56 0.53 Cook
484D 1.7 0.83 0.82 Packager
425 D 1.4 0.80 0.79 Packager
423 D 1.7 0.97 0.97 Packager
441 D 1.9 1.01 1.01 Packager
432 D 1.9 1.08 1.08 Packager
440 D 1.2 0.65 0.63 Packager
AA single bag was used for both morning and afternoon sampling with pump and bag removed from employee and pump turned off with bag inlet valve closed
during lunch break. B1.0 = 0.5% CO2 as 8-hr TWA; full shift sample, based on laboratory analysis results; unsampled time counted as zero CO2 exposure. CThis
sample included exposure known to exceed 4% CO2 . DUnsampled time at beginning or end of shift likely to include CO2 concentrations >0.5% due to elevated
CO2 concentrations in the open bay area near pneumatic machinery.
concentrations > 10% using the handheld NDIR detector, and shoveling dry ice into boxes in the facility’s icing room. The
at 9.79% by laboratory analysis of a grab sample. Regarding dry ice was kept in a large tote box, with no mechanical
the second IDLH exposure situation, grated drain channels ventilation present.
existed throughout the building and CO2 gas flowed through Dual port sampling pumps were not available for TWA CO2
a drain channel to enter a restroom adjacent to a kitchen area sampling during the first visit, and an exhaust fan was added
due to a small fan that continuously exhausted air from the to the icing room near floor level between the first and second
restroom. Using the handheld NDIR detector, CO2 concentra- site visits. This likely resulted in lower CO2 concentrations
tions > 4% were detected near the floor of the kitchen and being measured in that area and in the evisceration room using
within the drain channel while liquid CO2 was used, and this the handheld NDIR detector during the second visit, and field
caused CO2 concentrations > IDLH to exist in the restroom at FTIR and laboratory analysis results for personal 8-hr TWA
breathing zone height. Table VI provides data for IDLH grab CO2 samples during the second visit are provided in Table
sampling completed at the frozen food production facility. VIII. Laboratory analysis of these TWA samples did not show
As at the brewery, airborne CO2 concentrations > 4% were that the FSIS employees working in the evisceration room
expected beforehand and SCBA respiratory protection was were overexposed to CO2 when sampled during the second
used by the authors as needed to complete assessments in visit.
hazardous locations. In 1988 Jacobs and Smith described elevated CO2 expo-
sures at four different poultry processing facilities.(17) Detector
Poultry Processing Facility tubes were used as a field screening tool, and Tedlar bags were
R
Chickens were killed, cleaned, and butchered at the poul- used for personal TWA sampling. Their data showed that CO2
try processing facility. The lack of local exhaust ventilation was lost from the bags before analysis, as detector tube results
allowed gases and vapors emitted within the plant to travel obtained with air from bags in the field were much higher than
throughout the facility, including locations where three FSIS the laboratory results. When evacuated steel canisters were
inspectors stood while examining chicken carcasses moving used to ship whole-air samples initially collected into bags,
through the evisceration room. The carcasses moved on an measured 8-hr TWA CO2 concentrations for dry ice packaging
overhead conveyor to other areas of the plant for further workers were in better agreement with field detector tube
processing in several dip tanks, and finally were cut as needed results. Documentation for the OSHA sampling and analytical
to produce marketable products which were then packaged for method(10) describes the need to use a five-layer aluminized
shipment. bag to avoid sample losses similar to those described by these
Two visits occurred about two months apart, and CO2 authors.
was initially not a suspected stressor. Several real-time de-
tection tools were used, including the field-portable FTIR Usefulness of CO2 Field-Detection Tools
instrument, to detect or rule out acetic acid and ammonia as Fast-responding real-time measurements identified condi-
potential air contaminants. The spectral library used with the tions at the brewery where IDLH conditions could be rapidly
FTIR instrument contained reference spectra for these ana- attained, while 8-hr TWA sampling for CO2 failed to iden-
lytes, and while no evidence was seen for their presence, ele- tify a hazard (Tables II and III). The 8-hr TWA sampling
vated CO2 concentrations were detected throughout the facility at the frozen food production facility easily detected per-
(Table VII). Upon investigation, an employee was observed sonal 8-hr TWA overexposures to airborne CO2 , however,
fast-responding CO2 detection tools were still needed to deter- sample collected in the restroom of the frozen food production
mine that IDLH conditions were associated with certain tasks, facility, with the FTIR measuring airborne CO2 at 4.96% there
Downloaded by [152.180.134.66] at 07:45 04 November 2015
and the handheld NDIR detector allowed the path taken by the following the use of liquid CO2 in the adjoining kitchen (Table
CO2 gas detected in the restroom to be traced back through VI). Confirmation through laboratory analysis was not possible
the grated floor drain to CO2 use in the adjoining room. The due to a damaged bag that leaked during shipment. The FTIR
handheld NDIR detector also allowed the authors to identify reference spectra used during field analysis did not extend
CO2 gas operation of pneumatic machinery, identifying 8- to CO2 concentrations > 1.0%, but the instrument indicated
hr TWA sampling of workers near the machinery as a high CO2 concentrations outside this range and results were re-
priority. evaluated after creating additional reference spectra. Using
Even though the large gas-phase FTIR instrument rapidly the laboratory-generated CO2 reference spectra after the fact,
produces concentration information, it is not particularly use- a CO2 concentration value of 6.80% was obtained, which was
ful as a fast-responding field detection tool. It requires stable close to the concentration measured in real-time using the
electrical power for operation and thus would mostly be used handheld NDIR detector.
at a fixed location, with whole-air samples brought to the The detection of unexpectedly elevated CO2 concentra-
instrument. Also, this type of instrument is not as readily tions by FTIR spectroscopy in the poultry processing facility
available to most industrial hygienists as detector tubes or demonstrated the ability of the FTIR instrument to detect
a small handheld NDIR detector. However, the usefulness unexpected analytes and to produce quantitative results very
of the FTIR instrument was demonstrated for the whole-air close to those obtained from the subsequent GC analysis of
corresponding whole-air samples in the laboratory. Only a
zero-point calibration of the FTIR instrument was completed
before use each day using either ultra-high purity N2 or helium,
and close agreement between the field FTIR analyses and those
completed on the same whole-air samples in the laboratory
was possible without the need for further field calibration.
A comparison between field FTIR and laboratory analysis
results that fell very close to or within the range of reference
spectra used for field analyses is provided in Figure 5 which
combines data from Tables V, VII, and VIII. The data were
highly correlated, and the slope of the resulting line was near
unity.
CONCLUSIONS