Professional Documents
Culture Documents
Prama
Implant system 4
UTM surface 4
ZirTi surface 5
Prosthetic reccommendations 6
Key to the implant codes 7
Table of colour codes 7
The range 8
Prama 8
Prama RF 10
Surgical Instruments 12
Prama surgical Kit 12
Initial and intermediate drills 14
Optional cylindrical drill 16
Optional conical drill 17
Final cylindrical drills and related stops 18
Final conical drills and related stops 20
Reply: replies for Prama RF implants 22
Countersink drills 24
Drills for distal sectors 26
Bone taps 27
Osteotomes 28
Additional instruments 30
Drills for Prama Shorty implants 36
Drills for Prama RF Shorty implants 38
Prosthetic components 40
Standard healing abutments 40
Prama IN healing abutments 42
Impression and model phase 44
Temporary posts 46
Pre-made straight and angled posts 48
Standard and Prama IN preparable posts 50
Standard castable posts with base in gold alloy, titanium, and cobalt chrome 52
Prama IN castable posts with base in gold alloy or cobalt chrome 54
Fully castable posts 56
Dynamic Abutments posts 58
Prostheses on intermediate abutments 60
Prostheses on PLAIN abutments 62
P.A.D. (Disparallel Screwed Prosthesis) 64
P.A.D. prostheses for “D.P.F.” Technique (Direct Prosthetic Framework) 68
Titanium and cobalt chrome sleeves 70
Conoweld posts 72
ECHO custom-made prosthesis 76
T-Connect 80
Prama IN T-Connect 82
Interfase Dynamica 84
Locator Abutment 86
Accessories for overdenture on Locator 88
Overdentures anchored with ball attachments 90
Accessories for overdentures on ball attachments 92
Overdenture on bars 93
General Indication 94
Composition of the materials 94
UTM Surface
(Ultrathin Threaded Microsurface)
The convergent neck of the Prama implants is characterized by the UTM
(Ultrathin Threaded Microsurface) surface treatment, a particular
micro-threading that allows the perfect control over the connection diameter
and prevents the accumulation of plaque around the junction with the post.
Image of a section of UTM surface obtained using a confocal microscope: the regularity
of the micro-threading of the surface can be noted.
Arrangement of murine myofibroblasts on the UTM surface after 72 hours and observed with fluorescence microscopy
- in vitro testing.
Thanks to the kind concession of the Dentistry Department of the University of Parma.
4
IMPLANT SYSTEM
ZirTi Surface
(Zirconium Sand-Blasted Acid Etched Titanium)
Prama implants are characterized by ZirTi surface, sand-blasted with
zirconium oxide and etched with mineral acids, techniques that give to the
surface a characteristic micromorphology capable of significantly increasing
the bone-to-implant contact area and promoting osseointegration.
The roughness of ZirTi surface, with its Sa medium value of 1.3 μm, is
considered ideal to achieve osseointegration.
The validity of ZirTi surface is documented by numerous experimental and clinical studies. 5
IMPLANT SYSTEM
Prosthetic reccommendations
Prama convergent neck is characterized by a cylindrical part of 0.80 mm and a hyperbolic
part with no sharp edges with a height of 2.00 mm designed in order to guarantee an effective
continuity with the post, thus optimizing the contact between hard and soft tissues, according to
the different clinical and anatomical situations. This absence of sharp edges will allow the soft
tissues to adhere on the titanium without finding obstacles and to reach the profile established by
the prosthodontist. Moreover, it will facilitate the positioning of the prosthetic crown in any part of
the transgingival section.
2.00
0.80
The limits of the prosthetic loading of dental implants derive from mechanical fatigue tests
according to the UNI EN ISO 1480.
Regarding Prama implants it has been observed that in case of full arch and multiple structures,
no restrictions exist for:
• both anterior and distal sectors;
• both cemented and screw-retained prosthesis;
• prosthesis closing on the neck of the implant as well as closing on the post.
In case of both cemented or screw-retained single crowns there are no restrictions for anterior
sectors, while in distal sectors it is necessary to close the prosthesis on the neck of the implant
(Prama IN prosthetic components closing on the neck of the implant are shown in the following
pages).
6
IMPLANT SYSTEM
L: Prama implant A: cylindrical body ZT: ZirTi surface 380: 3.80 mm 060: 6.00 mm
S: RF tapered body 425: 4.25 mm 085: 8.50 mm
500: 5.00 mm 100: 10.00 mm
115: 11.50 mm
130: 13.00 mm
150: 15.00 mm
7
THE RANGE
Prama
The Prama implants have a cylindrical endosseous shape and a convergent neck characterized by
a first cylindrical section and an upper part with a hyperbolic geometry. The Prama implants are
are produced starting from cold-worked Gr. 4 titanium bars.
The implants are also available in the Shorty version for all the sytem
diameters. The height of only 6.00 mm allows them to be used as an
alternative to the maxillary sinus elevation and vertical regeneration
protocols. The surgical instruments required for the insertion of the
Prama implants with a height of 6.00 mm are included in the Shorty
Drilling Kit.
8
THE RANGE
diameter
ø 3.80 mm ø 4.25 mm ø 5.00 mm
height
13.00 13.00
13.00
Please note: The implant nominal length expresses the endosseous length of the implant. The total length of
the implant is 2.80 mm longer than the nominal one, due to the presence of the transgingival neck.
Each implant is sold with its own surgical cover screw. The surgical cover screws are also available on sale
individually in a sterile pack and must be tightened to 8-10 Ncm.
9
See technical characteristics of Gr. 4 titanium on page 94.
THE RANGE
Prama RF
The Prama RF (Root Form) implants have a conical endosseous shape and a convergent neck
characterized by a first cylindrical section and an upper part with a hyperbolic geometry. The
rounded apex makes them ideal for maxillary sinus elevation protocols. The Prama RF implants
are are produced starting from cold-worked Gr. 4 titanium bars.
10
THE RANGE
diameter
height
ø 3.80 mm ø 4.25 mm ø 5.00 mm
Please note: The implant nominal length expresses the endosseous length of the implant. The total length of
the implant is 2.80 mm longer than the nominal one, due to the presence of the transgingival neck.
Each implant is sold with its own surgical cover screw. The surgical cover screws are also available on sale
individually in a sterile pack and must be tightened to 8-10 Ncm.
11
See technical characteristics of Gr. 4 titanium on page 94.
SURGICAL INSTRUMENTS
A practical ratchet is also included that acts as a The kit consists of a practical box in Radel with
dynamometric key for checking the closing torque a surgical tray inside that is set-up to hold the
of the prosthetic screws and as a surgical key for instruments according to a guided procedure. The
inserting the implants. The ratchet has a very small sequences of use of the instruments are indicated by
head, making it easy to use even in distal sectors coloured marks
12
SURGICAL INSTRUMENTS
description code
13
SURGICAL INSTRUMENTS
the tip. LS
LS: Length of the tip. This measurement must
be calculated in addition to the length of
the preparation hole.
7.0 8.5 10.0 11.5 13.0 15.0 18.0
Please note: the initial drills always make a hole that is longer than the implant to be inserted.
The oversizing (LS) is equal to the height of the tip of the drill that is being used. See drawing above.
14
SURGICAL INSTRUMENTS
10.00 10.00
8.50 8.50
ø 2.80 ø 4.25
ø 2.40 ø 3.80
4.80 ø 2.00 ø 3.30
19.30
0.58
15
SURGICAL INSTRUMENTS
ø 2.50
19.50
0.72
*The cylindrical drill FFT3-250-LXS and the related stops are not included in the surgical kit. The whole set of drills and
stops can be ordered with the code KIT-INTEGRA-F250. Drills and stops are also available in single pack.
16
SURGICAL INSTRUMENTS
ø 2.50
18.00
0.50
*The conical drill SE-FK250 and the related stops are not included in the surgical kit. The whole set of drills and
stops can be ordered with the code KIT-INTEGRA-SE-FK250. Drills and stops are also available in single pack.
17
SURGICAL INSTRUMENTS
LL
LL: Total length of the working part, including LS
the tip.
LS: Length of the tip. This measurement must
be calculated in addition to the length of
the preparation hole.
7.0 8.5 10.0 11.5 13.0 15.0 18.0
Please note: the drills always make a hole that is longer than the implant to be inserted.
The oversizing (LS) is equal to the height of the tip of the drill that is being used. See drawing above.
18
SURGICAL INSTRUMENTS
19
SURGICAL INSTRUMENTS
LL
LL: Total length of the working part, including
LS
the tip.
LS: Length of the tip. This measurement must
be calculated in addition to the length of
the preparation hole.
Please note: the drills always make a hole that is longer than the implant to be inserted.
The oversizing (LS) is equal to the height of the tip of the drill that is being used. See drawing above.
20
SURGICAL INSTRUMENTS
18.00
15.00 15.00
21
SURGICAL INSTRUMENTS
22
SURGICAL INSTRUMENTS
Countersink drills
Countersink drills are available optionally. They allow to prepare the neck of Prama and Prama RF
implants in presence of a very thick cortical bone. The drills are characterized by an apical portion
that is guided into the hole created by the final cylindrical drill and by a standard height of the
working part of 6.60 mm.
24
SURGICAL INSTRUMENTS
The particular morphology of the countersink drills allows to widen the initial part of the hole bored
by the final cylindrical drills to prepare the cortical bone portion, that will be in contact with the
neck of the implant. The maximum recommended speed is 1,000 rpm.
Each drill has to be used only for the implant of equal diameter.
h 6.60 mm
h 6.60 mm
25
SURGICAL INSTRUMENTS
LL
the tip
LS: Length of the tip. This measurement must
be calculated in addition to the length of
the preparation hole
Please note: the drills always make a hole that is longer than the implant to be inserted.
The oversizing (LS) is equal to the height of the tip of the drill that is being used. See drawing above.
The drills for distal sectors are not included in any surgical kit and must be purchased separately.
They cannot be used with depth stops.
26
SURGICAL INSTRUMENTS
Bone taps
Prama and Prama RF implants are self-tapping implants with excellent cutting capacity and are
easy to insert. However, the use of a bone tap is recommended in all cases where the type of bone
requires it in order to make inserting the fixture easier. They are available both with right angle
shank and with a hexagonal connector for dynamometric ratchet, and have a specific design for
the two endosseous morphologies.
11.50
10.00 13.00
8.50 15.00
7.00 18.00
Osteotomes
Osteotomes for expansion protocols are available for each Prama and Prama RF endosseous
morphology, not included in the surgical kit. The laser-etched codes on the handles indicate the
osteotome diameter, to make it easier to recognize the correct surgical sequence. A practical
universal instrument case for storing and organising them is available and can be ordered
separately using the code OS-TRAY.
Prama osteotomes
Osteotomes are optional instruments that are not included in the surgical kit. They can be purchased separately and singularly.
11.50
10.00 13.00
8.50 15.00
7.00 18.00
description code
28
SURGICAL INSTRUMENTS
Prama RF osteotomes
Prama RF osteotomes are designed considering the height and diameter of the implant to be
inserted. They present two laser-etched markings: one corresponds to the height of the fixture and
the other one, at a lower height, is useful for an intermediate check during the preparation phase.
In the osteotome for 10.00 mm implants the laser markings correspond to the height 8.50 and
10.00 mm, therefore the instrument can also be used for inserting implants of height 8.50 mm.
ø 3.50 10.00 ø 3.50 10.00 ø 3.80 10.00 ø 3.80 10.00 ø 4.60 10.00 ø 4.60 10.00
8.50 8.50 8.50 8.50 8.50 8.50
ø 3.50 11.50 ø 3.50 11.50 ø 3.80 11.50 ø 3.80 11.50 ø 4.60 11.50 ø 4.60 11.50
10.00 10.00 10.00 10.00 10.00 10.00
ø 3.50 13.00 ø 3.50 13.00 ø 3.80 13.00 ø 3.80 13.00 ø 4.60 13.00 ø 4.60 13.00
10.00 10.00 10.00 10.00 10.00 10.00
ø 3.50 15.00 ø 3.50 15.00 ø 3.80 15.00 ø 3.80 15.00 ø 4.60 15.00 ø 4.60 15.00
Osteotomes are optional instruments that are not included in the surgical kit.
They can be purchased separately and singularly.
29
SURGICAL INSTRUMENTS
Additional instruments
All the additional instruments for inserting Prama and Prama RF implants, made of surgical
stainless steel, have been designed to offer maximum ergonomics and ease of use. All the
instruments have a laser-marked code for easy identification of the pieces. All the instruments
are available individually as spare parts.
30
SURGICAL INSTRUMENTS
Ratchet
31
SURGICAL INSTRUMENTS
Surgical screwdrivers
Mounter
32
SURGICAL INSTRUMENTS
33
SURGICAL INSTRUMENTS
Parallelism pins
34
implants,
description
description
description
depth gauge
Depth gauge
Spare O-ring
real dimensions
X-ray templates
dynamometric key
box with 5 spare o-rings for all
H H
2.80 mm H 2.80 mm
ø max 2.80 mm ø max
spira ø max spira
H H
nominale spira nominale
RIPRODUZIONE SCALA MAGGIORATA DEL 30% - DIMENSIONS ENLARGED BY 30% RIPRODUZIONE SCALA REALE - REAL DIMENSIONS
RIPRODUZIONE SCALA MAGGIORATA DEL 20% -DIMENSIONS ENLARGED BY 20%
FIXTURES PRAMA FIXTURES PRAMA
FIXTURES PRAMA
L-L120
L-L130
L-L100
PROF3
H 6.00 mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm H 6.00 mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm
H 6.00 mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm
LA-ZT-380-060 LA-ZT-380-085 LA-ZT-380-100 LA-ZT-380-115 LA-ZT-380-130 LA-ZT-380-150 LA-ZT-380-060 LA-ZT-380-085 LA-ZT-380-100 LA-ZT-380-115 LA-ZT-380-130 LA-ZT-380-150
LA-M-380-060 LA-M-380-085 LA-M-380-100 LA-M-380-115 LA-M-380-130 LA-M-380-150 LA-ZT-380-060 LA-ZT-380-085 LA-ZT-380-100 LA-ZT-380-115 LA-ZT-380-130 LA-ZT-380-150 LA-M-380-060 LA-M-380-085 LA-M-380-100 LA-M-380-115 LA-M-380-130 LA-M-380-150
LA-M-380-060 LA-M-380-085 LA-M-380-100 LA-M-380-115 LA-M-380-130 LA-M-380-150
ø 3.80 mm
ø 3.80 mm
ø 3.80 mm
ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira
3.85 mm 3.85 mm 3.85 mm 3.85 mm 3.85 mm 3.85 mm ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira 3.85 mm 3.85 mm 3.85 mm 3.85 mm 3.85 mm 3.85 mm
3.85 mm 3.85 mm 3.85 mm 3.85 mm 3.85 mm 3.85 mm
H 6.00mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm H 6.00mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm
H 6.00mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm
LA-ZT-425-060 LA-ZT-425-085 LA-ZT-425-100 LA-ZT-425-115 LA-ZT-425-130 LA-ZT-425-150 LA-ZT-425-060 LA-ZT-425-085 LA-ZT-425-100 LA-ZT-425-115 LA-ZT-425-130 LA-ZT-425-150
LA-M-425-060 LA-M-425-085 LA-M-425-100 LA-M-425-115 LA-M-425-130 LA-M-425-150 LA-ZT-425-060 LA-ZT-425-085 LA-ZT-425-100 LA-ZT-425-115 LA-ZT-425-130 LA-ZT-425-150 LA-M-425-060 LA-M-425-085 LA-M-425-100 LA-M-425-115 LA-M-425-130 LA-M-425-150
LA-M-425-060 LA-M-425-085 LA-M-425-100 LA-M-425-115 LA-M-425-130 LA-M-425-150
ø 4.25 mm
ø 4.25 mm
ø 4.25 mm
ORING180-088
ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira
4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira 4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm
4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm
C C
C
H 6.00 mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm H 6.00 mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm
CM
MY MY
ø 5.00 mm
ø 5.00 mm
MY
ø 5.00 mm
CY ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira CY ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira
5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm CY ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira 5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm
CMY 5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm CMY
CMY
H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm
H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm
LS-ZT-380-085 LS-ZT-380-100 LS-ZT-380-115 LS-ZT-380-130 LS-ZT-380-150 LS-ZT-380-085 LS-ZT-380-100 LS-ZT-380-115 LS-ZT-380-130 LS-ZT-380-150
LS-M-380-085 LS-M-380-100 LS-M-380-115 LS-M-380-130 LS-M-380-150 LS-ZT-380-085 LS-ZT-380-100 LS-ZT-380-115 LS-ZT-380-130 LS-ZT-380-150 LS-M-380-085 LS-M-380-100 LS-M-380-115 LS-M-380-130 LS-M-380-150
LS-M-380-085 LS-M-380-100 LS-M-380-115 LS-M-380-130 LS-M-380-150
ø 3.80 mm
ø 3.80 mm
ø 3.80 mm
ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira
3.85 mm 3.85 mm 3.85 mm 3.85 mm 3.85 mm ø max spira ø max spira ø max spira ø max spira ø max spira 3.85 mm 3.85 mm 3.85 mm 3.85 mm 3.85 mm
3.85 mm 3.85 mm 3.85 mm 3.85 mm 3.85 mm
H 6.00 mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm H 6.00 mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm
H 6.00 mm H 8.50 mm H 10.00 mm H 11.50 mm H 13.00 mm H 15.00 mm
LS-ZT-425-060 LS-ZT-425-085 LS-ZT-425-100 LS-ZT-425-115 LS-ZT-425-130 LS-ZT-425-150 LS-ZT-425-060 LS-ZT-425-085 LS-ZT-425-100 LS-ZT-425-115 LS-ZT-425-130 LS-ZT-425-150
LS-M-425-060 LS-M-425-085 LS-M-425-100 LS-M-425-115 LS-M-425-130 LS-M-425-150 LS-ZT-425-060 LS-ZT-425-085 LS-ZT-425-100 LS-ZT-425-115 LS-ZT-425-130 LS-ZT-425-150 LS-M-425-060 LS-M-425-085 LS-M-425-100 LS-M-425-115 LS-M-425-130 LS-M-425-150
LS-M-425-060 LS-M-425-085 LS-M-425-100 LS-M-425-115 LS-M-425-130 LS-M-425-150
ø 4.25 mm
ø 4.25 mm
ø 4.25 mm
ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira
4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira 4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm
4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm 4.25 mm
ø 5.00 mm
ø 5.00 mm
ø 5.00 mm
ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira
5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm ø max spira ø max spira ø max spira ø max spira ø max spira ø max spira 5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm
5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm 5.00 mm
kit
kit
kit
ZPRAMA-INT
ZPRAMA-INT
ZPRAMA-INT
ZPRAMA-INT
Not included in
the surgical kit,
available separately
SURGICAL INSTRUMENTS
35
SURGICAL INSTRUMENTS
Pilot drill
FPS-200 STOPS4-200-060
Shorty pilot drill Stop 6.00 mm for Shorty pilot drill
25.00
ø 2.00
FGS-200/300 FGS-340/425
ø 3.00 ø 4.25
ø 2.50 ø 3.80
ø 2.00 ø 3.40
Parallelism pin
parallelism pin for short drills with depth lines at 5.00, 6.00 and 7.00 mm
PPS-2/3
19.00 mm
8.30 mm
ø 2.80 mm ø 2.00 mm
5.00 5.00
6.00 6.00
7.00 7.00
36
SURGICAL INSTRUMENTS
LL
LL: Total length of the working part, including
the tip
Drills for Prama Shorty implants are contained in the Shorty Drilling Kit, as well as available separately.
The Shorty Drilling Kit is a set of drills for the insertion of short implants with cylindrical morphology.
description code
Important warning
The Shorty Drilling Kit contains only drills and two parallelism pins. It is not a complete surgical kit: to insert implants
must be used all the other instruments included in the standard surgical kit (driver, ratchet, screwdrivers, etc.).
37
SURGICAL INSTRUMENTS
SE-FP200 SE-STOP-FP200-060
short pilot drill Stop 6.00 mm for short pilot drill
ø 2.00
15.00
LL
38
SURGICAL INSTRUMENTS
LL
LL: Total length of the working part,
including the tip.
Drills for Prama RF Shorty implants are contained in the Syra Short Drilling Kit, as well as available separately.
description code
Important warning
The Syra Short Drilling Kit contains only drills. It is not a complete surgical kit: to insert Prama RF implants
must be used all the other instruments included in the standard surgical kit (driver, ratchet, screwdrivers, etc.).
39
PROSTHETIC COMPONENTS
40
PROSTHETIC COMPONENTS
description code
M 1.8
M 1.8
M 1.8
M 1.8
42
PROSTHETIC COMPONENTS
Pull-up transfer: the connection with tabs Implant analog: anodised according to the
clicks into the implant fixation hexagon, colour code to facilitate recognition and
without requiring a screw laboratory phases
44
PROSTHETIC COMPONENTS
M 1.8
ø 3.30
See technical characteristics of titanium Gr. 5 and PEEK on pages 95 and 97. 45
PROSTHETIC COMPONENTS
Temporary posts
The temporary posts can be used in a conventional way after the bone healing period, or
immediately after surgical insertion of the implants, if conditions exist for immediate loading.
They can also be used as an alternative to the traditional healing abutments for reconditioning the
soft tissues, depending on the prosthetic protocol adopted.
The Gr. 5 titanium temporary posts have been designed to give a resistant support in case both
of single crowns and multiple or full arch rehabilitations. The connection presents an hexagon
for single crowns, while it’s not indexed for multiple structures or full arch; the posts for multiple
structures present a morphology that facilitates the insertion and removal manoeuvres and it is
deep enough to guarantee more stability. PEEK temporary posts with a Gr. 5 titanium base are
also available, ideal to support single cemented crowns.
The REEF resin has a great Titanium offers great resistance in case of
chair-side use easiness allowing an long-term temporary rehabilitations
easy relining
46
PROSTHETIC COMPONENTS
description code
ø 3.30
ø 3.30
ø 3.30 1.80
ø 3.30 2.00
ø 3.30 2.00
ø 3.85
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity. It is not possible to use the screw with conical support with these
posts (code L-VMS-180).
47
See technical characteristics of Gr. 5 titanium, PEEK and REEF resin on pages 95, 96, 97 e 98.
PROSTHETIC COMPONENTS
The conical support of the screws used with the pre-made posts increases
resistance to unscrewing by 20% compared with the standard screws
Important warning
It is recommended not to use standard fixation screws (code VM2-180) with these posts because standard
screws do not have the conical support, so they do not interface exactly with the seat of the screw head
inside these prosthetic components. Failure to observe this warning leads to the risk of unscrewing or
breaking of the screw.
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity.
Pre-made posts
description code
Prama IN
preparable post Prama IN preparable post
incorporating incorporating the implant-post
the implant-post interface for a height of 1.50 mm
interface for a height
of 0.50 mm
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity.
50
PROSTHETIC COMPONENTS
Preparable posts
description code
ø 4.40
simple preparable posts A-MFS-330
engaging 9.50
very wide emergence
standard fixation screw included ø 3.30 0.80
Important warning
It is recommended not to use standard fixation screws (code VM2-180) with these posts because standard
screws do not have the conical support, so they do not interface exactly with the seat of the screw head
inside these prosthetic components. Failure to observe this warning leads to the risk of unscrewing or
breaking of the screw.
52
PROSTHETIC COMPONENTS
description code
ø 3.80
Recommended torque for castable posts with metal base: 20-25 Ncm.
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity. It is not possible to use the screw with conical support with these
posts (code L-VMS-180).
See technical characteristics of Gr. 5 titanium, PMMA, gold alloy “1” and cobalt chrome on pages 95, 98, 100 e 101. 53
PROSTHETIC COMPONENTS
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity.
Important warning
It is recommended not to use standard fixation screws (code VM2-180) with these posts because standard
screws do not have the conical support, so they do not interface exactly with the seat of the screw head
inside these prosthetic components. Failure to observe this warning leads to the risk of unscrewing or
breaking of the screw.
* The fixation screws with Full Head technology (L-VM-180) must be tightened with the appropriate drivers for
screws with Full Head technology contained with the Prama surgical kit.
54
PROSTHETIC COMPONENTS
Recommended torque for castable posts with metal base: 20-25 Ncm. 55
See technical characteristics of PMMA, gold alloy and cobalt chrome on pages 98, 100 and 101.
PROSTHETIC COMPONENTS
56
PROSTHETIC COMPONENTS
description code
M 1.8
standard fixation screw
supplied with the castable posts, it can also be
ordered separately as a spare
Recommended torque for posts obtained from the casting of fully castable posts: 20-25 Ncm.
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity. It is not possible to use the screw with conical support with these
posts (code L-VMS-180).
*Dynamic Abutment posts are medical devices manufactured and patented by Talladium España S.L.,
58 Avenida Blondel, 54 3°, 25002 Lleida, Spain. Dynamic Abutment is a registered trademark of this company.
PROSTHETIC COMPONENTS
description code
ø 3.30
ø 3.30
ø 3.30
ø 3.30
description code
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity.
60
PROSTHETIC COMPONENTS
description code
11.00
ø 3.30
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity.
The PLAIN abutments, made of Gr. 5 titanium, take advantage of the completely flat geometry of their
upper part which, by means of a very small guide, is joined to normal castable sleeves of PMMA. The
usefulness of these abutments is therefore that they maximize centring and repositioning operations
of structures screw retained onto several implants.
62
PROSTHETIC COMPONENTS
description code
5.00
ø 3.30
ø 3.30
ø 3.30 1.05
ø 3.30
11.00
ø 3.30
M 2.0
Recommended torque for Pick-up transfers and PLAIN healing caps: 8-10 Ncm.
Recommended torque for PLAIN castable sleeves: 20-25 Ncm.
Recommended torque for PLAIN abutments: 25-30 Ncm.
See technical characteristics of Gr. 5 titanium and PMMA on pages. 95 e 98. 63
PROSTHETIC COMPONENTS
15°
17°
15°
30°
64
PROSTHETIC COMPONENTS
description code
M 1.8
M 1.8
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity.
65
See technical characteristics of Gr. 5 titanium on page 95.
PROSTHETIC COMPONENTS
description code
ø 5.00
ø 5.00
ø 5.00
ø 5.00
M 1.4
M 1.4
12.00
ø 5.00
ø 5.00
description code
castable posts in PMMA with a pre-made base in gold alloy PAD-UC ø 3.80
type “1”, rotating, non engaging, for overcasting on
10.50
P.A.D. abutments
fixation screw included (code PAD-VP-140)
ø 5.00 3.20
description code
ø 5.00
ø 5.00
ø 5.00
ø 5.00
67
PROSTHETIC COMPONENTS
STEP 3
The product is cast in the laboratory,
obtaining the metal structure for making the
finished truss
STEP 1
The castable bar is fixed to the caps with a
photopolymerisable resin
STEP 2
After photo-polymerisation of the truss, the
product is removed from the oral cavity
68
PROSTHETIC COMPONENTS
description code
ø 5.00
70
PROSTHETIC COMPONENTS
description code
1.70
1.70
1.70
1.70
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity.
Important warning
It is recommended not to use standard fixation screws (code VM2-180) with these posts because standard
screws do not have the conical support, so they do not interface exactly with the seat of the screw head
inside these prosthetic components. Failure to observe this warning leads to the risk of unscrewing or
breaking of the screw.
See technical characteristics of Gr. 5 titanium and cobalt chrome on pages 95 e 101. 71
PROSTHETIC COMPONENTS
Conoweld posts
These abutments are made of Gr. 5 titanium for the purpose of using the Conoweld conometric
technique. This technique embraces the advantages of two protocols already widely established
in oral implantoprosthesis: intraoral welding and conometric retention, both for temporary and
final solution, employing cement-less prosthetic components and removable by the clinician.
72
PROSTHETIC COMPONENTS
The Conoweld range includes three different caps, which are universal in relation to the diameters
of the posts and the implant platforms: this is due to the fact that conometric retention is in the
most coronal portion of the post, which always has the same dimensions.
The two titanium caps differ in thickness: the cap designed for the construction of a welded
structure intraorally for the temporary stage is thicker in order to withstand the welding with the
titanium bars, without bonding with the underlying post, while the cap designed to anchor the final
prosthesis glued on is thinner, in order to reduce the impact on the anatomical morphology of the
prosthesis; it should not, therefore, be used for the welding.
A cap in PMMA is also available, which allows a precise impression to be taken even when working
without a intraoral welding machine and which can be used for modelling and casting a structure
entirely in cobalt chrome or other alloys, when the decision has been taken not to use luting
techniques for assembly.
5° 10° 15°
73
PROSTHETIC COMPONENTS
description code
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity. It is not possible to use the screw with conical support with these
posts (code L-VMS-180).
74
Recommended torque for Conoweld posts: 20–25 Ncm.
PROSTHETIC COMPONENTS
description code
17.00
See technical characteristics of Gr. 2 titanium, Gr. 5 titanium and PMMA on pages 95, 96 e 98. 75
COMPONENTI PROTESICHE
76
COMPONENTI PROTESICHE
77
COMPONENTI PROTESICHE
description code
ø 3.30
6.00
1.00
description code
ø 5.00
6.00
ø 5.00 1.00
T-Connect
T-Connect supports are made of Gr. 5 titanium and can be used to create custom-made posts
for single or multiple screw retained prostheses as Toronto which lean on the connection
platform. Posts and structures can be fabricated in titanium, zirconia, cobalt-chrome, PEEK and
biomedical resin, by means of open CAD CAM systems, including ECHO by Sweden & Martina,
without sacrificing the micrometric precision in coupling between platforms that can be obtained
with conventional components. ECHO users too can choose to use T-Connect supports: posts in
zirconium produced in this way have a small titanium base that prevents the contact between the
zirconium body and the implant platform. If posts are realized using ECHO software, see page 74 for
the scanbody codes to use.
For further information about compatible systems contact the Sweden & Martina CAD CAM product
specialists at +39 049 9124394.
80
PROSTHETIC COMPONENTS
description code
T-Connect A-BASTZR-S-330-4
engaging
cone h 4.00 mm 4.00
fixation screw included ø 4.20
0.40
T-Connect A-BASTZR-S-330-6
engaging
cone h 6.00 mm
fixation screw included 6.00
ø 4.20
0.40
T-Connect A-BASTZR-M-330-4
non engaging
cone h 4.00 mm 4.00
fixation screw included ø 4.20
0.40
T-Connect A-BASTZR-M-330-6
non engaging
cone h 6.00 mm 6.00
fixation screw included
ø 4.20
0.40
4.00
ø 4.20
6.00
ø 4.20
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity.
It is not possible to use the screw with conical support with these posts (code L-VMS-180).
Prama IN T-Connect
Prama IN T-Connect supports can be used to create custom-made posts for single prostheses
or Toronto multiple screw retained structures closing on the neck of the implant for a height
of 0.50 mm. Posts and structures can be made of titanium, zirconium, cobalt-chrome, PEEK and
biomedical resin, using open CAD CAM systems, including Sweden & Martina’s ECHO, without
sacrificing micrometric precision in coupling between platforms that can be obtained with
conventional components.
Prama IN T-Connect supports can be also used for luting crowns or multiple structures fabricated in
the laboratory with the traditional casting techniques.
For posts made with ECHO software, see pages 78-79 for the codes of the scanbodies to be used.
For further information on compatible systems please contact the Sweden & Martina CAD CAM
product specialists at +39 049 9124394.
82
PROSTHETIC COMPONENTS
Important warning
It is recommended not to use standard fixation screws (code VM2-180) with these posts because standard
screws do not have the conical support, so they do not interface exactly with the seat of the screw head
inside these prosthetic components. Failure to observe this warning leads to the risk of unscrewing or
breaking of the screw.
Important warning
It is recommended always to use test screws for the laboratory phases and to keep the new screw supplied
for the final fastening in the oral cavity.
Interfase Dynamica
Interfase Dynamica* supports are a patented solution that allows to produce prostheses in the
aesthetic zone designed with CAD CAM systems. Their morphology permits to move the screw
hole more palatally or lingually, so avoiding unsightly vestibular holes. The supports must be
tightened with a screwdriver with a special hexalobular design, which allow the head of the screw
to be engaged even in presence of extreme angulations, up to a maximum of 30°.
Interfase Dynamica supports are available in indexed version with hexagon as well as in rotating
version, and allow to produce crowns, posts and multiple structures with luting technique in
different materials with open CAD CAM systems, including ECHO by Sweden & Martina.
Interfase Dynamica
supports are
contained in the
libraries of the
softwares: Exocad,
Dental Wings and
3Shape
*Interfase Dynamica supports are medical devices manufactured and patented by Talladium España S.L., Avenida Blondel, 54 3°,
84 25002 Lleida, Spain. Interfase Dynamica is a registered trademark of the same company.
PROSTHETIC COMPONENTS
description code
description code
Locator Abutment
Locator Abutments* are a patented and versatile prosthetic solution for attaching overdentures
to dental implants easily and safely. The Locator system allows easily correcting misalignment
of divergent implants by up to 40° (20° for each implant) in limited occlusal spaces and, given
the limited amount of space occupied, is perfect for patients with a removable prosthesis.
The abutments are made of Gr. 5 titanium and are available in different transgingival heights; they
must be tightened at 25-30 Ncm, using the special drivers (see the codes on page 89).
See following pages for the list of accessories available.
*Locator abutments are medical devices manufactured and patented by Zest Anchors, Inc., 2061 Wineridge Place,
Escondido, CA 92029, USA. Locator is a registered trademark of Zest Anchors, Inc. The European Agent for the purposes
of MDD 93/42/EEC is Ventura Implant and Attachment Systems, 69 The Avenue, Ealing, London W13 8JR, England.
86
PROSTHETIC COMPONENTS
description code
M 1.8
M 1.8
M 1.8
88
PROSTHETIC COMPONENTS
description code
steel push rod for inserting the retainers in the caps 8397
Not necessary for those who already own or order the
complete Locator Core Tool separately
retention jacket for the driver (8390) for transferring the 8394
Locator abutments into the oral cavity
89
PROSTHETIC COMPONENTS
90
PROSTHETIC COMPONENTS
description code
ø 3.30 15.30
description code
description code
spare retention spring for titanium cap, soft, for progressive MOL2-CAP-TIT-1
adaptation of the prosthesis, steel, ø 3.20 mm
description code
cap in gold alloy 2, complete with plastic positioning ring for CAP-1
ball attachments ø 2.20 mm. The total height is 3.10 mm, and
the outside diameter is 3.50 mm
92
PROSTHETIC COMPONENTS
description code
Overdenture on bars
description code
93
GENERAL INDICATIONS
titanium remainder -
*This technical information complies with the express specifications of the regulations in force on the use of
Gr. 4 titanium in implantology:
• ASTM F67-13: Standard Specification for unalloyed titanium, for surgical implant applications.
• ISO 5832-2: 2012: Implants for surgery – Metallic materials – Part 2: Unalloyed titanium.
Please note: the use of cold-worked Gr. 4 titanium bars for the production of Sweden & Martina implants
allows the exploitation of mechanical characteristics higher than those required by applicable standards.
Furthermore, the excellent results documented since 1996 corroborate the choice of the cold-working
production process and of ZirTi surface treatments,which express and enhance the raw material potential
selected by Sweden & Martina.
94
GENERAL INDICATIONS
titanium a bilancio -
* This technical information complies with the express specifications of the regulations in force on the use of Gr. 5 titanium in
implantology:
• ASTM F 136-13: Standard Specification for wrought Titanium-6 Aluminium-4 Vanadium Eli (Extra low interstitial) Alloy for surgical
applications;
• ISO 5832-3:2012: Implants for surgery – Metallic materials – Part 3: Wrought Titanium-6 Aluminium-4 Vanadium Alloy.
95
GENERAL INDICATIONS
titanium remainder -
* This technical information complies with the express specifications of the regulations in force on the use of
titanium Gr. 2 in implantology.
Resina REEF
REEF resin
96
GENERAL INDICATIONS
PEEK
PEEK
thermal properties
chemical properties
97
GENERAL INDICATIONS
PMMA
PMMA
colour transparent
thermal properties
98
GENERAL INDICATIONS
POM
POM
thermal properties
chemical properties
99
GENERAL INDICATIONS
Gold alloy
composition % of reference
Au 60 % > 68.60 % 70 %
Pt 24 % 2.45 % 8.5 %
Pd 15 % 3.95 % -
Ir 1% 0.05 % 0.10 %
Ag - 11.85 % 13.40 %
Cu - 10.60 % 7.50 %
Zn - 2.50 % 0.50 %
Ru - - -
limit of elasticity 400 MPa (annealed) > 710 MPa 380 MPa (annealed)
700 (after deformation) 730 (after deformatione)
800 (after casting)
• Gold alloy “1”: all castable posts with a premade alloy base (e.g. VSR-UCR. etc).
• Gold alloy “2”: CAP-1 cap for ball attachments in gold alloy.
• Gold alloy “3”: CAV-375 bar attachment for round bars ø 2.20 mm
100
GENERAL INDICATIONS
C 0.10
Mn 1.00
Cr 26.00 ÷ 30.00
Ni 1.00
Mo 5.00 ÷ 7.00
N 0.25
Fe 0.75
Co remainder
elongation at yield 25 %
section reduction 23 %
hardness 30 HTc
thermal properties
101
BIBLIOGRAPHY ON SWEDEN & MARTINA IMPLANTS
Sweden& Martina offers to dentists the opportunity to promptly get an update on the
international scientific studies on its implant systems with the volumes SCIENTIFICA, where you
can find the abstracts of all publications in the most relevant international papers specialized in
implantology.
102
C-IMP-PRAMA-E
rev.11-18
Sweden & Martina S.p.A. Sweden & Martina Ltd Sweden & Martina Inc. Sweden & Martina Mediterranea S.L. - España
Via Veneto, 10 Unit 45 Distributor for U.S. info.es@sweden-martina.com
35020 Due Carrare (PD), Italy Basepoint Business Centre 80 Broad Street, 5th floor, Suite 502 Sweden & Martina Lda - Portugal
Tel. +39.049.9124300 Metcalf Way, Crawley, New York, NY 10004 info.pt@sweden-martina.com
Fax +39.049.9124290 West Sussex, RH11 7XX, UK Toll free 1-844-8MARTINA (1-844-862-7846)
info@sweden-martina.com Toll free 0800 1123575 info.us@sweden-martina.com
info.uk@sweden-martina.com
www.sweden-martina.com www.sweden-martinainc.com
The implants, standard prosthetic components and surgical instruments contained in this catalogue are Medical devices and are manufactured by Sweden & Martina S.p.A. They
conform to the ISO 9001 and ISO 13485 standards and are certified with the CE Mark (Class I) and CE 0476 mark (Class IIA and class IIB) in compliance with European Medical Device
Directive No. 93/42 and European Directive No. 2007/47/CE.
We have met the good manufacturing standards (GMP) set forth by many countries worldwide, including the United States FDA.
The contents are updated at the time of publication. Check with the company for any subsequent updates.