Professional Documents
Culture Documents
net/publication/279999743
Comparison between View Summing and Tube Pulsing in Ultra Low Dose CT
Acquisitions for PET Attenuation Correction
CITATIONS READS
0 880
5 authors, including:
Bruno De Man
GE Global Research
141 PUBLICATIONS 3,411 CITATIONS
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
GPU based Fast Projection and Backprojection model development View project
All content following this page was uploaded by Paul E Kinahan on 18 July 2015.
ABSTRACT
The image quality needed for CT-based attenuation correction (CTAC) is significantly lower than what is used currently
for diagnostic CT imaging. Consequently, the X-ray dose required for sufficient image quality with CTAC is relatively
small, potentially smaller than the lowest X-ray dose clinical CT scanners can provide. Operating modes have been
proposed in which the X-rays are periodically turned on and off during the scan in order to reduce X-ray dose. This study
reviews the different methods by which X-rays can be modulated in a CT scanner, and assesses their adequacy for low-
dose acquisitions as required for CTAC. Calculations and experimental data are provided to exemplify selected X-ray
pulsing scenarios. Our analysis shows that low-dose pulsing is possible but challenging with clinically available CT
tubes. Alternative X-ray tube designs would lift this restriction.
Keywords: Pulsing, X-rays, Computed tomography, CT-based attenuation correction, PET/CT, low-dose, sparse
sampling
Medical Imaging 2014: Physics of Medical Imaging, edited by Bruce R. Whiting, Christoph Hoeschen, Despina Kontos,
Proc. of SPIE Vol. 9033, 90332Z · © 2014 SPIE · CCC code: 1605-7422/14/$18 · doi: 10.1117/12.2043021
1 2
4
relative
1
0 X-ray
intensity
continuous pulsed
acquisition acquisition
X-ray
fan
beam
Figure 1. Illustration of two methods to reduce the dose. The left side – region 1 – shows a reduction in tube current. The
right side – region 2 – shows a pulsed approach, in which case the same total intensity of photons is divided into a subset of
the views with higher intensity. In this example the tube is off ¾ of the time for a 25% duty cycle, and on at 4x the intensity
of the continuous acquisitions. Thus total radiation dose is the same for both modes.
In this paper we focus on describing various methods for achieving tube current pulsing, and we present calculations and
experimental data for selected approaches.
3. PULSING REQUIREMENTS
In a state-of-the-art CT scanner, the X-ray tube rotates around the patient, at a typical speed of 0.35 s to 2 s per rotation,
and the equipment samples up to 2496 views per rotation [12], i.e. up to one view every 140 µs [12]. This is 2.5 times
more views per rotation [13] than a typical system with 1000 views per rotation [7]. Such a typical system would have a
view duration of 1 ms at a speed of one rotation per second.
In a standard CT system, X-rays are emitted continuously during the exposure, and the fast read-out electronics of the
detector sample the integrated signal at the end of each view duration. In that case, X-rays are turned on and off only
once for each scan, and rise and fall durations are of little importance.
However, for pulsing on a view-by-view basis, we want the on/off transition times to be at least one order of magnitude
shorter than the view duration. This implies that the on/off transition times for each pulse should be 10 µs to 100 µs,
since the view duration is on the order of 100 µs to 1 ms. For CTAC, operation at the slower end of these ranges is
sufficient, due to the relatively low image quality required for this application.
Tube Current
X-Ray Tube
Spectral Filter
Output
Window Collimator
Tube =
Voltage Target (Anode)
X-Rays
40kV…150kV Electron Beam
Grid = Emitter
Grid Electrode
Voltage (Cathode)
-1kV…-10kV
2A…10A
Filament Current
Figure 2. Schematic drawing of an X-ray tube. Gridding capability is added here – although currently only available for
interventional tubes.
In the next section we present an overview of tube pulsing and modulation technologies along with their strengths and
weaknesses.
Tube Current
Tube Voltage
Figure 3. Schematic drawing of tube current as a function of tube voltage, for an X-ray tube with a thermal emitter at a given
temperature. This type of dependency can be applied to high-voltage switching. Tube voltage here goes up to roughly
100 kV, while tube current is on the order of 100 mA.
Time
Figure 4. Schematic drawing of tube voltage and tube current waveforms of a high-voltage switched pulse. Note that the
tube current is a function of tube voltage, as illustrated in Fig. 3.
Tube Voltage
Tube Current
Time
Figure 5. Schematic drawing of tube voltage and tube current waveforms of a gridded pulse. Note that the tube voltage is
stable during the entire pulse duration.
In the next sections we provide a more detailed description of two pulsing technologies at opposite ends of the spectrum
in terms of maturity in clinical systems: high-voltage switching (that is available for implementation in CT systems
today, without hardware changes), and extraction grid as used in multi-source technologies (an elegant, hardware based
approach, demonstrated on research CT prototypes, but not available on clinical systems).
Anode
AC Power
Supply
Cathode
>
Y
60
50
cq) 40
> 30
á 20
~ 10
0
10
-0.2 0.0 0.2 0.4 0.6 0.8 1.0 12
Time (ms)
Figure 7: Measured data for a high-voltage switched tube pulse, nominally operating at 80kVp and 400mA [22].
In high-voltage switched pulsing, each pulse is a separate exposure with its own tail. During the tailing time, the tube
voltage decreases, and softer X-rays are created. These softer X-rays are more strongly absorbed by the patient, which
means that they contribute to patient dose, but only marginally to the image [16]. Softer X-rays are also created during
the voltage rise interval, but since the rise time is typically a lot shorter, we ignore that part in the following discussion.
The finite duration of the tail has two implications. First, there is a minimum amount of dose associated with each pulse,
hence a constraint on the number of pulses to remain below a maximum dose level. Second, there is a finite time
associated with each pulse and hence, for a given rotation time, a limit on the number of pulses that can be fired.
The charge and the size of the tail will be quantified in subsection 6.2. The dose constraints are studied in subsection 6.3
and the timing constraints in subsection 6.4.
Typical values of high-voltage capacitance C=1nF and tube voltage V=100kV result in a minimum charge per pulse of
Q=0.1mAs. As a sanity check, we compare the minimum pulse charge of Q=0.1mAs to the minimum pulse of a recent
vascular system, which in high-voltage switching mode is 0.2 mAs [18]. A vascular system typically has longer high-
voltage cables than a CT system, and therefore larger high-voltage capacitance, which leads to larger minimum pulses
and confirms the estimate.
On the other hand, each time the high voltage needs to be decreased, this high-voltage capacitance needs to be
discharged. As simple power supplies, including high-voltage generators, typically comprise a rectifier that only allows
an output current in one direction, this charge Q cannot go back to the power supply at the end of the pulse, but instead it
needs to be dissipated into the X-ray tube. This is done through the tube current I, with
t2
Q1 Q2 I t dt (2)
t1
where t1 and t2 denote the start and end of the voltage decay, respectively, and Q Q1 Q2 is the amount of charge
that needs to be discharged. Assuming constant tube current during the discharge, we get a lower bound of the discharge
time from
Q
tdischarge (3)
I
So for a charge of 0.1mAs and a tube current of 100mA, the discharge time is approximately 1ms.
The minimum dose in each pulse is the sum of the rise time dose and the discharge time dose. The charge delivered to
the X-ray tube during the rise time can be approximated as tube current multiplied by rise time. The charge delivered to
the X-ray tube during high-voltage discharge was derived above. Based on this minimum overall pulse charge we can
compute a maximum number of pulses, i.e. where the dose would reach parity with the continuous low mA scan. This is
represented by the solid blue curve in Fig. 8.
10 Time Limit
Dose Limit
Dose Limit with Additional Filtering
Max. # of Pulses
Max. # of Pulses with Add. Filtering
1
1 10 100 1000
Tube Current (mA)
Figure 8: Maximum number of pulses per rotation, with respect to the time limit (red line) and dose limit (blue lines). Time
limit is calculated for the case where the pulses follow each other without off-time. Dose limit is for same dose as a
continuous exposure at 10mA, 0.5s/rot. Other parameters used are tube voltage 100kV, cable capacitance 1nF, and pulse rise
time 0.2ms. The dashed blue line illustrates a case where spectral filtering reduces patient dose in the pulsed mode by a
factor of 10. The green circles correspond to the maximum number of pulses that satisfy both the time limit and the dose
limit, in the case without and with additional filtering.
In the next two subsections we are going to assess the performance of high-voltage switching with respect to two criteria:
reduction of patient dose, and shortest achievable pulse duration.
6.3 Maximum number of pulses for same dose as standard CT (assuming same spectral filtering)
Let us now compare high-voltage switching to an acquisition in standard CT mode, where X-rays are not pulsed, but
applied continuously. In such a continuous mode, the charge Q applied to the tube during a time t with a tube current I is
Q I t (4)
At a typical minimum tube current of 10 mA [10], and a typical fastest rotation speed of 0.35 s per rotation [10], the
charge per full rotation in continuous mode is 3.5 mAs. For Radiography/Fluoroscopy (R/F) systems, X-ray generators
with significantly lower minimum tube current are used, down to 0.2 mA [24]. If such generators were used in CT
systems, the minimum charge per full rotation would be even lower than 3.5 mAs.
We now want to use compressed sensing to lower detector noise while maintaining patient dose, in a scheme as shown in
Fig. 1. The baseline minimum charge per rotation without pulsing is 3.5 mAs. This means that in pulsed mode with a
minimum pulse charge of 0.1mAs, we can have a maximum of 35 pulses (solid blue line in Fig. 8). This may result in
severe under-sampling artifacts in the reconstruction, for which with a continuous exposure about 1000 views are
available.
It would therefore be desirable to lower the minimum charge per pulse by an order of magnitude, so that we can have up
to a few hundred pulses per rotation, while reducing or at least maintaining the overall charge of 3.5 mAs per rotation.
While the minimum pulse charge of 0.1 mAs may be too high for efficient dose reduction with compressed sensing with
today’s scanners, the addition of strong spectral filtering could significantly lower the patient dose at same minimum
pulse charge. An additional benefit of strong spectral filtering would be that the lower-energy X-rays during high-voltage
switching transitions would be filtered out even more efficiently, thereby increasing the dose-efficiency of high-voltage
switched pulses. The new dose limit with an assumed dose reduction by a factor of 10 through spectral filtering is shown
as blue dashed line in Fig. 8.
Existing CT systems already have a number of different spectral filters, and adding or replacing one of them with a
stronger filter is technically feasible. This relatively small hardware modification would make high-voltage switching a
good option for pulsing at low additional cost.
Only the first row in Table 3 satisfies our initial goal of a discharge duration of 100 µs, which had been set to make sure
that transition on/off times are short with respect to pulse durations for a one second gantry rotation speed. The high tube
current of 1000 mA associated with this short transition time shows us once more that strong spectral filtering will
generally be a prerequisite for making high-voltage switching a viable option for dose reduction.
Taking into account the additional time required for the tube voltage rise duration of approximately 0.2 ms (Fig. 7), a
reasonable trade-off may be found at an intermediate tube current of about a hundred mA (Fig. 8).
ECE
Figure 9: Electron gun structure designed at GE Global Research for Inverse Geometry CT [19]. Emitter = emitter/cathode,
Extraction grid = mesh grid, ECE = emittance compensating electrode (focusing electrode 1), Focusing plate = focusing
electrode 2. Simulation of the electron beam optics/focusing is illustrated here.
Focusing plate
ECE Extraction Emitter
grid
Figure 10: Electron gun built at GE Global Research for Inverse Geometry CT – the structure of the gun is similar with
Fig. 8. Emitter here is a dispenser cathode. Each electron gun has four different emitters (see the left hand side image). Each
individual beam line has the extraction and focusing scheme illustrated in Fig. 8.
200mA
ti
1µs
Figure 11: Example of tube pulsing using one cathode of the multi-source space charge limited electron gun demonstrated
for Inverse Geometry CT at GE Global Research. Green line is the ECE voltage (constant 6 kV, used for focusing). Blue
line is the pulsed extraction grid voltage: +266 V for extraction during a 3 µs pulse width, and -557 V outside the pulse to
suppress the DCE emission. Magenta trace is the current emitted by the dispenser cathode emitter (DCE): 953 mA during
the pulse.9
8. CONCLUSIONS
Clinically available CT systems deliver a minimum charge per pulse on the order of 0.1 mAs, which may be too high to
be useful in low-dose pulsing schemes when high-voltage switching is used. In vascular systems, much smaller pulses of
0.002 mAs are currently available, with added grid and corresponding power supply hardware. While this additional
gridding hardware could technically be implemented in a CT scanner, it is not included in clinically available CT scanner
models, as up to now there has been no clinical demand for such small pulses.
Alternative options for low-dose CT acquisitions with little or no hardware modifications of clinically available CT
systems include the addition of strong spectral filtering in order to lower patient dose at the minimum pulse charge
available, and continuous non-pulsed acquisitions at significantly reduced tube current.
In the absence of novel CT tube architectures that include gridding or space charge electron guns with extraction grids, a
high-voltage switching approach may be employed, with a combination of gantry speed optimization, spectral filtration
and faster pulsing at medium tube current. For example, a protocol with 1 s gantry rotation, 1 kHz sampling rate,
200 mA (discharge time of 0.5 ms), 100 active views and 900 inactive views may be used (Table 3). This paper focused
on the feasibility of tube pulsing technologies, and not the determination of the appropriate or optimal imaging protocol.
Such an analysis is the subject of future research.
ACKNOWLEDGEMENTS
Research reported in this publication was supported by Grant Number 1R01CA160253 from the National Cancer
Institute. The content is solely the responsibility of the authors and does not necessarily represent the official views of
the National Cancer Institute or the National Institutes of Health.
[1] Weber, W. A., “Assessing tumor response to therapy,” Journal of Nuclear Medicine 50, Suppl 1, 1S-10S (2009).
[2] Weber, W. A., “Positron emission tomography as an imaging biomarker,” Journal of Clinical Oncology 24, 3282-92
(2006).
[3] Kinahan, P. E., Townsend, D. W., Beyer, T. and Sashin, D., “Attenuation correction for a combined 3D PET/CT
scanner,” Medical Physics 25, 2046-53 (1998).
[4] Xia, T., Alessio, A. M., De Man, B., Manjeshwar, R., Asma, E., Kinahan, P. E., “Ultra-Low Dose CT Attenuation
Correction for PET/CT,” Physics in Medicine and Biology 57, 309-328 (2012).
[5] Colsher, J. G., Jiang, H., Thibault, J.-B., Lonn, A., Pan, T., Lokitz, S. J. and Turkington, T. G., “Ultra low dose CT
for attenuation correction in PET/CT,” Nuclear Science Symposium Conference Record, NSS '08, IEEE, 5506-5011
(2008). http://ieeexplore.ieee.org/stamp/stamp.jsp?arnumber=04774499
[6] Xia, T., Alessio, A. and Kinahan, P. E., “Spectral shaping for ultra-low dose CT attenuation correction in PET/CT,”
The First International Meeting on Image Formation in X-Ray Computed Tomography, Salt Lake City, Utah, 105-
108 (2010).
[7] Ulzheimer, S.. et al., “Multislice CT: Current Technology and Future Developments”, Springer, ISBN 978-3-540-
33124-7 (2009). http://www.springer.com/978-3-540-33124-7
[8] American Association of Physicists in Medicine, “Functionality and Operation of Fluoroscopic Automatic
Brightness Control/Automatic Dose Rate Control Logic in Modern Cardiovascular and Interventional Angiography
Systems”, AAPM Report No. 125 (2012). http://149.28.118.16/pubs/reports/RPT_125.pdf
[9] Belanger, B., Boudry, J., “Management of pediatric radiation dose using GE fluoroscopic equipment,” Pediatric
Radiology 36, Suppl 2, 204–211 (2006).
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2663641/pdf/247_2006_Article_228.pdf
[10] NHS Purchasing and Supply Agency, “Comparative specifications, 64 slice CT scanners”, CEP08027 (2009).
http://nhscep.useconnect.co.uk/ShowDocument.ashx?id=146&i=true
[11] Jørgensen J. S., Sidky, E. Y., Pan, X, “Quantifying admissible undersampling for sparsity-exploiting iterative image
reconstruction in X-ray CT,” IEEE Transactions on Medical Imaging 32(2), 460-73 (2013).
[12] Chandra, N., “CT Sampling Technology. Discovery CT750HD,” A GE Healthcare CT publication, Spring 2009,
http://www.gehealthcare.com/usen/ct/products/docs/ct_sampling_technology09.pdf (2009).
[13] GE Healthcare Website, “Discovery CT750 HD FREEdom Edition,”
http://www3.gehealthcare.com/en/Products/Categories/Computed_Tomography/Discovery_CT750_HD#tabs/tab95
D7BCF2EB2F481F9748C1B0B247AC92 (16 December 2013).
[14] Pemler, P. et al., “Influence of respiration-induced organ motion on dose distributions in treatments using enhanced
dynamic wedges,” Med. Phys. 28, 2234 (2001). http://dx.doi.org/10.1118/1.1410121
[15] General Electric Company, “AutomA / SmartmA Theory, TiP Training in Partnership,” "08-10495 01-08",
http://www3.gehealthcare.co.uk/en-GB/Education/Offering_by_Product/~/media/Downloads/us/Education/Product-
Education-Clinical/TiP-App-Library/GEHealthcare-Education-TiP-App-Library_CT-AutomA-SmartmA-Theory.pdf
(16 December 2013).
[16] Beland, R., “Discharge module for X-ray cable,” Patent, US 5056125 A (1991).
[17] Lu, Z., Nickoloff, E., Ruzal-Shapiro, C., So, J. and Dutta, A., “New automated fluoroscopic systems for pediatric
applications,” Journal of Applied Clinical Medical Physics 6(4), 88–105 (2005).
[18] GE Healthcare, “InnovaTM 2100-IQ, 3100-IQ, 4100-IQ Cardiovascular Imaging System Operator Manual,”
Direction 5391979-5-399 (EN), Revision 1 (2011).
[19] Uribe, J. et al., “Multisource inverse-geometry CT — Prototype system integration,” Nuclear Science Symposium
Conference Record (NSS/MIC), 2578 – 2581 (2010).
[20] McClelland, I., “X-ray equipment maintenance and repairs workbook for radiographers & radiological
technologists,” Appendix E: X-ray equipment operation, “Fig. E-21b”, World Health Organization (2004).
http://libdoc.who.int/publications/2004/9241591633_part5.pdf
[21] Maher, K.., “Basic Physics of Digital Radiography/The Source,” Fig. 2.16, Wikibook,
http://en.wikibooks.org/wiki/Basic_Physics_of_Digital_Radiography/The_Source#Automatic_Exposure_Control
(16 December 2013).
[22] Kinahan, P., “Enabling Ultra Low Dose PET/CT Imaging,” Grant application for FOA PAR-10-169 "Academic-
Industrial Partnerships for Translation of in vivo Imaging Systems for Cancer Investigations (R01)", University of
Washington (2010).
DownloadedViewFrom:
publicationhttp://spiedigitallibrary.org/
stats on 07/18/2015 Terms of Use: http://spiedl.org/terms