Professional Documents
Culture Documents
net/publication/253516749
Characterization of a novel photon counting detector for clinical CT: Count rate,
energy resolution, and noise performance
Article in Proceedings of SPIE - The International Society for Optical Engineering · February 2009
DOI: 10.1117/12.813915
CITATIONS READS
85 1,188
12 authors, including:
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Benjamin M. W. Tsui on 30 March 2015.
ABSTRACT
We report on a characterization study of a multi-row direct-conversion x-ray detector used to generate the first photon
counting clinical x-ray computed tomography (CT) patent images. In order to provide the photon counting detector with
adequate performance for low-dose CT applications, we have designed and fabricated a fast application specific
integrated circuit (ASIC) for data readout from the pixellated CdTe detectors that comprise the photon counting detector.
The cadmium telluride (CdTe) detector has 512 pixels with a 1 mm pitch and is vertically integrated with the ASIC
readout so it can be tiled in two dimensions similar to those that are tiled in an arc found in 32-row multi-slice CT
systems. We have measured several important detector parameters including the maximum output count rate, energy
resolution, and noise performance. Additionally the relationship between the output and input rate has been found to fit a
non-paralyzable detector model with a dead time of 160 nsec. A maximum output rate of 6 × 106 counts per second per
pixel has been obtained with a low output x-ray tube for CT operated between 0.01 mA and 6 mA at 140 keV and
different source-to-detector distances. All detector noise counts are less that 20 keV which is sufficiently low for clinical
CT. The energy resolution measured with the 60 keV photons from a 241Am source is ~12%. In conclusion, our results
demonstrate the potential for the application of the CdTe based photon counting detector to clinical CT systems. Our
future plans include further performance improvement by incorporating drift structures to each detector pixel.
Keywords: CdTe, ASIC, CT, photon counting
*william.barber@dxray.com; phone 1 818 280-0177; fax 1 818 709-7036; wbarber.webs.com
1. INTRODUCTION
The soft tissue contrast and spatial resolution of images which can be produced by clinical X-ray computed tomography
(CT) scanners have enabled CT to become one of the most widespread modalities for diagnostic imaging. Although used
clinically for several decades, CT continues to improve performance as new technologies are incorporated and so effort
to improve CT detectors continues to be an active area of research. Current CT scanners provide three-dimensional
images that accurately delineate tissues and help physicians make accurate diagnoses. However, there are two major
limitations in current CT technologies. These limitations are; large doses of radiation to patients, and insufficient
contrast resolution to discriminate tissues. Both these limitations are addressed by the development of a photon counting
CT technology. For example, in order to improve contrast at reduced dose, photon counting detectors have been
considered to develop alternatives to conventional x-ray detectors used in radiography and CT1,2,3. Photon counting x-ray
detectors applied to mammography and radiography have shown a dose reduction of 40% to 400% while maintaining
sufficient contrast for these applications4,5. Simulations have shown that photon counting along with optimal energy
weighting can increase dose efficiency up to 40% for digital mammography as compared to a conventional integrating
system6,7. One group has reported that the lesion detection signal to noise ratio is significantly affected by the energy
dependence of the detector’s quantum efficiency and the varying contrast carried by different energies in the beam
spectrum8. Recent studies have demonstrated that weighting photons by a factor of E-3 can lead to an improved SNR
compared to simple photon counting or energy weighting (as in intensity-integrating or current-mode detectors). In
another simulation study it was demonstrated that a photon counting detector can reduce beam hardening and improve
Medical Imaging 2009: Physics of Medical Imaging, edited by Ehsan Samei, Jiang Hsieh,
Proc. of SPIE Vol. 7258, 725824 · © 2009 SPIE
CCC code: 1605-7422/09/$18 · doi: 10.1117/12.813915
2. METHODOLOGY
Single photon counting detectors with energy discrimination capabilities have been developed for nuclear medicine
application such as positron emission tomography (PET) and single photon emission computed tomography
(SPECT)10,11. If photon counting detectors are applicable for x-ray CT imaging, they have the potential to address both of
the two major problems inherent in current mode CT imaging, namely, dose and contrast. Photon counting detectors will
not add noise; they count each and individual photon without weight; and they can maximize the contrast of tissues of
interest by retrospectively optimizing weights to the energy of interest12. In an effort to overcome the limits imposed by
current mode detectors, a number of research groups in hospitals, universities and commercial companies have tried to
develop photon counting systems for x-ray imaging applications13,14,15.
2.1 CdTe detectors
As mentioned previously, current CT systems typically use a detector comprised of Si photodiodes that are optically
coupled to a scintillator. In these detectors, the energy of an absorbed x-ray is converted to electrical charge by an
indirect method. The development our proposed photon counting CT detector makes use of direct conversion of the
energy deposited by each x-ray photon to charge and on counting each photon provided by the parallel pulse shaping
electronics for each pixel. Direct conversion with photon counting offers numerous advantages applicable in various x-
ray imaging applications. Single crystals of CdTe can effectively absorb the incoming x-ray radiation and convert it
directly into electrical signals, which can be read by an ASIC16. An absorber with bulk resistivity larger than 108 Ohm-
cm will minimize leakage current and inherent noise but can produce a large measurable signal. Thus the overall signal
to noise ratios can be significantly better than indirect detectors. Moreover, the efficiency of converting the x-ray signal
to an electrical signal (detector gain) can be an order of magnitude larger due to the basic underlying physics of the
energy transfer process in the direct detection approach. That is, the mean energy for creating an electron-hole pair in a
semiconductor detector (4.43 electron Volts per electron hole pair in CdTe) is typically an order of magnitude smaller
than the corresponding energy necessary to create an optical photon through the scintillation approach. This results
directly in an order of magnitude larger signal for the same incident X-ray with a direct detector approach. Another
important consideration is that charges generated by x-rays do not spread laterally (aside from the negligible diffusion),
but move along the applied electric field lines. Spreading of light in the indirect conversion scintillator approach that can
deteriorate spatial resolution can be compensated for by segmenting the scintillator. It also allows the construction of
thicker, more efficient detectors without any loss in resolution due to lateral spread in the detector.
Fig. 1. Pictures of a completed 32 row CT detector with a high voltage cable attached. The left side shows a sagittal view
with 32 rows and the right side shows an axial view with 16 pixels per row.
U)
U)
U)
C
(-)
U)
'I0
C
0
C)
1000
1000
Fig. 2. S-curves for various input flux from an X-ray tube for clinical CT. The left side shows a graph of the number of
counts above the threshold as a function of threshold (S-curve) for an input flux from the X-ray tube set at 65 kVp and
1mA. The right side shows S-curve for several max kVp setting on the X-ray tube. Data has been taken from 30 kVp to
140 kVp which covers the range of X-rays used in clinical CT.
1000
900
800
700
600
°-
300
200
100
0
0 20 40 60 80 100 120 140 160
max kVp (key)
Fig. 3. A graph of pulse height (preamplifier signal) versus energy (max kVp) for various X-ray tube setting from 30 keV to
140 keV. An extrapolation of the linear portion of the energy response (up to 60 keV) to low energy shows the noise
floor at 20 keV (250 mV pulse height above which no noise counts are seen).
6000 - 50000 -
S
a . S 00
a
S
0.
U
5000 -
. 40000
S
00
4000 30000 -
00
C 20000
g 3000
0
0
2000 10000-
1000
0 I 2 3 4 5
1 2 3 4 5
X-ray tube current (mA)
X-ray tube current (mA)
Fig. 4. Graphs of the output count rate (OCR) versus input count rate (ICR) as determined by acquiring data from 0.01 mA
the 6 mA with a clinical CT X-ray generator. pulse height (preamplifier signal) versus energy (max kVp) for various
X-ray tube setting from 30 keV to 140 keV.
600
300 -
600
200-
a0 400
100 - I
S 200
0- 0
0 200 400 600 600 1 GOC 0 200 400 600 800 100c
pulse height (mV) pulse heght mV)
Fig. 5. Graphs of differentiated un-calibrated S-curves for a 10 µCi 241Am source (left side) and a 6 mCi 57Co source (right
side).
800 -
300
600 -
FWHMr7keV Ce
200 =
0) FWHM 35 keV
C 11.6% @60 keV 400 -
C-,
100 200 -
\ 29.1%@l2OkeV
S
0
0 20 40 60 80 100 120 140 0 50 ff0 50 200 250 300
241 57
Fig. 6. Graphs of differentiated calibrated S-curves for a 10 µCi Am source (left side) and a 6 mCi Co source (right
side). Notice the noise increases rapidly below 20 keV.
The relatively poorer energy resolution at 120 keV as compared to 60 keV is due to a loss of resolution above 120 keV
as can be seem in the distortion of the peak above 120 keV on the right side of Figure 6. This distortion occurs as the
preamplifiers become saturated above 120 keV (see Figure 3). The degradation of energy resolution above 120 keV does
not limit performance for the dual-energy CT applications used clinically which separate the low and high energy images
below 100 keV.
For use in clinical CT, the detector system counts the x-ray photons transmitted through the patient and sorts the counts
into two energy bins. The two bins of count data are processed to produce two sets of raw data (high and low energy),
that are input to the CT reconstruction algorithm. The raw data sets are processed with a dual energy calibration
technique that enables projection space reconstruction and eliminates the beam hardening effect. The image
reconstruction generates mono-energetic equivalent CT images (with a user-selected energy) and “virtual-un-enhanced”
images where the iodine-based contrast media is identified and removed in the reconstruction process. The first clinical
trial with the prototype system was a prospective study on patients with known carotid artery disease. The study was
performed at the Rabin Medical Center in Israel and was reviewed by the hospital IRB committee. The clinical images
demonstrated the vascular and stenotic elements with good image quality, which is particularly notable in regard to the
low x-ray tube current technique used for these scans. These results demonstrated the potential use of energy information
to drastically improve clinical CT.
4. CONCLUSIONS
We have fabricated and tested novel detector array structures utilizing CdTe that have been specifically developed for
CT applications. The development of this CT detector has produced several important technological advancements.
These include the fabrication of CdTe detectors with an active area extending to the edge of the crystals (no guard rings)
which allows tiling with almost no dead space. Additionally we have developed packaging and encapsulation methods to
connect dense multi channel ASICs to the crystals and formed within the active area of the crystal to preserve tiling in
two dimensions. These detectors provide a signal response much faster than the transit time of carriers over the whole
detector thickness. At the same time the individual detector pixels exhibit good spectral performance at very short
peaking times. We have developed fast and low-noise ASIC readout electronics that are optimized for amplification and
processing of signals from CdTe detector arrays. Detector modules containing crystals and ASIC electronics were
specifically designed to fit into a CT scanner and replace conventional integrating scintillator detectors with fast and
highly efficient photon-counting and energy-dispersive detectors. A maximum output count rate of 6x106 cps per 1 mm2
ACKNOLEDGEMENTS
We would like to thank the National Institutes of Health for funding this research with the following grants numbered; 1
R43 EB008612, and 4 R44 EB008612. We would also like to thank the Norwegian Research Council for funding this
research with grant number BIA-180079/I40. Additionally, we thank the European Organization for Nuclear Research
(CERN) for collaborating on the development of DxRay’s fast photon counting ASIC.
REFERENCES
1
P.C. Johns, J. Dubeau, D.G. Gobbi, M.Li, and S. Dixit, Photon-Counting Detectors for Digital Radiography and X-Ray
Computed Tomography, SPIE, TD01, 367-369, (2002).
2
P.M. Shikhaliev, X. Tong, S. Molloi, Photon computed tomography: Concept and initial results, Med. Phys., 32(2),
427-436, (2005).
3
A. Maidment, C. Ullberg, K. Lindman, L. Adelow, J. Egerstrom, M. Eklund, T. Francke, U Jordung, T. Kristoffersson,
L. Lindqvist, D. Marchal, H. Olla, E. Penton, J. Rantanen, S. Solokov, N. Weber, H. Westerberg, Evaluation of a
photon-counting breast tomosynthesis imaging system, Medical Imaging 2005, Proceedings of the SPIE, 5745, 572-582
(2005).
4
M. Lundqvist, B. Cederstrom, V. Chmill, M. Danielsson, and B. Hasegawa, Evaluation of a photon-counting x-ray
imaging system, IEEE Trans. Nucl. Sci., 48(4), 1530-1536, (2001).
5
M. Chmeissani, C. Frojdh, First Experimental Tests with a CdTe Photon Counting Pixel Detector Hybridized with a
Medipix2 Readout Chip, IEEE Trans. Nucl. Sci., 51(5), 2379-2385 (2004).
6
M. Danielsson, and H. Bornefalk, Dose-efficient system or digiltal mammography, IEEE Trans. Nucl. Sci., 48(4),
1530-1536, (2001).
7
M. Lundqvist, B. Cederstrom, V. Chmill, M. Danielsson, and B. Hasegawa, Evaluation of a photon-counting x-ray
imaging system, IEEE Trans. Nucl. Sci., 48(4), 1530-1536, (2001).
8
Tapiovaara MJ, Wagner R., SNR and DQE analysis of broad spectrum X-ray imaging. Physics in Medicine and
Biology. 30(6):519-29, (1985).
9
Shikhaliev PM., Beam hardening artefacts in computed tomography with photon counting, charge integrating and
energy weighting detectors: a simulation study., Physics in Medicine and Biology. 50(24):5813-27, (2005).
16
W.C. Barber, K. Iwata, B.H. Hasegawa, P.R. Bennett, L.J. Cirignano, and K.S. Shah, Current mode operation of a
CdZnTe for CT imaging, Penetrating Radiation Systems and Applications IV, SPIE, 4786, 144-150, (2002).
17
A. Arodzero, W.C. Barber, M.Q. Damron, N.E. Hartsough, J.S. Iwanczyk N. Malakhov, E. Nygard, D. Moraes, P.
Weilhammer and P. Jarron, A System for the Characterization and Testing of CdZnTe/CdTe Pixel Detectors for X-ray
and Gamma-ray Imaging, IEEE Nucl. Sci. Symp. Conference Record, 3638-3642, (2006).
18
W.C. Barber, N. Malakhov, M.Q. Damron, N.E. Hartsough, D. Moraes, P, Weilhammer, E. Nygard and J.S. Iwanczyk,
Guard ring elimination in CdTe and CdZnTe detectors, IEEE Nucl. Sci. Symp. Conference Record, 2414-2416, (2006).
19
J. S. Iwanczyk, E. Nygard, O. Meirav, J. Arenson, W. C. Barber, N. E. Hartsough, N. Malakhov, and W. J. C., Photon
counting energy disparsive detector arrays for x-ray imaging, in IEEE Nuclear Science Symposium and Medical
Imaging Conference, 1 ed. vol. M09-04 Honolulu, HI: IEEE, 2741-2748, (2007).
20
D. Moraes, J. Kaplon, E. Nygard, CERN_DxCTA counting mode chip. NIM A 591, 167-170, (2008).
21
D. Moraes, E. Nygard and A. Rudge, Front-end electronics for the readout of CdZnTe sensors, Nuclear Science
Symposium Conference Record, IEEE, 1, 348-352, (2005).
22
J. Watt , D.W. Davidson , C. Johnston , C. Smith , L. Tlustos , B. Mikulec , K.M. Smith, and M. Rahman, Dose
reductions in dental X-ray imaging using Medipix. NIM, A 513, 65-69, (2003).
23
S. Kheddache, R. Kullenberg and E. Kivio-Carlsson, Dose reduction in pevimetry using a digital technique. Radiation
Protection Dosimetry, 80(1-3), 275-278, (1998).