You are on page 1of 5

IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, VOL. 3, NO.

1, FEBRUARY 2021 289

Design of a Low-Cost Miniature Robot to Assist the


COVID-19 Nasopharyngeal Swab Sampling
Shuangyi Wang , Kehao Wang , Ruijie Tang, Jingzhe Qiao,
Hongbin Liu , Member, IEEE, and Zeng-Guang Hou , Fellow, IEEE

Abstract—Nasopharyngeal (NP) swab sampling is an effective approach


for the diagnosis of coronavirus disease 2019 (COVID-19). Medical staffs
carrying out the task of collecting NP specimens are in close contact
with the suspected patient, thereby posing a high risk of cross-infection.
We propose a low-cost miniature robot that can be easily assembled and
remotely controlled. The system includes an active end-effector, a passive
positioning arm, and a detachable swab gripper with integrated force
sensing capability. The cost of the materials for building this robot is
55 USD and the total weight of the functional part is 0.23kg. The design
of the force sensing swab gripper was justified using Finite Element (FE)
modeling and the performances of the robot were validated with a sim-
ulation phantom and three pig noses. FE analysis indicated a 0.5mm
magnitude displacement of the gripper’s sensing beam, which meets the
ideal detecting range of the optoelectronic sensor. Studies on both the Fig. 1. Diagram showing the nasopharyngeal (NP) swab collection procedure.
phantom and the pig nose demonstrated the successful operation of the
robot during the collection task. The average forces were found to be
0.35N and 0.68N, respectively. It is concluded that the proposed robot is swab (also known as deep nasal swab) and anterior nares (nasal
promising and could be further developed to be used in vivo. swab) specimen, could be supervised onsite self-collection and home
Index Terms—COVID-19 diagnostic robot, nasopharyngeal swab robot,
self-collection [5]. During the conventional manually controlled swab
robot to combat COVID-19, interventional robot. sampling, medical staffs are inevitably in close contact with the sus-
pected patient, posing a high risk of cross-infection. The operating
skills and psychological states of medical workers may also affect
I. I NTRODUCTION the accuracy and quality of the results of swab collection.
The global outbreak of novel coronavirus pneumonia (NCP) As the global fight against COVID-19 may last for a long period of
caused by coronavirus disease 2019 (COVID-19) has spread rapidly. time with tens of thousands nasal swab samplings performed world-
Collection of specimens from the surface of the respiratory wide each day, robotic-assisted NP and OP swabbing with remote
mucosa with nasopharyngeal (NP) or Oropharyngeal (OP) swabs are operation capability may reduce the risk of infection and at the mean
treated as effective ways for the diagnosis and screening [1]. Several time free up staffs for other tasks. Comparing with human, the robot
recent studies have indicated that OP swabs are less effective than can be more thoroughly disinfected and those parts that are in close
NP swabs in detecting the COVID-19 virus [2], [3] and concluded contact with patients can be disassembled and replaced. The value of
that the use of NP may be more suitable, although a study also high- using robots in fighting against COVID-19 has been highlighted in [6]
lighted the data should be viewed with cautions [4]. According to US and a well-designed robot for OP collection was recently developed
CDC, both NP and OP should be performed by a healthcare profes- by the Guangzhou Institute of Respiratory Health and the Shenyang
sional. Other possible approaches, e.g., nasal mid-turbinate (NMT) Institute of Automation under the Chinese Academy of Sciences [7].
The robot consists of a snaking-shape robot arm for motion control
Manuscript received August 19, 2020; revised October 10, 2020; accepted and a binocular endoscope for visualization. The system has been
November 3, 2020. Date of publication November 6, 2020; date of current
version February 22, 2021. This article was recommended for publica- successfully tested in vivo and proved effective in clinical trials.
tion by Associate Editor K. Harada and Editor P. Dario upon evaluation In the present study, we aim to develop a low-cost, easy-to-assemble
of the reviewers’ comments. This work was supported in part by the robot with small footprint to assist NP swab collections. The NP swab
Institute of Automation, Chinese Academy of Sciences, and in part by the collection involves inserting a specifically manufactured swab into
National Natural Science Foundation of China under Grant 62003339 and
the patient’s nasal cavity [8]. The head of the patient is expected to
Grant 82000878. (Shuangyi Wang and Kehao Wang are co-first authors.)
(Corresponding author: Shuangyi Wang.) be tilted back (at approximate 70◦ ) so the nasal passage becomes
Shuangyi Wang, Ruijie Tang, and Zeng-Guang Hou are with the State Key straight and accessible. As illustrated in Figure 1, the swab is inserted
Laboratory of Management and Control for Complex Systems, Institute of through the nostril parallel to the palate, and to the nasopharynx
Automation, Chinese Academy of Sciences, Beijing 100190, China (e-mail: until resistance forces are identified. The swab is then left in place
shuangyi.wang@ia.ac.cn; ruijie.tang@ia.ac.cn; zengguang.hou@ia.ac.cn).
Kehao Wang is with the Beijing Advanced Innovation Center for several seconds for secretion absorptions and been rotationally
for Biomedical Engineering, Key Laboratory for Biomechanics and retracted from the nasal cavity slowly. In this article, we aim to present
Mechanobiology of Ministry of Education, School of Biological Science and the design concept, the implementation method, the simulation study,
Medical Engineering, Beihang University, Beijing 100191, China (e-mail: and the preliminary phantom and animal tests of the robot.
kehaowang@buaa.edu.cn).
Jingzhe Qiao is with the School of Mechanical Engineering and
II. D ESIGN AND I MPLEMENTATION OF THE ROBOT
Automation, Northeastern University, Shenyang 110819, China (e-mail:
qiaojz@yeah.net). A. Design Concepts
Hongbin Liu is with the School of Biomedical Engineering and Imaging
Sciences, King’s College London, London WC2R 2LS, U.K. (e-mail:
The proposed robot includes an active 2-degree of freedom (DOF)
hongbin.liu@kcl.ac.uk). end-effector (Figure 2) for actuating the swab and a generic 6-DOF
Digital Object Identifier 10.1109/TMRB.2020.3036461 passive arm for the global positioning. Within the supporting case,

c IEEE 2020. This article is free to access and download, along with rights for full text and data mining, re-use and analysis.

Authorized licensed use limited to: cao anh. Downloaded on December 14,2021 at 09:53:57 UTC from IEEE Xplore. Restrictions apply.
290 IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, VOL. 3, NO. 1, FEBRUARY 2021

Fig. 4. Implementation of the proposed robot with its passive arm, active
end-effector, add-on headrest, and the embedded electronics shown. The photo
was taken only for illustrational purpose to show the expected clinical setup
of the NP swab collection procedure.

used in this study. The selected sensor has been proved in our previous
Fig. 2. Schematic illustration of the proposed 2-DOF active end-effector studies to be able to achieve large output voltage variations without
with each component and the mechanism of the force sensing swab gripper using an amplifier and demonstrates low level of noises [9], [10]. The
shown.
overall size of the swab gripper is 38(L) × 33(W) × 5(D) mm.

B. Hardware and Software Implementation


In this research, we employed a 6-DOF passive arm for global posi-
tioning and insertion angle adjustments. The passive arm has a global
revolute joint (R1), two parallel mechanisms to adjust backwards-
forwards and up-down movements (R2 and R3), and a tilting joint
(R4) working together with a ball joint (R5 & R6) to fine-tune
the angulation of the end-effector, i.e., the insertion angle of the
swab. The setup of the 6-DOF passive arm and the 2-DOF active
Fig. 3. Movements of the proposed 2-DOF robotic end-effector during the end-effector is shown in Figure 4.
procedure of NP swab collection. The setup is to demonstrate the expected clinical scenery for the NP
swab collection procedure when using the robot. The passive arm is
clamped to the table with the active end-effector attached. An inertial
a leadscrew driven linear stage actuated by a stepper motor was
measurement unit (IMU, DT-06, Shenzhen Doctors of Intelligence &
mounted. A geared stepper motor is attached to the front end of
Technology Co., Ltd., Shenzhen, China) could be embedded into the
the linear stage to control the following rotation link. As illustrated
end-effector to measure the tilting angle. The add-on headrest has
in Figure 2, a specially designed swab gripper is attached to the rota-
a tilting joint that can be manually adjusted. The combined used
tion link with its extruded structure tightly fitted into the grove of the
of the headrest and the IMU sensor would help to align the angle
rotation link. The swab can be assembled to the swab gripper with
of the patient’s head and the initial insertion angle of the robot, at
its shaft constrained by the constraint hole and held by the gripping
approximate 70◦ .
cylinder of the swab gripper.
Additionally, a microcontroller (MCU) WeMos D1 mini with
The proposed individual structures allow easy attaching and
its power shield (core chip based on ESP8266, Espressif Systems,
detaching operations between the swab and the swab gripper that
Shanghai, China) and two stepper motor drivers (DRV8825, Texas
can be performed by either the patient or the on-site health
Instruments, Dallas, US) were utilized in the design. The electron-
professional, making the collection conveniently and both compo-
ics setup is shown in Figure 4. The whole system is powered by
nents recyclable. The overall size of the 2-DOF end-effector is
a regular 9V-3A power supply which can be directly plug into the
150(L) × 60(W) × 40(D) mm. The movements of the 2-DOF robotic
power shield. With the WiFi function of the MCU, we developed
end-effector is shown in Figure 3 with the motions of the prismatic
a mobile phone-based simple user interface (UI) using the Blynk
and the rotation joints demonstrated.
software (Blynk ver. 2.27.17, Android app) to remotely control the
The structure of the swab gripper was designed to also work as
stepper motors and display the sensor’s reading. Joysticks, buttons,
a single-axis force sensor based on the principle of compliant mecha-
and level displays were included into the UI with their functions
nism (CM). The compression force experienced by the swab when it
illustrated in Figure 5. Specifically, with the 2-axis joystick actuated
is in contact with the nasal cavity would result in the largest vertical
diagonally, the swab can be translated and rotated simultaneously, in
displacement segment of the structure at the center of the deflection
both directions of each axis to meet the clinical needs as explained
beam. As shown in Figure 2, the displacement is detected by an opto-
in Figure 1.
electronic sensor (QRE1113, Fairchild Semiconductor, California, the
The proposed active end-effector was mainly made from 3D print-
United States) which consists of a LED to generate infrared (IR) light
ing using Polylactic (PLA) acid and the swab gripper was 3D printed
and a photo-transistor to receive the reflected light. The measurement
using Polypropylene (PP) material. PLA is the widely used rigid
circuit is embedded into a breakout board (SparkFun Electronics,
printing material while PP is a semi-rigid flexible material with excel-
Colorado, the United States) to detect the output voltage, which varies
lent fatigue resistance. The manufacturing cost of the system is listed
depending on the amount of IR light reflected to the sensor with
in the Table I, with a total cost price of 400 RMB (∼55 USD).
changes of the distance between the sensor and the deflection beam.
Two parts of the characteristic curve (as illustrated in Figure 2) show
linear relationships between the measured distance and the output C. Safety Features
current. To guarantee a high linearity and sensitivity, the yellow 1) Motor Current Limiting and the Maximum Payload: The
area (measured distance in between 0 to 0.50 mm) in Figure 2 was DRV8825 stepper motor driver used in this study provides an active

Authorized licensed use limited to: cao anh. Downloaded on December 14,2021 at 09:53:57 UTC from IEEE Xplore. Restrictions apply.
IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, VOL. 3, NO. 1, FEBRUARY 2021 291

Fig. 5. Screenshot of the robot control software developed in Blynk.

TABLE I
C OST P RICE S UMMARY OF THE P ROPOSED ROBOT
Fig. 7. FE modeling of the swab gripper when subjected to external
forces: (a) the mesh results with the force, contacts, and the boundary
conditions shown, (b) the displacement results under 2.50N, and (c) the
force-displacement relationship with the forces applied from 0 to 2.50N.

noses (test facilities shown in Figure 9). The maximum force found
during the simulation procedure was found to be at the magnitude of
2.50N. Based on the manual experimental results, the maximum pay-
load of the robot was set to 3.00N with some additional allowances.
The maximum current setting, in real use, can be further adjusted
in an easy way once the in-vivo experiments using the manually
controlled swab holder are conducted.
2) Software Settings: The software settings for safety include
a watchdog between the microcontroller and the Blynk software which
continually monitors the connection by checking periodically refreshed
signals and various software logics to ensure that motor actuation
works in conjunction with the force sensors. The WiFi connection
between the robot and the control interface is also monitored by the
Blynk software based on its default built-in function. Communication
failures would be highlighted in the UI to inform the operator.
3) Intrinsic Features: In addition to above mentioned features, the
weight of the 2-DOF end-effector with all the electronics included
is only 0.23kg and the weight of the 6-DOF passive arm is 0.31kg.
This lightweight design intrinsically reduces the risks when using the
robot because the patient can easily move the robot away by pushing
the passive arm.

Fig. 6. (a) Setup for the current-force measurement experiment and the test III. F INITE E LEMENT A NALYSIS OF THE S WAB G RIPPER
results; (b) manually controlled tool for the maximum force measurements. Finite Element (FE) models of the swab gripper and the swab were
developed in ANSYS mechanical APDL (ver. 18.2) using geometric
current limiting method: each full-bridge is controlled by a fixed input of IGES file that was created by SOLIDWORKS (ver. 2018).
off-time PWM current control circuit that limits the load current The model is to analyze the displacement performances of the deflec-
to a desired value. The maximum value of current limiting can be tion beam of the swab gripper when the swab is subjected to external
adjusted by setting the reference pin and the motor’s maximum torque forces. To justify the design, the expected ranges and magnitudes
is proportional to ampere-turns. This feature limits the maximum of the displacements should meet the selected linear range of the
allowable force when the swab is in contact with the patient. Test optoelectronic sensor. In this study, Young’s modulus of 1.70 GPa
results shown in Figure 6(a) were obtained by adjusting the current and Poisson’s ratio of 0.43 were used for the swab gripper based
limiting setting from 0.2A to the maximum allowable value, 2.2A, on the information provided by the 3D printing manufacturer. Half
of the stepper motor driver in the linear axis and measuring the cor- model was considered as the whole configuration is symmetric about
responding maximum forces generated by the linear stage before its central plane. Both parts were meshed using quadratic tetrahe-
the stepper motor slips using a commercial force sensor (JHBM- dral elements (ANSYS element type: SOLID186). Mix u-p element
H3, JNSensor, Anhui, China). A regular daily-used shorter swab formulation method was chosen to avoid volumetric locking phe-
with wooden shaft was employed to avoid the buckling effects. nomenon given that the selected material has a Poisson’s ratio of
The experimental setup is shown in Figure 6(a). To decide the 0.43 which is close to that of an incompressible material (ν = 0.50).
value of the maximum current setting, a separate swab holder was The swab gripper was meshed using a total number of 32859 elements
designed and 3D printed to integrate with the commercial force sen- yielding a total of 55894 nodes and the swab has 4615 elements and
sor (Figure 6(b)). This swab holder allows for the measurement of 10027 nodes (Figure 7(a)). Contacts between the swab and the con-
the maximum force generated during the manually controlled NP straint hole was modelled as frictionless using Augmented Lagrange
swab collection procedure and was tested with a phantom and pig algorithm and the contact between the swab and the deflection beam

Authorized licensed use limited to: cao anh. Downloaded on December 14,2021 at 09:53:57 UTC from IEEE Xplore. Restrictions apply.
292 IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, VOL. 3, NO. 1, FEBRUARY 2021

Fig. 8. Measurement and processed results of the force calibration


experiment.

was considered as fully bonded. The boundary conditions were added Fig. 9. Experimental setups for (a) the phantom and (b) the pig nose tests.
to constrain the movement of the gripper in Y and Z-axes. 2.50N were
applied to the distal end of the swab in seven sub steps as this was
the maximum expected force when the swab is in the nasal cavity.
The swab has a thin long shaft and is subjected to buckling under
increasing load. However, as this only happens in a limited range
and the swab is constrained by the nasal cavity, the deformations
do not cause the collapse and will continue to transfer the load.
In this study, we focused on the load analysis. An example color
coded displacement profile in the vertical direction and the com-
parison between undeformed and deformed shapes under 2.50N are
shown in Figure 7(b), as the results of the FEA. It can be observed
that the maximum displacement occurs at the center point of the
deflection beam, resulting in an approximate 0.50 mm deformation.
The force-displacement curve is shown in Figure 7(c).

IV. P RELIMINARY E XPERIMENTAL R ESULTS


A. Force Calibration
Force calibration was performed using the commercial force sen-
sor. The setup is the same as the previous experiment (Figure 6(a)).
The linear stage was actuated slowly to generate the target contact
forces (from 0 to 2.50N, 0.50N interval) and the according measured
output voltages from the optoelectronic sensor were recorded. The
loading (blue line) and unloading (black line) curves are shown in
Figure 8, where the elastic hysteresis can be observed. As the cur-
Fig. 10. Results of force measurements: (a) time versus force using the
rent application does not require high accuracy when measuring the phantom; (b) time versus force using a pig nose; (c) typical force-displacement
force, the average values of the measured output voltage for gener- curves during the insertion process for the phantom and the pig nose.
ating the same amounts of force during the loading and unloading
phases were used for the second order polynomial curve fitting and In both experiments, the insertion angles of the swab were set with
the relationship (red line) between the force values, F, and the volt- the swab parallel to the palate. The phantom is stably positioned on
age V was: V = − 0.1184F2 + 1.4074F + 0.2588. Using the red the bench while the pig noses required manual supports to remain
line as the calibration result, the error of the measured samples is stable. The robot was actuated from the mobile app with the insertion
0.24 ± 0.16V (mean ± standard deviation). The calibration result first and then the retraction combining with the rotation. For each test,
was programmed into the MCU with running average implemented the insertion-retraction procedure was repeated for six times.
to smooth out the values from the sensor. To further analyze the per- Upon successful completions of both the phantom and the pig nose
formances, the measured voltages were set as inputs and the predicted experiments with the initial insertion axis aligned and the required
forces were calculated based on the calibration results. The predicted motions completed, the measured forces versus time for the phan-
forces were then compared with the real generated forces and the tom and an example measurement for the pig nose are shown in
root-mean-square error (RMSE) was found to be 0.27N. Figure 10(a) and Figure 10(b). By extracting the data of the insertion-
retraction procedures, the average force was found to be 0.35± 0.58N
B. Phantom and Animal Experiments (mean ± standard deviation) and 0.68 ± 0.64N while the maximum
To validate the performance of the robot and understand the force was 2.43 and 2.61N for the phantom and pig noses, respec-
required force for the NP swab sampling procedure, a commercial tively. The maximum identified forces were within the payload of
nasopharynx phantom (Simon Co., Shanghai, China) and three pig the robot and the detectable range of the customized sensor.
noses (from a food supplier, no ethics required) were utilized, as As can be seen from Figure 10, the force tends to follow a similar
shown in Figure 9. The phantom has realistic shapes and anatomical trend in each iteration and reach a peak at the similar magnitude when
structures of the human nose although the silicone material’s prop- tested with the phantom and the pig nose. Variations between the two
erties are different to human tissues. The pig noses were obtained tests are noted, which aroused from differences in anatomical structures
from freshly slaughtered pigs (preserved in an ice bag and delivered of the two test facilities and the initial insertion angles during the
two days later). Although the anatomical structures of pig nose are two experiments. The peaks were aroused from the contacts between
different to human nose and the pig noses were cut from the head the swab and the nasopharynx for the phantom and the contacts
thereby having shorter lengths, the tissue’s mechanical properties are between the swab and turbinates for the pig noses. Additionally,
likely to be close to human. typical force-displacement curves during the swab insertion process

Authorized licensed use limited to: cao anh. Downloaded on December 14,2021 at 09:53:57 UTC from IEEE Xplore. Restrictions apply.
IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, VOL. 3, NO. 1, FEBRUARY 2021 293

TABLE II
I MPORTANT F EATURES OF THE P ROPOSED ROBOT of the robot. Based on the current evidence, it is concluded that the
measured forces were below the payload of the robot, within the
detectable range of the sensing gripper, and has the similar trends
and ranges when tested using different facilities.
The initial insertion angle of the robot can be set to be parallel to
the palate using the passive arm with supports from visual inspection
or IMU sensor, and this ensures the swab go into the lower nasal pas-
sages of the phantom and pig noses although both are static objects.
Considering the possible motions induced by real patients, additional
active DOFs, as replacements of the current passive joints R5 and R6
(pitch and yaw motions) can be added to the robot design to allow
further flexibility, although this will increase complexity and cost of
the robot. Therefore, the necessity for doing this would need to be
further investigated by our future in-vivo tests. The present study was
limited to experiments on simulation phantom and pig noses, although
both cannot fully represent the real nasal cavity of human. Our next
focus will be pursuing ethical approval for in-vivo tests, improving
the current force sensor design, modelling the contact between swab
tip and soft tissue, and evaluating the effectiveness and robustness of
for the phantom and the pig nose are also shown in Fig. 10(c) based the robot via large-scale tests.
on the recorded motor steps, which reflect the swab insertion depths
in the nasal cavities. The motor was actuated with constant speed R EFERENCES
during the procedure.
[1] “Laboratory testing for coronavirus disease 2019 (COVID-19) in
suspected human cases: Interim guidance,” document WHO/COVID-
C. Summary 19/laboratory/2020.4, World Health Org., Geneva, Switzerland,
Mar. 2020.
Based on the design introduced in Section II and the preliminary
[2] X. Wang et al., “Comparison of nasopharyngeal and oropharyngeal
test results analyzed in Sections IV-A and IV-B, the features of the swabs for SARS-CoV-2 detection in 353 patients received tests with
proposed robot were summarized in Table II. both specimens simultaneously,” Int. J. Infect. Dis., vol. 94, pp. 107–109,
May 2020. [Online]. Available: https://doi.org/10.1016/j.ijid.2020.04.023
[3] Y. Yang et al. (Feb. 2020). Evaluating the Accuracy of Different
V. D ISCUSSIONS AND C ONCLUSION Respiratory Specimens in the Laboratory Diagnosis and Monitoring
This article introduces a low-cost miniature robot to assist NP sam- the Viral Shedding of 2019-nCoV Infections. [Online]. Available:
https://doi.org/10.1101/2020.02.11.20021493
pling. The robot can be made easily using 3D printing technique
[4] K. Carver and N. Jones, Comparative Accuracy of Oropharyngeal
together with other easy-to-find commonly used components on the and Nasopharyngeal Swabs for Diagnosis of COVID-19, Centre for
market. The robot is intended to be used in a simple way by assem- Evidence-Based Medicine (CEBM), Nuffield Dept. Primary Care
bling different pieces together and controlling from a mobile app. Health Sci., Univ. Oxford, Oxford, U.K., Mar. 2020. Accessed:
The costs for developing the robot can be further reduced if manu- May 24, 2020. [Online]. Available: https://www.cebm.net/covid-19/
comparative-accuracy-of-oropharyngeal-and-nasopharyngeal-swabs-for-
factured in large numbers (estimated to be less than 30 USD). The diagnosis-of-covid-19/
payload of the robot, when controlling a swab with a long thin shaft, [5] Interim Guidelines for Collecting, Handling, and Testing Clinical
is strictly limited by using small motors. Above the threshold, the Specimens for COVID-19, Dept. Health Human Services, Centers Dis.
motor would slip. To guarantee the safe use, active current limiting, Control Prevent., Atlanta, GA, USA, May 2020.
watchdog with periodical response checking were implemented, in [6] G.-Z. Yang et al., “Combating COVID-19—The role of robotics
in managing public health and infectious diseases,” Sci. Robot.,
addition to the intrinsic features of the robot. vol. 5, no. 40, Mar. 2020, Art. no. eabb5589. [Online]. Available:
The robot also has a detachable swab gripper which incorporates https://doi.org/10.1126/scirobotics.abb5589
the force sensing capability based on a simple CM and a low- [7] Q. Wang, China Develops Intelligent Robots for Throat Swab
cost optoelectronic sensor. This piece can be easily assembled to Sampling of Coronavirus Tests, Global Times, Beijing, China,
Mar. 2020. Accessed: May 24, 2020. [Online]. Available:
and removed from the robot, making it possible for the suspected https://www.globaltimes.cn/content/1182175.shtml
patient to do self-assembly and disassembly pre and post being tested. [8] F. M. Marty, K. Chen, and K. A. Verrill, “How to Obtain a nasopharyn-
The main detect range of 0-2.50N of the sensing gripper was jus- geal swab specimen,” New England J. Med., vol. 382, p. e76, Apr. 2020.
tified using the FE analysis. The calibration result indicated that [Online]. Available: https://doi.org/10.1056/nejmvcm2010260
the displacement range (0–0.50mm) of the deflection beam meets [9] Y. Noh et al., “A 2-piece six-axis force/torque sensor capable of mea-
suring loads applied to tools of complex shapes,” in Proc. IEEE/RSJ
the prediction of the FE model and falls into the expected linear IROS, Macau, China, 2019, pp. 7976–7981. [Online]. Available:
range of the optoelectronic sensor. The sensor’s design concept was https://doi.org/10.1109/IROS40897.2019.8967947
specifically for this low-cost application. Similar concepts of the low- [10] Y. Noh, S. Han, P. Gawenda, W. Li, S. Sareh, and K. Rhode, “A con-
cost 3D printable force sensor using optoelectronic sensors were also tact force sensor based on s-shaped beams and optoelectronic sensors
for flexible manipulators for minimally invasive surgery (MIS),” IEEE
found in several related studies [10]–[12].
Sensors J., vol. 20, no. 7, pp. 3487–3495, Apr. 2020. [Online]. Available:
Using the current available resources, we have tested this robot https://doi.org/10.1109/JSEN.2019.2945163
with a commercial nasopharynx phantom and three pig noses. The [11] N. Hendrich, F. Wasserfall and J. Zhang, “3D printed low-cost
successful working of the robot in terms of the required motions force-torque sensors,” IEEE Access, vol. 8, pp. 140569–140585, 2020.
was verified and the amounts of forces during the procedure have [Online]. Available: https://doi.org/10.1109/ACCESS.2020.3007565
[12] S. B. Kesner and R. D. Howe, “Design principles for rapid pro-
been preliminarily quantified, in conjunction with the study using totyping forces sensors using 3-D printing,” IEEE/ASME Trans.
a manually controlled tool with a commercial force sensor to mea- Mechatronics, vol. 16, no. 5, pp. 866–870, Oct. 2011. [Online].
sure the maximum forces and decide the maximum current setting Available: https://doi.org/10.1109/TMECH.2011.2160353

Authorized licensed use limited to: cao anh. Downloaded on December 14,2021 at 09:53:57 UTC from IEEE Xplore. Restrictions apply.

You might also like