Professional Documents
Culture Documents
Authorized licensed use limited to: cao anh. Downloaded on December 14,2021 at 09:53:57 UTC from IEEE Xplore. Restrictions apply.
290 IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, VOL. 3, NO. 1, FEBRUARY 2021
Fig. 4. Implementation of the proposed robot with its passive arm, active
end-effector, add-on headrest, and the embedded electronics shown. The photo
was taken only for illustrational purpose to show the expected clinical setup
of the NP swab collection procedure.
used in this study. The selected sensor has been proved in our previous
Fig. 2. Schematic illustration of the proposed 2-DOF active end-effector studies to be able to achieve large output voltage variations without
with each component and the mechanism of the force sensing swab gripper using an amplifier and demonstrates low level of noises [9], [10]. The
shown.
overall size of the swab gripper is 38(L) × 33(W) × 5(D) mm.
Authorized licensed use limited to: cao anh. Downloaded on December 14,2021 at 09:53:57 UTC from IEEE Xplore. Restrictions apply.
IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, VOL. 3, NO. 1, FEBRUARY 2021 291
TABLE I
C OST P RICE S UMMARY OF THE P ROPOSED ROBOT
Fig. 7. FE modeling of the swab gripper when subjected to external
forces: (a) the mesh results with the force, contacts, and the boundary
conditions shown, (b) the displacement results under 2.50N, and (c) the
force-displacement relationship with the forces applied from 0 to 2.50N.
noses (test facilities shown in Figure 9). The maximum force found
during the simulation procedure was found to be at the magnitude of
2.50N. Based on the manual experimental results, the maximum pay-
load of the robot was set to 3.00N with some additional allowances.
The maximum current setting, in real use, can be further adjusted
in an easy way once the in-vivo experiments using the manually
controlled swab holder are conducted.
2) Software Settings: The software settings for safety include
a watchdog between the microcontroller and the Blynk software which
continually monitors the connection by checking periodically refreshed
signals and various software logics to ensure that motor actuation
works in conjunction with the force sensors. The WiFi connection
between the robot and the control interface is also monitored by the
Blynk software based on its default built-in function. Communication
failures would be highlighted in the UI to inform the operator.
3) Intrinsic Features: In addition to above mentioned features, the
weight of the 2-DOF end-effector with all the electronics included
is only 0.23kg and the weight of the 6-DOF passive arm is 0.31kg.
This lightweight design intrinsically reduces the risks when using the
robot because the patient can easily move the robot away by pushing
the passive arm.
Fig. 6. (a) Setup for the current-force measurement experiment and the test III. F INITE E LEMENT A NALYSIS OF THE S WAB G RIPPER
results; (b) manually controlled tool for the maximum force measurements. Finite Element (FE) models of the swab gripper and the swab were
developed in ANSYS mechanical APDL (ver. 18.2) using geometric
current limiting method: each full-bridge is controlled by a fixed input of IGES file that was created by SOLIDWORKS (ver. 2018).
off-time PWM current control circuit that limits the load current The model is to analyze the displacement performances of the deflec-
to a desired value. The maximum value of current limiting can be tion beam of the swab gripper when the swab is subjected to external
adjusted by setting the reference pin and the motor’s maximum torque forces. To justify the design, the expected ranges and magnitudes
is proportional to ampere-turns. This feature limits the maximum of the displacements should meet the selected linear range of the
allowable force when the swab is in contact with the patient. Test optoelectronic sensor. In this study, Young’s modulus of 1.70 GPa
results shown in Figure 6(a) were obtained by adjusting the current and Poisson’s ratio of 0.43 were used for the swab gripper based
limiting setting from 0.2A to the maximum allowable value, 2.2A, on the information provided by the 3D printing manufacturer. Half
of the stepper motor driver in the linear axis and measuring the cor- model was considered as the whole configuration is symmetric about
responding maximum forces generated by the linear stage before its central plane. Both parts were meshed using quadratic tetrahe-
the stepper motor slips using a commercial force sensor (JHBM- dral elements (ANSYS element type: SOLID186). Mix u-p element
H3, JNSensor, Anhui, China). A regular daily-used shorter swab formulation method was chosen to avoid volumetric locking phe-
with wooden shaft was employed to avoid the buckling effects. nomenon given that the selected material has a Poisson’s ratio of
The experimental setup is shown in Figure 6(a). To decide the 0.43 which is close to that of an incompressible material (ν = 0.50).
value of the maximum current setting, a separate swab holder was The swab gripper was meshed using a total number of 32859 elements
designed and 3D printed to integrate with the commercial force sen- yielding a total of 55894 nodes and the swab has 4615 elements and
sor (Figure 6(b)). This swab holder allows for the measurement of 10027 nodes (Figure 7(a)). Contacts between the swab and the con-
the maximum force generated during the manually controlled NP straint hole was modelled as frictionless using Augmented Lagrange
swab collection procedure and was tested with a phantom and pig algorithm and the contact between the swab and the deflection beam
Authorized licensed use limited to: cao anh. Downloaded on December 14,2021 at 09:53:57 UTC from IEEE Xplore. Restrictions apply.
292 IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, VOL. 3, NO. 1, FEBRUARY 2021
was considered as fully bonded. The boundary conditions were added Fig. 9. Experimental setups for (a) the phantom and (b) the pig nose tests.
to constrain the movement of the gripper in Y and Z-axes. 2.50N were
applied to the distal end of the swab in seven sub steps as this was
the maximum expected force when the swab is in the nasal cavity.
The swab has a thin long shaft and is subjected to buckling under
increasing load. However, as this only happens in a limited range
and the swab is constrained by the nasal cavity, the deformations
do not cause the collapse and will continue to transfer the load.
In this study, we focused on the load analysis. An example color
coded displacement profile in the vertical direction and the com-
parison between undeformed and deformed shapes under 2.50N are
shown in Figure 7(b), as the results of the FEA. It can be observed
that the maximum displacement occurs at the center point of the
deflection beam, resulting in an approximate 0.50 mm deformation.
The force-displacement curve is shown in Figure 7(c).
Authorized licensed use limited to: cao anh. Downloaded on December 14,2021 at 09:53:57 UTC from IEEE Xplore. Restrictions apply.
IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, VOL. 3, NO. 1, FEBRUARY 2021 293
TABLE II
I MPORTANT F EATURES OF THE P ROPOSED ROBOT of the robot. Based on the current evidence, it is concluded that the
measured forces were below the payload of the robot, within the
detectable range of the sensing gripper, and has the similar trends
and ranges when tested using different facilities.
The initial insertion angle of the robot can be set to be parallel to
the palate using the passive arm with supports from visual inspection
or IMU sensor, and this ensures the swab go into the lower nasal pas-
sages of the phantom and pig noses although both are static objects.
Considering the possible motions induced by real patients, additional
active DOFs, as replacements of the current passive joints R5 and R6
(pitch and yaw motions) can be added to the robot design to allow
further flexibility, although this will increase complexity and cost of
the robot. Therefore, the necessity for doing this would need to be
further investigated by our future in-vivo tests. The present study was
limited to experiments on simulation phantom and pig noses, although
both cannot fully represent the real nasal cavity of human. Our next
focus will be pursuing ethical approval for in-vivo tests, improving
the current force sensor design, modelling the contact between swab
tip and soft tissue, and evaluating the effectiveness and robustness of
for the phantom and the pig nose are also shown in Fig. 10(c) based the robot via large-scale tests.
on the recorded motor steps, which reflect the swab insertion depths
in the nasal cavities. The motor was actuated with constant speed R EFERENCES
during the procedure.
[1] “Laboratory testing for coronavirus disease 2019 (COVID-19) in
suspected human cases: Interim guidance,” document WHO/COVID-
C. Summary 19/laboratory/2020.4, World Health Org., Geneva, Switzerland,
Mar. 2020.
Based on the design introduced in Section II and the preliminary
[2] X. Wang et al., “Comparison of nasopharyngeal and oropharyngeal
test results analyzed in Sections IV-A and IV-B, the features of the swabs for SARS-CoV-2 detection in 353 patients received tests with
proposed robot were summarized in Table II. both specimens simultaneously,” Int. J. Infect. Dis., vol. 94, pp. 107–109,
May 2020. [Online]. Available: https://doi.org/10.1016/j.ijid.2020.04.023
[3] Y. Yang et al. (Feb. 2020). Evaluating the Accuracy of Different
V. D ISCUSSIONS AND C ONCLUSION Respiratory Specimens in the Laboratory Diagnosis and Monitoring
This article introduces a low-cost miniature robot to assist NP sam- the Viral Shedding of 2019-nCoV Infections. [Online]. Available:
https://doi.org/10.1101/2020.02.11.20021493
pling. The robot can be made easily using 3D printing technique
[4] K. Carver and N. Jones, Comparative Accuracy of Oropharyngeal
together with other easy-to-find commonly used components on the and Nasopharyngeal Swabs for Diagnosis of COVID-19, Centre for
market. The robot is intended to be used in a simple way by assem- Evidence-Based Medicine (CEBM), Nuffield Dept. Primary Care
bling different pieces together and controlling from a mobile app. Health Sci., Univ. Oxford, Oxford, U.K., Mar. 2020. Accessed:
The costs for developing the robot can be further reduced if manu- May 24, 2020. [Online]. Available: https://www.cebm.net/covid-19/
comparative-accuracy-of-oropharyngeal-and-nasopharyngeal-swabs-for-
factured in large numbers (estimated to be less than 30 USD). The diagnosis-of-covid-19/
payload of the robot, when controlling a swab with a long thin shaft, [5] Interim Guidelines for Collecting, Handling, and Testing Clinical
is strictly limited by using small motors. Above the threshold, the Specimens for COVID-19, Dept. Health Human Services, Centers Dis.
motor would slip. To guarantee the safe use, active current limiting, Control Prevent., Atlanta, GA, USA, May 2020.
watchdog with periodical response checking were implemented, in [6] G.-Z. Yang et al., “Combating COVID-19—The role of robotics
in managing public health and infectious diseases,” Sci. Robot.,
addition to the intrinsic features of the robot. vol. 5, no. 40, Mar. 2020, Art. no. eabb5589. [Online]. Available:
The robot also has a detachable swab gripper which incorporates https://doi.org/10.1126/scirobotics.abb5589
the force sensing capability based on a simple CM and a low- [7] Q. Wang, China Develops Intelligent Robots for Throat Swab
cost optoelectronic sensor. This piece can be easily assembled to Sampling of Coronavirus Tests, Global Times, Beijing, China,
Mar. 2020. Accessed: May 24, 2020. [Online]. Available:
and removed from the robot, making it possible for the suspected https://www.globaltimes.cn/content/1182175.shtml
patient to do self-assembly and disassembly pre and post being tested. [8] F. M. Marty, K. Chen, and K. A. Verrill, “How to Obtain a nasopharyn-
The main detect range of 0-2.50N of the sensing gripper was jus- geal swab specimen,” New England J. Med., vol. 382, p. e76, Apr. 2020.
tified using the FE analysis. The calibration result indicated that [Online]. Available: https://doi.org/10.1056/nejmvcm2010260
the displacement range (0–0.50mm) of the deflection beam meets [9] Y. Noh et al., “A 2-piece six-axis force/torque sensor capable of mea-
suring loads applied to tools of complex shapes,” in Proc. IEEE/RSJ
the prediction of the FE model and falls into the expected linear IROS, Macau, China, 2019, pp. 7976–7981. [Online]. Available:
range of the optoelectronic sensor. The sensor’s design concept was https://doi.org/10.1109/IROS40897.2019.8967947
specifically for this low-cost application. Similar concepts of the low- [10] Y. Noh, S. Han, P. Gawenda, W. Li, S. Sareh, and K. Rhode, “A con-
cost 3D printable force sensor using optoelectronic sensors were also tact force sensor based on s-shaped beams and optoelectronic sensors
for flexible manipulators for minimally invasive surgery (MIS),” IEEE
found in several related studies [10]–[12].
Sensors J., vol. 20, no. 7, pp. 3487–3495, Apr. 2020. [Online]. Available:
Using the current available resources, we have tested this robot https://doi.org/10.1109/JSEN.2019.2945163
with a commercial nasopharynx phantom and three pig noses. The [11] N. Hendrich, F. Wasserfall and J. Zhang, “3D printed low-cost
successful working of the robot in terms of the required motions force-torque sensors,” IEEE Access, vol. 8, pp. 140569–140585, 2020.
was verified and the amounts of forces during the procedure have [Online]. Available: https://doi.org/10.1109/ACCESS.2020.3007565
[12] S. B. Kesner and R. D. Howe, “Design principles for rapid pro-
been preliminarily quantified, in conjunction with the study using totyping forces sensors using 3-D printing,” IEEE/ASME Trans.
a manually controlled tool with a commercial force sensor to mea- Mechatronics, vol. 16, no. 5, pp. 866–870, Oct. 2011. [Online].
sure the maximum forces and decide the maximum current setting Available: https://doi.org/10.1109/TMECH.2011.2160353
Authorized licensed use limited to: cao anh. Downloaded on December 14,2021 at 09:53:57 UTC from IEEE Xplore. Restrictions apply.