EPSILON 4

TECHNICAL SPECIFICATIONS

Energy dispersive X-ray fluorescence spectrometer for the elemental analysis range from F to Am and the
concentration range from ppm to 100 %.
637

0
53

510
Sample handling Data processing
Types of samples Powders (loose and pressed), fused beads, Built-in 4096 multi-channel analyser (MCA)
metals, liquids, films, large samples
Sample size Solids 27 - 51.5 mm diameter
Height 10 cm maximum Computer
Typically 10 mL of liquid or powder in a sample cup 32- or 64-bit based notebook or desktop PC with DVD-ROM and a
LSM: Large (un-cupped) sample mode minimum of 1 USB port.
Maximum 10 x 20 x 10 (H x W x D) cm Minimum of 2 GB RAM and 10 GB available hard disk space
Sample weight Maximum 200 g
Sample changer 10 position, removable tray
Software
Spinner Included
• Epsilon software for qualitative and quantitative analysis. Optional
software modules include: Omnian (standardless), Oil-Trace, Stratos,
Basic Upgrades Special FingerPrint and Enhanced Data Security.
configuration
• Interactive, operation under Windows 7 and Windows 8.1 Mouse or
X-ray tube touch-screen operation
Tube settings Software Software Software • Remote support (online and phone)
controlled: max controlled: max controlled: max

of this information and we shall not be liable for errors contained herein or for damages in connection with the use of this material. Malvern Panalytical reserves the right to change the content in this material at any time without notice.
50 kV, max 2 mA, 50 kV, max 3 mA, 50 kV, max 3 mA,

Disclaimer: Although diligent care has been used to ensure that the information in this material is accurate, nothing herein can be construed to imply any representation or warranty as to the accuracy, correctness or completeness
max 10 W max 15 W max 15 W Installation
Types Metal-ceramic, side-window Dimensions 270 x 530 x 510 mm (H x W x D)
Anode material Rh and Ag, specials on request Weight 47 kg
Tube filters Six, software selectable (Cu 300 µm, Cu 500 µm, Operating ambient 5 °C – 35 °C
Al 50 µm, Al 200 µm, Ti 7 µm, Ag 100 µm) temperature
Window 50 µm (Be) Operating humidity 20 % - 80 % non-condensing
Detector
Types High-resolution High-resolution High-resolution 100
Si drift detector Si drift detector Si drift detector Operating conditions
SDD10 SDD30 SDDUltra 90
Relative humidity (%)

Windows 8 µm (Be) 8 µm (Be) Ultra-thin


80

Copyright: © 2018 Malvern Panalytical. This publication or any portion thereof may not be copied or transmitted without our express written permission. 9498 707 60211 PN11180
window
Elemental F - Am F - Am C - Am 70
range
Resolution < 145 eV @ 5.9 keV cps 60
Typically 135 eV @ 5.9 keV cps
50
Max count rate 1500 kcps @ 50% dead time SDD10 / SDD30
input 40
SDDUltra
Electrical
30
HT generator Max 10 W Max 15 W Max 15 W
21 23 25 27 29 31 33 35 37
Main supply 90 - 264 V
voltage Optical chamber temperature (°C)
Frequency 47 - 63 Hz
Power 250 VA 100
consumption
90
Relative humidity (%)

Helium purge Safety


80 standards
Improves intensities of low atomic number elements such as C, N, O, F, Protection IEC1010/EN61010
Na, Mg and Al 70 CSA c22.2 No. 1010
Inlet pressure 0.8 bar IEC 529 IP20
60
EMC Directive 2004/108/EC
Gas purity 99.996 %
50 Machine Directive 2006/42/EC
Gas consumption 0.5 L/min G.O.S.
Reduction valve for gas bottle and 8 meters of tube included 40
Radiation Vollschutz according to Röntgenverordnung1987 in
SDD5
the version published 30 April 2003
30
21 23 25 27 29 31 33 35 37
Quality Optical chamber temperature (°C)
Developed, manufactured and supported according to ISO9001

MALVERN PANALYTICAL
Grovewood Road, Malvern Lelyweg 1, 7602 EA Almelo,
Worcestershire, UK, WR14 1XZ the Netherlands
P.O. Box 13, 7600 AA Almelo,
Tel +44 1684 892456 The Netherlands
Fax +44 1684 892789
Tel +31 546 534 444
Fax +31 546 534 598

info@malvernpanalytical.com
www.malvernpanalytical.com

www.panalytical.com/epsilon4