Professional Documents
Culture Documents
Digital Intraoral Radiographic Quality Assurance
Digital Intraoral Radiographic Quality Assurance
At present, the American Dental Association and the American Academy machine was tested for equipment performance and accuracy, and the
of Oral Maxillofacial Radiology have guidelines for the dental environ- computer monitor calibration was evaluated and adjusted as needed.
ment that include quality assurance and control of film-based radiog- The results confirm the continued need for updated QA procedures
raphy. Approximately 19%-30% of US dental offices use some form of in the dental office that include digital X-ray imaging. By implementing
digital intraoral radiography, and growth is expected to continue. It is these changes and practices, dentists should be able to improve the
anticipated that new tools and guidelines will be needed to aid in the diagnostic quality of radiographs while reducing the radiation exposure
development of quality assurance (QA) and control of digital intraoral of the patient.
radiographic images. Working with a representative sample of private Received: May 2, 2012
practice dental offices, this study examined and evaluated the entire Revised: October 31, 2012
digital intraoral radiographic system used in each operatory. The X-ray Accepted: February 19, 2013
I
n their textbook, Oral Radiology: (DENT), which found that 40% of the For example, the X-ray machine should be
Principles and Interpretation, White X-ray units evaluated were operating in inspected at regular intervals by a quali-
& Pharoah state: excess of the recommended ranges.2,3 fied expert, per state regulations. The film
In December 1979, the Bureau of processor should be evaluated daily using
A quality assurance (QA) program in Radiological Health published a set of 1 of 3 methods: sensitometry and dosim-
radiology is a series of procedures designed QA guidelines based on the recommenda- etry, stepwedge, and reference film. Image
to ensure optimal and consistent operation tions of their Medical Radiation Advisory receptor devices (such as film and inten-
of each component in the imaging chain. Committee.4 These guidelines were gen- sifying screens) along with the cassettes
When all components are functioning eral in nature, and covered all diagnostic should be inspected every 1-6 months.
properly, the result is consistently high radiology facilities, including dental Darkroom integrity should be checked
quality radiographs made with low expo- offices. In 1981, Valachovic et al outlined monthly, along with a leaded apron and
sure to patients and office personnel.1 a program for QA tests and frequency collar inspection.8 This protocol was also
of performance, tailored specifically for recommended by White & Pharoah.1
Beginning in the 1960s, several the private dental office.5 In 1983, the Many states require by law that dentists
national studies were initiated that Quality Assurance Committee of the have a written QA protocol for dental
pointed out the need for quality assur- American Academy of Dental Radiology radiography. For instance, Texas law
ance in medical and dental radiology. (now American Academy of Oral and §289.232(i)(7) states:
The Division of Radiologic Health, Maxillofacial Radiology) published a set
within the United States Public Health of QA guidelines based on the Bureau For all dental radiation machines, the
Service, conducted a program called of Radiological Health guidelines, but registrant shall perform, or cause to be
SurPak, the goal of which was to identify designed specifically for dental use.6 performed, tests necessary to ensure proper
X-ray generators that were lacking suf- The authors of the National Council on functioning of equipment with the indi-
ficient filtration and collimation.2 In Radiation Protection and Measurements’ cated standard for each item specified in
the 1970s, the Nashville Dental Project, (NCRP) Report No. 145, Radiation paragraph (6) (H)-(M) of this subsection.
conducted by the Bureau of Radiological Protection in Dentistry, referred to the After installation, the test listed shall be
Health (formerly the Division of protocol proposed by Valachovic et al performed every four years.9
Radiologic Health), was directed toward when making their own QA recommenda-
the elimination of unnecessary exposure tions.5,7 In September 2006, the American Historically, published QA and quality
to the dental patient. Also in the 1970s, Dental Association’s (ADA) Council on control (QC) protocols were intended for
the Bureau of Radiological Health con- Scientific Affairs updated its recommenda- film-based radiography. In fact, NCRP
ducted the Nationwide Evaluation of tions (based on NCRP Report No. 145) Report No. 145 refers to digital radiogra-
X-Ray Trends (NEXT), which focused on for the use of dental radiographs.8 The phy by acknowledging that “the required
patients’ exposure to ionizing radiation Council recommended developing and standards, apparatus and software for
during standard examinations. Dental implementing QA protocols in each dental dental systems do not currently exist.”7
radiology was included in this evalua- healthcare setting for the X-ray machine, With the advent of digital intraoral radi-
tion.2 Another project was the Dental imaging receptor, film processing, dark ography, the details are different but the
Exposure Normalization Technique room, leaded aprons, and thyroid collar.8 need for QA remains. In addition to the
if 5% of the squares on either end of the 70 kVp and 7 mA. This unit was operat- with only 6 visible steps. An exposure
grayscale were visible within the larger ing at 8% over the set kVp, the CV of reduction of approximately 20% resulted
grayscale squares, with 0% representing repeated exposures was 0.4%, and the in a phantom with 7 visible steps, 7 line
black and 100% representing white. A half-value layer of Al measured 2.6 mm. pairs/mm, 4 changing diameter wells, and
linear test pattern (a known standard for The second operatory used a Marksman 1 varying depth well. The fifth dental office
comparison) also was incorporated to X-ray unit (SS White Burs, Inc.) operating had 2 Schick CDR sensors used in 4 opera-
evaluate the monitors’ spatial resolution at 70 kVp and 7 mA. The measured kVp tories. These operatories were equipped
and distortion.12 In addition to evaluat- was 3% lower than the nominal kVp for with Gendex 770 X-ray units (Gendex
ing monitor calibration, the environment this unit. The CV of repeated exposures Dental Systems) operating at 70 kVp and
in which the monitor was viewed was was less than 2.2% and the half-value 7 mA. The measured kVp of all 4 units was
assessed for diagnostic purposes. layer of Al measured 1.9 mm. The third within 8% of the nominal value and a CV
operatory had a Preva unit operating at 60 of repeated exposures was <1.1% for all
Results kVp and 7 mA, with a measured kVp 2% units. The recorded half-value layer of Al
Every monitor evaluated was located in lower than the nominal kVp. The CV of for all units was ≥2 mm. The ESE for the
the operatories tested and was operating repeated exposures was 1.5% and the half- bitewing setting ranged from 68.2 to 79.9
within SMPTE standards. The lighting value layer of Al was 2 mm. mR, resulting in phantom images with 5-6
in the operatories was not subdued, and The ESE used for bitewings in the visible steps, 7-8 line pairs/mm, 5 changing
thus not ideal for diagnostic and treatment first operatory was 35.9 mR. Increasing diameter wells, and 2 changing depth wells.
planning purposes. Diagnosis and treat- the exposure time by 25% increased the Only 1 of the Gendex 770 units could pro-
ment planning were conducted in the doc- number of visible contrast wells. Exposures duce a phantom image with 7 visible steps
tor’s private office for all practices visited. used in the second and third operatories due to the limited available exposure times.
Only 2 of the computer monitors could were 30%-50% higher than optimum ESE was reduced by 30%-33% when the
be accessed; both were operating within exposure for this sensor, and reduced expo- phantom image was optimized within the
SMPTE standards. sure times were recommended. capability of the X-ray machines.
The third office had 2 Schick CDR sen- The sixth office used 1 Schick CDR
Offices using Schick sensors sors in 4 operatories equipped with Sirona sensor in 3 dental operatories. Three
All 6 offices using Schick sensors used X-ray units (Sirona Dental Systems, Inc.) X-ray machines (Heliodent 70, Siemens
the same practice management software operating at 60 kVp and 7 mA. All X-ray Corporation) operated at 70 kVp and 7 mA.
(Eaglesoft, Patterson Dental Supply, Inc.). units operated within 2% of the nominal The measured kVp of all 3 units was within
Schick proprietary software was used kVp. The exposures had a CV of <0.8% 5% of the nominal value. For all units, the
for image capture and viewing. The first and half-value Al layers ≥2.2 mm. The CV of repeated exposures was <1.7%, the
Schick office had 2 Schick Elite sensors ESEs for the bitewing setting ranged from half-value layer of Al was ≥2.4 mm, and the
in 5 operatories equipped with Preva 64.3-84.7 mR; as a result, the step wedge ESE for the bitewing setting ranged from
X-ray units (Progeny Dental) operating at image had only 6 visible steps. Decreasing 50.8 to 95.7 mR. The timer on 1 X-ray
60 kVp and 7 milliamperes (mA). For all exposure time by 20% extended the unit was off by 20%, which could explain
units, the measured kVp was within 7% dynamic range to all 7 steps. Using this the large discrepancy. The phantom images
of the nominal kVp, and the coefficient of sensor under optimal conditions, 8-9 line produced at this setting resulted in only 5
variation (CV) of repeated exposures was pairs/mm, 4 constant diameter wells, and visible steps for 1 unit; 6 steps were visible
<1.2%. The half-value layer of aluminum 2 varying depth wells were visible. In gen- for the other 2 units. Altering the exposure
(Al) for all units was greater than the eral, the Schick CDR sensor demonstrated times reduced the ESE by 30%-48% and
1.5 mm required by Texas state regulations. better image quality and 30% lower ESE produced phantom images with 7 steps,
The ESE used for bitewing radiographs in than the newer Schick Elite sensor. 7 line pairs/mm, 5 changing diameter
the various operatories ranged from 89.8 to The fourth Schick office had 2 Schick wells, and 2 changing depth wells.
101.6 milliroentgen (mR), with an average CDR sensors in 3 operatories. Two
of 95.9 mR. These parameters resulted in operatories had Heliodent DS X-ray units Offices using Dexis sensors
images of the phantom with an average (Sirona Dental Systems, Inc.) operat- The 6 offices using Dexis sensors used
of 7 line pairs/mm, 5 varying-diameter ing at 60 kVp and 7 mA; the third had Dentrix practice management software
wells, and 1 varying-depth well. A series a Heliodent MD unit (Sirona Dental (Henry Schein Dental). All offices used
of exposures (with decreasing exposure Systems, Inc.) operating at 70 kVp and Dexis proprietary image capture and
times) showed no significant deterioration 7 mA. The Heliodent DS units operated viewing software.
in image quality to exposures at approxi- within 2% of the nominal kVp, a CV for The first Dexis office had 1 Dexis
mately 60 mR. Adjustments to this level repeated exposures of <1.5%, and half- Platinum sensor and 1 operatory equipped
resulted in a dose reduction of 35%. value Al layers ≥2.4 mm. The kVp of the with a Bel Ray II X-ray unit (Belmont
The second Schick office had 1 Schick Heliodent MD unit was 5% lower than the Equipment) operating at 60 kVp and 7
Elite sensor used in 3 operatories. The nominal kVp; in addition, the unit’s CV of mA. The measured kVp was 2% lower
first operatory utilized an Image X 70 Plus repeated exposures was <0.1%. The bite- than the nominal kVp, while the CV of
X-ray machine (ImageWorks) operating at wing settings resulted in phantom images repeated exposures was <0.4%, and the
Discussion
Although they generally were not located
in an ideal environment for diagnosis and Fig. 2. Radiograph of a hemimandibular phantom Fig. 3. Radiograph of a hemimandibular phantom
treatment planning purposes, the opera- taken at 367.7 mR exposure. taken at 96.7 mR exposure.
tory viewing monitors were for the most
part consistent in their calibration quality,
with only 2 needing adjustment. It would
be convenient to have the medical view-
ing monitor SMPTE test pattern loaded NCRP-recommended diagnostic refer- patient exposure. A phantom allows den-
onto every computer in the office so the ence levels. One X-ray generator that had tists to accurately evaluate dynamic range,
calibration of a viewing monitor could be passed state certification 1 year earlier contrast/detail resolution, and spatial
checked periodically.16 had an exposure of 367 mR for bitewing resolution for all sensor/X-ray generator
All X-ray units tested had a half-value radiographs—twice the maximum recom- combinations and to select exposure set-
layer that complied with Texas regulations. mended exposure of 183 mR.10 Reducing tings based on these parameters.
A total of 5 X-ray units had a kVp >10%, the exposure time decreased patient expo- Table 1 shows the values measured in a
which was not acceptable per the manufac- sure radically without affecting the image representative dental operatory. The gray-
turers’ stated values and Texas state regula- quality (Fig. 2 and 3). This is an example scale density values for the various steps
tions. Four X-ray units were found to be of the type of situation referred to by of the wedge remain remarkably consis-
marginally within compliance boundaries. Lipoti that led to suggested changes in the tent for exposures ranging from 102.0
While the majority of X-ray units New Jersey radiation control program.13 to 249.3 mR. These values demonstrate
were compliant with state regulations, The DDQA phantom provides a means that a step wedge alone has no value for
in many cases, the settings used for for selecting exposure settings that provide determining optimal exposure settings in
routine radiographs were not within maximum image quality with minimal digital imaging. The purpose of the step
Chart. Distribution of skin exposures using different sensors before and after quality control (QC).
0.06
Schick CDR
before QC after QC
0.05 Schick Elite
before QC after QC
Dexis Platinum
0.04 before QC after QC
Dexis 601P
before QC after QC
Frequency
0.03
0.02
0.01
0
0 50 100 150 200 250 300 350
Exposure (mR)
wedge is to find a setting that allows for 102.0 mR, producing an image with used to compare changes to the QC for
visualization of the entire range of patient 10 line pairs/mm, 6 changing diameter exposures and deviation from the mean
(tissue) thickness encountered in routine wells, and 4 changing depth wells. A exposure for each of the 4 types of sen-
practice. When exposures exceed this 60% reduction in ESE resulted, with no sors. There were significant reductions in
level, 1 or more steps of the wedge cannot change in the radiographic image’s diag- exposure for all 4 types of sensors (aver-
be visualized. Having ensured that the nostic quality. age reduction = 34.6 mR, P < 0.05) after
entire dynamic range is registered, spatial The distribution of exposures cor- QCs were instituted. The Dexis 601P
and contrast resolutions are evaluated to responding to the original settings and Dexis Platinum had significantly
determine a satisfactory exposure time. used in the 12 offices were calculated higher reductions (P < 0.01) than the
In Table 1, the original exposure setting and compared to the distributions of Schick CDR and Schick Elite. There
in the dental office was 249.3 mR, with exposures at the optimized settings. was a significant reduction in terms of
a line pair resolution of 10 line pairs/ The results, segregated by sensor model, deviance from the average exposure for
mm and a contrast resolution showing are shown in the Chart. In each case, each instrument; averaging a reduction
6 changing diameter wells and 4 chang- an overall reduction in the mean ESE in spread of 53.2 mR. The Schick CDR
ing depth wells. An optimized setting was achieved, as was a narrowing in demonstrated the greatest reduction in
was selected with a new ESE setting of exposure range. One-way ANOVA was spread, followed by the Dexis 601, Dexis
20. Hellen-Halme K, Nilsson M, Petersson A. Digital radi- Patterson Dental Supply, Inc., Wood Dale, IL
ography in general dental practice: a field study. Den- 630.616.8202, www.pattersondental.com
tomaxillofac Radiol. 2007;36(5):249-255. Phillips Scientific, Mahwah, NJ
21. National Council on Radiation Protection and Mea- 201.934.8015, www.phillipsscientific.com
surements. NCRP Report No. 73. Radiological Protec-
Planmeca USA, Roselle, IL
tion and Safety in Medicine. Bethesda, MD: 1996.
630.529.2300, www.planmecausa.com
Progeny Dental, Lincolnshire, IL
Manufacturers 877.544.9672, www.progenydental.com
Belmont Equipment, Somerset, NJ
800.223.1192, dental.takarabelmont.com RaySafe, Hopkinton, MA
866.486.3677, www.raysafe.com
Dental Imaging Consultants, LLC, San Antonio, TX
210.347.7091, www.dentalimagingconsultants.com Sanko Co., Ltd., Kagawa, Japan
81.87.821.0035, www.sanko.kk.net
Eastman Kodak Company, Rochester, NY
800.698.3324, www.kodak.com Siemens Corporation, Washington, DC
800.743.6367, www.usa.siemens.com
Gendex Dental Systems, Hatfield, PA
800.323.8029, www.gendex.com Sirona Dental Systems, Inc., Long Island City, NY
718.937.5765, www.sirona.com
Henry Schein Dental, Melville, NY
800.372.4346, www.henryschein.com SS White Burs, Inc., Lakewood, NJ
732.905.1100, www.sswhiteburs.com
ImageWorks, Elmsford, NY
800.592.6666, www.imageworkscorporation.com THE YOSHIDA DENTAL MFG. COMPANY, LTD,
Tokyo, Japan
Microsoft, Redmond, WA
81.3.3631.2165, www.yoshida-net.co.jp
425.882.8080, www.microsoft.com